Comparison of Retinal ImageQuality betweenSBK and PRK
Allen Boghossian, D.O.Durrie Vision
Overland Park, KS
![Page 2: Comparison of Retinal Image Quality between SBK and PRKvoi.opt.uh.edu/.../2007/presentations/52RetinalQuality_Boghossian.pdf · Comparison of Retinal Image Quality between SBK and](https://reader031.vdocuments.site/reader031/viewer/2022031213/5bd5ebf509d3f2513e8cae41/html5/thumbnails/2.jpg)
Study Design
• Purpose: To compare optical quality and intraocular scatterbetween SBK and PRK using double-pass (DP) technology
• Analysis performed using the Optical Quality AnalysisSystem (OQAS) (Visiometrics SL, Tarrasa, Spain)
• Author has no commercial affiliation with Visiometrics
![Page 3: Comparison of Retinal Image Quality between SBK and PRKvoi.opt.uh.edu/.../2007/presentations/52RetinalQuality_Boghossian.pdf · Comparison of Retinal Image Quality between SBK and](https://reader031.vdocuments.site/reader031/viewer/2022031213/5bd5ebf509d3f2513e8cae41/html5/thumbnails/3.jpg)
• 25 patients (50 eyes)– One eye treated with SBK (25 eyes)– Fellow eye treated with PRK (25 eyes)– Subjects randomized according to dominant eye– All SBK flaps made using the 60 Hz IntraLase FS Laser– All eyes treated with wavefront guided Alcon LADAR6000– Average Age: 31 ± 5 years– Mean MRSE:
• SBK eyes: -4.18 ± 1.03 D• PRK eyes: -4.30 ± 1.04 D
– Measurements at pre-op, 1, 3, and 6 months
Study Design
![Page 4: Comparison of Retinal Image Quality between SBK and PRKvoi.opt.uh.edu/.../2007/presentations/52RetinalQuality_Boghossian.pdf · Comparison of Retinal Image Quality between SBK and](https://reader031.vdocuments.site/reader031/viewer/2022031213/5bd5ebf509d3f2513e8cae41/html5/thumbnails/4.jpg)
IntroductionWhat is SBK?
Sub-Bowman's Keratomileusis (SBK)
• Corneal flap just below the level of Bowman’s membranecreated by Intralase FS laser
• Optimal flap: 8.5 mm diameter, 100 microns or less,centered on visual axis
• Has advantage of both Lasik and surface ablation• Decreased post-op dryness, decreased loss of sensitivity,
and decreased higher order aberrations
![Page 5: Comparison of Retinal Image Quality between SBK and PRKvoi.opt.uh.edu/.../2007/presentations/52RetinalQuality_Boghossian.pdf · Comparison of Retinal Image Quality between SBK and](https://reader031.vdocuments.site/reader031/viewer/2022031213/5bd5ebf509d3f2513e8cae41/html5/thumbnails/5.jpg)
Part of larger SBK vs. PRK comparative study:1. Visual acuity, refractive Outcome2. Corneal sensitivity3. Dry eye signs/symptoms4. Endothelial cell count5. Keratocyte cell density – ARVO 20076. Corneal biomechanical properties – ASCRS 20077. Intraocular pressure comparison:
a. Goldmann Applanation Tonometer (GAT)b. Ocular Response Analyzer (ORA)c. Pascal Dynamic Tonometer (PDT)
8. Wavefront analysis9. Patient questionnaire10. OQAS analysis
Introduction
![Page 6: Comparison of Retinal Image Quality between SBK and PRKvoi.opt.uh.edu/.../2007/presentations/52RetinalQuality_Boghossian.pdf · Comparison of Retinal Image Quality between SBK and](https://reader031.vdocuments.site/reader031/viewer/2022031213/5bd5ebf509d3f2513e8cae41/html5/thumbnails/6.jpg)
IntroductionOQAS
Optical Quality Analysis System (OQAS)• Objective measurement of visual quality• Uses double-pass (DP) technology to measure light
reflection from retina including light scatter (LS) and higherorder aberrations (HOA)
![Page 7: Comparison of Retinal Image Quality between SBK and PRKvoi.opt.uh.edu/.../2007/presentations/52RetinalQuality_Boghossian.pdf · Comparison of Retinal Image Quality between SBK and](https://reader031.vdocuments.site/reader031/viewer/2022031213/5bd5ebf509d3f2513e8cae41/html5/thumbnails/7.jpg)
IntroductionHow does OQAS work?
