coco80 step by step tutorial

Upload: sugumar-sar-durai

Post on 03-Jun-2018

242 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/12/2019 CoCo80 Step by Step Tutorial

    1/25

    4633 Old Ironsides Drive, Suite 304, Santa Clara, CA 95054, USA

    1

    Step by Step Tutorial for CoCo-80

    This tutorial demonstrates a step by step approach to set up the signal display, analysis parameters, input and output

    channels, recording, signal save and other basic operations. It takes approximately 1 hour to complete the entire

    tutorial. The tutorial is divided into the following sections:

    1. Start up and Display Operations2. Parameters Settings3. Record the Time Streams and Save the Signals4. Set up the Trigger for Transient Capture5. Cursor Operation

    User Operation CoCo Screen and comments

    Part 1: Start up and Display Operation

    Part 1 demonstrates how to start up the CoCo, open a project file, add signals to a window, scale and pan within thewindow and change the display format.

    Press the Power Buttonand

    wait for the device to initialize.

    Press the Analysis Button.

    Use the Up/DownButtonsto

    select Frequency Response

    Folderand press the Enter

    Button.

    Use the Up/Down Buttonsto

    select

    FRF_COH_CPS_APS.csa

    then press theEnter Button.

    This project computes the auto-power spectra of each channels and the FRF

    between ch2, ch3 and ch4 to ch1. Ch1 is the excitation channel.

  • 8/12/2019 CoCo80 Step by Step Tutorial

    2/25

    4633 Old Ironsides Drive, Suite 304, Santa Clara, CA 95054, USA

    2

    The CSA will open and the

    display will show the live

    signals. Press the F6 Hold

    Button to stop the display.

    Press the F1 TracesButton. Depending on the project setup during previous use, the screen may show different

    signals. It can show auto-power spectra, FRF, or time domain signals. A time

    domain signals is shown in the picture below.

    Use the Up/DownButtonsto

    select Traceand Window

    Settings and press the Enter

    Button.

    Press the F1 Add Window.

  • 8/12/2019 CoCo80 Step by Step Tutorial

    3/25

    4633 Old Ironsides Drive, Suite 304, Santa Clara, CA 95054, USA

    3

    This step creates a new window and assigns the signals into the window.

    Use the Up/DownButtonsto

    select Add a Window with

    one Trace And pressthe

    Enter Button.

    A window may contain one or two traces.

    To display continuous time

    streams, highlight the signal

    ch1 and press the Enter

    Buttonto select it.

    Use the Up and Down Arrow

    Buttonsto select ch2 and

    press the Enter Button.

    Only signals with the same type can be overlaid together.

    Press the F6 OK Button.

    Now a new Window is created.

    Press the F4Buttonto switch

    from Autoto ZOOM.

  • 8/12/2019 CoCo80 Step by Step Tutorial

    4/25

    4633 Old Ironsides Drive, Suite 304, Santa Clara, CA 95054, USA

    4

    When ZOOM is active then the navigation Buttons can be used to manually adjust

    the display scale. When Auto is active then the display scale is automatically

    controlled to fit all the data in the window.

    Press the UpButtona few

    times to expand the Y axis.

    Press the Down Buttona few

    times to reduce the Y axis.

    Press the Right Arrow Button

    a few times to expand the X

    axis.

  • 8/12/2019 CoCo80 Step by Step Tutorial

    5/25

    4633 Old Ironsides Drive, Suite 304, Santa Clara, CA 95054, USA

    5

    Press the LeftArrow Buttona

    few times to reduce the X axis.

    Press the ShiftButtononce to

    switch from ZOOMto Move

    mode.

    When Move is active then the UP and Down Arrow Buttons move the signal plot

    up or down, left or right to view a specific part of the signal.

    Press the Up Arrow Buttona

    few times to move the Y axis.

    Press DownArrow Buttona

    few times to move the Y axis.

