MDD 93/42/EEC May 1999 8200590 Quint-7000 User's Guide We reserve the right to make changes. Manufactured by: FINNDENT OY Teollisuustie 5 FIN-07230 MONNINKYLÄ Tel. +358 19 5221500 Fax +358 19 643457 Dental Unit Dental Unit Dental Unit Dental Unit Dental Unit User's Guide User's Guide User's Guide User's Guide User's Guide

Upload: others

Post on 27-Mar-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

I

MDD93/42/EEC

May 19998200590 Quint-7000 User's Guide

We reserve the right to make changes.

Manufactured by:FINNDENT OY

Teollisuustie 5FIN-07230 MONNINKYLÄ

Tel. +358 19 5221500Fax +358 19 643457

Dental UnitDental UnitDental UnitDental UnitDental UnitUser's GuideUser's GuideUser's GuideUser's GuideUser's Guide

Administrator
Tel. +358 (0)207 43 5115 Fax. +358 (0)207 43 5116 Email:[email protected]
Administrator
Kenkäpellontie 4 FIN-04600 MÄNTSÄLÄ
Page 2: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

II

Page 3: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

III

Contents

1. Introduction1. Introduction1. Introduction1. Introduction1. Introduction .............................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................. 11111Scope of this manual ..........................................................................................................1Associated documents .......................................................................................................1Warnings and precautions ..................................................................................................2

2. The main parts2. The main parts2. The main parts2. The main parts2. The main parts ..................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................... 33333Quint 7000 A1 Dental Unit ................................................................................................3Quint 7000 A2 Dental Unit ................................................................................................4Quint 7000 E2 Dental Unit ................................................................................................4Dentist's Control Panel ......................................................................................................5Assistant's Control Panel ...................................................................................................5The foot control .................................................................................................................6

3. Using the dental unit3. Using the dental unit3. Using the dental unit3. Using the dental unit3. Using the dental unit ................................................................................................................................................................................................................................................................................................................................................................................................................................................... 77777Switching the dental unit on ...............................................................................................7Switching the dental unit off ..............................................................................................7Rotating the cuspidor .........................................................................................................8Positioning the instrument bridge .......................................................................................8Positioning the instrument cart ...........................................................................................9Using the instruments .......................................................................................................10

The Micromotor ..........................................................................................................10The Turbine.................................................................................................................12The Scaler...................................................................................................................13The Polymerisation light (Satelec)..............................................................................15The Polymerisation light (Lysta) .................................................................................15The Syringe - 3-way ...................................................................................................16The Syringe - 6-way ...................................................................................................16

Dentist's control panel .....................................................................................................17Water bottle .....................................................................................................................18Tray holder ......................................................................................................................18The Suction Arm ..............................................................................................................18Positioning the suction arm ..............................................................................................19

Using the suction nozzles ............................................................................................19The Syringe .................................................................................................................20Suction arm controls ...................................................................................................21

The Dental Light ..............................................................................................................22

Page 4: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

IV

4. Cleaning, disinfecting and sterilising the unit4. Cleaning, disinfecting and sterilising the unit4. Cleaning, disinfecting and sterilising the unit4. Cleaning, disinfecting and sterilising the unit4. Cleaning, disinfecting and sterilising the unit ................................................................................................................................................................................................................................. 2323232323Suction system ............................................................................................................23Suction hoses ..............................................................................................................23Instruments and suction nozzles...................................................................................24Unit surfaces ...............................................................................................................24

5. Maintenance5. Maintenance5. Maintenance5. Maintenance5. Maintenance ............................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................... 2727272727Oil collector ...............................................................................................................27Tightening instrument holders .....................................................................................27Instrument water adjustment ........................................................................................28

6. 6. 6. 6. 6. ServiceServiceServiceServiceService................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................. 2929292929

7. Approved accessories7. Approved accessories7. Approved accessories7. Approved accessories7. Approved accessories ......................................................................................................................................................................................................................................................................................................................................................................................................................................... 3030303030

8. T8. T8. T8. T8. Technical Specificationsechnical Specificationsechnical Specificationsechnical Specificationsechnical Specifications ..................................................................................................................................................................................................................................................................................................................................................................................................................... 3232323232Symbols and abbreviations ..............................................................................................33Dimensions ......................................................................................................................34

