Download - Liquisys (CCM253)_Chlorine_analyser
BA 209C/07/en/07.02No. 51517599Software version 2.30 or later
Supplement toBA 193C/07/enBA 194C/07/enBA 199C/07/enBA 200C/07/enBA 214C/07/en
PROFIBUS-PA/-DPField Communication forLiquisys M CXM 223/253Operating Instructions
Endress HauserThe Power of Know How
Quality made byEndress+Hauser
ISO 9001
ENDRESS+HAUSERLIQUISYS M
ENDRESS+HAUSERLIQUISYS M
You need information on the instrument?Please read the following chapters:
SafetyGeneralinformation
You wish to install and start up the instrument.The required steps are described in these chapters:
Installation
You wish to operate or reconfigure the instrument:
Start-up
Communication
Technical data
Index
Accessories % Ω°C% Ω°C
XM53PE0.CHP
TABLE OF CONTENTS1 General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21.1 Symbols used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32.1 Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32.2 General safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
3 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43.1 System equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43.2 Electrical connection PA instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53.3 Electrical connection DP instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
4 Start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94.1 Setting of device address . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94.2 Device data and type files. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
5 Communication. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125.1 Cyclical data exchange (Data_Exchange) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125.2 Acyclical data exchange. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145.3 Operation via Commuwin II . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
6 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
7 Technical data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
8 Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Liquisys M PROFIBUS-PA/-DP
Endress+Hauser 1
1 General information
These operating instructions were designedspecifically for use with transmitters of thefamily Liquisys M CXM 223/253. They containspecific information for instruments with thePROFIBUS-PA interface (Process Field Bus –Process Automation) or the PROFIBUS-DPinterface (Process Fieldbus DecentralizedPeriphery).
Please refer to the corresponding standardoperating instructions for information oninstallation and general transmitter operation:BA 193C/07/en (CLM 223/253)BA 194C/07/en (CPM 223/253)BA 199C/07/en (COM 223/253)BA 200C/07/en (CUM 223/253)BA 214C/07/en (CCM 223/253).
1.1 Symbols used
Warning:
This symbol alerts to hazards whichmay cause serious injuries as well asdamage to equipment if ignored.
Note:
This symbol indicates importantitems of information.Ignoring this information may resultin malfunction.
General information Liquisys M PROFIBUS-PA/-DP
2 Endress+Hauser
2 Safety
2.1 Intended use
Operation via PROFIBUS-PA interface
PROFIBUS is an open field bus standardaccording to DIN 19 245 / EN 50 170. It hasbeen created specifically for process controlpurposes and permits connection of severalmeasuring instruments to a bus line.
The transmission method according toIEC 1158-2 guarantees safe signaltransmission. The PROFIBUS-PA interfacepermits operation of the instrument from thePC via the Commuwin II operating program.
2.2 General safety instructions
Warning:
• The notes and warnings in theseoperating instructions must bestrictly adhered to.
• The notes and warnings containedin the standard operatinginstructions for the instrument inquestion (193C, 194C, 199C,200C, 214C) must be strictlyadhered to.
XM53PE2.CHP
Liquisys M PROFIBUS-PA/-DP Safety
Endress+Hauser 3
3 Installation
3.1 System equipment
A complete system equipment comprises thefollowing components:
• Transmitter Liquisys M CXM 223/253-xx3• Segment coupler (only with PA)• Programmable logic controller (PLC) or PC
with PROFIBUS interface and Commuwin IIoperating program
• PROFIBUS-PA terminating resistor• Cabling incl. bus distributor.
The maximum number of transmitters in a bussegment is determined by their powerconsumption, the bus coupler output and therequired bus length.
Note:
For detailed information on functionand connection of a PROFIBUSsystem, please refer to TechnicalInformation TI 260F/00/en, which isavailable via the Endress+Hausersales agencies (see back cover ofthese operating instructions).
PROFIBUS-PA
PROFIBUS-DP
ENDRESS+HAUSERLIQUISYS SENDRESS+HAUSERLIQUISYS S
ENDRESS+HAUSERLIQUISYS
CAL REL
REL1
REL1
ALARMALARMREL2
REL2E
ENDRESS+HAUSERLIQUISYS
CAL REL
REL1
REL1
ALARMREL2
REL2E
Segmentcoupler
PC with Commuwin IIoperating program
PLC
XM253-02.CDRFig. 3.1Measuring systems withPROFIBUS interface
e.g. Liquisys M CPM 223with CPA 140
e.g. Liquisys M COM 253with COS 4
Installation Liquisys M PROFIBUS-PA/-DP
4 Endress+Hauser
3.2 Electrical connection PA instrument
Bus cable connection to field instrument
The bus cable connection to the fieldinstrument can be performed with or withoutM12 plug. Connect the bus cable as follows:
• Unscrew the four recessed-head screwsand remove the instrument cover.
• Thread the cable through the openedcable entry into the connectioncompartment.
• Connect the wires of the bus cable to theterminal block according to the figurebelow. Polarity reversal of the connectionsPA+ and PA– does not affect operation.
• Tighten the cable entry.
Connection of several field instruments
565554 51 52 53 5957 58 47 48 49 41 42 43
3231 33 34 9385 86 94 81 82
565554 51 52 53 5957 58 47 48 49 41 42 43
3231 33 34 9385 86 94 81 82
XM253AN.CDR
Pg couplingM12 plug
Connection terminal block
bk bn blPA+ PA–
Fig. 3.2
Bus cable connectionto field instrumentLiquisys M-PACPM 253 / CLM 253inductive / COM 253 /CUM 253 / CCM 253
left:Connection withM12 plug
right:Connection withoutM12 plug
565554 51 52 53 5957 58 47 48 49 41 42 43
3231 33 34 9385 86 94 81 82
565554 51 52 53 5957 58 47 48 49 41 42 43
3231 33 34 9385 86 94 81 82
Pg couplingM12 plug
PA+ PA–
bn bl bk Connection terminal block
PA+ PA–
XM253AN.CDR
screen
Fig. 3.3
Bus cable connectionto field instrumentLiquisys M-PACLM 253 conductive
left:Connection withM12 plug
right:Connection withoutM12 plug
Supplyinstrument /segmentcoupler
Connection box Connection box
Condensatorsmax. 10 nF/250 V AC
Field instrument 1 Field instrument 2
XM253A5.CDR Fig. 3.4Connection of severalfield instruments
Screen PA+ PA–
XM53PE3.CHP
Liquisys M PROFIBUS-PA/-DP Installation
Endress+Hauser 5
Bus cable connection to panel-mounted instrument
The bus cable connection to the panel-mounted instrument is performed directly at the terminalblock on the rear of the instrument.
Connection of several panel-mounted instruments
For the connection of several panel-mounted instruments to the bus cable, the connection linesmust be bridged according to the figure below.
82819493868594 78959899
XM253A2.CDR
PA+ PA– Screen
Connection terminal block
Fig. 3.5
Bus cable connection topanel-mounted instrumentLiquisys M-PACPM 223 / CLM 223inductive / COM 223 /CUM 223 / CCM 223
82819493868594 78959899Connection terminal block
PA+ PA– Screen
XM253A4.CDR
bn bl
Fig. 3.6
Bus cable connection topanel-mounted instrumentLiquisys M-PACLM 223 conductive
XM253-03.CDR
Panel-mountedinstrument 1
Panel-mountedinstrument 2
Connect and ground shields
Fig. 3.7
Connection of severalpanel-mountedinstruments
bn bl
Installation Liquisys M PROFIBUS-PA/-DP
6 Endress+Hauser
3.3 Electrical connection DP instrument
Bus cable connection to field instrument
Bus termination
Install a bus termination for each instrumentconnected.
565554 51 52 53 5957 58 47 48 49 41 42 43
3231 33 34 9385 86 94 81 82
Connection terminal block
S B A
XM253A6.CDR Fig. 3.8
Bus cable connectionto field instrumentLiquisys M-DPCPM 253 / CLM 253inductive / COM 253 /CUM 253 / CCM 253
565554 51 52 53 5957 58 47 48 49 41 42 43
3231 33 34 9385 86 94 81 82
Connection terminal block
B A S
XM253A7.CDR Fig. 3.10
Bus cable connectionto field instrumentLiquisys M-DPCLM 253 conductive
31(99)
32(98)
33(67)
390 Ω 220 Ω 390 Ω
34(68)
XM253A9.CDR Fig. 3.9 Bus terminating resistor
XM53PE3.CHP
Liquisys M PROFIBUS-PA/-DP Installation
Endress+Hauser 7
Bus cable connection panel-mounted instrument
The bus cable connection to the panel-mounted instrument is performed via a plug-in blockintegrated in the cover frame.
82819493868568 69679899Connection terminal block
gn ye br gr wt
Plug-in block
XM253A8.CDRFig. 3.11
Bus cable connection topanel-mounted instrumentLiquisys M-DPCPM 253 / CLM 253ind. / cond. / COM 253 /CUM 253 / CCM 253
Installation Liquisys M PROFIBUS-PA/-DP
8 Endress+Hauser
4 Start-up
4.1 Setting of device address
The addresses of all PROFIBUS-PA devicesmust be set. An incorrect address setting willprevent detection of the measuring instrumentby the control system.
All instruments are set to address 126 at thefactory. This address can be used to test thefunction of the instrument and for connectionto a PROFIBUS-PA network. Subsequently,this address must be changed to connnectadditional devices. The device address canbe set via
• field operation• PROFIBUS communication• DIL switch at the instrument.
Note:
• Valid device address range:0 ... 126.
• Each address may only beassigned once in a PROFIBUS-PAnetwork.
• The double arrow in the displayindicates communication viaPROFIBUS.
Setting the device address via field operation (software setting)
The device address is set via keys inthe menu field Ι1 of the function groupINTERFACE.
Note:
The address can only be set viasoftware if DIL switch 8 is set tosoftware position. DIL switch 8 is setto software position at the factory.The opposite figure shows thedefault setting of the switch.
Setting the device address via PROFIBUS communication
The device address is set via the Set Slave Adr service.
Setting the device address via DIL switch (hardware setting)
The DIL switch is situated at the electronics module above the display. To set the deviceaddress, proceed as follows:
• Field instrument:– Unscrew the four recessed-head
screws and remove the instrumentcover. The electronics module withDIL switch is situated in the instrumentcover.