1. A motorized optometer lenses correctsfor patient’s defocus
2. First pass, the system sends a diodelaser light (780 nm) into the eye
3. Produces an image of a circular spoton the retina
4. Second pass, the reflection of the spotis measure by a light camera
5. Directly measures the point spreadfunction (PSF) of the optical system
6. Using Fourier analysis, the PSF data istranslated to a modulation transferfunction (MTF) value
7. Measurement at 4mm undilated pupils
Double-Pass Technology
![Page 8: Comparison of Retinal Image Quality between SBK and PRKvoi.opt.uh.edu/.../2007/presentations/52RetinalQuality_Boghossian.pdf · Comparison of Retinal Image Quality between SBK and](https://reader031.vdocuments.site/reader031/viewer/2022031213/5bd5ebf509d3f2513e8cae41/html5/thumbnails/8.jpg)
• The main advantage of this instrument compared withaberrometers is that it directly obtains the actual retinalimages including lower, higher order aberrations andintraocular scattering.
• Current aberrometers mathematically calculate MTF values• OQAS measures MTF directly
IntroductionOQAS
![Page 9: Comparison of Retinal Image Quality between SBK and PRKvoi.opt.uh.edu/.../2007/presentations/52RetinalQuality_Boghossian.pdf · Comparison of Retinal Image Quality between SBK and](https://reader031.vdocuments.site/reader031/viewer/2022031213/5bd5ebf509d3f2513e8cae41/html5/thumbnails/9.jpg)
• MTF is the square root of the modulus of the Fouriertransformation of the average double-pass image
• Correlates to the loss ofcontrast in the retinal imagefor each spatial frequency
• Average value is 30.0• The higher the MTF value the
higher the quality of image
IntroductionMTF Value
![Page 10: Comparison of Retinal Image Quality between SBK and PRKvoi.opt.uh.edu/.../2007/presentations/52RetinalQuality_Boghossian.pdf · Comparison of Retinal Image Quality between SBK and](https://reader031.vdocuments.site/reader031/viewer/2022031213/5bd5ebf509d3f2513e8cae41/html5/thumbnails/10.jpg)
• OQAS numericalmeasurement of contrastsensitivity
• Average value is 1.0• Higher the contrast value the
higher the quality of image
IntroductionContrast Value
![Page 11: Comparison of Retinal Image Quality between SBK and PRKvoi.opt.uh.edu/.../2007/presentations/52RetinalQuality_Boghossian.pdf · Comparison of Retinal Image Quality between SBK and](https://reader031.vdocuments.site/reader031/viewer/2022031213/5bd5ebf509d3f2513e8cae41/html5/thumbnails/11.jpg)
ResultsUncorrected Vision – 1 month
0%10%20%30%40%50%60%70%80%90%
100%
20/16 orbetter
20/20 orbetter
20/25 orbetter
20/40 orbetter
SBKPRK
N = 25
![Page 12: Comparison of Retinal Image Quality between SBK and PRKvoi.opt.uh.edu/.../2007/presentations/52RetinalQuality_Boghossian.pdf · Comparison of Retinal Image Quality between SBK and](https://reader031.vdocuments.site/reader031/viewer/2022031213/5bd5ebf509d3f2513e8cae41/html5/thumbnails/12.jpg)
64%
28%
80%
44%
88%76%
100% 92%
0%10%20%30%40%50%60%70%80%90%
100%
20/16 orbetter
20/20 orbetter
20/25 orbetter
20/40 orbetter
SBKPRK
N = 25
ResultsUncorrected Vision – 1 month
![Page 13: Comparison of Retinal Image Quality between SBK and PRKvoi.opt.uh.edu/.../2007/presentations/52RetinalQuality_Boghossian.pdf · Comparison of Retinal Image Quality between SBK and](https://reader031.vdocuments.site/reader031/viewer/2022031213/5bd5ebf509d3f2513e8cae41/html5/thumbnails/13.jpg)
67%75%
92% 88%100% 100%
0%10%20%30%40%50%60%70%80%90%
100%
20/16 orbetter
20/20 orbetter
20/25 orbetter
20/40 orbetter
SBKPRK
N = 25
ResultsUncorrected Vision – 3 months
![Page 14: Comparison of Retinal Image Quality between SBK and PRKvoi.opt.uh.edu/.../2007/presentations/52RetinalQuality_Boghossian.pdf · Comparison of Retinal Image Quality between SBK and](https://reader031.vdocuments.site/reader031/viewer/2022031213/5bd5ebf509d3f2513e8cae41/html5/thumbnails/14.jpg)
