ct 64 dinamica de medellin
TRANSCRIPT
![Page 1: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/1.jpg)
GE Company confidential - For internal use only
Nuevos Tópicos
En TC Multicorte
Medellin Colombia Octubre 2012
Marta lucia Villa Muñoz
Especialista en Aplicaciones en CT GEHC L.A
![Page 2: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/2.jpg)
2 / GE /
GE confidential, for internal use only
TEMAS • Evolucion de la Tomografia
• Factores que afectan la dosis
& calidad • Avances en CT
• Aplicaciones Avanzadas
1. Angio CT Coronaria 2. CT Perfusion TiP Training in Partnership
![Page 3: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/3.jpg)
3 / GE /
GE confidential, for internal use only
GE Healthcare
Evolucion de la Tomografia
TiP Training in Partnership
![Page 4: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/4.jpg)
4 / GE /
GE confidential, for internal use only
Evolucion de la tecnologia...
![Page 5: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/5.jpg)
5 / GE /
GE confidential, for internal use only
Godfrey N. Hounsfield Introdujo la Tomografia Axial Computarizada en 1.972 como un metodo de diagnostico
![Page 6: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/6.jpg)
6 / GE /
GE confidential, for internal use only
LightSpeed Plus 1st 4 Cortes CT con
sub-seg y velocidades variables
2000
2001
ProSpeed
LightSpeed QX/i 1st 4 Cortes CT
1992-1998
2002
LightSpeed16
1st 16 Cortes CT
MicroVoxel
LightSpeed VCT 1st 64 Cortes CT
2005
LightSpeed VCT XT 1st 64 Cortes CT con tecnología
Snapshot Pulse y Volume shuttle
2006
EVOLUCION CT
LightSpeed Ultra 1st 8 Cortes CT
![Page 7: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/7.jpg)
7 / GE /
GE confidential, for internal use only
AVANCES EN CT
Dual Head-E/CAM,
Siemens-SPECT
Optima 660 EcoImagination 2010
LightSpeed CT 750 HD 1st High-Definition CT scanner
2008
![Page 8: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/8.jpg)
8 / GE /
GE confidential, for internal use only
GE Healthcare
Factores que afectan la dosis & calidad
TiP Training in Partnership
![Page 9: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/9.jpg)
9 / GE /
GE confidential, for internal use only
Patient Dose Image Noise
As Low As Reasonably Achievable (ALARA)
Dose Vs. Image Noise
![Page 10: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/10.jpg)
10 / GE /
GE confidential, for internal use only
En resumen . . .
No penetra la piel La piel puede detenerla
Riesgo interno
Detenida por el papel
Penetra piel, ojo e implica riesgo interno
Detenido por plástico
![Page 11: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/11.jpg)
11 / GE /
GE confidential, for internal use only
Efectos Agudos
Generalizados:
Accidentes de reactores nucleares
Produce daño hematológico, digestivo y cutáneo
Localizados:
Por aplicación indiscriminada de rayos X
Quemaduras cutáneos y destrucción de partes profundas
Puede generalizarse
Efectos Biológicos
![Page 12: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/12.jpg)
12 / GE /
GE confidential, for internal use only
Efectos Crónicos Generalizados:
Exposición a fuente externa (X o gamma): Leucemias, Anemias aplásicas. Leucopenia, Púrpura trombocitopénica, Poliglobulia
Exposición a fuentes intrernas: Sarcomas óseos, Ca. Pulmonar (inhalación de radón en minas de uranio)
Localizados:
Dermatitis crónica en manos, no reversible
Ca. Cutáneo que complica dermatosis de manos
Cataratas
Oligospermia
Efectos Biológicos
![Page 13: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/13.jpg)
13 / GE /
GE confidential, for internal use only
JULIO /2001 U.C. - S.C.M.T. - MINTRABAJO - S.G.R.P. -SOCIEDAD COLOMBIANA DE MEDICINA DEL
TRABAJO – UNIVERSIDAD DE CALDAS – SGRP - MINTRABAJO
Efectos Estocásticos
Presentación Aleatoria
Relación Dosis - Efecto
La célula conserva su capacidad reproductiva
Son diferidos en el tiempo
Efectos Biológicos
![Page 14: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/14.jpg)
14 / GE /
GE confidential, for internal use only
JULIO /2001 U.C. - S.C.M.T. - MINTRABAJO - S.G.R.P. -SOCIEDAD COLOMBIANA DE MEDICINA DEL TRABAJO –
UNIVERSIDAD DE CALDAS – SGRP - MINTRABAJO
Efectos Determinísticos
La gravedad del efecto depende proporcionalmente de la dosis
Relación de Causalidad determinística
Posibilidad de reparación celular
Afectan la capacidad reproductiva
No son diferidos en el tiempo
Efectos agudos
Efectos Biológicos
![Page 15: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/15.jpg)
15 / GE /
GE confidential, for internal use only
EFECTOS BIOLOGICOS EFECTOS NO ESTOCÁSTICOS
•AGUDOS:
•Globales: Aplasia medular
•Parciales: Radiodermatitis agudas
exudativas, alopecia, oligospermia
•TARDIOS:
•Ojo: Catarata,
•Piel: Radiodermatitis crónica atrófica
•Embrión: Malformaciones
![Page 16: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/16.jpg)
16 / GE /
GE confidential, for internal use only
Minimizar la Dosis con Buenas Prácticas (Principio ALARA)
- TIEMPO - reducir tiempo de exposición
- DISTANCIA - incrementar distancia
- BARRERAS - uso de barreras
![Page 17: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/17.jpg)
17 / GE /
GE confidential, for internal use only
MARCO TEORICO
CONTROLES EN LA PERSONA
•Capacitación y entrenamiento
•EPP plomados, de acuerdo con el tipo de
exposición
•Vigilancia individual a la exposición.
![Page 18: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/18.jpg)
18 / GE /
GE confidential, for internal use only
Límites actuales en el mundo: Para los trabajadores de la salud: 20 msv por año promediados en 5 años Para el públIco en general: 1 msv por año
Cual es el Limite de Dosis en la actualidad a nivel mundial?
![Page 19: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/19.jpg)
19 / GE /
GE confidential, for internal use only
RADIACIÓN DE FONDO
3 mSv/año por persona
Cósmico 0.3 mSv
Alimentos 0.4 mSv Terrestre 0.3 mSv
Radón 2 mSv
Air
![Page 20: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/20.jpg)
20 / GE /
GE confidential, for internal use only
OTRAS FUENTES DE RADIACIÓN 0.6 mSv / año
Viaje aéreo transatlántico 0.05 mSv
Fuentes hospitalarias
0.5 mSv Plantas nucleares y precipitaciones
radiactivas < 0.05 mSv
![Page 21: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/21.jpg)
GE Company confidential - For internal use only
Dosis de Radiación
CTDIvol
DLP (dose length product)
Z-axis dose efficiency
Projected DLP
Accumulated DLP
100.0
CTDIvol
Pantalla con Información de Dosis
![Page 22: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/22.jpg)
GE Company confidential - For internal use only
Dosis de Radiación
Scan Range
CTDIvol
DLP (dose length product)
CTDI phantom diameter
Imagen con Informe de Dosis
![Page 23: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/23.jpg)
GE Company confidential - For internal use only
Dosis de Radiación Dosis Efectiva - ¿ Cómo Calcular?
Método DLP / EUR16262 EN
Exposure Length (not image range if helical) x CTDIvol DLP =
(mGycm) (cm)
(mGy)
Helical Exposure
Length =
Exposure time Table speed x
(cm) (s) (cm/s)
![Page 24: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/24.jpg)
GE Company confidential - For internal use only
Dosis de Radiación
Método DLP / EUR16262 EN
Exposure Length x CTDIvol DLP =
DLPEE DLP
Head 0.0023
Neck 0.0054
Chest 0.017
Abdomen 0.015
Pelvis 0.019
150 25 6
2.55 mSv 150 0.017
¿Cual es la dosis efectiva en un estudio de TC de tórax donde el largo de exposición fue de 25 cm y el CTDIvol fue de 6 mGy?
![Page 25: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/25.jpg)
25 / GE /
GE confidential, for internal use only
Una exploración del CT - generalmente una exposición parcial del cuerpo con la dosis a algunos órganos
Región de la dosis para una exploración del
pulmón
Órganos del torax que reciben la
dosis, el pecho, un poco de hueso y
algo piel
![Page 26: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/26.jpg)
26 / GE /
GE confidential, for internal use only
Las dosis del órgano se pueden cargar y agregar para obtener la dosis efectiva
1990 ICRP Report 60 - recommendations of the international commission on radiological protection
Sum of (organ dose x weight factor)
Convierte la dosis parcial a una dosis efectiva del cuerpo entero
![Page 27: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/27.jpg)
27 / GE /
GE confidential, for internal use only
LINK PARA ESTIMADO DE RIESGO DE CANCER POR RADIACIONES IONIZANTES www.faculty.washington.edu/aalessio/doserisk2
Tener en cuenta para ingresar:
• Total DLP
• Phantoma (16cm) o (32cm)
• Edad del paciente
• Sexo (Female/Male)
![Page 28: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/28.jpg)
28 / GE /
GE confidential, for internal use only
A continuación se muestra una comparación de las dosis de radiación efectivas con la exposición de fondo para varios procedimientos radiológicos descritos en
![Page 29: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/29.jpg)
29 / GE /
GE confidential, for internal use only
![Page 30: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/30.jpg)
30 / GE /
GE confidential, for internal use only
GE Healthcare
PROTOCOLOS PEDIATRICOS
TiP Training in Partnership
![Page 31: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/31.jpg)
31 / GE /
GE confidential, for internal use only
Protocolos Pediatricos
![Page 32: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/32.jpg)
32 / GE /
GE confidential, for internal use only
![Page 33: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/33.jpg)
33 / GE /
GE confidential, for internal use only
![Page 34: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/34.jpg)
34 / GE /
GE confidential, for internal use only
Protocolo Pediatrico
Puntos para Recordar!
