dräger cc-vision - draeger · dräger cc-vision is available in two software versions: dräger...

40
Dräger CC-Vision Dräger CC-Vision E-Cal Dräger CC-Vision X-am 1/2/5000 Dräger Pac Vision Configuration and calibration of Dräger gas detection instruments Instructions for Use 1

Upload: others

Post on 23-May-2020

13 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

Dräger CC-Vision Dräger CC-Vision E-Cal Dräger CC-Vision X-am 1/2/5000 Dräger Pac Vision Configuration and calibration of Dräger gas detection instruments Instructions for Use

1

Page 2: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

2

Table of Contents Safety Instructions ..................................................................... 3

Compatibility .......................................................................... 3 Intended Use .......................................................................... 4

Use...................................................................................... 4 Benefits for the User............................................................ 4

Installing Dräger CC-Vision .................................................. 4 Notices on How to Use these Instructions.......................... 5

Conventions Used in this Manual........................................ 5 Starting Dräger CC-Vision .................................................... 6

Help System........................................................................ 6 Working with Dräger CC-Vision ........................................... 7

The Buttons of Dräger CC-Vision........................................ 7 Data Transmission between Measuring Instrument and Dräger CC-Vision ................................................................... 8

Reading Data from the Measuring Instrument .................... 8 Setting Up the Interface....................................................... 8 Sending Data to the Measuring Instrument......................... 9 Instrument Window.............................................................. 9 Operating Service Life....................................................... 10 Instrument Identification .................................................... 10 Battery ............................................................................... 10 Status ................................................................................ 10 Datalogger......................................................................... 10 Quick Menu ....................................................................... 10 Date & Time ...................................................................... 10 Calibrating Sensor(s)......................................................... 11 Comments ......................................................................... 11 Instrument ......................................................................... 11 Changing Gas & Sensor with Dräger X-am 5000/5600 ... 11 Gas Test (GT) ................................................................... 12 Languages......................................................................... 12 Sensor (Description of Sensor) ......................................... 13 Application Mode (X-am 7000 only) .................................. 13

Dräger X-zone 5000 ............................................................. 16 Alarms ............................................................................... 16 Life signal .......................................................................... 17 Status signal...................................................................... 17 Radio ................................................................................. 17

Calibrating with Dräger CC-Vision ..................................... 18

Calibrating Options............................................................ 18 Calibration Gas Settings.................................................... 19 Important Notes on Calibration.......................................... 19 Calibration Procedure........................................................ 20 Carry Out Gas Test ........................................................... 20

Protocols in Dräger CC-Vision ........................................... 21 Menu...................................................................................... 22

File Menu........................................................................... 22 Edit Menu .......................................................................... 23 Communication Menu........................................................ 23 Mobile Printer .................................................................... 24 Autocalibration Menu......................................................... 24 Tools Menu........................................................................ 25

Registration .......................................................................... 26 Registration Settings ......................................................... 26 User Data .......................................................................... 27 Inventory Data ................................................................... 27 Comments ......................................................................... 28 Instrument Check-Out ....................................................... 28 Instrument Check-In .......................................................... 29

Working with Dräger CC-Vision E-Cal ............................... 30 Checklists ............................................................................. 31 Menu with E-Cal Station ...................................................... 31 Autocalibration Menu .......................................................... 32 Commissioning of Dräger CC-Vision E-Cal....................... 33

System Configuration ........................................................ 33 Configuring Gases............................................................. 33 Go-Button Configuration.................................................... 34 Gas Test Configuration ..................................................... 35 Quicktest Configuration ..................................................... 35 Testing and Calibrating with the E-Cal Station.................. 36 Display of Calibration Results ........................................... 36

Protocols in Dräger CC-Vision E-Cal ................................ 37 Network Configuration ........................................................ 37 Database Function............................................................... 38 E-Cal Error List..................................................................... 39

Page 3: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

Safety Instructions Strictly follow the instructions for use! Any use of the PC-program CC-Vision requires complete understanding and strict observation of the user instructions below. The program is only to be used for the purposes explicitly specified in this manual. Safety symbols used in these Instructions for Use These Instructions for Use have a number of warnings for risks and hazards which might occur when using the instrument. These warnings contain "signal words" that will alert you to the degree of hazard you may encounter. These signal words and corresponding hazards are as follows:

CAUTION Indicates a potentially hazardous situation which, if not avoided, could result in physical injury to you, or damage to the product. It may also be used to alert against unsafe practices.

3

Notice

Additional information on how to use the instrument.

Compatibility Minimum System Requirements Operating system: Windows 2000, Windows XP, Vista Processor: Intel Pentium Working memory: 16 MB RAM Interfaces: a free COM port (CC-Vision) two free COM ports or a free USB port (CC-Vision E-Cal) a free USB port (Pac Vision) Compatible Instruments Multiwarn II as of software version 6.00 MiniWarn as of software version 2.06 Pac III E as of software version 3.30 Pac Ex 2 as of software version 1.2 X-am 3000 as of software version 1.2 X-am 7000 as of software version 1.03 Pac 1000 - 5000 as of software version 2.2 Pac 3500/5500 as of software version 3.0 Pac 7000 as of software version 2.2 X-am 1100 - 2000 as of software version 3.0 X - am 5000 as of software version 3.0 X - am 5600 as of software version 3.5 X-zone 5000 as of software version 1.08 Driver USB-DIRA USB-Milo (communication module Pac 1000 - 7000) USB E-Cal USB printer The corresponding drivers are available on the installation CD.

Page 4: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

4

Intended Use Use The CC-Vision program helps configure and calibrate the following portable gas detection instruments made by Dräger Safety AG & Co. KGaA. - Multiwarn II - MiniWarn - Pac III - Pac Ex 2 - X-am 3000 - X-am 7000 - Pac 1000 – 7000 - Pac 3500/5500 - X-am 1100 - 2000 - X-am 5000/5600 - X-zone 5000 In addition, the Dräger CC-Vision E-Cal software version can be combined with an E-Cal station for automatic calibration of the above-mentioned measuring devices. CE marking electromagnetic compatibility (Directive 89/336/EEC) Benefits for the User With Dräger CC-Vision, a comfortable and timesaving calibration, configuration, documentation and verification of the above mentioned gas detection instruments is possible.

