GO Systemelektronik GmbH [email protected] +49 431 5 80 80-0 www.go-sys.de +49 431 5 80 80-11 Introduction Fluorescence Oil in Water Calibration BlueTrace Product Solutions Functions & Features Application Areas BlueTrace Fluorescence Sensor for BTEX, PAH & Refined Oils Solutions by GO Systemelektronik

Upload: others

Post on 26-Apr-2022

5 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Booklet BlueTrace 2 en

GO Systemelektronik GmbH [email protected]+49 431 5 80 80-0 www.go-sys.de+49 431 5 80 80-11GO Systemelektronik GmbH [email protected]+49 431 5 80 80-0 www.go-sys.de+49 431 5 80 80-11

Introduction Fluorescence Oil in Water Calibration

BlueTrace Product Solutions

Functions & Features

Application Areas

BlueTraceFluorescence Sensor for BTEX, PAH & Refined Oils

Solutions by GO Systemelektronik

Page 2: Booklet BlueTrace 2 en

GO Systemelektronik GmbH 1 www.go-sys.de

Fluorescence spectral data

There is no universal fluorescence spectrum for all oils. Rather, the spectrum depends on the composition of the oil. Refined oils consist mainly of aromatic hydrocarbons, which in solution in water are often indicated in the BTEX collection parameter. The graph shows examples for typical fluorescence spectra of some refined oils. The BlueTrace oil in water sensor is suitable for the measurement of refined oils, BTEX and aromatic hydrocarbons.

Excitation

Hydraulic oil

Rela

tive

light

inte

nsity

Wavelength (nm)250 275 300 325 350 375 400

0

1

S1

Ener

gy

Ligh

t sou

rce

Abso

rptio

n

Fluorescence

Relaxation

Relaxation

Phosphorescence

IntercombinationT1

S0

Light source Detector

Oil particles

When light of a certain wavelength hits an oil particle, the oil emits light of a different wavelength shortly after excitation. This effect is called fluorescence. Fluo-rescence occurs not only in oils, but also in other substances. The BlueTrace oil in water sensor uses this effect to determine the concentration of refined oils in water.A transmitter installed in the sensor emits light at around 280 nm. The oil particles in the water absorb this energy and then emit light in a range from 300 to 400 nm. This light is measured by a detector.

The Jablonski diagram shows the fluorescence effect in detail. The oil particle absorbs the energy of the light, changes to a higher, unstable energy level and then falls back to the lower energy level. Part of the energy is released by the fluorescence effect. The intensity of the fluorescence is directly dependent on the concentration. By measuring the intensity at the detector, the concentration of the oil in the water can be determined.

Introduction Fluorescence - Oil in Water

Rela

tive

light

inte

nsity

Wavelength (nm)250 275 300 325 350 375 400

0

1

Excitation

DieselCooling lubricant

Rela

tive

light

inte

nsity

Wavelength (nm)250 275 300 325 350 375 400

0

1

Excitation

GO Systemelektronik GmbH 1 www.go-sys.de

Fluorescence spectral data

There is no universal fluorescence spectrum for all oils. Rather, the spectrum depends on the composition of the oil. Refined oils consist mainly of aromatic hydrocarbons, which in solution in water are often indicated in the BTEX collection parameter. The graph shows examples for typical fluorescence spectra of some refined oils. The BlueTrace oil in water sensor is suitable for the measurement of refined oils, BTEX and aromatic hydrocarbons.

Excitation

Hydraulic oil

Rela

tive

light

inte

nsity

Wavelength (nm)250 275 300 325 350 375 400

0

1

S1

Ener

gy

Ligh

t sou

rce

Abso

rptio

n

Fluorescence

Relaxation

Relaxation

Phosphorescence

IntercombinationT1

S0

Light source Detector

Oil particles

When light of a certain wavelength hits an oil particle, the oil emits light of a different wavelength shortly after excitation. This effect is called fluorescence. Fluoafter excitation. This effect is called fluorescence. Fluo-rescence occurs not only in oils, but also in other substances. The BlueTrace oil in water sensor uses this effect to determine the concentration of refined oils in water.A transmitter installed in the sensor emits light at around 280 nm. The oil particles in the water absorb this energy and then emit light in a range from 300 to 400 nm. This light is measured by a detector.400 nm. This light is measured by a detector.

