AK 95 Op. man. HCEN9542. Rev. 08.2000Program version 5.xx
8:1
Alarm
Chapter 8
Alarm list, Attention list,Technical message list
![Page 2: ch08en](https://reader038.vdocuments.site/reader038/viewer/2022100601/557202b34979599169a3f6a0/html5/thumbnails/2.jpg)
8:2 AK 95 Op. man. HCEN9542. Rev. 08.2000Program version 5.xx
Mute time Additional information
2 min Low limitBlood pump stopsVenous clamp closesMinimum UF rate (when blood isdetected)When pressure is within limitsagain, the blood pump starts andthe venous clamp opens
High limitBlood pumps stopsMinimum UF rate (when blood isdetected)When pressure is within limitsagain, the blood pump starts andthe venous clamp opens
If the Venous pressure button ispressed for 3 s the low limitalarm is overridden for 5 – 120* s
2 min Alarm limits are too wideAfter 2 min an attention alarmappears
2 min
2 min
2 min Dialysis fluid bypassWhen the temperature is withinlimits again, the fluid enters thedialyzer automatically
Button
Flashing
Alarm
* Can be preset
Venouspressure
Venouspressure
Bloodflow
SNdata
Temp
![Page 3: ch08en](https://reader038.vdocuments.site/reader038/viewer/2022100601/557202b34979599169a3f6a0/html5/thumbnails/3.jpg)
AK 95 Op. man. HCEN9542. Rev. 08.2000Program version 5.xx
8:3
Alarm
Possible cause
Low alarmThe blood line has separated from the filter.The blood line has separated at the arterial or venousside of the access.The blood position of the venous needle has changed.Clotting before or in the filter.
High alarmThe venous blood line is kinked or clamped.The position of the venous needle has been changed.Clotting has occured after the venous pressuremeasurement e.g. in the blood line, in the venous dripchamber or in the needle.
The blood flow has been lower than the low alarm limitfor more than the preset time.The blood pump has been left in stop position too longdue to an alarm.The operator has left the blood pump in stop positiontoo long.
The stroke volume during SN is lower than the set lowalarm limit.
Low alarmThe machine has not yet stabilized the temperatureafter an interruption of the water supply.
High alarmThe temperature of the incoming water is too high.The machine has not yet stabilized the temperatureafter the function check.
![Page 4: ch08en](https://reader038.vdocuments.site/reader038/viewer/2022100601/557202b34979599169a3f6a0/html5/thumbnails/4.jpg)
8:4 AK 95 Op. man. HCEN9542. Rev. 08.2000Program version 5.xx
Button Mute time Additional information
2 min Dialysis fluid bypassWhen the conductivity is withinlimits again, the fluid enters thefilter automatically
2 min Low limitBlood pump stopsVenous clamp closesMin. UF rate (when blood is detected)When pressure is within limitsagain, the blood pump starts and thevenous clamp opens
High limitBlood pump stopsVenous clamp closesMin. UF rate (when blood is detected)When pressure is within limitsagain, the blood pump starts andthe venous clamp opens
If the arterial pressure button ispressed for 3 s the low limit and highalarms is overridden for 5–120* s
2 min Alarm limits are too wideAfter 2 min an attention alarmappears
Alarm
Flashing
* Can be preset
Cond
Arterialpressure
Arterialpressure
![Page 5: ch08en](https://reader038.vdocuments.site/reader038/viewer/2022100601/557202b34979599169a3f6a0/html5/thumbnails/5.jpg)
AK 95 Op. man. HCEN9542. Rev. 08.2000Program version 5.xx
8:5
Possible cause
Low alarmThe concentrate container/BiCart cartridge is empty.The concentrate has not been connected.Incorrect concentrate has been connected.The proper concentrate has not been selected on thedisplay.There is an interruption in the concentrate administra-tion to the machine e.g. the connector is not properlyconnected to the pick-up tube, the concentrate line iskinked or occluded.Air is drawn into the machine instead of concentrate.The filter(s) of the concentrate(s) is/are clogged.The content of the concentrate container does notcorrespond to the specifications given.
High alarmIncorrect concentrate has been connected.The proper concentrate has not been selected on thedisplay.The content of the concentrate container does notcorrespond to the specifications given.
