detector de gases explosivos portátil
TRANSCRIPT
-
7/31/2019 Detector de gases explosivos porttil
1/2
s
www.gasmonitors.comEmail: @bwtnet.com
GasAlertMicro 5 PIDVOCs (PID), H2S, CO, O2, SO2, PH3, NH3,
NO2, HCN, Cl2, ClO2, O3, and Combustibles
Five-Gas Detector
Protect Yourself Smaller and over 50% lighter than competitive instruments
Water-resistant design in an integral concussion-proof enclosure
Continuous LCD shows simultaneous gas concentrations for five gase
0 - 1000 ppm measurement of VOCs (Volatile Organic Compounds
Integral motorized pump option for remote sampling
Two power options: AA alkaline or rechargeable hot-swappablebattery packs
Provides 95 dB alarm tone and two bright wide-angled alarm bars
Equipped with internal vibrating alarm for high noise areas
Multi-language support in English, French, German, Spanish, andPortuguese
Three backlight options: standardactivates in low light (auto), in
alarm (auto) and reactivates on-demand; alarm onlyactivates inalarm conditions and reactivates on-demand; stealth mode
Four alarm levels: instantaneous Low and High Alarm for all gasesTWA (time-weighted average) and STEL (short-term exposure limifor VOCs and toxic gases
Simple automatic calibration procedure; compatible with BWMicroDock II automatic test and calibration station
Records TWA, STEL, and maximum exposures to gas and displaysreadings on command
User selectable field options include: PID correction factor,language selection, confidence beep, span level, set TWA/STELcalculation choice, set calibration due date, pass code protection,latching alarm, SAFE display function, stealth mode, andselectable ppm resolution for some gases (e.g. SO2 0.1 ppm)
Full function self-test of: sensor, battery and circuit integrity; andaudible/visual alarms
Ordering InformationGasAlertMicro 5 PID is delivered ready for use with sensors (as specified),sensor compartment cover for diffusion operation, battery (as specified),stainless steel alligator clip, calibration cap/hose, 2 extra electrode stacks,instructions in local languages and interactive training CD.
Battery/Charger OptionsThree AA alkaline batteriesRechargeable battery pack and desktop/wall mountable slip-in cradle charger
User-Downloadable Datalogger OptionDatalogging versions include 32 MB MMC card, software and an IR portfor use with the MicroDock II automated calibration and test station.
Pump OptionIncludes detector, standard accessories plus integrally attached motorizedpump and cover (installed), 12 in./30 cm sample probe with hydrophobicbarrier (water trap) and particulate filters, 5 ft./1.5 m sampling tubing, fivespare particulate filters, and one spare water trap filter.
Value Added Confined Space KitsKits include detector, standard accessories plus a hard-sided, foam-linedcarrying case, 10 ft./3 m sampling hose, sample probe, battery power (asspecified), 34 L calibration gas cylinders(s) with regulator(s).
AccessoriesMMC USB Reader Vehicle ChargerManual Aspirator Pump Carrying HolsterMicroDock II Automatic Calibration and Test Station
n ATEXXgand
others
www.clean.com.brEmail: [email protected]
GasAlertMicro 5 PID
with integrally attached pump
Optional rechargeable batteryand slip-in cradle charger
-
7/31/2019 Detector de gases explosivos porttil
2/2
SPECIFICATIONS DETECTOR: 1, 2, 3, 4, or 5-gas detector
GasMeasuring
rangeResolution TWA STEL LOW HIGH
O2 0-30.0% 0.1 N/A N/A 19.5 23.5
CO 0-999 ppm 1.0 35 50 35 200
H2S 0-300 ppm 1.0 10 15 10 15
PH3 0-5.0 ppm 0.1 0.3 1.0 0.3 1.0
SO2 0-100 ppm 0.1 2 5 2 5
NO2 0-99.9 ppm 0.1 2.0 5.0 2.0 5.0
HCN 0-30.0 ppm 0.1 4.7 10.0 4.7 10.0
Cl2 0-50.0 ppm 0.1 0.5 1.0 0.5 1.0
NH3 0-100 ppm 1.0 25 35 25 50
ClO2 0-1.00 ppm 0.01 0.1 0.3 0.1 0.3
O3 0-1.00 ppm 0.01 0.1 0.1 0.1 0.1
PID 0-1000 ppm 1.0 50 100 50 100
LEL 0-100% LEL
0-5% v/v1% High Range
ResolutionN/A N/A 10% 20%
Auto zero: Yes
Calibration: Automatic
Backlight: Standard Mode: low light (auto); alarm condition (auto); reactivation on demand
Alarm Only Mode: alarm condition (auto); reactivation on demand
Stealth Mode: backlight inactive; audible and visual alarms silenced
Sensor: PID (VOC) 10.6 eV lamp; plug-in, electrochemical cell (toxic and oxygen); catalytic (LEL)
User field options: Confidence beep, set STEL period, force calibration on start-up, enable / disable
sensor, pass code protection, latching alarms, span level, passed calibration user-
lockout option, SAFE display function, stealth mode, fast pump, oxygen auto-cal onstart-up, alarm only backlight mode, combustible gas measurement (% LEL or % by
