LABORATORY 1: BIOREACTOR-BASICS
OBJECTIVE
- To understand the basic design and operation features of the bioreactor.
THEORY
Key parts of the bioreactor are described below:
Bioreactor – can be described as a vessel which has provision of cell cultivation under sterile
condition & control of environmental conditions e.g., pH, Temperature, Dissolved oxygen
etc. It can be used for the cultivation of microbial plant or animal cells. A typical bioreactor
consists of following parts.
Agitator – This facilitates the mixing of the contents of the reactor which eventually keeps
the “cells” in the perfect homogenous condition for better transport of nutrients and oxygen
for adequate metabolism of cell to the desired product(s).
The agitator can be top driven or bottom which could be basically magnetic / mechanically
driven. The bottom driven magnetic /mechanical agitators are preferred as opposed to top
driven agitators as it saves adequate space on the top of the vessel for insertion of essential
probes (Temperature, pH, dissolved oxygen foam, Co2 etc) or inlet ports for acid, alkali, foam,
fresh media inlet /exit gases etc. However mechanical driven bottom impellers need high
quality mechanical seals to prevent leakage of the broth.
Baffle – The purpose of the baffle in the reactor is to break the vortex formation in the vessel,
which is usually highly undesirable as it changes the centre of gravity of the system and
consumes additional power.
Sparger – In aerobic cultivation process the purpose of the sparger is to supply oxygen to the
growing cells. Bubbling of air through the sparger not only provide the adequate oxygen to
the growing cells but also helps in the mixing of the reactor contents thereby reducing the
power consumed to achieve a particular level of (mixing) homogeneity in the culture.
Jacket – The jacket provides the annular area for circulation of constant temperature water
which keeps the temperature of the bioreactor at a constant value. The desired temperature of
the circulating water is maintained in a separate Chilled Water Circulator which has the
provision for the maintenance of low/high temperature in a reservoir. The contact area of
jacket provides adequate heat transfer area wherein desired temperature water is constantly
circulated to maintain a particular temperature in the bioreactor.
Basic control systems for the operation of the bioreactor are described below:
Temperature Measurement and control – The measurement of the temperature of the
bioreactor is done by a thermocouple or Pt -100 sensor which essentially sends the signal to
Fermentation Technology Lab Bioreactor-Basics Ms. Phoeurk Chanrith
2
the Temperature controller. The set point is entered in the controller which then compares the
set point with the measured value and depending on the error, either the heating or cooling
finger of the bioreactor is activated to slowly decrease the error and essentially bring the
measured temperature value close to the set point.
pH measurement and control – The measurement of pH in the bioreactor is done by the
autoclavable pH probe. The measured signal is compared with the set point in the controller
unit which then activates the acid or alkali to bring the measured value close to the set point.
However before the pH probe is used, it needs to be calibrated with two buffers usually in the
pH range which is to be used in the bioreactor cultivation experiment. The probe is first
inserted in (let us say) pH 4 buffer and the measured value is corrected by the zero knob of the
controller. Thereafter the probe is put in pH 7 buffer and if needed the measured value is
corrected by the asymmetry knob of the controller. The pH probe is now ready for use in the
range 0-7 pH range.
