barington three axis magnetometer

Upload: madheshvaran-loganathan

Post on 03-Jun-2018

230 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/12/2019 Barington Three Axis Magnetometer

    1/22

    Operation Manual forMag-03 Three-Axis Magnetic Field Sensors

    Innovation in Magnetic Measuring Instruments

  • 8/12/2019 Barington Three Axis Magnetometer

    2/22

    BARTINGTON INSTRUMENTS

    Page 2 of 22 OM1004/26

    Table of Contents

    1. Legal Notices 4

    1.1. Copyright 4

    1.2. Trademarks 4

    2. How to Use this Manual 4

    2.1. Symbols glossary 4

    3. Safe Use 5

    4. Introduction 5

    5. General Description 5

    6. Enclosures 6

    7. Compatible Power Supply and Data Acquisition Units 6

    8. Cables 6

    9. Accessories 7

    9.1. Mag-03 Calibration Unit 7

    9.2. Mating Connectors 7

    9.3. Mounting accessories 7

    10. Mounting 8

    10.1. Mag-03MC, Mag-03MCES, Mag-03MCT, Mag-03MCFL & Mag-03IE8

    10.2. Mag-03MS and Mag-03MSES 8

    10.3. Mag-03MSS 8

    11. Operation 9

    11.1. Connector Pin Allocation 9

    11.2. Interface 9

    11.3. Power Supplies 9

    11.4. Signal/Power Ground 10

    11.5. Cabling 11

    11.6. Connecting Power 11

    11.7. Electromagnetic Compatibility 11

  • 8/12/2019 Barington Three Axis Magnetometer

    3/22

    BARTINGTON INSTRUMENTS

    Page 3 of 22 OM1004/26

    12. Performance 12

    12.1. Frequency Response 12

    12.2. Noise 12

    12.3. Over Range 12

    13. Signal Processing 13

    14. Troubleshooting, Care and Maintenance 14

    14.1. Troubleshooting 14

    14.2. Care and Maintenance 14

    15. Storage & Transport 14

    16. Disposal 14

    15.1. Waste Electrical and Electronic Equipment (WEEE) Regulations 14

    Appendix 1: Mag03MSS-TC addendum 15

    Appendix 2: Magnetic Units and Measurements 16

    A1.1. Conventions 16

    A1.2. Measurement Units 16

    A1.3. Conversion Table 17

    A1.4. Vector Measurements 17

    Appendix 3: Figures 18

  • 8/12/2019 Barington Three Axis Magnetometer

    4/22

    BARTINGTON INSTRUMENTS

    Page 4 of 22 OM1004/26

    1. Legal Notices

    1.1. Copyright

    The copyright of thisdocument is the property of Bartington Instruments Ltd.

    1.2. Trademarks

    Bartington is a trademark of Bartington Instruments Ltd in Australia, Brazil, Canada, China, the

    European Community, India, Japan, the countries of the Madrid Agreement & Protocol, Norway

    and the United States of America.

    2. How to Use this Manual

    This manual describes the installation, operation and maintenance of the Mag-03 range of three

    axis magnetic field sensors.

    Take the time to get well acquainted with your sensor by reading this manual. Knowing and

    understanding the equipment will ensure you experience the most reliable operation.

    When service or maintenance is required, please contact Bartington Instruments or your local

    agent company.

    Technical, mechanical, electrical, environmental and performance specifications for this product

    can be found in the product brochureon the Bartington Instruments website.

    2.1. Symbols glossary

    The following symbols used within this manual call your attention to specific types of information:

    WARNING:Indicates a situation in which serious bodily injury or death could result if thewarning is ignored.

    Caution:Indicates a situation in which bodily injury or damage to your instrument, or both,

    could result if the caution is ignored.

    Identifies items that must be disposed of safely to prevent unnecessary damage to the

    environment.

    Note:Provides useful supporting information on how to make better use of your purchase.

  • 8/12/2019 Barington Three Axis Magnetometer

    5/22

    BARTINGTON INSTRUMENTS

    Page 5 of 22 OM1004/26

    3. Safe Use

    WARNING: These products are not qualified for use in explosive atmospheres or life support

    systems. Consult Bartington Instruments for advice.

    4. Introduction

    These compact, high performance sensors with integral electronics provide measurements

    of static and alternating magnetic fields in three axes. The sensors, also described as

    magnetometers, convert magnetic flux density, measured in three axes, into a bipolar analogue

    voltage. Analogue output voltages Vx, Vy and Vz vary linearly with magnetic flux density.

    In designing the Mag-03 series, the policy has been to provide a high performance sensor havinga flat amplitude response and a small, predictable phase lag over a wide bandwidth. In order to

    offer maximum flexibility and not degrade the performance, the sensor has no internal filters.