The Jablonski diagram shows the fluorescence effect The Jablonski diagram shows the fluorescence effect in detail. The oil particle absorbs the energy of the in detail. The oil particle absorbs the energy of the light, changes to a higher, unstable energy level and light, changes to a higher, unstable energy level and light, changes to a higher, unstable energy level and then falls back to the lower energy level. Part of the energy is released by the fluorescence effect. The intensity of the fluorescence is directly dependent on the concentration. By measuring the intensity at the detector, the concentration of the oil in the water can be determined.

Introduction Fluorescence - Oil in Water

Rela

tive

light

inte

nsity

Wavelength (nm)250 275 300 325 350 375 400

0

1

Excitation

DieselCooling lubricant

Rela

tive

light

inte

nsity

Wavelength (nm)250 275 300 325 350 375 400

0

1

Excitation

Page 3: Booklet BlueTrace 2 en

GO Systemelektronik GmbH 2 www.go-sys.de

Specific calibration to specific oil typesSince the fluorescence spectrum is extremely depend-ent on the refined oil, the sensor functions optimally with a specific calibration to this oil. Also, the absorp-tion of the water has an influence on the measure-ment, so that for the calibration the oil should be mixed in the water, which is to be expected locally. In low concentrations a linear behaviour can be expect-ed, therefore, a two-point calibration can be sufficient. However, at least a three-point calibration is recom-mended.

Easy calibrationThe BlueTrace can easily be optimally calibrated to the specific application. All you have to do is hold the sensor in the prepared samples and then perform a multi-point calibration. For this purpose, only the reference values have to be entered into the BlueBox controller or the corresponding software.

Calibration

Dark non-reflective surface

Prepared sample

Sensor

Magnetic stirrer

Oil in water concentration (ppm)

Out

put s

igna

l of d

etec

tor (

mV)

0

0 50 100 150 200 250 300

450

400

350

300

250

200

150

100

50

Functions & Features

Selectable Measuring Range

The sensitivity of the receiver can easily be changed either directly on the controllers of GO Systemelek-tronik, or with the help of the freely available PC program. Thus, the sensor can optimally be adjusted to the required measuring range.

Modbus Interface

The BlueTrace features a Modbus RTU interface. This means that the sensor can not only be connected to a GO Systemelektronik controller, but can also be inte-grated into third-party controllers or directly to a PLC. The necessary Modbus protocol is freely available. Settings or calibrations are stored directly on the sensor and can be adapted with the freely available PC program. The sensor only needs to be connected to a PC, for example with a Modbus USB converter.

Robust Design

The BlueTrace is available in stainless-steel and titani-um. For most applications, stainless steel is sufficient. However, if the sensor is used in corrosive media or harsh environmental conditions, it is recommended to use the BlueTrace in a titanium housing.

Hydraulic oil

Diesel

Cooling lubricant

www.go-sys.de

The BlueTrace is available in stainless-steel and titani-um. For most applications, stainless steel is sufficient. However, if the sensor is used in corrosive media or harsh environmental conditions, it is recommended to

GO Systemelektronik GmbH

Specific calibration to specific oil typesSince the fluorescence spectrum is extremely dependent on the refined oil, the sensor functions optimally with a specific calibration to this oil. Also, the absorption of the water has an influence on the measurement, so that for the calibration the oil should be mixed in the water, which is to be expected locally. In low concentrations a linear behaviour can be expected, therefore, a two-point calibration can be sufficient. However, at least a three-point calibration is recommended.