Installing Dräger CC-Vision Proceed as follows: • Insert installation CD in CD-ROM drive • If the installation routine does not start automatically: launch setup.exe • Select language • Follow the installation instructions Result: 1. Dräger CC-Vision is automatically installed in the folder

C:\Draeger\CC-Vision. Any existing version in the folder will be overwritten.

2. Links to the program will be copied to the Desktop and the Start menu.

Dräger Safety AG & Co. KGaA offers a free-of-charge update on the following website: http://www.draeger.com/ST/internet/DE/de/Produkte/Detection/Tragbare/Calibration/software/pd_cal_software.jsp Updates will work with any installed full version as of version 5.5.0

Page 5: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

5

Notices on How to Use these Instructions The IFUs at hand include a description of functions for Dräger CC-Vision, Dräger CC-Vision E-Cal, Dräger CC-Vision X-am 1/2/5000 and Pac Vision. So please use these IFUs in combination with the respective sensor data sheets and test gas guidelines and be sure to read the operating manuals and the technical user manuals which were supplied with the gas detection instruments you intend to use. PDF files of instrument operating manuals and sensor data sheets are stored in the respective instrument windows and sensor windows. In addition, users of the E-Cal station need to read the chapter on E-Ca functions which is at the back of the manual.

Conventions Used in this Manual Identification marking of Work step sequence: • ...

• ... • ...

Listings for the (completeness) check: 1. ... 2. ... 3. ...

Handling options:

a) ... b) ... c) ...

Buttons and menu items:

Selection

Page 6: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

6

Starting Dräger CC-Vision Proceed as follows: • You can start Dräger CC-Vision in two ways: a) by double-clicking the CC-Vision icon b) by choosing the CC-Vision menu item in the Windows menu

under Start - Programs - Dräger • After installation, the password protection is automatically

deactivated. If the password protection is activated, you must enter the user name and password. The standard settings are preconfigured: User name: Admin Password: Admin

Help System In addition to the operating manual that comes as a PDF file on the CD-ROM, a comprehensive Help system is implemented in the Dräger CC-Vision software. You can call the Help function by clicking on the question mark in the menu. The operating manual content is also available in the Dräger CC-Vision Help.

Page 7: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

Working with Dräger CC-Vision Dräger CC-Vision can be used to configure, calibrate and test the portable Dräger gas detection instruments specified under 'intended use'. Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas

detection instruments to a COM port of the computer. - The version Dräger CC-Vision E-Cal is available for users

of the E-Cal station. In this chapter, buttons are shown as they are available in Dräger CC-Vision on the user interface. The corresponding buttons of Dräger CC-Vision E-Cal can be found in chapter "Working with Dräger CC-Vision E-Cal" The Buttons of Dräger CC-Vision First, all configuration data of a connected measuring instrument is transferred to the PC. For this, many buttons have been provided on the button bar. The button bar allows quick access to important program functions by means of convenient button icons.

0022

3615

.tif

receive MiniWarn data

0032

3615

.tif

receive Multiwarn II data

0042

3615

.tif

receive PAC III data

0052

3615

.tif

receive PacEx II data

0062

3615

.tif

receive X-am 3000 data

0072

3615

.tif

receive X-am 7000 data

0232

3615

.tif

receive Pac X000 data

0242

3615

.tif

receive X-am 1100 - 2000 data

0242

3615

.tif

receive X-am 5000/5600 data

0472

3615

.jpg

receive X-zone 5000 data

After data transmission, the unit setup appears in the instrument window. Now you can view and edit instrument parameters, or perform a calibration or exposure to gas test. In connection with the editing, the instrument data can be sent back to the measuring device. A button is provided even for this:

0092

3615

.tif

Send data to the instrument

Instrument or sensor data can also be saved as file. Consequently, it is possible to create configuration files, to save on the PC, and if needed, to transfer to measuring devices and sensors.

7

Page 8: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

Data Transmission between Measuring Instrument and Dräger CC-Vision Reading Data from the Measuring Instrument Proceed as follows: • Connect and switch on measuring instrument: Multiwarn II: connect data cable, power up

Multiwarn II. MiniWarn: set up in line of sight of Dräger

infrared adapter (DIRA or USB-DIRA), power up MiniWarn.

Pac III: connect to Pac III interface. Pac Ex 2: connect to Pac Ex 2 interface. X-am 3000: connect to X-am 3000 interface. X-am 7000: set up in line of sight of Dräger

infrared adapter (DIRA or USB-DIRA), power up X -am 7000.

Pac 1000 - 7000: connect to the corresponding communication module.

X-am 1100 - 2000: set up in line of sight of Dräger infrared adapter (DIRA or USB-DIRA).

X-am 5000/5600 set up in line of sight of Dräger infrared adapter (DIRA or USB-DIRA).

X-zone 5000 adapt USB-DIRA in recess. • launch Dräger CC-Vision

Setting Up the Interface • Select Extras - COM Ports menu and required interface • Read configuration (either from menu or with button): a) Select Receive data in the Communication menu b) Press corresponding button

0022

3615

.bm

p or

0032

3615

.bm

p or

0042

3615

.bm

p or

0052

3615

.bm

p or

0062

3615

.bm

p or

0312

3615

.bm

p or

0322

3615

.bm

p or

0242

3615

.tif

or

0242

3615

.tif

or

0472

3615

.jpg

Result: The device window appears with the measuring device data.

8

Page 9: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

Sending Data to the Measuring Instrument Requirements: The measuring instrument is connected and powered up, and the COM port connected to the instrument is selected. Data that is to be sent to the instrument is displayed in the open and active instrument window. Proceed as follows: • Send configuration to instrument

Either from menu or with button: a) Select Send data in the Communication menu b) Press button

0092

3615

.tif

Result: Parameters from Dräger CC-Vision are transferred to the measuring instrument, previous settings of the measuring instrument are overwritten.

Instrument Window The instrument window presents the data received from the measuring instrument or the measuring instrument data called from a file (file type *.gdt). The instrument window menu is visible in a tree structure on the left side of the instrument window. By selecting the required parameters, the corresponding data is displayed on the right side and can be changed here too if needed. Values shown in grey cannot be changed. The manuals of the respective instrument types are available as PDF files.

0102

3615

.tif

9

Page 10: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

10

The Datalogger settings (data log for measured values and events) are displayed and can be changed.

