DIFFERENTIAL SCANNING CALORIMETER Chip-DSC 100 Chip-DSC 10 Chip-DSC 1 THERMAL ANALYSIS

Upload: others

Post on 10-Mar-2020

27 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: DIFFERENTIAL Chip-DSC 100 Chip-DSC 10 SCANNING Chip-DSC 1 … · 2019-11-13 · 5 The DSC Principle The Differential Scanning Calorimetry (DSC) is the most popular thermal analysis

DIFFERENTIAL SCANNING

CALORIMETER

Chip-DSC 100Chip-DSC 10Chip-DSC 1

T H E R M A L A N A L Y S I S

Page 2: DIFFERENTIAL Chip-DSC 100 Chip-DSC 10 SCANNING Chip-DSC 1 … · 2019-11-13 · 5 The DSC Principle The Differential Scanning Calorimetry (DSC) is the most popular thermal analysis

2

Since 1957 LINSEIS Corporation has been deliv-

ering outstanding service, know-how and lead-

ing innovative products in the field of thermal

analysis and thermophysical properties.

Customer satisfaction, innovation, flexibility

and high quality are what LINSEIS represents.

Thanks to these fundamentals our company

enjoys an exceptional reputation among the

leading scientific and industrial organizations.

LINSEIS has been offering highly innovative

benchmark products for many years.

The LINSEIS business unit of thermal analysis

is involved in the complete range of thermo-

analytical equipment for R&D as well as qual-

ity control. We support applications in sectors

such as polymers, chemical industry, inorganic

building materials and environmental analytics.

In addition, thermophysical properties of solids,

liquids and melts can be analyzed.

LINSEIS provides technological leadership. We

develop and manufacture thermoanalytic and

thermophysical testing equipment to the high-

est standards and precision. Due to our innova-

tive drive and precision, we are a leading manu-

facturer of thermal analysis equipment.

The development of thermoanalytical testing

machines requires significant research and a

high degree of precision. LINSEIS Corp. invests

in this research to the benefit of our customers.

Claus Linseis Managing Director

Page 3: DIFFERENTIAL Chip-DSC 100 Chip-DSC 10 SCANNING Chip-DSC 1 … · 2019-11-13 · 5 The DSC Principle The Differential Scanning Calorimetry (DSC) is the most popular thermal analysis

3

InnovationWe want to deliver the latest and best techno-

logy for our customers. LINSEIS continues to

innovate and enhance our existing thermal ana-

lyzers. Our goal is constant development of new

technologies to enable continuous discovery in

science.

German engineeringThe strive for the best due diligence and ac-

countability is part of our DNA. Our history is af-

fected by German engineering and strict quality

control.

Page 4: DIFFERENTIAL Chip-DSC 100 Chip-DSC 10 SCANNING Chip-DSC 1 … · 2019-11-13 · 5 The DSC Principle The Differential Scanning Calorimetry (DSC) is the most popular thermal analysis

4

DIFFERENTIAL SCANNINGCALORIMETER

Chip-DSC 10Chip-DSC 1

Chip-DSC 100

Page 5: DIFFERENTIAL Chip-DSC 100 Chip-DSC 10 SCANNING Chip-DSC 1 … · 2019-11-13 · 5 The DSC Principle The Differential Scanning Calorimetry (DSC) is the most popular thermal analysis

5

The DSC PrincipleThe Differential Scanning Calorimetry (DSC) is the

most popular thermal analysis technique to measu-

re endothermic and exothermic transitions as a

function of temperature.

The instrument is used to characterize polymers,

pharmaceuticals, foods/biologicals, organic and

inorganic chemicals. Transitions measured include

Tg, melting, crystallization, curing, cure kinetics, oxi-

dation induction time and heat capacity.

Unsurpassed performance

• Revolutionary sensor design – combined

heat flux sensor and furnace in one chip.

• Unsurpassed sensitivity – for detection of

melts and weak transitions

• Benchmark resolution – precise separation of

close lying events

• Reliable automation – up to 96 position

autosampler

• Widest temperature range – from -180 °C to

600°C in one measurement

The LINSEIS Differential Scanning Calorimeters

(DSC) operate in agreement with national and in-

ternational standards such as: ASTM C 351, D 3417,

D 3418, D 3895, D 4565, E 793,

E 794, DIN 51004, 51007, 53765, 65467, DIN EN 728,

ISO 10837, 11357, 11409.

