compartment syndrome and the stryker intra-compartmental pressure monitor system

20
Compartment Syndrome and the Stryker Intra- Compartmental Pressure Monitor System

Upload: ralf-warren

Post on 16-Dec-2015

223 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Compartment Syndrome and the Stryker Intra-Compartmental Pressure Monitor System

Compartment Syndrome and the Stryker Intra-Compartmental

Pressure Monitor System

Page 2: Compartment Syndrome and the Stryker Intra-Compartmental Pressure Monitor System

Compartment Syndrome

When pressure is elevated within a confined space, capillary blood flow is compromised. The resulting edema within the soft tissue from ischemia results in further swelling and increased pressure.

Page 3: Compartment Syndrome and the Stryker Intra-Compartmental Pressure Monitor System

Terms• Acute compartment syndrome: An elevation of intercompartmental

pressure to a level and for a duration that without decompression will cause tissue necrosis.

• Exertional compartment syndrome: Elevation of intercompartmental pressure during exercise causing ischemia, pain, and possibly neurologic symptoms and signs. There is resolution with rest, but it may progress to acute compartment syndrome.

• Volkmann ischemic contracture: Irreversible muscle necrosis leading to ischemic contractures.

• Crush syndrome: The systemic result of muscle necrosis commonly caused by prolonged external compression of an extremity. Muscle necrosis is established by the time of presentation, but intracompartmental pressure may rise as a result of intracompartmental edema, causing a superimposed acute compartment syndrome.

Page 4: Compartment Syndrome and the Stryker Intra-Compartmental Pressure Monitor System

History• 1850: First report attributed to Hamilton by Hildebrand

• 1881: Richard von Volkmann published a summary of his findings: paralysis and contractures occurred after tight bandaging and were caused by prolonged blocking of arterial blood. He recognized that muscle cannot survive for longer than six hours with complete occlusion of blood and not for longer than 12 hours with partial occlusion.

• 1888: Peterson recognized that ischemic contracture can occur in the absence of bandaging.

• Early twentieth century: The first description of fasciotomy and the importance of its early application were suggested.

Page 5: Compartment Syndrome and the Stryker Intra-Compartmental Pressure Monitor System

• World War I: The belief propagated that compartment syndrome was result of arterial injury and spasm. The excision of the “damaged” artery yielded successful results. Of course, the fascia was released during the exposure.

• So….can you have compartment syndrome and normal peripheral pulses?

History

Page 6: Compartment Syndrome and the Stryker Intra-Compartmental Pressure Monitor System

• Seddon17 challenged the arterial injury theory as the sole cause of compartment syndrome, noting normal pulses. In 1966, early and gross swelling in the compartments and pressure that was released with fasciotomy were noted.

• Nolan and McQuillan18 described a vicious circle of increasing tension in an enclosed compartment causing venous outflow obstruction and subsequent reduction in arterial inflow. They concluded that delay in fasciotomy was the single cause of failure of treatment.

Page 7: Compartment Syndrome and the Stryker Intra-Compartmental Pressure Monitor System

EpidemiologyUnderlying conditions associated with injury causing acute compartment syndrome presenting to an orthopaedic trauma unit (percentage of cases):

• Tibial diaphyseal fracture (36.0%)• Soft-tissue injury (23.2%)• Distal radial fracture (9.8%)• Crush syndrome (7.9%)• Diaphyseal fracture forearm (7.9%)• Femoral diaphyseal fracture (3.0%)• Tibial plateau fracture (3.0%)• Hand fracture(s) (2.5%)• Tibial pilon fractures (2.5%)• Foot fracture(s) (1.8%)• Ankle fracture (0.6%)• Elbow fracture dislocation (0.6%)• Pelvic fracture (0.6%)• Humeral diaphyseal fracture (0.6%)

Page 8: Compartment Syndrome and the Stryker Intra-Compartmental Pressure Monitor System

Compartment Syndrome Etiology

Compartment Size• Tight dressing (bandage or cast)• Localized external pressure, lying on limb• Closure of fascial defects

Compartment Content• Bleeding, fixation, vascular injury, bleeding

disorders• Capillary permeability: ischemia, trauma, burns,

exercise, snake bite, drug Injection, in vitro fertilization

Page 9: Compartment Syndrome and the Stryker Intra-Compartmental Pressure Monitor System

Compartment SyndromeEtiology

• Fractures (closed and open)