Low alarmThe position of the arterial needle has been changed.The arterial blood line is kinked or clotted between thearterial access and the arterial pressure measurement.The set blood flow rate exceeds the blood flow that thearterial access has possibility to deliver.The position of the arterial needle has changed.
High alarmThe blood line has separated at the arterial side of theaccess.The position of the arterial needle has changed.Air or infusion enters the arterial line between thearterial access and the blood pump e.g. when a preblood pump infusion is connected.
Alarm
![Page 6: ch08en](https://reader038.vdocuments.site/reader038/viewer/2022100601/557202b34979599169a3f6a0/html5/thumbnails/6.jpg)
8:6 AK 95 Op. man. HCEN9542. Rev. 08.2000Program version 5.xx
Alarm
Button Mute time Additional information
2 min The TMP is outside the preset limits
2 min Alarm limits are too wideAfter 2 min an attention alarmappears
2 min
2 min An attention appears simultaneously
2 min An attention appears simultaneously
2 min Blood pump stopsArterial clamp closesVenous clamp closesIf the Air detect button is pressedfor 3 s the alarm is overridden for5 – 120* s
2 min Appears after an Air detect alarmPress the Air detect button toconfirm
Flashing
Flashing
Flashing
* Can be preset
TMP
TMP
UFvolume
UFrate
Airdetect
Airdetect
..................
![Page 7: ch08en](https://reader038.vdocuments.site/reader038/viewer/2022100601/557202b34979599169a3f6a0/html5/thumbnails/7.jpg)
AK 95 Op. man. HCEN9542. Rev. 08.2000Program version 5.xx
8:7
Alarm
Possible cause
Low alarmThe set value for the treatment time has beenincreased or the set value for the UF volume has beendecreased during treatment.Venous pressure decreases during treatment.
High alarmThe set value for the treatment time has been decrea-sed or the set value for the UF volume has beenincreased.Venous pressure increases during treatment.Partly clogged dialyzer.
The connection to the external computer system hasbeen interrupted.
No UF volume has been set when the machine detectsblood
Calculated set UF rate is lower than minimum UFrate set value.
Blood line(s) not properly connected to e.g. the arte-rial needle connection, the filter connection(s), theinfusion port(s).Blood line(s) or the filter not properly de-aired duringpriming.An empty infusion container with air inlet is con-nected to the blood line(s).The arterial needle is dislocated.
![Page 8: ch08en](https://reader038.vdocuments.site/reader038/viewer/2022100601/557202b34979599169a3f6a0/html5/thumbnails/8.jpg)
8:8 AK 95 Op. man. HCEN9542. Rev. 08.2000Program version 5.xx
2 min Blood pump stopsVenous clamp closesThe alarm action is delayed 3 sec-onds.
2 min The pH is outside the limits
2 min Blood pump stopsArterial clamp closesUF rate decreases and the dialysisfluid is bypassedIf the Blood leak button is pressedfor 3 s the alarm is overridden for5 – 120* sAfter treatment when the primingdetector does not detect blood, theblood leak alarm is delayed 30 sec.
2 min Appears after a Blood leak alarm.Press the Blood leak button toconfirm
2 min An attention appears simultaneously
Indefinite An attention appears simultaneously
Button Mute time Additional information
Flashing
Flashing
Flashing
Alarm
* Can be preset
Arterialguard
pH
Bloodleak
Bloodleak
Time
![Page 9: ch08en](https://reader038.vdocuments.site/reader038/viewer/2022100601/557202b34979599169a3f6a0/html5/thumbnails/9.jpg)
AK 95 Op. man. HCEN9542. Rev. 08.2000Program version 5.xx
8:9
Possible cause
The position of the arterial needle has been changed.The arterial blood line is kinked or clotted between thearterial access and the arterial pressure measurement.The set blood flow rate exceeds the blood flow that thearterial access has possibility to deliverThe position of the arterial needle has changed.
The content of the concentrate container does notcorrespond to the specifications given.Incorrect concentrate has been connected.Either the acid container or the bicarbonate container/BiCart cartridge has been emptied during bicarbonatetreatment.