• Panel-mounted instrument:– The DIL switch is accessible through
a break in the instrument above thedisplay.
• Set the device address (0 ... 126) atswitches 1 to 7, e.g. 18 = 2 + 16.
• For setting the address via DIL switch→ set switch 8 to OFF.
• For field instrument, close the instrumentcover again.
126 I1
Address
1 2 3 4 5 6 7
ON
OFF
8
ON = software address
XM253-06.CDR
1
1
2
2
3
4
4
8
5
16
6
32
7
64
ON
OFF
8
OFF = hardware address
XM253-04.CDR
XM53PE4.CHP
Liquisys M PROFIBUS-PA/-DP Start-up
Endress+Hauser 9
4.2 Device data and type files
The device data file (GSD) is required toconfigure a PROFIBUS-DP network. The GSD(a simple text file) describes e.g., which datatransfer rate is supported by the device orwhich digital information in which format thePLC receives from the device.
Each device is assigned an ID number by thePROFIBUS user organization (PNO). Thename of the device data file (GSD) is derivedfrom this ID number.
Note:
• For Endress+Hauser devices,the ID number always starts with“15XX”.
The GSD files for all Endress+Hauser devicescan be obtained from the following sources:
– via the Internet:E+H: http://www.endress.com(Product Avenue → DownloadStreet → Field CommunicationStreet)PNO: http://www.profibus.com(GSD library)
– on floppy disk from E+H:order no. 943157-0000
Contents of download file from the Internetand of floppy disk:
• All Endress+Hauser GSD files• All Endress+Hauser bitmap files• Useful information on the devices
Name of device ID no. GSD Type file Bitmaps
Liquisys MCLM 223/253-xx3 1515 hex EH_ _1515.gsd EH_1515x.200
EH1515_d.bmpEH1515_n.bmpEH1515_s.bmp
Liquisys MCPM 223/253-xx3 1516 hex EH_ _1516.gsd EH_1516x.200
EH1516_d.bmpEH1516_n.bmpEH1516_s.bmp
Liquisys MCUM 223/253-xx3 1517 hex EH_ _1517.gsd EH_1517x.200
EH1517_d.bmpEH1517_n.bmpEH1517_s.bmp
Liquisys MCOM 223/253-xx3 1518 hex EH_ _1518.gsd EH_1518x.200
EH1518_d.bmpEH1518_n.bmpEH1518_s.bmp
Liquisys MCCM 223/253-xx3 1519 hex EH_ _1519.gsd EH_1519x.200
EH1519_d.bmpEH1519_n.bmpEH1519_s.bmp
Liquisys MCLM 223/253-xx4 1521 hex EH_ _1521.gsd EH_1521x.200
EH1521_d.bmpEH1521_n.bmpEH1521_s.bmp
Liquisys MCPM 223/253-xx4 1520 hex EH_ _1520.gsd EH_1520x.200
EH1520_d.bmpEH1520_n.bmpEH1520_s.bmp
Liquisys MCUM 223/253-xx4 151f hex EH_ _151f.gsd EH_151fx.200
EH151f_d.bmpEH151f_n.bmpEH151f_s.bmp
Liquisys MCOM 223/253-xx4 151e hex EH_ _151e.gsd EH_151ex.200
EH151e_d.bmpEH151e_n.bmpEH151e_s.bmp
Liquisys MCCM 223/253-xx4 151d hex EH_ _151d.gsd EH_151dx.200
EH151d_d.bmpEH151d_n.bmpEH151d_s.bmp
Start-up Liquisys M PROFIBUS-PA/-DP
10 Endress+Hauser
Working with the GSD / type files
The GSD files must be copied to a specificsubdirectory of the PROFIBUS-DPconfiguration software of your PLC.
Note:
• The subdirectory to be used forthe Siemens STEP7 configurationsoftware of the SiemensS7-300/400 PLC is\siemens\step7\s7data\gsd.
• If you are using a differentconfiguration software, then contactthe manufacturer of your PLC forthe correct subdirectory to use.
• If you have a Siemens S5 PLC andif the PROFIBUS-PA network isconfigured with the COM ET200configuration software, then youwill need type files (x.200 files).
• The GSD directory contains asubdirectory which holds the GSDfiles with the standard identifier(0x94). For example, these GSDfiles are intended for use with anAllen-Bradley PLC5.
XM53PE4.CHP
Liquisys M PROFIBUS-PA/-DP Start-up
Endress+Hauser 11
5 Communication
The Liquisys M CXM 223/253 transmitter supplies the measured values using the cyclicalservice according to the PROFIBUS-DP protocol. Other parameters are made available usingthe acyclical service.
5.1 Cyclical data exchange (Data_Exchange)
The Data_Exchange service enables a PLC to read input data from the Liquisys M CXM223/253 transmitter. The cyclical data message for the maximum configuration of the LiquisysM CXM 223/253 is as follows:
Maximum configuration Liquisys M CXM 223/25
IEEE 754 floating point number
The Liquisys M CXM 223/253 transmitter uses a subset of the status codes defined by thePROFIBUS user organization (PNO). (See also: PNO specification “PROFIBUS-PA Profile forProcess Control Devices; General Requirements V2.0”.)
Input dataindex Data Access Data format /
remarksUnit
(factory setting)
0, 1, 2, 3 Main parameter read 32-bit floating pointnumber (IEEE 754)
CPM 223/253: pH; mV; %CLM 223/253: mS/cm; %; MΩCUM 223/253: FNU; ppm; g/l;%COM 223/253: mg/l; % SATCCM 223/253: mg/l
4 Main parameterstatus read See status codes –
5, 6, 7, 8 Temperature read 32-bit floating pointnumber (IEEE 754) °C
9 Temperaturestatus read See status codes –
10,11,12,13 pH / ORP read 32-bit floating pointnumber (IEEE 754)
pH or mV(dep. on operating mode)
14 pH / ORP status read See status codes –Tab. 5.1 Maximum configuration
D15 D14 D13 D12 D11 D10 D9 D8 D7 D6 D5 D4 D3 D2 D1 D0
S Exponent (E) Fraction (F)
27 26 25 24 23 22 21 20 2–1 2–2 2–3 2–4 2–5 2–6 2–7
Fraction (F)
2–8 2–9 2–10 2–11 2–12 2–13 2–14 2–15 2–16 2–17 2–18 2–19 2–20 2–21 2–22 2–23Tab. 5.2
IEEE 754 floating pointnumber
Communication Liquisys M PROFIBUS-PA/-DP
12 Endress+Hauser
Status codes for main parameter (pH, conductivity, turbidity, oxygen) and temperature
If you do not wish to use all cyclical data provided by the Liquisys M CXM 223/253, then youcan eliminate cyclical data blocks from the cyclical message with the aid of the deviceconfiguration data string (Chk_Cfg) using the PLC software. You should only activate those datablocks which are actually processed by the system. This will improve the data throughput of aPROFIBUS network.
On the Liquisys M CXM 223/253 transmitter, two data blocks can be activated for cyclical dataexchange. The following table shows the configuration data needed for this (h means that thenumber is a hexadecimal number). The shortened identifier 94h can also be used. Theextended format is to be preferred, however, because it includes information for interpretationof the data.
Liquisys M CXM 223/253 configuration data
The following configurations are useful for the data blocks available with the Liquisys M CXM223/253:
Data blocks that are not activated must be identified by a zero as a placeholder in theconfiguration data string if other configuration data follow. Zeros at the end of the configurationdata may be omitted.
Status code Meaning Device status
hex. (dec.)
08h (8)Not connected(no meas. valueavailable)
BAD
0Ch (12) Device failure BAD
80h (128) O.K. GOOD Tab. 5.5
Status codes formain parameterand temperature
Cyclical data block Sequencein Chk_Cfg
Extended format Short formatData block active active Data block inactive
Main parameter 1 42h, 84h, 08h, 05h 94h 00h
Temperature 2 42h, 84h, 08h, 05h 94h 00h
pH / ORP withCCM 223/253 3 42h, 84h, 08h, 05h 94h 00h
Tab. 5.3 Configuration data
Activated data blocks Configuration data string(Chk_Cfg) Length (Chk_cfg)
Main parameter with status+ temperature with status
42h, 84h, 08h, 05h,42h, 84h, 08h, 05halternatively: 94h, 94h
8 bytes(2 bytes)
Main parameter with status 42h, 84h, 08h, 05h, (00h)alternatively: 94h (00h)
4 ... 5 bytes(1 ... 2 bytes)
Temperature with status 00h, 42h, 84h, 08h, 05halternatively: 00h, 94h
5 bytes(2 bytes)
pH/ORP with CCM 223/253 42h, 84h, 08h, 05h 4 BytesTab. 5.4 Useful configurations
XM53PE5.CHP
Liquisys M PROFIBUS-PA/-DP Communication
Endress+Hauser 13
5.2 Acyclical data exchange
Acyclic data transmission is used to transfer parameters during commissioning, duringmaintenance or to display other measured variables that are not contained in the useful datatraffic.
The software of the Liquisys M CXM 223/253 transmitter includes the so-called physical blockcontaining instrument-specific information, such as measuring point designation (tag), softwareversion, etc., and a block with manufacturer-specific data used for instrument operation.
Generally, a distinction is made between Class 1 and Class 2 master connections.The Liquisys M transmitter only supports Class 2 master connections.
• Two Class 2 masters are permitted with Liquisys M. This means that two Class 2 masterscan access Liquisys M at the same time. However, you must make certain that they do notboth attempt to write to the same data. Otherwise the data consistency can lo longer beguaranteed.
• When a Class 2 master reads parameters, it sends a request telegram to the transmitterspecifiying the device address, the slot/index and the expected record length.The transmitter answers with the requested record if the record exists and is the correctlength (byte).
• When a Class 2 master writes parameters, it transmits the address of the transmitter, theslot/index, length information (byte) and the record. The transmitter acknowledges this writejob after completion. A Class 2 master can access the blocks mentioned above.
The device management and physical block are identical for all parameters measured.Please refer to the following tables for the index (all parameters are situated in slot 1).