•Solo realice los estudios de CT, si son necesarios.
•Barrer solo el organo/región solicitada
•Minimize examenes Multifasicos.
•Centre apropiadamente al paciente con el eje del isocentro.
•Use bajo mA para examenes de toráx y hueso
•Solo use tecnicas altas para visualizar estrucutras pequeñas de CTA.
•Ajuste parametros por tamaño e indicaciones.
•Use Auto-mA y bajo Kv si es posible
•Incremente el Pitch
•Use Small SFOV cuando sea posible.
![Page 35: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/35.jpg)
35 / GE /
GE confidential, for internal use only
Protocolos Pediatricos
ZONA ROSADA:La rutina o la mayoría de las situaciones o las indicaciones de urgencias
ZONA VERDE :Dosis baja o estudios de control
ZONA GRIS: Angiografía del CT (excepto CTA coronarios)
Carteles disponibles de GE Healthcare.com CT 4Kids
http://www.gehealthcare.com/usen/ct/products/ct_4kidsftrs.html
![Page 36: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/36.jpg)
36 / GE /
GE confidential, for internal use only
![Page 37: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/37.jpg)
37 / GE /
GE confidential, for internal use only
![Page 38: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/38.jpg)
38 / GE /
GE confidential, for internal use only
![Page 39: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/39.jpg)
39 / GE /
GE confidential, for internal use only
Caso 1. Centro de Imaginologia UNEME. Villahermosa-Tabasco. Mexico
ANGIO-TC DE TORAX
CONEXION VENOSA PULMONAR ANOMALA TOTAL,
TIPO: SUPRACARDIACA. LAS 4 VENAS PULMONARES SE CONECTAN CON LA VENA BRAQUIOCEFALICA IZQUIERDA A TRAVES DE UN COLECTOR VERTICAL.
TOTAL EXAM DLP: 70.81 mGy-cmSCAN
RANGE I132.25 - S2.750
CTDIvol (mGy) = 4.53Phantom = cm = body 32
![Page 40: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/40.jpg)
40 / GE /
GE confidential, for internal use only
Caso 1.
Colector
Aorta
![Page 41: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/41.jpg)
41 / GE /
GE confidential, for internal use only
![Page 42: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/42.jpg)
42 / GE /
GE confidential, for internal use only
Caso 2. Centro de Imaginologia UNEME. Villahermosa-Tabasco. Mexico
![Page 43: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/43.jpg)
43 / GE /
GE confidential, for internal use only
Caso 2.
UROTAC CON UN RIÑON ECTOPICO (PELVICO). SE REALIZARON 2 FASES:
NEFROGRAFICA Y DE ELIMINACION.
2 AÑOS
TOTAL DLP: 144.85
PHANTOM BODY 32 CM
CTDIvol (mGY) DE 3.05 EN AMBAS FASES.
DLP (mGy-CM): 55.65 Y 89.20 RESPECTIVAMENTE.
SCAN RANGE:
F. NEF: S23.250-I126.750
F. ELIM: S23.250-I236.750
![Page 44: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/44.jpg)
44 / GE /
GE confidential, for internal use only
![Page 45: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/45.jpg)
45 / GE /
GE confidential, for internal use only
![Page 46: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/46.jpg)
46 / GE /
GE confidential, for internal use only
![Page 47: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/47.jpg)
47 / GE /
GE confidential, for internal use only
GE Healthcare
Aplicaciones Avanzadas
TiP Training in Partnership
![Page 48: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/48.jpg)
48 / GE /
GE confidential, for internal use only
Cabeza y cuello - CTA*
- Carotids
- Orbitas / senos
-Perfusion
Torax - Angio TEP
- Aorta dissección
Corazón - Angio Coronaria
- Cuantificación de calcio Abdomen Pelvis - Colonografía
- Renales
- 3 Fases Higado
- Pelvis Extremidades CTA
Miembros Inf
- Rodilla
- Tobillo
- Pie
CT Areas de aplicación
![Page 49: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/49.jpg)
49 / GE /
GE confidential, for internal use only
Abdomen CTA
Protocol
- rotation time: 0.8sec
- scan configuration: 1.25*16 - kv:120kv - mA: 50 to 350mA -Noise index : 9.6 - pitch: 1.75 - kernel: STD
- Scan time : 25 sec
![Page 50: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/50.jpg)
50 / GE /
GE confidential, for internal use only
Peripheral CTA
Protocol - rotation time: 0.8sec
- scan configuration: 1.25*16 - kV:120kv
- mA: 50 to 350mA - Noise index : 9.6 - pitch: 1.75 - kernel: STD
- Scan time : 25 sec
Curved left 3D MIP
![Page 51: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/51.jpg)
51 / GE /
GE confidential, for internal use only
![Page 52: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/52.jpg)
52 / GE /
GE confidential, for internal use only
Carotidas con Baja Dosis: 0,91 mSv !
* Obtained by EUR-16262 EN, using a neck factor of 0.0054*DLP
Adquisición en 4,2 s
acquisition cortes
submilimetricos
DLP = 170 mGy.cm
Eqv dosis = 0,91mSv*
![Page 53: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/53.jpg)
53 / GE /
GE confidential, for internal use only
Cortesia de CCN, St Denis - France
RunOff: Excepcional Calidad de Imagen y Baja Dosis!
* Obtained by EUR-16262 EN, using an adult RunOff factor of 0.009*DLP
138 cm cobertura DLP = 835 mGy.cm
Eqv dosis = 7,5mSv*
![Page 54: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/54.jpg)
54 / GE /
GE confidential, for internal use only
Cortesia de CCN, St Denis - France
Arterias Renales Rápida adquisición a Baja Dosis!
52 cm cobertura en 6,4 seg DLP = 877 mGy.cm
Eqv dosis = 4,4 mSv*
* Obtained by EUR-16262 EN, using an adult abdomen factor of 0.015*DLP and an adult pelvis factor of 0,019*DLP
![Page 55: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/55.jpg)
55 / GE /
GE confidential, for internal use only
“Direccionado a nuevos diseños en la aplicacion,
explorando el poder de la Alta resolucion, y
cobertura volumetrica.”
VCAR
olu
me
om
pute
r
ssis
ted
ead
ing
![Page 56: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/56.jpg)
56 / GE /
GE confidential, for internal use only
VCARTM – LECTURA volumen interactivo : un diseño innovador
que aprovecha plenamente el tremendo poder de alta
resolución, imágenes de volumen CT y PET/CT. ¿El resultado?
Diagnostico… informe
QUE ES VCAR?
Volume Computer Assisted Reading (VCAR) La herramienta para cuantificar la enfermedad desde un volumen de datos encontrar - equipo asistido por visualización aislar - precisa segmentación medida - potente análisis cuantitativo interactivo
![Page 57: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/57.jpg)
57 / GE /
GE confidential, for internal use only
1. DE QUE HABLAMOS
El nódulo pulmonar solitario (NPS), se define como la imagen radiológica redondeada de aumento de densidad, rodeada de tejido pulmonar y menor de tres centímetros que no se acompaña de adenopatías ni atelectasias.
Los principales objetivos de su estudio son: realizar un diagnóstico precoz del cáncer de pulmón y determinar con la mayor precisión posible, las posibilidades de malignidad o benignidad de la imagen, con objeto de evitar el mayor número de pruebas cruentas para el paciente
![Page 58: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/58.jpg)
58 / GE /
GE confidential, for internal use only
OBJETIVOS
•POSIBILIDADDE UN DIAGNÓSTICO PRECOZ
•POR DIAGNÓSTICO RADIOLÓGICO DETERMINAR CON MAYOR PRECISIÓN MALIGNIDAD O BENIGNIDAD
![Page 59: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/59.jpg)
59 / GE /
GE confidential, for internal use only
Lung VCAR
Analisis Automática de nódulos: • % de Crecimiento • Tiempo de duplicación • Los volúmenes (de componentes sólidos, de componentes no sólido, Volumen Total) • Máxima distancia en 3D (eje longitudinal x, y, z) • Las mediciones 2D • unidades Hounsfield (Min, Max, de media)
![Page 60: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/60.jpg)
60 / GE /
GE confidential, for internal use only
Segments All Nodule Types
Solid Part Solid Non Solid
![Page 61: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/61.jpg)
61 / GE /
GE confidential, for internal use only
Lung VCAR •Comparación de inicial y seguimiento del
estudio
exámenes anteriores y actuales son
registrados automáticamente y se
muestra una comparación
correlacionados.
