AK 95 Op. man. HCEN9542. Rev. 08.2000Program version 5.xx

8:1

Alarm

Chapter 8

Alarm list, Attention list,Technical message list

Page 2: ch08en

8:2 AK 95 Op. man. HCEN9542. Rev. 08.2000Program version 5.xx

Mute time Additional information

2 min Low limitBlood pump stopsVenous clamp closesMinimum UF rate (when blood isdetected)When pressure is within limitsagain, the blood pump starts andthe venous clamp opens

High limitBlood pumps stopsMinimum UF rate (when blood isdetected)When pressure is within limitsagain, the blood pump starts andthe venous clamp opens

If the Venous pressure button ispressed for 3 s the low limitalarm is overridden for 5 – 120* s

2 min Alarm limits are too wideAfter 2 min an attention alarmappears

2 min

2 min

2 min Dialysis fluid bypassWhen the temperature is withinlimits again, the fluid enters thedialyzer automatically

Button

Flashing

Alarm

* Can be preset

Venouspressure

Venouspressure

Bloodflow

SNdata

Temp

Page 3: ch08en

AK 95 Op. man. HCEN9542. Rev. 08.2000Program version 5.xx

8:3

Alarm

Possible cause

Low alarmThe blood line has separated from the filter.The blood line has separated at the arterial or venousside of the access.The blood position of the venous needle has changed.Clotting before or in the filter.

High alarmThe venous blood line is kinked or clamped.The position of the venous needle has been changed.Clotting has occured after the venous pressuremeasurement e.g. in the blood line, in the venous dripchamber or in the needle.

The blood flow has been lower than the low alarm limitfor more than the preset time.The blood pump has been left in stop position too longdue to an alarm.The operator has left the blood pump in stop positiontoo long.

The stroke volume during SN is lower than the set lowalarm limit.

Low alarmThe machine has not yet stabilized the temperatureafter an interruption of the water supply.

High alarmThe temperature of the incoming water is too high.The machine has not yet stabilized the temperatureafter the function check.

Page 4: ch08en

8:4 AK 95 Op. man. HCEN9542. Rev. 08.2000Program version 5.xx

Button Mute time Additional information

2 min Dialysis fluid bypassWhen the conductivity is withinlimits again, the fluid enters thefilter automatically

2 min Low limitBlood pump stopsVenous clamp closesMin. UF rate (when blood is detected)When pressure is within limitsagain, the blood pump starts and thevenous clamp opens

High limitBlood pump stopsVenous clamp closesMin. UF rate (when blood is detected)When pressure is within limitsagain, the blood pump starts andthe venous clamp opens

If the arterial pressure button ispressed for 3 s the low limit and highalarms is overridden for 5–120* s

2 min Alarm limits are too wideAfter 2 min an attention alarmappears

Alarm

Flashing

* Can be preset

Cond

Arterialpressure

Arterialpressure

Page 5: ch08en

AK 95 Op. man. HCEN9542. Rev. 08.2000Program version 5.xx

8:5

Possible cause

Low alarmThe concentrate container/BiCart cartridge is empty.The concentrate has not been connected.Incorrect concentrate has been connected.The proper concentrate has not been selected on thedisplay.There is an interruption in the concentrate administra-tion to the machine e.g. the connector is not properlyconnected to the pick-up tube, the concentrate line iskinked or occluded.Air is drawn into the machine instead of concentrate.The filter(s) of the concentrate(s) is/are clogged.The content of the concentrate container does notcorrespond to the specifications given.

High alarmIncorrect concentrate has been connected.The proper concentrate has not been selected on thedisplay.The content of the concentrate container does notcorrespond to the specifications given.

Low alarmThe position of the arterial needle has been changed.The arterial blood line is kinked or clotted between thearterial access and the arterial pressure measurement.The set blood flow rate exceeds the blood flow that thearterial access has possibility to deliver.The position of the arterial needle has changed.

High alarmThe blood line has separated at the arterial side of theaccess.The position of the arterial needle has changed.Air or infusion enters the arterial line between thearterial access and the blood pump e.g. when a preblood pump infusion is connected.

