digital mammography detector qc introduction introduction
TRANSCRIPT
1
Slide 1
Digital Mammography
Detector QC
Eric A. Berns, Ph.D.Eric A. Berns, Ph.D.
[email protected]@gmail.com
University of Colorado at Denver and HSCUniversity of Colorado at Denver and HSCDenver, CODenver, CO
Slide 2
What is FFDM QC and why is it What is FFDM QC and why is it
important?important?
What to know before you startWhat to know before you start
Overview and compare QC testsOverview and compare QC tests
Key take home pointsKey take home points
IntroductionIntroduction
Slide 3
Accredited FFDM units
Certified Facilities with FFDM units
Total Accredited Mammo Units
Total Certified MammoFacilities
+1,845
+1,293
-575
-123
Difference
6,964
4,659
12,875
8,714
July 1, 2008
5,119
3,366
13,450
8,837
July 1, 2007
Certified Statistics – Past Year
IntroductionIntroduction 5,911 SFM Units
Slide 4
MQSA MQSA
–– Mammography Quality Standards ActMammography Quality Standards Act
ACRACR
–– American College of RadiologyAmerican College of Radiology
IntroductionIntroduction
Slide 5
ScreenScreen--FilmFilm
Slide 6
FFDMFFDM
In FFDM, the manufacturer designs and mandates their In FFDM, the manufacturer designs and mandates their own QC program and action limitsown QC program and action limitsIn FFDM, you must the manufacturersIn FFDM, you must the manufacturers’’ QC programQC program
2
Slide 7
AGFA IMPAX MA3000AGFA IMPAX MA3000–– Clean monitor surfaceClean monitor surface–– Measure display white & blackMeasure display white & black–– Measure quality levelMeasure quality level–– Measure uniformityMeasure uniformity–– Calibrate displaysCalibrate displays–– View SMPTE patternView SMPTE pattern–– Viewing conditionsViewing conditions
GE Seno AdvantageGE Seno Advantage–– Viewing conditions check and settingViewing conditions check and setting–– Monitor calibrationMonitor calibration–– Image quality Image quality –– SMPTE patternSMPTE pattern–– Analysis of screen uniformityAnalysis of screen uniformity
Siemens Siemens MammoReportMammoReport PlusPlus–– Geometric distortionGeometric distortion–– Reflection Reflection –– Luminance ResponseLuminance Response–– Luminance UniformityLuminance Uniformity–– ResolutionResolution–– NoiseNoise–– Overall EvaluationOverall Evaluation
Slide 8
BarcoBarco–– II--Guard CheckGuard Check–– Quality LevelQuality Level–– Display WhiteDisplay White–– Calibration Settings CheckCalibration Settings Check
EizoEizo–– Pattern CheckPattern Check–– Luminance CheckLuminance Check–– GrayscaleGrayscale CheckCheck–– Uniformity CheckUniformity Check
PlanarPlanar–– Monitor CleaningMonitor Cleaning–– Viewing Conditions CheckViewing Conditions Check–– Monitor Calibration CheckMonitor Calibration Check–– Image Quality Image Quality –– SMPTESMPTE–– Visual Inspection for Display Visual Inspection for Display
DefectsDefects
Slide 9
Quality Control: Why?Quality Control: Why?
Reduce exposure to patients and Reduce exposure to patients and
personnelpersonnel
Consistent image qualityConsistent image quality
Detect and correct for potential problems, Detect and correct for potential problems,
before they impact image qualitybefore they impact image quality
Slide 10
What is Quality Control?What is Quality Control?
Determination of what is Determination of what is ““NormalNormal””
Detection of what is Detection of what is ““AbnormalAbnormal””
Understanding of how to return to Understanding of how to return to ““NormalNormal””from from ““AbnormalAbnormal””
In particular, in FFDM, how do you know In particular, in FFDM, how do you know what you are seeing is what it is supposed to what you are seeing is what it is supposed to be?be?