Lung VCAR proporciona tamaño del
nódulo de seguimiento junto con el% de
aumento y tiempo de duplicación para ayudar a manejar los nódulos en el
tiempo
![Page 62: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/62.jpg)
62 / GE /
GE confidential, for internal use only
Mujer de 65 años, fumadora. Le detectaron nodulo por
Radiografía de tórax.
1st examen calcularon volume 769 mm3.
En la imagen axial del segundo examen el nodulo no muesttra aparente cambio en el tamaño, realizado 3 meses y 19 dias mas
tarde. La variacion de la medida manual del diametro fue < 1
mm.
La medidad de volumen del software en la 2a CT adquisicion fue
1038 mm3, representando un incremento 35% con tiempo promedio al doble de su tamaño en 255 dias, sugeriendo
malignidad. Examen patologico despues de la reseccion
quirurgica demostro un adenocarcinoma pulmonar estadio I
Revel MP, Bissery A, Bienvenu M, Aycard L, Lefort C and Frija G
1st exam volume:769 mm3
Evaluacion del Nodulo
2nd exam Volume: 1038 mm3
Doubling time: 255 days % growth: 35%
![Page 63: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/63.jpg)
63 / GE /
GE confidential, for internal use only
![Page 64: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/64.jpg)
64 / GE /
GE confidential, for internal use only
Lung VCAR Ubicacion automatica de Nodulos
Digital Contrast Agent
DCA Iluminacion de estructuras de interés para su revisión
Con el tamaño de filtro ajustable para disminuir el tiempo de revisión.
Mejorar su productividad Complementos de la lectura visual, ofreciendo automáticamente la visualización y el análisis cuantitativo
![Page 65: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/65.jpg)
65 / GE /
GE confidential, for internal use only
Cancer de Colon.. #2 Causa de Muerte
El cáncer colorrectal es tratable cuando se
detecta en una etapa temprana localizada
sin embargo del 90%, sólo el 39% de los cánceres de colon son diagnosticado en
esta etapa
Limitantes
Las barreras de los pacientes:
Percibir que el procedimiento doloroso,
incómodo, invasivo, vergonzoso, etc...
Mala preparacion.
Los pacientes que se consideran saludable evitar
exámenes que les hará sentir incómodos o
imponer fuerte estrés
![Page 66: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/66.jpg)
66 / GE /
GE confidential, for internal use only
Colon Procedures Patient Compliance
Patient Feedback
FECAL BLOOD TESTING
SIGMOIDOSCOPY
DOUBLE CONTRAST BARIUM ENEMA
COLONOSCOPY
CT COLONOGRAPHY
Compliance Comparison
![Page 67: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/67.jpg)
67 / GE /
GE confidential, for internal use only
Procedimientos del Colon Test Fecal () Provee una prueba efectiva de Sangre en las heces
() Seguro y Economico
() Solo es positivo cuando sangra (La mayogranria de los polipos)
() 30-40% sensibilidad
Sigmoidoscopia () Procedimiento Ambulatorio
() Menos costoso que la Colonoscopia
() Solo explora el recto-sigmoide (y la izquierda del colon)
() 40-65% sensibilidad
CT Colonography () Facil procedimiento de limpieza
() Excelente como screening
() preparacion e insuflacion completa del colon
() 75-85% sensitivity (polipos 5-9mm)
La sensibilidad de la Colonografia es el
standard de Oro. . . CT
Colonography deberia avanzar para mejorar la experiencia del paciente
y la sensibilidad del diagnostico
Colonoscopia Optica () 90% sensitivity (polipos menores 5-9mm)
() Permite realizar Muestras Bx
() Preparacion y parcial insuflacion del colon puede requerir anestesia
() Incompleta en el 15-20% de los casos
DC Barium Enema () Estudio Completo del colon
() Baja Dosis
() 50-70% sensitivity (polipos mayores a 10mm)
![Page 68: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/68.jpg)
68 / GE /
GE confidential, for internal use only
Beneficios para el Paciente
Más comfortable… menos ansiedad… una mejor experiencia del paciente. alternativa no invasiva a colonoscopia toma mucho menos tiempo que una colonoscopia no requiere sedación
![Page 69: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/69.jpg)
69 / GE /
GE confidential, for internal use only
Pacientes 50 años de edad y mayor Pacientes 40 años de edad. Historia Personal o familiar de cáncer de colon: colonoscopia incompleta Contraindicaciones (CO) colonoscopia optica es actualmente todavía el procedimiento de elección para pacientes de alto riesgo
Indicaciones para CT Colonography
![Page 70: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/70.jpg)
70 / GE /
GE confidential, for internal use only
CTC Beneficios
Completamente sincronizado, orientado y personalizado en la lectura protocolos prono y supino Automatica y rapida segmentacion Visualizacion de diseccion 360 del colon Reporte Integrado
![Page 71: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/71.jpg)
71 / GE /
GE confidential, for internal use only
Colon VCAR
Detección automática de posibles lesiones umbral de detección de color mejora en 3D - Auto disección en el actual proceso
![Page 72: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/72.jpg)
72 / GE /
GE confidential, for internal use only
Colon VCAR EC Electronic Cleansing
•Lectura asistida por computador… aprovechalos beneficios de alta resolución •software LightSpeed VCT • ofrece potente análisis cuantitativo inteligente(biopsia virtual)
![Page 73: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/73.jpg)
73 / GE /
GE confidential, for internal use only
COLON VCAR
Detectar temprano ... puede hacer una diferencia
![Page 74: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/74.jpg)
74 / GE /
GE confidential, for internal use only
GE Healthcare
CT PERFUSION “Porque el Tiempo.... Es Cerebro”
TiP Training in Partnership
![Page 75: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/75.jpg)
75 / GE /
GE confidential, for internal use only
Cada 40 segundos, alguien sufre un infarto cerebral en America. Cada 4 minutos alguien muere por infarto cerebral
• Tercera causa de muerte, despues de las enfermadades cardiovasculares y todas las formas de cancer.
• Principal causa de incapacidad en los U.S.
• 700,000 personas sufren un stroke por primer vez o son recurrentes:
500,000 son nuevos y 200,000 son infartos recurrentes.
• De los Pacientes con stroke el 40% son hombres, 60% son Mujeres.
Source: American Stroke Association
Source: American Stroke Association
![Page 76: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/76.jpg)
76 / GE /
GE confidential, for internal use only
Impactos del Stroke
Discapacidad producida por el Stroke:
• 50% Hemiparesia
• 30% Discapacidad para caminar sin asistencia.
• 26% dependientes de por vida
• 19% afasia
• 35% Sintomatologia depresiva.
• 26% Requierean asistencia medica en casa.
Source: American Stroke Association
![Page 77: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/77.jpg)
77 / GE /
GE confidential, for internal use only
![Page 78: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/78.jpg)
78 / GE /
GE confidential, for internal use only Image courtesy of Genentech, Inc.
Localizacion para Perfusion
Prescribir de inferior a superior para incluir los ganglios basales en el nivel de la cápsula interna..
Asegúrese de que habrá una arteria y una vena que se pueden detectar en el rango de corte establecido.
![Page 79: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/79.jpg)
79 / GE /
GE confidential, for internal use only
Cerebral Blood Volume (CBV)
Define: Flujo de sangre a través de la vasculatura
Medido: ml/100 g de tejido humedo Rampa Escalas 00-100 Imagen: por defecto de color en pista es de 3 colores o talio Puh (depende del software)
![Page 80: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/80.jpg)
80 / GE /
GE confidential, for internal use only
Cerebral Hemodynamics
MTT
CBV/CBF
100
50
0
CBF
0
-50
-100
CBV
100
50
0
PHASE 0 Normal
PHASE 2 Early Failure
PHASE 3 End-Stage
Failure
PE
RC
EN
TAG
E
CH
AN
GE
PHASE 1
Unstable
Powers WJ, Ann Neurol. 1991;29;231-240
Normal Values:
*LCPP - 10 to 12 min-1
CBV - 4 ml.100g-1
CBF - 50 to 100 ml.min-1.100g-1 MTT - 5 - 6 s
Exhaustion of Reserve:
*Hemodynamic LCPP 7 min-1
(CBV) MTT 9 s
*Perfusion LCPP 4 min-1
(OEF) MTT 15 s
*The patient’s local cerebral perfusion
pressure (LCPP) has an impact on statistical
information. CT Perfusion does not account
for changes in LCPP.
![Page 81: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/81.jpg)
81 / GE /
GE confidential, for internal use only
Normal: 50-55ml/100gr/min Oligohemia: -50ml/100gr/min Penumbra:-10 a 20ml/100gr/min Centro Isquémia: -10ml/100gr/min
VALORES CBF
![Page 82: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/82.jpg)
82 / GE /
GE confidential, for internal use only
Symon L. Et al J clin Pathol suppl 1997< Austrup J, et al. Stroke 1981. Baron JC Brain 2001
![Page 83: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/83.jpg)
83 / GE /
GE confidential, for internal use only
Protocolo de Inyeccion
![Page 84: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/84.jpg)
84 / GE /
GE confidential, for internal use only
Protocolo de Perfusion
![Page 85: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/85.jpg)
GE Company confidential - For internal use only
40mm coverage 80mm coverage
Perfusion Dose vs. Coverage
Axial Scan DLP: 1261.71mGy*cm
VolumeShuttle DLP: 1645.71mGy*cm
![Page 86: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/86.jpg)
86 / GE /
GE confidential, for internal use only
CT Perfusion Axial Shuttle Blood Flow
Blood Volume
Mean Transit Time
![Page 87: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/87.jpg)
87 / GE /
GE confidential, for internal use only
Cerebral Blood Flow (CBF)
Define: Flujo de sangre a través de la vasculatura incluidos los buques de la conductancia (arterias, arteriolas, vasos capilares, vénulas, venas, senos paranasales) Medido: ml/100 g de tejido húmedo / min Rampa Escalas 000-100 Imagen: por defecto de color en pista es de 3 colores o talio Puh (depende del software)
![Page 88: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/88.jpg)
88 / GE /
GE confidential, for internal use only
Mean Transit Time (MTT) Define: La sangre atraviesa el sistema vascular a traves de diferentes longitudes de trayectoria.