Identification of pH controller control settings for Bio-Engineering AG (Switzerland)
bioreactor – For this specific pH controller one has to suitably identify the right control
action setting for the addition of certain concentration of acid / alkali in the desired
fermentation broth which can give quick control action with-out any oscillations/offset of
measured value around the set point. The controller panel and the different knobs are
adequately described in the following figure –
Before the start of autoclaving of the broth for any cultivation experiment, it is
essential to calibrate the pH probe (as described above). Thereafter the set point (say 5) and p-
band (1.0) is entered on the controller. This essentially means that now the pH controller will
control the pH value in the range 4.0 to 6.0 For example if the measured pH value in the
Fermentation Technology Lab Bioreactor-Basics Ms. Phoeurk Chanrith
3
bioreactor is 4.5 then the controller will trigger alkali addition to the reactor to bring the
measured value to 5.0 and similarly if the pH value is 5.5 it activates the acid pump to bring
down the pH value to the set point. It is ensured that the p-band value is not kept too small or
else it may lead to oscillation of the measured value around the set point similarly if the p-
band is too large then it may give rise to offset between the measured value and set point of
the controller. It may also be noted that the activation of acid/alkali pumps is done in a phased
manner. For example if the controller is adding alkali to bring down the pH, the addition of
the alkali is not done in one shot. In fact the alkali addition pump is kept on for some time and
then it is off for some time. This ensures adequate mixing of first installment of acid/alkali in
the broth before next installment is added. This strategy avoids over addition of acid/alkali for
the pH control. The On/Off time of the controller has to be adjusted by separate experiments
and will depend on the buffering capacity of the broth, concentration of acid/alkali etc.
However it is absolutely essential to identify & maintain these setting before the start of the
experiment in order to have efficient control action of the control which features stable quick
control action with-out oscillations &/or off set around set point. There is another knob td in
the control panel which provides the setting of the safety time during which if the control
action is not achieved it raises the alarm for the operator.
Dissolved oxygen controller – The dissolved oxygen in the bioreactor broth is measured by a
dissolved oxygen probe which basically generates some potential corresponding to the
dissolved oxygen diffused in the probe. Before the measurement can be done by the probe it is
to be calibrated for its zero and hundred percent values. The zero of the probe is set by (zero
knob) the measured value of the dissolved oxygen when the broth is saturated with nitrogen
purging. Similarly the hundred percent of the instrument is calibrated by the measured value
of dissolved oxygen when broth is saturated with purging air in it. After calibration the
instrument is ready for the measurement of the dissolved oxygen in the broth. In the event of
low oxygen in the fermentation broth, more oxygen can be purged in the bioreactor &/or
stirrer speed can be increased to enhance the beating of the bubbles which essentially
enhances the oxygen transfer area and net availability of oxygen in the fermentation broth.
Foam control – The fermentation broth contains a number of organic compounds and the
broth is vigorously agitated to keep the cells in suspension and ensure efficient nutrient
transfer from the dissolved nutrients and oxygen. This invariably gives rise to lot of foam. It is
essential that control of the foam is done as soon as possible.
PROCEDURE AND SIMULATOR
1. Speed control system
The purpose of Speed Control System is to provide mixing of reactor contents
(Microorganism and Medium).
To start the operation of the reactor enter “Set Lower Limit” as some value let us say 100.
Then enter “Set Upper Limit” as some value Let us say 400.
Then enter “Set Actual Speed” as some value Let us say 200.
Then press RUN.
The agitator will start rotating at 200 RPM.
To stop the agitator press STOP button (Do not do it now!!).
Fermentation Technology Lab Bioreactor-Basics Ms. Phoeurk Chanrith
4
2. Speed control system
The purpose of Speed Control System is to provide mixing of reactor contents
(Microorganism and Medium).
To start the operation of the reactor enter “Set Lower Limit” as some value let us say 100.
Then enter “Set Upper Limit” as some value Let us say 400.
Then enter “Set Actual Speed” as some value Let us say 200.
Then press RUN.
The agitator will start rotating at 200 RPM.
To stop the agitator press STOP button (Do not do it now!!).
Press enter NEXT button to go to Gas supply system
Fermentation Technology Lab Bioreactor-Basics Ms. Phoeurk Chanrith
5
3. Temperature Control System
The purpose of Temperature control system is to maintain the right cultivation temperature in
the bioreactor. The temperature is measured by the temperature sensor (Blue). This value is
then compared with the set point value and depending on the error the temperature
manipulation is done by passing either cooling (blue dots) or hot water (red dots) in the jacket
of the bioreactor.
4. pH measurement and Control System
Enter any pH value say 5.0 in the set point and press RUN.
The pH in-side the bioreactor is measured by the sensor (orange) and is compared with the set
point and depending on the error either acid (green) or base (orange) peristaltic pump is
triggered from the controller.
