I
MDD93/42/EEC
May 19998200590 Quint-7000 User's Guide
We reserve the right to make changes.
Manufactured by:FINNDENT OY
Teollisuustie 5FIN-07230 MONNINKYLÄ
Tel. +358 19 5221500Fax +358 19 643457
Dental UnitDental UnitDental UnitDental UnitDental UnitUser's GuideUser's GuideUser's GuideUser's GuideUser's Guide
II
III
Contents
1. Introduction1. Introduction1. Introduction1. Introduction1. Introduction .............................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................. 11111Scope of this manual ..........................................................................................................1Associated documents .......................................................................................................1Warnings and precautions ..................................................................................................2
2. The main parts2. The main parts2. The main parts2. The main parts2. The main parts ..................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................... 33333Quint 7000 A1 Dental Unit ................................................................................................3Quint 7000 A2 Dental Unit ................................................................................................4Quint 7000 E2 Dental Unit ................................................................................................4Dentist's Control Panel ......................................................................................................5Assistant's Control Panel ...................................................................................................5The foot control .................................................................................................................6
3. Using the dental unit3. Using the dental unit3. Using the dental unit3. Using the dental unit3. Using the dental unit ................................................................................................................................................................................................................................................................................................................................................................................................................................................... 77777Switching the dental unit on ...............................................................................................7Switching the dental unit off ..............................................................................................7Rotating the cuspidor .........................................................................................................8Positioning the instrument bridge .......................................................................................8Positioning the instrument cart ...........................................................................................9Using the instruments .......................................................................................................10
The Micromotor ..........................................................................................................10The Turbine.................................................................................................................12The Scaler...................................................................................................................13The Polymerisation light (Satelec)..............................................................................15The Polymerisation light (Lysta) .................................................................................15The Syringe - 3-way ...................................................................................................16The Syringe - 6-way ...................................................................................................16
Dentist's control panel .....................................................................................................17Water bottle .....................................................................................................................18Tray holder ......................................................................................................................18The Suction Arm ..............................................................................................................18Positioning the suction arm ..............................................................................................19
Using the suction nozzles ............................................................................................19The Syringe .................................................................................................................20Suction arm controls ...................................................................................................21
The Dental Light ..............................................................................................................22
IV
4. Cleaning, disinfecting and sterilising the unit4. Cleaning, disinfecting and sterilising the unit4. Cleaning, disinfecting and sterilising the unit4. Cleaning, disinfecting and sterilising the unit4. Cleaning, disinfecting and sterilising the unit ................................................................................................................................................................................................................................. 2323232323Suction system ............................................................................................................23Suction hoses ..............................................................................................................23Instruments and suction nozzles...................................................................................24Unit surfaces ...............................................................................................................24
5. Maintenance5. Maintenance5. Maintenance5. Maintenance5. Maintenance ............................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................... 2727272727Oil collector ...............................................................................................................27Tightening instrument holders .....................................................................................27Instrument water adjustment ........................................................................................28
6. 6. 6. 6. 6. ServiceServiceServiceServiceService................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................. 2929292929
7. Approved accessories7. Approved accessories7. Approved accessories7. Approved accessories7. Approved accessories ......................................................................................................................................................................................................................................................................................................................................................................................................................................... 3030303030
8. T8. T8. T8. T8. Technical Specificationsechnical Specificationsechnical Specificationsechnical Specificationsechnical Specifications ..................................................................................................................................................................................................................................................................................................................................................................................................................... 3232323232Symbols and abbreviations ..............................................................................................33Dimensions ......................................................................................................................34
Quint 7000 Dental Unit 1
1. Introduction1. Introduction1. Introduction1. Introduction1. Introduction
Scope of this manualScope of this manualScope of this manualScope of this manualScope of this manualThis manual describes how to use and look after alldental units in the FD 7000 range. The rangeincludes the following versions:- 7000A1
Chair mounted cuspidor, whip arminstruments
- 7000A2Chair mounted cuspidor, hanging tubeinstruments
- 7000B1Floor mounted cuspidor, whip arm instruments
- 7000B2Floor mounted cuspidor, hanging tubeinstruments
- 7000C2No cuspidor, instrument cart with hangingtube instruments
- 7000D2Chair mounted cuspidor, instrument cart withhanging tube instruments
- 7000E2Floor mounted cuspidor, instrument cart withhanging tube instruments
Operating differences between the differentversions of the unit are described whereappropriate.