Operating Service Life The instrument service life can be set and activated using the Instrument menu. An error message will be issued when the instrument reaches the end of its operating service life. This error message will be issued if the instrument is in operation or not! The operating service life can at any time be extended or deactivated using Dräger CC-Vision. For instruments (such as X-am 1700, Pac 3x00/5x00) with an operating service life counter, the maximum service life value can be set to coincide with the remaining overall service life. Instrument Identification All instrument data that clearly identify the measuring instrument will be displayed under the instrument description (uppermost selection in instrument window menu). Moreover, you can also view information on the configuration. Various instrument passwords are stated and can be edited. The operating manual of the respective instrument type can be opened as PDF file. Battery The battery data is displayed according to corresponding battery type. The serial number and the part number of the battery can be entered (these numbers can be found on the battery unit). Status Faults and warnings, related to the measuring instrument and to individual sensors are displayed under Status. Notice on colour markings: Red markings in the instrument window menu refer to existing faults or warnings. The selection text appears in red; and in addition, all branches belonging to the relevant point are marked with a red checkmark. Datalogger

Quick Menu This item can be used to determine which functions and menu items are to be displayed in the uppermost menu level of the measuring instrument (to a large extent, this corresponds with the measuring instrument function "Menu without Password" which is used for quick menu configuration). When activated, each function of the left column which is checkmarked will be displayed in the uppermost menu level of the measuring instrument, making it accessible and executable without needing a password. In addition, it is possible to determine whether or not the branch menus Maintenance, Configuration and Service are to be displayed in the uppermost menu level. Date & Time Here you can call up and change date, date format, time and time format for the measuring instrument. The current data is read and displayed by pressing the Receive button on the instrument. On pressing the Send button, PC-date, PC-time and the specified formats are transferred to the gas detection instrument.

Notice The Pac 1000 - 7000 instruments can only send time and date, not be read.

Page 11: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

11

Calibrating Sensor(s) The calibration parameters of all active sensors are displayed: next calibration date, gas and gas unit. The instrument window menu can be used to select various calibration functions: zero point, span, Fresh Air Cal and one button calibration. For more details on calibration, see chapter "Calibrating with Dräger CC-Vision " Comments Here you can enter 10 lines of comments on the instrument with up to 80 characters respectively. The lettering of the comment lines can be determined in the Extra - Options menu. Instrument Display of instrument settings. - operating signal: acoustic signal ON/OFF - silent alarm ON/OFF - track alarm ON/OFF - OK mode ON/OFF - switch off mode - Max. operation time in days, non-alterable. - adjustable operation time in days, alterable.

Changing Gas & Sensor with Dräger X-am 5000/5600 With the X-am 5000, you can change or deactivate the sensors. To do so, use the device window to select the menu item "Change sensor(s)", then - step by step - follow the instructions of the assistant.

Notice Do not forget to calibrate the instrument! In addition, the sensors Kat Ex 125, OV, OV-A, PH3 and Chlorine can be set to measure other gases and/or to determine other calibration gases. To do so, use the device window to select the menu item "Change gas", then - step by step - follow the instructions of the assistant.

Notice Do not forget to calibrate the instrument!

Page 12: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

12

Gas Test (GT) The gas test is used to check the proper sensor function. A protocol of the results can be sent to the printer. Up to 400 test results can be stored using the instrument file which can be called up via Test Results submenu. A protocol of can be displayed and sent to the printer. The result of the last gas test saved in the instrument is displayed under Last State. Depending on the instrument, the following settings can be edited and sent. To do so, it might be necessary to call up the GT Configuration submenu. - gas test mode: Manual

In this mode, a concentration is entered which needs to be exceeded during the gas test. The chosen concentration must be above the set A1 alarm level (with O2 below A1).

- gas test mode: Auto in addition to the concentration, the permissible gas test tolerance rate is taken into account.

- gas test interval in days The instrument will display a warning if the interval is exceeded. If necessary, the error message can be triggered with a delay. In addition, you need to enter the number of days the error message is to be delayed.

If the instrument is used in combination with the Dräger Bumptest Station, a score of additional options which were not mentioned in the settings described above can be activated. - Auto. Bumptest Station Detection

This option allows the instrument to start the gas test automatically as soon as it is put inside the Bumptest Station.

- Activate span cal. after negative gas test Provided that this option is supported by the instrument, the calibration will be launched automatically after a negative gas test.

In addition, see: " Carry Out Gas Test".

Languages Here you can choose languages supported by the instrument.

Page 13: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

13

Sensor (Description of Sensor) Display of sensor-specific data. The sensor-specific data includes: sensor type, measuring device slot, last and next calibration date, as well as display range and unit. Sensors can be deactivated with the Deactivate button. A deactivated sensor is marked "inactive" in red. Sensor - Evaluation Display of the sensor measuring range as well as the alarm thresholds A1 and A2 required for alarm analysis. Selection of an Exposure Evaluation: TWA: This alarm indicates that the gas concentration exceeds the permissible exposure limit during operation time. STEL: This alarm indicates that the short-term exposure value (STEV) exceeds the maximum short-term exposure limit (STEL), or that the short-term exposure value (STEV) exceeds the maximum short-term exposure period (DUR). Sensor - Calibrate You can use this menu to calibrate zero point and span of individual sensors. (See chapter "Calibrating with CC-Vision".)

Notice Not possible with Pac Vision. Sensor - Comment In this window, you can enter a comment for each sensor (10 lines with maximum 80 characters each).

Notice Not possible with Pac Vision.

Sensor - Alarms Here, you can call up and edit all settings for the alarms of the sensor used. - Alarm threshold A1 and A2 - Latching / non-latching (Latching: The alarm status is

retained, even if the measured value is no longer near the alarm threshold.)

- Acknowledgeable / non-acknowledgeable (Acknowledgeable: The alarm can be acknowledged so that the acoustic alarm falls silent.)

- Ascending / descending (Ascending: The alarm is triggered when the alarm threshold is exceeded. / Descending: Alarm is triggered when the alarm threshold falls below value.)

Application Mode (X-am 7000 only) You can use the Dräger CC-Vision software (as of version 5.3.3) to customise the application mode. For example, you might want to use the application mode to be able to apply the Dräger X-am 7000 in applications with various instrument- and sensor-specific settings. The application mode always calls up all saved instrument settings for the respective application in one go. All together, you can use the Dräger CC-Vision software to save 5 pre-set application modes. You need the Dräger CC-Vision software to configure and activate the application mode.

Notice Configuration via instrument menu is no longer possible once the Dräger X-am 7000 is in application mode.

Page 14: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

Configuration The modes can be either activated or deactivated.

