D-R300 / D-R320 / D-R 800
DATASHEET
JUNHO 2013
Measuring device for smoke spot number / dust concentration
D-R 300: Extremely sensitive device for measuring smoke spot number. D-R 300-40: Especially sensitive measuring device for the smallest concentration of dust, in particular in waste incineration plants.
Measuring principleThe D-R 300 / D-R 300-40 meters operate ac-cording to the scattered light method. The modulated light from a Halogen lamp illumi-nates the dust particles in the exhaust duct. The scattered light reflected from these particles is measured and assessed.
Specifications D-R 300
● Measuring range: smoke spot number 0-3. Limit value II can be set as delayed discon-nected contact
● Calibration: VDI 2066, page 8.
D-R 300-40 ● Minimum measuring range 0-1 mg/m3 dust
concentration. Other measuring ranges up to 300 mg/m3 can be set (optional automatic switching of measuring ranges 1:3:9)
● Calibration: VDI 2066, page 7.
System components ● Measuring head ● Mounting flange ● Control and display unit ● light trap(s) ● Purge air unit.
Options ● Automatic measuring range selection for
dust measurement in accordance with 17. BImSchV (D-R 300-40)
● Weather protection covers ● Quick-closing shutter to protect measuring
device in the event of failure of the purge air supply
● Halar flange coating.
Features ● In-situ measurement directly in the flue gas
flow ● Automatic system tests and correction of
measured values ● Self-calibration in 4-h cycle ● Optics and electronics in a hermetically
sealed unit - no smoke gas can enter device ● Maintenance-friendly, thanks to optimal
purge air conduction in front of the heated optical end plates
● Direct access to all parameters via the opera-tor display
● Automatic measuring range selection in ac-cordance with 17. BImSchV. in relation 1:3:9*.
Applications ● Smoke spot number measurement in fur-
naces designed for light fuel oil ● Fine dust measurements in processes ● Waste incineration ● Filter monitoring and emission-value
measuring in waste incineration and power plants*.
Approvals ● Suitability-tested by the TÜV Cologne,
test report 936/800002 ● Suitability-tested by the TÜV Cologne,
test report 936/801004* ● Itemized in the list of suitable measuring
devices for continuous emission measuring ● MCERTS.
*D-R 300-40
light trap
2nd light trap (D-R 300 only)
quick-closing shutter (optional)
measuring head
control and display unit
purge air unit
measurements smoke spot number (D-R 300) dust concentration (D-R 300-40) detection limit <1% measuring range
measuring ranges
D-R 300: smoke spot number 1–3 (5) D-R 300-40: dust concentration 0–1 mg/m3 ... 0–30 mg/m3 1),
optional 0–300 mg/m3 with range switching
reference point drift <0.2% of measuring range/month
measuring principle back scattering zero point drift <0.2% of measuring range/month
flue gas temperature
above dew point up to 320°C, optional up to 600°C, depending on application
supply voltage 115 / 230 VAC, 50 / 60 Hz, 50 VA
flue gas pressure -50 up to +20 hPa dimensions (h x w x d)
measuring head 565 x 310 x 200 mm
duct diameter 0.3 to 4 m weight 18 kg
ambient temperature -20 up to +50°C remarks 1) after gravimetric calibration
protection IP65 purge air supply
measuring outputs 2 x 0/4–20 mA/500 Ohm, optional measuring range switching purge air quantity approx. 80 m3/h
digital outputs 3 relay outputs, permissable load 48 V / 0.5 A supply voltage 115 / 230 VAC, 50 / 60 Hz,
0.37 / 0.43 kW
digital inputs 1 potential free inputdimensions
(h x w x d) weight
350 x 550 x 500 mm 12 kg
accuracy <1% of measuring range protection IP54
320D-R
Stray light measurement redefined
The light of a red laser diode (eye safe) illu-minates the dust particles in the measuring volume of the flue gas duct. The light is scat-tered by the particles and also generates background light in the stack. Scattered light plus background light are received in one half of the integrated dual detector.
At the same time the other half of the detec-tor receives only the background light from the same area of the stack, without the stray light portion.