Communication Liquisys M PROFIBUS-PA/-DP
14 Endress+Hauser
Slot / index list for device management and physical block
Parameter Matrix(Standard)
Slot /Index
Sizein byte(record)
Type read/write
Device Management
Directory Object Header 1 / 0 12 UNSIGNED16 R
Composite List Directory Entries 1 / 1 24 UNSIGNED16 R
Physical Block Parameter
PB Block Objekt 1 / 14 20 DS-32 R
PB Static Revision 1 / 15 2 UNSIGNED16 R
PB Tag Description VAH0 1 / 16 32 OSTRING R / W
PB Strategy 1 / 17 2 UNSIGNED16 R / W
PB Alert Key 1 / 18 1 UNSIGNED8 R / W
PB Target Mode 1 / 19 1 UNSIGNED8 R / W
PB Mode Block 1 / 20 3 DS-37 R
PB Alarm Summary 1 / 21 8 DS-42 R
PB Software Revision 1 / 22 16 OSTRING R
PB Hardware Revision 1 / 23 16 OSTRING R
PB Device Manufacturer ID 1 / 24 2 UNSIGNED16 R
PB Device ID 1 / 25 16 OSTRING R
PB Device Serial Number 1 / 26 16 OSTRING R
PB Diagnosis 1 / 27 4 OSTRING R
PB Diagnosis Extension 1 / 28 6 OSTRING R
PB Diagnosis Mask 1 / 29 4 OSTRING R
PB Diagnosis Mask Extension 1 / 30 6 OSTRING R
PB Device Certification 1 / 31 16 OSTRING R / W
PB Security Locking 1 / 32 2 UNSIGNED16 R / W
PB Factory Reset 1 / 33 2 UNSIGNED16 R / W
PB Descriptor 1 / 44 32 OSTRING R / W
PB Device Message 1 / 45 32 OSTRING R / W
PB Device Installation Date 1 / 46 8 OSTRING R / W
PB Actual Errorcode V8H0 1 / 62 2 UNSIGNED16 R
PB UpDown Features Supported 1 / 64 1 OSTRING R
PB UpDown Control Parameter 1 / 65 1 UNSIGNED8 W
PB UpDown Parameter 1 / 66 20 UPDOWNDATA R / W
PB Device Bus Address VAH1 1 / 67 1 SIGNED8 R
PB Device & Software Number 1 / 68 2 UNSIGNED16 R
PB View_1 1 / 74 17 OSTRING RTab. 5.6
Slot / index list fordevice managementand physical block
XM53PE5.CHP
Liquisys M PROFIBUS-PA/-DP Communication
Endress+Hauser 15
Slot / index list for pH / redox
Parameter Matrix(Standard)
Slot /Index
Sizein byte(record)
Type read/write
Manufacturer specific Block
Block Object 1 / 100 20 DS-32 R
Main measured value V0H0 1 / 108 4 FLOAT R
Measured temperature value V0H1 1 / 109 4 FLOAT R
Operating mode V0H2 1 / 110 1 UNSIGNED8 R / W
Connection type V0H3 1 / 111 1 UNSIGNED8 R / W
Electrode type V0H5 1 / 112 1 UNSIGNED8 R / W
Measured value damping V0H6 1 / 113 1 UNSIGNED8 R / W
Measured value in mV V1H7 1 / 114 4 FLOAT R
Current measured / offset V0H8 1 / 115 4 FLOAT R / W
Calculated offset V0H9 1 / 116 4 FLOAT R
Temperature compensation type V1H0 1 / 117 1 UNSIGNED8 R / W
Entry of MTC temperature V1H1 1 / 118 4 FLOAT R / W
Temp. compensation for calibration V1H2 1 / 119 1 UNSIGNED8 R / W
Entry of process temperature V1H8 1 / 120 4 FLOAT R / W
Entry of temperature offset V1H9 1 / 121 4 FLOAT R / W
Beg. of meas. (bar graph) V2H3 1 / 122 4 FLOAT R
End of meas. (bar graph) V2H4 1 / 123 4 FLOAT R
Contact selection V3H0 1 / 124 1 UNSIGNED8 R / W
Contact configuration V3H1 1 / 125 1 UNSIGNED8 R / W
Switch contact on or off V3H2 1 / 126 1 UNSIGNED8 R / W
Switch-on point (LC) V4H0 1 / 127 4 FLOAT R / W
Switch-off point (LC) V4H1 1 / 128 4 FLOAT R / W
Pickup delay (LC) V4H2 1 / 129 2 UNSIGNED16 R / W
Dropout delay (LC) V4H3 1 / 130 2 UNSIGNED16 R / W
Alarm threshold (LC) V4H4 1 / 131 4 FLOAT R / W
Setpoint (PID / neutra) V4H0 1 / 132 4 FLOAT R / W
Control gain Kp (PID / neutra) V4H1 1 / 133 4 FLOAT R / W
Integral action time Tn (PID / neutra) V4H2 1 / 134 4 FLOAT R / W
Derivative action time Tv (PID / neutra) V4H3 1 / 135 4 FLOAT R / W
Control characteristic (PID) V4H4 1 / 136 1 UNSIGNED8 R / W
Operating mode of controller (PID / neutra) V4H5 1 / 137 1 UNSIGNED8 R / W
Pulse period (PID / neutra) V4H6 1 / 138 4 FLOAT R / W
Pulse frequency (PID / neutra) V4H7 1 / 139 2 UNSIGNED16 R / W
Minimum ON time (PID / neutra) V4H8 1 / 140 4 FLOAT R / W
Rinse time (timer) V4H0 1 / 141 2 UNSIGNED16 R / W
Pause time (timer) V4H1 1 / 142 2 UNSIGNED16 R / W
Minimum pause time (timer) V4H2 1 / 143 2 UNSIGNED16 R / W
Start pulse (cleaning) V4H0 1 / 144 1 UNSIGNED8 R / W
Pre-rinse time (cleaning) V4H1 1 / 145 2 UNSIGNED16 R / W
Cleaning time (cleaning) V4H2 1 / 146 2 UNSIGNED16 R / W
Post-rinse time (cleaning) V4H3 1 / 147 2 UNSIGNED16 R / W
Repeat cycles (cleaning) V4H4 1 / 148 1 UNSIGNED8 R / W
Pause time (cleaning) V4H5 1 / 149 2 UNSIGNED16 R / W
Minimum pause time (cleaning) V4H6 1 / 150 2 UNSIGNED16 R / W
Economy (cleaning) V4H7 1 / 151 1 UNSIGNED8 R / WTab. 5.7Slot / index list forpH / redox
Communication Liquisys M PROFIBUS-PA/-DP
16 Endress+Hauser
Slot / index list for pH / redox (continued)
Parameter Matrix(Standard)
Slot /Index
Sizein byte(record)
Type read/write
Temperatursensor
V6H0
V0H4
1 / 160
1 / 152
1
1
UNSIGNED8
UNSIGNED8
R / W
R / W
Messfrequenz
V6H1
V0H7
1 / 161
1 / 153
1
1
UNSIGNED8
UNSIGNED8
UNSIGNED8
R / W
R / W
Transmitter type
V6H2 1 / 162
1 / 154
4
4
FLOAT R / W
Limit contact status
V6H3
V4H5
1 / 163
1 / 155
1
1
UNSIGNED8 R / W
Serial number
V8H1
VAH3
1 / 164
1 / 188
1
1
UNSIGNED8
OSTRING
R / W
Basic load
V8H2
V4H9
1 / 165
1 / 189
1
1
UNSIGNED8
UNSIGNED8
R / W
R / W
Controller type
V8H3
V5H8/V4H8
1 / 166
1 / 198
2
2
UNSIGNED16
UNSIGNED8
UNSIGNED8
R / W
R / W
Set current input active
Current input off delay
Current input on delay
Current input switch-off direction
Current input feedforward control
Current input gain
Current input available
Current input switch-off point
V8H4
V5H0
1 / 167
1 / 199
1
1
UNSIGNED8 R / W
R / W
R / W
V8H5
V5H1
1 / 168
1 / 200
1
1
UNSIGNED8 R
V8H6
V5H2
1 / 169
1 / 201
1
1
UNSIGNED8 R / W
R / W
V8H8
V5H3
1 / 170
1 / 202
1
1
UNSIGNED8
UNSIGNED8
UNSIGNED8
R / W
R / W
V8H9
V5H4
1 / 171
1 / 203
2
2
UNSIGNED16
UNSIGNED16
UNSIGNED16
R / W
R / W
V9H0
V5H5
1 / 172
1 / 204
1
1
UNSIGNED8
UNSIGNED8
UNSIGNED8
R / W
R / W
V9H1
V5H6
1 / 173
1 / 205
2
2
UNSIGNED16 R / W
R / W
V9H2 1 / 174
1 / 206
1
1
UNSIGNED8
UNSIGNED8
R
R
Alarm threshold monitoring
Process monitoring
Alarm delay
Max. period lower limit exceeded
Max. period upper limit exceeded
Lower alarmthreshold
Upper alarmthreshold
Limit of process monitoring
V9H3
V7H1
V7H5
1 / 175
1 / 207
1 / 211
1
1
1
UNSIGNED8
UNSIGNED8
UNSIGNED8
R
V9H4
V7H2
V7H6
1 / 176
1 / 208
1 / 212
1
1
1
UNSIGNED8 R / W
R / W
R / W
R / W
R / W
R / W
R / W
R / W
R / W
VAH2
V7H3
V7H4
V7H8
V7H7
1 / 177
1 / 209
1 / 213
2
2
2
UNSIGNED16
FLOAT
FLOAT
FLOAT
UNSIGNED16
UNSIGNED16
UNSIGNED16
R
R
R
R
1 / 178
1 / 210
1 / 214
1
1
1
UNSIGNED8 R
SCS alarm measuring electrode
SCS alarm reference electrode
SCS alarm threshold
PCS alarm
Alarm contact type
Unit alarm delay
Alarm delay
Error number selection
Error status
Alarm contact effective
Cleaning trigger
Release code
Hold configuration
Hold dwell period
Plus package released
Chemoclean released
Reset of instrument
Software version
Number of relays
Tab. 5.8Slot index list forpH / redox (continued)
XM53PE5.CHP
Liquisys M PROFIBUS-PA/-DP Communication
Endress+Hauser 17
Slot / index list for conductivity
Parameter Matrix(Standard)
Slot /Index
Sizein byte(record)
Type read/write
Manufacturer specific Block
Block Object 1 / 100 20 DS-32 R