Notice You need Dräger CC-Vision software to activate / deactivate the application mode.

14

0422

3615

.tif

However, track mode, confidence beep and silent alarm can still be activated / deactivated in the respective application mode. The application names can be customised (max. 9 characters).

0432

3615

.tif

Page 15: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

You can configure the selected sensor in the selected application mode. - sensor activated/deactivated - sensor hidden and masked in the display – but not

deactivated - full figures– display shows figures without decimals - alarm level A1 - alarm level A1 acknowledgeable - alarm level A2 - alarm level A2 acknowledgeable - measurement unit

0442

3615

.tif

15

The complete dataset is sent to the instrument after editing the settings.

0422

3615

.tif

Notice

Before activating the application mode, you need to let users know the contents of each individual configuration!

Page 16: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

16

Dräger X-zone 5000 Dräger CC-Vision is used to configure the Dräger X-zone 5000. The current configuration can be read out and parameters for the following functions can be altered here: - Alarms, - Life signal, - Status signal, - Radio. Alarms Pattern Optional pattern, from which one can be selected, are held here. The corresponding pattern for the triggering device and the receiving devices are also indicated. If a sample is to be tested under Test Options, an alarm type must be selected and the Save and test button pressed. The test is ended by pressing the button again. After the Save and test button is pressed, the last sample selected is in each case the one used. Volume The alarms can be switched on and off and the volume changed and tested for operation here. The volume can be adjusted with the slide control. Pressing the Save and test button stores the new value and allows the test alarm signal to be heard. The test is ended by pressing the button again. After the Save and test button is pressed, the last volume level selected is in each case the one used.

Forwarding (triggering Dräger X-zone 5000)

Note A distinction is made between A1 and A2 alarm types and alarms due to equipment faults. A further distinction is made as to whether the alarm is acknowledged (confirmed) or not. Here it is possible to determine which alarms are to be forwarded from the triggering Dräger X-zone 5000 to the receiving units. Combination (receiving Dräger X-zone 5000)

Note A distinction is made between A1 and A2 alarm types and alarms due to equipment faults. A further distinction is made as to whether the alarm is acknowledged (confirmed) or not. Here it is possible to determine how the receiving Dräger X-zone 5000 should respond to forwarded alarms. The horn and the LEDs are used as the outputs for this. Alarm contact Here it is possible to determine how the alarm contact should respond to forwarded alarms (see also alarm types).

Page 17: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

17

Life signal Settings The LED and horn function can be switched on or off here. When the function is switched on, the time interval (in seconds) can be set. Pattern A predefined sample can be selected here. Additionally, the length of the life signal in milliseconds can be displayed or changed. The volume can also be adjusted. Pressing the Save and test button stores the new value and allows a test life signal to be seen and heard. The test is ended by pressing the button again. After the Save and test button is pressed, the last parameters selected are in each case the ones used. Volume The life signal can be switched on and off and the volume changed and tested for operation here. The volume can be adjusted with the slide control. Pressing the Save and test button stores the new value and allows the test alarm signal to be heard. The test is ended by pressing the button again. After the Save and test button is pressed, the last volume level selected is in each case the one used. Status signal Volume The volume of the status signal can be changed here. The value can only be changed by means of the Send button on the program button bar on the unit. The status signal is used during a pump calibration, among other things.

Radio The radio settings display whether the radio mode is activated or not, the transmit power, which network number is set and what frequency is being used to transmit. The radio mode, the network number and the transmit power can be changed and can be transmitted to the unit by means of the Send button. The standard setting for the network number is 0, so that each unit will receive all the alarms. If different networks are to be set up so that they are not alarmed with respect to one another, the units must be provided with their own network numbers in each case. An alarm will only be forwarded between units with the same network numbers. Up to 255 different networks can be set up.

Note More detailed information on the radio mode and network numbers can be found in the Dräger X-zone 5000 Instructions for Use, in the chapter "Establishing a wireless connection".

Page 18: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

18

Calibrating with Dräger CC-Vision The calibrate functions in Dräger CC-Vision can be accessed from the instrument window menu (also see chapter "The Instrument Window"). You can choose between two function groups: a) To calibrate several sensors which belong to one measuring

instrument, select: Calibrate sensor(s).

b) Should you prefer direct access to certain sensor, select: Sensor (Description of Sensor) - Calibrate.

b) Not possible with Pac Vision The various calibrating options can be selected under these function groups. Calibrating Options Depending on the functionality of the gas detection instrument, Dräger CC-Vision offers four calibrating options: Calibration Parameters (Dräger X-am 5000 only) - Warning following calibration interval timeout:

If this option is activated, a warning will be issued after a calibration interval has timed out at the instrument.

- Fault following calibration interval timeout: If this option is activated, a fault message will be issued with a delay after a calibration interval has timed out at the instrument. The sensor then switches to channel error and will no longer display a measurement value.

- delay to fault: Delay in days until the sensor switches to channel error.

- Activation of span calibration for CO compensation in the H2 sensor: This option is only available with the Dräger Sensor XXS CO H2-CP. In this case, the span calibration for the H2 compensation electrode is activated.

Calibration History Calibration results are saved in the calibration history. Calibration date, name of tester and the sensors are listed with their calibration results. The individual results of the calibration history can be accessed with the arrow keys. Zero Calibration For the zero calibration, sensors of a connected gas detection instrument are calibrated with so-called zero gas to the values that correspond to their exact display at 0% measured gas. Please observe: During zero calibration, the sensor that needs to be calibrated may be loaded only with zero gas and not with measurement gas, this also applies for the O2 sensor!

Notice Not possible with Pac Vision. Span Calibration For the span calibration, the sensors are calibrated to display the exact target value of the stipulated calibration gas. Before the span calibration of a sensor, a zero calibration must be carried out. To run a span calibration of the catalytic sensor (CAT Ex Sensor) with the thermal conductivity measuring method, you need CH4 measuring gas at a level that is not relevant to LEL. Fresh Air Cal Fresh Air Cal is used for ambient air calibration of the sensors. The ambient air must be free of measuring gases and any other disturbance variables. The span of the EC-O2 sensor is set to 20.9 Vol.%, the zero point of the CO2 sensor is set to 0.03 Vol.%, the zero point of all other sensors is set to 0. One Button Calibration The one button calibration is used to trigger the span calibration for all instrument sensors.

Notice Not possible with Pac Vision.