A leakage, which allows blood to flow from the bloodcompartment to the dialysis fluid compartment, hasoccured.Air bubbles or particles, which can cause a false bloodleak alarm, have entered the blood leak detector.
The heparin rate has been set to zero when the machinedetects blood.
Treatment time has expired.
Alarm
![Page 10: ch08en](https://reader038.vdocuments.site/reader038/viewer/2022100601/557202b34979599169a3f6a0/html5/thumbnails/10.jpg)
8:10 AK 95 Op. man. HCEN9542. Rev. 08.2000Program version 5.xx
Attention list
The attentions are described in alphabetic order
Mute time Additional information Buzzer sound
A - PICK UP TUBE 5 min Buzzer is delayed 2 min Soft buzzerIMPROPER POSITION
ADJUST PUMPHEAD TO TUBE XXX 2 min Appears after a change of blood Soft buzzerPress BLOOD FLOW key to confirm pump segments has been done
The occlusion of the blood pump headmust be manually adjusted
AIRDETECTOR NOT ACTIVATED No buzzerPress AIRDETECTOR key
AIRLEAKAGE 2 min Buzzer is delayed 2 min Soft buzzerCheck fluid tubes connections Appears when dialysate tubes are
not properly connected to filterand priming detector detects bloodCaution: incorrect weight loss mayoccur if the machine is left for alonger period with this attention
ART PRESSURE LIMITS TOO WIDE 2 min Alarm limits are too wide Soft buzzerPress ARTERIAL key to adjust
AUTO PRIMING Appears during autopriming, press No buzzerWaiting for dialysis fluid the Fluid bypass button
B /ACETATE - PICK UP TUBE 5 min Buzzer is delayed 2 min Soft buzzerIMPROPER POSITION
BATTERY FAILURE Indefinite Press the Select arrow button to Soft buzzerChange battery or press SELECT confirm the information (in F.Ch)
No buzzer(in disinf.)
BiCart emptied Appears when the emptying of the No buzzerpress SELECT to confirm BiCart is completed
BiCart HOLDER NOT CLOSED 2min Buzzer is delayed 2 min Soft buzzerClose or check BiCart holder
BiCart HOLDER NOT INSTALLED 2min Soft buzzerCheck concentrate selection
BiCart NOT ATTACHED TO HOLDER Indefinite Buzzer is delayed 2 min Soft buzzerAttach or check BiCart
BLOOD DETECTED IN AUTO PRIMING Indefinite May occur during autopriming Soft buzzer
Attention
![Page 11: ch08en](https://reader038.vdocuments.site/reader038/viewer/2022100601/557202b34979599169a3f6a0/html5/thumbnails/11.jpg)
AK 95 Op. man. HCEN9542. Rev. 08.2000Program version 5.xx
8:11
BLOOD IN PRIMING DETECTOR Indefinite Priming detector detects blood ContinuousFunction check stopped during function check Clean the buzzer
window of the priming detector
BLOOD LEAKAGE DETECTOR FAILURE Indefinite Will only appear during Fch Soft buzzerClean and press BLOOD LEAK key
BLOOD PATH PRESSURE TEST FAILED Indefinite May occur during autopriming Soft buzzerPress SELECT key to continue if the leakage test of the dialyzer
fails
BLOOD PUMP COVER NOT CLOSED 2 min Blood pump stops Soft buzzerClose blood pump cover
BLOOD PUMP OVERLOAD 2 min Blood pump stops Soft buzzerPress BLOOD PUMP key
BLOOD PUMP OVERLOAD IN FCH Indefinite Soft buzzerCheck the pump and press SELECT
BLOOD PUMP STOP TIME EXPIRED 2 min Appears 3 min after blood ContinuousStart Blood pump pump is stopped buzzer
CENTRAL CHEM DISINF. COMPLETED Indefinite Will appear when the preset contactPress RINSE/DRAIN time in central disinfection has
passed.