• Blunt trauma• Temporary vascular

occlusion• Cast or dressing• Closure of fascial defects• Electrical burns

• Exertional states• Gunshot wound• Intravenous A lines• Hemophilia and

coagulation• Intraosseous infusion

(infant)

• Snake bite

Page 10: Compartment Syndrome and the Stryker Intra-Compartmental Pressure Monitor System

Measurement of pressures with the Stryker Intra-Compartmental Pressure

Monitor• Turn on.• Assemble the needle, transducer, and syringe.• Seat into chamber and close lid.• Tilt 45° and purge chamber and needle of air.• Prep skin (not on pig).• Just before the needle enters the skin, zero the Stryker Intra-

Compartmental Pressure System, and don’t change the angle after this.

• Inject <1/3 mL of fluid provided by the manufacturer in the syringe to clear the side port.

• Allow time for the reading to stabilize; it may take 15 to 20 seconds.

Page 11: Compartment Syndrome and the Stryker Intra-Compartmental Pressure Monitor System

Compartment SyndromePathophysiology

• Normal tissue pressure– 0 to 4 mm Hg – 8 to 10 mm Hg with exertion

• Absolute pressure theory– 30 mm Hg (Mubarak et al.20)

• Pressure gradient theory– <20 mm Hg of diastolic pressure (Whitesides et

al.8 and McQueen and Court-Brown21)

Page 12: Compartment Syndrome and the Stryker Intra-Compartmental Pressure Monitor System

Compartment Syndrome: A Clinical Diagnosis

• Pain out of proportion• Palpably tense compartment• Pain with passive stretch• Paresthesia or hypoesthesia• Paralysis• Pulselessness or pallor

Page 13: Compartment Syndrome and the Stryker Intra-Compartmental Pressure Monitor System

Beware

• Epidurals• Regional blocks• Unconscious• Insensate

Page 14: Compartment Syndrome and the Stryker Intra-Compartmental Pressure Monitor System

Medical Management• Make sure that the patient is normotensive.

Hypotension reduces perfusion pressure and facilitates further tissue injury.

• Remove the circumferential bandages and cast.

• Maintain the limb at the level of the heart as elevation reduces the arterial inflow and the arteriovenous pressure gradient on which perfusion depends.

• Administer supplemental oxygen.

Page 15: Compartment Syndrome and the Stryker Intra-Compartmental Pressure Monitor System

Compartment SyndromePressure Measurements

• Suspected compartment syndrome

• Equivocal or unreliable examination

• Clinical adjunct ONLY after discussing it with the attending physician

Page 16: Compartment Syndrome and the Stryker Intra-Compartmental Pressure Monitor System

When NOT to check pressures

• Compartment syndrome is diagnosed on a clinical basis.

Page 17: Compartment Syndrome and the Stryker Intra-Compartmental Pressure Monitor System

Threshold for Decompression

• 30 mm Hg: close to capillary blood pressure20,25

• 40 mm Hg13,26

• 50 mm Hg in tibia fixation and normotensive27

• <10 to 30 mm Hg: difference between diastolic and tissue pressure (delta P)8

• Difference between mean arterial pressure and tissue pressure of <30 mm Hg in normal muscle or <40 mm Hg in traumatized muscle22-

24,28

Page 18: Compartment Syndrome and the Stryker Intra-Compartmental Pressure Monitor System

Fasciotomy Principles

• Make an early diagnosis.• Make long extensile incisions.• Release all fascial compartments.• Preserve neurovascular structures.• Debride necrotic tissues.• Provide coverage within seven to ten days.

Page 19: Compartment Syndrome and the Stryker Intra-Compartmental Pressure Monitor System

Compartment SyndromeLower Leg

• Four compartments

– Lateral: peroneus longus and peroneus brevis (PB)

– Anterior: extensor hallucis longus, extensor digitorum longus (EDL), tibialis anterior (TA), peroneus tertius

– Superficial posterior: gastrocnemius (G), soleus (S)

– Deep posterior: tibialis posterior (TP), flexor hallucis (FH) longus, flexor digitorum longus

Page 20: Compartment Syndrome and the Stryker Intra-Compartmental Pressure Monitor System

Compartment SyndromeForearm

• Three anatomical compartments– Mobile wad: brachioradialis,

extensor carpi radialis longus, and extensor carpi radialis brevis

– Volar: superficial and deep flexors

– Dorsal: extensorsThe pronator quadratus is

described as a separate compartment.