compensation, negative controls and the optimized choice

35
1 Compensation, Negative Controls and the Optimized Choice of Reagents for Multicolor Flow Cytometry

Upload: others

Post on 26-May-2022

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Compensation, Negative Controls and the Optimized Choice

1

Compensation, Negative Controls and the Optimized Choice of Reagents for Multicolor

Flow Cytometry

Page 2: Compensation, Negative Controls and the Optimized Choice

2

Overview

• CompensationIntroductionPrevention of compensation related artifacts

• Negative ControlsFluorescence Minus One (FMO)Isotype controls Transfection controls

• Combination of Reagents for Multicolor Flow Experiments

• Characteristics of Fluorochromes

Page 3: Compensation, Negative Controls and the Optimized Choice

3

Overview

• CompensationIntroductionPrevention of compensation related artifacts

• Negative ControlsFluorescence Minus One (FMO)Isotype controls Transfection controls

• Combination of Reagents for Multicolor Flow Experiments

• Characteristics of Fluorochromes

Page 4: Compensation, Negative Controls and the Optimized Choice

4

Compensation:Introduction

• Photons reaching a specific detector are coming from the fluorochrome specific for this detector

optical background (especially cell-type specific auto-fluorescence)

photon spill over from all fluorochromes present in experiment

650 700

PerCP-Cy5.5695/40

500 600

FITC530/30

Rel

ativ

e In

tens

ity

550

PE585/42

650 700

PerCP-Cy5.5695/40

500 600

FITC530/30

Rel

ativ

e In

tens

ity

Wavelength (nm)550

PE585/42

750n 800

Page 5: Compensation, Negative Controls and the Optimized Choice

5

Compensation:Introduction

• Compensation is a procedure to subtracts from all photons that reach a detector the

1) nonspecific electronic signal from cellular autofluorescenceApplication of “Instrument Settings” for one specific (type of) sample

2) nonspecific electronic signal from fluorochrome spill over

Page 6: Compensation, Negative Controls and the Optimized Choice

6

Compensation:Introduction

• Compensation is a procedure to subtracts from all photons that reach a detector the

1) nonspecific electronic signal from cellular autofluorescenceApplication of “Instrument Settings” for one specific (type of) sample

2) nonspecific electronic signal from fluorochrome spill over

Page 7: Compensation, Negative Controls and the Optimized Choice

7

Compensation:Introduction

• 1) Generate sample-specific “Application Settings”according to SDEN given in CS&T Baseline ReportRefer to Handout #3: Generation of Application SettingsRefer to the Presentation: QC for Digital Instruments: BD CS&TTM

• 2) Perform an Automated Compensation using the BD FACS DivaTM software Refer to Handout #4: Automated Compensation

Page 8: Compensation, Negative Controls and the Optimized Choice

8

Compensation:Prevention of compensation related artifacts

• To avoid errors in compensation 1) never change the PMT V of fluorescent parameters (Instrument

Settings) after a compensation has been calculated

2) only the “median-related” = “statistic-based” compensation will result in accurate compensation values: Never compensate “by eye”

3) use only compensation controls that show in minimum equally high signal intensities than the highest signal you expect in your sample

4) BUT, the signal intensities of the compensation controls have to lie within the maximum linearity range of the scales (refer to the CS&T baseline report)

5) it is not recommended to re-use compensations if a tandem dye is included in the panel

Page 9: Compensation, Negative Controls and the Optimized Choice

9

Compensation:Prevention of compensation related artifacts

• Digital Compensation is a software-based compensation in that linear numbers are subtracted

• But if each optimal compensation is a subtraction that results in “0”, why is a compensation value of 20% worse than 0.2%?

A/D

DSPFPGA

A/D

FPGA

A/D

FPGA

A/D

FPGA

DSP1Comp Embedded Computer

Master board

FSCSSC A B C D E F G H I J

RAM RAM RAM RAM

A/D

DSPFPGA

A/D

FPGA

A/D

FPGA

A/D

FPGA

RAM RAM RAM RAM

A/D

DSPFPGA

A/D

FPGA

A/D

FPGA

A/D

FPGA

RAM RAM RAM RAM

DSP2gating

Linear Amplifiers

39,27139,27139,27139,271Event 1Event 2Event 3

FSC SSC FITC PE00 6060 120120 55

1010 160160 6565

3030 650650 160160

Uncompensated List-Mode Data

23Time

8989 675675

39.27139.271 30.62130.621

Page 10: Compensation, Negative Controls and the Optimized Choice

10

Compensation:Prevention of compensation related artifacts

• High compensation values induce “data-spread” effects that reduce substantially the resolution between populations

