breath alcohol analysis and the lion intoxilyzer® … alcohol analysis and the lion intoxilyzer®...

6
Breath alcohol analysis and the lion intoxilyzer® 6000: in vivo variations in breath alcohol concentration Paul M Williams, Helen E Cortis and David S Josty Lion Laboratories pic, Barry, Wales CF63 2BE, United Kingdom INTRODUCTION At midnight on 5 May 1983 evidential breath alcohol analysis was bom in the United Kingdom as the chief method of dealing with drink-drive suspects. Sampling and analysis were passed into the hands of the Police, who had been trained in the operational use of the then recently Approved evidential instruments - th e 'Lion Intoximeter® 3000’ and the 'Camic Breath Analyser'. Each of these is a 3 micron infrared-based system, with no mouth alcohol detection software, but to guard against acetone interference the 3000 is fitted with a secondary semiconductor device. The prescribed breath alcohol limit was set at 35 micrograms of alcohol per 100 millilitres [35|ig/100ml] - equivalent to 0.35mg/l BrAC, or 0.08% BAC. However, as a safeguard, instructions were issued to the Police that drivers should not be prosecuted unless their breath contained at least 40|tg/l 00ml of alcohol, with an additional blood/urine option being available to those drivers whose lower breath alcohol reading was not more than 50pg/100ml. British law requires the subject to provide TWO separate specimens of breath for analysis, and it is the lower of the two readings so produced that is used as the basis of the procedure, and evidence in Court. And so, in 1983, the stage was set for the British legal fraternity to embark on a crusade against the new system: trying to gain acquittals for their clients on very often quite spurious grounds, even when the defendant had unarguably been driving with a dangerously high level of alcohol. One of the main areas of challenge was in respect of 'differences between paired breath readings', which, claimed certain defence experts, was clear evidence of an instrument error, even though the device had accurately checked its calibration using a known alcohol vapour [from a ’wet-bath' simulator] both before and after the analysis of each subject's pair of breath specimens. - 305-

Upload: dangthu

Post on 19-May-2018

278 views

Category:

Documents


2 download

TRANSCRIPT

Breath alcohol analysis and the lion intoxilyzer® 6000:

in vivo variations in breath alcohol concentration

Paul M Williams, H elen E C o rtis an d D avid S Jo s ty

Lion Laboratories pic, Barry, Wales CF63 2BE, United Kingdom

INTRODUCTION

At midnight on 5 May 1983 evidential breath alcohol analysis was bom in the United Kingdom

as the chief method of dealing with drink-drive suspects. Sampling and analysis were passed

into the hands o f the Police, who had been trained in the operational use o f the then recently

Approved evidential instruments - th e 'L io n In tox im eter® 3000’ and the 'Camic Breath Analyser'. Each o f these is a 3 micron infrared-based system, with no mouth alcohol

detection software, but to guard against acetone interference the 3000 is fitted with a secondary

semiconductor device.

The prescribed breath alcohol limit was set at 35 micrograms of alcohol per 100 millilitres

[35|ig/100ml] - equivalent to 0.35mg/l BrAC, or 0.08% BAC. However, as a safeguard,

instructions were issued to the Police that drivers should not be prosecuted unless their breath

contained at least 40|tg/l 00ml of alcohol, with an additional blood/urine option being available

to those drivers whose lower breath alcohol reading was not more than 50pg/100ml. British

law requires the subject to provide TWO separate specimens of breath for analysis, and it is the

low er o f the two readings so produced that is used as the basis of the procedure, and evidence

in Court.

And so, in 1983, the stage was set for the British legal fraternity to embark on a crusade against

the new system: trying to gain acquittals for their clients on very often quite spurious grounds,

even when the defendant had unarguably been driving with a dangerously high level o f alcohol.

One o f the main areas of challenge was in respect o f 'differences between paired breath

readings', which, claimed certain defence experts, was clear evidence of an instrument error,

even though the device had accurately checked its calibration using a known alcohol vapour

[from a ’wet-bath ' simulator] both before and a fter the analysis o f each subject's pair of

breath specimens.

- 305-

BREATH ALCOHOL D IF F E R E N C E S AND T H E LION INTOXIMETER® 3000

The original reason for requiring the subject to provide two breath specimens was to use any

large decrease from the first to second as an indication of mouth alcohol. But no official

tolerance was ever established and each expert had his or her own view. This was partly

resolved by Sir William Paton in his report [Cobb and Dabbs, 1985], who recommended a

maximum difference o f 2 0 % [as a percentage o f the lower figure], but this was never enshrined

in any regulation or power.