Easy calibrationThe BlueTrace can easily be optimally calibrated to the specific application. All you have to do is hold the sensor in the prepared samples and then perform a multi-point calibration. For this purpose, only the reference values have to be entered into the BlueBox controller or the corresponding software.

CalibrationCalibration

Prepared sample

Out

put s

igna

l of d

etec

tor (

mV)

Out

put s

igna

l of d

etec

tor (

mV)

Out

put s

igna

l of d

etec

tor (

mV)

0

0 50

450

400

350

300

250

200

150

100

50

Page 4: Booklet BlueTrace 2 en

GO Systemelektronik GmbH 3 www.go-sys.deGO Systemelektronik GmbH

Application Areas - BlueTrace

■ Cooling water (e.g. in power plants)■ Leakage detection

Process Monitoring

■ Detection of contamination in rivers and lakes■ Maritime applications

Environmental Monitoring

■ Influent of dirnking water plants■ Influent to desalination plants

Drinking Water

■ Influent of WWTP■ Monitoring of wastewater of industrial plants (e.g. refineries, tire manufacturers, petrochemical industry)

Wastewater

Application Areas - BlueTraceApplication Areas - BlueTrace

www.go-sys.de3

Application Areas - BlueTrace

■ Cooling water Cooling water (e.g. in power plants)■ Leakage detection

Process Monitoring

■ Detection of contamination in rivers and lakes■ Maritime applications

Environmental Monitoring

■ Influent of dirnking water plants■ Influent to desalination plants

Drinking Water

■ Influent of WWTP■ Monitoring of wastewater of industrial plants(e.g. refineries, tire manufacturers, petrochemical industry)

Wastewater

Application Areas - BlueTrace

Page 5: Booklet BlueTrace 2 en

GO Systemelektronik GmbH 4 www.go-sys.de

Power supply 10 - 32 V DC

Power consumption (typical) 0.5 W

Measuring principle Fluorescence measurementEvaluation (300 - 400 nm)

Measuring range (typical) 0 - 30 / 100 / 300 ppm

Measuring accuracy 3 % FS

Detection limit (typical) 0.1 ppm

Measuring interval ≥ 1 s

Light source < 300 nm

Material Stainless steel 1.4404 / Titanium [optional]

Operation temperature range 0 °C to +55 °C

Weight 0.8 kg

Dimensions Length 150 mm; Ø 38 mm

Interface Modbus [RTU]

Maximum pressure 6 bar

Art. no. 461 6200

Technical dataParameters■ BTEX■ PAH■ Refined oil in water

Application Areas

Environmental Monitoring■ River water■ Surface water

Wastewater■ Effluent monitoring■ Trend analysis■ Early detection of discharge

Process Measurement & Control Technology■ Process monitoring in industrial facilities■ Control of process water treatment■ Process optimization

Drinking Water■ Quality control■ Alarm systems

The BlueTrace Oil in Water Sensor is a compact fluorescence probe for the measurement of BTEX, PAH and refined oil in water. The robust design of the Blue-Trace Sensor allows for applications in harsh condi-tions, such as corrosive media or at high pressure.

BlueTrace Oil in Water Sensor

Main Functions & Features

Selectable Measuring Range

Robust & Non-corrosive Modbus InterfaceCompact Design

Easy Calibration

Low Power Demand

Insensitive to Ambiant Light High Pressure Resistance

GO Systemelektronik GmbH 4 www.go-sys.de

Power supply 10 - 32 V DC

Power consumption (typical) 0.5 W

Measuring principle Fluorescence measurementEvaluation (300 - 400 nm)

Measuring range (typical) 0 - 30 / 100 / 300 ppm

Measuring accuracy 3 % FS

Detection limit (typical) 0.1 ppm

Measuring interval ≥ 1 s

Light source < 300 nm

Material Stainless steel 1.4404 / Titanium [optional]

Operation temperature range 0 °C to +55 °C

Weight 0.8 kg

Dimensions Length 150 mm; Ø 38 mm

Interface Modbus [RTU]