Medida: En Segundos Rampa Escalas 0-15 Imagen: por defecto de color del arco iris rampa es inversa
![Page 89: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/89.jpg)
89 / GE /
GE confidential, for internal use only
Tissue Classification
Define: Analizar los tejidos afectados por un derrame cerebral. Este protocolo es el promedio calculado de los mapas de CFB y el CVB. Medido: ml/100 g de tejido
![Page 90: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/90.jpg)
GE Company confidential - For internal use only
VolumeShuttle Para estudios de neuro funcionales
Amplia la cobertuta para estudios de CTA dinámicos de neuro y análisis de perfusión en una sola adquisición 2x mayor cobertura: 80mm
Hasta 24% menos dosis con una sola inyección de contraste comparado con el método tradicional de perfusión con 40mm
Location 1
Location 2
![Page 91: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/91.jpg)
GE Company confidential - For internal use only
VolumeShuttle Para Estudio Neurológico Funcional
Incrementa la cobertura del scan para angio CT dinamica y perfusion en una sola adquisición
Alterna dos localizaciones contigua en un examen y con una inyección de contraste 2v mas cobertura llegando a 80mm Hasta 24% menos dosis comparado con la perfusion tradicional de 40mm
Localización 1
Localización 2
![Page 92: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/92.jpg)
92 / GE /
GE confidential, for internal use only
Soluciones para Extender la Cobertura Dinamica en la Imagen
120 mm
Volume Helical Shuttle: Para Imagen Neuro Cubre desde los ganglios basales a la parte superior de los ventriculos laterales
120 mm
Courtesy of Dr. Murakami, Dr. Tanaka Kinki University, Japan
VHS passes:7
Pitch: 0.984:1
Rotacion: 0.4s
Mid-scan sample rate:
3.0s
Time/pass:1.7s
Corte: 0.625mm
Visualización Dinámica de Flujo Sanguineo Cerebral
![Page 93: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/93.jpg)
93 / GE /
GE confidential, for internal use only
* Obtained by EUR-16262 EN, using an adult brain factor of 0.023*DLP
DLP = 607 mGy.cm
Eqv dosis = 1.39 mSv*
VHS: permite una secuencia dinámica para Angio CT
Hipoperfusion
![Page 94: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/94.jpg)
95 / GE /
GE confidential, for internal use only
Diagnóstico de Muerte Encefalica
Caso 1.
Tres (3) días despúes del ingreso se le realizo una CT Perfusión y angio CT, con las cuales se pudo concluir por imagen Dxca, M.E.
![Page 95: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/95.jpg)
96 / GE /
GE confidential, for internal use only
Caso 1. Paciente de 60 años de edad, hallado en la calle inconciente, es trasladado al servcio de urgencias por T.E.C. con Glaswow 3, lo remiten a Tomografía para la valoración Dxca.
La CT demuestra imágenes de hematomas intraparenquimatosos, hemorragia subaracnoidea e intraventricular, edema axonal difuso.
![Page 96: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/96.jpg)
GE Company confidential - For internal use only
Muerte Encefalica!!
![Page 97: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/97.jpg)
98 / GE /
GE confidential, for internal use only
Ayudas Diagnósticas en la Muerte Cerebral por CT
•CT Simple de rutina
•Angio Digital
•AngioCT
•Perfusion Cerebral por CT
![Page 98: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/98.jpg)
99 / GE /
GE confidential, for internal use only
CT Angio
Seguidamente del CT Perfusión, se adquirió la angioCT desde el cayado aortico hasta el poligono de willis, el cual demostro oclusion total de las carotidas a nivel de la base del cráneo y oclusión total de las vertebrales a nivel de C4.
Caso 1.
![Page 99: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/99.jpg)
100 / GE /
GE confidential, for internal use only
CT Perfusión x evaluación de M.E
Se realizó un protocolo de ct perfusión, el cuál demostro un BF, BV y TTM totalmente alterado, cualitativo y cuantitativa.
Caso 1.
![Page 100: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/100.jpg)
101 / GE /
GE confidential, for internal use only
Tabla Perfusión Caso 1.
![Page 101: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/101.jpg)
102 / GE /
GE confidential, for internal use only
Angio CT Caso 1.
![Page 102: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/102.jpg)
103 / GE /
GE confidential, for internal use only
GE Healthcare
Cardio IQ Cardio Function
TiP Training in Partnership
![Page 103: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/103.jpg)
GE Company confidential - For internal use only
LigthSpeed VCT XT
SnapShot Pulse Prospectivo
![Page 104: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/104.jpg)
105 / GE /
GE confidential, for internal use only
Why did we develop SnapShot Pulse? Appropriateness Criteria for Cardiac CTA*
Radiation Dose is one of the most important factors for Cardiac CTA imaging.
* JACC, October 2006
![Page 105: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/105.jpg)
106 / GE /
GE confidential, for internal use only
Con dosis bajas, es posible brindar un buen Diagnostico?
LightSpeed VCT XT Con SnapShot Pulse
Dose: 0.8 mSv* BMI= 19.1 kg/m²
100 kV, 400 mA Scan: 0.7 sec, 105 mm DLP 48 mGy.cm
*Obtained by EUR-16262 EN, using a chest factor of 0.017*DLP
The low-dose results for this patient may not be typical of a broad patient population
![Page 106: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/106.jpg)
107 / GE /
GE confidential, for internal use only
SnapShot Pulse Hasta 83% Menos Dosis y Calidad Mejorada
Adquisición cardiaca Prospective axial gatillada con ECG
•40mm de cobertura de alta resolución
•Monitoreo y gatillado en tiempo real de la FC
•Completa el scan via modo step-and-shoot
•No es una fase única, sino un rango de fase
•Menos Dosis y Mejorada Calidad de Imagen
•Resolución Temporal de 175msec.
![Page 107: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/107.jpg)
108 / GE /
GE confidential, for internal use only
108
SNAP SHOT
SEGMENT
SNAP SHOT
BURST
PULSE
De 5 a 95 %
De 65 a 95 %
![Page 108: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/108.jpg)
109 / GE /
GE confidential, for internal use only
5
GE Healthcare LightSpeed VCT XT
LPM Cardiacos
Hasta 83% reduccion de dosis con aumento de QI.
2.2 mSv *
*Obtained by EUR-16262 EN, using a chest factor of 0.017*DLP
![Page 109: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/109.jpg)
110 / GE /
GE confidential, for internal use only
Calidad mejorada de la imagen: Comparación clínica
Patient:
• Hypertension, high cholesterol, diabetic
• Heart rate: 52-54 bpm
Helical Scan
• Tube: 700 mA w/ 40% modulation & 120 kVp
• Gantry Rotation: 0.35 sec
SnapShot Pulse Scan
• Tube: 750 mA & 120 kVp
• Gantry Rotation: 0.35 sec
Same or Better IQ
62% dose reduction
Images Courtesy Dr. Dowe, Atlantic Imaging
Helical SnapShot Pulse
Mejorado IQ
83% menos Dosis
* Obtained by EUR-16262 EN, using a chest factor of 0.017*DLP
Z axis resolution
In Plane resolution
Low contrast resolution
Dose* (mSv)
![Page 110: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/110.jpg)
111 / GE /
GE confidential, for internal use only
![Page 111: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/111.jpg)
112 / GE /
GE confidential, for internal use only
El comité de BEIR calcula que una dosis de mSv 10 causaría cáncer en un promedio de uno de cada 1.000 , mientras que la Comisión Internacional en la Protección Radiológica estima que una dosis de mSv 10 causaría en cáncer en un promedio de uno de cada 2.000.