Main measured value V0H0 1 / 108 4 FLOAT R
Measured temperature value V0H1 1 / 109 4 FLOAT R
Operating mode V0H2 1 / 110 1 UNSIGNED8 R / W
Display unit conc. V0H3 1 / 111 1 UNSIGNED8 R / W
Display format V0H4 1 / 112 1 UNSIGNED8 R / W
Display unit cd. / res. V0H5 1 / 113 1 UNSIGNED8 R / W
Measured value damping V0H6 1 / 114 1 UNSIGNED8 R / W
Uncompensated meas. value V0H7 1 / 115 4 FLOAT R
Temperature sensor V1H0 1 / 116 1 UNSIGNED8 R / W
Temperature compensation type V1H1 1 / 117 1 UNSIGNED8 R / W
Temperature coefficient alpha V1H2 1 / 118 4 FLOAT R / W
Process temperature V1H3 1 / 119 4 FLOAT R / W
Cell constant V1H4 1 / 120 4 FLOAT R / W
Cable resistance V1H5 1 / 121 4 FLOAT R / W
Entry of process temperature V1H8 1 / 122 4 FLOAT R / W
Entry of temperature offset V1H9 1 / 123 4 FLOAT R / W
Beg. of meas. (bar graph) –
–
1 / 124 4 FLOAT R
End of meas. (bar graph) 1 / 125 4 FLOAT R
Contact selection V3H0 1 / 126 1 UNSIGNED8 R / W
Contact configuration V3H1 1 / 127 1 UNSIGNED8 R / W
Switch contact on or off V3H2 1 / 128 1 UNSIGNED8 R / W
Switch-on point (LC) V4H0 1 / 129 4 FLOAT R / W
Switch-off point (LC) V4H1 1 / 130 4 FLOAT R / W
Pickup delay (LC) V4H2 1 / 131 2 UNSIGNED16 R / W
Dropout delay (LC) V4H3 1 / 132 2 UNSIGNED16 R / W
Alarm threshold (LC) V4H4 1 / 133 4 FLOAT R / W
Setpoint (PID / neutra) V4H0 1 / 134 4 FLOAT R / W
Control gain Kp (PID / neutra) V4H1 1 / 135 4 FLOAT R / W
Integral action time Tn (PID / neutra) V4H2 1 / 136 4 FLOAT R / W
Derivative action time Tv (PID / neutra) V4H3 1 / 137 4 FLOAT R / W
Control characteristic (PID) V4H4 1 / 138 1 UNSIGNED8 R / W
Operating mode of controller (PID / neutra) V4H5 1 / 139 1 UNSIGNED8 R / W
Pulse period (PID / neutra) V4H6 1 / 140 4 FLOAT R / W
Pulse frequency (PID / neutra) V4H7 1 / 141 2 UNSIGNED16 R / W
Minimum ON time (PID / neutra) V4H8 1 / 142 4 FLOAT R / W
Rinse time (timer) V4H0 1 / 143 2 UNSIGNED16 R / W
Pause time (timer) V4H1 1 / 144 2 UNSIGNED16 R / W
Minimum pause time (timer) V4H2 1 / 145 2 UNSIGNED16 R / W
Start pulse (cleaning) V4H0 1 / 146 1 UNSIGNED8 R / W
Pre-rinse time (cleaning) V4H1 1 / 147 2 UNSIGNED16 R / W
Cleaning time (cleaning) V4H2 1 / 148 2 UNSIGNED16 R / W
Post-rinse time (cleaning) V4H3 1 / 149 2 UNSIGNED16 R / W
Repeat cycles (cleaning) V4H4 1 / 150 1 UNSIGNED8 R / W
Pause time (cleaning) V4H5 1 / 151 2 UNSIGNED16 R / W
Minimum pause time (cleaning) V4H6 1 / 152 2 UNSIGNED16 R / W
Economy (cleaning) V4H7 1 / 153 1 UNSIGNED8 R / WTab. 5.9Slot / index list forconductivity
Communication Liquisys M PROFIBUS-PA/-DP
18 Endress+Hauser
Slot / index list for conductivity (continued)
Parameter Matrix(Standard)
Slot /Index
Sizein byte(record)
Type read/write
V5H0 1 / 154 1 UNSIGNED8 R / W
V5H1 1 / 155 1 UNSIGNED8 R / W
V5H2 1 / 156 1 UNSIGNED8 R / W
V5H3 1 / 157 1 UNSIGNED8 R / W
V5H4 1 / 158 1 UNSIGNED8 R / W
V5H5 1 / 159 4 FLOAT R / W
V5H6 1 / 160 4 FLOAT R / W
V5H7 1 / 161 4 FLOAT R / W
V5H8 1 / 162 1 UNSIGNED8 R / W
V6H0 1 / 163 1 UNSIGNED8 R / W
V6H1 1 / 164 1 UNSIGNED8 R / W
V6H2 1 / 165 1 UNSIGNED8 R / W
V6H3 1 / 166 4 FLOAT R / W
V6H4 1 / 167 4 FLOAT R / W
V6H5 1 / 168 1 UNSIGNED8 R / W
V7H0 1 / 169 1 UNSIGNED8 R / W
V4H5 1 / 170 1 UNSIGNED8 R
V8H1 1 / 171 1 UNSIGNED8 R / W
V8H2 1 / 172 1 UNSIGNED8 R / W
V8H3 1 / 173 2 UNSIGNED16 R / W
V8H4 1 / 174 1 UNSIGNED8 R / W
V8H5 1 / 175 1 UNSIGNED8 R
V8H6 1 / 176 1 UNSIGNED8 R / W
V8H8 1 / 177 1 UNSIGNED8 R / W
V8H9 1 / 178 2 UNSIGNED16 R / W
VAH5 1 / 179 1 UNSIGNED8 R / W
VAH6 1 / 180 2 UNSIGNED16 R / W
VAH7 1 / 181 1 UNSIGNED8 R
V9H7 1 / 182 1 UNSIGNED8 R
V9H8 1 / 183 1 UNSIGNED8 R / W
V9H3 1 / 184 2 UNSIGNED16 R
1 / 185 1 UNSIGNED8 R
V1H6 1 / 186
4
4
4
4
4 FLOAT R
Reference temperature V1H3 1 / 187 R / W
R / WVAH3
V4H9
1 / 188 11 OSTRING
UNSIGNED81 / 189
2
Serial number
Basic load
Controller type
UNSIGNED8V9H6 1 / 206
1
1
R / W
R1 / 199 11
V9H0 1 / 200 1 R / W
V9H1 1 / 201
UNSIGNED8
UNSIGNED8
R / W
V9H2 1 / 202 2
2 UNSIGNED16
UNSIGNED16 R / W
V9H3 1 / 203 1
1
UNSIGNED8
UNSIGNED8
R / W
V9H4
V9H5
1 / 204 R / WR / W1 / 205 1 UNSIGNED8
Alarm threshold monitoring
Process monitoring
Alarm delay
Max. period lower limit exceeded
Max. period upper limit exceeded
Lower alarm threshold
Upper alarm threshold
Limit of process monitoring
V7H1
V7H5
1 / 190
1 / 194
1
1
UNSIGNED8
UNSIGNED8
UNSIGNED8
V7H2
V7H6
1 / 191
1 / 195
R / W
R / W
R / W
R / W
R / W
R / WR / W
R / W
R / W
V7H3
V7H4
V7H8
V7H7
1 / 192
1 / 196
22
FLOAT
FLOAT
FLOAT
FLOAT
UNSIGNED16
UNSIGNED16
UNSIGNED16
R
1 / 193
1 / 1971 / 198
Active concentration curve
Edit table (conc)
Table option (conc)
Number of value pairs (conc)
Selection of value pair (conc)
Uncomp. cd. value (conc)
Concentration value (conc)
Temperature value (conc)
Table status (conc)
Table option (alpha)
Number of table value pairs (alpha)
Selection of table value pair (alpha)
Temperature value (alpha)
Temperature coefficient (alpha)
Table status (alpha)
Polarisation detection (cond)
Limit contact status
Alarm contact type
Unit of alarm delay
Alarm delay
Error number selection
Error status
Alarm contact effective
Cleaning trigger
Release code
Hold configuration
Hold dwell period
Plus package released
Chemoclean released
Reset of instrument
Software version
Number of relays
Installation factor
Current input off delay
Current input on delay
Current input switch-off direction
Current input feedforward control
Current input gain
Current input available
Current input switch-off point
Set current input active
Tab. 5.10Slot / index list forconductivity (continued)
XM53PE5.CHP
Liquisys M PROFIBUS-PA/-DP Communication
Endress+Hauser 19
Slot / index list for turbidity
Parameter Matrix(Standard)
Slot /Index
Sizein byte(record)
Type read/write
Manufacturer specific Block
Block Object 1 / 100 20 DS-32 R
Main measured value V0H0 1 / 108 4 FLOAT R
Measured temperature value V0H1 1 / 109 4 FLOAT R
Operating mode V0H2 1 / 110 1 UNSIGNED8 R / W
Display unit V0H3 1 / 111 1 UNSIGNED8 R / W
Display format V0H4 1 / 112 1 UNSIGNED8 R / W
Sensor type V0H5 1 / 113 1 UNSIGNED8 R / W
Measured value damping V0H6 1 / 114 1 UNSIGNED8 R / W
Measured value in FNU V0H7 1 / 115 4 FLOAT R
Current measured value / offset V0H8 1 / 116 4 FLOAT R / W
Calculated offset V0H9 1 / 117 4 FLOAT R
Wiper on/off V1H0 1 / 118 1 UNSIGNED8 R / W
Wiping time V1H1 1 / 119 2 UNSIGNED16 R / W
Pause time V1H2 1 / 120 2 UNSIGNED16 R / W
Calibration data set V1H3 1 / 121 1 UNSIGNED8 R / W
Copy calibration data set V1H4 1 / 122 1 UNSIGNED8 R / W
Installation adaption V1H5 1 / 123 1 UNSIGNED8 R / W
Entry of process temperature V1H8 1 / 124 4 FLOAT R / W
Entry of temperature offset V1H9 1 / 125 4 FLOAT R / W
Beg. of meas. (bar graph) – 1 / 126 4 FLOAT R
End of meas. (bar graph) – 1 / 127 4 FLOAT R
Contact selection V3H0 1 / 128 1 UNSIGNED8 R / W
Contact configuration V3H1 1 / 129 1 UNSIGNED8 R / W
Switch contact on or off V3H2 1 / 130 1 UNSIGNED8 R / W
Switch-on point (LC) V4H0 1 / 131 4 FLOAT R / W