Slide 11
Golden Rules for FFDM QCGolden Rules for FFDM QC–– Must use manufacturerMust use manufacturer’’s QC proceduress QC procedures
•• Mandate action limitsMandate action limits
–– ManufacturersManufacturers’’ QC may refer to Monitor & Printer QC may refer to Monitor & Printer
ManufacturersManufacturers’’ QCQC
–– Multimodality Workstations have own separate QCMultimodality Workstations have own separate QC
–– Printers may have their own QCPrinters may have their own QC
–– Most failures result in stopping clinical imaging Most failures result in stopping clinical imaging
until failure can be correcteduntil failure can be corrected
Before You Begin QCBefore You Begin QCSlide 12
Obtain proper training & CE credits (8 hours)Obtain proper training & CE credits (8 hours)
–– HandsHands--on training on on training on actualactual unit:unit:
•• MechanicsMechanics
•• SoftwareSoftware
•• ArtifactsArtifacts
•• Learn vendor specific tests and tricks Learn vendor specific tests and tricks
Before You BeginBefore You Begin
3
Slide 13
Must have proper continuing experience:Must have proper continuing experience:
–– 2 facilities2 facilities
–– 6 units6 units
–– Within last 2 yearsWithin last 2 years
Before You BeginBefore You BeginSlide 14
Before You BeginBefore You Begin
Note:Note:––For new unitFor new unit: Must use most current : Must use most current
versionversion––For renewal unitFor renewal unit: Can use older : Can use older
version (version used when tested version (version used when tested previously)previously)
Slide 15
ACR AccreditationACR Accreditation -- www.acr.orgwww.acr.org–– Physics formsPhysics forms
•• http://acr.org/accreditation/mammography/mammo_qc_forms.aspxhttp://acr.org/accreditation/mammography/mammo_qc_forms.aspx
–– GEGE SenographeSenographe 2000D, DS, and Essential2000D, DS, and Essential–– Fischer Fischer SenoscanSenoscan–– LoradLorad SeleniaSelenia–– Siemens Siemens NovationNovation–– Fuji Fuji FCRmFCRm
•• NoteNote: if using unit for both screen: if using unit for both screen--film & CR, you must accredit for film & CR, you must accredit for bothboth
FDA AccreditationFDA Accreditation–– www.fda.gov/cdrh/mammographywww.fda.gov/cdrh/mammography//–– Other vendors when approvedOther vendors when approved
Before You BeginBefore You BeginSlide 16
Before You BeginBefore You Begin
Slide 17
Before You BeginBefore You BeginSlide 18
Before You BeginBefore You Begin
4
Slide 19
Before You BeginBefore You BeginSlide 20
Before You BeginBefore You Begin
Slide 21
ScreenScreen--Film MammoFilm Mammo–– Assure XAssure X--ray field aligns with the light fieldray field aligns with the light field
–– Collimator allows full coverage of receptorCollimator allows full coverage of receptor–– Chest wall edge of compression paddle aligns with chest Chest wall edge of compression paddle aligns with chest
wall of receptorwall of receptor
FFDMFFDM–– SameSame
–– Multiple target/filters and/or multiple paddle positionsMultiple target/filters and/or multiple paddle positions
–– Chest Wall Missed Tissue Chest Wall Missed Tissue –– Compression Plate Overlap on the Chest Wall SideCompression Plate Overlap on the Chest Wall Side
CollimationCollimationSlide 22
ScreenScreen--Film MammoFilm Mammo–– LineLine--pair test toolpair test tool
•• Focal spotFocal spot
FFDMFFDM–– LineLine--pair test toolpair test tool
•• Focal spotFocal spot
•• Detector & imaging chainDetector & imaging chain
Limiting Spatial ResolutionLimiting Spatial Resolution
Slide 23
Spatial ResolutionSpatial Resolution
XXXXMTFMTF
XXXXXXMeasure using LP Measure using LP phantom on top of phantom on top of acrylicacrylic
XXSame as ScreenSame as Screen--Film Film (Focal Spot (Focal Spot EvalEval))
Siemens Siemens NovationNovation
LoradLoradSeleniaSelenia
Fischer Fischer SenoscanSenoscan
GE GE 2000D, 2000D,
DS, DS, EssentialEssential
Fuji Fuji FCRmFCRm
Slide 24
ScreenScreen--Film MammoFilm Mammo–– Measure optical densitiesMeasure optical densities
•• Different thicknesses using clinical modesDifferent thicknesses using clinical modes
•• Different density settings (Different density settings (--2, 2, --1, 0, +1, +2, etc.)1, 0, +1, +2, etc.)
FFDMFFDM–– Measure resultant techniquesMeasure resultant techniques
–– Measure signal, exposure, & SNR valuesMeasure signal, exposure, & SNR values
AEC PerformanceAEC Performance
5
Slide 25
AECAEC
XXXXXXXXImage Mode TrackingImage Mode Tracking
XXACR ReproducibilityACR Reproducibility
XXXXXXSNRSNR
XXCNR CNR
XXXXXXDensity Control Density Control FunctionFunction
Same as ScreenSame as Screen--FilmFilm
Siemens Siemens NovationNovation
LoradLoradSeleniaSelenia
Fischer Fischer SenoscanSenoscan
GE GE 2000D, 2000D,
DS, DS, EssentialEssential
Fuji Fuji FCRmFCRm
Slide 26
AOP Mode and SNR CheckAOP Mode and SNR Check–– Variable thicknesses of acrylic Variable thicknesses of acrylic –– Std, AutoStd, Auto–– Evaluate:Evaluate:
•• Correct techniques?Correct techniques?•• Adequate SNR?Adequate SNR?