Page 6: DIFFERENTIAL Chip-DSC 100 Chip-DSC 10 SCANNING Chip-DSC 1 … · 2019-11-13 · 5 The DSC Principle The Differential Scanning Calorimetry (DSC) is the most popular thermal analysis

6

INTEGRATED CHIPTECHNOLOGY

Revolutionary Sensor Concept

The novel Chip-DSC integrates all essential parts

of DSC: furnace, sensor and electronics in a mi-

niaturized arrangement. The chip-arrangement

comprises the heater and temperature sensor

in a chemically inert ceramic arrangement with

metallic heater and temperature sensor.

This arrangement allows superior reproducibili-

ty and due to the low thermal mass outstanding

temperature control and heating rates of up to

1000 K/min. The integrated sensor is easily user

exchangeable and available at a low cost.

The integrated design of the chip-sensor deli-

vers superior raw data, which enables a direct

analysis without pre- or post-processing of heat

flow data.

Reinventing Differential Scanning Calorimetry (DSC) – Integrated chip technology for your application needs

Chip-DSC 10

New Chip-Sensor

Page 7: DIFFERENTIAL Chip-DSC 100 Chip-DSC 10 SCANNING Chip-DSC 1 … · 2019-11-13 · 5 The DSC Principle The Differential Scanning Calorimetry (DSC) is the most popular thermal analysis

7

reference

sample

furnace

temperature sensors

heater

Usual DSCNew chip

technology

Chip reduction: similar to the memory cards

Small Footprint

The compact construction leads to a signifi-

cant reduction in production cost which can

be passed on to our customers. The low energy

consumption and unrivaled dynamic response

results in unsurpassed performance of this revo-

lutionary DSC-concept.

new technology allows for DSC miniaturization

Page 8: DIFFERENTIAL Chip-DSC 100 Chip-DSC 10 SCANNING Chip-DSC 1 … · 2019-11-13 · 5 The DSC Principle The Differential Scanning Calorimetry (DSC) is the most popular thermal analysis

8

COOLING OPTIONSPeltier-cooling system (0 – 600°C)

This is a Peltier cooled heat exchanger. The

simple to install accessory reduces the star-

ting temperature of the DSC-sensor by 20°C in

regards to room temperature. Due to the low

thermal mass of the sensor the DSC can now

reach linear heating from 10°C onwards. With

this starting temperature 90% of polymer appli-

cations can be covered.

Closed-loop intracooler (-100 – 600°C)

Closed cycle refrigerated cooling system. Can

cool down to -100°C. This intracooler eliminates

the need of refilling LN2 for cooling purposes.

LN2-cooling system (-150 – 600°C)

Controlled cooling system for ultra-low tempe-

rature application down to -150°C. This acces-

sory provides the openest flexibility and coo-

ling capacity of all available options.

Quench-cooling-system, (-180 – 600°C)

The quench-cooling accessory provides an

open cooling container surrounding sensor

and sample. Coolant dependent, e.g. dry ice

or LN2, sample temperature can go down to

-180°C. This system does not allow defined gas

atmospheres while measuring, as outgasing

will surround the sample.

Quench-Cooling-System

Page 9: DIFFERENTIAL Chip-DSC 100 Chip-DSC 10 SCANNING Chip-DSC 1 … · 2019-11-13 · 5 The DSC Principle The Differential Scanning Calorimetry (DSC) is the most popular thermal analysis

9

ACCESSORIESDSC-sample-press

For optimum sample preparation of aluminum

crucibles a ergonomic sample press is available.

Crucibles

Various crucibles made of aluminum, alumina,

copper, gold, platinum and sapphire are availa-

ble for measurements with the Chip-DSC. Other

crucibles are available on request.

Sensor change

The new user exchangable sensor can be re-

placed within a few seconds. This innovative

concept reduces the maintenance costs drasti-

cally.

Sample robot

The Chip-DSC 100 can be equipped with a

rugged and reliable sample robot. Up to 96

samples increase the efficiency significantly.

Page 10: DIFFERENTIAL Chip-DSC 100 Chip-DSC 10 SCANNING Chip-DSC 1 … · 2019-11-13 · 5 The DSC Principle The Differential Scanning Calorimetry (DSC) is the most popular thermal analysis

10

HP DSC

The 50/150 bar high pressure cell enables OIT

stability tests to monitor aging of polymers, oils

and fats.