Switch-off point (LC) V4H1 1 / 132 4 FLOAT R / W
Pickup delay (LC) V4H2 1 / 133 2 UNSIGNED16 R / W
Dropout delay (LC) V4H3 1 / 134 2 UNSIGNED16 R / W
Alarm threshold (LC) V4H4 1 / 135 4 FLOAT R / W
Setpoint (PID / neutra) V4H0 1 / 136 4 FLOAT R / W
Control gain Kp (PID / neutra) V4H1 1 / 137 4 FLOAT R / W
Integral action time Tn (PID / neutra) V4H2 1 / 138 4 FLOAT R / W
Derivative action time Tv (PID / neutra) V4H3 1 / 139 4 FLOAT R / W
Control characteristic (PID) V4H4 1 / 140 1 UNSIGNED8 R / W
Operating mode of controller (PID / neutra) V4H5 1 / 141 1 UNSIGNED8 R / W
Pulse period (PID / neutra) V4H6 1 / 142 4 FLOAT R / W
Pulse frequency (PID / neutra) V4H7 1 / 143 2 UNSIGNED16 R / W
Minimum ON time (PID / neutra) V4H8 1 / 144 4 FLOAT R / W
Rinse time (timer) V4H0 1 / 145 2 UNSIGNED16 R / W
Pause time (timer) V4H1 1 / 146 2 UNSIGNED16 R / W
Minimum pause time (timer) V4H2 1 / 147 2 UNSIGNED16 R / WTab. 5.11 Slot / index list for turbidity
Communication Liquisys M PROFIBUS-PA/-DP
20 Endress+Hauser
Slot / index list for turbidity (continued)
Parameter Matrix(Standard)
Slot /Index
Sizein byte(record)
Type read/write
V4H0 1 / 148 1 UNSIGNED8 R / W
V4H1 1 / 149 2 UNSIGNED16 R / W
V4H2 1 / 150 2 UNSIGNED16 R / W
V4H3 1 / 151 2 UNSIGNED16 R / W
V4H4 1 / 152 1 UNSIGNED8 R / W
V4H5 1 / 153 2 UNSIGNED16 R / W
V4H6 1 / 154 2 UNSIGNED16 R / W
V4H7 1 / 155 1 UNSIGNED8 R / W
V5H0 1 / 156 1 UNSIGNED8 R / W
V5H1 1 / 157 1 UNSIGNED8 R / W
V5H2 1 / 158 1 UNSIGNED8 R / W
V5H3 1 / 159 1 UNSIGNED8 R / W
V5H4 1 / 160 1 UNSIGNED8 R / W
V5H5 1 / 161 4 FLOAT R / W
V5H6 1 / 162 4 FLOAT R / W
V5H7 1 / 163 1 UNSIGNED8 R / W
V1H6 1 / 164 2 UNSIGNED8 R / W
V8H1 1 / 165 1 UNSIGNED8 R / W
V8H2 1 / 166 1 UNSIGNED8 R / W
V8H3 1 / 167 2 UNSIGNED16 R / W
V8H4 1 / 168 1 UNSIGNED8 R / W
V8H5 1 / 169 1 UNSIGNED8 R
V8H6 1 / 170 1 UNSIGNED8 R / W
V8H8 1 / 171 1 UNSIGNED8 R / W
V8H9 1 / 172 2 UNSIGNED16 R / W
V9H0 1 / 173 1 UNSIGNED8 R / W
V9H1 1 / 174 2 UNSIGNED16 R / W
V9H2 1 / 175 1 UNSIGNED8 R
V9H3 1 / 176 1 UNSIGNED8 R
V9H4 1 / 177 1 UNSIGNED8 R / W
VAH2 1 / 178 2 UNSIGNED16 R
1 / 179 1 UNSIGNED8
UNSIGNED8
R
R1
4
4
4
Limit contact status V1H3 1 / 180
R / WVAH3
V4H9
1 / 188 11 OSTRING
UNSIGNED81 / 189
2
UNSIGNED8V9H6 1 / 215
1
1
R / W
R1 / 208 1
1
V9H0 1 / 209 1 R / W
V9H1 1 / 210
UNSIGNED8
UNSIGNED8
R / W
V9H2 1 / 211 2
2 UNSIGNED16
UNSIGNED16 R / W
V9H3 1 / 212 1
1
UNSIGNED8
UNSIGNED8
R / W
V9H4
V9H5
1 / 213 R / WR / W1 / 214 1 UNSIGNED8
V7H1
V7H5
1 / 199
1 / 203
1
1
UNSIGNED8
UNSIGNED8
UNSIGNED8
V7H2
V7H6
1 / 200
1 / 204
R / W
R / W
R / W
R / W
R / W
R / W
R / W
R / W
R / W
V7H3
V7H4
V7H8
V7H7
1 / 201
1 / 205
22
FLOAT
FLOAT
FLOAT
UNSIGNED16
UNSIGNED16
UNSIGNED16
R
1 / 202
1 / 206
1 / 198
Serial number
Basic load
Controller type
Alarm threshold monitoring
Process monitoring
Alarm delay
Max. period lower limit exceeded
Max. period upper limit exceeded
Lower alarm threshold
Upper alarm threshold
Limit of process monitoring
Current input off delay
Current input on delay
Current input switch-off direction
Current input feedforward control
Current input gain
Current input available
Current input switch-off point
Set current input active
Start pulse (cleaning)
Pre-rinse time (cleaning)
Cleaning time (cleaning)
Post-rinse time (cleaning)
Repeat cycles (cleaning)
Pause time (cleaning)
Minimum pause time (cleaning)
Economy (cleaning)
Active concentration curve
Edit table (conc)
Table option (conc)
Number of value pairs (conc)
Selection of value pair (conc)
Entry of turbidity value (conc)
Concentration value (conc)
Table status (conc)
Gas bubble barrier
Alarm contact type
Unit of alarm delay
Alarm delay
Error number selection
Error status
Alarm contact effective
Cleaning trigger
Release code
Hold configuration
Hold dwell period
Plus package released
Chemoclean released
Reset of instrument
Software version
Number of relays
Tab. 5.12Slot / index list for turbidity(continued)
XM53PE5.CHP
Liquisys M PROFIBUS-PA/-DP Communication
Endress+Hauser 21
Slot / index list for oxygen
Parameter Matrix(Standard)
Slot /Index
Sizein byte(record)
Type read/write
Manufacturer specific Block
Block Object 1 / 100 20 DS-32 R
Main measured value V0H0 1 / 108 4 FLOAT R
Measured temperature value V0H1 1 / 109 4 FLOAT R
Operating mode V0H2 1 / 110 1 UNSIGNED8 R / W
Pressure compensation V0H4 1 / 111 1 UNSIGNED8 R / W
Altitude V0H5 1 / 112 2 UNSIGNED16 R / W
Measured value damping V0H7 1 / 113 1 UNSIGNED8 R / W
Sensor current in nA V0H8 1 / 114 4 FLOAT R
Salinity V1H0 1 / 115 4 FLOAT R / W
Entry of process temperature V1H8 1 / 116 4 FLOAT R / W
Entry of temperature offset V1H9 1 / 117 4 FLOAT R / W
Beg. of meas. (bar graph) – 1 / 118 4 FLOAT R
End of meas. (bar graph) – 1 / 119 4 FLOAT R
Contact selection V3H0 1 / 120 1 UNSIGNED8 R / W
Contact configuration V3H1 1 / 121 1 UNSIGNED8 R / W
Switch contact on or off V3H2 1 / 122 1 UNSIGNED8 R / W
Switch-on point (LC) V4H0 1 / 123 4 FLOAT R / W
Switch-off (LC) V4H1 1 / 124 4 FLOAT R / W
Pickup delay (LC) V4H2 1 / 125 2 UNSIGNED16 R / W
dropout delay (LC) V4H3 1 / 126 2 UNSIGNED16 R / W
Alarm threshold (LC) V4H4 1 / 127 4 FLOAT R / W
Setpoint (PID / neutra) V4H0 1 / 128 4 FLOAT R / W
Control gain Kp (PID / neutra) V4H1 1 / 129 4 FLOAT R / W
Integral action time Tn (PID / neutra) V4H2 1 / 130 4 FLOAT R / W
Derivative action time Tv (PID / neutra) V4H3 1 / 131 4 FLOAT R / W
Control characteristic (PID) V4H4 1 / 132 1 UNSIGNED8 R / W
Operating mode of controller (PID / neutra) V4H5 1 / 133 1 UNSIGNED8 R / W
Pulse period (PID / neutra) V4H6 1 / 134 4 FLOAT R / W
Pulse frequency (PID / neutra) V4H7 1 / 135 2 UNSIGNED16 R / W
Minimum ON time (PID / neutra) V4H8 1 / 136 4 FLOAT R / W
Rinse time (timer) V4H0 1 / 137 2 UNSIGNED16 R / W
Pause time (timer) V4H1 1 / 138 2 UNSIGNED16 R / W
Minimum pause time (timer) V4H2 1 / 139 2 UNSIGNED16 R / W
Start pulse (cleaning) V4H0 1 / 140 1 UNSIGNED8 R / W
Pre-rinse time (cleaning) V4H1 1 / 141 2 UNSIGNED16 R / W
Cleaning time (cleaning) V4H2 1 / 142 2 UNSIGNED16 R / W
Post-rinse time (cleaning) V4H3 1 / 143 2 UNSIGNED16 R / W
Repeat cycles (cleaning) V4H4 1 / 144 1 UNSIGNED8 R / W
Pause time (cleaning) V4H5 1 / 145 2 UNSIGNED16 R / W
Minimum pause time (cleaning) V4H6 1 / 146 2 UNSIGNED16 R / W
Economy (cleaning) V4H7 1 / 147 1 UNSIGNED8 R / WTab. 5.13 Slot / index list for oxygen
Communication Liquisys M PROFIBUS-PA/-DP
22 Endress+Hauser
Slot / index list for oxygen (continued)
Parameter Matrix(Standard)
Slot /Index
Sizein byte(record)
Type read/write
V7H0 1 / 148 1 UNSIGNED8 R / W
V7H1 1 / 149 2 UNSIGNED16 R / W
V7H2 1 / 150 4 FLOAT R / W
V7H3 1 / 151 4 FLOAT R / W
V7H4 1 / 152 1 UNSIGNED8 R / W
V7H5 1 / 153 2 UNSIGNED16 R / W
V7H6 1 / 154 2 UNSIGNED16 R / W
V7H7 1 / 155 4 FLOAT R / W
V8H1 1 / 156 1 UNSIGNED8 R / W
V8H2 1 / 157 1 UNSIGNED8 R / W
V8H3 1 / 158 2 UNSIGNED16 R / W
V8H4 1 / 159 1 UNSIGNED8 R / W
V8H5 1 / 160 1 UNSIGNED8 R