2 User's Manual
Associated documentsAssociated documentsAssociated documentsAssociated documentsAssociated documentsUser's manual supplied with the amalgamseparating system that is installed in the dentalunit.
User's manuals for the instruments that aresupplied with the dental unit.
WWWWWarnings and precautionsarnings and precautionsarnings and precautionsarnings and precautionsarnings and precautions• Read this user's manual and the associated
documentation before using the dental unit orany of its instruments or accessories. Makesure that you know how to use the dental unitbefore starting to use it.
• Do not let children or unauthorized people playwith or use the unit or any of its instruments oraccessories.
• Only use instruments and accessories listed inthis manual with the unit.
• Only carry out the maintenance procedureslisted in this manual. All other maintenanceMUST be carried out by authorized servicepersonnel.
• For the dental unit to function efficiently itmust be regularly serviced.
• If the quality of the water being used with theunit is poor, an additional water filter mayhave to be installed. (Refer to the installationmanual)
• The compressed air being used with the unitmust be dry and oil free. The air compressormust be fitted with an air dryer and/orcondensation valve. (Refer to the installationmanual)
Quint 7000 Dental Unit 3
2. The main parts2. The main parts2. The main parts2. The main parts2. The main parts
Quint 7000 Quint 7000 Quint 7000 Quint 7000 Quint 7000 A1 Dental UnitA1 Dental UnitA1 Dental UnitA1 Dental UnitA1 Dental UnitChair mounted cuspidor with whip arm instruments
NOTEWith the dental units Quint 7000 BI, 7000 B2 and 7000 E2the cuspidor is attached to the floor box via a metal tube.
Instrument arm
Floor box
Instrument tray
Dental light
Cuspidor
Foot control
Suction arm andassistant's control panel
Instrument bridge and dentist'scontrol panel
Water bottle(Option)
4 User's Manual
Quint 7000 Quint 7000 Quint 7000 Quint 7000 Quint 7000 A2 Dental UnitA2 Dental UnitA2 Dental UnitA2 Dental UnitA2 Dental UnitChair mounted cuspidor with hanging tube instruments
Quint 7000 E2 Dental UnitQuint 7000 E2 Dental UnitQuint 7000 E2 Dental UnitQuint 7000 E2 Dental UnitQuint 7000 E2 Dental Unit Floor mounted cuspidor with instrument cart
Hanging tube instruments
Instrument cart with hanging tube instruments
Metal tube
Quint 7000 Dental Unit 5
Dentist's Control PanelDentist's Control PanelDentist's Control PanelDentist's Control PanelDentist's Control Panel
Assistant's Control PanelAssistant's Control PanelAssistant's Control PanelAssistant's Control PanelAssistant's Control Panel
Glass fill key
Bowl rinse key
Door lock/Nurse call key
Turbine spray indicator
Accessory key
Bowl rinse key
Glass fill key
Turbine spray key
Door lock/Nurse call key
Scaler power key
Reverse rotation indicator
Micromotor spray indicator
Scaler power indicators
6 User's Manual
The foot controlThe foot controlThe foot controlThe foot controlThe foot control
Option switch (For example - nurse call)
Reverse direction switchFoot leverInstrument on/off switchInstrument speed/power controlChip blow switch
Micromotor spray switchLifting handle
Quint 7000 Dental Unit 7
3. Using the dental unit3. Using the dental unit3. Using the dental unit3. Using the dental unit3. Using the dental unit
Switching the dental unit onSwitching the dental unit onSwitching the dental unit onSwitching the dental unit onSwitching the dental unit onSwitch the dental unit on with the ON/OFF switchthat is located on the side of the floor box. Whenyou switch the dental unit on you will hear anaudible signal and the switch light will come on.