2.5 cm 2.5 cm AcrylicAcrylic
4.0 cm 4.0 cm AcrylicAcrylic
6.0 cm Acrylic6.0 cm Acrylic
Acrylic Thickness
Target-Filter
Selected kVp
Selected mAs
2.5 cm Mo-Mo 27 kVp 20-60
4.0 cm Mo-Rh 28 kVp 35-90
6.0 cm Rh-Rh 32 kVp 35-90
Each Each ““rawraw”” image must have a measured SNR of image must have a measured SNR of at least 50at least 50
GE 2000DGE 2000D
Slide 27
GE DS & EssentialGE DS & Essential
DS specific QC testsDS specific QC tests–– AOP Mode and SNR CheckAOP Mode and SNR Check
•• Use builtUse built--in software for image acquisitionin software for image acquisition
•• AOP: STD/AutoAOP: STD/Auto
31314545--9595Rh/RhRh/Rh606029294040--9090Rh/RhRh/Rh505026262020--6060Mo/MoMo/Mo2525kVkVmAsmAsTrack/FilterTrack/Filter
Exposure ParametersExposure ParametersFor AOP STD mode For AOP STD mode olyoly
Acrylic Thickness Acrylic Thickness (mm)(mm)
SNR must be > 50SNR must be > 50
Slide 28
LoradLorad SeleniaSelenia
Automatic Exposure Control Automatic Exposure Control –– AUTOAUTO--FILTER FILTER –– 2, 4, 6, 8 cm2, 4, 6, 8 cm
–– 4 cm (4 cm (--5 thru +5, or 5 thru +5, or --3 to +4)3 to +4)
–– Use ROI to measure mean pixel value Use ROI to measure mean pixel value
inside virtual AEC detectorinside virtual AEC detector
Performance CriteriaPerformance Criteria–– Pixel value should not vary by more Pixel value should not vary by more
than 10% of meanthan 10% of mean
–– Exposure compensation steps shall Exposure compensation steps shall
meet requirements in pixel value tablemeet requirements in pixel value table
Slide 29
AEC Image Stability and Reproducibility and SNRAEC Image Stability and Reproducibility and SNR
–– ACR Phantom ACR Phantom
–– Mo/Mo, 28 kV, Mo/Mo, 28 kV, ““HH””, sensor 1, sensor 1
–– Record Record mAsmAs, SNR, Mean, Entrance Exposure (5 times), SNR, Mean, Entrance Exposure (5 times)
Action Limits:Action Limits:
–– CoefCoef. of . of VarVar for for mASmAS and X < 5%and X < 5%
–– SNR and Mean shall not vary by SNR and Mean shall not vary by ++ 15% of Mean15% of Mean
Siemens Siemens NovationNovationSlide 30
AEC Thickness TrackingAEC Thickness Tracking
–– 2, 4, and 6 cm2, 4, and 6 cm
–– ““HH”” modemode
–– Max deviation = (Max difference / mean value ) * 100Max deviation = (Max difference / mean value ) * 100
Siemens Siemens NovationNovation DRDR
6
Slide 31
Density Control FunctionDensity Control Function
–– 4 cm acrylic, ACR phantom technique4 cm acrylic, ACR phantom technique
–– Repeat at Repeat at --2, 2, --1, 0, +1, +2 etc. density1, 0, +1, +2 etc. density
–– Record Record mAsmAs
–– mAsmAs change should be 5 to 15% per stepchange should be 5 to 15% per step
Fuji Fuji FCRmFCRmSlide 32
Reproducibility & Image Mode TrackingReproducibility & Image Mode Tracking–– 4 cm acrylic with clinical technique 4 cm acrylic with clinical technique –– Position ion chamber in beamPosition ion chamber in beam–– Record Record mAsmAs and Xand X–– Repeat 3 timesRepeat 3 times–– Repeat in each mode (small, large, Repeat in each mode (small, large, magmag, no grid), no grid)–– Action LimitsAction Limits: :
•• Coefficient of variation for Exposure or Coefficient of variation for Exposure or mAsmAs must not exceed must not exceed 0.050.05
•• No significant difference in exposure between small and large No significant difference in exposure between small and large buckybucky when using similar gridswhen using similar grids
Fuji Fuji FCRmFCRm
Slide 33
CNR Per Object ThicknessCNR Per Object Thickness
–– CNR using clinical technique for 2 cmCNR using clinical technique for 2 cm
–– Repeat for 4 and 6 cmRepeat for 4 and 6 cm
–– Action LimitsAction Limits::
•• CNR of 2 cm relative to 4 cm must be CNR of 2 cm relative to 4 cm must be >> 100%100%
•• CNR of 6 cm relative to 4 cm must be CNR of 6 cm relative to 4 cm must be >> 75%75%
Fuji Fuji FCRmFCRmSlide 34
ScreenScreen--Film MammoFilm Mammo
–– Measure if ODMeasure if OD’’s are consistents are consistent
–– Check for Check for artifactsartifacts
FFDMFFDM
–– FlatFlat--field uniformityfield uniformity
–– Detector calibrationDetector calibration
–– Pixel correction testPixel correction test
–– CR Reader Scanner PerformanceCR Reader Scanner Performance
Uniformity of Screen Speed & Uniformity of Screen Speed & Detector PerformanceDetector Performance
Slide 35
Detector PerformanceDetector Performance
XXPixel CorrectionPixel Correction
XXGeometric DistortionGeometric Distortion
XXDynamic RangeDynamic Range
XXCR Reader Scanner CR Reader Scanner PerformancePerformance
XXXXGhostingGhosting
XXInterInter--Plate Plate ConsistencyConsistency
XXPrimary ErasurePrimary Erasure