ResultsMRSE and Cylinder
-0.38
-0.27
-0.08
-0.17
1 month
-0.40
-0.18
-0.15
-0.12
3 months
-0.61PRK
-0.70SBK
-4.30PRK
-4.18SBK
Pre-OP
MR
SEC
ylin
der
![Page 15: Comparison of Retinal Image Quality between SBK and PRKvoi.opt.uh.edu/.../2007/presentations/52RetinalQuality_Boghossian.pdf · Comparison of Retinal Image Quality between SBK and](https://reader031.vdocuments.site/reader031/viewer/2022031213/5bd5ebf509d3f2513e8cae41/html5/thumbnails/15.jpg)
Which Eye Has Better Vision?Through 3 Months
0%10%20%30%40%50%60%70%80%90%
100%
1 Month 3 Months
SBK Better VisionPRK Better VisionBoth Same
N=25
![Page 16: Comparison of Retinal Image Quality between SBK and PRKvoi.opt.uh.edu/.../2007/presentations/52RetinalQuality_Boghossian.pdf · Comparison of Retinal Image Quality between SBK and](https://reader031.vdocuments.site/reader031/viewer/2022031213/5bd5ebf509d3f2513e8cae41/html5/thumbnails/16.jpg)
OQASAverage MTF
0.00
5.00
10.00
15.00
20.00
25.00
30.00
35.00
40.00
Pre-Op 1 Month 3 Month 6 Month
SBKPRK
N=25
MTF
![Page 17: Comparison of Retinal Image Quality between SBK and PRKvoi.opt.uh.edu/.../2007/presentations/52RetinalQuality_Boghossian.pdf · Comparison of Retinal Image Quality between SBK and](https://reader031.vdocuments.site/reader031/viewer/2022031213/5bd5ebf509d3f2513e8cae41/html5/thumbnails/17.jpg)
OQASAverage MTF
30.09 29.39
0.00
5.00
10.00
15.00
20.00
25.00
30.00
35.00
40.00
Pre-Op 1 Month 3 Month 6 Month
SBKPRK
N=25
MTF
![Page 18: Comparison of Retinal Image Quality between SBK and PRKvoi.opt.uh.edu/.../2007/presentations/52RetinalQuality_Boghossian.pdf · Comparison of Retinal Image Quality between SBK and](https://reader031.vdocuments.site/reader031/viewer/2022031213/5bd5ebf509d3f2513e8cae41/html5/thumbnails/18.jpg)
OQASAverage MTF
30.09 29.3931.85 30.88 29.93
0.00
5.00
10.00
15.00
20.00
25.00
30.00
35.00
40.00
Pre-Op 1 Month 3 Month 6 Month
SBKPRK
N=25
MTF
![Page 19: Comparison of Retinal Image Quality between SBK and PRKvoi.opt.uh.edu/.../2007/presentations/52RetinalQuality_Boghossian.pdf · Comparison of Retinal Image Quality between SBK and](https://reader031.vdocuments.site/reader031/viewer/2022031213/5bd5ebf509d3f2513e8cae41/html5/thumbnails/19.jpg)
OQASAverage MTF
30.09 29.3931.85
25.61
30.88 29.49 29.9327.81
0.00
5.00
10.00
15.00
20.00
25.00
30.00
35.00
40.00
Pre-Op 1 Month 3 Month 6 Month
SBKPRK
N=25
MTF
![Page 20: Comparison of Retinal Image Quality between SBK and PRKvoi.opt.uh.edu/.../2007/presentations/52RetinalQuality_Boghossian.pdf · Comparison of Retinal Image Quality between SBK and](https://reader031.vdocuments.site/reader031/viewer/2022031213/5bd5ebf509d3f2513e8cae41/html5/thumbnails/20.jpg)
OQASContrast Value
0.00
0.20
0.40
0.60
0.80
1.00
1.20
1.40
Pre-Op 1 Month 3 Month 6 Month
SBKPRK
N=25
Con
trast
Val
ue
![Page 21: Comparison of Retinal Image Quality between SBK and PRKvoi.opt.uh.edu/.../2007/presentations/52RetinalQuality_Boghossian.pdf · Comparison of Retinal Image Quality between SBK and](https://reader031.vdocuments.site/reader031/viewer/2022031213/5bd5ebf509d3f2513e8cae41/html5/thumbnails/21.jpg)
OQASContrast Value
1.04 1.03
0.00
0.20
0.40
0.60
0.80
1.00
1.20
1.40
Pre-Op 1 Month 3 Month 6 Month
SBKPRK
N=25
Con
trast
Val
ue
![Page 22: Comparison of Retinal Image Quality between SBK and PRKvoi.opt.uh.edu/.../2007/presentations/52RetinalQuality_Boghossian.pdf · Comparison of Retinal Image Quality between SBK and](https://reader031.vdocuments.site/reader031/viewer/2022031213/5bd5ebf509d3f2513e8cae41/html5/thumbnails/22.jpg)
OQASContrast Value
1.04 1.031.13 1.06 1.04
0.00
0.20
0.40
0.60
0.80
1.00
1.20
1.40
Pre-Op 1 Month 3 Month 6 Month
SBKPRK
