duti 254 d

2
250 SERIES IN PROCESS TO EXCELLENCE D U T I ensity llage emperature nterface detector - - - PORTABLE GAUGING INSTRUMENT DUTI -254 Digital Gauging Device DUTI-254 Ullage level Interface level Density & Temperature 1 Density & Temperature 3 Density & Temperature 2 Density & Temperature N process

Upload: neilc5703

Post on 12-Jun-2015

204 views

Category:

Documents


0 download

DESCRIPTION

New Product show case at the ILTA show in Houston

TRANSCRIPT

Page 1: Duti 254 D

250 SERIES

IN PROCESS TO EXCELLENCE

D U T Iensity llage emperature nterface detector - - -

PORTABLE GAUGING INSTRUMENT

DUTI -254

Digital Gauging Device

DUTI-254

Ullage level

Interface level

Density &Temperature 1

Density &Temperature 3

Density &Temperature 2

Density &Temperature N

process

Page 2: Duti 254 D

Specifications250 SERIES www.lemis-process.com

Advantages

Applications

Chemical industry

Pharmaceutical industry

Cosmetic industries

Petroleum industry

Food & Beverages

For more information please visit www.lemis-process.com

Ethanol production

Direct density and level measurement

Record spot density, average per tank

ATEX Hazloc certifcation

Safe operation, low maintenance

At any depths up to 30 meters

They’re economical and easy to operate

Measures highly viscous liquids up to 2000 cP

Rigid construction for heavy duty

Local results storage and Bluetooth

outdoor operation

data transfer

Able to download the measurements to PC; Multifunctional software allows to proceed the

measurements results in user-convenient form;

Compatible for a Windows 98/ME/2000/NT/XP.

Immediately printout the measurements; No need for PC.

Remote data transfer; Useful in field conditions; Software for data processing.

SOFTWARE BLUETOOTH PRINTER POCKET PC

Automatic temperature/viscosity

No sampling required

compensation

and depth of the sensor embedding

Measuring range:0… 3 g/cm³ (0… 3000 kg/m³)Density

Density Standard 0.6… 1.2 g/cm³ (600… 1200 kg/m³)

±0.0003 or g/cm³ ( 0.3 or kg/m³)±0.0005 ± ±0.5Accuracy:

Density

±0.00015 or g/cm³ ( or kg/m³)±0.00025 ±0.15 ±0.25Repeatability:

Density

Resolution:0.0001 g/cm³ (0.1 kg/m³)Density

Temperature 0.01 0.02°F°C ( )

Ambient temperature

Depth of submersion Up to 30 meters (up to 100 ft)

Intrinsically safe: Controller Level block with sensor

ATEX II (2G) EEx ib [ia] IIB T4ATEX II 1G EEx ia IIB T4

Power supply 2 x NiMH 3.6V-1200 mAh

Operating time without charging Appr. 12 hours

Dimensions, weight: ControllerLevel block with sensor

180 x 80 x 40 mm (7.1 x 3.2 x 1.6 in), 0.6 kg (1.3 lb) 420 x 245 x 140 mm, x 5.5 in(16.5 x 9.7 ), 4 kg (8.8 lb)

Temperature compensation Automatic

Data handling

Backlighted LCD displayLocal memory for 998 results with date/time stampedBuild in bluetooth for data transfer to printer or PCOptional Windows - based software

Delivery Delivered in compact carrying case

Supported measuring unitsReferred density: at 15° , 20°C, 60°F; API60; SG60CTables ASTM D1250

Real density: g/cm³, kg/m³, lb/gal, lb/ft³; API; SG

Tape graduation - Metric /English Temperature in °C or °F

-40… +85°C (-40... +185 )°F

Viscosity compensation Automatic

Temperature -40… +85°C (-40... +185 )°FLevel Up to 30 m (up to 100 ft)

±0.1 ±0. °F°C ( 2 )Temperature± m ±2 m ( 0.008”)Ullage/Interface detector

±0.1 ±0.2°F°C ( )Temperature±1 mm ± ( 1/16”)Level

Level 1 mm (1/16’)