XXXXXXDetector CalibrationDetector Calibration
XXXXFlatFlat--Field UniformityField Uniformity
Same as ScreenSame as Screen--Film Film (Screen Uniformity)(Screen Uniformity)
Siemens Siemens NovationNovation
LoradLoradSeleniaSelenia
Fischer Fischer SenoscanSenoscan
GE GE 2000D, 2000D,
DS, DS, EssentialEssential
Fuji Fuji FCRmFCRm
Slide 36
ScreenScreen--Film MammoFilm Mammo–– ACR Phantom ScoresACR Phantom Scores
–– Optical Density & ContrastOptical Density & Contrast
FFDMFFDM–– ACR Phantom ScoresACR Phantom Scores
•• Pass/fail requirements differ by vendorPass/fail requirements differ by vendor
–– SignalSignal--toto--Noise Ratio (SNR)Noise Ratio (SNR)
–– ContrastContrast--toto--Noise Ratio (CNR)Noise Ratio (CNR)
Image QualityImage Quality
7
Slide 37
Image QualityImage Quality
XXXXXXXXXXCNRCNR
PartialPartialXXXXClinical TechniqueClinical Technique
XXXXManual TechniquesManual Techniques
Same as ScreenSame as Screen--FilmFilm
Siemens Siemens NovationNovation
LoradLoradSeleniaSelenia
Fischer Fischer SenoscanSenoscan
GE GE 2000D, 2000D,
DS, DS, EssentialEssential
Fuji Fuji FCRmFCRm
Slide 38
ACR Phantom ImagingACR Phantom Imaging
4433MassesMasses
4433SpecksSpecks
5544FibersFibers
LoradLorad & & SiemensSiemens
GE & Fischer & GE & Fischer & Fuji Fuji
Image QualityImage Quality
Slide 39
ACR Phantom ImagingACR Phantom Imaging
–– Manual technique (Mo/Mo, 26 Manual technique (Mo/Mo, 26
kVp, 125 kVp, 125 mASmAS))
–– Score the Score the processedprocessed imageimage
–– Acquisition workstationAcquisition workstation
–– Each monitor of the RWSEach monitor of the RWS
–– Laser imagerLaser imager
GE 2000DGE 2000DSlide 40
ContrastContrast--toto--Noise Test (CNR)Noise Test (CNR)
–– To examine consistency of CNR To examine consistency of CNR
ratio measured over timeratio measured over time
–– Use the Use the rawraw imageimage
–– ++ 20% of baseline20% of baseline
Background ROIBackground ROI Mass ROIMass ROI
CNR = (CNR = (MeanMeanbackgroundbackground -- MeanMeanmassmass)/SD)/SDbackgroundbackground
GE 2000DGE 2000D
Slide 41
GE DSGE DS
DS specific QC testsDS specific QC tests
–– Phantom Image QualityPhantom Image Quality
•• Manual technique: Manual technique: Rh/RhRh/Rh, 29 kVp, 56 , 29 kVp, 56 mAsmAs
–– MTF and CNR MeasurementMTF and CNR Measurement
•• Use IQST test toolUse IQST test tool
•• Use builtUse built--in software for image acquisitionin software for image acquisition
•• Manual technique: Manual technique: Rh/RhRh/Rh 30 kVp, 56 30 kVp, 56 mAsmAs
•• Results are automatically displayed (pass/fail)Results are automatically displayed (pass/fail)
•• Same action limits as 2000DSame action limits as 2000D
Slide 42
Phantom Image QualityPhantom Image Quality–– Select clinical exposure mode (i.e. AUTOSelect clinical exposure mode (i.e. AUTO--FILTER)FILTER)–– Print film Print film –– Measure background OD and density difference Measure background OD and density difference ––
Plot on tech worksheetsPlot on tech worksheets•• Background must be Background must be >> 1.20 OD 1.20 OD ++ 0.200.20
•• DD must be DD must be >> 0.40 0.40 ++ 0.050.05
–– Score on each Soft Copy WorkstationScore on each Soft Copy Workstation•• 5 fibers5 fibers
•• 4 speck groups4 speck groups
•• 4 masses4 masses
LoradLorad SeleniaSelenia
8
Slide 43
Phantom Image QualityPhantom Image Quality–– Position phantom 1 cm Position phantom 1 cm
over chest wall edgeover chest wall edge
–– Select: 28 kV, AECSelect: 28 kV, AEC--Auto, Auto,
Mo/MoMo/Mo
–– Score on RWS or FilmScore on RWS or Film
–– Fiber Fiber >> 55
–– Specks Specks >> 44
–– Masses Masses >> 44
Siemens Siemens NovationNovation DRDRSlide 44
SNR and CNR MeasurementsSNR and CNR Measurements
–– SNRSNR at least at least >> 4040
–– CNRCNR should stay within should stay within ±±15% of baseline15% of baseline
•• Obtained during acceptance testingObtained during acceptance testing
LoradLorad & Siemens& Siemens
Slide 45
ContrastContrast--toto--Noise Test (CNR)Noise Test (CNR)–– To examine consistency of CNR ratio measured over timeTo examine consistency of CNR ratio measured over time
–– Use 4 cm acrylic & 0.2 mm AlUse 4 cm acrylic & 0.2 mm Al
–– Manual technique (Mo/Mo, 26 kVp, 125 Manual technique (Mo/Mo, 26 kVp, 125 mAsmAs))
–– Calculate CNR using softwareCalculate CNR using software
–– ++ 20% of baseline20% of baseline
Fuji Fuji FCRmFCRmSlide 46
ScreenScreen--Film MammoFilm Mammo–– Dose for single CC view of ACR phantom shall Dose for single CC view of ACR phantom shall
not exceed 3.0 not exceed 3.0 mGymGy per exposure per FDAper exposure per FDA
FFDMFFDM–– SameSame
DoseDose
Slide 47
ScreenScreen--Film MammoFilm Mammo