V8H6 1 / 161 1 UNSIGNED8 R / W
V8H8 1 / 162 1 UNSIGNED8 R / W
V8H9 1 / 163 2 UNSIGNED16 R / W
V9H0 1 / 164 1 UNSIGNED8 R / W
V9H1 1 / 165 2 UNSIGNED16 R / W
V9H2 1 / 166 1 UNSIGNED8 R
V9H3 1 / 167 1 UNSIGNED8 R
V9H4 1 / 168 1 UNSIGNED8 R / W
V9H3
–
–
–
–
1 / 169 2 UNSIGNED16 R
1 / 170 1 UNSIGNED8
UNSIGNED8
R
Range switching V0H6 1 / 171 1
Pressure measurement available 1 / 172 1 UNSIGNED8
UNSIGNED8
R
R1Limit contact status V4H5 1 / 173
R / WVAH3
V4H9
V0H3
1 / 188 11 OSTRING
UNSIGNED81 / 189
Serial number
Basic load
Controller type
Unit of measured value
UNSIGNED8
UNSIGNED8
V9H6 1 / 205
1 / 205
1
R / W
R
R1 / 199 1
1
1
V9H0 1 R / W
R / W
V9H1 1 / 200UNSIGNED8
R / W
V9H2 1 / 201 2
2 UNSIGNED16
UNSIGNED16 R / W
V9H3 1 / 20 1
1
UNSIGNED8
UNSIGNED8
R / W
V9H4
V9H5
1 / 203 R / WR / W1 / 204 1 UNSIGNED8
1
1
UNSIGNED8
UNSIGNED8
R / W
R / W
R
1 / 198
1 / 190
Alarm threshold monitoring
Alarm delay
Lower alarm threshold
Upper alarm threshold
Process monitoring
Limit value of process monitoring
Alarm contact type
Unit of alarm delay
Alarm delay
Error number selection
Error status
Alarm contact effective
Cleaning trigger
Release code
Hold configuration
Hold dwell period
Plus package released
Chemoclean released
Reset of instrument
Software version
Number of relays
Current input off delay
Current input on delay
Current input switch-off direction
Current input feedforward control
Current input gain
Current input available
Current input switch-off point
Set current input active
Max. period lower limit exceeded
Max. period upper limit exceeded
Tab. 5.14Slot / index list for oxygen(continued)
XM53PE5.CHP
Liquisys M PROFIBUS-PA/-DP Communication
Endress+Hauser 23
Slot / index list for chlorine
Parameter Matrix(Standard)
Slot /Index
Sizein byte(record)
Type read/write
1 / 100 20 DS-32 R
V0H0 1 / 108 4 FLOAT R
Measured value pH / ORPV0H1
V0H2
V0H3
1 / 109
1 / 110
4
4
FLOAT
FLOAT
R
R
Controller switch-on delay
Controller switch-off delay
Controller switch-off
Type of sensor
V0H5 1 / 112
1 / 111
1
1
UNSIGNED8
UNSIGNED8
R / W
R / W
V0H6 1 / 113 2 UNSIGNED16 R / W
V0H7 1 / 114 2 UNSIGNED16 R / W
Selection of digital input 1 V0H8 1 / 115 1 UNSIGNED8
UNSIGNED8
UNSIGNED8
UNSIGNED8
R / W
Messwertdämpfung V0H9 1 / 116 1 R
Operating mode pH / ORP V1H0 1 / 117 1 R / W
pH compensation
Entry of process temperature
V1H8
V1H8
1 / 118
1 / 120
1
4 FLOAT
R / W
R / W
Manual pH compensation
Entry of temperature offset
V1H2
V1H9
1 / 119
1 / 121
4
4
FLOAT
FLOAT
R / W
R / W
V3H0 1 / 122 1 UNSIGNED8 R / W
V3H1 1 / 123 1 UNSIGNED8 R / W
V3H2 1 / 124 1 UNSIGNED8 R / W
V4H0 1 / 125 4 FLOAT R / W
V4H1 1 / 126 4 FLOAT R / W
V4H2 1 / 127 2 UNSIGNED16 R / W
V4H3 1 / 128 2 UNSIGNED16 R / W
V4H4 1 / 129 4 FLOAT R / W
V4H0 1 / 130 4 FLOAT R / W
V4H1 1 / 131 4 FLOAT R / W
V4H2 1 / 132 4 FLOAT R / W
V4H3 1 / 133 4 FLOAT R / W
V4H4 1 / 134 1 UNSIGNED8 R / W
V4H5 1 / 135 1 UNSIGNED8 R / W
V4H6 1 / 136 4 FLOAT R / W
V4H7 1 / 137 2 UNSIGNED16 R / W
V4H8 1 / 138 4 FLOAT R / W
V4H0 1 / 139 2 UNSIGNED16 R / W
V4H1 1 / 140 2 UNSIGNED16 R / W
V4H2 1 / 141 2 UNSIGNED16 R / W
V4H0 1 / 142 1 UNSIGNED8 R / W
V4H1 1 / 143 2 UNSIGNED16 R / W
V4H2 1 / 144 2 UNSIGNED16 R / W
V4H3 1 / 145 2 UNSIGNED16 R / W
V4H4 1 / 146 1 UNSIGNED8 R / W
V4H5 1 / 147 2 UNSIGNED16 R / W
V4H6 1 / 148 2 UNSIGNED16 R / W
V4H7 1 / 149 1 UNSIGNED8 R / W
Switch-on point (LC)
Switch-off (LC)
Pickup delay (LC)
Dropout delay (LC)
Alarm threshold (LC)
Setpoint (PID / neutra)
Control gain Kp (PID / neutra)
Integral action time Tn (PID / neutra)
Derivative action time Tv (PID / neutra)
Control characteristic (PID)
Operating mode of controller (PID / neutra)
Pulse period (PID / neutra)
Pulse frequency (PID / neutra)
Minimum ON time (PID / neutra)
Rinse time (timer)
Pause time (timer)
Minimum pause time (timer)
Start pulse (cleaning)
Pre-rinse time (cleaning)
Post-rinse time (cleaning)
Repeat cycles (cleaning)
Pause time (cleaning)
Minimum pause time (cleaning)
Economy (cleaning)
Cleaning time (cleaning)
Contact selection
Contact configuration
Switch contact on or off
Manufacturer specific Block
Block Object
Main measured value
Measured temperature value
Tab. 5.15 Slot / index list for chlorine
Communication Liquisys M PROFIBUS-PA/-DP
24 Endress+Hauser
Slot / index list for chlorine (continued)
Parameter Matrix(Standard)
Slot /Index
Sizein byte(record)
Type read/write
Selection of monitoring parameters
V7H1
V7H0
1 / 151
1 / 150
1
1
UNSIGNED8
UNSIGNED8
R / W
R / W
V7H2 1 / 152 2 UNSIGNED16 R / W
V7H3 1 / 153 4 FLOAT R / W
V7H4 1 / 154 4 FLOAT R / W
V7H5 1 / 155 1 UNSIGNED8 R / W
V7H6 1 / 156 2 UNSIGNED16 R / W
V7H7 1 / 157 2 UNSIGNED16 R / W
V7H8 1 / 158 4 FLOAT R / W
V8H1 1 / 159 1 UNSIGNED8 R / W
V8H2 1 / 160 1 UNSIGNED8 R / W
V8H3 1 / 161 2 UNSIGNED16 R / W
V8H4 1 / 162 1 UNSIGNED8 R / W
V8H5 1 / 163 1 UNSIGNED8 R
V8H6 1 / 164 1 UNSIGNED8 R / W
V8H8 1 / 165 1 UNSIGNED8 R / W
V8H9 1 / 166 2 UNSIGNED16 R / W
V9H0 1 / 167 1 UNSIGNED8 R / W
V9H1 1 / 168 2 UNSIGNED16 R / W
V9H2 1 / 169 1 UNSIGNED8 R
V9H3 1 / 170 1 UNSIGNED8 R
V9H4 1 / 171 1 UNSIGNED8 R / W
Hardware version
VAH2
–
–
1 / 172
1 / 173
2
2
2
UNSIGNED16 R
R
1 / 174 1
1
UNSIGNED8
UNSIGNED8
R
Motor run time (3-pt. step controller)
Neutral zone (3-pt. step controller)
V4H5
V4H6
1 / 176
1 / 177
UNSIGNED8
R
V4H5
R / W
VAH3
V4H9
V0H4V5H9/V4H8
OSTRING
UNSIGNED8
Display unit
UNSIGNED8V5H6 1 / 184
1 / 185
1 / 188
1 / 189
1 / 190
1 / 198
R / W
1 / 178
1
1
V5H0 1 R / W
R / W
R / W
V5H1 1 / 179UNSIGNED8
R / WR / W
R / W
R / W
R / W
V5H2 1 / 180 2
2 UNSIGNED16
UNSIGNED16
UNSIGNED16UNSIGNED16
V5H3 1 / 181 1 UNSIGNED8
UNSIGNED8
R / W
V5H4
V5H5
1 / 182R / W1 / 183 1 UNSIGNED8
1
11
1
1
14
UNSIGNED8
UNSIGNED8
UNSIGNED8
R / W
R
1 / 175
Limit contact status
Serial number
Basic load
Controller type
Alarm threshold monitoring
Alarm delay
Lower alarm threshold
Upper alarm threshold
Process monitoring
Limit value of process monitoring
Alarm contact type
Unit of alarm delay
Alarm delay
Error number selection
Error status
Alarm contact effective
Cleaning trigger
Release code
Hold configuration
Hold dwell period
Plus package released
Chemoclean released
Reset of instrumentSoftware version
Number of relays
Current input off delay
Current input on delay
Current input switch-off direction
Current input feedforward control
Current input gain
Current input available
Current input switch-off point
Set current input active
Max. period lower limit exceeded
Max. period upper limit exceeded
Tab. 5.16Slot / index list for chlorine(continued)
XM53PE5.CHP
Liquisys M PROFIBUS-PA/-DP Communication
Endress+Hauser 25
5.3 Operation via Commuwin II
PROFIBUS devices can be operated usingthe Commuwin II operating program. Theconnection to Commuwin II is established viathe PROFIBUS server. The device list showsall devices connected to the selected bussegments. The profile parameters areconfigured either via the operating matrix orvia the graphical user interface.