Page 5: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

Quint 7000 Dental Unit 1

1. Introduction1. Introduction1. Introduction1. Introduction1. Introduction

Scope of this manualScope of this manualScope of this manualScope of this manualScope of this manualThis manual describes how to use and look after alldental units in the FD 7000 range. The rangeincludes the following versions:- 7000A1

Chair mounted cuspidor, whip arminstruments

- 7000A2Chair mounted cuspidor, hanging tubeinstruments

- 7000B1Floor mounted cuspidor, whip arm instruments

- 7000B2Floor mounted cuspidor, hanging tubeinstruments

- 7000C2No cuspidor, instrument cart with hangingtube instruments

- 7000D2Chair mounted cuspidor, instrument cart withhanging tube instruments

- 7000E2Floor mounted cuspidor, instrument cart withhanging tube instruments

Operating differences between the differentversions of the unit are described whereappropriate.

Page 6: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

2 User's Manual

Associated documentsAssociated documentsAssociated documentsAssociated documentsAssociated documentsUser's manual supplied with the amalgamseparating system that is installed in the dentalunit.

User's manuals for the instruments that aresupplied with the dental unit.

WWWWWarnings and precautionsarnings and precautionsarnings and precautionsarnings and precautionsarnings and precautions• Read this user's manual and the associated

documentation before using the dental unit orany of its instruments or accessories. Makesure that you know how to use the dental unitbefore starting to use it.

• Do not let children or unauthorized people playwith or use the unit or any of its instruments oraccessories.

• Only use instruments and accessories listed inthis manual with the unit.

• Only carry out the maintenance procedureslisted in this manual. All other maintenanceMUST be carried out by authorized servicepersonnel.

• For the dental unit to function efficiently itmust be regularly serviced.

• If the quality of the water being used with theunit is poor, an additional water filter mayhave to be installed. (Refer to the installationmanual)

• The compressed air being used with the unitmust be dry and oil free. The air compressormust be fitted with an air dryer and/orcondensation valve. (Refer to the installationmanual)

Page 7: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

Quint 7000 Dental Unit 3

2. The main parts2. The main parts2. The main parts2. The main parts2. The main parts

Quint 7000 Quint 7000 Quint 7000 Quint 7000 Quint 7000 A1 Dental UnitA1 Dental UnitA1 Dental UnitA1 Dental UnitA1 Dental UnitChair mounted cuspidor with whip arm instruments

NOTEWith the dental units Quint 7000 BI, 7000 B2 and 7000 E2the cuspidor is attached to the floor box via a metal tube.

Instrument arm

Floor box

Instrument tray

Dental light

Cuspidor

Foot control

Suction arm andassistant's control panel

Instrument bridge and dentist'scontrol panel

Water bottle(Option)

Page 8: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

4 User's Manual

Quint 7000 Quint 7000 Quint 7000 Quint 7000 Quint 7000 A2 Dental UnitA2 Dental UnitA2 Dental UnitA2 Dental UnitA2 Dental UnitChair mounted cuspidor with hanging tube instruments

Quint 7000 E2 Dental UnitQuint 7000 E2 Dental UnitQuint 7000 E2 Dental UnitQuint 7000 E2 Dental UnitQuint 7000 E2 Dental Unit Floor mounted cuspidor with instrument cart

Hanging tube instruments

Instrument cart with hanging tube instruments

Metal tube

Page 9: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

Quint 7000 Dental Unit 5

Dentist's Control PanelDentist's Control PanelDentist's Control PanelDentist's Control PanelDentist's Control Panel

Assistant's Control PanelAssistant's Control PanelAssistant's Control PanelAssistant's Control PanelAssistant's Control Panel

Glass fill key

Bowl rinse key

Door lock/Nurse call key

Turbine spray indicator

Accessory key

Bowl rinse key

Glass fill key

Turbine spray key

Door lock/Nurse call key

Scaler power key

Reverse rotation indicator

Micromotor spray indicator

Scaler power indicators

Page 10: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

6 User's Manual

The foot controlThe foot controlThe foot controlThe foot controlThe foot control

Option switch (For example - nurse call)

