Mike McEvoy, PhD, RN, CCRN, NRP Senior Staff RN – Cardiothoracic Surgical ICUs Albany Medical College – Albany, New York Chair – Resuscitation Committee – Albany Medical Center EMS Coordinator – Saratoga County, New York EMS Editor – Fire Engineering magazine Check a Pulse! When to Question SpO 2 , NIBP & EtCO 2 Readings

Upload: race

Post on 15-Jan-2016

43 views

Category:

Documents


3 download

DESCRIPTION

Check a Pulse! When to Question SpO 2 , NIBP & EtCO 2 Readings. Mike McEvoy, PhD, RN, CCRN, NRP Senior Staff RN – Cardiothoracic Surgical ICUs Albany Medical College – Albany, New York Chair – Resuscitation Committee – Albany Medical Center EMS Coordinator – Saratoga County, New York - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: Check a Pulse!  When to Question SpO 2 , NIBP & EtCO 2  Readings

Mike McEvoy, PhD, RN, CCRN, NRPSenior Staff RN – Cardiothoracic Surgical ICUs

Albany Medical College – Albany, New YorkChair – Resuscitation Committee – Albany Medical Center

EMS Coordinator – Saratoga County, New YorkEMS Editor – Fire Engineering magazine

Check a Pulse! When to Question SpO2, NIBP & EtCO2 Readings

Page 2: Check a Pulse!  When to Question SpO 2 , NIBP & EtCO 2  Readings

Learning Objectives

Upon completion of the presentation the participant will:

1. Recall two common sources of user error in non-invasive vital sign measurement

2. Discuss the methodology used to obtain a non-invasive blood pressure reading

3. State one response of a pulse oximeter when unable to detect a pulse

Talk Code = 711

Page 3: Check a Pulse!  When to Question SpO 2 , NIBP & EtCO 2  Readings

Case # 1 - Desaturation• While charting…

• SpO2 alarms 74%

• Patient in no distress, good color

• Repositioning sensor yields same 74% sat

• ABG shows 98% sat

Page 4: Check a Pulse!  When to Question SpO 2 , NIBP & EtCO 2  Readings

Well appearing patient, 74% SpO2

•Why me?

Page 5: Check a Pulse!  When to Question SpO 2 , NIBP & EtCO 2  Readings

Case # 2 – O2 Sat Out Of Nowhere…

• Patient discharged 2 hours ago

• Mysteriouswaveformand 100% sat

Page 6: Check a Pulse!  When to Question SpO 2 , NIBP & EtCO 2  Readings

Model of Light Absorption At Measurement Site Without Motion

ACAC Variable light absorption due pulsatile volume of arterial blood

DCDC Constant light absorption due to non-pulsatile arterial blood.

DCDC Constant light absorption due to venous blood.

DC Constant light absorption due to tissue, bone, ...

Abs

orpt

ion

TimeTime

Page 7: Check a Pulse!  When to Question SpO 2 , NIBP & EtCO 2  Readings

Model of Light Absorption At Measurement Site With Motion

ACAC Variable light absorption due pulsatile volume of arterial blood

DCDC Constant light absorption due to non-pulsatile arterial blood.

AC Variable light absorption due to moving venous blood

DCDC Constant light absorption due to venous blood.

DC Constant light absorption due to tissue, bone ...

TimeTime

Abs

orpt

ion

Page 8: Check a Pulse!  When to Question SpO 2 , NIBP & EtCO 2  Readings

Influence of Perfusion on Accuracy of Conventional Pulse Oximetry During Motion

Good Perfusion (Conventional PO)

SpaO2=98

SpvO2=88SpO2=93

Poor Perfusion (Conventional PO)

SpO2=74

SpaO2=98

SpvO2=50

Page 9: Check a Pulse!  When to Question SpO 2 , NIBP & EtCO 2  Readings

Post Processor

R & IRDigitized, Filtered &

NormalizedR/IR

MEASUREMENT

CONFIDENCE % Saturation% Saturation

Conventional Pulse Oximetry Algorithm

3 options during motion or low perfusion:

1. Freeze last good value

2. Lengthen averaging cycle

3. Zero out

Page 10: Check a Pulse!  When to Question SpO 2 , NIBP & EtCO 2  Readings

Next Generation Pulse Oximetry

Page 11: Check a Pulse!  When to Question SpO 2 , NIBP & EtCO 2  Readings

Next Generation Pulse Oximetry

Page 12: Check a Pulse!  When to Question SpO 2 , NIBP & EtCO 2  Readings

Masimo SET: Signal Extraction Technology

SET “Parallel Engines”