N=25
Con
trast
Val
ue
![Page 23: Comparison of Retinal Image Quality between SBK and PRKvoi.opt.uh.edu/.../2007/presentations/52RetinalQuality_Boghossian.pdf · Comparison of Retinal Image Quality between SBK and](https://reader031.vdocuments.site/reader031/viewer/2022031213/5bd5ebf509d3f2513e8cae41/html5/thumbnails/23.jpg)
OQASContrast Value
1.04 1.031.13
0.921.06 1.03 1.04
0.96
0.00
0.20
0.40
0.60
0.80
1.00
1.20
1.40
Pre-Op 1 Month 3 Month 6 Month
SBKPRK
N=25
Con
trast
Val
ue
![Page 24: Comparison of Retinal Image Quality between SBK and PRKvoi.opt.uh.edu/.../2007/presentations/52RetinalQuality_Boghossian.pdf · Comparison of Retinal Image Quality between SBK and](https://reader031.vdocuments.site/reader031/viewer/2022031213/5bd5ebf509d3f2513e8cae41/html5/thumbnails/24.jpg)
-1%+3%+6%--SBK
-5% 0%-13%--PRK
6 Month3 Month1 MonthPre-Op
Change in MTF Value
OQASResults
-7%-1%-13%--PRK
-1%+2%+8%--SBK6 Month3 Month1 MonthPre-Op
Change in Contrast Value
![Page 25: Comparison of Retinal Image Quality between SBK and PRKvoi.opt.uh.edu/.../2007/presentations/52RetinalQuality_Boghossian.pdf · Comparison of Retinal Image Quality between SBK and](https://reader031.vdocuments.site/reader031/viewer/2022031213/5bd5ebf509d3f2513e8cae41/html5/thumbnails/25.jpg)
MTF
MTF
6 month3 month1 monthPre-Op
SBK
PRK
OQAS Sample Point Spread
38.79
39.43
24.51
42.11
35.89
41.67 42.56
36.24
![Page 26: Comparison of Retinal Image Quality between SBK and PRKvoi.opt.uh.edu/.../2007/presentations/52RetinalQuality_Boghossian.pdf · Comparison of Retinal Image Quality between SBK and](https://reader031.vdocuments.site/reader031/viewer/2022031213/5bd5ebf509d3f2513e8cae41/html5/thumbnails/26.jpg)
Retinal Image QualityClinical Application of OQAS
![Page 27: Comparison of Retinal Image Quality between SBK and PRKvoi.opt.uh.edu/.../2007/presentations/52RetinalQuality_Boghossian.pdf · Comparison of Retinal Image Quality between SBK and](https://reader031.vdocuments.site/reader031/viewer/2022031213/5bd5ebf509d3f2513e8cae41/html5/thumbnails/27.jpg)
Retinal Image QualityNormal Eye
• 33 year old male• BSCVA 20/15
• MTF = 43.58• Contrast Value = 1.48
![Page 28: Comparison of Retinal Image Quality between SBK and PRKvoi.opt.uh.edu/.../2007/presentations/52RetinalQuality_Boghossian.pdf · Comparison of Retinal Image Quality between SBK and](https://reader031.vdocuments.site/reader031/viewer/2022031213/5bd5ebf509d3f2513e8cae41/html5/thumbnails/28.jpg)
Retinal Image QualityCortical Cataract
• 62 year old male• BSCVA 20/20-2
• MTF = 14.50• Contrast Value = 0.48
![Page 29: Comparison of Retinal Image Quality between SBK and PRKvoi.opt.uh.edu/.../2007/presentations/52RetinalQuality_Boghossian.pdf · Comparison of Retinal Image Quality between SBK and](https://reader031.vdocuments.site/reader031/viewer/2022031213/5bd5ebf509d3f2513e8cae41/html5/thumbnails/29.jpg)
Retinal Image QualityNuclear Sclerosis
• 66 year old male• BSCVA 20/25
• MTF = 8.92• Contrast Value = 0.3
![Page 30: Comparison of Retinal Image Quality between SBK and PRKvoi.opt.uh.edu/.../2007/presentations/52RetinalQuality_Boghossian.pdf · Comparison of Retinal Image Quality between SBK and](https://reader031.vdocuments.site/reader031/viewer/2022031213/5bd5ebf509d3f2513e8cae41/html5/thumbnails/30.jpg)
Conclusion
• Double pass technique can be used to assess theoverall image quality not possible with snellenvisual acuity charts
• At 1 month, the retinal image quality decreasedsignificantly in PRK eyes compared to SBK eyes
• Retinal image quality was higher in all post-opvisits for SBK eyes compared to PRK eyes
• There was a slight decrease in retinal imagequality from 3 month to 6 month in all eyes
• Further study needed on importance of scatter onoverall quality of vision