Processes under high pressures can be simu-

lated and optimized, e.g. sorption, chemical

reactions etc.

Optical DSC

The Chip-DSC 100 can be equipped with a

CCD camera to observe the sample during

the measurement. The visualization of the

sample gives a much deeper insight to phase

transitions and decomposition processes.

Photo DSC

The Photo cell allows measurements under UV

light to investigate UV curing systems. Due to

the very short time constant of the chip sensor,

also fast UV curing reactions in the subsecond

time scale can be measured.

HARDWARE OPTIONS

Page 11: DIFFERENTIAL Chip-DSC 100 Chip-DSC 10 SCANNING Chip-DSC 1 … · 2019-11-13 · 5 The DSC Principle The Differential Scanning Calorimetry (DSC) is the most popular thermal analysis

11

RAMAN DSC

The coupling of the Chip-DSC with a Raman

spectrometer can be realized very cost effec-

tively. In the Raman spectra the amorphous and

crystalline phases can be detected very precise-

ly and in situ.

Sample Robot DSC

The sample robot for up to 96 samples increases

the productivity significantly. The instrument

can run automatically overnight or at the week-

end. Together with the intuitive and intelligent

software it reduces labour costs and saves time.

Low temperature DSC

The cooling can be realized with a variety of coo-

ling options: intracooler, liquid nitrogen cooling

or a Peltier cooling system. The available tempe-

rature range at subambient conditions can be

extended down to -180°C.

Page 12: DIFFERENTIAL Chip-DSC 100 Chip-DSC 10 SCANNING Chip-DSC 1 … · 2019-11-13 · 5 The DSC Principle The Differential Scanning Calorimetry (DSC) is the most popular thermal analysis

12

SOFTWARE

The software greatly enhances your workflow as

the intuitive data handling only requires mini-

mum parameter input.

AutoEval offers a valuable guidance for the user

when evaluating standard processes such as

glass transitions or melting points. The optional

thermal library product identification tool, pro-

vides a database with 600 polymers permitting

an automatic identification tool for your tested

polymer.

Instrument control and/or surveillance through

mobile devices gives you control wherever you

are.

• Software packages are compatible with latest

Windows operating system

• Set up menu entries

• All specific measuring parameters (User, Lab,

Sample, Company, etc.)

• Optional password and user levels

• Undo and redo function for all steps

• infinite heating, cooling or dwell time seg-

ments

• multiple language versions such as English,

German, French, Spanish, Chinese, Japanese,

Russian, etc. (user selectable)

• Evaluation software features a number of

functions enabling a complete evaluation of

all types of data

• Multiple smoothing models

• Complete evaluation history (all steps can be

undone)

• Data acquisition and evaluation can be per-

formed simultaneously

• Data can be corrected using zero and calibra-

tion correction

• Data evaluation includes: peak separation soft-

ware signal correction and smoothing, first

and second derivative, curve arithmetic, data

peak evaluation, glass point evaluation, slope

correction, zoom / individual segment display,

multiple curve overlay, annotation and draw-

ing tools, copy to clipboard function, multiple

export features for graphic and data export,

reference based correction

Smart Software Solutions from LINSEIS

Page 13: DIFFERENTIAL Chip-DSC 100 Chip-DSC 10 SCANNING Chip-DSC 1 … · 2019-11-13 · 5 The DSC Principle The Differential Scanning Calorimetry (DSC) is the most popular thermal analysis

13

Multi-Instrument All LINSEIS instruments DSC, DIL, STA, HFM, LFA,

etc. can be controlled from one software tem-

plate.

Report GeneratorConvenient template selection to generate cu-

stomized measurement reports.

Multi-Lingual Our software is available in many different user

exchangeable languages, such as: English,

Spanish, French, German, Chinese, Korean,

Japanese, etc.

Multi-User The administrator can generate different user

levels providing different rights to operate the

instrument. An optional Log file is available, too.

Thermal Library

The LINSEIS Thermal Library software package

comes as an option for the well-known, user

friendly LINSEIS evaluation software that is

integrated in almost all our instruments. The

Measured and matching reference curves (green) and reference curves with poor fit (red).

Thermal Library allows the comparison of the

complete curves with a data base providing

hundreds of references and standard materials

within only seconds.