Uncompensated Compensated

compensation = 21.5%

compensation = 0.2%

CD45 - FITC

CD45 - APC

PE

2.0%10.5%

4.5%12.7%

CD3 – APC-H7C

D56

–PE

Gated on CD45 – FITC

Gated on CD45 – APC

Page 11: Compensation, Negative Controls and the Optimized Choice

11

Overview

• CompensationIntroductionPrevention of compensation related artifacts

• Negative ControlsFluorescence Minus One (FMO)Isotype controls Transfection controls

• Combination of Reagents for Multicolor Flow Experiments

• Characteristics of Fluorochromes

Page 12: Compensation, Negative Controls and the Optimized Choice

12

Negative Controls:Fluorescence Minus One (FMO)

• Indicates the data-spread induced by “high”compensation values

compensation = 21.5%

FITC

PE

CD45 FITC

IgG

1PE

FMO

Gated on PBMC

CD45 FITC

CD

56 P

E

Sample

? !

Page 13: Compensation, Negative Controls and the Optimized Choice

13

Negative Controls:Fluorescence Minus One (FMO)

• Fluorochrome-combinations with “high” compensation values?

• Get experience: compare of multicolor compensation matrices.

Page 14: Compensation, Negative Controls and the Optimized Choice

14

Negative Controls:Fluorescence Minus One (FMO)

• Does every fluorochrome combination with “high”compensation values requests an FMO Control?

• Not if the fluorochromes bind to different cells!

CD19 FITC

+ CD45 AmCyanw/o CD45

+ CD19-AmCyanw/o CD19-AmCyan

CD3-FITC

Page 15: Compensation, Negative Controls and the Optimized Choice

15

Negative Controls:Isotype Control

• Indicates specificity of antigen-expression in unknown samples

• Indicates the nonspecific binding of antibodies to membranes

0.025μg antibody per 100 μl whole blood

MFI = 48 MFI = 163IgG2b - PEIgG2a - PEunstained IgG1 - PE

MFI = 102 MFI = 156

CD4 is THE T-helper cell marker

?!

Page 16: Compensation, Negative Controls and the Optimized Choice

16

Negative Controls:Isotype Control

• Indicates specific binding of antibodies to very low and/or “smeared” expressed antigens

• But (if available and known) “biological controls” are betterCD4 PerCP

CD

25-P

E ?!

IgG1 PerCP

IgG

1-PE

Isotype Control

CD4 PerCP

Display CD8+ T Cytotox

CD

25-P

E

IgG

1-PE

Page 17: Compensation, Negative Controls and the Optimized Choice

17

Negative Controls:Isotype Control

• Isotype controls HAVE toshow perfect match in isotype subtype

come from the same species

be labeled to the same fluorochrome

show equalized concentrations between specific antibody and isotype control

be specifically purified from free fluorochromes for intra-cellular staining

show comparable antibody to fluorochrome ratios

Page 18: Compensation, Negative Controls and the Optimized Choice

18

Negative Controls:Transfection Control (“Mock Control”)

• Transfection methods using Lipofectin and similar substances enhance the autofluorescence due to the binding of plasmid-DNA to cell membranes

• Nontransfected cells are for this reason never a proper control!

• In this example the transfection-efficiency is 0% as the plasmid-promotor shows a deletion that prevents gene-expression!

untransfected transfected Mock-transfected

Page 19: Compensation, Negative Controls and the Optimized Choice

19

Overview

• CompensationIntroductionPrevention of compensation related artifacts

• Negative ControlsFluorescence Minus One (FMO)Isotype controls Transfection controls

• Combination of Reagents for Multicolor Flow Experiments

• Characteristics of Fluorochromes

Page 20: Compensation, Negative Controls and the Optimized Choice

20

Combination of Reagents for Multicolor Flow Experiments

• Basic rule for the selection of antibody-fluorochrome-combinations:Match the fluorochrome brightness with the antigen-expression levels

• For further information see in Appendix 4:http://www.bdbiosciences.com/documents/Multicolor_AppNote.pdf