The operating structures of the variousmeasuring parameters are shown on thenext pages.
Note:
• Operation via Commuwin II isdescribed in the operatinginstructions BA 124F/00/en.
• If the off-line parameterisationdoes not correspond with theversion of the device(e.g., no release code for pluspackage), then error 03 will be setafter completion of the download,and the transmitter will not switchto the normal operating status.In this case, please repeat thedownload with correct data, orperform a device reset (VH 94).
Example of a Commuwin II matrix (e.g. CLM 2X3)
Communication Liquisys M PROFIBUS-PA/-DP
26 Endress+Hauser
6 Accessories
Fieldbus connection box for PAFieldbus connection box for direct mountingto the instrument. Aluminium housing, ingressprotection IP 67. With 4-pole connector anda bus termination via a jumper. 2 cableglands Pg 9.Order no.: 017 481-0130
Fieldbus connection box for PAFieldbus connection box for direct mountingto the instrument. Aluminium housing, ingressprotection IP 67. With 4-pole connector anda bus termination via a jumper. 2 cableglands Pg 9. Internal ground condensator.Order no.: 017 481-0110
Fieldbus M12 instrument plug for PA4-pole metal plug for mounting to theinstrument for connection box or cable socketconnection. Cable length: 150 mm.Order no.: 51502184
Fieldbus Y adapterMetal Y adapter with 2 cable glands Pg 13.5.Order no.: 51502183
E+H
PROFIBUS-PA
PROFIBUS-DP
ENDRESS+HAUSERLIQUISYS SENDRESS+HAUSERLIQUISYS S
ENDRESS+HAUSERLIQUISYS
CAL REL
REL1
REL1
ALARMREL2
REL2E
ENDRESS+HAUSERLIQUISYS
CAL REL
REL1
REL1
ALARMREL2
REL2E
ENDRESS+HAUSERLIQUISYS
CAL REL
REL1
REL1
ALARMREL2
REL2E
PLCconfiguration
tool Device data file
PLC
Segment coupler
CXM 223separatemounting
CXM 253separatemountingCXM 253
directmounting
Fieldbus connection box
M12 plug
Bus termination box
XM253-07.CDR
CXM 253separatemounting
Fig. 6.1Instrumentation overviewLiquisys M CXM 2X3
n.c.
A (PA-)
B (PA+)
S
C2
58
64
18.5
17.5
18.5
37
n.c. A (PA-)
SB (PA+)
A (PA-)SB (PA+)
45 150/300 7
62
Pg 13.5
65
Pg 13.5
XM253D07.CHP
Liquisys M PROFIBUS-PA/-DP Accessories
Endress+Hauser 27
Device data file• via the Internet:
E+H: http://www.endress.comProduct Avenue → Download Street →Field Communication Street)PNO: http://www.profibus.com(GSD library)
• Floppy disk by E+H:Order no.: 943157-0000
Commuwin IIGraphical operation software for allPROFIBUS instruments. Connection via theDDE server. Supported systems are Windows3.1/3.11, WIN 95 and WIN NT 4.0.Order no.: FXS 113-A7B
• 16 bit ISA plug-in board for PCOrder no.: 016 570-5300
• PCMCIA plug-in board for notebooksOrder no.: 016 570-5200
Cable for PATerminated cable with M12 plug andM12 coupling made of polyurethane andnickel-plated brass joints. Ingress protectionIP 67. Screen is connected to the threadedjoint. PVC sheath. Drilled and screenedconductor pair, 2/18 AWG.Temperature range – 40... +70°C.Lengths 1, 2, 5, 10 meter.
Cable type Order no.Cable terminated, yellow, 1 m 52001025
Cable terminated, yellow, 2 m 52001040
Cable terminated, yellow, 5 m 52001041
Cable terminated, yellow, 10 m 52001042
For further accessories, please refer to theSpecial Documentation SD 096F/00/en,available from the Endress+Hauser salesagencies (see back cover of these operatinginstructions).
Accessories Liquisys M PROFIBUS-PA/-DP
28 Endress+Hauser
7 Technical data
General specifications Manufacturer Endress+Hauser
Product designation Liquisys M CXM 223/253-xx3
Output PROFIBUS-PA Output signal Digital communication signal,PROFIBUS-PA
PA function Slave
Transfer rate 31.25 kbit/s
Signal coding Manchester II
Response time of slave approx. 20 ms
Failure signal Signal: status bit is set, last valid measuredvalue is maintainedDisplay: error code
Physical layer IEC 1158-2
Bus voltage 9 ... 32 V
Current consumption of bus 10 mA ± 1 mA
Output PROFIBUS-DP Output signal PROFIBUS-DP acc. to EN 50 170 Part 3
DP function Slave
Transfer rate 9.6 kBit/s, 19.2 kBit/s, 93.75 kBit/s,187.5 kBit/s, 500 kBit/s, 1.5 MBit/s,3 MBit/s, 6 MBit/s, 12 MBit/s
Signal coding NRZ code
Failure signal Signal: status bit is set, last valid measuredvalue is maintainedDisplay: error code
Physical layer RS 485
Display and user interface Field operation via +, –, E, CAL, REL, AUTO keys
PC operation via PROFIBUS with Commuwin IIoperating program
Bus address Setting via DIL switch orvia operating menu orvia Set Slave Adr service
Communication interface PROFIBUS-PA/-DP
Standards and guidelines PROFIBUS EN 50 170, part 2; DIN 19 245, parts 1-3
PROFIBUS-DP EN 50 170, part 3; DIN 19 245, part 3PNO guidelines for PROFIBUS-DP
PROFIBUS-PA EN 50 170, part 4; DIN 19 245, part 4PNO guidelines for PROFIBUS-PA
Physical layer EN 61 158-2; IEC 1158-2, RS 485
Subject to modifications.
% Ω°C% Ω°C
XM53PE8.CHP
Liquisys M PROFIBUS-PA/-DP Technical data
Endress+Hauser 29
8 Index
AAccessories . . . . . . . . . . . . . . . . . . . . . . 27-28Acyclical data exchange . . . . . . . . . . . . . . . 14Application . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
BBus cable connection to field instrument . . . . 5Bus cable connection to panel-mountedinstrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
CCommunication. . . . . . . . . . . . . . . . . . . . 12-26Commuwin II . . . . . . . . . . . . . . . . . . . . . . . . . 26Configuration. . . . . . . . . . . . . . . . . . . . . . . . . 13Configuration software . . . . . . . . . . . . . . . . . 11Cyclical data exchange . . . . . . . . . . . . . . . . 12
DData block. . . . . . . . . . . . . . . . . . . . . . . . . . . 13Data_Exchange . . . . . . . . . . . . . . . . . . . . . . 12Device address. . . . . . . . . . . . . . . . . . . . . . . . 9Device data file . . . . . . . . . . . . . . . . . . . . 10-11
EElectrical connection . . . . . . . . . . . . . . . . . 5-8
FFloating point number. . . . . . . . . . . . . . . . . . 12
GGeneral . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2General safety instructions . . . . . . . . . . . . . . . 3GSD. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-11
HHazards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
IInstallation . . . . . . . . . . . . . . . . . . . . . . . . . 4-8Intended use . . . . . . . . . . . . . . . . . . . . . . . . . 3
MMaximum configuration . . . . . . . . . . . . . . . . 12
PPhysical block . . . . . . . . . . . . . . . . . . . . . . . 14Physical layer . . . . . . . . . . . . . . . . . . . . . . . . 29PLC. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11PNO . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10, 12PROFIBUS user organization. . . . . . . . . . . . 10Programmable logic controller . . . . . 4, 12-13
SSafety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Safety instructions . . . . . . . . . . . . . . . . . . . . . 3Segment coupler . . . . . . . . . . . . . . . . . . . . . . 4Slot / index list . . . . . . . . . . . . . . . . . . . . 15-25Start-up. . . . . . . . . . . . . . . . . . . . . . . . . . . 9-11Status codes . . . . . . . . . . . . . . . . . . . . . . . . 13Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2Symbols used. . . . . . . . . . . . . . . . . . . . . . . . . 2System equipment . . . . . . . . . . . . . . . . . . . . . 4