Reverse direction switchFoot leverInstrument on/off switchInstrument speed/power controlChip blow switch

Micromotor spray switchLifting handle

Page 11: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

Quint 7000 Dental Unit 7

3. Using the dental unit3. Using the dental unit3. Using the dental unit3. Using the dental unit3. Using the dental unit

Switching the dental unit onSwitching the dental unit onSwitching the dental unit onSwitching the dental unit onSwitching the dental unit onSwitch the dental unit on with the ON/OFF switchthat is located on the side of the floor box. Whenyou switch the dental unit on you will hear anaudible signal and the switch light will come on.

NoteNoteNoteNoteNoteIf the dental chair is connected to the power supplyvia the dental unit (i.e. the dental chair does nothave its own power supply), the dental chair willautomatically come on when the unit is switchedon.Note that the chair's ON/OFF switch must be in theON position for this to happen.

Switching the dental unit offSwitching the dental unit offSwitching the dental unit offSwitching the dental unit offSwitching the dental unit offSwitch the dental unit off with the same switch.When the unit is switched off the switch light willgo out.ALWAYS switch the unit off at the end of each workday and at weekends.

Off

On

Page 12: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

8 User's Manual

Rotating the cuspidorRotating the cuspidorRotating the cuspidorRotating the cuspidorRotating the cuspidorThe cuspidor can be rotated approximately 80°away from the chair. This may be necessary if theassistant needs additional working space whenspecial dental procedures are being carried out.

Positioning the instrument bridgePositioning the instrument bridgePositioning the instrument bridgePositioning the instrument bridgePositioning the instrument bridgeMove the instrument bridge in the requiredposition with the handles on the sides of the bridge.The height of the instrument bridge can beadjusted by lifting it up or pushing it down.If you wish to lock the instrument bridge at aparticular height turn the height locking knobclockwise.To unlock the bridge turn the knob anticlockwise.

Cuspidor

Height locking knob

Handle

Page 13: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

Quint 7000 Dental Unit 9

If the instrument bridge "drifts" out of position, thefriction of the arm joints needs to be increased.Contact your service technician for assistance.

Positioning the instrument cartPositioning the instrument cartPositioning the instrument cartPositioning the instrument cartPositioning the instrument cartMove the instrument cart in the required positionwith the handles on the side of the instrumentbridge.

Instrument cart

Handle

Page 14: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

10 User's Manual

Using the instrumentsUsing the instrumentsUsing the instrumentsUsing the instrumentsUsing the instrumentsWhen an instrument is lifted from its holder itbecomes the "active" instrument and no otherinstrument, except the syringe and polymerisationlight, will operate while the active instrument is inuse.

The MicromotorThe MicromotorThe MicromotorThe MicromotorThe Micromotor

Switching it on and offSwitching it on and offSwitching it on and offSwitching it on and offSwitching it on and offLift the micromotor from its holder and press andhold down the foot lever. The micromotor will start.To stop the micromotor, remove your foot from thefoot lever.

Adjusting the speedAdjusting the speedAdjusting the speedAdjusting the speedAdjusting the speedSlide the foot lever to the right to increase thespeed of the micrometer.Slide the foot lever to the left to decrease the speed.

Page 15: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

Quint 7000 Dental Unit 11

Activating the sprayActivating the sprayActivating the sprayActivating the sprayActivating the sprayPress the right-hand foot switch once to switch theinstrument spray on. When the spray is on thegreen M light on the dentist's display will come on.Press the right-hand foot switch a second time toswitch the spray off. The green light will go out.

Chip blowerChip blowerChip blowerChip blowerChip blowerDouble press and hold down the foot lever. Chipblow will start. To stop chip blow remove your footfrom the foot switch.

Reversing the rotationReversing the rotationReversing the rotationReversing the rotationReversing the rotationPress the left-hand foot switch down once to reversethe direction of rotation of the micromotor (anti-clockwise). The red M light on the dentist's displaywill come on when the micromotor is set to rotate inthe reverse direction.Press the foot switch a second time to switch thedirection of rotation of the micromotor back to thenormal (clockwise) direction. The red M light will goout.