    The analogue outputs may require external filters to optimise the performance, depending on the

    application. See Section 13 on Signal Processing.

    5. General Description

    This section describes the features common to the Mag-03 range of sensors. Where there are

    exceptions they are described in the product brochuredetailing the different types of enclosures.

    Three fluxgate sensing elements are mounted orthogonally at one end of an enclosure, which also

    contains the electronic circuitry. The connector is mounted at the opposite end of the enclosure.

    The position and direction of each sensing element is shown on the outside of the sensor, together

    with the product code, measuring range and serial number.

    Details of the enclosures, mounting, connector dimensions, connector pin allocation and the

    position of the sensing elements relative to the enclosure are given in outline drawings on

    the product webpage. The sensor elements are precisely aligned along the centre lines of the

    package.

    The sensors provide three high precision analogue outputs, proportional to the magnetic field

    along each axis. The relationship between the magnetic field and the analogue output is extremely

    linear.

    The low output impedance of the sensor ensures it can be operated over long cables and permits

    it to be interfaced to low impedance data acquisition systems. The zero field offset error, scale

    factor, orthogonality and frequency response are individually calibrated.

  • 8/12/2019 Barington Three Axis Magnetometer

    6/22

    BARTINGTON INSTRUMENTS

    Page 6 of 22 OM1004/26

    6. Enclosures

    The sensors are available in a variety of enclosures, with five measuring ranges. A full list of

    sensors with specifications is provided in the brochure.

    Note:Several of our enclosures are awarded IP (Ingress Protection) ratings as determined

    by the IEC 60529 standard. These ratings are shown in brochure DS0013, available from

    Bartington Instruments at www.bartington.com/mag-03.html. Users are advised to consult

    this standard, available from the IEC website at www.iec.ch, for the precise definitions and

    limitations of each IP rating.

    Note: Using your sensor in an environment that exceeds its rating may result in the need for

    repair at the customers expense.

    7. Compatible Power Supply and Data Acquisition Units

    A number of other Bartington Instruments products will work with the Mag-03 as power supply

    and/or data acquisition units. These are listed with their specifications on the Bartington

    Instruments website at http://www.bartington.com/data-acquisition-and-conditioning-units.html .

    These include, but are not necessarily limited to:

    the PSU1 Power Supply Unit

    the DecaPSU

    the Magmeter Power Supply and Display Unit

    the Spectramag-6 Data Acquisition Unit

    the SCU1 Signal Conditioning Unit

    the Mag-03DAM Data Acquisition Module

    the Decaport Analogue Interface module

    the DAS1 Data Acquisition System.

    8. Cables

    Cables are available to connect the range of Mag-03 sensors to the range of suitable Bartington

    Instruments power supply and data acquisition units. Specifications for each of the cables are

    given in the product brochure.

    Note: Cables must be ordered separately.

  • 8/12/2019 Barington Three Axis Magnetometer

    7/22

    BARTINGTON INSTRUMENTS

    Page 7 of 22 OM1004/26

    Note: Customers manufacturing their own cables must ensure the cables are shielded to

    prevent them picking up EM (electromagnetic) interference.

    9. Accessories

    9.1. Mag-03 Calibration Unit

    The Mag-03MS-CU is a battery powered unit which produces a sinusoidal alternating magnetic

    field of defined frequency and magnitude. The unit provides a reference magnetic field for

    checking the calibration of the Mag-03MS sensors, which have a square enclosure. For calibration

    checking of Mag-03MC and Mag-03IE sensors, adaptors are available .

    A separate calibration unit, Mag-03MSS-CU, is available for the Mag-03MSS submersible sensors.

    The units operate by passing a temperature stabilised, constant current through a single axis

    Helmholtz coil, with guides to align each of the sensor axes in turn. More details are provided in

    the product brochure.

    9.2. Mating Connectors

    All sensors, except for the Mag-03MSS, are supplied with a non-magnetic mating connector if no

    cable for connection to a power supply or data acquisition module is purchased.

    Note: The Mag-03MSS mating connector must be purchased separately.

    9.3. Mounting accessories

    Mounting options available are:

    Mag-03MC-BR Mounting bracket, for use with the cylindrical range of sensors

    Mag-03-T Tripod

    Mag-03-TA tripod adaptor

    Mag-03-MR Mounting rack

    Mag-03-LP Levelling platform, for use with the tripod adaptor and mounting rack.

    These are described further in the product brochure.

    See also Section 10below.

  • 8/12/2019 Barington Three Axis Magnetometer

    8/22

    BARTINGTON INSTRUMENTS

    Page 8 of 22 OM1004/26

    10. Mounting

    The method of mounting will depend on the application and the enclosure. For details of the

    mounting arrangements for each sensor, refer to the relevant outline drawing available on the

    product page of the Bartington Instruments website.

