dsc: differential scanning calorimetry a bulk analytical technique
TRANSCRIPT
What Does a DSC Measure?
A DSC measures the difference in heat flow rate (mW = mJ/sec) between a sample and inert reference as a function of time and temperature
-0.4
-0.3
-0.2
-0.1
0.0
0.1
He
at
Flo
w (
W/g
)
0 25 50 75 100 125 150
Temperature (°C)Exo Up
Endothermic Heat Flow
• Heat FlowEndothermic: heat flows into the sample as
a result of either heat capacity (heating) or some endothermic process (glass transition, melting, evaporation, etc.)
-0.1
0.0
0.1
Heat F
low
(W
/g)
0 20 40 60 80 100 120 140 160
Temperature (°C)Exo Up
Exothermic Heat Flow
• Heat FlowExothermic: heat flows out of the sample
as a result of either heat capacity (cooling) or some exothermic process (crystallization, cure, oxidation, etc.)
Temperature
• What temperature is being measured and displayed by the DSC?Sensor Temp: used by most DSCs. It is
measured at the sample platform with a thermocouple (transducer), thermopile (series of thermocouples) or PRT (Platinum Resistance Thermometers)
Constantan Body
Chromel Wire
Chromel Area Detector
Constantan Wire
Chromel Wire
Base Surface
Thin Wall Tube
Sample Platform
Reference Platform
Temperature
• What temperature is being measured and displayed by the DSC?Pan Temp: calculated by TA Q1000 based
on pan material and shape Uses weight of pan, resistance of pan, &
thermoconductivity of purge gasWhat about sample temperature?
The actual temperature of the sample is never measured by DSC
Temperature
• What other temperatures are not typically being displayed?Program Temp: the set-point temperature
is usually not recorded. It is used to control furnace temperature
Furnace Temp: usually not recorded. It creates the temperature environment of the sample and reference
Understanding DSC Signals
Heat Flow• Relative Heat Flow: measured by many
DSCs. The absolute value of the signal is not relevant, only absolute changes are used.
• Absolute Heat Flow: used by TA’s Q1000. Dividing the signal by the measured heating rate converts the heat flow signal into a heat capacity signal
DSC Heat Flow
t)(T,dt
dT Cp
dt
dHf
signal flowheat DSC dt
dH
Weight SampleHeat x Specific Sample
CapacityHeat Sample Cp
Rate Heating dt
dT
(kinetic) re temperatuabsolutean at
timeoffunction is that flowHeat t)(T, f
Tzero Heat Flow Equation
Rs Rr
qs qr
Cs Cr
Tr
T0
Ts
Heat FlowSensor Model
d
TdC
d
dTCC
RRT
R
Tq r
ssr
rsr
110
How do we calculate these?
Besides the three temperatures (Ts, Tr, T0); what other values do we need to calculate Heat
Flow?
Measuring the C’s & R’s
• Tzero™ Calibration calculates the C’s & R’s• Calibration is a misnomer, THIS IS NOT A
CALIBRATION, but rather a measurement of the Capacitance (C) and Resistance (R) of each DSC cell
• After determination of these values, they can be used in the Four Term Heat Flow Equation showed previously
Measuring the C’s & R’s
• Preformed using Tzero™ Calibration Wizard
1. Run Empty Cell
2. Run Sapphire on both Sample & Reference side
Measuring the C’s & R’s
Empty DSC constant heating rate
Assume: 0 rs qq
Heat balance equations give sensor time constants
ddT
TRC
ssss
0
dTd
ddT
TTRC
srrr
0
Measuring the C’s & R’s
Repeat first experiment with sapphire disks on sample and reference (no pans)
Assume:d
dTcmq s
sapphss d
dTcmq r
sapphrr
Use time constants to calculate heat capacities
10
ss
sapphss
ddT
Tcm
C
10
rs
sapphrr
dTd
ddT
TTcm
C
Measuring the C’s & R’s
Use time constants and heat capacities to calculate thermal resistances
s
ss C
R
r
rr C
R
A few words about the Cs and Rs
• The curves should be smooth and continuous, without evidence of noise or artifacts
• Capacitance values should increase with temperature (with a decreasing slope)
• Resistance values should decrease with temperature (also with a decreasing slope)
• It is not unusual for there to be a difference between the two sides, although often they are very close to identical
Before Running Tzero™ Calibration
• System should be dry• Dry the cell and the cooler heat exchanger
using the cell/cooler conditioning template and the default conditions (2 hrs at 75°C) with the cooler offPreferably enable the secondary purgeDo not exceed 75°C cell temperature with
the cooler off, although the time can be extended indefinitely
Stabilization before Calibration
• System must be stable before Tzero™ Calibration
• Stabilization is achieved by cycling the baseline over the same temperature range and using the same heating rate as will be used for the subsequent calibration
• Typical systems will stabilize after 3-4 cycles, 8 cycles recommended to ensure that the system has stabilized
Example of Typical Results
40
50
60R
efer
ence
Res
ista
nce
(°C
/Wat
t)
0.02
0.03
0.04
Ref
eren
ce C
apac
itanc
e (J
oule
/°C
)
0.01
0.02
0.03
0.04
0.05
Sam
ple
Cap
acita
nce
(Jou
le/°
C)
30
40
50
60
70
Sam
ple
Res
ista
nce
(°C
/Wat
t)
-200 -100 0 100 200 300
Temperature (°C)
Characteristics of the thermal resistances and heat capacities:Both curves should be smooth, with no steps, spikes or inflection points.Thermal resistances should always have negative slope that gradually decreases.Heat capacities should always have positive slope that gradually decreases.