Page 19: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

Calibration Gas Settings Select one of the calibration options (zero calibration, span calibration, Fresh Air Cal, one button calibration) to display a window with the calibration gas settings of the sensors.

19

Besides selecting the sensors to calibrate and the span gas concentrations, it is possible to make the following settings: Mixed Gas Activate this option when using mixed gas. Dräger CC-Vision will now calibrate all relevant sensors at the same time. The concentration curve of the selected sensors is displayed simultaneously. Automatic Valve Activate this option when using an automatic valve. You will not be prompted to expose the sensor to gas, nor to stop exposing the sensor to gas.

Notice The automatic valve can only be used with a gas detection instrument which is equipped with a pump. Expert Mode You will not be prompted to expose the sensor to gas, nor to stop exposing the sensor to gas. The minimum calibration time setting is not taken into account. Tolerance The measurement value can deviate for the minimum calibration time only by this tolerance value, otherwise a calibration is not possible. Minimum Calibration Time The permissible period is between 3 and 420 seconds. At the earliest, a calibration is possible after the set period has elapsed.

Important Notes on Calibration

CAUTION Danger to health! Never inhale calibration or test gas! Use commercially available calibration gas.

CAUTION Danger to health! To minimise the danger to health, select adequate span gas concentration.

Notice Observe date and time setting in the measuring instrument: Calibration is only possible when date and time are set correctly in the instrument.

Notice Before any calibration, the relevant sensor must have gone through a sufficient warm-up! Sufficient sensor warm-up time has passed when the Notice icon in the measuring instrument display winks out.

Notice Observe sequence: Always calibrate zero point before span. Always calibrate EC sensors first, then CAT-Ex sensors.

Notice Use non-condensing calibration gases.

Page 20: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

Calibration Procedure

CAUTION Before you begin with the calibration, observe the safety advise and notes in "Important Notes on Calibration"!

20

First, one of the calibration options is selected from the instrument window menu, then the calibration gas settings are made. The calibration procedure is started by pressing the Start button in the instrument window. The measurement values for the time for all sensors selected before are now displayed in a diagram. The presentation of many selected sensors takes place one after the other except when using mixed gas. In addition to the signal course graph, for each sensor, there is a field that contains the set points and actual values as numerical value. Over it is the Confirmation control panel respectively. This field can be activated only when the minimum calibration time set before lapses (does not apply for expert mode). Activate the confirmation field only when the measured value is stable over a specific period. The OK button starts the calibration of the confirmed channels. The calibration of sensors is carried out and is displayed through corresponding messages. The message showing which sensors were calibrated successfully or where errors occurred is displayed.

Carry Out Gas Test The gas test is used to check the proper function of measuring instrument sensors. To this end, the sensors need to be loaded with test gas so that the gas concentration measured by the sensor can be compared with the test gas concentration. In Dräger CC-Vision, the function 'exposure to gas' can be accessed from the instrument window menu (also see chapter "The Instrument Window"). Select Gas Test. The data window for test gas settings which is now displayed has the same configuration and the same options as the window for the calibration gas settings (for more information, refer to chapter "Calibration Gas Settings"). The gas test for the selected sensors is triggered by actuating the Start button. Every gas test is logged. The individual tests can be accessed via Test Results: The display shows test date, name of examiner and battery condition, as well as the measured actual value and the set point.

Page 21: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

Protocols in Dräger CC-Vision Protocols with configuration data and procedure data can be created for documentation purposes.

0122

3615

.tif

Print ...

When the printer is connected, the Print button is active and protocols can be printed.

0132

3615

.tif

Print Previews

The print preview provides a page view on the screen. Under Tools - Options menu, you can define the protocol headings. Dräger CC-Vision protocols can be prepared from the following data: - instrument configuration - instrument status - sensor configuration - calibration result from the calibration history - exposure to gas test result from the exposure to gas history - search results

21

Page 22: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

Menu The Dräger CC-Vision menu is located in the program window header. Frequently used functions are also available as buttons in a button bar. Some of these buttons are adaptive, hence are active only when corresponding devices are connected. File Menu Open ...

0142

3615

.tif

The selection window for opening a file appears. Select the required file, and press the Open button. The file type is determined in the File type field. Here it is possible to choose between instrument files (*.gdt) and sensor files (*.sdt). Search ...

0152

3615

.tif

The selection window for direct search of instrument files (e.g. instrument type, measuring point, next calibration date). Close The active window is closed. If the last changes of the window are not yet saved, you are asked if you want to save the data or not before the window closes.

Save

0162

3615

.tif

The data in the active window is saved in a file. The filename suggested automatically is composed of the instrument type and serial number. This filename can be changed. The data saved in ASCII format can be imported to other programs. Save as ... Enables saving data under another filename. The filename suggested automatically is composed of the instrument type and serial number. This filename can be changed. Print Previews

0132

3615

.tif

A window appears with the protocol print preview. The function Print Preview requires a properly installed Windows printer driver. Print

0122

3615

.tif

The protocol is printed on the printer that was defined as default printer. Printer setup If many printers are connected, you can select a printer here. List of last opened files Shows the list with the 8 last opened files. At the top, is the last opened file.

22

Page 23: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

Exit

0172

3615

.tif

Exits the Dräger CC-Vision program. If files still containing changes that have not yet been saved are opened, you are asked if you want to save the data or not before the window closes. Edit Menu Copy

0182

3615

.tif

Complete instrument data records as well as individual sensor data records can be copied. Mark the corresponding selection in the instrument window menu and choose Copy from the Edit menu. Paste

0192

3615

.tif

Pastes the copied instrument and sensor data in the active window. The instrument type (only for instrument data), the part number of sensors and the gas name must be identical.

Communication Menu Receive data

0022

3615

.bm

p

0032

3615

.bm

p

0042

3615

.bm

p

0052

3615

.bm

p

0062

3615

.bm

p

0072

3615

.bm

p

0312

3615

.bm

p

0232

3615

.tif

0242

3615

.tif

0472

3615

.jpg

The configuration of the connected gas detection instrument is read and presented in an instrument window.

Send data

0092

3615

.tif

The data from the active instrument window is sent to the connected gas detection instrument.

23

Page 24: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

24

Mobile Printer The Dräger Mobile Printer can be used to save the gas test results coming from a Dräger Bump Test Station. The menu entries can be used to load, display and save this data. When communicating, make sure that the Dräger Mobile Printer has either been connected to the corresponding external mains power supply or has been switched to battery operation!