Data BaseState of the art data base design enables easy

data handling.

Kinetic softwareKinetic analysis of DSC, DTA, TGA, EGA (TG-MS,

TG-FTIR) data for the study of the thermal beha-

vior of raw materials and products.

Material Identification

DSC curve recognition and

interpretation system

Page 14: DIFFERENTIAL Chip-DSC 100 Chip-DSC 10 SCANNING Chip-DSC 1 … · 2019-11-13 · 5 The DSC Principle The Differential Scanning Calorimetry (DSC) is the most popular thermal analysis

14

Chip-DSC 1 Chip-DSC 10 Chip-DSC 100

Temperature range RT up to 450 °C (no cooling option available)

RT up to 600°C-180 up to 600°C (with optional LN2 quench cooling)

–150°C up to +600°C(depending on optional coo-ling systems: Peltier, closed-loop intracooler, LN2)

Heating and cooling rates

0.001 up to 100 K/min 0.001 up to 300 K/min 0.001 up to 1000 K/minballistic cooling with up to 500 K/min

Temperature accuracy +/- 0.2K +/- 0.2K +/- 0.2K

Temperature precision +/- 0.02K +/- 0.02K +/- 0.02K

Cool down time400°C to 50 °C

2.7 min 2.7 min 2.7 min

Digital resolution 16.8 million points 16.8 million points 16.8 million points

Resolution 0.03 µW 0.03 µW 0.03 µW

Atmospheres inert, oxidizing (static, dyna-mic)

inert, oxidizing (static, dyna-mic)

inert, oxidizing (static, dyna-mic), vacuum

Measuring range +/-2.5 up to +/-250mW +/-2.5 up to +/-250mW +/-2.5 up to +/-250mW

Calibration materials included included included

Calibration recommended 6-month inter-val

recommended 6-month inter-val

recommended 6-month inter-val

SPECIFICATIONS

Up to 80% less power consumption than standard DSC-instruments!

Page 15: DIFFERENTIAL Chip-DSC 100 Chip-DSC 10 SCANNING Chip-DSC 1 … · 2019-11-13 · 5 The DSC Principle The Differential Scanning Calorimetry (DSC) is the most popular thermal analysis

15

10

0

-10

-20

-30

-40

-50

-60

-70

-80

-90

DTA

[µV

]

100 105 110 115 120 125 130 135 140 145Temperature [°C]

TAWN Test – Resolution

APPLICATION TAWN TEST

Scope

4,4´-Azoxyanisole reference material was used

to perform the so called TAWN test. The sub-

stance forms a liquid crystalline phase at 120°C

which transforms into a further liquid phase at

134°C with a small activation energy barrier in a

second step.

This “double peak” is used in the TAWN test to

investigate sensitivity and resolution of a DSC

system and compare different DSC instruments

with each other.

For the test an open aluminum crucible has

to be used. The atmosphere must be air, ar-

gon or nitrogen, in this case we used air. The

resolution is investigated with 5 mg of test

substance and a heating rate of 20 K/min. The

distance between the baseline to the mini-

mum of the curve between the two peaks is

measured. The ratio of this height to the height

of the second peak is the resolution. The result

of 0.13 is an excellent one.

LINSEIS Chip-DSC TAWN I Resolution Test 20K/min

Page 16: DIFFERENTIAL Chip-DSC 100 Chip-DSC 10 SCANNING Chip-DSC 1 … · 2019-11-13 · 5 The DSC Principle The Differential Scanning Calorimetry (DSC) is the most popular thermal analysis

16

15

10

5

0

-5

-10

-15

-20

-25

-30

-35

-40

-45

Hea

t Fl

ow

[mW

]

25 50 75 100 125 150 175 200 225 250 275 300Temperature [°C]

79.8°C148.0°C

230.6°C

Glass transition

Crystallisation

Melting

Measurement of PET granulate

APPLICATIONS

The analysis of polymers is one of the main ap-

plications of DSC. Effects like glass transitions,

melting and crystallization points are of interest

and often very hard to detect. The new LINSEIS

Chip-DSC provides high resolution and sensiti-

vity, making it an ideal instrument for this kind

of analysis. As an example, a PET granulate was

heated, quench cooled to freeze the amorphous

state and afterwards analyzed by Chip-DSC with

a linear heating rate of 50 K/min. The curve

shows a significant glass transition around 80°C,

followed by a cold crystallization of the amor-

phous parts starting around 148°C and a mel-

ting peak at 230°C.