CHEM DIS FILLUP PHASE COMPLETED Indefinite Soft buzzerPlace pick up tube in machine
CHEMICAL FILLUP NOT SUFFICIENT Indefinite Check pick-up tube and disinfectant Soft buzzerCheck supply of disinfectant container
CleanCart FILL COMPLETED Indefinite CleanCart disinf. with the Soft buzzerOpen upper latch and press HEAT latches closed after fill phase
CleanCart FILL COMPLETED Indefinite CleanCart rinse with the Soft buzzerOpen upper latch and press RINSE latches closed after fill phase
CleanCart NOT ATTACHED TO HOLDER Indefinite Buzzer is delayed 1 min Soft buzzerAttach or check CleanCart
CLEANING REQUIRED Indefinite Will disappear as soon as Soft buzzerxx hours since last cleaning priming is activated
Connect concentrate when time to Indefinite No concentrate(s) added Soft buzzerprepare machine for treatment
DECALCIFICATION REQUIRED Indefinite Will disappear as soon as Soft buzzerxx hours since last decalcific. priming is activated
DIALYSIS FLUID READY Appears after FCh, press the No buzzerPRESS BYPASS KEY bypass button to flush the dialysis
fluid tubes for 30 seconds.
Mute time Additional information Buzzer sound
Attention
![Page 12: ch08en](https://reader038.vdocuments.site/reader038/viewer/2022100601/557202b34979599169a3f6a0/html5/thumbnails/12.jpg)
8:12 AK 95 Op. man. HCEN9542. Rev. 08.2000Program version 5.xx
DISINFECTION REQUIRED Indefinite Will disappear as soon as Soft buzzerxx days since last disinfection priming is activated
DISCONNECT ARTERIAL PRESSURE 5 min Occurs during function check Soft buzzerPress ARTERIAL key to confirm if a arterial pressure more than
±2 mmHg is measured
DISCONNECT VENOUS PRESSURE 5 min Occurs during function check Soft buzzerPress VENOUS key to confirm if a venous pressure more than
±2 mmHg is mesured
EXIT AUTO PRIMING Indefinite Appears after autopriming when Soft buzzerPlace arterial line in clamp the venous drip chamber level
adjustment has been done.
EXTRA PRIMING VOLUME ACHIEVED 2 min If this attention appears during Soft buzzerinitiation of treatment, check thatvenous line is properly put into thepriming detector.
FLUID BYPASS TIME EXPIRED 2 min Appears 5 min after dialysis Soft buzzerPress BYPASS key fluid is bypassed
FLUID PATH OBSTRUCTION 2 min An obstruction has been ContinuousCheck fluid tubes and dialyzer detected Check if the obstruc- buzzer
tion is external i.e. in the fluidtubes or in the filter bypressing Fluid Bypass buttonIf the attention disappears theobstruction is external
FLUID TUBES IN SAFETY BYPASS 2 min The fluid tubes are still in safety Soft buzzerConnect fluid tubes to dialyzer bypass when blood is detected
by the priming detector
FLUID TUBES IN SAFETY BYPASS 2 min The fluid tubes are returned to SupersoftConnect fluid tubes to dialyzer safety bypass during treatment buzzer
Buzzer delayed 5 min
FLUID TUBES NOT IN SAFETY BYPASS 2 min Buzzer is delayed 2 min Soft buzzerFunction check is stoppeddisinf/rinse can not be started
FLUID TUBE SENSOR TEST PART 1(2) 2 min Soft buzzerRemove tubes, press SELECT, wait
FLUID TUBE SENSOR TEST PART 2(2) 2 min Soft buzzerAttach fluid tubes, press SELECT
HEATER EFFICIENCY NOT SUFFICIENT - Insufficient heat during the heat No buzzerFOR PROPER DISINFECTION program. Press Select arrow button
HEPARIN PUMP OVERLOAD 2 min Heparin pump stops Soft buzzerCheck syringe and line
Mute time Additional information Buzzer sound
Attention
![Page 13: ch08en](https://reader038.vdocuments.site/reader038/viewer/2022100601/557202b34979599169a3f6a0/html5/thumbnails/13.jpg)
AK 95 Op. man. HCEN9542. Rev. 08.2000Program version 5.xx
8:13
Mute time Additional information Buzzer sound
HEPARIN SET TO ZERO 2 min Buzzer is delayed 2 min Soft buzzerPress HEPARIN key to confirm Appears when priming detector
detects blood
High conductivity not obtained Indefinite Appears during Fch Soft buzzerCheck concentrate Check pick-up tube and container