![Page 112: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/112.jpg)
113 / GE /
GE confidential, for internal use only
![Page 113: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/113.jpg)
114 / GE /
GE confidential, for internal use only
M38, 1.69m, 65 kg - BMI= 22.7 kg/m² Patient with atypical angina and risk factors 100 kVp, 600 mA X-Ray exposure: 1.5 sec, 139 mm Dose: 2.6 mSv* / DLP= 153 mGy.cm
Stenosis on soft plaque on LAD1 (50-60%) Occlusion of RCA
2.6 mSv * Coronary CTA with SnapShot Pulse
LAD
LAD
RCA
RCA
LAD
D2
Courtesy of Dr. Sablayrolles, CCN , France * Obtained by EUR-16262 EN, using a chest factor of 0.017*DLP
![Page 114: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/114.jpg)
115 / GE /
GE confidential, for internal use only
M77, 170cm, 55 kg-BMI = 19 kg/m² Patient with atypical chest pain and
negative stress thalium 100 kV, 600 mA X-Ray exposure: 0.7 sec, 105 mm Dose: 1.2 mSv** / DLP= 72 mGy.cm
LAD
1.2 mSv **
Coronary CTA with SnapShot Pulse
LAD
Courtesy of Dr. Sablayrolles, CCN , France Severe stenosis of the LAD2 (90%)
LAD
RCA
LCx
* Obtained by EUR-16262 EN, using a chest factor of 0.017*DLP ** The low dose results for this patient may not be typical of a broad patient population
![Page 115: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/115.jpg)
116 / GE /
GE confidential, for internal use only
VCT XTe: Ultra Baja dosis Cardiaca con ASIR 0.34 mSv
Protocol
• 20 mGy.cm (0.34 mSv)
• kV: 100
• BMI: 21
• ASIR recon time < 30s
*Obtained by EUR-16262 EN, using a chest factor of 0.017*DLP
Images courtesy of Dr. James Earls
![Page 116: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/116.jpg)
117 / GE /
GE confidential, for internal use only
Dosis Baja en CT Cardiaco
CT750HD – Reduccion de Dosis Menos Dosis
• 50% dose reduccion en cuerpo entero
• 83% reduccionde dosis para CT Cardiaco
Provee Calidad en la Imagen Reduce Artefactos
CT750 HD, ASIR w/ SSP
provides more than 83% less dose
2/2009
Non-HD, CCTA
8/2006
DLP = 115 mGy
Dose = 1.9 mSv
DLP = 1419 mGy
Dose = 19.8 mSv Obtained by ICRP using a chest factor of 0.017*DLP
Calcium Body Habitus Stenosis % Identify restenosis High heart rates Dose
![Page 117: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/117.jpg)
118 / GE /
GE confidential, for internal use only
0.46mSv*
Courtesy of Dr Leipsic St. Paul’s Vancouver Canada Obtained by ICRP using a chest factor of 0.014DLP*
CardIQ Reveal - Relative perfusion color map
Mayor Resolucion Cardiaca en la Industria*
- Reduce Artefacto de Calcio
- Mejora Confiabilidad Diagnóstica
- Mejora visualización de stent
- Visualiza vasos nativos e injertos
*Based upon internal test data comparing Discovery CT750 HD cardiac half-scan spatial resolution to data from Advanced CT Scanners for Coronary Angiography, ImPACT Report CEP10043, March, 2010, available at http://www.impactscan.org
CardIQ Reveal - PlaqueID
High Definition stent imaging
Ultra low dose imaging
Alta Definición Cardiaca – 18.4 lp/cm
Discovery CT750 HD
![Page 118: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/118.jpg)
119 / GE /
GE confidential, for internal use only
Resolucion Espacial Mejora la capacidad Diagnostica en CT
Reduce artificio de Calcio
Visualizacion dentro del Stent
Detecccion de Re-Estenosis
Discovery™ CT750 HD
Calcium Body Habitus Stenosis % Identify restenosis High heart rates Dose
![Page 119: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/119.jpg)
120 / GE /
GE confidential, for internal use only
LightSpeed VCT 07/22/2008
Discovery™ CT750 HD 7/23/2008
Ver maswith Higher Resolution Reduce Artificio de Calcio– PLB Lesion
Calcium Body Habitus Stenosis % Identify restenosis High heart rates Dose
![Page 120: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/120.jpg)
121 / GE /
GE confidential, for internal use only
DLP 276 Dose: 3.84mSv** Dose: 4.7mSv*
3.84mSv**
BAJA DOSIS SnapShot w/ASIR Anatomia & Funcional 3.84 mSv
* Obtained by EUR-16262 EN, using a chest factor of 0.017*DLP
** Obtained by ICRP using a chest factor of 0.014*DLP
Calcium Body Habitus Stenosis % Identify restenosis High heart rates Dose
![Page 121: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/121.jpg)
122 / GE /
GE confidential, for internal use only
Discovery CT750 HD Coronary CTA SnapShot Pulse w/ASIR 0.67mSv**
M60 chest pain 100 kV, 250mA HR 49-52 bpm ASIR 40%
DLP 47 mGy/cm Dose: 0.81mSv* Dose: 0.67mSv** Obtained by EUR-16262 EN, using a chest factor of 0.017DLP*
Obtained by ICRP using a chest factor of 0.014DLP**
Courtesy of James Earls M.D. Fairfax Virginia
![Page 122: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/122.jpg)
123 / GE /
GE confidential, for internal use only
GE Healthcare
Cardio IQ Cardio Function
TiP Training in Partnership
![Page 123: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/123.jpg)
124 / GE /
GE confidential, for internal use only
CardIQ Function Xpress
Simplicidad. Automatico, facil
![Page 124: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/124.jpg)
125 / GE /
GE confidential, for internal use only
Vista tipo IVUS Con un solo clic aparece una imagen transversal con códigos de colores que revelan una placa no calcificada de lumen y calcio para mostrar el nivel de arterosclerosis.
![Page 125: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/125.jpg)
126 / GE /
GE confidential, for internal use only
16-Oct-12 126
VALVULA AORTICA
VALVULA CALCIFICADA
![Page 126: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/126.jpg)
127 / GE /
GE confidential, for internal use only
Proceso Automatico
• LV, RV & LA Fraccion de Eyeccion
• Analisis del volumen todas la camaras
• Analisis de la pared miocardica
Facil revision e interpretacion
• Cargar datos de múltiples fases
• Generación de informes con analisis importados
Mejorar el tiempo de procesamiento de lectura & Revisión en tiempo real
![Page 127: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/127.jpg)
128 / GE /
GE confidential, for internal use only
Automatico Fraccion de Eyeccion Izq & Der Selección de protocolo proporciona:
• Automatica deteccion de las fases de ED/ES
• Deteccion de SV (Stroke volume)
• Basal, medio, distal y dos camaras en vista de eje largo y eje corto l
Fraccion de Eyeccion RV :
•Útil en anomalías congénitas y defectos del Septo auricular
• Diagnosticar los síntomas de la enfermedad pulmonar (EPOC, hipertensión pulmonar)
![Page 128: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/128.jpg)
129 / GE /
GE confidential, for internal use only
Volume Analisis 4 Camaras
Carga automática para su revisión:
•Contorno de cámaras individuales
•Conjuntos de datos único o múltiples fases
•Diferentes cámaras del corazón para el análisis de volumen
LV RV LA RA
LV Normal Values*
EDV=150ml
ESV=75ml
SV=<80ml
EF=50-65%
Wall Thickness=10 mm
Cardiac Output= 5-6 l/min
![Page 129: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/129.jpg)
130 / GE /
GE confidential, for internal use only
Analisis de la pared del Miocardio Automatizado de análisis funcional cardíaca, incluyendo:
•LV myocardial thickness
•Myocardial mass analysis
•Wall motion
•Bullseye graphs for wall motion abnormalities
Auto Deteccion de las paredes del Endo y EpiCardio
![Page 130: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/130.jpg)
131 / GE /
GE confidential, for internal use only
Freedom Edition
![Page 131: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/131.jpg)
132 / GE /
GE confidential, for internal use only
SNAPSHOT FREEZE
![Page 132: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/132.jpg)
133 / GE /
GE confidential, for internal use only
GE Healthcare
Avances en CT
TiP Training in Partnership
![Page 133: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/133.jpg)
134 / GE /
GE confidential, for internal use only
134 of 38 12/2009
FBP Vs. ASiR
Que es ASiR?
![Page 134: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/134.jpg)
135 / GE /
GE confidential, for internal use only
Imagen a menor dosis sin compromiso ASIR
Adaptive Statistical Iterative Recon (ASIR) entrega un proceso sofisticado de reconstrucción iterativa
Hasta 50% menos dosis sin perdidad de IQ
Hasta 30% de mejoria en low contrast detectability
Reducción del Ruido
Mejor Ejecución Diagnostica a baja dosis
Menor Dosis en Imagen de Rutina
Menor Dosis en Imagen Pediátrica
ASIR (40% dose reduction)
non-ASIR
![Page 135: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/135.jpg)
136 / GE /
GE confidential, for internal use only
ASiR
• Mejora de la calidad de
imagen, • menos ruido
• mejor contraste especialmente el área de pelvis.