–– Measure optical densitiesMeasure optical densities
•• Density difference, midDensity difference, mid--density, density, base+fogbase+fog
•• Measures consistencyMeasures consistency
FFDMFFDM
–– ManufacturersManufacturers’’ recommendationsrecommendations
–– Some refer to printer manufacturersSome refer to printer manufacturers’’ recommendationsrecommendations
–– Typically identical to ACR SFM ManualTypically identical to ACR SFM Manual
Film ProcessingFilm ProcessingSlide 48
Film Processor QCFilm Processor QC
XXXXFollow Follow FFDMsFFDMs’’ QCQC
XXXXXXIf Not, Use TheirsIf Not, Use Theirs
XXXXXXFollow Printer Follow Printer Manufacturers QCManufacturers QC
Siemens Siemens NovationNovation
LoradLoradSeleniaSelenia
Fischer Fischer SenoscanSenoscan
GE GE 2000D, 2000D,
DS, DS, EssentialEssential
Fuji Fuji FCRmFCRm
9
Slide 49
MidMid--DensityDensity
Film ProcessingFilm Processing
Northwestern Memorial Hospital - Processor ID: 2 - MAR 2006 Mid-Density
1.30
1.40
1.50
1.60
1.70
1.80
1.90
2.00
2.10
2.20
2.30
1-Mar-0
6
2-Mar-0
6
3-Mar-
06
6-Mar-
06
7-Mar-
06
8-Mar-
06
9-Mar-
06
10-M
ar-06
13-M
ar-06
14-M
ar-06
servi
ce 3/14
15-M
ar-06
16-M
ar-06
17-M
ar-06
20-M
ar-06
21-Mar-
06
22-Mar-
06
23-M
ar-06
24-M
ar-06
27-M
ar-06
28-M
ar-06
29-M
ar-06
30-Mar-
06
31-M
ar-06
Date
Mid-Density Upper Control Limit = 2.06Mid-Density Daily ValueMid-Density Daily Value Operating Level = 1.91Mid-Density Lower Control Limit = 1.76
MD Step=12
Northwestern Memorial Hospital - Processor ID: Kodak 8900 - Oct 2006 Mid-Density
0.70
0.80
0.90
1.00
1.10
1.20
1.30
1.40
1.50
1.60
1.70
October
2, 20
06
October
3, 200
6
October 4,
2006
October
5, 20
06
October
6, 20
06
October
9, 200
6
October 10
, 200
6
October
11, 2
006
October
12, 2
006
October
13, 2
006
October 1
6, 20
06
October
17, 2
006
October
18, 2
006
October
19, 20
06
October 2
0, 20
06
October
23, 2
006
October
24, 2
006
October
25, 20
06
October 2
6, 20
06
October
27, 2
006
October
30, 2
006
Date
Mid-Density Upper Control Limit = 1.48
Mid-Density Daily Value
Mid-Density Daily Value Operating Level = 1.33
Mid-Density Lower Control Limit = 1.18
Wet Processing Laser Printer
Slide 50
Density DifferenceDensity Difference
Film ProcessingFilm Processing
Northwestern Memorial Hospital - Processor ID: 2 - MAR 2006 Density Difference
1.50
1.60
1.70
1.80
1.90
2.00
2.10
2.20
2.30
2.40
2.50
2.60
1-Mar-
06
2-Mar-
06
3-Mar-0
6
6-Mar-
06
7-Mar-0
6
8-Mar-0
6
9-Mar-
06
10-Mar-
06
13-Mar-
06
14-Mar-
06
service
3/14
15-M
ar-06
16-M
ar-06
17-Mar-
06
20-Mar-
06
21-Mar-
06
22-Mar-
06
23-M
ar-06
24-M
ar-06
27-M
ar-06
28-Mar-
06
29-Mar-
06
30-Mar-
06
31-Mar-
06
Date
Density Difference Upper Control Limit = 2.25
Density Difference Daily Value
Density Difference Daily Value Operating Level = 2.1
Density Difference Lower Control Limit = 1.95
HD Step=13 LD Step=10
Northwestern Memorial Hospital - Processor ID: Kodak 8900 - Oct 2006 Density Difference
1.00
1.10
1.20
1.30
1.40
1.50
1.60
1.70
1.80
1.90
2.00
October
2, 200
6
October 3, 2
006
October
4, 200
6
October 5,
2006
Octobe
r 6, 2
006
October 9,
2006
October 1
0, 2006
October 1
1, 200
6
October
12, 20
06
October 13, 2
006
October
16, 20
06
October 17, 2
006
October
18, 20
06
October 19
, 2006
October 2
0, 200
6
October 23
, 2006
October 2
4, 200
6
October 25
, 2006
October 2
6, 200
6
October
27, 20
06
October 3
0, 200
6
Date
Density Difference Upper Control Limit = 1.77
Density Difference Daily Value
Density Difference Daily Value Operating Level = 1.62
Density Difference Lower Control Limit = 1.47
Wet Processing Laser Printer
Slide 51
DDmaxmax
Film ProcessingFilm Processing
Northwestern Memorial Hospital - Processor ID: Kodak 8900 - Oct 2006 Dmax
2.60
2.70
2.80
2.90
3.00
3.10
3.20
3.30
3.40
3.50
3.60
3.70
3.80
3.90
4.00
Octobe
r 2, 2
006
Octobe
r 3, 2
006
Octobe
r 4, 2
006
October
5, 20
06
October
6, 20
06
October
9, 20
06
October
10, 2
006
October
11, 2
006
October
12, 2
006
Octobe
r 13, 2
006
Octobe
r 16,
2006
Octobe
r 17,
2006
Octobe
r 18, 2
006
Octobe
r 19,
2006
October
20, 2
006
October
23, 2
006
Octobe
r 24,
2006
October
25, 2
006
Octobe
r 26,
2006
Octobe
r 27,
2006
Octobe
r 30, 2
006
Octobe
r 31,
2006
Date
Dmax Daily Value
Dmax - Operating Level = 3.44
Dmax Action Limit = 3.19
Laser Printer
Slide 52
ScreenScreen--Film MammoFilm Mammo
–– Evaluate cassettes Evaluate cassettes
FFDMFFDM
–– Evaluate detectorEvaluate detector
–– CR: Imaging PlatesCR: Imaging Plates
ArtifactsArtifacts
Slide 53
XXXXXXXXXXExpose Using AcrylicExpose Using Acrylic
Same as ScreenSame as Screen--FilmFilm
Siemens Siemens NovationNovation