After subtraction of the background light the pure amount of scattered light is avail-able. Since both parts of the integrated dou-ble detector share the same optics unit, no adjustment of the beam paths is necessary. The D-R 320 is ready for operation directly after mounting and can easily be operated at another measurement point.
The new DURAG dust monitor D-R 320 is based on the backscattering principle and is ideally suited for monitoring smallest to medium dust concentrations. The measurement takes place continuously in the flue gas stream. Naturally the D-R 320 is compliant with the high require-ments of the new European standards and therefore suitable for measurements according to official regulations.
Startup without adjustment due to unique optical design
The D-R 320 does not require a light trap. The background light in the stack is detected by means of a special optical system with dual de-tector and automatically compensated without the need of any manual adjustment.Advantage of this new design is the quick and uncomplicated installation, completely without adjustment at site. So the D-R 320 is ready for operation directly after installation and flexible for different applications. Even after a conver-sion to a different measurement point the de-vice is operational immediately again.
The new scattered light dust monitor D-R 320High-sensitivity measurement even at the lowest concentrations of dust
Features
●● Continuous measurement of dust concentration
●● Easy installation on standard flanges●● Easy startup without adjustment●● Automatic background light compensa-
tion, no light trap●● Convenient operation,
remote access possible●● Automatic control functions●● Integrated purge air monitoring
and control●● Long service intervals●● Certified acc. to EN 15267●● Integrated fail safe shutter as an option●● Data transfer via Modbus compliant with
VDI 4201-3
Applications
Continuous emission monitoring, e.g.:●● on incineration plants generally●● on plants according to IED 2010/75/EU●● on plants according to 2001/80/EC LCPD●● on waste, hazardous waste and sewage
sludge incineration plants●● on cement manufacturing plants●● on power plants with gas, oil, coal or co-
firing●● on plants for the incineration of biomass
Monitoring of:●● ventilation units●● filter units
measuring head
measurements dust concentration
measuring range min: 0–5 mg/m3, max: 0–200 mg/m3
measuring principle back scattering
flue gas temperature
0…600 °C
flue gas pressure -50 ... +50 hPa, optional higher
duct diameter > 0.7 m
ambient temperature
-40 up to +60 °C
protection IP 65, Ex optional
measuring outputs 0 / 4–20 mA / 400 Ohm, Modbus RTU bi-directional
digital outputs 2 relay outputs, permissible load 60 VDC / 30 VAC / 0.5 A
supply voltage 24 VDC, 0.5 A
dimensions (h x w x d)
200 x 190 x 260/410 mm
weight 15 kg
supply unit
purge air supply integrated blower
supply unit 115/230 VAC, 50/60 Hz, 0.37/0.43 kW
dimensions 480 x 450 x 320 mm
weight 12 kg
protection IP 65
Long life due to integrated purge air monitoring and control
The D-R 320 is continuously purged and there-fore the optical system is not in contact with the exhaust gas. The integrated purge air control ensures the optimum amount of purge air. This ensures a long life and long maintenance inter-vals. A purge air failure is immediately indicated by the device. The maintenance of the device is minimal.
Convenient operation, readings anywhere accessible
The operation of the D-R 320 is comfortable and extremely flexible. Different operating options are available:
●● local via USB interface●● by means of the external operation unit
D-ISC 100●● remote access with RS-485 interface via
MODBUS RTU / MODBUS TCP protocol (acc. to VDI 4201-3)
Automatic control functions ensure reliable measuring values
Reliability has the highest priority when meas-uring emissions. For that reason the D-R 320 automatically performs zero and span check as well as contamination check in regular in-tervals and on demand. The device features an automatic contamination correction. Need for maintenance is immediately indicated by the electronics.
Options
●● Weather protection covers●● Fully integrated failsafe shutter as protection
for the measuring device in case of a purge air failure.
With the integrated fail safe shutter the D-R 320 is optimally protected for reliable use even un-der overpressure conditions.
Universal Operation Unit D-ISC 100
The universal operation unit D-ISC 100 allows for the connection of up to 8 DURAG measuring heads and enables the combination of dust and flow measurements. The D-ISC 100 offers con-venient operation and parameterization of the connected devices. The display provides an im-mediate overview of the current measurements and the status of the devices.