Page 17: DIFFERENTIAL Chip-DSC 100 Chip-DSC 10 SCANNING Chip-DSC 1 … · 2019-11-13 · 5 The DSC Principle The Differential Scanning Calorimetry (DSC) is the most popular thermal analysis

17

0

50

100

150

200

250

300

350

400

450

0 50 100 150 200 250 300Temperature [°C]

Hea

t fl

ow

[mW

]

50°C/min

100°C/min

300°C/min

600°C/min

800°C/min

PET sample: High heating ratesThe Chip-DSC has a very low thermal mass and therefore can achieve very fast heating rates. The experiment shows a PET measurement with that was performed with heating rates of 50, 100, 300, 600 and 800K/min after an initial very fast quench cooling of the sample. The cold crystallization peak can only be observed using rela-tively slow heating rates. At very high heating rates, the crystallization is too slow and the corresponding peak disappears. As the sample remains in an amorphous state and as only crystalline parts can melt, no melting peak is observed anymore.

-10

-20

-30

-40

-50

-60

-70

-80

-90

-100

50 100 150 200 250 300Temperature [°C]

Hea

t fl

ow

[mW

]

16.47mg in aluminium panall curves: 50 K/min heating rateblue curve after balistic coolinggreen curve after 80 K/min cooling rateyellow curve after 60 K/minred curve after 40 K/min

PET CoolingDepending on the cooling rate, the grade of crystallinity of the polymer changes significantly. During a fol-lowing heating run, cold crystalliza-tion can only be observed if crystal-lization was not completed during a fast cooling run. If the previous coo-ling was slow, no glass transition or crystallization can be observed. Wit-hout the need of any active cooling such experiments with high cooling rates can be performed.

Page 18: DIFFERENTIAL Chip-DSC 100 Chip-DSC 10 SCANNING Chip-DSC 1 … · 2019-11-13 · 5 The DSC Principle The Differential Scanning Calorimetry (DSC) is the most popular thermal analysis

18

400

350

300

250

200

150

100

50

Tem

per

atu

re [°

C]

Co

olin

g s

pee

d [K

/min

]

1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5Time [min]

1.0 min | 400°C 100

50

0

-50

-100

-150

-200

-250

-300

-350

-400

-450

-500Max. cooling rate: -503K/min

2.24 min | 100°CCooling rate: -87 K/min

5.0 min | 29.4°CRT: 24.1°C

High cooling rates without active coolingThe LINSEIS Chip-DSC allows fastest possible ballistic cooling rates wit-hout any active coo- ler needed. Due to the low thermal mass and innova-tive sensor design, cooling rates up to 500 K/min from 400°C on can be re-ached. Even cooling to 100°C can be done with cooling rates of up to 90 K/min. A cooling from 400°C down to 30°C can be done in 4 minutes just by ballistic colling without need of any additional cooling devices.Of course, the signal can still be eva-luated during that cooling segment and does not lose sensitivity or accu-racy.This means that a complete analysis including heating and cooling can be done in only 10 minutes without any need for optional cooling devices.

Temperature [°C]120 140 160 180 200 220 240 260 280 300 320

Enthalpy: -28.45 J/gHeatingrate: 5 K/min

Enthalpy: -28.42 J/gHeatingrate: 50 K/min

Enthalpy: -28.59 J/gHeatingrate: 100 K/min

Enthalpy: -28.48 J/gHeatingrate: 200 K/min

Enthalpy: -28.71 J/gHeatingrate: 300 K/min

Enthalpy: -28.55 J/gHeatingrate: 500 K/min

Indium sample: High heating ratesExtremely high heating rates up to 1000K/min can be achieved while the reproducibility of the melting enthal-py remains excellent. The example shows the melting point of Indium measured with diffe-rent heating rates (5 K/min; 50 K/min; 100K/min; 200K/min; 300 K/min and 500K/min).