Page 16: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

12 User's Manual

The TThe TThe TThe TThe Turbineurbineurbineurbineurbine

Switching it on and offSwitching it on and offSwitching it on and offSwitching it on and offSwitching it on and offLift the turbine from its holder and press and holddown the foot lever. The T light will come on andthe turbine and spray will start. To stop the turbineremove your foot from the foot lever. The T lightwill go out.Note that there is no speed control with the turbine.

Chip blowerChip blowerChip blowerChip blowerChip blowerDouble press and hold down the foot lever. Chipblow will start. To stop chip blow remove your footfrom the foot switch.

Page 17: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

Quint 7000 Dental Unit 13

The ScalerThe ScalerThe ScalerThe ScalerThe Scaler

Switching it on and offSwitching it on and offSwitching it on and offSwitching it on and offSwitching it on and offLift the scaler from its holder and press and holddown the foot lever. The scaler will start. To stop thescaler, remove your foot from the foot lever.

Selecting the power levelSelecting the power levelSelecting the power levelSelecting the power levelSelecting the power levelThe maximum power level of the scaler can be setat either 100%, 50% or 30%. The lights next to theSP key indicate what power level has been selected.To change the power levels press the SP keyrepeatedly until the light combination for thepower level you require appears. The power level/light combinations are as follows:- 100% power = no lights on- 50% power = top light on- 30% power = bottom light on

Note that the scaler power level will automaticallyreturn to the 100% level, i.e. no lights on, wheneverthe unit is switched off and then on again.

No lights Top light Bottom light 100% 50% 30%

Page 18: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

14 User's Manual

Adjusting the power levelAdjusting the power levelAdjusting the power levelAdjusting the power levelAdjusting the power levelTo increase the power of the scaler slide the footlever to the right to decrease the power slide thefoot lever to the left.Note that when the foot control is moved to themaximum power position, the maximum powerlevel will be the level selected using the SP key.For example, if you selected the 50% power levelthe power level can adjusted from 0% to amaximum of 50%.

Page 19: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

Quint 7000 Dental Unit 15

The Polymerisation light (Satelec)The Polymerisation light (Satelec)The Polymerisation light (Satelec)The Polymerisation light (Satelec)The Polymerisation light (Satelec)Lift the polymerisation light from its holder andpress the button on the side of the instrument. Thelight will come on. The light will automatically goout after 30 seconds. During the time the light is onyou will hear a "beep" every ten seconds. If youwish to turn the light out before the 30 secondselapses, press the button a second time.

For more information on how to operate and servicethe Satelec polymerisation light refer to thedocumentation supplied with the light.

The Polymerisation light (LThe Polymerisation light (LThe Polymerisation light (LThe Polymerisation light (LThe Polymerisation light (Lysta)ysta)ysta)ysta)ysta)Lift the polymerisation light from its holder andpress the button on the top of the instrument. Thelight will come on. The light will automatically goout after 35 seconds. After seventeen seconds youwill hear a beep. If you wish to turn the light outbefore the 35 seconds elapses, press the button asecond time.

For more information on how to operate and servicethe Lysta polymerisation light refer to thedocumentation supplied with the light.

Activation switch

Page 20: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

16 User's Manual

Water

Spray

Air

The Syringe - 3-wayThe Syringe - 3-wayThe Syringe - 3-wayThe Syringe - 3-wayThe Syringe - 3-wayPress the left-hand button to activate the waterspray. Press the right-hand button to activate theair. Press both buttons together to activate the air/water spray.

The Syringe - 6-wayThe Syringe - 6-wayThe Syringe - 6-wayThe Syringe - 6-wayThe Syringe - 6-wayThere is a heater in this syringe that allows you tohave either hot air and water or cold. To switch theheater on twist the collar at the end of the handleanticlockwise - the indicator spot will be green. Toswitch the heater off twist the collar at the end ofthe handle clockwise - the indicator spot will beblack.Press the left-hand button to activate the waterspray. Press the right-hand button to activate theair. Press both buttons together to activate the air/water spray.