Fermentation Technology Lab Bioreactor-Basics Ms. Phoeurk Chanrith
6
5. Dissolved Oxygen measurement and Control Enter value of Dissolved oxygen level in the bioreactor say (30% Saturation) and press
RUN.
The Dissolved oxygen inside the bioreactor is measured by magenta sensor, the value
is compared with the set point and depending on error the controller action is activated. For
example if the Dissolved oxygen in the bioreactor is registered low the controller speeds up
the air supply and/or the agitation speed to increase the level of dissolved oxygen in the
bioreactor. Similarly if the level of the dissolved oxygen is measured low in the bioreactor it
will reduce the rate of supply of oxygen &/or RPM in the bioreactor. This way the set point
value of 30 % saturation level of dissolved oxygen is maintained in the bioreactor.
6. Foam Control system
Foam is defined as dispersion of liquid in gas and it is invariably produced in any cell
cultivation primarily due to presence of nitrogen and protein material in the medium. Though
it is essential to keep the nitrogenous material in the medium as high as possible because it is
essentially required to give good growth however the foam is on the other hand highly
undesirable as it can lead to heavy loss of medium from the bioreactor. The foam is detected
by Red sensor and antifoam addition is triggered through peristaltic pump to kill the foam.
Peristaltic pump adding the antifoam solution (blue bottle) is made to work at a very slow
speed primarily to add as minimum amount as possible. Over addition of antiform may create
serious problem in isolation of the product from the fermentation broth.
Fermentation Technology Lab Bioreactor-Basics Ms. Phoeurk Chanrith
7
REFFERENCE
1. iitd.vlab.co.in,. (2011). Bioreactor- Basics. Retrieved 20 September 2015, from
iitd.vlab.co.in/?sub=63&brch=177&sim=647&cnt=1
2. Biochemical Engineering, By S. Aiba, A.E. Humphrey and N.F. Millis, 2nd Edition,
University of Tokyo Press, Japan
3. Biochemical Engineering Fundamentals By J.E. Bailey, D.F. Ollis 2nd Edition, McGraw
Hill Book Company, New Delhi
4. Biochemical Engineering By J.M. Lee Prentice Hall, Englewood Cliffs, New Jersey
5. Bioprocess Engineering, Basic Concepts By M.L. Shuler, F. Kargi Prentice Hall,
Englewood Cliffs, New Jersey
6. Bioprocess Engineering Systems, Equipment & Facilities By B.K. Lydersen, N.A. D'Elia
& K.L. Nelsen John Wiley and Sons Inc. New York
7. Biological Reaction Engineering, I.I. Dunn, E. Heinzle, J. Ingham, J.E. Prenosil, VCH
Publishers, Weinheim.
Fermentation Technology Lab Bioreactor-Basics Ms. Phoeurk Chanrith
8
8. Process Biotechnology Fundamentals, S.N. Mukhopadhyay, Viva Books Pvt. Limited,
New Delhi.
9. Fermentation and Enzyme Technology By D.I.C. Wang, C.L. Cooney, A.L. Demain, P.
Dunnill, A.E. Humphrey & M.D. Lilly John Wiley and sons, New York
10. Principles of Fermentation Technology By P.F. Stanbury & A. Whitaker Pergamon
Press, New York
11. Biotechnology : A textbook of Industrial Microbiology By W Crueger & Crueger,
Sinauer Associates Inc, Sunderland, MA & Science Tech, Inc. Madison, WI
12. Basic Biotechnology By J.Bulock & B. Kristiansen Academic Press, New York (U.S.A.)
13. Concepts in Biotechnology By D. Balasubramanian, C.F.A. Bryce, K Dharmalingam, J.
Green, K Jayaraman. University Press, Hyderabad
14. Manual of Industrial Microbiology and Biotechnology, Arnold L. Demain, Nadine A.
Solomon, American Society for Microbiology, D.C.
15. Introduction to Industrial Sterilization, Richards