COLPOSCOPE MODEL 935 OWNER’S MANUAL PRECISION MICROSCOPES A Division of Seiler Instrument Company PRECISION MICROSCOPES A Division of Seiler Instrument Company 3433 Tree Court Industrial Blvd. St. Louis, MO 63122 (314) 968-2282 Toll-Free: (800) 489-2282 Fax: (314) 968-3601 E-mail: [email protected] www.seilerinst.com

Upload: others

Post on 11-Jul-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: COLPOSCOPE MODEL 935 - Shopify · The Objective lens can be protected by use of the Sterilizable Protector Cap. It is fitted on the external diameter of the objective lens and protects

COLPOSCOPEMODEL 935

OWNER’S MANUAL

PRECISION MICROSCOPESA Division of Seiler Instrument Company

PRECISION MICROSCOPESA Division of Seiler Instrument Company

3433 Tree Court Industrial Blvd.St. Louis, MO 63122

(314) 968-2282Toll-Free: (800) 489-2282

Fax: (314) 968-3601E-mail: [email protected]

www.seilerinst.com

Page 2: COLPOSCOPE MODEL 935 - Shopify · The Objective lens can be protected by use of the Sterilizable Protector Cap. It is fitted on the external diameter of the objective lens and protects

OPTICAL AND FIELD MAGNIFICATION TABLE

The optical magnifications achieved with the Colposcope are determined by 3 variables: the objective focal distance, the position of the Magnification Selector, and by the eyepiece used.

The table shows the optical magnification and the observed field diameters, in millimeters, depending on those variables. At the bottom line of the table the illu-minated field is shown, corresponding to the light disk incident on the field, with a diameter that depends only on the focus of the work objective.The magnification selector has 3 positions: 1, 2, and 3. The 2nd position is repeated in the selector. The working position is the one aligned to the black dot mark.

BINOCULAR HEAD FOCAL LENGTH = 170

MGNIFICATION FIELD MGNIFICATION FIELD MGNIFICATION FIELD MGNIFICATION FIELD MGNIFICATION FIELD

1(0.6) 5.83 30.88 5.10 35.29 4.08 44.12 3.40 52.94 2.55 70.59

2(1.0) 9.71 18.53 8.50 21.18 6.80 26.47 5.67 31.76 4.25 42.35

3(1.6) 15.54 11.58 13.60 13.24 10.88 16.54 9.07 19.85 6.80 26.47

1(0.6) 7.29 27.45 6.38 31.37 5.10 39.22 4.25 47.06 3.19 62.75

2(1.0) 12.14 16.47 10.63 18.82 8.50 23.53 7.08 28.24 5.31 37.65

3(1.6) 19.43 10.29 17.00 11.76 13.60 14.71 11.33 17.65 8.50 23.53

1(0.6) 9.33 27.45 8.16 31.37 6.53 39.22 5.44 47.06 4.08 62.75

2(1.0) 15.54 16.47 13.60 18.82 10.88 23.53 9.07 28.24 6.80 37.65

3(1.6) 24.87 10.29 21.76 11.76 17.41 14.71 14.51 17.65 10.88 23.53

2(0.6) 11.66 20.59 10.20 23.53 8.16 29.41 6.80 35.29 5.10 47.06

3(1.0) 19.43 12.35 17.00 14.12 13.60 17.65 11.33 21.18 8.50 28.24

4(1.6) 31.09 7.72 27.20 8.82 21.76 11.03 18.13 13.24 13.60 17.65

illuminated field 65

EYEPIECEPOSITION OF BUTTON

OBJECTIVE LENSES

175 200

MAGNIFICATION TABLE FOR 935.

10x/18

12.5x/16

16x/16

20x/12

250 300 400

72 90 108 144

10

TECHNICAL DATA

A - Colposcope Head Microscope Galilean System Binoculars Straight, F=170mm (optional: Inclined, F=135) Objective 300mm (optional: 200-400mm) Eyepieces WF 12.5x (optional: 10x/20x) Magnification Refer to Table Interpupillary Distance Adjustable between 42mm & 75mm Illumination System Halogen 15V/150W (100-120V/220-240) (optional: Metal Halide 50W) Light Conductor Coaxial fiber optic cable with 6mm active diameter, 812mm length Filter Green Focusing Fine Focus: 11mm

B - Colposcope Stand

Angular Rotation 360 Degrees Distance between the 1016mm-1193mm floor and the objective Base Five casters, two with brakes Total Weight 18.14 kg

Page 3: COLPOSCOPE MODEL 935 - Shopify · The Objective lens can be protected by use of the Sterilizable Protector Cap. It is fitted on the external diameter of the objective lens and protects

INDICATIONS FOR USE:

The Seiler Colposcope is intended to provide magnified visualization of the tissues of the vulva, vagina, cervix, and anogenital area. It is used to evaluate these tissues, select areas for biopsy, as necessary, and to facilitate related procedures, e.g., LEEP, conization, etc.

Caution: Federal law restricts this device to sale by or on the order of a physician or practitioner trained in its use.

Presentation and Use

The sections contained in this Manual were especially grouped to serve the owners of the COLPOSCOPE MODEL 935 from Seiler Precision Microscopes.Read this Manual before setting up the equipment, in order to avoid damage from improper usage.

For safe use of the colposcope, it is necessary to be familiar with the details ofits operation prior to use.