ASiR Conventional
![Page 136: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/136.jpg)
137 / GE /
GE confidential, for internal use only
ASiR reduces noise and provides better lesion visualization
With ASiR (both images with original slice thickness)
120 kV, 180 mAs, pitch 1.75 6 s acquisition
![Page 137: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/137.jpg)
138 / GE /
GE confidential, for internal use only
1992 1998 2002
Helical
Multi-
Slice
Isotropic
sub-mm
Multi-
phase
liver
CAP
Exam
Vascular
Dual Source
Axial Area Detector
Advantages •Temp resolution •Dual Energy
(Mechanical)
VCT Gemstone Technology
Advantages
•Resolution •Noise reduction •Artifact reduction •101 Spectrums of energy •Lower Dose
Advantages
•Perfusion •Coverage
Evolucion de la Tecnologia en CT
Ceramic Based
Gemstone Based
Head
1976
Axial
![Page 138: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/138.jpg)
139 / GE /
GE confidential, for internal use only
GSI/Dual Source
1 tube & 1 detector
Fast kVp switching
2 tubes & 2 detectors
1 tube & 1 detector
Rotate/rotate (WIP)
80 kVp 140 kVp
140 kVp
80
kV
p
![Page 139: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/139.jpg)
140 / GE /
GE confidential, for internal use only
Adquisición de GSI y Descomposición Material Interleaved High- and Low-kVp Projections GSI Data Acquisition
140 kVp
MD Iodine MD Water
Image Reconstruction
....)()()()()()()()()( 1
2
1
2
1111 iPiPiPiPiPiiPiiP highlowhighlowhighlow ....)()()()()()()()()( 2
2
2
2
2222 iPiPiPiPiPiiPiiP highlowhighlowhighlow iodine
water
Attenuation-to-material density transformation
Iodine Projections Water Projections
70 keV
Monochromatic Generation
2 data sets
Low kVp Projections High kVp Projections
GE Company Confidential - For internal training use only
80 kVp
Image reconstruction
![Page 140: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/140.jpg)
141 / GE /
GE confidential, for internal use only
GE Gemstone
GOS 3.0ms
0.030ms
GE - 100 Veces más rapido
GE Gemstone
GOS 0.004%
0.001%
GE - 4 Veces más rápido
Velocidad Primaria (IQ & Speed)
Tiempo de respuesta del cintilador a los rayos X.
Imagen subsegundo de Alta Resolución
Imagen de Energía Dual con un rápido y único Detector
Afterglow (IQ)
Tiempo de recuperación del cintilador cuando los rayos X se apagan
Incrementa la resolución espacial y reduce los artefactos
Si… el detector es realmente hecho de Gemstones Incrementada Resolución Espacial e Imagen de Energía Dual
Mejora la Resolución Espacial hasta 47%
7.3 kHz Vistas – Mejora la calidad de imagen fuera del centro
La velocidad del Gemstone permite la “Imagen Espectral”
Discovery CT750 HD
![Page 141: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/141.jpg)
142 / GE /
GE confidential, for internal use only
Resolución de 0.23 mm Mejor resolución espacial cardiaca en la industria – 18.2 lp/cm Hasta 312.5 mm (Equivalente a 500 slices) de cobertura en
imagen dinámica y alta cobertura en perfusión Imagen Espectral – Energia Dual con 0.5msec rápido switch de
kV ASiR – Reconstrucción Iterativa sobre los datos crudos con
hasta 35 fps Hasta 50% menos dosis de radiacion vs equipo tradicional
Ver mas Resolución Espacial
Conocer mas Cobertura
Conocer mas Imagen Espectral
Baja Dosis
Presionando las fronteras en todas las dimensiones
* Veo is 510(k) pending. Not available for sale in the U.S.
Discovery CT750 HD
![Page 142: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/142.jpg)
143 / GE /
GE confidential, for internal use only
Caracteriza Pequeñas Lesiones ayuda a caracterizar lesiones resaltadas y no resaltadas
CT Cuantitativo – caracteriza la composición química como en los calculos renales
Reduce Artefactos de Metal y de Dispersión del Rayo para una mas clara
visualización anatómica
Optimiza la vizualización del contraste con la imagen monocromática
GSI Gemstone Spectral Imaging- Beneficios Clínicos Para interpretaciones diagnósticas mas informadas
Discovery CT750 HD
![Page 143: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/143.jpg)
144 / GE /
GE confidential, for internal use only
Imagen Espectral Gemstone
CT750 HD – Saber Mas
Imagen Espectral
Unico Tubo con switch de energía dual ultra rápido para registro helicoidal casi perfecto
Sub msec de switch entre las dos energías (0,5msec)
Total 50 cm FOV
Detector Gemstone permite la separación Material
Identificación Material
Verdaderos CT #’s
Procesamiento Basado en “Datos Crudos” vs Datos de Imagen
Optimizada relacion contraste ruido
Imagenes Monocromáticas (espectral)
Reduce Artefactos de Dispersión del Rayo
165 veces más rapido que la Tecnología de Energia Dual
existente
![Page 144: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/144.jpg)
145 / GE /
GE confidential, for internal use only
CT750 HD – Conocer Mas Extiende su cobertura – Equivalente a 500 cortes
CT Dinamico - Cobertura
![Page 145: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/145.jpg)
146 / GE /
GE confidential, for internal use only
La Cobertura es NO limitada por el Detector Volume Helical Shuttle
Cobertura Extendida para Perfusion y Angiografia Dinamica por CT
4D CTA: Hasta 312.5mm
Perfusion: Hasta 140mm (cuerpo)
Hasta 120mm (cabeza)
Evalua la extensión de la enfermedad tanto en metodos anatómicos como funcionales
Reconstruccion & Control de Scan en Tiempo Real
Adquisición Helicoidal de Velocidad Variable
Reconstrucción de Pitch Dinamico
ASIR y auto mA (accel/decel) para manejar dosis
![Page 146: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/146.jpg)
GE Company confidential - For internal use only
Gracias
![Page 147: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/147.jpg)
148 / GE /
GE confidential, for internal use only
GE Healthcare
TiP Training in Partnership
![Page 148: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/148.jpg)
149 / GE /
GE confidential, for internal use only
¿¿¿ Protocolos, Post - proceso ???
![Page 149: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/149.jpg)
150 / GE /
GE confidential, for internal use only
Angio CT Coronaria… CT Perfusion??
![Page 150: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/150.jpg)
151 / GE /
GE confidential, for internal use only
Angiotomografía de arterias coronarias.
Evolución tecnológica
• 1998: cuatro detectores.
• 2002: 16 detectores.
• 2004: 64 detectores.
Es capaz de evaluar todos los segmentos de relevancia clínica descritos AHA, +1.5 mm*
*Circulation 2005; 112: 2318-2323.
![Page 151: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/151.jpg)
152 / GE /
GE confidential, for internal use only
4 detectores 16 detectores 64 detectores
Sensibilidad 78-85 % 95 %
94 %
Especificidad 76-95 % 86 % 97 %
VPN 99 %
VPP 87 %
Radiographics. 2003;23:S111-S125 Eur Heart J 2005; 26: 1482-87
![Page 152: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/152.jpg)
153 / GE /
GE confidential, for internal use only
Score calcico:
La presencia de calcio coronario es un factor de riesgo que se correlaciona de manera estrecha con el volumen y extensión de la placa ateroesclerosa.
Unidades Agatston Riesgo para enfermedad coronarias Categoría
(Rumberger)
0 Muy bajo riesgo I
0-10 Bajo Riesgo II
10-100
Riesgo moderado III
100-400 Riesgo moderadamente alto IV
+ 400 Riesgo alto
V
Radiology 2008;246:90-98
![Page 153: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/153.jpg)
154 / GE /
GE confidential, for internal use only
Método de Evaluación
![Page 154: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/154.jpg)
155 / GE /
GE confidential, for internal use only
![Page 155: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/155.jpg)
156 / GE /
GE confidential, for internal use only
Test Positivo Calcio Coronario
Agatston absoluto
1. < 10 Mínimo
2. 10 – 99 Moderado
3. 100 – 400 Aumentado
4. > 400 Extenso
![Page 156: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/156.jpg)
157 / GE /
GE confidential, for internal use only
Que Hacer con Agatston ???
Score de Agatston y Recomendaciones
Score “0” No placa Aterosclerótica
Ejercicio, dieta, normal
Score 1 a 10 Riesgo placa mínima
Vida normal, control FRC.
Score 11 a 100
Placa leve
Control FRC, recomendado de NCEP
ASA. /día, recomendada
Score 101 a 400
Placa moderada
Control FRC, test isquemia no invasivo
ASA + estatina/día recomendadas
Score > 401
Placa cargada significativa > 75%
Modificar “agresiva” FRC
ASA + estatinas/dia recomendadas. Estudio funcional
![Page 157: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/157.jpg)
158 / GE /
GE confidential, for internal use only
META-ANALISIS: 1,286 PACIENTES PATIENT-BASED DETECTION > 50% STENOSIS 64 DETECTORES
SENSIBILIDAD = 99% ESPECIFICIDAD = 89% VALOR PREDICTIVO POSITIVO = 81%
- 56 to 100% depending on pre-test probability
- Number of trials exclude segment with high calcium - Selected patient populations VALOR PREDICTIVO NEGATIVO = 100%
“64-Slice Computed Tomography Angiography in the Diagnosis and Assessment of Coronary Artery Disease: A Systematic Review and Meta-analysis” G Mowatt Heart 2008;94:1386-1393
Cual es la confiabilidad de un Angio CT Coronario?
*This is a cross vendor view and does not represent the performance of any specific CT product
![Page 158: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/158.jpg)
159 / GE /
GE confidential, for internal use only
Cuáles son las razones para un bajo VPP?
Actuales Desafios en Imagen de CT Coronario :
1. Calcio
2. Stents Intracoronarios
3. Moderada a Severa Estenosis
4. Altas Frecuencias Cardiacas
5. Bypass grafts
- Vasos Nativos…muy calcificados
6. Habitos Corporales
Resolución Espacial
Resolución Espacial
Resolución Espacial
Resolución Temporal
Resolución Espacial
Ruido Imagen /Dosis
![Page 159: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/159.jpg)
160 / GE /
GE confidential, for internal use only
Alta Sensibilidad Menor Especificidad Alto VPN ……………………………….. Manos Experimentadas
![Page 160: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/160.jpg)
161 / GE /
GE confidential, for internal use only
Angio Coronaria
Permite evaluar de forma no invasiva las arterias coronarias.