TTechnical data . . . . . . . . . . . . . . . . . . . . . . . 29Terminating resistor . . . . . . . . . . . . . . . . . . . . 4Transmission method. . . . . . . . . . . . . . . . . . . 3Type file . . . . . . . . . . . . . . . . . . . . . . . . . 10-11
Index Liquisys M PROFIBUS-PA/-DP
30 Endress+Hauser
Europe
Austria Endress+Hauser Ges.m.b.H.WienTel. ++43 (1) 88056-0, Fax (1) 8 8056-35
BelarusBelorgsintezMinskTel. ++3 75 (1 72) 26 3166, Fax (172) 2631 11
Belgium / Luxembourg Endress+Hauser S.A./N.V.BrusselsTel. ++32 (2) 24806 00, Fax (2) 2 4805 53
BulgariaINTERTECH-AUTOMATIONSofiaTel. ++3 59 (2) 6648 69, Fax (2) 9 6313 89
Croatia Endress+Hauser GmbH+Co.ZagrebTel. ++3 85 (1) 6 63 7785, Fax (1) 663 7823
CyprusI+G Electrical Services Co. Ltd.NicosiaTel. ++3 57 (2) 4847 88, Fax (2) 48 4690
Czech Republic Endress+Hauser GmbH+Co.PrahaTel. ++4 20 (26) 678 4200, Fax (26) 67841 79
Denmark Endress+Hauser A/SSøborgTel. ++45 (70) 13 1132, Fax (70) 13 2133
EstoniaElvi-AquaTartuTel. ++3 72 (7) 4227 26, Fax (7) 42 2727
Finland Endress+Hauser OyEspooTel. ++3 58 (9) 8 59 6155, Fax (9) 859 6055
France Endress+HauserHuningueTel. ++33 (3) 89 6967 68, Fax (3) 89 69 4802
Germany Endress+Hauser Meßtechnik GmbH+Co.Weil am RheinTel. ++49 (7621) 975 01, Fax (76 21) 9 755 55
Great Britain Endress+Hauser Ltd.ManchesterTel. ++44 (161) 286 5000,Fax (161) 998 18 41
GreeceI & G Building Services Automation S.A.AthensTel. ++30 (1) 92415 00, Fax (1) 9 2217 14
HungaryMile Ipari-ElektroBudapestTel. ++36 (1) 26155 35, Fax (1) 2 6155 35
IcelandVatnshreinsun HFReykjavikTel. ++3 54 (5) 6196 16, Fax (5) 61 9617
IrelandFlomeaco Company Ltd.KildareTel. ++3 53 (45) 8686 15, Fax (45) 8681 82
Italy Endress+Hauser Italia S.p.A.Cernusco s/N MilanoTel. ++39 (02) 92 106421,Fax (02) 92 10 7153
LatviaRaita Ltd.RigaTel. ++3 71 (7) 3128 97, Fax (7) 31 2894
LithuaniaAgava Ltd.KaunasTel. ++3 70 (7) 2024 10, Fax (7) 20 7414
Netherlands Endress+Hauser B.V.NaardenTel. ++31 (35) 6 9586 11, Fax (35) 6 95 8825
Norway Endress+Hauser A/STranbyTel. ++47 (32) 85 98 50, Fax (32) 85 9851
Poland Endress+Hauser Polska Sp. z o.o.WarszawyTel. ++48 (22) 7 201090, Fax (22) 720 1085
PortugalTecnisis - Tecnica de Sistemas IndustriaisLinda-a-VelhaTel. ++351 (1) 417 2637, Fax (1) 418 52 78
RomaniaRomconseng SRLBucharestTel. ++40 (1) 4 1016 34, Fax (1) 4 101634
Russia Endress+Hauser Moscow OfficeMoscowTel.++7 09 (5) 1 5875 71, Fax (5) 1 589864
Slovak RepublicTranscom Technik s.r.o.BratislavaTel. ++421 (74) 488 86 84, Fax (74) 4 8871 12
Slovenia Endress+Hauser D.O.O.LjubljanaTel. ++386 (61) 159 22 17, Fax (61) 1 5922 98
Spain Endress+Hauser S.A.BarcelonaTel. ++34 (93) 4 803366, Fax (93) 473 3839
Sweden Endress+Hauser ABSollentunaTel. ++46 (8) 55 511600, Fax (8) 5551 1600
Switzerland Endress+Hauser AGReinach/BL 1Tel. ++41 (61) 7 157575, Fax (61) 711 1650
TurkeyIntek Endüstriyel Ölcü ve Kontrol SistemleriIstanbulTel. ++90 (212) 275 13 55,Fax (2 12) 26627 75
UkraineIndustria UkraïnaKievTel. ++380 (44) 268 81, Fax (44) 2 69 08
YugoslaviaMeris d.o.o.BeogradTel. ++381 (11) 444 61 64, Fax (11) 4 4419 66
Africa
EgyptAnasiaHeliopolis/CairoTel. ++20 (2) 4 179 00, Fax (2) 4 179 00
MoroccoOussama S.A.CasablancaTel. ++212 (2) 2413 38, Fax (2) 40 26 57
NigeriaJ F Technical Invest. Nig. Ltd.LagosTel. ++234 (1) 6223 45 46, Fax (1) 62 2345 48
South Africa Endress+Hauser Pty. Ltd.SandtonTel. ++27 (11) 4 441386, Fax (11) 444 1977
TunisiaControle, Maintenance et RegulationTunisTel. ++216 (1) 7930 77, Fax (1) 78 85 95
America
Argentina Endress+Hauser Argentina S.A.Buenos AiresTel. ++54 (1) 1 4522 79 70,Fax (1) 1 452279 09
BoliviaTritec S.R.L.CochabambaTel. ++591 (42) 569 93, Fax (42) 5 09 81
Brazil Samson Endress+Hauser Ltda.Sao PauloTel. ++55 (11) 5 03134 55,Fax (11) 50 3130 67
Canada Endress+Hauser Ltd.Burlington, OntarioTel. ++1 (905) 681 92 92,Fax (9 05) 6 8194 44
ChileDIN Instrumentos Ltda.SantiagoTel. ++56 (2) 2 050100, Fax (2) 2 25 8139
ColombiaColsein Ltd.Bogota D.C.Tel. ++57 (1) 2 367659, Fax (1) 6 10 7868
Costa RicaEURO-TEC S.A.San JoseTel. ++506 (2) 961542, Fax (2 ) 961542
EcuadorInsetec Cia. Ltda.QuitoTel. ++593 (2) 269148, Fax (02) 4618 33
GuatemalaACISA Automatizacion Y Control Industrial S.A.Ciudad de Guatemala, C.A.Tel. ++502 (3) 345985, Fax (2) 3274 31
Mexico Endress+Hauser I.I.Mexico CityTel. ++52 (5) 56 89 65, Fax (5) 56 8418
ParaguayIncoel S.R.L.AsuncionTel. ++595 (21) 21 3989, Fax (21) 226583
UruguayCircular S.A.MontevideoTel. ++598 (2) 925785, Fax (2) 9291 51
USA Endress+Hauser Inc.Greenwood, IndianaTel. ++1 (317) 535 71 38,Fax (3 17) 5 3584 89
VenezuelaH. Z. Instrumentos C.A.CaracasTel. ++58 (2) 9 440966, Fax (2) 9 44 4554
Asia
China Endress+Hauser Shanghai
Instrumentation Co. Ltd.ShanghaiTel. ++86 (21) 5490 2300,Fax (21) 5490 23 03
Endress+Hauser Beijing OfficeBeijingTel. ++86 (10) 6834 4058,Fax (10) 6834 40 68
Endress+Hauser (H.K.) Ltd.Hong KongTel. ++852 (2) 528 31 20, Fax (2) 86541 71
India Endress+Hauser India Branch OfficeMumbaiTel. ++91 (22) 8 52 1458, Fax (22) 852 19 27
IndonesiaPT Grama BazitaJakartaTel. ++62 (21) 7 97 5083, Fax (21) 797 50 89
Japan Sakura Endress Co., Ltd.TokyoTel. ++81 (4 22) 54 0611, Fax (4 22) 55 02 75
Malaysia Endress+Hauser (M) Sdn. Bhd.Petaling Jaya, Selangor Darul EhsanTel. ++60 (3) 7 334848, Fax (3) 7 33 8800
PakistanSpeedy AutomationKarachiTel. ++92 (21) 7 72 2953, Fax (21) 773 68 84
Papua New GuineaSBS Electrical Pty LimitedPort MoresbyTel. ++675 (3) 25 1188, Fax (3) 2595 56
PhilippinesBrenton Industries Inc.Makati Metro ManilaTel. ++63 (2) 6 38 8041, Fax (2) 638 8042
Singapore Endress+Hauser (S.E.A.) Pte., Ltd.SingaporeTel.++65 (5) 66 8222, Fax (2) 6668 48
South Korea Endress+Hauser (Korea) Co., Ltd.SeoulTel. ++82 (2) 6 58 7200, Fax (2) 659 2838
TaiwanKingjarl CorporationTaipei R.O.C.Tel. ++886 (2) 2718 39 38, Fax (2) 27 1341 90
Thailand Endress+Hauser Ltd.BangkokTel. ++66 (2) 9 96 7811 20, Fax (2) 9 967810
VietnamTan Viet Bao Co. Ltd.Ho Chi Minh CityTel. ++84 (8) 8 33 5225, Fax (8) 833 5227
IranTelephone Technical Services Co. Ltd.TehranTel. ++98 (21) 874 6750, Fax (21) 87372 95
IsraelInstrumetrics Industrial Control Ltd.Tel-AvivTel. ++972 (3) 64802 05, Fax (3) 6 4719 92
JordanA.P. Parpas Engineering S.A.AmmanTel. ++962 (6) 46432 46, Fax (6) 4 6457 07
Kingdom of Saudi ArabiaAnasiaJeddahTel. ++966 (2) 67100 14, Fax (2) 6 7259 29
KuwaitKuwait Maritime & Mercantile Co. K.S.C.SafatTel.++9 65 (2) 4414 81, Fax (2) 44 1486
LebanonNabil IbrahimJbeilTel. ++961 (3) 25 4052, Fax (9) 5480 38
Sultanate of OmanMustafa & Jawad Science & Industry Co.L.L.C.RuwiTel. ++968 (60) 20 09, Fax (60) 70 66
United Arab EmiratesDescon Trading EST.DubaiTel. ++971 (4) 65 3651, Fax (4) 6532 64
YemenYemen Company for Ghee and Soap IndustryTaizTel. ++976 (4) 23 0664, Fax (4) 2123 38
Australia + New Zealand
AustraliaALSTOM Australia Ltd.SydneyTel. ++61 (2) 9722 4777, Fax (2) 972248 88
New ZealandEMC Industrial InstrumentationAucklandTel. ++64 (9) 4 15 5110, Fax (9) 415 5115
All other countries
Endress+Hauser GmbH+Co.Instruments International
D-Weil am RheinGermanyTel. ++49 (76 21) 9 7502, Fax (76 21) 97 5345
BA 209C/07/en/07.02Printed in Germany / CV5 / DT
Members of the Endress+Hauser group
Endress HauserThe Power of Know How51517599