NoteNoteNoteNoteNoteIf the dental chair is connected to the power supplyvia the dental unit (i.e. the dental chair does nothave its own power supply), the dental chair willautomatically come on when the unit is switchedon.Note that the chair's ON/OFF switch must be in theON position for this to happen.
Switching the dental unit offSwitching the dental unit offSwitching the dental unit offSwitching the dental unit offSwitching the dental unit offSwitch the dental unit off with the same switch.When the unit is switched off the switch light willgo out.ALWAYS switch the unit off at the end of each workday and at weekends.
Off
On
8 User's Manual
Rotating the cuspidorRotating the cuspidorRotating the cuspidorRotating the cuspidorRotating the cuspidorThe cuspidor can be rotated approximately 80°away from the chair. This may be necessary if theassistant needs additional working space whenspecial dental procedures are being carried out.
Positioning the instrument bridgePositioning the instrument bridgePositioning the instrument bridgePositioning the instrument bridgePositioning the instrument bridgeMove the instrument bridge in the requiredposition with the handles on the sides of the bridge.The height of the instrument bridge can beadjusted by lifting it up or pushing it down.If you wish to lock the instrument bridge at aparticular height turn the height locking knobclockwise.To unlock the bridge turn the knob anticlockwise.
Cuspidor
Height locking knob
Handle
Quint 7000 Dental Unit 9
If the instrument bridge "drifts" out of position, thefriction of the arm joints needs to be increased.Contact your service technician for assistance.
Positioning the instrument cartPositioning the instrument cartPositioning the instrument cartPositioning the instrument cartPositioning the instrument cartMove the instrument cart in the required positionwith the handles on the side of the instrumentbridge.
Instrument cart
Handle
10 User's Manual
Using the instrumentsUsing the instrumentsUsing the instrumentsUsing the instrumentsUsing the instrumentsWhen an instrument is lifted from its holder itbecomes the "active" instrument and no otherinstrument, except the syringe and polymerisationlight, will operate while the active instrument is inuse.
The MicromotorThe MicromotorThe MicromotorThe MicromotorThe Micromotor
Switching it on and offSwitching it on and offSwitching it on and offSwitching it on and offSwitching it on and offLift the micromotor from its holder and press andhold down the foot lever. The micromotor will start.To stop the micromotor, remove your foot from thefoot lever.
Adjusting the speedAdjusting the speedAdjusting the speedAdjusting the speedAdjusting the speedSlide the foot lever to the right to increase thespeed of the micrometer.Slide the foot lever to the left to decrease the speed.
Quint 7000 Dental Unit 11
Activating the sprayActivating the sprayActivating the sprayActivating the sprayActivating the sprayPress the right-hand foot switch once to switch theinstrument spray on. When the spray is on thegreen M light on the dentist's display will come on.Press the right-hand foot switch a second time toswitch the spray off. The green light will go out.
Chip blowerChip blowerChip blowerChip blowerChip blowerDouble press and hold down the foot lever. Chipblow will start. To stop chip blow remove your footfrom the foot switch.
Reversing the rotationReversing the rotationReversing the rotationReversing the rotationReversing the rotationPress the left-hand foot switch down once to reversethe direction of rotation of the micromotor (anti-clockwise). The red M light on the dentist's displaywill come on when the micromotor is set to rotate inthe reverse direction.Press the foot switch a second time to switch thedirection of rotation of the micromotor back to thenormal (clockwise) direction. The red M light will goout.
12 User's Manual
The TThe TThe TThe TThe Turbineurbineurbineurbineurbine
Switching it on and offSwitching it on and offSwitching it on and offSwitching it on and offSwitching it on and offLift the turbine from its holder and press and holddown the foot lever. The T light will come on andthe turbine and spray will start. To stop the turbineremove your foot from the foot lever. The T lightwill go out.Note that there is no speed control with the turbine.