  • 8/12/2019 CoCo80 Step by Step Tutorial

    6/25

    4633 Old Ironsides Drive, Suite 304, Santa Clara, CA 95054, USA

    6

    Press the F4Buttonto change

    from the Move/Zoommode to

    Auto mode.

    This concludes the tutorial on scaling and moving around the window.

    Next, display the frequency

    signals by pressing the F1

    TracesButton and select

    Trace and Window Settings.

    Press the F1 Add Window

    Button and selectAdd a

    Window with one Trace.

    Next, display the auto power

    spectra by scrolling down the

    list, highlighting the signal

    APS(ch1)and pressing the

    EnterButtonto select it.

  • 8/12/2019 CoCo80 Step by Step Tutorial

    7/25

    4633 Old Ironsides Drive, Suite 304, Santa Clara, CA 95054, USA

    7

    Scroll down one row to the

    next signal, highlight

    APS(ch2)and press the Enter

    Button.

    Press the F6 OKButton.

    After OK is pressed, you should see the auto-power spectra display.

    Next, change the view mode by

    pressing the F1 TracesButton.

  • 8/12/2019 CoCo80 Step by Step Tutorial

    8/25

    4633 Old Ironsides Drive, Suite 304, Santa Clara, CA 95054, USA

    8

    Use the DownArrow Button

    and highlight Select View

    Mode.

    Press the EnterButton.

    Select LogMag.

    Press the Enter Button.

  • 8/12/2019 CoCo80 Step by Step Tutorial

    9/25

    4633 Old Ironsides Drive, Suite 304, Santa Clara, CA 95054, USA

    9

    Next create a Window with

    two traces by pressing the F1

    TracesButton.

    Press the F1Add Window

    with Two Traces Button.

    Scroll down the list, highlight

    the signal H(ch2, ch1)and

    press the EnterButtonto

    select it.

    Press the ShiftButtonto select

    the bottom trace.

  • 8/12/2019 CoCo80 Step by Step Tutorial

    10/25

    4633 Old Ironsides Drive, Suite 304, Santa Clara, CA 95054, USA

    10

    Scroll down the list, highlight

    the signal FRF(ch2, ch1)and

    press the Enter Button to

    select it.

    Press the F5 View Mode

    Button.

    Select Phaseand press theEnter Button.

  • 8/12/2019 CoCo80 Step by Step Tutorial

    11/25

    4633 Old Ironsides Drive, Suite 304, Santa Clara, CA 95054, USA

    11

    Press the F6 OKButton.

    Then a window with two traces

    will be displayed.

    At this stage, if you have created too many windows under the Traces menu. You

    can go back to delete some of them so the windows left are those that you

    frequently want to see.

  • 8/12/2019 CoCo80 Step by Step Tutorial

    12/25

    4633 Old Ironsides Drive, Suite 304, Santa Clara, CA 95054, USA

    12

    Part 2: Set the ParametersPart 2 demonstrates how to change the sensitivity, units and output channel settings.

    Press the F2 Param.Button,

    Select Sampling Rate to set

    appropriate sampling rate.

    Press the EnterButtonto

    accept the sampling rate.

    Press the F2 Param.Button,

    use the UP and Down Arrow

    Buttonsto highlight Input

    Channelsand press the Enter

    Button.

    Use the Up and Down Arrow

    and EnterButtonsto select

    and set the quantity, sensitivity

    and engineering unit (details

    not shown).

  • 8/12/2019 CoCo80 Step by Step Tutorial

    13/25

    4633 Old Ironsides Drive, Suite 304, Santa Clara, CA 95054, USA

    13

    Press the F6 ApplyButtonto

    accept the changes.

    The F1 Button, Apply to All Chapplies the same setting to all input channels in

    one step.

    Press the F2 Param.Button,

    Use the Up and Down Arrow

    Buttonsto highlight Output

    Channel andpress the Enter

  • 8/12/2019 CoCo80 Step by Step Tutorial

    14/25

    4633 Old Ironsides Drive, Suite 304, Santa Clara, CA 95054, USA

    14

    Button.