volume methane), LCD language option, STEL calculation choice, TWA calculation
choice, combustible correction factor options, user-settable calibration gas level, clock
ALARM INDICATORS: Clearlyadvises alarm conditions with audible, visual, and vibrator alarms Visual alarms: Two flashing alarm bars visible from all angles
LCD indicates gas present and alarm levels encountered
Audible alarm: Typically 95 dB at 1 ft. / 30 cm variable pulsed audible alarm
Vibrator alarm: Pulses warning in gas alarm conditions, or for any status alarms
Status alarms: Low battery, over range, multiple gas / pump and failed or missing sensor
ALARM LEVELS: Instantaneous Low Alarm for all gases; Instantaneous High Alarm for all gases;TWA (time-weighted average) and STEL (short-term exposure limit) for toxics/VOCs
LCD: Continuous,alphanumeric gas readout and status display advises:Gases monitored: VOCs (PID), H2S, CO, O2, SO2, PH3, NO2, NH3, HCN, Cl2, ClO2, O3 and/or
combustibles (0-100% LEL or 0-5.0% methane)
Gas concentration: Simultaneously and continuously displays gas concentration(s) for all sensors
(in ppm, % LEL, or % v/v)
Peak (max)/STEL/
TWA ppm or %: Records exposures and displays on demand
Alarm setpoints: Displayed on activation and on demand
Calibration due date: In days remaining (on start-up)
Clock: Real-time clock setting displayed on start-up
TESTS:Full function self-test: Sensor integrity, circuitry, battery and audible/visual alarms on activation
Battery: Continuous
GENERAL SPECIFICATIONS: Rugged, composite material; two-shot molded polymer case with integral anti-shock bootOperating temperature: 14 to 104oF / -10 to 40oC VOC (PID); -4 to 122oF / -20 to 50oC other gases
Humidity: 0 - 95% RH (non-condensing)
EMI/RFI: Complies with EMC Directive 89/336/EEC
Protection: Highly water resistant and dust proof housing
Battery Options: 12 hours on three AA alkaline cells or 10 hours on rechargeable hot-swappable
battery pack when equipped with PID, oxygen, combustible and TwinTox
(H2S/CO) sensor
Size: 5.7 x 2.9 x 1.5 in. / 14.5 x 7.4 x 3.8 cm
Weight: 13.1 oz. / 370 g
WARRANTY: Full 2-year warranty including all sensors (1 year PID lamp and NH3 sensor) DATALOGGER OPTIONS: User-downloadable datalogger
Data recorder: All events and occurrences (calibration zero adjustment, calibration span adjustment,
gas readings, TWA, STEL, alarm condition, failures detected by the microprocessor).
Sampling rate: Factory set at 5-second intervals and user-settable from 1 to 127-second intervals
Storage: 2 months of continuous data at 5 second intervals via 32 MB Flash MMC card
(64 MB and 128 MB cards also available); when full, the wrap-around feature
replaces oldest data with most recent data
GasAlertMicro 5 PIDVOCs (PID), H2S, CO, O2,
SO2, PH3, NH3, NO2, HCN, Cl2,
ClO2, O3, and Combustibles
ve-Gas Detector
asAlertMicro5 PID portable gas
tector simultaneously monitors up to
e atmospheric hazards including volatile
ganic compounds (VOCs), oxygen,
mbustible gas (% LEL), and a wideection of individual electrochemical
xic sensors. Compact and lightweight,
asAlertMicro5 PID activates audible,
ual and vibrating alarms in the event of
ow, high, TWA or STEL alarm condition.
OCs present multiple potential threats
the workplace environment. VOCs
e characterized by their tendency to
aporate easily at room temperature
d include substances such as solvents,
nt thinners, benzene, butadiene,
xane, toluene, as well as vapors
sociated with fuels such as gasoline,sel, heating oil, kerosene, and jet fuel.
e GasAlertMicro5 PIDs integrated
otoionization detector (PID) enables
nsitive broad spectrum monitoring of
atile organic contaminants.
ghly configurable to suit a range of
plications, GasAlertMicro5 PID comes
a diffusion or pumped instrument; with
her alkaline or NiMH rechargeable
t-swappable battery pack and has
tional datalogging capabilities. Standard
atures include confidence beep,
tomatic calibration, full function self-t, and an extended selection of user-
ttable field options.
BW1105-01-01-00-8.25x11-20050503-5789-3
DUE TO ONGOING RESEARCH AND PRODUCT IMPROVEMENT, SPECIFICATIONS ARE SUBJECT TO CHANGE WITHOUT NOTICE.
RATINGSAND CERTIFICATIONS:
(Intrinsic safety)
ATEX: X g II 1 G T4**
EEx ia IIC T4**
IECEx: Ex ia IIC T4**
**in progress
n Class I, Div. 1, Gr. A, B, C, DAmerican Bureau of Shipping**rua Rggio Nbrega, 129 Campinas, SP 13025-310
Tel (19) 3794-2900 Fax (19) 3794-2919
www.clean.com.br [email protected]
ribuidor no Brasil: