Beam Test Result. Jeewon SEO World lab fellow. T10 Beam Test . 14. Oct. ~ 20. Nov. 2009 ALICE TOF 10 gap chamber and 24 gap MRPC was tested with different percentages of SF 6 . SF 6 Property : Highly electronegative Disadvantages Expensive cost Global warming gas. 10 Gap MRPC. - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: Beam Test Result

Beam Test Result

Jeewon SEO

World lab fellow

Page 2: Beam Test Result

T10 Beam Test 14. Oct. ~ 20. Nov. 2009ALICE TOF 10 gap chamber and 24 gap MRPC

was tested with different percentages of SF6.SF6

◦ Property : Highly electronegative◦ Disadvantages

Expensive cost Global warming gas

Page 3: Beam Test Result

10 Gap MRPC Gap size : 250 ㎛

Aluminum box volume : 200ℓ Gas flow rate : 20ℓ/h (O2 content was kept at 100 ppm)

Page 4: Beam Test Result

Beam Test Setup

-5GeV/c beam of π-, μ- . Trigger selected 10 ㎜ⅹ 10 ㎜ beam spot

(130 triggers/spill in every 20 seconds)

Page 5: Beam Test Result

Data Acquisition

Page 6: Beam Test Result

Slewing Correction

Time = Tchamber – Tave Tave : average time of the timing scintillators

Page 7: Beam Test Result

Time Resolution

σres = √( σch2-σsc

2)

σres : time resolution σch : measured time of signals σsc : jiter of scintillators

Ave(S1,S2)-Ave(S3,S4)

Page 8: Beam Test Result

Efficiency and Streamer Plateau

SF6 0 % 2 % 4 % 6 % 8 % 10 %Streamer

free and ef -ficient >90%

operation range

10.6~11.6 kV

10.8~13 kV

11~13.2 kV

11.2~13.4 kV

11.4~13.6 kV

11.6~13.8 kV

1 kV 2.2 kV 2 .2kV 2.2 kV 2.2 kV 2.2 kV

The knee of the effi-ciency plateau moves to high voltage when the percentage of SF6 in-crease.

Breakdown at 11.8 kV with the pure freonBreak down at

11.8kVwith pure freon

Page 9: Beam Test Result

Time resolution with various percentage of SF6

Time resolution is independent of the fraction of SF6.

Page 10: Beam Test Result

24 Gap MRPC

160 ㎛ gaps (total thickness : 3.84 ㎜ )

Gas flow rate : 20ℓ/h (box volume : 30ℓ)

-> O2 content was kept at

20 ppm

Threshold : 40 fC

Page 11: Beam Test Result

Data Acquisition

NINO mounted on chamber inside gas volume

Page 12: Beam Test Result

Compare time of one MRPC with another

Time resolution obtained by measuring time difference between two chambers

Timing scintillators are not used.

Page 13: Beam Test Result

Efficiency and Time resolution

Tdiff = (Tmean)MRPC1-(Tmean)MRPC2

The best resolution is 20.8 ps with pure freon and 19.1 ps with added 5% SF6.

Efficiency plateau is shifted to a lower voltage with the pure freon.

The streamer-free and efficient operation range

Pure freon : 1.6 kV (10.6kV ~ 12.2kV)

5% SF6 : 2 kV (11kV ~ 13 kV)

Page 14: Beam Test Result

Conclusion 10 gap MRPC

◦ Break down at 11.8kV with pure freon.◦ Work with 2% SF6, obtaining the time resolution below 50 ps with a

2.2 kV range.-> Seems to need the SF6 to avoid breakdown

24 gap MRPC◦ Work satisfactorily with pure freon

SF6 effect◦ Shift the efficiency plateau to a higher voltage ◦ Shift in streamer production to higher voltage◦ For 24 gap MRPC, the efficient operation range without streamer is

broader when SF6 added.

Time resolution is independent of SF6 fraction.