Notice If you want to connect the Dräger Mobile Printer with a network, you need to use a compact ConLine V24 device server. Device Window (Mobile Printer List) - Add: A new Mobile Printer can be added here, and the Edit

dialog (see below) is called up. - Edit: The existing configuration of a Mobile Printer can be

edited here, and the Edit dialog (see below) is called up. However, the printer must be selected from the mobile printer list in advance.

- Delete: A Mobile Printer can be deleted here. However, the printer must be selected from the mobile printer list in advance.

- Read: The available data can be read here. If the data is saved in a file, the file name and the corresponding directory will be displayed. But the printer will also display the fact that no data is available, and that hence no data will be saved. The CC-Vision folder is used as default directory.

- Test: Communications check with Printer without Data Transfer. We recommend this test every time you install a new printer.

- DB Save: Instead of saving a file, the data is sent to a database, provided that one is available.

- Save & Close: The current Dräger Mobile Printer List is saved and the dialog is closed.

Dialog: Edit Instrument Data (Edit Mobile Printer) - Serial and part number are used to identify the instrument.

This data will be automatically corrected as soon as a line of communication is open.

- Interface: There are two ways of communicating with the printer. Either via RS232 (via USB interface) or via Ethernet (compact ConLine V24 device server).

- Interface Settings: If you want to use an RS232 interface, you need to select the printer baud rate and its COM port (verification via Windows device manager with "Dräger Mobile Printer") . The default baud rate is 115200 baud. If you are using an Ethernet interface to go through a network, you will need to assign an IP address for the compact device server (for more information, refer to "Network Configuration") and the port address, which has a standard setting of 3000.

Show Data Select File A file which was read out in the "Device Window (Mobile Printer List)" can be selected here. Format: "serialnumber_time_date.TXT". Data, which was saved to the database cannot be loaded here. Dialog: Results of Bumptest Presentation of the gas test results. The data can be sent to the printer. Autocalibration Menu In the Dräger CC-Vision menu, the menu item Autocalibration is only displayed for selection after connecting the E-Cal Station and the PC. For more information, refer to chapter "The Menu with E-Cal Station".

Page 25: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

Tools Menu COM Ports The COM port to which the gas detection instrument is connected is selected under this menu item. Options Here you can determine the lettering for the comment fields, which can be accessed from Comment in the instrument window. Letterings of the comment lines are limited to 11 characters. Letterings in already saved files cannot be changed. In addition, headings for protocols can be determined. Headings for protocols are entered in the Print headings and Footer field. The text length for report headings is limited to 15 characters. If the "Overwrite mode" is activated, when saving the calibration and test history in the field, the instrument settings are always updated, otherwise settings during the first save are retained. Use Directory to choose a default directory for data storage. In addition, a graphic, e.g. your company logo, can be included into the protocol. User Administration You can use the User function to create a user rights administration with user-specific user rights. Certain functions can be assigned of denied certain users. Creating a new User Use the New button to create a new user, by entering the user name of the new user.

Notice This name will from then on be a valid user name and a password will be demanded when the new user fires up Dräger CC-Vision.

25

Administration of User Rights You can only administrate user rights after launching the CC-Vision software with administrator rights (default setting: user name and password both "Admin"). To select and/or delete

various user rights, you can use the arrow keys to navigate the "All Rights" list in the User menu.

selects individual rights

deletes individual items from the list of assigned rights

selects all rights

deletes all items from the list of assigned rights

The new scope of user rights will be activated the next time the Dräger CC-Vision software is launched. Activate Password Protection With this function, a password protection is set, so that you are asked to enter a user name and password each time you start the Dräger CC-Vision program. In the input window, enter the user name and password (default setting "Admin"). To rule out any write errors when defining the password, you must enter the password twice. To edit these changes, terminate and relaunch the Dräger CC-Vision software. When launched, the Dräger CC-Vision software will now prompt for password. User name and password will both correspond with the name used to implement the user. Deactivate Password Protection To deactivate the password protection, launch the Dräger CC-Vision software using "Admin" as user name and password, then select the menu item Deactivate Password Protection. When user name and password ("Admin") are entered in this field, you are no longer asked for the password the next time you start Dräger CC-Vision.

Page 26: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

Registration Registration Settings

0252

3615

.tif

Datapath: Path used to save instrument data Name of Archive File: file name of archive file used to log instrument check-out and instrument check-in

Default Instrument: Includes the instrument settings which were used for instrument check-out and instrument check-in. The instrument will be detected automatically after selecting "Autodetect". This will add some time to the readout process. Barcode Type: Setting of the selected bar code to be created when printing. Entry of User Data: If "Select from List" is activated before instrument check-out and instrument check-in, the user must be manually selected from the list of available user data. If "Bar Code" is selected, the user is selected when scanning the bar code.

26

Page 27: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

User Data

0262

3615

tif

This menu can be used for user database administration. New is used to call up a blank entry mask. After entering the user data, actuating Confirm will add the new user to the database. You can use the arrows to navigate the database. Use Print to send the bar code of the current user to the printer. Use Print All to send the bar codes of all users to the printer.

Inventory Data

0292

3615

.tif

Easy Registration will allow you to administrate items where inventory data is not readily available on the PC (e.g. part number, serial number). For these items, you can create a bar code which can then be scanned during instrument registration. New is used to call up a blank entry mask. After entering the inventory data, actuating Confirm will add the new item to the database. You can use the arrows to navigate the database. Use Print to send the bar code of the current item to the printer. Use Print All to send the bar codes of all items to the printer.

27

Page 28: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

Comments

0302

3615

.tif

This menu can be used for comment database administration. New is used to call up a blank entry mask. After entering the user data, actuating Confirm will add the new comment to the database. You can use the arrows to navigate the database. Use Print to send the bar code of the current comment to the printer. Use Print All to send the bar codes of all comments to the printer.

Instrument Check-Out

0272

3615

.tif

Use Readout to log the instrument check-out data. Select the user data corresponding to configuration from the list or take over via bar code. You can use the Accept command to check out the instrument. The instrument data and the user then receive the status "checked out" in a database.

28

Page 29: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

Instrument Check-In

0282

3615

.tif

Use Readout to log the instrument check-in data. Select the user data corresponding to configuration from the list or take over via bar code. You can use the Accept command to check in the instrument. The instrument data and the user then receive the status "checked in" in a database.