LoradLoradSeleniaSelenia
Fischer Fischer SenoscanSenoscan
GE GE 2000D, 2000D,
DS, DS, EssentialEssential
Fuji Fuji FCRmFCRm
ArtifactsArtifacts ArtifactsArtifacts
10
DescriptionDescription: :
GE Good FlatGE Good Flat--field field
ImageImage
ArtifactsArtifacts
DescriptionDescription: :
HologicHologic Good Flat Good Flat
FieldField
ArtifactsArtifacts
DescriptionDescription: Acceptable : Acceptable
ACR Phantom imageACR Phantom image
Raw imageRaw image
WW = 200WW = 200
ArtifactsArtifactsDescriptionDescription: Good : Good imageimage
ArtifactsArtifacts
Artifact evaluation Artifact evaluation –– Contact Mode GE 2000DContact Mode GE 2000D
Mo/MoMo/Mo Mo/Mo/RhRh Rh/RhRh/Rh
ArtifactsArtifactsArtifact evaluation Artifact evaluation -- windowingwindowing
Window width = 30Window width = 30 Window width = 200Window width = 200 Window width = 1000Window width = 1000
ArtifactsArtifacts
11
DescriptionDescription: Non: Non--
uniform backgrounduniform background
Possible CausePossible Cause: :
Calibration fileCalibration file
SolutionSolution: Recalibrate: Recalibrate
ArtifactsArtifacts
DescriptionDescription: Low : Low
contrast overallcontrast overall
Possible CausePossible Cause: :
Window width too Window width too
wide wide –– (~750)(~750)
SolutionSolution: Re: Re--window window
ArtifactsArtifacts
DescriptionDescription: Poor : Poor
contrast, fibers and contrast, fibers and
masses failingmasses failing
Possible CausePossible Cause: Ultra low : Ultra low
dose dose –– Mo/Mo, 29 kVp, Mo/Mo, 29 kVp,
25 25 mAsmAs, 0.58 , 0.58 mGymGy
SolutionSolution: Increase : Increase
techniquestechniques
ArtifactsArtifacts
DescriptionDescription: Vertical : Vertical
and horizontal linesand horizontal lines
Possible CausePossible Cause: :
Gridlines, grid Gridlines, grid
artifacts in in cal fileartifacts in in cal file
SolutionSolution: Check grid : Check grid
mechanism, mechanism,
recalibraterecalibrate
ArtifactsArtifacts
ArtifactsArtifacts
DescriptionDescription::
White vertical bandsWhite vertical bands
Possible CausePossible Cause: :
Calibration fileCalibration file
SolutionSolution: :
Recalibrate detectorRecalibrate detector
ArtifactsArtifacts
12
DescriptionDescription: : Magnification image, Magnification image, several small flecks, several small flecks, grayscale gradationgrayscale gradation
Possible CausePossible Cause: : Object calibrated into Object calibrated into Cal file, grayscale Cal file, grayscale gradation ~ normalgradation ~ normal
SolutionSolution: Clean : Clean detector and tube detector and tube head, recalibrate head, recalibrate detectordetector
ArtifactsArtifacts
DescriptionDescription: White : White
pixelspixels
Possible CausePossible Cause: :
Detector going badDetector going bad
SolutionSolution: New : New
DetectorDetector
ArtifactsArtifacts
2x2xArtifactsArtifacts 4x4xArtifactsArtifacts
6x6xArtifactsArtifacts 8x8xArtifactsArtifacts
13
DescriptionDescription: :
Gridlines, black flecksGridlines, black flecks
Possible CausePossible Cause: Grid : Grid
failure, calibration filefailure, calibration file
SolutionSolution: Grid repair, : Grid repair,
clean the imaging clean the imaging
chain, and recalibratechain, and recalibrate
ArtifactsArtifacts
DescriptionDescription: Image : Image
processing around processing around
wax insertwax insert
Possible CausePossible Cause: :
Image processing Image processing
algorithmalgorithm
SolutionSolution: None: None
ArtifactsArtifacts
DescriptionDescription: Washed out Image: Washed out Image
CauseCause: Phantom not at chest wall edge : Phantom not at chest wall edge
SolutionSolution: Re: Re--position the phantomposition the phantom
ArtifactsArtifacts
DescriptionDescription: :
Linear and Linear and
rectangular rectangular
bandingbanding
Possible CausePossible Cause: :
Calibration fileCalibration file
SolutionSolution: Re: Re--
calibratecalibrate
ArtifactsArtifacts
DescriptionDescription: Shadows just : Shadows just
outside of breast skin lineoutside of breast skin line
Possible CausePossible Cause: Saturated : Saturated
detector due to overexposure detector due to overexposure
from dense breastfrom dense breast
SolutionSolution: Siemens : Siemens
recommends to increase kVp recommends to increase kVp
to reduce exposure time to reduce exposure time
ArtifactsArtifacts
DescriptionDescription: CR : CR --
Noisy Background, Noisy Background,
dark speckdark speck
Possible CausePossible Cause: :
Background Background
somewhat normalsomewhat normal
SolutionSolution: Check : Check
screen and system to screen and system to
identify black speckidentify black speck
ArtifactsArtifacts