HIGH MONITOR TEMPERATURE ALARM Indefinite Appears when electronic circuit Soft buzzerTurn off the monitor board temperature exceeds 55 ½C
IMPOSSIBLE AUTO START SETTING 2 min Soft buzzerCheck auto heat/rinse ready time
IMPOSSIBLE TIME OR UF SETTING 2 min Buzzer is delayed 2 min Soft buzzerAdjust TIME or UF VOLUME The high UF rate limit is reached
Adjust Time or UF volume untilcalculated UF rate is lower thanhigh limit
INCORRECT ACIDIC CONCENTRATE 2 min Soft buzzerCheck concentrate
INCORRECT BICARBONATE CONC 2 min Soft buzzerCheck concentrate
INCORRECT CONCENTRATE 2 min Soft buzzerCheck concentrate
INSUFFICIENT INLET PRESSURE 2 min Buzzer is delayed 5 min in Soft buzzerPlease check water supply priming phase but not delayed in
disinf. or treatment phaseThe blood pump rate is decreasedduring the alarm if a plate dialyzeris used
LOW UF RATE CAN NOT BE OBTAINED 5 min The positive pressure limit is Soft buzzerVenous pressure too high reached (negative TMP) Check
the dialyzer (clotted?) and thevenous pressure (high pressurekink between dialyzer and venouspressure measurement?) Correctaccordingly Adjust Time andUF volume until measuredUF rate is lower than calculated
NO CHANGE IN BLOOD PRESSURE 2 min Venous pressure is stable Soft buzzerCheck cause, press BLOODPUMP key during SN
Attention
![Page 14: ch08en](https://reader038.vdocuments.site/reader038/viewer/2022100601/557202b34979599169a3f6a0/html5/thumbnails/14.jpg)
8:14 AK 95 Op. man. HCEN9542. Rev. 08.2000Program version 5.xx
Mute time Additional information Buzzer sound
Attention
NEGATIVE UF RATE 2 min Relative negative UF volume of Soft buzzerCheck and press UF volume 500 ml or negative UF rate >50 ml/min
for 5 min. Check for "blockage" ofdraining tube. After correction ofthe fault, press the UF volume buttonfor 3 sec. for confirmation. Check theUF rate before leaving the machine,if this is not correct call for technicalservice.Caution: incorrect weight loss mayoccur if confirmation is done withouthaving corrected the cause for theattention.
NOT ENOUGH DISINFECTANT - May appear if a working solution is No buzzerselected but not if a concentratedsolution is selected
ONLY BLOOD MONITOR IS RUNNING Indefinite Only the blood unit starts after SupersoftPress SELECT to start fluid part a power failure. The fluid unit buzzer
has to be started manually
PLEASE WAIT 30 SEC. TO 2 min Appears if the bypass button is Soft buzzerBYPASS DIALYSIS FLUID* pressed during the flushing of the
dialysis fluid tubes
PRIMING VOLUME ACHIEVED Indefinite The blood pump stops. No buzzerIf the attention appears duringinitiation of treatment, check thatvenous line is properly put into thepriming detector.
READY FOR VENOUS PRESSURE TEST – – No buzzerPress VENOUS key to start
RECOVERED FROM POWER FAILURE 2 min Time display shows ”rec” until Soft buzzerPress TIME key to confirm Time button is pressed
REMOTE CONTROL FAILURE Indefinite Contact to CAM90 system failed Soft buzzerCheck parameters/connections
RINSE BACK VOLUME ACHIEVED 5 min Occurs during ”automatic” Soft buzzerrinse back
RUNTIME METER FAILURE Not implemented yetPress SELECT key to confirm
SAGS TEST FAILED Indefinite Check if the AK 95 is connected Soft buzzerto the water supply
* Can be preset
![Page 15: ch08en](https://reader038.vdocuments.site/reader038/viewer/2022100601/557202b34979599169a3f6a0/html5/thumbnails/15.jpg)
AK 95 Op. man. HCEN9542. Rev. 08.2000Program version 5.xx
8:15
Mute time Additional information Buzzer sound
Attention
TOO HIGH TMP REQUIRED 2 min Buzzer is delayed 2 min Soft buzzerIncrease time or decrease UF vol The negative pressure limit is
reached (high positive TMP)Check the dialyzer (too small/partly clotted?) and correctaccordingly Adjust Time andUF volume until TMP decreases
TMP LIMITS TOO WIDE 2 min Alarm limits are too wide Soft buzzerPress TMP key to adjust
To empty BiCart, press SELECT Appears when emptying of the No buzzerBiCart has been initiated by pressingthe Bicarbonte button twice.