Alarm

Page 6: ch08en

8:6 AK 95 Op. man. HCEN9542. Rev. 08.2000Program version 5.xx

Alarm

Button Mute time Additional information

2 min The TMP is outside the preset limits

2 min Alarm limits are too wideAfter 2 min an attention alarmappears

2 min

2 min An attention appears simultaneously

2 min An attention appears simultaneously

2 min Blood pump stopsArterial clamp closesVenous clamp closesIf the Air detect button is pressedfor 3 s the alarm is overridden for5 – 120* s

2 min Appears after an Air detect alarmPress the Air detect button toconfirm

Flashing

Flashing

Flashing

* Can be preset

TMP

TMP

UFvolume

UFrate

Airdetect

Airdetect

..................

Page 7: ch08en

AK 95 Op. man. HCEN9542. Rev. 08.2000Program version 5.xx

8:7

Alarm

Possible cause

Low alarmThe set value for the treatment time has beenincreased or the set value for the UF volume has beendecreased during treatment.Venous pressure decreases during treatment.

High alarmThe set value for the treatment time has been decrea-sed or the set value for the UF volume has beenincreased.Venous pressure increases during treatment.Partly clogged dialyzer.

The connection to the external computer system hasbeen interrupted.

No UF volume has been set when the machine detectsblood

Calculated set UF rate is lower than minimum UFrate set value.

Blood line(s) not properly connected to e.g. the arte-rial needle connection, the filter connection(s), theinfusion port(s).Blood line(s) or the filter not properly de-aired duringpriming.An empty infusion container with air inlet is con-nected to the blood line(s).The arterial needle is dislocated.

Page 8: ch08en

8:8 AK 95 Op. man. HCEN9542. Rev. 08.2000Program version 5.xx

2 min Blood pump stopsVenous clamp closesThe alarm action is delayed 3 sec-onds.

2 min The pH is outside the limits

2 min Blood pump stopsArterial clamp closesUF rate decreases and the dialysisfluid is bypassedIf the Blood leak button is pressedfor 3 s the alarm is overridden for5 – 120* sAfter treatment when the primingdetector does not detect blood, theblood leak alarm is delayed 30 sec.

2 min Appears after a Blood leak alarm.Press the Blood leak button toconfirm

2 min An attention appears simultaneously

Indefinite An attention appears simultaneously

Button Mute time Additional information

Flashing

Flashing

Flashing

Alarm

* Can be preset

Arterialguard

pH

Bloodleak

Bloodleak

Time

Page 9: ch08en

AK 95 Op. man. HCEN9542. Rev. 08.2000Program version 5.xx

8:9

Possible cause

The position of the arterial needle has been changed.The arterial blood line is kinked or clotted between thearterial access and the arterial pressure measurement.The set blood flow rate exceeds the blood flow that thearterial access has possibility to deliverThe position of the arterial needle has changed.

The content of the concentrate container does notcorrespond to the specifications given.Incorrect concentrate has been connected.Either the acid container or the bicarbonate container/BiCart cartridge has been emptied during bicarbonatetreatment.

A leakage, which allows blood to flow from the bloodcompartment to the dialysis fluid compartment, hasoccured.Air bubbles or particles, which can cause a false bloodleak alarm, have entered the blood leak detector.

The heparin rate has been set to zero when the machinedetects blood.

Treatment time has expired.

Alarm

Page 10: ch08en

8:10 AK 95 Op. man. HCEN9542. Rev. 08.2000Program version 5.xx

Attention list

The attentions are described in alphabetic order

Mute time Additional information Buzzer sound

A - PICK UP TUBE 5 min Buzzer is delayed 2 min Soft buzzerIMPROPER POSITION

ADJUST PUMPHEAD TO TUBE XXX 2 min Appears after a change of blood Soft buzzerPress BLOOD FLOW key to confirm pump segments has been done

The occlusion of the blood pump headmust be manually adjusted

AIRDETECTOR NOT ACTIVATED No buzzerPress AIRDETECTOR key

AIRLEAKAGE 2 min Buzzer is delayed 2 min Soft buzzerCheck fluid tubes connections Appears when dialysate tubes are

not properly connected to filterand priming detector detects bloodCaution: incorrect weight loss mayoccur if the machine is left for alonger period with this attention

ART PRESSURE LIMITS TOO WIDE 2 min Alarm limits are too wide Soft buzzerPress ARTERIAL key to adjust

AUTO PRIMING Appears during autopriming, press No buzzerWaiting for dialysis fluid the Fluid bypass button

B /ACETATE - PICK UP TUBE 5 min Buzzer is delayed 2 min Soft buzzerIMPROPER POSITION

BATTERY FAILURE Indefinite Press the Select arrow button to Soft buzzerChange battery or press SELECT confirm the information (in F.Ch)

No buzzer(in disinf.)