Page 21: Compensation, Negative Controls and the Optimized Choice

21

Combination of Reagents for Multicolor Flow Experiments

• Match fluorochrome brightness with antigen-expression levels

Fluorochrome brightness?• The “Stain Index”

• Stain indices of BD specific fluorochromes obtained on a BD FACSCantoTM II

W2W1

WD(SI)IndexStain =

D = Δ between positive and negative peak mediansW = Spread of the background peak

Bright fluorochromes

Intermediate-bright fluorochromes

Dim fluorochromes

90.0BD HorizonTM V450

AmCyanBDTM APC-H7BD HorizonTM V500

24.218.012.0

Page 22: Compensation, Negative Controls and the Optimized Choice

22

Combination of Reagents for Multicolor Flow Experiments

• Match fluorochrome brightness with antigen-levelsAntigen Expression level?

+

+

+++

+

++

+++

++

++

Compare examples of PE-staining given in TDS

CD3 - PE CD19 - PE

CD45 - PECD80 - PE

Page 23: Compensation, Negative Controls and the Optimized Choice

23

Combination of Reagents for Multicolor Flow Experiments

• Match fluorochrome brightness with antigen-levelsAntigen Expression level?

+

+

+++

+

++

+++

++

++

Compare examples of PE-staining given in TDS

CD3 - PE CD19 - PE

CD45 - PECD80 - PE

Example for one proper Antibody-Fluorochrome-Match for a BD FACS AriaTM

405nm / 688nm / 635nm

BD HorizonTM V450FITCBDTM APC-H7 PerCP-CyTM5.5APCBD HorizonTM V500 PE-CyTM7PE

Or use literature-data about # of antigen-molecules per cell

ExpressionLevel

2.000 + CD127

10.000 + CD56

200.000 +++CD45

3.000 +CD25

18.000 ++ CD19

144.000 +++CD14

100.000 +++CD4

80.000 ++CD3

Antigen-DensityAntigen

Antigen-expression High / Intermediate / Low:+++ / ++ / +

Page 24: Compensation, Negative Controls and the Optimized Choice

24

Overview

• CompensationIntroductionPrevention of compensation related artifacts

• Negative ControlsFluorescence Minus One (FMO)Isotype controls Transfection controls

• Combination of Reagents for Multicolor Flow Experiments

• Characteristics of Fluorochromes

Page 25: Compensation, Negative Controls and the Optimized Choice

25

Characteristics of Fluorochromes

• Select fluorochromes according to instrument configuration

Possible fluorochrome-excitation by

Pe-Cy™7Pe-Cy™7785nm

APC-Cy™7 or BDTM APC-H7

785nm or 783nm

AF®700723nm

Pe-Cy™5 orPerCP or PerCP-

CyTM5.5

Pe-Cy™5 orPerCP or PerCP-

CyTM5.5

667nm or678nm or 695nm

APC orAF®647

660nm or 668nm

PEPE578nm

AF®488 or FITC519nm or 520nm

AmCyan orBD HorizonTM

V500

491nm

461nm

BD HorizonTM V450 orPacific BlueTM

448nm or 452nm

Red 635nmY-G 561nmBlue 488nmViolet 405nmMax Emission

Page 26: Compensation, Negative Controls and the Optimized Choice

26

Characteristics of Fluorochromes

• Select fluorescent dyes according to instrument configuration

dsRed586nm

DRAQ5683nm

SYTOX Red658nm

Possible fluorochrome-excitation by

mPlum650nm

PI or 7-AAD617nm or 647nm

mCherry610nm

mTomato581nm

mOrange562nm

SYBR-Green or SYTOX-Green or

EYFP

522nm or 524nm or 528nm

CFSE or EGFP505 or 510nm

Dapi or SYTOX-Blue or ECFP

461nm or 470nmor 477nm

Red 635nmY-G 561nmBlue 488nmViolet 405nmMax Emission

Page 27: Compensation, Negative Controls and the Optimized Choice

27

Characteristics of Fluorochromes

• BD-supported fluorochromes

3Groups of

Fluorochromes

Small Organic Molecules: BDTM Horizon V450, Pacific BlueTM, FITC,

Alexa Fluor (AF)® 350, 488 / 647 / 700

Large ProteinsAmCyan, PerCP, PE,

APC

Tandem ConjugatesPerCP-Cy TM 5.5,

PE-CyTM5, PE-CyTM7, APC-CyTM7, BDTM APC-

H7

Page 28: Compensation, Negative Controls and the Optimized Choice

28

Large size (240.000 D) is NOT a disadvantage!