Differences o f over 30% were observed, which resulted in much Court debate, and, in some

cases, a lingering doubt as to the reliability of the instrument involved. Some o f the observed

differences were interpreted in interesting ways, but with wrongful conclusions [Dossett,

1984],

By analysing data from a great many cases it became clear that the reasons for breath differences

when the Lion Intoxim eter® 30 0 0 was used were attributable to two principal types of

factor - these became known as in s tru m e n t in d u ced fa c to r s and su b jec t in d u c e d factors'.

Instrument induced factors - were broadly found to be:

acetone sensor misfimg - resulting in a false reduction of one [or both] breath alcohol

readings;

sampling system setting error - which allows the subject to supply breath of top lung air only; and

analogue/digital conversion error - resulting in an under-reading o f the breath alcohol

concentration.

Subject induced factors - were broadly found to be:

difference in expired volume - so one breath specimen is composed o f more deep lung air

than the other;

hyperventilation prior to blowing - resulting in a lowering of the alcohol content in that

specimen; mouth alcohol - from a recent drink or use of a mouth spray deodoriser, or

after regurgitation or vomiting;

mouth irrigation - following a drink of cold water;

a deliberate variation in blowing technique - such as ’puffing', rather than blowing

continuously; and

rapidly dissipated interfering substances present in breath - especially those derived from

solvent abuse.

- 306-

LION INTOXIM ETER® 3000: FEATURES W HICH IN CREA SE THE

PO SSIBILITY OF BREATH DIFFERENCES

This instrument was designed in the late 1970's and first described in 1980 [Forrester, 1980],

Some o f its features which have undoubtedly increased the potential for breath differences are:

Breath sam pling system

the instrument uses a flow sensor only, with no software to monitor the breath alcohol

expirogram ['] during expiration. There is a minimum discard volume requirement o f 1.5 litres,

but no maximum: this means that two successive breath specimens can be supplied from quite

different depths of the lungs, with a resulting difference in their alcohol concentrations. The

sampling system also allows a quasi-continuous [rapidly puffed] specimen to be accepted, so

that its alcohol level will be lower than if it had been supplied properly.

Secondary sem iconductor sensor for acetone d iscrim ination

this falsely reduces the alcohol reading on some breath specimens, so causing a difference if the

other specimen in the pair is not similarly affected [Cobb and Dabbs, 1985].

M outh alcohol

is not detected as there is no expirogram monitoring software.

N o n -sp e c if ic ity

so some solvent vapours may be read wrongly as alcohol.

SPECIFICATIO N AND DESIG N OF THE REPLACEM ENT IN ST R U M E N T :

TH E 'LIO N INTOXILYZER® 6000'

In October 1994 the British Authorities issued Guidelines for an instrument to replace the Lion

Intoximeter® 3000 [Home Office and Forensic Science Service UK, 1994]. A new requirement

is for the breath analysis to be aborted if the difference between the two readings is more than

15% [based on the lower reading], or by more than 5|ig/100ml, whichever is the greater.

The lion intoxilyzer® 6000 was submitted for evaluation and successfully completed all the

required technical procedures. These evaluations by the Forensic Science Service involved

testing using precisely controlled simulator devices only [in v itro ], with few tests carried out

on subjects.

1 the 'breath alcohol expirogram' is defined as the change in breath alcohol concentration during an expiiation, measured against elapsed time, discard volume and breath flow rate.

- 307 -

A research program was therefore commenced at Lion to investigate what magnitude o f breath

differences subjects could achieve in practice - in v iv o . We were particularly interested to

investigate whether the following aspects of the design of the new 6 0 0 0 , by comparison to the

3000 which it will soon replace, would minimise the possibility of significant breath

differences:

Breath sampling system

in the lion intoxilyzer® 60 0 0 the subject must provide a minimum volume o f 1.2 litres,

followed by software monitoring of the expirogram to detect deep lung air. The flow sensor

requires breath to be blown in a fully continuous way, with even the shortest break causing the

analysis to abort. This should reduce most of those subject-induced factors which are capable of

causing significant breath differences in the Lion Intoxim eter® 3 0 0 0 . In v iv o subject tests

were carried out on this aspect, as described in Section 5 o f this paper.

Misfiring of the secondary sensor

there is no secondary sensor in the 6000, so no possibility exists for breath differences caused

by misfiring o f such a device.

Residual m outh alcohol

the 6000's expirogram monitoring software will detect breath specimens which are polluted by

residual alcohol in the mouth or upper respiratory tract, and provide an appropriate warning to

the operator, who would then be unable to use those breath readings for Court purposes.