Chip blowerChip blowerChip blowerChip blowerChip blowerDouble press and hold down the foot lever. Chipblow will start. To stop chip blow remove your footfrom the foot switch.
Quint 7000 Dental Unit 13
The ScalerThe ScalerThe ScalerThe ScalerThe Scaler
Switching it on and offSwitching it on and offSwitching it on and offSwitching it on and offSwitching it on and offLift the scaler from its holder and press and holddown the foot lever. The scaler will start. To stop thescaler, remove your foot from the foot lever.
Selecting the power levelSelecting the power levelSelecting the power levelSelecting the power levelSelecting the power levelThe maximum power level of the scaler can be setat either 100%, 50% or 30%. The lights next to theSP key indicate what power level has been selected.To change the power levels press the SP keyrepeatedly until the light combination for thepower level you require appears. The power level/light combinations are as follows:- 100% power = no lights on- 50% power = top light on- 30% power = bottom light on
Note that the scaler power level will automaticallyreturn to the 100% level, i.e. no lights on, wheneverthe unit is switched off and then on again.
No lights Top light Bottom light 100% 50% 30%
14 User's Manual
Adjusting the power levelAdjusting the power levelAdjusting the power levelAdjusting the power levelAdjusting the power levelTo increase the power of the scaler slide the footlever to the right to decrease the power slide thefoot lever to the left.Note that when the foot control is moved to themaximum power position, the maximum powerlevel will be the level selected using the SP key.For example, if you selected the 50% power levelthe power level can adjusted from 0% to amaximum of 50%.
Quint 7000 Dental Unit 15
The Polymerisation light (Satelec)The Polymerisation light (Satelec)The Polymerisation light (Satelec)The Polymerisation light (Satelec)The Polymerisation light (Satelec)Lift the polymerisation light from its holder andpress the button on the side of the instrument. Thelight will come on. The light will automatically goout after 30 seconds. During the time the light is onyou will hear a "beep" every ten seconds. If youwish to turn the light out before the 30 secondselapses, press the button a second time.
For more information on how to operate and servicethe Satelec polymerisation light refer to thedocumentation supplied with the light.
The Polymerisation light (LThe Polymerisation light (LThe Polymerisation light (LThe Polymerisation light (LThe Polymerisation light (Lysta)ysta)ysta)ysta)ysta)Lift the polymerisation light from its holder andpress the button on the top of the instrument. Thelight will come on. The light will automatically goout after 35 seconds. After seventeen seconds youwill hear a beep. If you wish to turn the light outbefore the 35 seconds elapses, press the button asecond time.
For more information on how to operate and servicethe Lysta polymerisation light refer to thedocumentation supplied with the light.
Activation switch
16 User's Manual
Water
Spray
Air
The Syringe - 3-wayThe Syringe - 3-wayThe Syringe - 3-wayThe Syringe - 3-wayThe Syringe - 3-wayPress the left-hand button to activate the waterspray. Press the right-hand button to activate theair. Press both buttons together to activate the air/water spray.
The Syringe - 6-wayThe Syringe - 6-wayThe Syringe - 6-wayThe Syringe - 6-wayThe Syringe - 6-wayThere is a heater in this syringe that allows you tohave either hot air and water or cold. To switch theheater on twist the collar at the end of the handleanticlockwise - the indicator spot will be green. Toswitch the heater off twist the collar at the end ofthe handle clockwise - the indicator spot will beblack.Press the left-hand button to activate the waterspray. Press the right-hand button to activate theair. Press both buttons together to activate the air/water spray.