    Use the Up and Down Arrow

    Buttonsto move down to

    Swept Sine.

    Use the RightArrow Button

    to set the focus to the right

    pane. You can change the

    output parameters here (details

    not shown).

    Press the F4Buttonto enable

    the output waveform generator

    then press the Display Button.

    Connect the signal source to

    Channel 1 using the provided

    output cable.

    After this operation with the signal source is fed into the channel 1, you should

    see live sine wave in the signal plot.

    Press the F1Traces Button,

    select Window: ch1,ch2and

    press the Enter Button.

  • 8/12/2019 CoCo80 Step by Step Tutorial

    15/25

    4633 Old Ironsides Drive, Suite 304, Santa Clara, CA 95054, USA

    15

    Press the F2 Param.Button.

    Use the Up and Down Arrow

    Buttonsto highlight the

    Analysis Parameters and

    press the Enter Button. You

    can change each of the

    Analysis Parameters here

    (details not shown).

  • 8/12/2019 CoCo80 Step by Step Tutorial

    16/25

    4633 Old Ironsides Drive, Suite 304, Santa Clara, CA 95054, USA

    16

    Part 3: Record the Time Streams and Save the SignalsPart 3 demonstrates how to record time streams, save data frames and review saved data on the CoCo.

    Press the F2 Params. Button,

    use Up and Down Arrow

    Buttonsto highlight the Time

    Streams Setup andpress the

    EnterButton.

    Highlight ch1and press the

    EnterButton to check it.

    Highlight ch2and press theEnterButtonto check it.

    Press the F6 OK Button.

    This setup is designed for the convenience of time stream recording.

    Press the Rec./StopButton

    once, to start time stream

    recording.

  • 8/12/2019 CoCo80 Step by Step Tutorial

    17/25

    4633 Old Ironsides Drive, Suite 304, Santa Clara, CA 95054, USA

    17

    Press the Rec./StopButtona

    second time to stop time stream

    recording.

    See the figure above.

    Press the FileButton.

    Press the F1 FilesButtonto

    see a list of recorded timestreams signals or saved

    frequency signals.

    Use the Up and Down Arrow

    Buttonsto highlight a file andpress the F5View Files

    Buttonto view the highlighted

    file.

    Press the F1Buttonto delete

    the all the files newer than the

    file that is highlighted files;

    Press the F2Buttonto delete

    all the files.

    Press the F4Buttonto copy

    the highlighted files to the SD

    card.

  • 8/12/2019 CoCo80 Step by Step Tutorial

    18/25

    4633 Old Ironsides Drive, Suite 304, Santa Clara, CA 95054, USA

    18

    Press the DisplayButtonto go

    back to the signal display.

    Press the F2 Params. Button,

    use theUP and Down Arrow

    Buttons to highlight the Signal

    Snapshot Save Setupand

    press the EnterButton.

    Highlight Block(ch1)and press

    the Enter Button. Highlight

    H(ch2,ch1)and press the

    EnterButton(or select more

    signals to save).

    Press the F6 OK Button.

    Press the SaveButtontwice to

    save the signals.

    The first time you press the Save Button, a list of commands is shown. Simply

    press it a second time to save the defined signals.

    Now that the signals have been saved you can press the File Button to review or

    recall the signals.

  • 8/12/2019 CoCo80 Step by Step Tutorial

    19/25

    4633 Old Ironsides Drive, Suite 304, Santa Clara, CA 95054, USA

    19

  • 8/12/2019 CoCo80 Step by Step Tutorial

    20/25

    4633 Old Ironsides Drive, Suite 304, Santa Clara, CA 95054, USA

    20

    Part 4: Set up the Trigger for Transient CapturePart 4 demonstrates how to set up a trigger to capture a transient event including setting the trigger mode, level, pre-

    trigger and accept/reject mode. Make sure the signal source is still on and ch1 is connected to the source with a

    cable. If the signal source is not on, please go back to Output Channel setup and enable it

    Press the F2 Params. Button,

    use the Up and Down Arrow

    Buttonsto highlight the

    Acquisition Modeand press

    the EnterButton.