BiCart emptied Appears when the emptying of the No buzzerpress SELECT to confirm BiCart is completed

BiCart HOLDER NOT CLOSED 2min Buzzer is delayed 2 min Soft buzzerClose or check BiCart holder

BiCart HOLDER NOT INSTALLED 2min Soft buzzerCheck concentrate selection

BiCart NOT ATTACHED TO HOLDER Indefinite Buzzer is delayed 2 min Soft buzzerAttach or check BiCart

BLOOD DETECTED IN AUTO PRIMING Indefinite May occur during autopriming Soft buzzer

Attention

Page 11: ch08en

AK 95 Op. man. HCEN9542. Rev. 08.2000Program version 5.xx

8:11

BLOOD IN PRIMING DETECTOR Indefinite Priming detector detects blood ContinuousFunction check stopped during function check Clean the buzzer

window of the priming detector

BLOOD LEAKAGE DETECTOR FAILURE Indefinite Will only appear during Fch Soft buzzerClean and press BLOOD LEAK key

BLOOD PATH PRESSURE TEST FAILED Indefinite May occur during autopriming Soft buzzerPress SELECT key to continue if the leakage test of the dialyzer

fails

BLOOD PUMP COVER NOT CLOSED 2 min Blood pump stops Soft buzzerClose blood pump cover

BLOOD PUMP OVERLOAD 2 min Blood pump stops Soft buzzerPress BLOOD PUMP key

BLOOD PUMP OVERLOAD IN FCH Indefinite Soft buzzerCheck the pump and press SELECT

BLOOD PUMP STOP TIME EXPIRED 2 min Appears 3 min after blood ContinuousStart Blood pump pump is stopped buzzer

CENTRAL CHEM DISINF. COMPLETED Indefinite Will appear when the preset contactPress RINSE/DRAIN time in central disinfection has

passed.

CHEM DIS FILLUP PHASE COMPLETED Indefinite Soft buzzerPlace pick up tube in machine

CHEMICAL FILLUP NOT SUFFICIENT Indefinite Check pick-up tube and disinfectant Soft buzzerCheck supply of disinfectant container

CleanCart FILL COMPLETED Indefinite CleanCart disinf. with the Soft buzzerOpen upper latch and press HEAT latches closed after fill phase

CleanCart FILL COMPLETED Indefinite CleanCart rinse with the Soft buzzerOpen upper latch and press RINSE latches closed after fill phase

CleanCart NOT ATTACHED TO HOLDER Indefinite Buzzer is delayed 1 min Soft buzzerAttach or check CleanCart

CLEANING REQUIRED Indefinite Will disappear as soon as Soft buzzerxx hours since last cleaning priming is activated

Connect concentrate when time to Indefinite No concentrate(s) added Soft buzzerprepare machine for treatment

DECALCIFICATION REQUIRED Indefinite Will disappear as soon as Soft buzzerxx hours since last decalcific. priming is activated

DIALYSIS FLUID READY Appears after FCh, press the No buzzerPRESS BYPASS KEY bypass button to flush the dialysis

fluid tubes for 30 seconds.

Mute time Additional information Buzzer sound

Attention

Page 12: ch08en

8:12 AK 95 Op. man. HCEN9542. Rev. 08.2000Program version 5.xx

DISINFECTION REQUIRED Indefinite Will disappear as soon as Soft buzzerxx days since last disinfection priming is activated

DISCONNECT ARTERIAL PRESSURE 5 min Occurs during function check Soft buzzerPress ARTERIAL key to confirm if a arterial pressure more than

±2 mmHg is measured

DISCONNECT VENOUS PRESSURE 5 min Occurs during function check Soft buzzerPress VENOUS key to confirm if a venous pressure more than

±2 mmHg is mesured

EXIT AUTO PRIMING Indefinite Appears after autopriming when Soft buzzerPlace arterial line in clamp the venous drip chamber level

adjustment has been done.

EXTRA PRIMING VOLUME ACHIEVED 2 min If this attention appears during Soft buzzerinitiation of treatment, check thatvenous line is properly put into thepriming detector.