Water

Spray

Air

Heater on

Heater off

Page 21: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

Quint 7000 Dental Unit 17

Dentist's control panelDentist's control panelDentist's control panelDentist's control panelDentist's control panel

Bowl rinse - on/offBowl rinse - on/offBowl rinse - on/offBowl rinse - on/offBowl rinse - on/offBriefly press the bowl rinse key on the dentist'scontrol panel to switch the bowl rinse on. Rinsingwill automatically stop after the preset rinseduration has elapsed. If you wish to stop the bowlrinse before the preset rinse time has elapsed, pressthe bowl rinse key a second time.

Bowl rinse - setting the rinse durationBowl rinse - setting the rinse durationBowl rinse - setting the rinse durationBowl rinse - setting the rinse durationBowl rinse - setting the rinse durationPress and hold down the bowl rinse key for theamount of time that you wish the bowl rinse tooperate (minimum one second, maximum tenseconds). Now, when the bowl rinse key is pressed torinse the bowl it will remain on for the sameamount of time as the bowl rinse key was helddown.

Glass fill - on/offGlass fill - on/offGlass fill - on/offGlass fill - on/offGlass fill - on/offBriefly press the glass fill key on the dentist'scontrol panel to switch the glass fill on. Filling willautomatically stop after the preset fill duration haselapsed. If you wish to stop the glass fill before thepreset fill duration has elapsed, press the glass fillkey a second time.

Glass fill - setting the fill durationGlass fill - setting the fill durationGlass fill - setting the fill durationGlass fill - setting the fill durationGlass fill - setting the fill durationPress and hold down the glass fill key for theamount of time that you wish the glass fill tooperate (minimum one second, maximum tenseconds). Now, when the glass fill key is pressed tofill the glass it will remain on for the same amountof time as the glass fill key was held down.

Door lock (Option)Door lock (Option)Door lock (Option)Door lock (Option)Door lock (Option)If a remote door lock is in use and has beenconnected to the dental unit this key is used toactivate it.

Auxiliary (Aux) keyAuxiliary (Aux) keyAuxiliary (Aux) keyAuxiliary (Aux) keyAuxiliary (Aux) keyIf any other equipment is being used with the unitthis key can be used to activate it.

Page 22: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

18 User's Manual

WWWWWater bottleater bottleater bottleater bottleater bottleTo remove and refill the water bottle unscrew it.

TTTTTray holderray holderray holderray holderray holderThe tray holder is attached to a tray holder armwhich is attached to the underside of the instru-ment bridge on either the left- or the right-handside. The arm is angled allowing the tray to bepositioned either high or low depending on whichway round the arm is attached.To position the tray holder loosen the two tray armknobs, move the tray holder to the required posi-tion and then tighten the two knobs.

The tray holder will hold one standard size tray ortwo half size trays. A double size tray holder isavailable as an option.

Instrument bridge

Tray holder

Tray holder arm

Attaching knobs

Water bottle

Page 23: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

Quint 7000 Dental Unit 19

The Suction The Suction The Suction The Suction The Suction ArmArmArmArmArm

Positioning the suction armPositioning the suction armPositioning the suction armPositioning the suction armPositioning the suction armThe height of the suction arm can be adjusted bysqueezing the locking handle, on the underside ofthe top part of the suction arm, and pulling thesuction arm up or pushing it down. When it is atthe required height, release the locking handle.

WWWWWarningarningarningarningarningDo not leave the suction arm under the chairbackrest as the arm could be damaged if you lowerthe chair.

Using the suction nozzlesUsing the suction nozzlesUsing the suction nozzlesUsing the suction nozzlesUsing the suction nozzles

Starting and stopping the suctionStarting and stopping the suctionStarting and stopping the suctionStarting and stopping the suctionStarting and stopping the suctionWhen a suction nozzle is lifted from its holdersuction will automatically start. When the suctionnozzle is replaced suction will stop.

Suction arm

Locking handle

Page 24: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

20 User's Manual

Adjusting the suctionAdjusting the suctionAdjusting the suctionAdjusting the suctionAdjusting the suctionSlide the knob on the side of the nozzle upwards toincrease the suction and downwards to decrease thesuction.