Water
Spray
Air
Heater on
Heater off
Quint 7000 Dental Unit 17
Dentist's control panelDentist's control panelDentist's control panelDentist's control panelDentist's control panel
Bowl rinse - on/offBowl rinse - on/offBowl rinse - on/offBowl rinse - on/offBowl rinse - on/offBriefly press the bowl rinse key on the dentist'scontrol panel to switch the bowl rinse on. Rinsingwill automatically stop after the preset rinseduration has elapsed. If you wish to stop the bowlrinse before the preset rinse time has elapsed, pressthe bowl rinse key a second time.
Bowl rinse - setting the rinse durationBowl rinse - setting the rinse durationBowl rinse - setting the rinse durationBowl rinse - setting the rinse durationBowl rinse - setting the rinse durationPress and hold down the bowl rinse key for theamount of time that you wish the bowl rinse tooperate (minimum one second, maximum tenseconds). Now, when the bowl rinse key is pressed torinse the bowl it will remain on for the sameamount of time as the bowl rinse key was helddown.
Glass fill - on/offGlass fill - on/offGlass fill - on/offGlass fill - on/offGlass fill - on/offBriefly press the glass fill key on the dentist'scontrol panel to switch the glass fill on. Filling willautomatically stop after the preset fill duration haselapsed. If you wish to stop the glass fill before thepreset fill duration has elapsed, press the glass fillkey a second time.
Glass fill - setting the fill durationGlass fill - setting the fill durationGlass fill - setting the fill durationGlass fill - setting the fill durationGlass fill - setting the fill durationPress and hold down the glass fill key for theamount of time that you wish the glass fill tooperate (minimum one second, maximum tenseconds). Now, when the glass fill key is pressed tofill the glass it will remain on for the same amountof time as the glass fill key was held down.
Door lock (Option)Door lock (Option)Door lock (Option)Door lock (Option)Door lock (Option)If a remote door lock is in use and has beenconnected to the dental unit this key is used toactivate it.
Auxiliary (Aux) keyAuxiliary (Aux) keyAuxiliary (Aux) keyAuxiliary (Aux) keyAuxiliary (Aux) keyIf any other equipment is being used with the unitthis key can be used to activate it.
18 User's Manual
WWWWWater bottleater bottleater bottleater bottleater bottleTo remove and refill the water bottle unscrew it.
TTTTTray holderray holderray holderray holderray holderThe tray holder is attached to a tray holder armwhich is attached to the underside of the instru-ment bridge on either the left- or the right-handside. The arm is angled allowing the tray to bepositioned either high or low depending on whichway round the arm is attached.To position the tray holder loosen the two tray armknobs, move the tray holder to the required posi-tion and then tighten the two knobs.
The tray holder will hold one standard size tray ortwo half size trays. A double size tray holder isavailable as an option.
Instrument bridge
Tray holder
Tray holder arm
Attaching knobs
Water bottle
Quint 7000 Dental Unit 19
The Suction The Suction The Suction The Suction The Suction ArmArmArmArmArm
Positioning the suction armPositioning the suction armPositioning the suction armPositioning the suction armPositioning the suction armThe height of the suction arm can be adjusted bysqueezing the locking handle, on the underside ofthe top part of the suction arm, and pulling thesuction arm up or pushing it down. When it is atthe required height, release the locking handle.
WWWWWarningarningarningarningarningDo not leave the suction arm under the chairbackrest as the arm could be damaged if you lowerthe chair.
Using the suction nozzlesUsing the suction nozzlesUsing the suction nozzlesUsing the suction nozzlesUsing the suction nozzles
Starting and stopping the suctionStarting and stopping the suctionStarting and stopping the suctionStarting and stopping the suctionStarting and stopping the suctionWhen a suction nozzle is lifted from its holdersuction will automatically start. When the suctionnozzle is replaced suction will stop.
Suction arm
Locking handle
20 User's Manual
Adjusting the suctionAdjusting the suctionAdjusting the suctionAdjusting the suctionAdjusting the suctionSlide the knob on the side of the nozzle upwards toincrease the suction and downwards to decrease thesuction.
Suction tipSuction tipSuction tipSuction tipSuction tipTo attach the suction tip push it into the opening atthe end of the suction nozzle.