Therefore, it is advisable that personnel operating the colposcope read this manual prior to using the equipment.

INDEX

Colposcope Floor Stand 3The Colposcope Head 4How to clean and focus 5-6Illumination System and Lamp Replacement 7,8,9Optical and Field Magnification Table 10Contact Us 11

2

6) Slide bulb back into bulb tray - Metal Halide bulb tray has a guide for proper seating of the bulb - If bulb is not seated correctly in the bulb tray light output will be affected. 7) Replace front panel and lock down thumb locks 8) Plug in the power cord and/or turn on the power button.

Metal Halide Light Source

Halogen Light Source

Light Source Change Switch

Rheostat (brightness control)

Power Switch

DC Power Source

9

Page 4: COLPOSCOPE MODEL 935 - Shopify · The Objective lens can be protected by use of the Sterilizable Protector Cap. It is fitted on the external diameter of the objective lens and protects

COLPOSCOPE FLOOR STAND

Fig. 10

NOMENCLATURE OF THE NUMBERED PARTS

1) Colposcope Head2) Head block knob (Caution: Optics will drop when knob is loosened.)3) Column height adjusting knob4) Cold Light Generator (1) Metal Halide (SMR50) / (1) Halogen 15V-150W (64634)5) Base with casters

A knob located at the stand tube (item 3) fixes the vertical movement of the column.

By turning the knob (item 3) counterclockwise or clockwise, the system can be released or fixed, and the height of the column be adjusted to achieve the correct height for the patient examination.

1

2

5

4

3

3

4) Grab the bulb by the ceramic disc and slide out of the bulb holder tray. BE CAREFUL to make sure that the light bulb is not HOT. 5a) For changing the METAL HALIDE bulb:

- Remove the bulb from the tray and separate the bulb with the wire from the connector wire.

- Connect the new bulb into the connector wire

5b) To change the HALOGEN bulb:

- Remove the bulb from the tray and separate the BULB from the Ceramic Disc

- Add new bulb by inserting new bulb pins into the Ceramic Disc

Ceramic Disc

Bulb holder tray

Connector Wire

Bulb Connector

8

Page 5: COLPOSCOPE MODEL 935 - Shopify · The Objective lens can be protected by use of the Sterilizable Protector Cap. It is fitted on the external diameter of the objective lens and protects

THE COLPOSCOPE HEAD

Composition

The Colposcope Head is composed by:• The Binoculars,• One Objective,• An illumination system, and• A floor stand.

OPTIC POD COMPONENTS

4

Bulb Change Instructions:

2) Unscrew the two thumb locks on the front of the light source and pull out the light bulb tray.

Thumb Locks

3) Be Cautious when

removing the front light bulb tray. The light bulbs are attached to the electronics within the light source.

1) TURN OFF POWER TO THE LIGHT SOURCE OR REMOVE THE POWER CORD

Page 6: COLPOSCOPE MODEL 935 - Shopify · The Objective lens can be protected by use of the Sterilizable Protector Cap. It is fitted on the external diameter of the objective lens and protects

CLEANINGThe Objective, at the bottom of the Colposcope body, may be exposed to blood and other bodily fluids. Spots on the objective dim the passage of light.

Alcohol or warm water in a mild detergent, applied with a clean cotton swab can be used to remove all smudges. Apply with soft and CIRCULAR movements. Follow your laboratory’s procedures to remove bloodstains or other contaminants from your equipment.

If the objective is too blotched, change to another cotton swab for each circular movement to avoid spreading the impurities again.

The Objective lens can be protected by use of the Sterilizable Protector Cap. It is fitted on the external diameter of the objective lens and protects it against mechanical damage and against contamination. (catalog #6132000)

Clean the metallic parts (chrome-plated or painted) with cotton, alcohol or mild detergent.

Lubrication It is recommended to lubricate the Colposcope head after 5 years of use.

HOW TO FOCUS THE COLPOSCOPE

Preliminarily:

Displace the Articulated Arm of the stand set where the colposcope is fixed, and point its objective to the direction of the field to be observed. Note that the objective, at front of the Colposcope, be positioned at a distance equal the focus engraved in millimeters in the external face of the metallic hoop of the objective. Example: With a 300mm objective, keep a distance of about 300mm between the objective and the specimen.

5

Focus the observation field

Interpupillary distance adjustment: Look through the two eyepieces with both eyes open and adjust the interpupillary distance as explained below in the items “inclined binoculars” and “straight binoculars”, until they are juxtaposed to the two pupils and the vision of the field appears as only one disk, clear and stereo-scopic. With the inclined binoculars grasp each tube of the of the binoculars and move the tubes in or out, so that, as it was already said, the eyepieces are juxtaposed to the observer’s pupils.

With the straight binoculars the action of separation or approach is made in a soft and semi-rotation movement, in both of the prism boxes.

You should now, using both eyes, see one clear stereoscopic image.

Parfocalizing Procedure:

1) Depress the orange tab on each eyepiece and rotate the zero (0) marking on the index line on the eyepiece tube.

2) Turn the magnification drum to the highest setting (5) and focus sharply on a target, such as a piece of paper with an “X” drawn on it, by using the focusing knobs.

3) The turn the magnification drum to the lowest setting (1) and focus by turning eyepieces clockwise or counter clockwise until you reach sharp focus. Be sure to depress the orange tab when turning the eyepiece.

4) Now the microscope is in focus for your eyes throughout all the magnification settings.

6