Suction tipSuction tipSuction tipSuction tipSuction tipTo attach the suction tip push it into the opening atthe end of the suction nozzle.

The SyringeThe SyringeThe SyringeThe SyringeThe SyringePress the green button to activate the water spray.Press the blue button to activate the air. Pressboth buttons together to activate the air/waterspray.

Water

Spray

Air

Suction nozzle

Knob

Increase

Decrease

Page 25: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

Quint 7000 Dental Unit 21

Suction arm controlsSuction arm controlsSuction arm controlsSuction arm controlsSuction arm controls

Bowl rinse - on/offBowl rinse - on/offBowl rinse - on/offBowl rinse - on/offBowl rinse - on/offBriefly press the bowl rinse key on the assistant'scontrol panel to switch the bowl rinse on. Rinsingwill automatically stop after the preset rinseduration has elapsed. If you wish to stop the bowlrinse before the preset rinse time has elapsed, pressthe bowl rinse key a second time.

Bowl rinse - setting the rinse durationBowl rinse - setting the rinse durationBowl rinse - setting the rinse durationBowl rinse - setting the rinse durationBowl rinse - setting the rinse durationPress and hold down the bowl rinse key for theamount of time that you wish the bowl rinse tooperate (minimum one second, maximum tenseconds). Now, when the bowl rinse key is pressed torinse the bowl it will remain on for the sameamount of time as the bowl rinse key was helddown.

Glass fill - on/offGlass fill - on/offGlass fill - on/offGlass fill - on/offGlass fill - on/offBriefly press the glass fill key on the assistant'scontrol panel to switch the glass fill on. Filling willautomatically stop after the preset fill duration haselapsed. If you wish to stop the glass fill before thepreset fill duration has elapsed, press the glass fillkey a second time.

Glass fill - setting the fill durationGlass fill - setting the fill durationGlass fill - setting the fill durationGlass fill - setting the fill durationGlass fill - setting the fill durationPress and hold down the glass fill key for theamount of time that you wish the glass fill tooperate (minimum one second, maximum tenseconds). Now, when the glass fill key is pressed tofill the glass it will remain on for the same amountof time as the glass fill key was held down.

Door lock (Option)Door lock (Option)Door lock (Option)Door lock (Option)Door lock (Option)If a remote door lock is in use and has beenconnected to the dental unit this key is used toactivate it.

Page 26: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

22 User's Manual

The Dental LightThe Dental LightThe Dental LightThe Dental LightThe Dental Light

Switching it on and offSwitching it on and offSwitching it on and offSwitching it on and offSwitching it on and offThe switch on the underside of the arm that holdsthe dental light is used to switch the light on andoff.

Adjusting the light intensityAdjusting the light intensityAdjusting the light intensityAdjusting the light intensityAdjusting the light intensity.....The switch on the underside of the dental light isused to adjust the intensity of the light.

For more information on using and maintainingthe dental light, refer to the documentationsupplied with the light.

On/off switch

Intensity switch

Page 27: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

Quint 7000 Dental Unit 23

4. Cleaning, disinfecting and sterilising the unit4. Cleaning, disinfecting and sterilising the unit4. Cleaning, disinfecting and sterilising the unit4. Cleaning, disinfecting and sterilising the unit4. Cleaning, disinfecting and sterilising the unit

Suction systemSuction systemSuction systemSuction systemSuction systemFor information on how to clean the suction system,refer to the documentation supplied with thesystem

Suction hosesSuction hosesSuction hosesSuction hosesSuction hosesAfter every patient allow one glass of water to besucked through each suction hose.

Twice a day, midday and evening, allow one glass ofsuction hose cleaning agent to be sucked througheach hose. Make sure that you prepare and use it inaccordance with the manufacturer's instructions.

Twice a day, midday and evening, empty and rinsethe suction hose filters. The suction hose filters arelocated inside the connectors that connect thesuction tubes to the cuspidor. To empty the filter,pull the suction hose connector from the cuspidor.Lift the filter out and rinse it. Also empty anydebris that is inside the connector. Replace thefilter and reconnect the suction hose connector tothe cuspidor.