Bright reagents: IC/FCMLow backgrounds: IC/FCM1st choice for intracellular protein/cytokine detection!Good for quantification:1 PE per Ig (BDTMQuantiBrite)

3. PE

One of the dullestfluorochromes

Very photo- and pH-stable (superior for microscopy)

2. AlexaFluor® 488

One of the dullest fluorochromesPhotobleaching (relevant only for microscopy)pH sensitive

Most widely usedStable, long-lasting conjugatesEasy conjugation (3-5 per Ig)Cheapest fluorochrome

1. FITC

DisadvantagesAdvantages

Characteristics of Fluorochromes: The “blue laser fluorochromes (I)”

Page 29: Compensation, Negative Controls and the Optimized Choice

29

Lot to lot differences in compensationLight sensitive (just keep samples in the dark)

Bright fluorochrome: Preferable above PE-Cy5

7. PE-CyTM7

Unspecific binding to FcReceptorsVery strong spill over to APC: has to be adjusted often by titrationLot to lot differences in compensationLight sensitive (just keep samples in the dark)

Bright fluorochromeUseful additional

fluorochrome for the LSR II

6. PE-CyTM5

DisadvantagesAdvantages

Characteristics of Fluorochromes: The “blue laser fluorochromes (III)”

Page 30: Compensation, Negative Controls and the Optimized Choice

30

Lot to lot differences in compensationLight sensitive (just keep samples in the dark)

Intermediate bright fluorochrome (PerCP is dim)

5. PerCP-CyTM5.5

Dull (equivalent to FITC)Sensitive to photo-bleaching (relevant only for microscopy or high energy blue lasers. See BD FACSTM Vantage)

Minimal Spill over to PE4. PerCP

DisadvantagesAdvantages

Characteristics of Fluorochromes: The “blue laser fluorochromes (II)”

Page 31: Compensation, Negative Controls and the Optimized Choice

31

Filter set not included in standard BD FACSAriaTM

configuration, but easy to obtain additionally

Very photostable (superior for microscopy)Additional fluorochrome for the red laser

3. AlexaFluor® 700

/Equivalent to APC, BUT Very photo- and pH-stable (superior for microscopy)

2. AlexaFluor® 647

Large size (106.000 D) is NOT a disadvantage!

Bright fluorochromeLow background: 2nd choice for intracellular protein / cytokine detection!

1. APC

DisadvantagesAdvantages

Characteristics of Fluorochromes: The “red laser fluorochromes (I)”

Page 32: Compensation, Negative Controls and the Optimized Choice

32

DimLot to lot differences in compensation

Photo-stableFixative-stable: Preferable above APC-Cy7

5. BDTM

APC-H7

DimPhoto-instableFixative-sensitiveLot to lot differences in compensation

Additional red laser excited fluorochrome

4. APC-CyTM7

DisadvantagesAdvantages

Characteristics of Fluorochromes: The “red laser fluorochromes (II)”

Page 33: Compensation, Negative Controls and the Optimized Choice

33

dimLittle Spill over to FITC1. Pacific BlueTM

/Little Spill over to FITCIntermediate bright

2. BD HorizonTM

V450

DisadvantagesAdvantages

Characteristics of Fluorochromes: The “violet laser fluorochromes (I)”

Page 34: Compensation, Negative Controls and the Optimized Choice

34

dimextremely strong spill over to FITCFixative-sensitiveLight-sensitive

Before BD HorizonTM V500 and BD HorizonTM V450 the only violet-excited fluorochrome from BD

3.AmCyan

DimLittle spill over to FITC: highly preferable alternative to AmCyanFixative-stable

4. BD Horizon V500

DisadvantagesAdvantages

Characteristics of Fluorochromes: The “violet laser fluorochromes (II)”

Page 35: Compensation, Negative Controls and the Optimized Choice

35

• Summary:There are no “bad” fluorochromes!

Considering limitations, all fluorochromes are usefull designing multicolor flow experiments

But some fluorochromes are easier to handle for multicolor-combinations than others!

Characteristics of Fluorochromes