Combatting Corrosion in Refineries:Using Ion Exchange Chromatography to Measure Free Cyanide in Sour Water
Jay Sheffer, Jay Gandhi
Metrohm USA
Tuesday, October 16, 2018
Gulf Coast Conference
Abstract # 186
2
•An interesting view of a
refinery’s purpose…
“Process or generate as much
product as possible from a
refinery before it all falls down.”
- unknown
Much Ado About Corrosion
3
•Corrosion is of great concern to
refineries and industry
Costs ~ $4 billion annually
•Halides, sulfur, cyanide in
feedstocks and products break
down to form acids
Corrosive to refinery, equipment, and
engines
Corrosive species include: fluoride,
chloride, sulfur, cyanide
Concerns for Corrosion
4
Crude Refinement
5
Crude Refinement
• Desalting• Desulfurization
(Sour Water Stripping)
Why?
Prevent corrosion
Sour Water Stripping
Refinery PAT
• Steam Stripping Process
Sour Water Stripping
Refinery PAT
• Parameters of Interest• Sulfide
• Ammonia
• Phenol
• Cyanide
• Parameters of Interest• Sulfide
• Ammonia
• Phenol
• Cyanide• pH
• Conductivity
• Alkalinity & Hardness
HCN/CN- is very corrosive
https://scindeks-clanci.ceon.rs/data/pdf/0042-8469/2016/0042-84691604048P.pdf
• 930 Compact IC Flex - Cation
• A SUPP 10 250mm x 4mm column
• A SUPP 10 Guard/4.0 guard
• Anion-exchange column
• Eluent: 100mM NaOH + 0.01mM Na2EDTA
• Flow: 0.8mL/min
• Column temperature: 45°C
• Sample loop: 20mL
• 858 ProfIC Sample Processor
Cyanide by Ion Exchange Chromatography
2.930.2160
Compact IC Flex Oven/Deg2.858.0020
ProfIC Sample Processor
•Amperometric detection• Ag working electrode
• Pd solid-state reference electrode
• Mode: direct current (DC), 0.0 volts
• Detector temperature: 40°C
Cyanide by Ion Exclusion Chromatography
2.850.9110
Amperometric Detector
11
Calibration standards
• Standards were prepared in ppb
range to accommodate sample
dilution (1:1000)
Standard CN- (mg/L)
# 1 2.5 ppb
# 2 5.0 ppb
# 3 10.0 ppb
# 4 32.5 ppb
# 5 50 ppb
Correlation 0.99986
RSD (%) 1.804 %
Cyanide (CN-)
Calibration Plot
AW IC US6-0257-022018
Calibration Standards
A SUPP 10 250mm x 4mm column
A SUPP 10 Guard/4.0 guard
Eluent: 100mM NaOH + 0.01mM Na2EDTA
Flow: 0.8mL/min
Column temperature: 45°C
Sample loop: 20mL
Overlay of Calibration Standards
2.5ppb to 50ppb
Cyanide
AW IC US6-0257-022018
No Interference with Sulfide Matrix
• Some methods encounter sulfide interference
• Flow-injection analyzers (FIA)
• Other techniques require sulfide removal using toxic salts
1) Precipitation with Pd, Cd, or other heavy metals
• Produces additional toxic waste streams
2) Centrifugation
3) Filtration
Sulfide is easily separated with ion chromatography
No sample prep required
AW IC US6-0257-022018
No Interference with Sulfide Matrix
Sample:
• 3% sulfide matrix
• Cyanide = 12ppm
Sample + spike:
• 3% sulfide matrix
• Cyanide = 12ppm
+ 10ppm CN- spike
AW IC US6-0257-022018
Cyanide, ppm
(Sample)
Cyanide, ppm
(Sample + spike)
Recovery
%
Replicate 1 0 1.73 86.5
Replicate 2 0 1.78 89.0
Replicate 3 0 1.79 89.5
Replicate 4 0 1.82 91.0
Replicate 5 0 1.77 88.5
Replicate 6 0 1.85 92.5
Average - 1.79 89.5
Std Dev. - 0.0415 -
Results – Spike Recovery
Low-level spike
Spike = 2.0ppm cyanide
Cyanide, ppm
(Sample)
Cyanide, ppm
(Sample + spike)
Recovery
%
Replicate 1 12.1 22.3 109
Replicate 2 12.1 21.9 104
Replicate 3 12.1 22.8 114
Replicate 4 12.1 22.1 106
Replicate 5 12.1 23.1 117
Replicate 6 12.1 21.8 103
Average - 22.3 109
Std Dev. - 0.516 -
Spike = 9.4ppm cyanide
High-level spike
AW IC US6-0257-022018
Summary
• Ion chromatography provides simple and reliable
measurements of cyanide in sour water
No interference from 3% sulfide matrix
No sample preparation is required; therefore, no
additional toxic waste is generated nor requires
disposal
High-capacity anion exchange column readily
separates sulfide from cyanide
Cyanide detection down to 2ppm
Metrohm at GCC 2018
Abstract # Title Time
188 Simple, Efficient Petrochemical Analysis with NIR and Raman
Spectroscopy
10:30am – 11:00am
190 Fast Hydroxyl Number Analysis 1:30pm – 2:00pm
191 Moisture Analysis of Solids, Liquids, and Gases 2:00pm – 2:30pm
Technical Presentations: Tuesday 10/16, Wysteria
Metrohm at GCC 2018
Abstract # Title Time
187 Burning Through the Confusion: See How Combustion IC
Provides Superior Halide Analysis
10:00am – 10:30am
251 Simple, Efficient Petrochemical Analysis with NIR and Raman
Spectroscopy
10:30am – 11:00am
193 Better Surfactant Analysis and a New Approach for Cloud
Point
1:30pm – 2:00pm
194 Robotic Titration for Consolidating Petrochemical Titrations 2:00pm – 2:30pm
Technical Presentations: Wednesday 10/17, Wysteria
Metrohm at GCC 2018
Abstract # Title Time
196 OMNIS: Titration for Faster, Safer and Easier Analysis 10:00am – 12:30pm
1976 Acid Number of Crude Oils and Petroleum Products by
Catalytic Thermometric Titration using ASTM D8045
1:30pm – 4:00pm
Technical Posters: Wednesday 10/17, South Lobby
Titration Ion Chromatography Electrochemistry
Process Monitoring Spectroscopy
Visit Metrohm at Booth #817