    Press the F1 Mode Button.

    Use the Up and Down Arrow

    Buttons to highlight the

    Manual-Arm Triggerand,

    press the Enter Button.

    Manual Re-Armmeans that in the transient capture, operator you will have a

    chance to accept or reject the captured time blocks.

    Press the F2 Trigger Source

    Button.

  • 8/12/2019 CoCo80 Step by Step Tutorial

    21/25

    4633 Old Ironsides Drive, Suite 304, Santa Clara, CA 95054, USA

    21

    Highlight ch1and press the

    EnterButton.

    The screen will show a live signal of containing the trigger source.

    Press the F4Set Trigger Level

    Button.

    This selection puts the navigation buttonButton into the functions of setting trigger

    level.

    Use the Upand DownArrow

    Buttonsand select Set Trigger

    Level. Then use the Up and

    Down Arrow Buttonsto

    change thetrigger thresholdlevel. Choose a level that is

    below the maximum value of

    the sine wave as shown in the

    fiture.

  • 8/12/2019 CoCo80 Step by Step Tutorial

    22/25

    4633 Old Ironsides Drive, Suite 304, Santa Clara, CA 95054, USA

    22

    Press the Leftand Right

    Arrow Buttonsto set the pre-

    trigger percentage.

    Press the F4Set Buttonand

    select Fixed Scale.

    This operation put changes the navigation buttonButtons mode to functions of

    setting the scaling window instead of setting up the trigger threshold.

    Use the Up and /Down

    Buttonsto scale the Y axis

    within the window.

    Press the F6 ApplyOK

    Buttonto accept the trigger

    setup.

    Use the Left/RightArrow

  • 8/12/2019 CoCo80 Step by Step Tutorial

    23/25

    4633 Old Ironsides Drive, Suite 304, Santa Clara, CA 95054, USA

    23

    buttonButtons to switch

    between Hide, Acceptand

    Reject. Select Move to Reject

    and press the Enter Button.

    Next select Acceptand press

    the EnterButton.

    This must be the second capture. Accept means that the signals will be accepted for

    FFT transform and averaged into the spectra and FRF.

    Select Hideand press the

    Enter Button.

    Press the EnterButtonagain. The trigger operation window will be shown again.

  • 8/12/2019 CoCo80 Step by Step Tutorial

    24/25

    4633 Old Ironsides Drive, Suite 304, Santa Clara, CA 95054, USA

    24

    Press the F2 Params. Button,

    use the Up and Down Arrow

    Buttonsto highlight the

    Acquisition Mode and, press

    the Enter Button. ;

    Press the F1 Mode Button.

    Use the Up and Down Arrow

    Buttons to highlight Free

    Run, and press the Enter

    Button. Press theDisplay

    Button.

    This operation will put change the CoCo back to Free Run mode. In Free-run

    mode, means the time capture occurs will be done as fast as possible with

    overlapping.

  • 8/12/2019 CoCo80 Step by Step Tutorial

    25/25

    4633 Old Ironsides Drive, Suite 304, Santa Clara, CA 95054, USA

    Part 5: Use CursorsPart 5 demonstrates how to add cursors to a window and move them to analyze a signal. Make sure the screen

    contains is in any signal display. If not, press the DisplayButton.

    Press the F5 Cursors Button.

    This operation puts the navigation buttonButtons into the cursor control mode.

    Highlight Add Cursor X1and

    press the Enter Button.

    Use the Left/RightArrow

    Buttonsto move the cursor.

    Note the value of the signals is displayed on the right.

    Press the F5Cursors Button

    and select Remove Cursor X1

    to remove it.