Cuspidor

Suction hoses

Suction hose filter

Connector

Page 28: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

24 User's Manual

Instruments and suction nozzlesInstruments and suction nozzlesInstruments and suction nozzlesInstruments and suction nozzlesInstruments and suction nozzlesFor information on how to clean and sterilize theinstruments, refer to the documentation suppliedwith them.

CautionCautionCautionCautionCautionSwitch the dental unit off before you remove anyinstruments. Press the syringe air and waterbuttons to relieve any pressure before removing thesyringe.

Suction nozzles can be disconnected from the hosesand autoclaved.

Unit surfacesUnit surfacesUnit surfacesUnit surfacesUnit surfacesUnit surfaces must cleaned and disinfected at leastonce a day with a suitable cleaning and disinfectingagents. Make sure that you prepare and use themin accordance with the manufactures instructions.Do not uses abrasive cleaning agents.

Note that the rubber instrument support can beremoved from the instrument bridge andautoclaved.

Instrument support

Instrument bridge

Page 29: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

Quint 7000 Dental Unit 25

The rollers and holders at the end of the whip armscan be removed so that the instrument tubes can becleaned.To remove the roller and holder slide the holder upslightly until the roller can be removed, and thenremove the holder.

When replacing the roller and holder note that thetwo flats on the roller axle must engage in the slotin the holder, so that the holes line up, before theroller and holder can be slide back onto the whiparm.

On units that are fitted with hanging tubeinstruments the instrument holders can beremoved.To do this tilt the instrument holder forward as faras it will go, and then lift the holder up slightlyuntil it disengages from the holding pin. The holdercan then be removed.

Slot

Flat

Roller

Holder

Whip arms

Flat

Roller axle

Instrument holderTilt forward Lift and remove

Page 30: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

26 User's Manual

To replace the instrument holder place the instru-ment holder so that the holding pin in theinstrument bride goes into the hole in the holderbracket. Push the instrument holder down to lock itin position.

Holder bracket

Holding pinInstrument holder

Hole

Page 31: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

Quint 7000 Dental Unit 27

5. Maintenance5. Maintenance5. Maintenance5. Maintenance5. Maintenance

Oil collectorOil collectorOil collectorOil collectorOil collectorThe cotton pad in the turbine oil collector must bereplace monthly or more frequently if free oil startsto appear. The oil collector on the underside of theinstrument bridge. To remove and empty the oilcollector unscrew it.

Tightening iTightening iTightening iTightening iTightening instrument holdersnstrument holdersnstrument holdersnstrument holdersnstrument holdersOn units that are fitted with hanging tubeinstruments the instrument holders can tilted tomake it easier to remove and replace instruments.If a holder becomes loose it can be tightened withscrews on the side of the holder.

Instrument bridge

Oil collector

Screws

Instrument holder

Page 32: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

28 User's Manual

Instrument water adjustmentInstrument water adjustmentInstrument water adjustmentInstrument water adjustmentInstrument water adjustmentThe supply of water to the instruments can beadjusted. This is done by turning the regulatorscrews located on the side of the instrument bridgewith a small screwdriver. The side on which the regulators are located willdepend on whether the unit is left- or right-handed.

Micromotor 1 water Syringe water Micromotor 2 water Scaler water

Turbine water

Water adjustment regulators

Page 33: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

Quint 7000 Dental Unit 29

6. ServiceServiceServiceServiceServiceEvery six months a FINNDENT approved servicetechnician must carry out the following inspectionand service tasks:• check the output pressures of the air and water

control valves• check the air pressure of the instruments• check to see if the filters in the air and water

control valves need replacing• check the operation of the instruments and if

necessary replace and instrument O-rings orseals

CautionCautionCautionCautionCautionSwitch the dental unit off before you removeany instruments. Press the syringe air andwater buttons to relieve any pressure beforeremoving the syringe.

Page 34: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

30 User's Manual

7. 7. 7. 7. 7. Approved accessoriesApproved accessoriesApproved accessoriesApproved accessoriesApproved accessoriesThe following list contains instruments that can beused with the Quint 7000 dental unit. Some of thelisted instruments are not part of the Finndentproduct range.