Luz coronaria como la pared arterial y estructuras circundantes, trayecto, origen, variantes anatómicas, placas ateromatosas, composición de la placa ateromatosa, grado de
estenosis. Score calcico.
![Page 161: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/161.jpg)
162 / GE /
GE confidential, for internal use only
La angio Tc permite detectar el calcio en etapas tempranas, cuando aun no hay estenosis luminal, evidenciando el crecimiento excéntrico de la placa.
Este hallazgo ha denotado la importancia de estudiar a pacientes asintomático con riesgo de enfermedad ateromatosa coronaria.
![Page 162: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/162.jpg)
163 / GE /
GE confidential, for internal use only
Protocolo de estudio
Evaluación cardiológico, sospecha clínica de enfermedad coronaria.
Valoración radiológica: preparación del paciente.
Beta bloqueador: FCF 60 latidos por minuto.
Vasodilatador sublingual.
Medio de contraste de alta concentración: 60 -80ml, 350mg/ml (20ml Bolus Timing) o Smartprep
![Page 163: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/163.jpg)
164 / GE /
GE confidential, for internal use only
Protocolos
Score calcico.
Angiotomografía de arterias coronarias.
Tiempo en sala 10 minutos.
Adquisición angiográfica: 8 segundos.
Sincronización cardiaca.
Reconstrucciones retrospectivas.
Postproceso de la imagen, en estaciones de trabajo.
![Page 164: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/164.jpg)
165 / GE /
GE confidential, for internal use only
Consideraciones Tecnicas para la angio Coronaria:
Preparacion del Paciente Seleccion del Protocolo • Retropectivo (SS Segment/SS Burst/SS Burst Plus • Prospectivo ( Snapshot Pulse) Inyector – Medio de Contraste IMC
![Page 165: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/165.jpg)
166 / GE /
GE confidential, for internal use only
166
SNAP SHOT
SEGMENT
SNAP SHOT
BURST
PULSE
De 5 a 95 %
De 65 a 95 %
![Page 166: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/166.jpg)
GE Company confidential - For internal use only
Z lo
ca
tio
n
Time
Indeseable fase cardiaca
Dosis de Radiación Adquisición cardiaca en el modo helicoidal
![Page 167: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/167.jpg)
GE Company confidential - For internal use only
Z lo
cati
on
Time
mA Constant % Ma Modulation
Dosis de Radiación Adquisición cardiaca con modulación de mA
![Page 168: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/168.jpg)
GE Company confidential - For internal use only
Z lo
cati
on
Time
En el eje “Z” sin rayos X en la fase
indeseable.
Dosis de Radiación Adquisición cardiaca con Snapshot Pulse
![Page 169: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/169.jpg)
170 / GE /
GE confidential, for internal use only
170
250ms
R-R Interval
Trigger
Delay
Q
R
S
Systole Diastole
75%
5% 95%
5 %
15%
25%
35%
45%
55%
65%
75%
85%
95%
75%
![Page 170: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/170.jpg)
171 / GE /
GE confidential, for internal use only
Cómo saber si usted tiene sobrepeso u obesidad
171
GE Health Care, Mexico
Srvicio Medico
El Indice de Masa Corporal es un índice del peso de una persona en relacion con su altura. A pesar de que no hace distinción entre los componentes grasos y no grasos de la masa corporal total, éste es el método más práctico para evaluar el grado de riesgo asociado con la obesidad
![Page 171: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/171.jpg)
172 / GE /
GE confidential, for internal use only
El Índice de Masa Corporal (IMC) que relaciona el peso con la altura para una determinada edad y sexo
Condición de peso
IMC
Bajo peso
18.5 o menor
Normal
Entre 18.6 y 24.9
Sobrepeso
Entre 25.0 y 29.9
Obesidad moderada
Entre 30.0 y 39.9
Obesidad extrema
40.0 o mayor
El IMC resulta de la división de la masa en kilogramos entre el cuadrado de la estatura expresada en metros.
![Page 172: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/172.jpg)
173 / GE /
GE confidential, for internal use only
PESO KG
48 50 52 54 56 58 60 62 64 66 68 70 72 74 76 78 80 82 84 86 88 90 92 94 96 98 100 102 104 106
148
150
152
154
156
158
160
162
T 164
A 166
M 168
A 170
Ñ 172
O 174
176
CM 178
180
182
184
186
188
190
192
194
196
198
200
202
Ma 600 Ma 650 Ma 700 Ma 700 Ma 800
Kv 100 Kv 100 Kv 100 Kv 120 Kv 120
Iode 300 Iode 300 Iode 350 Iode 350 Iode 350
Dlp 160 Dlp 175 Dlp 205 Dlp 325 Dlp 350
Msv 2.75 Msv 3 Msv 3.5 Msv 5.5 Msv 6 173
Indice de masa Corporal
![Page 173: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/173.jpg)
174 / GE /
GE confidential, for internal use only
16-Oct-12 174
EQUIPO Y MATERIAL
CARDIAC TRIGGER MONITOR
INYECTOR AUTOMATICO DE DOBLE CABEZAL
JERINGAS CON INYECCION SIMULTANEA
MEDIO DE CONTRASTE NO IONICO 350
SOLUCION FISIOLOGICA 5 %
CARRO DE EMERGENCIA ELECTRODOS
![Page 174: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/174.jpg)
175 / GE /
GE confidential, for internal use only
SUP 60
INF 300 INF 300
Realizar scout o imagen digital con los parámetros y la posición del paciente correspondientes para la realización de este estudio
![Page 175: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/175.jpg)
176 / GE /
GE confidential, for internal use only
VCT 64
Bomba Injetora Monitor Cardíaco
![Page 176: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/176.jpg)
177 / GE /
GE confidential, for internal use only
![Page 177: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/177.jpg)
178 / GE /
GE confidential, for internal use only
EVALUACION DE HEMODUCTOS Y STENTS
Identifica el trayecto de los hemoductos desde su nacimiento hasta su anastomosis distal.
Evalúa el estado de la pared y la permeabilidad de la luz arterial.
RadioGraphics 2006;26:887-904
![Page 178: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/178.jpg)
179 / GE /
GE confidential, for internal use only
![Page 179: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/179.jpg)
180 / GE /
GE confidential, for internal use only
![Page 180: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/180.jpg)
181 / GE /
GE confidential, for internal use only
![Page 181: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/181.jpg)
182 / GE /
GE confidential, for internal use only
![Page 182: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/182.jpg)
183 / GE /
GE confidential, for internal use only
![Page 183: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/183.jpg)
184 / GE /
GE confidential, for internal use only
![Page 184: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/184.jpg)
185 / GE /
GE confidential, for internal use only
Abdomen CTA
Protocol
- rotation time: 0.8sec
- scan configuration: 1.25*16 - kv:120kv - mA: 50 to 350mA -Noise index : 9.6 - pitch: 1.75 - kernel: STD
- Scan time : 25 sec
![Page 185: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/185.jpg)
186 / GE /
GE confidential, for internal use only
Peripheral CTA
Protocol - rotation time: 0.8sec
- scan configuration: 1.25*16 - kV:120kv
- mA: 50 to 350mA - Noise index : 9.6 - pitch: 1.75 - kernel: STD
- Scan time : 25 sec
Curved left 3D MIP
![Page 186: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/186.jpg)
187 / GE /
GE confidential, for internal use only
![Page 187: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/187.jpg)
188 / GE /
GE confidential, for internal use only
“Direccionado a nuevos diseños en la aplicacion,
explorando el poder de la Alta resolucion, y
cobertura volumetrica.”
VCAR
olu
me
om
pute
r
ssis
ted
ead
ing
![Page 188: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/188.jpg)
189 / GE /
GE confidential, for internal use only
NPS
1. DE QUE HABLAMOS?
2. COMO SE DIAGNÓSTICA
3. QUE APORTA EL LUNG VCAR?
![Page 189: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/189.jpg)
190 / GE /
GE confidential, for internal use only
OBJETIVOS
•POSIBILIDADDE UN DIAGNÓSTICO PRECOZ
•POR DIAGNÓSTICO RADIOLÓGICO DETERMINAR CON MAYOR PRECISIÓN MALIGNIDAD O BENIGNIDAD
![Page 190: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/190.jpg)
191 / GE /
GE confidential, for internal use only
1. DE QUE HABLAMOS
El nódulo pulmonar solitario (NPS), se define como la imagen radiológica redondeada de aumento de densidad, rodeada de tejido pulmonar y menor de tres centímetros que no se acompaña de adenopatías ni atelectasias.
Los principales objetivos de su estudio son: realizar un diagnóstico precoz del cáncer de pulmón y determinar con la mayor precisión posible, las posibilidades de malignidad o benignidad de la imagen, con objeto de evitar el mayor número de pruebas cruentas para el paciente
![Page 191: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/191.jpg)
192 / GE /
GE confidential, for internal use only
Lung VCAR
Analisis Automática de nódulos: • % de Crecimiento • Tiempo de duplicación • Los volúmenes (de componentes sólidos, de componentes no sólido, Volumen Total) • Máxima distancia en 3D (eje longitudinal x, y, z) • Las mediciones 2D • unidades Hounsfield (Min, Max, de media)
![Page 192: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/192.jpg)
193 / GE /
GE confidential, for internal use only
Segments All Nodule Types
Solid Part Solid Non Solid
![Page 193: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/193.jpg)
194 / GE /
GE confidential, for internal use only
Lung VCAR •Comparación de inicial y seguimiento del
estudio
exámenes anteriores y actuales son
registrados automáticamente y se
muestra una comparación
correlacionados.