Maximum pressure 6 bar

Art. no. 461 6200

Technical dataParameters■ BTEX■ PAH■ Refined oil in water

Application Areas

Environmental Monitoring■ River water■ Surface water

Wastewater■ Effluent monitoring■ Trend analysis■ Early detection of discharge

Process Measurement & Control Technology■ Process monitoring in industrial facilities■ Control of process water treatment■ Process optimization

Drinking Water■ Quality control■ Alarm systems

The The BlueTrace Oil in Water Sensor is a BlueTrace Oil in Water Sensor is a BlueTrace Oil in Water Sensor compact fluorescencefluorescence probe for the measurement of BTEX, PAH and refined oil in water. The robust design of the Blueand refined oil in water. The robust design of the Blueand refined oil in water. The robust design of the Blue-Trace Sensor allows for applications in harsh condi-tions, such as corrosive media or at high pressure.

BlueTrace Oil in Water SensorBlueTrace Oil in Water Sensor

Main Functions & Features

Selectable Measuring Range

Robust & Non-corrosive Modbus InterfaceCompact Design

Easy Calibration

Low Power Demand

Insensitive to Ambiant Light High Pressure Resistance

Page 6: Booklet BlueTrace 2 en

GO Systemelektronik GmbH [email protected]+49 431 5 80 80-0 www.go-sys.de+49 431 5 80 80-11

Main Functions & Features

Cloud Data Service

SD Card Data Logger

Modular & Expandable

PID-Controller FunctionMonitoring Function

CAN bus, Modbus & ProfibusIntelligent Event Handling

1

2 Sensor Inputs

Parameters / Sensors■ Blue-green algae■ Chlorine■ Chlorine dioxide■ Chlorophyll a■ Conductivity■ Dissolved organic substances■ Dissolved oxygen■ ISE■ Level■ Oil in water

■ Ozone■ pH■ Redox (ORP)■ Salinity■ Temperature■ Turbidity■ Water pressure

Application Areas

Environmental Monitoring■ River water■ Surface water

Wastewater■ Effluent monitoring■ Trend analysis■ Early detection of discharge (fingerprint)

Process Measurement & Control Technology■ Process monitoring in industrial facilities■ Control of process water treatment■ Process optimization

Drinking Water■ Quality control■ Alarm systems

The BlueSense transducer facilitates the processing and calculation of measurement values of physical and chemical sensors. In addition, the transducer features a full controller function that can cover all process-oriented tasks.The BlueSense transducer is compatible with analo-gue and digital sensors and actuators of all manufac-turers.

BlueSense Transducer

Page 7: Booklet BlueTrace 2 en

GO Systemelektronik GmbH [email protected]+49 431 5 80 80-0 www.go-sys.de+49 431 5 80 80-11

Technical data

Power consumption (typical) 8 W

Dimensions (wxhxd) 23.5 x 18.5 x 11.9 cm

IP protection class IP 65

Graphic touch display 240 x 128 pixel

12 V DC (9 - 18 V)Power supply24 V DC (18 - 36 V)

230 V AC (90 - 260 V)

Ambient temperature -10 to +45 °C

Optional interfaces1x RS-232, RS-485, var. protocols e.g. Modbus

1x Ethernet [TCP/IP], Modbus [TCP/IP]

Profibus

GPRS / UMTS / LTE modem

1x CAN bus for connection to ISA, BlueBox, BlueMon & MPS

Inputs2x Sensor input for selectable parameters

2x Pulse-In (PNP/NPN selectable)

2x Digital-In (static) potential-free contacts

Outputs2x Current output 4-20 mA

2x Relay with a switching capacity of 24 V AC /DC; 0,5 A

2x Relay with a switching capacity of 230 V AC/DC; 2 A

Features & functionsPID controller

Signal compensation

Data storage on SD card

Multipoint calibration

The BlueSense transducer can optimally be adapted to the customer specific application through the modular config-uration of the sensor inputs.

Art. no. 485 0001BlueSense

Product configuration