Micromotor drillsMicromotor drillsMicromotor drillsMicromotor drillsMicromotor drillsKavo Intramatic Lux - Models 190, 192 and 196

Satelec Servotorque - Models IML, IS, ES, AV, S

Quint (Satelec Servotorque IML)

Bien Air - Isolite 100, 300, MC40IR and MC20IR

TTTTTurbine drillsurbine drillsurbine drillsurbine drillsurbine drillsKavo - 630 630B, 640B and 650B

Bien Air - Black Pearl, Ondine, Gyro Long Life andPrestige S30L

W&H Toplight - all models

Air micromotorsAir micromotorsAir micromotorsAir micromotorsAir micromotorsKavo - all models

Bien Air - all models

W&H - all models

Light probesLight probesLight probesLight probesLight probesKavo - Dia-Lux 2300L

W&H - LS 1 Diagnostic Light

Bien - Air Translite

Page 35: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

Quint 7000 Dental Unit 31

TTTTTartar scalersartar scalersartar scalersartar scalersartar scalersSatelec Suprasson SP 4055

Amdent 930

Quint (manufactured by Amdent)

Polymerisation lightsPolymerisation lightsPolymerisation lightsPolymerisation lightsPolymerisation lightsLysta LC-85

Satelec Servolite SL 4055

Satelec Acta

SyringesSyringesSyringesSyringesSyringesQuint 3-way syringe

Luzzani 6-way syringe

WWWWWater heatersater heatersater heatersater heatersater heatersSIE-Dental - water heating element 50/100W, 24V

Amalgam separatorsAmalgam separatorsAmalgam separatorsAmalgam separatorsAmalgam separatorsCattani - Suction separator

Metasys Multisystem Type MST 1

Metasys Multisystem MST 1 ECO

Dental lightsDental lightsDental lightsDental lightsDental lights

Faro - S90, S2000 and SC2000 Forma(230 V or 24 V)

Page 36: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

32 User's Manual

8. T8. T8. T8. T8. Technical Specificationsechnical Specificationsechnical Specificationsechnical Specificationsechnical SpecificationsPROTECTION CLASS: IPROTECTION GRADE: BPROTECTION AGAINST FLUIDS: IPX0

OPERATIONAL SAFETY GRADE:- Standard: The unit is not suitable for use in an

environ ment with a mixture ofinflammable

anaesthesia gas and air, or a mixture ofinflammable anaesthesia gas and oxygen

or laughing gas (nitrous oxide).

MODE OF OPERATION: Continuous operation with short-term load.

ALLOWED OPERATINGCONDITIONS:

- Temperature: 15oC - 40oC- Relative humidity: 45% - 75%

ELECTRICAL FEATURES:- Power supply: 230 V, 50 Hz- Maximum power consumption: 1200 W (includes power consumption of

possible accessories connected to the unit)- Main fuse: 6.3 AT.- Primary fuse of transformer: 2AT

WATER CONNECTION: 3 bar - 6 barparticle size < 10 um

COMPRESSED AIRCONNECTION: 5.5 bar - 8 bar

WEIGHT : 78 kg without accessories90 kg with accessories125 kg in package with accessories

MANUFACTURER:FINNDENT OYTEOLLISUUSTIE 507230 MONNINKYLÄFINLANDTEL. +358 19 - 5221500FAX +358 19 - 643457

Page 37: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

Quint 7000 Dental Unit 33

Symbols and abbreviationsSymbols and abbreviationsSymbols and abbreviationsSymbols and abbreviationsSymbols and abbreviationsv Voltage

A Ampere

W Watt

~~~~~ Alternating current

FI Approved by Fimko

Symbol for protection class B

European Community approval mark

AT Amperes, slow blow fuse

Control symbols are described where appropriate inthis guide.

Page 38: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

34 User's Manual

DimensionsDimensionsDimensionsDimensionsDimensions

Page 39: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

Quint 7000 Dental Unit 35

Page 40: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

36 User's Manual

Page 41: Dental Unit User's Guide Tel. +358 (0)207 43 5115 Fax ...finndent.de/wp-content/uploads/2017/01/FD-7000_User_Manual.pdf · Associated documents .....1 Warnings and precautions

Quint 7000 Dental Unit 37