Lung VCAR proporciona tamaño del
nódulo de seguimiento junto con el% de
aumento y tiempo de duplicación para ayudar a manejar los nódulos en el
tiempo
![Page 194: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/194.jpg)
195 / GE /
GE confidential, for internal use only
Protocolo
Protocolo Recomendado
• Cortes 1.25 mm o 0.625mm
• High resolution recon kernal: bone
• 80 MAS
• 120kV
• Sub-segundo rotacion (0.5)
• Una sola apnea
![Page 195: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/195.jpg)
196 / GE /
GE confidential, for internal use only
Mujer de 65 años, fumadora. Le detectaron nodulo por
Radiografía de tórax.
1st examen calcularon volume 769 mm3.
En la imagen axial del segundo examen el nodulo no muesttra aparente cambio en el tamaño, realizado 3 meses y 19 dias mas
tarde. La variacion de la medida manual del diametro fue < 1
mm.
La medidad de volumen del software en la 2a CT adquisicion fue
1038 mm3, representando un incremento 35% con tiempo promedio al doble de su tamaño en 255 dias, sugeriendo
malignidad. Examen patologico despues de la reseccion
quirurgica demostro un adenocarcinoma pulmonar estadio I
Revel MP, Bissery A, Bienvenu M, Aycard L, Lefort C and Frija G
1st exam volume:769 mm3
Evaluacion del Nodulo
2nd exam Volume: 1038 mm3
Doubling time: 255 days % growth: 35%
![Page 196: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/196.jpg)
197 / GE /
GE confidential, for internal use only
![Page 197: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/197.jpg)
198 / GE /
GE confidential, for internal use only
Lung VCAR Ubicacion automatica de Nodulos
Digital Contrast Agent
DCA Iluminacion de estructuras de interés para su revisión
Con el tamaño de filtro ajustable para disminuir el tiempo de revisión.
Mejorar su productividad Complementos de la lectura visual, ofreciendo automáticamente la visualización y el análisis cuantitativo
![Page 198: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/198.jpg)
199 / GE /
GE confidential, for internal use only
COLON VCAR
Detectar temprano ... puede hacer una diferencia
![Page 199: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/199.jpg)
200 / GE /
GE confidential, for internal use only
Cancer de Colon.. #2 Causa de Muerte
El cáncer colorrectal es tratable cuando se
detecta en una etapa temprana localizada
sin embargo del 90%, sólo el 39% de los cánceres de colon son diagnosticado en
esta etapa
Limitantes
Las barreras de los pacientes:
Percibir que el procedimiento doloroso,
incómodo, invasivo, vergonzoso, etc...
Mala preparacion.
Los pacientes que se consideran saludable evitar
exámenes que les hará sentir incómodos o
imponer fuerte estrés
![Page 200: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/200.jpg)
201 / GE /
GE confidential, for internal use only
Colon Procedures Patient Compliance
Patient Feedback
FECAL BLOOD TESTING
SIGMOIDOSCOPY
DOUBLE CONTRAST BARIUM ENEMA
COLONOSCOPY
CT COLONOGRAPHY
Compliance Comparison
![Page 201: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/201.jpg)
202 / GE /
GE confidential, for internal use only
Procedimientos del Colon Test Fecal () Provee una prueba efectiva de Sangre en las heces
() Seguro y Economico
() Solo es positivo cuando sangra (La mayogranria de los polipos)
() 30-40% sensibilidad
Sigmoidoscopia () Procedimiento Ambulatorio
() Menos costoso que la Colonoscopia
() Solo explora el recto-sigmoide (y la izquierda del colon)
() 40-65% sensibilidad
CT Colonography () Facil procedimiento de limpieza
() Excelente como screening
() preparacion e insuflacion completa del colon
() 75-85% sensitivity (polipos 5-9mm)
La sensibilidad de la Colonografia es el
standard de Oro. . . CT
Colonography deberia avanzar para mejorar la experiencia del paciente
y la sensibilidad del diagnostico
Colonoscopia Optica () 90% sensitivity (polipos menores 5-9mm)
() Permite realizar Muestras Bx
() Preparacion y parcial insuflacion del colon puede requerir anestesia
() Incompleta en el 15-20% de los casos
DC Barium Enema () Estudio Completo del colon
() Baja Dosis
() 50-70% sensitivity (polipos mayores a 10mm)
![Page 202: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/202.jpg)
203 / GE /
GE confidential, for internal use only
Beneficios para el Paciente
Más comfortable… menos ansiedad… una mejor experiencia del paciente. alternativa no invasiva a colonoscopia toma mucho menos tiempo que una colonoscopia no requiere sedación
![Page 203: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/203.jpg)
204 / GE /
GE confidential, for internal use only
Pacientes 50 años de edad y mayor Pacientes 40 años de edad. Historia Personal o familiar de cáncer de colon: colonoscopia incompleta Contraindicaciones (CO) colonoscopia optica es actualmente todavía el procedimiento de elección para pacientes de alto riesgo
Indicaciones para CT Colonography
![Page 204: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/204.jpg)
205 / GE /
GE confidential, for internal use only
CTC Beneficios
Completamente sincronizado, orientado y personalizado en la lectura protocolos prono y supino Automatica y rapida segmentacion Visualizacion de diseccion 360 del colon Reporte Integrado
![Page 205: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/205.jpg)
206 / GE /
GE confidential, for internal use only
VCARTM – LECTURA volumen interactivo : un diseño innovador
que aprovecha plenamente el tremendo poder de alta
resolución, imágenes de volumen CT y PET/CT. ¿El resultado?
Diagnostico… informe
QUE ES VCAR?
Volume Computer Assisted Reading (VCAR) La herramienta para cuantificar la enfermedad desde un volumen de datos encontrar - equipo asistido por visualización aislar - precisa segmentación medida - potente análisis cuantitativo interactivo
![Page 206: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/206.jpg)
207 / GE /
GE confidential, for internal use only
Colon VCAR
Detección automática de posibles lesiones umbral de detección de color mejora en 3D - Auto disección en el actual proceso
![Page 207: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/207.jpg)
208 / GE /
GE confidential, for internal use only
Colon VCAR EC Electronic Cleansing
•Lectura asistida por computador… aprovechalos beneficios de alta resolución •software LightSpeed VCT • ofrece potente análisis cuantitativo inteligente(biopsia virtual)
![Page 208: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/208.jpg)
209 / GE /
GE confidential, for internal use only
Patient Size Shape Centering Anatomy
patient
X-ray kV mA Scan time Helical pitch Geometry
detector
Detection Slice width
X-ray filter
Bowtie filter
Filtration Bowtie Beam quality
mA modulation regulates only the tube current (mA)
Factors Affecting Image Noise
DAS Recon- struction
Recon Algorithm Helical Slice width
Helical weighting Special filters (AAR)
Post Processing Image space filters 3D, MPR, MIPs
Dose Vs. Image Noise
![Page 209: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/209.jpg)
210 / GE /
GE confidential, for internal use only
¿Por qué centrar el paciente importante?
AutomA & SmartmA
Ruido de la imagen puede aumentar considerablemente si el paciente es mis-centered. Esto ocurre porque La radiacion efectiva es la región de atenuación máxima si el paciente está centrado. Si el paciente es mis-centered, hay que menos radiacion sobre la estructura a evaluar y, por tanto, aumentará el ruido de la imagen.
![Page 210: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/210.jpg)
211 / GE /
GE confidential, for internal use only
Mis-centering affects noise and radiation dose
1
20cm
Fan angle
50%
90%
0%
High noise streaks due to mis-centering
Increased dose to
breast
% Effective mA reduction contours
Centering
error
Excessive dose
Dose too low
Patient 6 cm too
low
La mitad de la energía del tubo se puede perder con un error de 4 cm
Effective
SD noise inc. mA loss
Cent 6.47 0 % 0%
4 cm 8.40 22% 50%
6 cm 9.22 30 70%
![Page 211: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/211.jpg)
GE Company confidential - For internal use only
1
20cm
Fan angle
50%
90%
0%
High noise streaks due to mis centering
Increased dose to
breast
% Effective mA reduction contours
Centering
error
Excessive dose
Dose too low
Patient 6 cm too
low
Average elevation error 2.5 cm low
Efecto en la dosis en un paciente mal centrado
Centering Error Distribution
0
10
20
30
40
50
60
70
-40
-30
-20
-10 0 10 20 30 40 50 60
- Erro r ( mm)
Elevation
Lateral
Dosis de Radiación
![Page 212: Ct 64 Dinamica de Medellin](https://reader033.vdocuments.site/reader033/viewer/2022050816/553182d64a7959a2318b4996/html5/thumbnails/212.jpg)
213 / GE /
GE confidential, for internal use only
Mean scout projection
Within abdomen
Mean centering adj: 27.5 mm
(calculated from centroid of mean scout
projection)
Auto-Patient Centering
Center of detector
Center of patient