FLUID BYPASS TIME EXPIRED 2 min Appears 5 min after dialysis Soft buzzerPress BYPASS key fluid is bypassed

FLUID PATH OBSTRUCTION 2 min An obstruction has been ContinuousCheck fluid tubes and dialyzer detected Check if the obstruc- buzzer

tion is external i.e. in the fluidtubes or in the filter bypressing Fluid Bypass buttonIf the attention disappears theobstruction is external

FLUID TUBES IN SAFETY BYPASS 2 min The fluid tubes are still in safety Soft buzzerConnect fluid tubes to dialyzer bypass when blood is detected

by the priming detector

FLUID TUBES IN SAFETY BYPASS 2 min The fluid tubes are returned to SupersoftConnect fluid tubes to dialyzer safety bypass during treatment buzzer

Buzzer delayed 5 min

FLUID TUBES NOT IN SAFETY BYPASS 2 min Buzzer is delayed 2 min Soft buzzerFunction check is stoppeddisinf/rinse can not be started

FLUID TUBE SENSOR TEST PART 1(2) 2 min Soft buzzerRemove tubes, press SELECT, wait

FLUID TUBE SENSOR TEST PART 2(2) 2 min Soft buzzerAttach fluid tubes, press SELECT

HEATER EFFICIENCY NOT SUFFICIENT - Insufficient heat during the heat No buzzerFOR PROPER DISINFECTION program. Press Select arrow button

HEPARIN PUMP OVERLOAD 2 min Heparin pump stops Soft buzzerCheck syringe and line

Mute time Additional information Buzzer sound

Attention

Page 13: ch08en

AK 95 Op. man. HCEN9542. Rev. 08.2000Program version 5.xx

8:13

Mute time Additional information Buzzer sound

HEPARIN SET TO ZERO 2 min Buzzer is delayed 2 min Soft buzzerPress HEPARIN key to confirm Appears when priming detector

detects blood

High conductivity not obtained Indefinite Appears during Fch Soft buzzerCheck concentrate Check pick-up tube and container

HIGH MONITOR TEMPERATURE ALARM Indefinite Appears when electronic circuit Soft buzzerTurn off the monitor board temperature exceeds 55 ½C

IMPOSSIBLE AUTO START SETTING 2 min Soft buzzerCheck auto heat/rinse ready time

IMPOSSIBLE TIME OR UF SETTING 2 min Buzzer is delayed 2 min Soft buzzerAdjust TIME or UF VOLUME The high UF rate limit is reached

Adjust Time or UF volume untilcalculated UF rate is lower thanhigh limit

INCORRECT ACIDIC CONCENTRATE 2 min Soft buzzerCheck concentrate

INCORRECT BICARBONATE CONC 2 min Soft buzzerCheck concentrate

INCORRECT CONCENTRATE 2 min Soft buzzerCheck concentrate

INSUFFICIENT INLET PRESSURE 2 min Buzzer is delayed 5 min in Soft buzzerPlease check water supply priming phase but not delayed in

disinf. or treatment phaseThe blood pump rate is decreasedduring the alarm if a plate dialyzeris used

LOW UF RATE CAN NOT BE OBTAINED 5 min The positive pressure limit is Soft buzzerVenous pressure too high reached (negative TMP) Check

the dialyzer (clotted?) and thevenous pressure (high pressurekink between dialyzer and venouspressure measurement?) Correctaccordingly Adjust Time andUF volume until measuredUF rate is lower than calculated

NO CHANGE IN BLOOD PRESSURE 2 min Venous pressure is stable Soft buzzerCheck cause, press BLOODPUMP key during SN

Attention

Page 14: ch08en

8:14 AK 95 Op. man. HCEN9542. Rev. 08.2000Program version 5.xx

Mute time Additional information Buzzer sound

Attention

NEGATIVE UF RATE 2 min Relative negative UF volume of Soft buzzerCheck and press UF volume 500 ml or negative UF rate >50 ml/min

for 5 min. Check for "blockage" ofdraining tube. After correction ofthe fault, press the UF volume buttonfor 3 sec. for confirmation. Check theUF rate before leaving the machine,if this is not correct call for technicalservice.Caution: incorrect weight loss mayoccur if confirmation is done withouthaving corrected the cause for theattention.

NOT ENOUGH DISINFECTANT - May appear if a working solution is No buzzerselected but not if a concentratedsolution is selected

ONLY BLOOD MONITOR IS RUNNING Indefinite Only the blood unit starts after SupersoftPress SELECT to start fluid part a power failure. The fluid unit buzzer

has to be started manually

PLEASE WAIT 30 SEC. TO 2 min Appears if the bypass button is Soft buzzerBYPASS DIALYSIS FLUID* pressed during the flushing of the

dialysis fluid tubes

PRIMING VOLUME ACHIEVED Indefinite The blood pump stops. No buzzerIf the attention appears duringinitiation of treatment, check thatvenous line is properly put into thepriming detector.