Masimo SET “Parallel Engines”

R/IR(Conventional Pulse

Oximetry)

Confidence Based

Arbitrator

0 50% 66% 97% 100%SpO2% SpO2%

Post ProcessorDigitized,

Filtered & Normalized

% Saturation

SSTTM

Proprietary Algorithm 4

DST DST SET – 97%SET – 97%

DSTTM

FSTTM

MEASUREMENT

CONFIDENCE

MEASUREMENT

CONFIDENCE

MEASUREMENT

CONFIDENCE

MEASUREMENT

CONFIDENCE

MEASUREMENT

CONFIDENCE

R & IR

Page 13: Check a Pulse!  When to Question SpO 2 , NIBP & EtCO 2  Readings

A Solution for Patient Motion Discrete Saturation Transform (DST)

0 50% 66% 86% 97% 100%

SpOSpO2% %

Measure Through Motion Pulse Oximetry

Separating - accurate SpO2

Conventional Pulse Oximetry

0 50% 66% 86% 97% 100%

SpOSpO2% %

Averaging - inaccurate SpO2

Variable

Constant

Variable

Constant

In the presence of motion, SET separates the venous and arterial saturation values resulting in accurate saturation readings without false alarms (compared to conventional oximetry that averages the values to produce a reading)

Page 14: Check a Pulse!  When to Question SpO 2 , NIBP & EtCO 2  Readings

Certainty…

Page 15: Check a Pulse!  When to Question SpO 2 , NIBP & EtCO 2  Readings

Case # 3 – Smoke Inhalation

ED Triage Desk:

• 35 yo male presents with diff breathing

• States, “My furnace exploded.”

• Soot in mouth/nares

•O2 sat 98%

Page 16: Check a Pulse!  When to Question SpO 2 , NIBP & EtCO 2  Readings

Carbon Monoxide (CO)

• Gas:• Colorless• Odorless• Tasteless• Nonirritating

• Physical Properties:• Vapor Density = 0.97• LEL/UEL = 12.5 – 74%• IDLH = 1200 ppm

Page 17: Check a Pulse!  When to Question SpO 2 , NIBP & EtCO 2  Readings

Limitations of Pulse Oximetry

Barker SJ, Tremper KK. The Effect of Carbon Monoxide Inhalation on Pulse Oximetry and Transcutaneous PO 2. Anesthesiology 1987; 66:677-679

SpCO-SpO2 Gap:

The fractional difference between actual SaO2 and display of SpO2 (2 wavelength oximetry)

in presence of carboxyhemoglobin

SpCO-SpO2 Gap:

The fractional difference between actual SaO2 and display of SpO2 (2 wavelength oximetry)

in presence of carboxyhemoglobin

From Conventional Pulse Oximeter

From invasive CO-Oximeter Blood

Sample

[Blood]

Conventional pulse oximetry can not distinguish between COHb, and O2Hb

Page 18: Check a Pulse!  When to Question SpO 2 , NIBP & EtCO 2  Readings

CO: The Leading Cause of Poisoning Deaths

30-50 % of CO-exposed patients presenting to Emergency Departments are misdiagnosed

Barker MD, et al. J Pediatr. 1988;1:233-43

Barret L, et al. Clin Toxicol. 1985;23:309-13

Grace TW, et al. JAMA. 1981;246:1698-700

Page 19: Check a Pulse!  When to Question SpO 2 , NIBP & EtCO 2  Readings

Pulse CO-oximetry

Page 20: Check a Pulse!  When to Question SpO 2 , NIBP & EtCO 2  Readings

Hgb Signatures: CO, Met, Hgb…

Page 21: Check a Pulse!  When to Question SpO 2 , NIBP & EtCO 2  Readings

14,438 Patient Brown University Study

• Partridge and Jay (Rhode Island Hospital, Brown University Medical School), assessed carbon monoxide (CO) levels of 10,856 ED patients

• 11 unsuspected cases of CO Toxicity (COT) were discovered.Overall mean SpCO was 3.60%

• Occult COT was 4 in 10,000 during cold, 1 in 10,000 during warm months

• They concluded “unsuspected COT may be identified using noninvasive COHb screening and the prevalence of COT may be higher than previously recognized”

Non-Invasive Pulse CO-Oximetry Screening in the Emergency Department Identifies Occult Carbon Monoxide Toxicity. Suner S, Partridge R, Sucov A, Valente J, Chee K, Hughes A, Jay G. J Emerg Med 2008 Department of Emergency Medicine, Rhode Island Hospital, Brown Medical School, Providence, RI.