29

Page 30: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

Working with Dräger CC-Vision E-Cal To facilitate and automate the calibration of a large amount of gas detection instruments, Dräger offers a calibration station which can be equipped with up to 10 different calibration modules and a purge module. To this end, the Dräger CC-Vision E-Cal software version has an additional menu item called Autocalibration. In addition, a sequence of functions can be assigned to a configurable Go button, which is provided in this software version. For each E-Cal module, a measuring device button is displayed in toolbar. The Dräger CC-Vision software will detect a measuring device on an E-Cal module. The corresponding button is activated and can then be selected to transmit instrument data.

0222

3615

.tif

The instrument configuration procedure is the same as with basic version of the Dräger CC-Vision software. Therefore, please remember to read the corresponding chapter in the Dräger CC-Vision manual. During autocalibration, an instrument file is created for each instrument, or - if the instrument in question already has a file - the data in that file is updated. The automatically assigned file name is made up of: "instrument_type_fabrication_date.gdt".

30

Page 31: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

Checklists A regular inspection of the most important instrument components of the E-Cal Station is necessary to ensure a faultless verification and calibration of portable gas detection instruments. The following checklists indicate the components which need to be checked during installation and prior to commissioning. Checks when Installing or Recommissioning the E-Cal Station 1. Is the instrument equipped with power plugs approved by

Dräger? 2. Were the power plugs correctly connected? 3. Is the system operated in unpressurised condition? (Only if

"on demand" control valves were not used.) 4. Are the hoses between gas cylinders and master station less

than 10 meters long? 5. Are the hoses used to connect gas cylinders and master

station made of PVC, Viton, Tygon or Teflon and do they have an inside diameter of 4 mm?

6. Were the hoses used to connect modules and master station supplied by Dräger?

7. Are the gas outlet hoses less than 2 meters long? (Only if no purge module is installed.)

8. Are the device module gas outlets all equipped with connection pieces which have a transversal hole? (Only if a purge module is installed.)

9. Have all parts of the packaging and adhesive films been removed from the system?

If all questions listed above were answered with Yes, there should be nothing to prevent proper operation.

Carry out the following checks before commissioning the E-Cal Station: 1. Visual inspection of sensor gasket in individual modules. In

case of heavy contamination or visible defects, the gasket must be replaced.

2. Visual inspection of all installed hoses. Hoses with visible defects must be replaced.

3. Visual inspection of gas outlet hose. This hose must be properly connected and unclogged.

Menu with E-Cal Station The E-Cal version of the Dräger CC-Vision Software comes with an additional menu item called 'Autocalibration'. The toolbar has an additional, configurable button.

0202

3615

.tif

31

Page 32: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

32

Autocalibration Menu This menu item has all functions and settings for the automatic calibration and configuration of the Go button. 1 System Configuration Initialisation of the system interfaces. The serial numbers of the E-Cal modules (upper label) are entered in the corresponding fields. The COM ports for master and modules are determined. 2 Gas Configuration Gas configuration for the available valves. Gas, concentration and unit are entered using the selection lists. 3 Setup Go-Button The Go button is configurable. A window is displayed showing a selection of functions which can be assigned to this button. 4 Sensor List List of all possible sensors with tolerance specification for the gas test in %. The tolerance value can be edited here. 5 Sensor List (Quicktest) List of all sensors which can be assigned to the Quicktest. Here, sensors can be added and deleted, incl. tolerance and time values. 6 Zero Calibration An automatic zero calibration is carried out for the sensors of all powered up instruments in the E-Cal station. 7 Fresh Air Cal An automatic Fresh Air Cal is carried out for the sensors of all powered up instruments in the E-Cal station. 8 Span Calibration An automatic span calibration is carried out for the sensors of all powered up instruments in the E-Cal station.

9 Gas Test A gas test is carried out for the sensors of all powered up instruments in the E-Cal station. 10 Quicktest The quicktest is a simple sensor function test (for more information, refer to "Quicktest Configuration"). 11 Datalogger Readout Used to read out the contents of the datalogger. 12 Datalogger Reset Used to delete the contents of the datalogger. 13 Switch Off All Instruments Following the calibration, all measuring instruments in the modules are switched off. 14 Date/Time Synchronisation Synchronisation of date and time between measuring instrument and PC. 15 Ex Cal. Gas You can use this function with instruments which normally only display "Ex" (for 'EX sensor') to have them show the default calibration test gas for the EX sensor.

Page 33: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

33

Commissioning of Dräger CC-Vision E-Cal System and gas configuration must be completed prior to calibration and gas test automation. System Configuration The system configuration can be accessed using the menu Autocalibration - System Configuration. Enter the COM ports for master and modules. Enter the serial numbers of the E-Cal modules (upper label) in the corresponding fields. While doing so, pay attention to the sequence of entry to ensure that the button position on the toolbar and the position of the E-Cal module in the E-Cal station correspond. The "adapter mode" must be selected when using a module adapter. It is then only possible to configure a single instrument module. It is not possible to configure master and purge module in "adapter mode". Purge modules need to be activated and configured prior to use. USB interfaces need to be activated prior to use with the E-Cal station. A test button is used to test the communication with all configured E-Cal modules.

Notice The communication with gas detection instruments which might be inserted is not tested.

Configuring Gases The Autocalibration - Gas Configuration menu can be used to access calibration and test gas configuration settings. Enter gas, concentration and unit for the available valves. State if you are using mixed gas and if the gas is intended for gas test and/or span calibration.

Page 34: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

Go-Button Configuration Various functions can be assigned to the Go button, so that actuation of the button will trigger an automatic cycle. The configuration of the Go button can be accessed using the Autocalibration - Setup Go-Button menu.

0212

3615

.tif

34

Page 35: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

Gas Test Configuration The gas test is used to verify sensor calibration. The permissible tolerance is specified in the Autocalibration/Sensor List menu. The tolerance settings for each sensor can be edited as needed to meet individual requirements. Quicktest Configuration The quicktest is a simple sensor function test. The only thing tested is whether the measuring instrument exceeds a certain measurement value. If necessary, the quicktest can be carried out using a replacement gas. Provided that it is possible, the gas test is to preferred over the quicktest.

0412

3615

.tif

The quicktest limit value is stated as a percentage of the full scale value (e.g. XS H2S 6809110 full scale value = 100 ppm): With an indicated limit value of 10 %, the sensor must reach or exceed a measurement signal of 10 ppm.