●● Connection of up to 8 measuring heads●● Local operation with keypad, through USB
port or remotely through RS-485 interface.●● Modular design, many extension modules
available
Single-sided installation on standard flanges without light trap
The D-R 320 allows single sided installation at the stack and requires just a small wall opening. It can be mounted in any orientation via a rotatable connection on existing DN 100 PN 6 or ANSI 4“ 150 lbs standard flanges. This reduces time and costs for installation.
Universal mounting: DN 100 PN 6 DIN flange
Universal mounting: ANSI 4“ 150 lbs standard flange
Universal mounting:rotatable, for any flange orientation
System components
●● Measuring head●● Mounting flange
●● Universal control unit D-ISC 100
●● Supply unit with inte-grated purge air supply
Technical Data
Dust monitor
Innovative measuring device with laser technology to monitor small to medium dust emission according to the new European regulations.
Features
●● Integrated display: Measuring value, threshold value, parameter in probe
●● In-situ measuring procedure with continu-ous measurement
●● High sensitivity
●● Easy installation on one side of the duct
●● Can also be deployed in thick-walled stone/insulated channels
●● Long lifetime, as there are no moving parts inside the duct
●● Hermetically sealed electronic housing against exhaust gas
●● Parameterisation and operation with key-board plus easily readable display directly on device or via bus interface
●● Automatic function test with soiling correc-tion
●● Two analog outputs with adjustable measur-ing ranges
●● Automatic switching of measuring ranges according to 17. BImSchV.
Applications
●● Power stations
●● Cement plants, the metallurgy and wood industries, chemical industry etc.
●● Waste incineration plants
●● Monitoring of dust filter plants.
Approvals
●● Suitability-tested by the TÜV Cologne, test report 936/21205307/A
●● Itemized in the list of suitable measuring devices for continuous emission measuring
●● MCERTS pending.
System components
●● Measuring lance
●● Supply unit with integrated purge air supply
●● Mounting flange 130 / 240 / 500 mm.
Options
●● Weather protection cover
●● Temperature compensation through ad-ditional analog input.
Measuring principle
The D-R 800 device works according to the prin-ciple of forward scattering. The concentrated and modulated light of a laser diode penetrates the measuring volume. The forward-scattered light largely reflected from dust particles is measured and assessed.
measuring lance
evaluation unit
supply unit with integrated purge air
supply
RS 485 and analog current output
measurements dust concentration accuracy <1% of measuring range limit
measuring ranges 0–10 mg/m3 ... 0–200 mg/m3 1) detection limit <0.5% measuring range
measuring principle forward scattering reference point
drift <0.7% of measuring range/month
flue gas temperature above dew point up to 220 °C zero point drift <0.15 % of measuring range/
month
flue gas pressure -50 up to +10 hPa supply voltage 85–264 VAC, 47–63 Hz, 50 VA
duct diameter 0.4–8 m dimensions (h x w x d)
measuring lance: 160 x 160 x 600 / 1000 mm supply unit: 380 x 300 x 210 mm
probe length (from flange) 400 / 800 mm weight measuring lance: 7 kg
supply unit: 13 kg
ambient temperature -20 up to +50°C purge air supply integrated into supply unit
protection IP65
measuring outputs 2 x 0 / 4–20 mA / 500 Ohm, Modbus RTU (RS485)
digital outputs 4 relay outputs, programmable, permissable load 24 V / 25 VA
digital inputs 2 potential free inputs, programmable remarks 1) after gravimetric calibration
pending
www.durag.de
Contactos/Contacts:
Comercial/Commercial:
Luís Ferreira da Costa
e-mail: luiscosta@bhb,pt
Tel: (+351) 21 843 64 00
Fax: (+351) 21 843 64 07
Assistência/Service:
Joaquim Picante
e-mail: [email protected]
Tel: (+351) 21 843 64 00
24 Horas: (+351) 96 523 73 93
Note:
DURAG, Inc., the owner of this document, reserves the right to make technical changes or modify the contents of this document without prior notice. With regard to purchase orders, the agreed particulars shall prevail. DURAG, Inc. does not accept any responsibility whatsoever for potential errors or possible lack of information in this document.
Copyright© 2011
DURAG, Inc. All rights reserved