Page 19: DIFFERENTIAL Chip-DSC 100 Chip-DSC 10 SCANNING Chip-DSC 1 … · 2019-11-13 · 5 The DSC Principle The Differential Scanning Calorimetry (DSC) is the most popular thermal analysis

19

3.0

2.5

2.0

1.5

1.0

0.5

0

HD

SC [µ

V]

Cp

[J/g

∙K]

0 5 10 15 20 25 30 35Time [min]

0.01

0

-0.01

-0.02

-0.03

-0.04

-0.05

-0.06

Sapphire lit (Cp)

Sapphire measured (Cp)

Sapphire raw data)Customer sample raw data

Correction raw data

Customer sample Cp (Cp)

noise at start because of temperature control

sample evaporates and Cp gets lower

Cp determinationBeside the caloric measurements of phase transitions and reactions, the Chip-DSC is also able to determine the specific heat capacity. The Chip-DSC can realize this with only one crucible by using a definite tempera-ture- modulated heating rate. There-fore a calibration measurement with a reference material (like sapphire) is recorded, followed by the measure-ment of the unknown sample that can be evaluated using this calibra-tion.The measurement shows the modu-lated measurement of heat capacity of sapphire at a heating rate of 10 K/min with amplitude of 3 K. Cp was de-termined with an error of 2%, repre-senting an outstanding performance.

200

180

160

140

120

100

80

60

40

20

Time [min]

1400

1200

1000

800

600

400

200

0

-200

Diffe

rent

ial T

empe

ratu

re [m

K]

Tem

pera

ture

[°C]

0 10 20 30 40

9.2min117.4°C

17.3min198.2°C

Temperature transitiondynamic/static

after 3mintemperature equilibriumchange from Ar to O2

21.14min 201.3°Cpoint of first deviationfrom stationary condition

OIT– Oxidation induction timeThe oxidation and aging behavior can be checked using the OIT test.A small piece of LDPE was heated with 10K/min at constant argon flow rate of 10 l/h. At 200°C, an isothermal segment was applied and later the gas flow was switched from argon to oxygen at 10 l/h.At the point where the heating ramp goes into the isothermal segment, there is a response « step » in the he-atflow signal. After the signal has become stable again, the atmosphere was changed to oxygen. Seven minutes after that, the heat flow shows a significant change, which symbolizes the onset of the oxidation.

Page 20: DIFFERENTIAL Chip-DSC 100 Chip-DSC 10 SCANNING Chip-DSC 1 … · 2019-11-13 · 5 The DSC Principle The Differential Scanning Calorimetry (DSC) is the most popular thermal analysis

20

Temperature [°C]

10

5

0

-5

-10

-15

-20

-25

-30

-35

Hea

t fl

ow

[mW

]

130 140 150 160 170 180

Thermochromism

In classical DSC instruments the sample cannot

be observed during the measurement. Obser-

vation can give further useful information (for-

ming of bubbles, fumes, change of color, etc.).

The graphic above shows an example of a ther-

mochromic material, showing an endothermic

Gradient of a sample during a measurement

phase transition between 150 and 160°C. The

phases have different colors and the color chan-

ge from red to yellow can be seen through the

transparent cover. A camera option for recor-

ding images is available.

Page 21: DIFFERENTIAL Chip-DSC 100 Chip-DSC 10 SCANNING Chip-DSC 1 … · 2019-11-13 · 5 The DSC Principle The Differential Scanning Calorimetry (DSC) is the most popular thermal analysis

21

Temperature [°C]

450

400

350

300

250

200

150

100

50

0

Hea

t fl

ow

[mW

]

50 100 150 200 250 300 350

Onset point: 279,2°C O�set point: 288.7°CPoint of reaction: 205.721 mW at 282.0°CPeak maximum: 446.625 mW at 287.3°CEnthalpy: 1076.69 J/g

Highly energetic materials

Highly energetic materials are used in airbags,

as propellants, blasting materials etc.

For any type of DSC instrument, there is a risk

of damaging the sensor and even the furnace.

In the Chip-DSC, the chip (integrating sensor

and furnace) can be easily replaced by the ope-

rator at low cost and in a short time. In case of

a damage to the instrument, the downtime of

the instrument is reduced drastically. Change

oft he sensor needs only some seconds and the

calibartion can be done in approx. 15 minutes.

The example shows the DSC diagram of 2,8 mg

of airbag igniter.