The SyringeThe SyringeThe SyringeThe SyringeThe SyringePress the green button to activate the water spray.Press the blue button to activate the air. Pressboth buttons together to activate the air/waterspray.
Water
Spray
Air
Suction nozzle
Knob
Increase
Decrease
Quint 7000 Dental Unit 21
Suction arm controlsSuction arm controlsSuction arm controlsSuction arm controlsSuction arm controls
Bowl rinse - on/offBowl rinse - on/offBowl rinse - on/offBowl rinse - on/offBowl rinse - on/offBriefly press the bowl rinse key on the assistant'scontrol panel to switch the bowl rinse on. Rinsingwill automatically stop after the preset rinseduration has elapsed. If you wish to stop the bowlrinse before the preset rinse time has elapsed, pressthe bowl rinse key a second time.
Bowl rinse - setting the rinse durationBowl rinse - setting the rinse durationBowl rinse - setting the rinse durationBowl rinse - setting the rinse durationBowl rinse - setting the rinse durationPress and hold down the bowl rinse key for theamount of time that you wish the bowl rinse tooperate (minimum one second, maximum tenseconds). Now, when the bowl rinse key is pressed torinse the bowl it will remain on for the sameamount of time as the bowl rinse key was helddown.
Glass fill - on/offGlass fill - on/offGlass fill - on/offGlass fill - on/offGlass fill - on/offBriefly press the glass fill key on the assistant'scontrol panel to switch the glass fill on. Filling willautomatically stop after the preset fill duration haselapsed. If you wish to stop the glass fill before thepreset fill duration has elapsed, press the glass fillkey a second time.
Glass fill - setting the fill durationGlass fill - setting the fill durationGlass fill - setting the fill durationGlass fill - setting the fill durationGlass fill - setting the fill durationPress and hold down the glass fill key for theamount of time that you wish the glass fill tooperate (minimum one second, maximum tenseconds). Now, when the glass fill key is pressed tofill the glass it will remain on for the same amountof time as the glass fill key was held down.
Door lock (Option)Door lock (Option)Door lock (Option)Door lock (Option)Door lock (Option)If a remote door lock is in use and has beenconnected to the dental unit this key is used toactivate it.
22 User's Manual
The Dental LightThe Dental LightThe Dental LightThe Dental LightThe Dental Light
Switching it on and offSwitching it on and offSwitching it on and offSwitching it on and offSwitching it on and offThe switch on the underside of the arm that holdsthe dental light is used to switch the light on andoff.
Adjusting the light intensityAdjusting the light intensityAdjusting the light intensityAdjusting the light intensityAdjusting the light intensity.....The switch on the underside of the dental light isused to adjust the intensity of the light.
For more information on using and maintainingthe dental light, refer to the documentationsupplied with the light.
On/off switch
Intensity switch
Quint 7000 Dental Unit 23
4. Cleaning, disinfecting and sterilising the unit4. Cleaning, disinfecting and sterilising the unit4. Cleaning, disinfecting and sterilising the unit4. Cleaning, disinfecting and sterilising the unit4. Cleaning, disinfecting and sterilising the unit
Suction systemSuction systemSuction systemSuction systemSuction systemFor information on how to clean the suction system,refer to the documentation supplied with thesystem
Suction hosesSuction hosesSuction hosesSuction hosesSuction hosesAfter every patient allow one glass of water to besucked through each suction hose.
Twice a day, midday and evening, allow one glass ofsuction hose cleaning agent to be sucked througheach hose. Make sure that you prepare and use it inaccordance with the manufacturer's instructions.
Twice a day, midday and evening, empty and rinsethe suction hose filters. The suction hose filters arelocated inside the connectors that connect thesuction tubes to the cuspidor. To empty the filter,pull the suction hose connector from the cuspidor.Lift the filter out and rinse it. Also empty anydebris that is inside the connector. Replace thefilter and reconnect the suction hose connector tothe cuspidor.