Specificityif a non-alcoholic substance is present in the breath at such a level that it affects the ethanol

analysis outside defined tolerances then the readings obtained cannot be used and the operator is

informed accordingly. This means that rapidly eluted materials, such as hydrocarbons and

toluene, as may be present in the breath after solvent abuse, will be detected and cause the

breath procedure to be aborted, rather than cause an apparently unexplainable significant breath

difference.

IN VIVO STUDIES ON THE LION INTOXILYZER® 6000

We investigated whether subjects could manipulate their expiration technique so as to produce

significant breath differences: that is, more than 15% or 5pg/100ml, whichever is the greater.

Volunteer subjects [n = 83, males and females], were given alcoholic beverages o f their choice

in such proportion as to raise their deep lung breath alcohol concentrations above the UK legal

limit o f 35|ig/100ml. O f the 83 subjects selected, 40 were classified as having 'normal'

- 308 -

respiratory function; 17 as having impaired lung function [asthma or the effects o f long term

heavy smoking]; while 26 were physically fit oarsmen, each with a larger than average lung

capacity.

In each case, having allowed at least 20 minutes for the dispersal of mouth alcohol, the subject

was instructed to deliver a full breath specimen into a lion in to x ily ze r® 6 0 0 0 so as to get a

true measurement of their deep lung breath alcohol concentration. He or she was then asked to

expire again into the device [within one minute o f the first specimen], this time in a manner only

just sufficient for the instrument to accept, in terms of flow or volume. The results were:

S u b j e c t G r o u p S i z e D i f f e r e n c e s D i f f e r e n c e s ' D i f f e r e n c e '

G r o u p [ n ] O v e r 5 U n i t s O v e r 1 5 % M e s s a g e s

'Normal' 4 0 7 2 1

Impaired 1 7 1 1 1

Large/Fit 2 6 3 1 1

Nine of the 'normal' subjects were then asked to vary their manner of blowing or pre-delivery

respiration, as much as they could, in an attempt to reduce the alcohol level and hence the

reading on their second specimen. In most cases these 'cheated' specimens were rejected by the

instrument, but o f those that were not - following prolonged and deliberate hyperventilation

[which was very obvious to the observer] - only two subjects could produce differences greater

than 15% [16.4% and 23.6%]. In the other 7 cases the breath differences that were produced

were too small to be classed as 'significant' or to cause the relevant message to be generated.

Hyperventilation cools the mouth and upper respiratory tract, so reducing the breath alcohol

concentration on expiration by both condensation and reversed secondary equilibration.

Although it has been reported [Schoknecht, 1992; Schoknecht and Stock, 1995] that breath

temperature monitoring systems can be used to correct for such an effect, it is the generally held

view outside Germany that this approach is flawed on both scientific a n d legal grounds.

- 309 -

CONCLUSION

Careful, overall design of the lion intoxilyzer® 6 0 0 0 instrument's sampling system and

controlling software means that significant breath differences caused only by subject-induced

factors should now occur. These factors will be attributable only to deliberate variation by the

subject o f his breathing technique, or to pre-breath delivery activity which, in most cases, will

be obvious to the Police Officer in charge of the instrument. It is concluded, therefore, that

significant breath differences should be rare in the operational use o f this new evidential breath

alcohol analyser.

REFERENCES

Cobb PGW and Dabbs MDG (1985), Foreword by Sir William Paton FRS. Report on the

performance o f the Lion intoximeter 3000 and Comic Breath Analyser evidential breath alcohol

measuring instruments during the period 16 April 1984 to 15 October 1984. HMSO, London

Dossett JA (1984). The significance of a difference between two successive readings on the

Lion Intoximeter 3000. The Law Society's Gazette 2840-1.

Forrester MR (1980). Advances in infrared technology for breath alcohol testing with particular

reference to the Intoximeter Model 3000. Proceedings o f the 8th International Conference on

Alcohol, Drugs and Traffic Safety, Stockholm, Sweden f Goldberg (Ed)] 2: 677-85.

Home Office and Forensic Science Service UK (1994). Evidential breath alcohol testing

instruments: a guide to type approval procedures fo r evidential breath alcohol testing

instruments used fo r road traffic law enforcement in Great Britain. HMSO, London.

SchoknechtG (1992). The influence of temperature on breath alcohol analysis. Proceedings o f

the 12th International Conference on Alcohol, Drugs and Traffic Safety, Cologne, Germany

[Utzelmann, Berghaus and Kroj (Eds)] 1: 392 -7.

Schoknecht G and Stock B (1995). The technical concept for evidential breath testing in

Germany. Proceedings o f the 13th International Conference on Alcohol, Drugs and Traffic

Safety, Adelaide, Australia [Kloeden and McLean (Eds)] 1: 134-40.

- 3 1 0 -