This cell is very well balanced. It is acceptable and usual to have larger differences between sample and reference.
-0.4
-0.2
0.0
0.2
0.4
0.6
He
at F
low
(m
W)
-100 0 100 200 300 400Temperature (°C)
Conventional BaselineT zero Baseline
Tzero™ vs Conventional Baseline
Keeping the DSC Cell Clean
• One of the first steps to ensuring good data is to keep the DSC cell clean
• How do DSC cells get dirty?Decomposing samples during DSC runsSamples spilling out of the panTransfer from bottom of pan to sensor
How do we keep DSC cells clean?
• DO NOT DECOMPOSE SAMPLES IN THE DSC CELL!!!
• Run TGA to determine the decomposition temperature Stay below that temperature!
• Make sure bottom of pans stay clean• Use lids• Use hermetic pans if necessary
Cleaning Cell
• If the cell gets dirtyClean w/ brush
Brush gently both sensors and cell if necessary
Be careful with the Tzero™ thermocouple
Blow out any remaining particles
Sample Shape
• Cut sample to make thin, don’t crush• If pellet, cut cross section
• If powder, spread evenly over the bottom of the pan
Using Sample Press
• When using crimped pans, don’t over crimp• Bottom of pan should remain flat after crimping
• When using Hermetic pans, a little more pressure is needed
• Hermetic pans are sealed by forming a cold wield on the Aluminum pans
Crimped Pans Hermetic Pans
Good Bad SealedNot Sealed
Sample Size
• Larger samples will increase sensitivity
but…………….• Larger samples will decrease resolution
• Goal is to have heat flow of 0.1-10mW going through a transition
Sample Size
• Sample size depends on what you are measuringIf running an extremely reactive sample (like
an explosive) run very small samples (<1mg)Pure organic materials, pharmaceuticals
(1-5mg)Polymers - ~10mgComposites – 15-20mg
Effect of Sample Size on Indium Melt
Size: 0.4900 mg
Size: 1.2100 mg
Size: 5.7010 mg
-25
-20
-15
-10
-5
0
He
at
Flo
w (
mW
)
150 152 154 156 158 160 162 164
Temperature (°C)
Weight Onset Peak Width(mg) (°C) (°C) (°C)0.49 156.41 156.56 0.171.21 156.45 156.76 0.295.70 156.61 157.17 0.55
Purge Gas• Purge gas should always be used during DSC
experimentsProvides dry, inert atmosphereEnsures even heatingHelps sweep away any off gases that might
be released• Nitrogen
Most commonIncreases Sensitivity Typical flow rate of 50ml/min
Purge Gas• Helium
Must be used with LNCSHigh Thermo-conductivityIncreases ResolutionUpper temp limited to 350°CTypical flow rate of 25ml/min
• Air or OxygenUsed to view oxidative effectsTypical flow rate of 50ml/min
Sample Temperature Range
• Rule of ThumbHave 2-3 minutes of baseline before and
after transitions of interest - if possible DO NOT DECOMPOSE SAMPLES
IN DSC CELLTemperature range can affect choice of pansJust because the instrument has a
temperature range of –90°C to 550°C (with RCS) doesn’t mean you need to heat every sample to 550°!
Start-up Hook
Do not attempt to interpret transitionsbefore Heating rate has stabilized
9.56mg PET @ 10°C/min
0
2
4
6
8
10
12
De
riv.
Te
mp
era
ture
(°C
/min
)
-0.25
-0.15
-0.05
He
at
Flo
w (
W/g
)
-5 5 15 25 35
Temperature (°C)Exo Up
Heating Rate
• Faster heating rates increase sensitivity
but…………….• Faster heating rates decrease resolution
• Good starting point is 10°C/min
Thermal History
• The thermal history of a sample can and will affect the results
• The cooling rate that the sample undergoes can affect :Crystallinity of semi-crystalline materialsEnthalpic recovery at the glass transition
• Run Heat-Cool Heat experiments to see effect of & eliminate thermal historyHeat at 10°C/minCool at 10°C/minHeat at 10°C/min