Cuspidor
Suction hoses
Suction hose filter
Connector
24 User's Manual
Instruments and suction nozzlesInstruments and suction nozzlesInstruments and suction nozzlesInstruments and suction nozzlesInstruments and suction nozzlesFor information on how to clean and sterilize theinstruments, refer to the documentation suppliedwith them.
CautionCautionCautionCautionCautionSwitch the dental unit off before you remove anyinstruments. Press the syringe air and waterbuttons to relieve any pressure before removing thesyringe.
Suction nozzles can be disconnected from the hosesand autoclaved.
Unit surfacesUnit surfacesUnit surfacesUnit surfacesUnit surfacesUnit surfaces must cleaned and disinfected at leastonce a day with a suitable cleaning and disinfectingagents. Make sure that you prepare and use themin accordance with the manufactures instructions.Do not uses abrasive cleaning agents.
Note that the rubber instrument support can beremoved from the instrument bridge andautoclaved.
Instrument support
Instrument bridge
Quint 7000 Dental Unit 25
The rollers and holders at the end of the whip armscan be removed so that the instrument tubes can becleaned.To remove the roller and holder slide the holder upslightly until the roller can be removed, and thenremove the holder.
When replacing the roller and holder note that thetwo flats on the roller axle must engage in the slotin the holder, so that the holes line up, before theroller and holder can be slide back onto the whiparm.
On units that are fitted with hanging tubeinstruments the instrument holders can beremoved.To do this tilt the instrument holder forward as faras it will go, and then lift the holder up slightlyuntil it disengages from the holding pin. The holdercan then be removed.
Slot
Flat
Roller
Holder
Whip arms
Flat
Roller axle
Instrument holderTilt forward Lift and remove
26 User's Manual
To replace the instrument holder place the instru-ment holder so that the holding pin in theinstrument bride goes into the hole in the holderbracket. Push the instrument holder down to lock itin position.
Holder bracket
Holding pinInstrument holder
Hole
Quint 7000 Dental Unit 27
5. Maintenance5. Maintenance5. Maintenance5. Maintenance5. Maintenance
Oil collectorOil collectorOil collectorOil collectorOil collectorThe cotton pad in the turbine oil collector must bereplace monthly or more frequently if free oil startsto appear. The oil collector on the underside of theinstrument bridge. To remove and empty the oilcollector unscrew it.
Tightening iTightening iTightening iTightening iTightening instrument holdersnstrument holdersnstrument holdersnstrument holdersnstrument holdersOn units that are fitted with hanging tubeinstruments the instrument holders can tilted tomake it easier to remove and replace instruments.If a holder becomes loose it can be tightened withscrews on the side of the holder.
Instrument bridge
Oil collector
Screws
Instrument holder
28 User's Manual
Instrument water adjustmentInstrument water adjustmentInstrument water adjustmentInstrument water adjustmentInstrument water adjustmentThe supply of water to the instruments can beadjusted. This is done by turning the regulatorscrews located on the side of the instrument bridgewith a small screwdriver. The side on which the regulators are located willdepend on whether the unit is left- or right-handed.
Micromotor 1 water Syringe water Micromotor 2 water Scaler water
Turbine water
Water adjustment regulators
Quint 7000 Dental Unit 29
6. ServiceServiceServiceServiceServiceEvery six months a FINNDENT approved servicetechnician must carry out the following inspectionand service tasks:• check the output pressures of the air and water
control valves• check the air pressure of the instruments• check to see if the filters in the air and water
control valves need replacing• check the operation of the instruments and if
necessary replace and instrument O-rings orseals
CautionCautionCautionCautionCautionSwitch the dental unit off before you removeany instruments. Press the syringe air andwater buttons to relieve any pressure beforeremoving the syringe.
30 User's Manual
7. 7. 7. 7. 7. Approved accessoriesApproved accessoriesApproved accessoriesApproved accessoriesApproved accessoriesThe following list contains instruments that can beused with the Quint 7000 dental unit. Some of thelisted instruments are not part of the Finndentproduct range.