READY FOR VENOUS PRESSURE TEST – – No buzzerPress VENOUS key to start

RECOVERED FROM POWER FAILURE 2 min Time display shows ”rec” until Soft buzzerPress TIME key to confirm Time button is pressed

REMOTE CONTROL FAILURE Indefinite Contact to CAM90 system failed Soft buzzerCheck parameters/connections

RINSE BACK VOLUME ACHIEVED 5 min Occurs during ”automatic” Soft buzzerrinse back

RUNTIME METER FAILURE Not implemented yetPress SELECT key to confirm

SAGS TEST FAILED Indefinite Check if the AK 95 is connected Soft buzzerto the water supply

* Can be preset

Page 15: ch08en

AK 95 Op. man. HCEN9542. Rev. 08.2000Program version 5.xx

8:15

Mute time Additional information Buzzer sound

Attention

TOO HIGH TMP REQUIRED 2 min Buzzer is delayed 2 min Soft buzzerIncrease time or decrease UF vol The negative pressure limit is

reached (high positive TMP)Check the dialyzer (too small/partly clotted?) and correctaccordingly Adjust Time andUF volume until TMP decreases

TMP LIMITS TOO WIDE 2 min Alarm limits are too wide Soft buzzerPress TMP key to adjust

To empty BiCart, press SELECT Appears when emptying of the No buzzerBiCart has been initiated by pressingthe Bicarbonte button twice.

To end treatment, press isol UF Appears if treatment time is turnedand zero time down to zero and there is still time

left under Isol UF.

To obtain total UF volume Indefinite Appears when the remaining total Soft buzzerISOL UF must be started time is =set isolated UF time +2 min

To set a negative TMP low limit: Press the TMP button within 10 s No buzzerPress TMP key and adjust and turn the Set knob

To start final flush, Appears in autopriming No buzzerpress BLOOD PUMP key when final flush is to be done

To start the air detector test: The venous drip chamber must be No buzzerFill the venous drip chamber filled with priming solution in order

to start the air detector test and tofinalize the function test

TREATMENT STARTED NO RESET MADE Indefinite Soft buzzerPress SELECT if OK, else RESET

TREATMENT TIME EXPIRED Indefinite SupersoftPress time for 3 seconds to confirm buzzer

UF RATE LIMIT REACHED 2 min Buzzer is delayed 2 min Soft buzzerRe-adjust TIME or UF VOLUME The ± 20% limit on UF rate has been

reached The cause may be astopped UF for too long AdjustTime and UF volume or allow for anew UF rate limit calculation (presseither Time or UF volume and thenSelect Arrow button)

UF VOLUME SET TO ZERO 2 min Buzzer is delayed 2 min Soft buzzerPress UF VOLUME key to confirm Appears when the priming detector

detects blood

Page 16: ch08en

8:16 AK 95 Op. man. HCEN9542. Rev. 08.2000Program version 5.xx

UF RATE LOWER THAN MINIMUM SET 2 min Buzzer is delayed 2 min Soft buzzerAdjust TIME, UF VOLUME or MIN UF Calculated set UF rate is lower

than minimum UF rate set value

UF STOP TIME EXPIRED 5 min Appears 5 min after START/STOP SupersoftPress START/STOP key button is pressed buzzer

VENOUS LOW LIMIT BELOW +10 mmHg 2 min Appears 10 min (presettable 2 - 20) ContinousPress Venous key to adjust after the low limit has been set buzzer

below +10 mmHg

VENOUS PRESSURE LIMITS TOO WIDE 2 min Alarm limits are too wide Soft buzzerPress VENOUS key to adjust

VENOUS PRESSURE TEST INTERRUPTED 2 min – Soft buzzerPress SELECT to confirm

Wrong disinfectant agent 2 min A too high or too low disinfectant Soft buzzerPress SELECT to restart disinf. concentration has been detected,

check cause and press SELECTkey shortly to repeat fill phase

Mute time Additional information Buzzer sound

Attention

Page 17: ch08en

AK 95 Op. man. HCEN9542. Rev. 08.2000Program version 5.xx

8:17

Technical message list

Mute time Additional information

TECHNICAL MESSAGE CALL SERVICE A technical message has appeared,x xxx xxx xxx call for a service technician

Depending on the message, the machinesometimes enters a patient safe statewhich makes it impossible to continue

Technical alarm