Page 22: Check a Pulse!  When to Question SpO 2 , NIBP & EtCO 2  Readings

Pulse Oximetry

Problems:

•Accuracy

•Motion & artifact

•Dyshemoglobins

Page 23: Check a Pulse!  When to Question SpO 2 , NIBP & EtCO 2  Readings

Case # 4 – Which Pressure Is Right?

78 yo trauma patient BP • A-line = 70/42 (50)• NIBP = 90/50 (52)

Page 24: Check a Pulse!  When to Question SpO 2 , NIBP & EtCO 2  Readings

Blood Pressure Monitoring

Direct

Pressure

vs

Indirect

Flow

Page 25: Check a Pulse!  When to Question SpO 2 , NIBP & EtCO 2  Readings

Errors in BP Measurement

Cuff Size:• Too large = BP• Too small = BP• 2/3 extremity length

Mid Heart Level:• Higher = BP• Lower = BP• Best sitting, arm @ side

Page 26: Check a Pulse!  When to Question SpO 2 , NIBP & EtCO 2  Readings

How does NIBP work?

• Measures flow (pulsatile)

• Determines HR and MAP

• By formula, calculatesSBP and DBP

• Subject to same interferences as auscultated BP

• Important to confirm HR (if wrong, SBP and DBP wrong)

Page 27: Check a Pulse!  When to Question SpO 2 , NIBP & EtCO 2  Readings

Mean Arterial Pressure (MAP)

• A clinical parameter useful in assessing perfusion

• Represents the average pressure within the arterial system throughout the cardiac cycle

• MAP = 2 (diastolic) + systolic

3

• 2/3 time in diastole only when HR = 70

Page 28: Check a Pulse!  When to Question SpO 2 , NIBP & EtCO 2  Readings

•28

150

90

60

Page 29: Check a Pulse!  When to Question SpO 2 , NIBP & EtCO 2  Readings

Waveform CapnographyAvailable for spontaneously breathing and for intubated patients

Page 30: Check a Pulse!  When to Question SpO 2 , NIBP & EtCO 2  Readings

Case # 5 – Bad Day in OR

• 37 yo male cholecystectomy• No significant PMH, smooth induction• Shortly after incision, EtCO2 gradually declines• Manual BVM with good compliance & chest rise• ???

Page 31: Check a Pulse!  When to Question SpO 2 , NIBP & EtCO 2  Readings

Circulation

The heart and lungs

are inextricably

linked together

Page 32: Check a Pulse!  When to Question SpO 2 , NIBP & EtCO 2  Readings

Cardiac Arrest!• Little O2 delivery or consumption

• Little CO2 production or venous return

• Little O2 delivery or consumption

• Little CO2 production or venous return

Page 33: Check a Pulse!  When to Question SpO 2 , NIBP & EtCO 2  Readings

In other words: CO2 production is largely

dependent on oxygen

consumption!

CO2 Clearance Reflects Perfusion

Page 34: Check a Pulse!  When to Question SpO 2 , NIBP & EtCO 2  Readings

Case # 6 – Misplaced ETT?

• Cardiac arrest on med-surg floor• CRNA intubates without difficulty, visualizes

tube pass through cords • EtCO2 circuit connected = flatline• ???

Page 35: Check a Pulse!  When to Question SpO 2 , NIBP & EtCO 2  Readings

Circuit Connector

Page 36: Check a Pulse!  When to Question SpO 2 , NIBP & EtCO 2  Readings

Case # 7 – EtCO2 ≠ PaCO2

• Post CABG patient EtCO2 drops to 6

• ABG PaCO2 = 48 mmHg

•Why?

Page 37: Check a Pulse!  When to Question SpO 2 , NIBP & EtCO 2  Readings

Another Cause of Low EtCO2

• Profound metabolic acidosis• pH = 6.93

Page 38: Check a Pulse!  When to Question SpO 2 , NIBP & EtCO 2  Readings

Questions?