To end treatment, press isol UF Appears if treatment time is turnedand zero time down to zero and there is still time
left under Isol UF.
To obtain total UF volume Indefinite Appears when the remaining total Soft buzzerISOL UF must be started time is =set isolated UF time +2 min
To set a negative TMP low limit: Press the TMP button within 10 s No buzzerPress TMP key and adjust and turn the Set knob
To start final flush, Appears in autopriming No buzzerpress BLOOD PUMP key when final flush is to be done
To start the air detector test: The venous drip chamber must be No buzzerFill the venous drip chamber filled with priming solution in order
to start the air detector test and tofinalize the function test
TREATMENT STARTED NO RESET MADE Indefinite Soft buzzerPress SELECT if OK, else RESET
TREATMENT TIME EXPIRED Indefinite SupersoftPress time for 3 seconds to confirm buzzer
UF RATE LIMIT REACHED 2 min Buzzer is delayed 2 min Soft buzzerRe-adjust TIME or UF VOLUME The ± 20% limit on UF rate has been
reached The cause may be astopped UF for too long AdjustTime and UF volume or allow for anew UF rate limit calculation (presseither Time or UF volume and thenSelect Arrow button)
UF VOLUME SET TO ZERO 2 min Buzzer is delayed 2 min Soft buzzerPress UF VOLUME key to confirm Appears when the priming detector
detects blood
![Page 16: ch08en](https://reader038.vdocuments.site/reader038/viewer/2022100601/557202b34979599169a3f6a0/html5/thumbnails/16.jpg)
8:16 AK 95 Op. man. HCEN9542. Rev. 08.2000Program version 5.xx
UF RATE LOWER THAN MINIMUM SET 2 min Buzzer is delayed 2 min Soft buzzerAdjust TIME, UF VOLUME or MIN UF Calculated set UF rate is lower
than minimum UF rate set value
UF STOP TIME EXPIRED 5 min Appears 5 min after START/STOP SupersoftPress START/STOP key button is pressed buzzer
VENOUS LOW LIMIT BELOW +10 mmHg 2 min Appears 10 min (presettable 2 - 20) ContinousPress Venous key to adjust after the low limit has been set buzzer
below +10 mmHg
VENOUS PRESSURE LIMITS TOO WIDE 2 min Alarm limits are too wide Soft buzzerPress VENOUS key to adjust
VENOUS PRESSURE TEST INTERRUPTED 2 min – Soft buzzerPress SELECT to confirm
Wrong disinfectant agent 2 min A too high or too low disinfectant Soft buzzerPress SELECT to restart disinf. concentration has been detected,
check cause and press SELECTkey shortly to repeat fill phase
Mute time Additional information Buzzer sound
Attention
![Page 17: ch08en](https://reader038.vdocuments.site/reader038/viewer/2022100601/557202b34979599169a3f6a0/html5/thumbnails/17.jpg)
AK 95 Op. man. HCEN9542. Rev. 08.2000Program version 5.xx
8:17
Technical message list
Mute time Additional information
TECHNICAL MESSAGE CALL SERVICE A technical message has appeared,x xxx xxx xxx call for a service technician
Depending on the message, the machinesometimes enters a patient safe statewhich makes it impossible to continue
Technical alarm
![Page 18: ch08en](https://reader038.vdocuments.site/reader038/viewer/2022100601/557202b34979599169a3f6a0/html5/thumbnails/18.jpg)
8:18 AK 95 Op. man. HCEN9542. Rev. 08.2000Program version 5.xx
This page is left intentionally blank