EE 4BD4 Lecture 20

Therapeutic Stimulation

Page 2: EE 4BD4 Lecture 20 Therapeutic Stimulation 1. 2 Weak and Strong Stimulation Recording Amplifier AC Coupled (bad choice) 3

2

Page 3: EE 4BD4 Lecture 20 Therapeutic Stimulation 1. 2 Weak and Strong Stimulation Recording Amplifier AC Coupled (bad choice) 3

3

Weak and Strong StimulationRecording Amplifier AC Coupled (bad choice)

Page 4: EE 4BD4 Lecture 20 Therapeutic Stimulation 1. 2 Weak and Strong Stimulation Recording Amplifier AC Coupled (bad choice) 3

4

Extracellular Stimulation(at cathode)

Page 5: EE 4BD4 Lecture 20 Therapeutic Stimulation 1. 2 Weak and Strong Stimulation Recording Amplifier AC Coupled (bad choice) 3

5

Page 6: EE 4BD4 Lecture 20 Therapeutic Stimulation 1. 2 Weak and Strong Stimulation Recording Amplifier AC Coupled (bad choice) 3

6

Extracellular Stimulation

• Stimulation threshold – minimum strength of stimulus required to excite the membrane

• Defined by current pulse amplitude and duration

• For unmyelinated nerve fibers and muscle fibers can be anywhere along fiber

• For myelinated depolarization only at the Node of Ranvier

Page 7: EE 4BD4 Lecture 20 Therapeutic Stimulation 1. 2 Weak and Strong Stimulation Recording Amplifier AC Coupled (bad choice) 3

7

Page 8: EE 4BD4 Lecture 20 Therapeutic Stimulation 1. 2 Weak and Strong Stimulation Recording Amplifier AC Coupled (bad choice) 3

8

Extracellular Stimulation Clinical Applications• Cardiac Pacing• Cardiac defibrillation• Functional electrical stimulation (next slide)• Diagnostic stimulation for nerve conduction

velocities• Innervated or denervated muscle maintenance• Therapeutic brain stimulation (ECT, vagal or direct

brain stimulation for Parkinsonism or epilepsy, rTMS – repetitive transcranial magnetic stimulation, tDCS – transcranial direct current stimulation, etc.)

• Artificial cochlea and vision prosthesis

Page 9: EE 4BD4 Lecture 20 Therapeutic Stimulation 1. 2 Weak and Strong Stimulation Recording Amplifier AC Coupled (bad choice) 3

9

Functional Electrical Stimulation

Page 10: EE 4BD4 Lecture 20 Therapeutic Stimulation 1. 2 Weak and Strong Stimulation Recording Amplifier AC Coupled (bad choice) 3

10

Stimulator Design• Constant voltage (current determined by electrode and

tissue impedance)• Constant current (more modern)• Typical pulse durations are 50 µs to 200 µs (some

applications use 1 ms and longer• Cardiac or smooth muscle require these longer durations

because membrane gates take longer to depolarize• Implantable constant current stimulators can deliver ma’s

with voltages < 10V but should have bipolar shapes to minimize charge buildup and tissue damage.

• External stimulators can be monopolar but require higher current amplitude (>10 ma) and voltage output

Page 11: EE 4BD4 Lecture 20 Therapeutic Stimulation 1. 2 Weak and Strong Stimulation Recording Amplifier AC Coupled (bad choice) 3

11

Page 12: EE 4BD4 Lecture 20 Therapeutic Stimulation 1. 2 Weak and Strong Stimulation Recording Amplifier AC Coupled (bad choice) 3

12

Page 13: EE 4BD4 Lecture 20 Therapeutic Stimulation 1. 2 Weak and Strong Stimulation Recording Amplifier AC Coupled (bad choice) 3

13

Rat Muscle Stimulator

MicrocontrollerPC USB Optocoupler

Optocoupler

H-bridge

Constant current source

+15v

To CCS

CNTRL 1 High

CNTRL 2 Low

CNTRL 1 Low

CNTRL 2 High

DAC

From optocoupler Rsense

From H-Bridge

VinCurrent through

rat = Vin______

Rsense

5V Power Isolator

To constant current source and H-bridge

components

Inside the H-bridge

Inside the constant current

source (CCS)

+

-

5V power isolator

From H-bridge