14
DescriptionDescription: : Horizontal, vertical, Horizontal, vertical, and curved lines and curved lines multiple density multiple density differences differences –– printed printed from from ““flatflat--fieldfield”” menu menu on on LoradLorad SeleniaSeleniaPossible CausePossible Cause: Dirty : Dirty spinner assembly in spinner assembly in optics casting optics casting shadowsshadowsSolutionSolution: Clean : Clean spinner assemblyspinner assembly
Printer
ArtifactsArtifacts
Collimator needs adjustment at chest wall edge
ArtifactsArtifacts
Readout Line Artifact
DescriptionDescription: :
Horizontal and Horizontal and
curving linescurving lines
Possible CausePossible Cause: :
Readout error, Readout error,
ghostingghosting
SolutionSolution: Recalibrate, : Recalibrate,
let detector sit idle, let detector sit idle,
replace detectorreplace detector
ArtifactsArtifactsProcessing Steps for Digital ImagesProcessing Steps for Digital Images
RawRaw ProcessedProcessedImage Detection
Image Correction
Image Processing
Image Display
Calibration File
ArtifactsArtifacts
15
Summary on Artifacts•Most artifacts due to calibration file
•Window/level adjustments can appear as AEC and/or exposure problems
•Printers can cause fine, linear streaking artifacts - rare
•Look at technique factors, breast thickness, and breast density for clues
•Objects on bucky, mag stand, and up in tube head often make their way into the image – dust on accessories
•Detectors can, and do, fail
Slide 86
Obtain relevant handsObtain relevant hands--on trainingon training
Must perform manufacturer specific QC testsMust perform manufacturer specific QC tests
Artifacts Artifacts –– most problems can be seen heremost problems can be seen here
ReRe--booting and/or rebooting and/or re--calibrating fixes most calibrating fixes most problemsproblems
Laser PrinterLaser Printer–– DDmaxmax at least 3.5 ODat least 3.5 OD–– MidMid--density about 1.5 ODdensity about 1.5 OD
Key Take Home PointsKey Take Home Points
Slide 87
SAMsSAMs QuestionsQuestions
Slide 88
In digital mammography, who In digital mammography, who mandates the pass/fail criteria for site mandates the pass/fail criteria for site QC?QC?
20%
20%
20%
20%
20%
10
1.1. The American College of RadiologyThe American College of Radiology2.2. The FDAThe FDA3.3. The FFDM unit manufacturerThe FFDM unit manufacturer4.4. NEMANEMA5.5. MQSAMQSA
Slide 89
Answer: 3 Answer: 3 -- The FFDM unit The FFDM unit manufacturer manufacturer
ReferencesReferences
MQSA Regulations 900.12(e)(6)MQSA Regulations 900.12(e)(6)http://www.fda.gov/CDRH/MAMMOGRAhttp://www.fda.gov/CDRH/MAMMOGRAPHY/frmamcom2.html#s90012 PHY/frmamcom2.html#s90012
Slide 90
Answer 3: The FFDM unit Answer 3: The FFDM unit manufacturer manufacturer
ExplanationExplanation
““the quality assurance program shall be the quality assurance program shall be substantially the same as the quality substantially the same as the quality assurance program recommended by the assurance program recommended by the image receptor manufacturer, except that image receptor manufacturer, except that the minimum allowable dose for screenthe minimum allowable dose for screen--film systems in this sectionfilm systems in this section””
16
Slide 91
How do you accredit a Fuji CR How do you accredit a Fuji CR Mammography system which uses Mammography system which uses both CR and screenboth CR and screen--film on the same film on the same xx--ray system?ray system?
25%
25%
25%
25%
10
1.1. As 1 mammography unit?As 1 mammography unit?
2.2. As 2 mammography units?As 2 mammography units?
3.3. You cannot use CR and film on the same You cannot use CR and film on the same unit.unit.
4.4. CR is exempt from accreditationCR is exempt from accreditation
Slide 92
Answer: 2 Answer: 2 –– As 2 mammography As 2 mammography units units
ReferencesReferences
http://www.acr.org/accreditation/mammography/http://www.acr.org/accreditation/mammography/mammo_faq/mammo_faq_mamac.aspx#8.0.1mammo_faq/mammo_faq_mamac.aspx#8.0.1
Slide 93
Answer: 2 Answer: 2 -- As 2 mammography As 2 mammography unitsunits
ExplanationExplanation
As of November 15, 2006 facilities using As of November 15, 2006 facilities using both screenboth screen--film and CR on the same film and CR on the same mammography units must accredit these mammography units must accredit these 2 systems as 2 separate units.2 systems as 2 separate units.
Slide 94
What are the minimum passing ACR What are the minimum passing ACR phantom scores for the Siemens phantom scores for the Siemens NovationNovationDR for weekly QC?DR for weekly QC?