Page 22: DIFFERENTIAL Chip-DSC 100 Chip-DSC 10 SCANNING Chip-DSC 1 … · 2019-11-13 · 5 The DSC Principle The Differential Scanning Calorimetry (DSC) is the most popular thermal analysis

22

500

400

300

200

100

0

70

65

60

55

50

45

40

35

30

0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0Time [min]

Hea

t fl

ow

[mW

]

Tem

per

atu

re [°

C]

Onset point:Offset point:

Point of reaction:Peak maximum:

Enthalpy:

50.2°C (2min)64.0°C (2min)195.257 mW at 53.7°C (2min)500.680 mW at 63.7°C (2min)355.15 J/g

Onset point:Offset point:

Point of reaction:Peak maximum:

Enthalpy:

50.3°C (9min)51.8°C (9min)36.479 mW at 51.6°C (9min)85.987 mW at 53.9°C (9min)64.36 J/g

UV curing of polymers

The Linseis Chip DSC in combination with an

UV/LED source is a powerful tool for the cha-

racterization of fast photocurable resin systems

[L. Gonzales, University of Bayreuth].

In this specific case, a mixture of a photocura-

ble acrylate and thermally curable epoxy was

used and around 10 mg of material was placed

in an open crucible and irradiated with several

UV light pulses until no change in peak area was

detected.

The difference between the first irradiation

peak and the last one, (when the area under the

peak reaches a plateau, thus it is assumed that

no reaction is taking place), is calculated to de-

termine the heat of reaction for the UV-curing

part, and to derive a conversion curve.

Page 23: DIFFERENTIAL Chip-DSC 100 Chip-DSC 10 SCANNING Chip-DSC 1 … · 2019-11-13 · 5 The DSC Principle The Differential Scanning Calorimetry (DSC) is the most popular thermal analysis

23

0

-20

0

-10

0

-20

-40 -30 -20 -10 0 10 20 30 40 50 60 70 80

0

-10

0

-5

-10

Temperature [°C]

Hea

t fl

ow

[mW

]

black cream chocolate

white cream chocolate

solid black white mix chocolate

manufactured milk chocolate

industrial milk chocolate

15.6°C

18.0°C

47.8°C

35.5°C

8.8°C

17.7°C

47.2°C

38.9°C30.9°C

34.4°C42.0°C

52.1°C

28.5°C

32.6°C

28.9°C

43.9°C

33.5°C44.4°C

51.2°C

Thermal analysis of chocolate

A very interesting application for DSC is the

analysis of chocolate. It helps to investigate the

melting behavior, melting temperature and fat

content of different mixtures and therefore it is

crucial for the design of a tasty, creamy product.

The figure shows DSC profiles of five different

chocolate samples. Each sample was precooled

with liquid nitrogen to -40°C and then measured

using the same conditions and similar sample

mass.

The result shows that nearly all tested choco-

lates show a double peak that is caused by low

melting milk fats and higher melting vegetable

plant fats.

The quality of emulsifiers and melting tempe-

rature can be compared and used for QC and

product design. In this example, the handmade

chocolate shows a much more homogeneous

melting behavior than cheaper industrial cho-

colate.

Page 24: DIFFERENTIAL Chip-DSC 100 Chip-DSC 10 SCANNING Chip-DSC 1 … · 2019-11-13 · 5 The DSC Principle The Differential Scanning Calorimetry (DSC) is the most popular thermal analysis

www.linseis.com

Products: DIL, TG, STA, DSC, HDSC, DTA, TMA, MS/FTIR, In-Situ EGA, Laser Flash, Seebeck Effect, Thin Film Analyzer, Hall-Effect

Services: Service Lab, Calibration Service

10/19

LINSEIS GmbH Germany

Vielitzerstr. 43

95100 Selb

Tel.: (+49) 9287 880 0

E-mail: [email protected]

LINSEIS China

Kaige Scientific Park 2653 Hunan Road

201315 Shanghai

Tel.: (+86) 21 5055 0642

Tel.: (+86) 10 6223 7812

E-mail: [email protected]

LINSEIS Poland

ul. Dabrowskiego 1

05-800 Pruszków

Tel.: (+48) 692 773 795

E-mail: [email protected]

LINSEIS Inc. USA

109 North Gold Drive

Robbinsville, NJ 08691

Tel.: (+1) 609 223 2070

E-mail: [email protected]

LINSEIS France

1 Route de Trévoux

69250 Neuville/Saone

Tel.: (+33) 6.24.72.33.31

E-mail: [email protected]