0412

3615

.tif

The default quicktest is carried out using measuring gas. However, you can use Set Gas List to define a replacement gas for the quicktest of certain sensors. As of now, the following replacement gases are approved: methane, propane and pentane (for EX sensors), carbon monoxide (for XS OV and OVA) and ammonia (for XS Amine). For all other sensors, the measuring gas needs to be used for the quicktest.

35

Page 36: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

36

Testing and Calibrating with the E-Cal Station The Dräger E-Cal can be used to perform the following functions: gas test, quicktest, calibration of zero point, fresh air and span. The corresponding functions can be launched either using the Go button or via Autocalibration menu. Before launching the autocalibration, make sure: 1. that the system and gas configuration settings are correct. 2. that the measuring instruments are correctly positioned in

the holders of the E-Cal modules. 3. that the measuring instruments are powered up. For more important information on how to calibrate, please refer to chapter "Calibrating with Dräger CC-Vision". After launching the autocalibration, the calibration window is opened to show the current calibration process status.

Display of Calibration Results The green background colour of a result field indicates that all tested sensors were calibrated successfully. The red background colour indicates that an error has occured with the calibration of at least one sensor. A sensor shown in grey in the calibration window indicates that this sensor is not being calibrated. Reasons for this could be: - the sensor is inactive - the sensor is defective - the sensor type is not supported by E-Cal - The calibration gas required for this sensor is not available. The following sensors are currently supported: Sensor Supported Calibration / Test Gas CAT-Ex C CAT-Ex 125 Smart CAT-Ex Smart CAT-Ex HC Smart CAT-Ex FR

methane, propane, butane, pentane and H2

CAT-Ex 2 Smart IR Ex IR Ex HC

methane and propane

Smart IR CO2 Smart IR CO2 HC

carbon dioxide

Smart PID isobutylene XS-2 XS-R

H2S, CO, and O2

XS H2S, CO, O2, NO, NO2, Nh3, HCN, PH3, CO2, H2, COCI2

XXS H2S, CO, O2, NO2, NH3, HCN, PH3, CO2, H2, EO

Further sensors to follow. Upon request, Technical Support will provide information.

Page 37: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

37

Protocols in Dräger CC-Vision E-Cal In addition to the protocols provided by Dräger CC-Vision (see chapter "Protocols in Dräger CC-Vision"), Dräger CC-Vision E-Cal offers further data protocols. Special Protocols in Dräger CC-Vision E-Cal - System Configuration - Gas Configuration - Calibration History

Network Configuration This dialog uses a wizard designed to identify and configure all connected ConLine V24 compact device servers. Should the system fail to detect a connected compact device server, you can alternately enter the MAC address (which is specified on the rear identification tag). If the wizard succeeds in finding the compact device server, you can enter an alternate IP address which will be assigned and valid for the corresponding local area network. A valid IP address is usually provided by the LAN operator, which in most cases will be the internal IT department. Once the IP address has been saved, you can add the Dräger Mobile Printer connected to this compact device server to the mobile printer list, use the IP address for configuration and establish communication.

Page 38: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

38

Database Function Instead of using a text file, the calibration and test results can be archived in a database. The preferred archiving method is to be determined during installation. Prerequisites Up to now, the only prerequisite required for database operation is the database software: - MS SQL Server 2000 (incl. MDAC Version 2.6 runtime or

higher on the client) - MS Access Installing Dräger CC-Vision During program setup, database installation will be prompted. If database installation is affirmative, the next query will be to determine if an installation is already available. If this is not the case, a new database will be created. To this end, the so-called "Connection" string is initialised. In this context, it is important to verify that the corresponding database programs have been installed on the computer and that a connection has been established. After installation, it is no longer possible to switch to a version without database without reinstalling the program. Operation Dräger CC-Vision can be launched after the database connection has been established during installation. The connection will be tested every time the program is launched. The results of every calibration or test will be sent to the database. It is not possible to save the entire instrument parameters with this database version. DB Search The search dialog of this database version can be used to select instruments, sensors and calibration and/or test procedures. In addition, there are various search and filter options. Auto-Configuration

Notice Currently only available with Dräger Miniwarn.

The auto-configuration makes monitoring and setting of certain device configurations possible, provided these features have been activated in the GO-Button setup.

Page 39: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

39

E-Cal Error List Error Message Meaning Cause Action / Remedy Initialisation error Internal error during

calibration sensor warms up, instrument problem

Wait until sensor was warmed up, then repeat calibration. Repeat calibration and inform Service, if necessary.

Error during execution Internal error during calibration

Instrument problem Repeat calibration and inform Service, if necessary.

Error when stopping. Internal error during calibration

Instrument problem Repeat calibration and inform Service, if necessary.

Calibration not possible. Sensor warms up

See description Sensor warms up Wait until sensor was warmed up, then repeat calibration.

unknown error - unknown cause Repeat action and inform Service, if necessary.

Wrong value after calibration

Verification after the calibration is outside tolerance

Sensor not stable, deviation of entered gas concentration from the actual concentration. Sensor defective

Repeat calibration, Replace sensor

gas flow error the flow of the module does not correspond with the predefined set point

do not open cylinder valve, valve master defective, hose bent or clogged, empty cylinder

Check flow, if visual inspection of gas path is inconclusive, call Service

Communication error Transmission of data incorrect or not possible

Hardware or software problem Check cabling (check driver USB adapter)

concentration out of tolerance

gas test measurement value is outside preset tolerance

sensor drifts, sensor not calibrated, incorrect test gas concentration

calibrate sensor, check gas supply

insufficient battery capacity

insufficient battery capacity for selected action

insufficient battery capacity load instrument, replace battery

measurement value is not stable

measurement value fluctuates

sensor defective (Cat-Ex poisoned) sensor opening contaminated

visual inspection of sensor, replace sensor

Date/time of instrument are wrong or not set

cannot be calibrated Date/time of instrument are wrong or not set

Set correct date and time

Page 40: Dräger CC-Vision - Draeger · Dräger CC-Vision is available in two software versions: Dräger CC-Vision is used to directly connect individual gas detection instruments to a COM

40

Dräger Safety AG & Co. KGaA Revalstrasse 1 D-23560 Lübeck, Germany Phone: +49 451 8 82-0 Fax: 1+49 451 8 82-20 80 www.draeger.com 90 23 615 GA 4634.820 en © Dräger Safety AG & Co. KGaA Edition 07 – March 2010 (Edition 01 – September 1999) Subject to alterations