20%20%
20%20%20%
10
1.1. 5 Fibers, 4 Speck Groups, 3 Masses5 Fibers, 4 Speck Groups, 3 Masses2.2. 4 Fibers, 3 Speck Groups, 3 Masses4 Fibers, 3 Speck Groups, 3 Masses3.3. 5 Fibers, 4 Speck Groups, 4 Masses5 Fibers, 4 Speck Groups, 4 Masses4.4. 4 Fibers, 4 Speck Groups, 3 Masses4 Fibers, 4 Speck Groups, 3 Masses5.5. 4 Fibers, 4 Speck Groups, 4 Masses4 Fibers, 4 Speck Groups, 4 Masses
Slide 95
Answer 3Answer 3: : 5 Fibers, 4 Speck Groups, 4 5 Fibers, 4 Speck Groups, 4 MassesMasses
ReferencesReferences
Siemens. Siemens Quality Control Siemens. Siemens Quality Control Manual Version 05, Manual Version 05, ErlanganErlangan, Germany. , Germany. 20072007
Slide 96
Answer 3Answer 3: : 5 Fibers, 4 Speck Groups, 4 5 Fibers, 4 Speck Groups, 4 MassesMasses
ExplanationExplanation
The Siemens The Siemens NovationNovation DR QC manual states that DR QC manual states that the ACR Phantom minimum passing scores the ACR Phantom minimum passing scores are as follows:are as follows:
5 Fibers5 Fibers4 Speck Groups 4 Speck Groups 4 Masses4 Masses
17
Slide 97
For FFDM, the exposure for a single CC For FFDM, the exposure for a single CC view of the ACR Phantom shall not exceed?view of the ACR Phantom shall not exceed?
20%
20%
20%
20%
20%
10
1.1. 0.75 0.75 mGymGy/exposure/exposure2.2. 1.25 1.25 mGymGy/exposure/exposure3.3. 2.00 2.00 mGymGy/exposure/exposure4.4. 3.00 3.00 mGymGy/exposure/exposure5.5. 4.00 4.00 mGymGy/exposure/exposure
Slide 98
Answer 4Answer 4: 3.00 : 3.00 mGymGy/exposure/exposure
ReferencesReferences
900.12(e)(5)(vi): 900.12(e)(5)(vi): DosimetryDosimetry. The . The average glandular dose delivered during average glandular dose delivered during a single a single craniocaudalcraniocaudal view of an FDAview of an FDA--accepted phantom simulating a standard accepted phantom simulating a standard breast shall not exceed 3.0 breast shall not exceed 3.0 milligraymilligray((mGymGy) (0.3 ) (0.3 radrad) per exposure. The dose ) per exposure. The dose shall be determined with technique shall be determined with technique factors and conditions used clinically for factors and conditions used clinically for a standard breast. a standard breast.
Slide 99
Answer 4Answer 4: 3.00 : 3.00 mGymGy/exposure/exposure
ExplanationExplanation
The FDA requires:The FDA requires:
3.00 3.00 mGymGy/exposure/exposure
Slide 100
To meet the FDA requirement for continuing To meet the FDA requirement for continuing experience, how many mammography experience, how many mammography facilities and mammography unit surveys facilities and mammography unit surveys must be performed within the previous 24 must be performed within the previous 24 months?months?
20%
20%
20%
20%
20%
10
1.1. 6 Facilities and 2 mammography units6 Facilities and 2 mammography units2.2. 4 Facilities and 12 mammography units4 Facilities and 12 mammography units3.3. 2 Facilities and 6 mammography units2 Facilities and 6 mammography units4.4. 1 Facilities and 6 mammography units1 Facilities and 6 mammography units5.5. 2 Facilities and 12 mammography units2 Facilities and 12 mammography units
Slide 101
Answer 3Answer 3: : 2 Facilities and 6 Mammography 2 Facilities and 6 Mammography UnitsUnits
ReferencesReferences
http://www.fda.gov/CDRH/MAMMOGRAPHY/robohelp/http://www.fda.gov/CDRH/MAMMOGRAPHY/robohelp/START.HTMSTART.HTM900.12(a)(3)(iii)(B): Continuing experience. Following 900.12(a)(3)(iii)(B): Continuing experience. Following the second anniversary date of the end of the calendar the second anniversary date of the end of the calendar quarter in which the requirements of paragraphs quarter in which the requirements of paragraphs (a)(3)(i) and (a)(3)(ii) of this section were completed or (a)(3)(i) and (a)(3)(ii) of this section were completed or of April 28, 1999, whichever is later, the medical of April 28, 1999, whichever is later, the medical physicist shall have surveyed at least two physicist shall have surveyed at least two mammography facilities and a total of at least six mammography facilities and a total of at least six mammography units during the 24 months immediately mammography units during the 24 months immediately preceding the date of the facilitypreceding the date of the facility’’s annual MQSA s annual MQSA inspection or the last day of the calendar quarter inspection or the last day of the calendar quarter preceding the inspection or any date in between the preceding the inspection or any date in between the two. The facility shall choose one of these dates to two. The facility shall choose one of these dates to determine the 24determine the 24--month period. No more than one month period. No more than one survey of a specific facility within a 10survey of a specific facility within a 10--month period or month period or a specific unit within a period of 60 days can be a specific unit within a period of 60 days can be counted towards this requirement. counted towards this requirement.
Slide 102
Answer 3Answer 3: : 2 Facilities and 6 Mammography 2 Facilities and 6 Mammography UnitsUnits
ExplanationExplanation
The FDA requires:The FDA requires:
2 Facilities2 Facilities6 Mammography Units6 Mammography UnitsWithin past 24 monthsWithin past 24 months
18
Slide 103
Thank YouThank You