Micromotor drillsMicromotor drillsMicromotor drillsMicromotor drillsMicromotor drillsKavo Intramatic Lux - Models 190, 192 and 196
Satelec Servotorque - Models IML, IS, ES, AV, S
Quint (Satelec Servotorque IML)
Bien Air - Isolite 100, 300, MC40IR and MC20IR
TTTTTurbine drillsurbine drillsurbine drillsurbine drillsurbine drillsKavo - 630 630B, 640B and 650B
Bien Air - Black Pearl, Ondine, Gyro Long Life andPrestige S30L
W&H Toplight - all models
Air micromotorsAir micromotorsAir micromotorsAir micromotorsAir micromotorsKavo - all models
Bien Air - all models
W&H - all models
Light probesLight probesLight probesLight probesLight probesKavo - Dia-Lux 2300L
W&H - LS 1 Diagnostic Light
Bien - Air Translite
Quint 7000 Dental Unit 31
TTTTTartar scalersartar scalersartar scalersartar scalersartar scalersSatelec Suprasson SP 4055
Amdent 930
Quint (manufactured by Amdent)
Polymerisation lightsPolymerisation lightsPolymerisation lightsPolymerisation lightsPolymerisation lightsLysta LC-85
Satelec Servolite SL 4055
Satelec Acta
SyringesSyringesSyringesSyringesSyringesQuint 3-way syringe
Luzzani 6-way syringe
WWWWWater heatersater heatersater heatersater heatersater heatersSIE-Dental - water heating element 50/100W, 24V
Amalgam separatorsAmalgam separatorsAmalgam separatorsAmalgam separatorsAmalgam separatorsCattani - Suction separator
Metasys Multisystem Type MST 1
Metasys Multisystem MST 1 ECO
Dental lightsDental lightsDental lightsDental lightsDental lights
Faro - S90, S2000 and SC2000 Forma(230 V or 24 V)
32 User's Manual
8. T8. T8. T8. T8. Technical Specificationsechnical Specificationsechnical Specificationsechnical Specificationsechnical SpecificationsPROTECTION CLASS: IPROTECTION GRADE: BPROTECTION AGAINST FLUIDS: IPX0
OPERATIONAL SAFETY GRADE:- Standard: The unit is not suitable for use in an
environ ment with a mixture ofinflammable
anaesthesia gas and air, or a mixture ofinflammable anaesthesia gas and oxygen
or laughing gas (nitrous oxide).
MODE OF OPERATION: Continuous operation with short-term load.
ALLOWED OPERATINGCONDITIONS:
- Temperature: 15oC - 40oC- Relative humidity: 45% - 75%
ELECTRICAL FEATURES:- Power supply: 230 V, 50 Hz- Maximum power consumption: 1200 W (includes power consumption of
possible accessories connected to the unit)- Main fuse: 6.3 AT.- Primary fuse of transformer: 2AT
WATER CONNECTION: 3 bar - 6 barparticle size < 10 um
COMPRESSED AIRCONNECTION: 5.5 bar - 8 bar
WEIGHT : 78 kg without accessories90 kg with accessories125 kg in package with accessories
MANUFACTURER:FINNDENT OYTEOLLISUUSTIE 507230 MONNINKYLÄFINLANDTEL. +358 19 - 5221500FAX +358 19 - 643457
Quint 7000 Dental Unit 33
Symbols and abbreviationsSymbols and abbreviationsSymbols and abbreviationsSymbols and abbreviationsSymbols and abbreviationsv Voltage
A Ampere
W Watt
~~~~~ Alternating current
FI Approved by Fimko
Symbol for protection class B
European Community approval mark
AT Amperes, slow blow fuse
Control symbols are described where appropriate inthis guide.
34 User's Manual
DimensionsDimensionsDimensionsDimensionsDimensions
Quint 7000 Dental Unit 35
36 User's Manual
Quint 7000 Dental Unit 37