    Note: The use of magnetic materials in the mounting arrangement must be avoided. All

    mounting components should be checked before installation by introducing the component

    within the immediate vicinity of the sensing elements of a working magnetometer and

    observing any variation in the background field.

    The analogue output is positive for conventional flux direction, South to North, in the direction of

    the arrow shown on the label for each axis; i.e. the maximum positive output will be obtained from

    any axis when the arrow points towards magnetic north along the total field vector.

    10.1. Mag-03MC, Mag-03MCES, Mag-03MCT, Mag-03MCFL & Mag-03IE

    These sensors may be supported by the Mag-03-TA tripod adaptor and the Mag-03MC-BR

    mounting bracket. The bracket (see Figure 1) clamps around the cylindrical enclosure. In the

    case of the Mag-03IE family it clamps around the body of the sensor, not the sensor heads.

    The bracket is manufactured from reinforced epoxy resin and supplied complete with nylon

    mounting screws.

    Note: The label area of the sensor is recessed and should not be used for clamping.

    10.2. Mag-03MS and Mag-03MSES

    These sensors may be supported by the Mag-03-TA tripod adaptor. These sensors have threaded

    holes tapped in the base which is also the datum face. The sensors can be mounted on any flat,

    non-magnetic surface, including the top of the mounting bracket described in Section 9.2, using

    the two brass screws supplied. A thin gasket or a suitable sealant should be used to seal the base

    of the units against water penetration.

    Caution: The absolute maximum screw penetration depth within the body is 16 mm and this

    must not be exceeded.

    10.3. Mag-03MSS

    This sensor may be supported by the Mag-03-TA tripod adaptor. The Mag-03MSS has a square

    section pressure housing with three mounting holes, 4 mm in diameter, drilled through the body

    and counterbored for cheesehead screws. Screws are not provided due to the variable nature of

    the environmental service conditions which may be encountered.

  • 8/12/2019 Barington Three Axis Magnetometer

    9/22

    BARTINGTON INSTRUMENTS

    Page 9 of 22 OM1004/26

    11. Operation

    11.1. Connector Pin Allocation

    The connector pin or cable colour allocation for the connection to each package type is shown on

    the appropriate outline drawing available on the product page.

    11.2. Interface

    A simplified interface schematic for the Mag-03 series is shown in Figure 2. The analogue outputs

    for the X, Y and Z axes are buffered to give a low output impedance, enabling the unit to be

    operated over long cables and interfaced to low impedance data acquisition systems.

    11.3. Power Supplies

    The normal power supply of the sensors is between 12V and 17V. The ideal power supply units

    are those referenced in Section 7. Alternatively, users may wish to provide their own supply. This

    would normally provide 12V and, for the lowest noise applications, ripple in the power supply

    should be in the region of a few mV.

    Note: adequate performance of the sensor cannot be guaranteed if used with non-

    Bartington Instruments products.

    The nominal current requirements are +35mA and -6mA for the standard and basic versions

    and +26mA and -6mA for the low noise versions with an additional current in proportion to the

    measured field. The additional current is 1.4mA per 100mT per axis and will be drawn from the

    positive or negative supply depending on the direction of the field.

    In the temperature range -40C to +70C, and with an external load of 10k, the maximum output

    voltage will be less than the supply voltage by the value given in Table 1 below. All parameters,

    other than the output voltage range, remain unaffected for supply voltage (Vs

    ) changes in the

    range from 8 to 15V.

    Nominal range (T) 100 250 500 1000

    Output swing Vs- 0.5 V

    s- 0.7 V

    s- 0.8 V

    s- 1.0

    Table 1. Typical output voltage swing

    The current drain is independent of the power supply voltage and the unit will operate with supply

    voltages down to 8V. For a unit with a full scale range of 100mT the output voltage for each axisis 0.1V/mT of the field in the direction of that axis. As the output voltage swing is limited to slightly

  • 8/12/2019 Barington Three Axis Magnetometer

    10/22

    BARTINGTON INSTRUMENTS

    Page 10 of 22 OM1004/26

    less than the supply voltage, for a supply of 8V the output will operate normally with any output

    between +7.3V and -7.3V representing a field of 0.73 of the full scale value in each direction. The

    scaling factor and linearity will remain at the normal value up to this saturation point. The output

    will remain at the saturation level if the field is increased beyond this point. Asymmetric supplies

    may be used provided that the minimum and maximum voltages are not exceeded for either

    polarity.

    11.4. Signal/Power Ground

    Note: The two signal/power ground conductors are connected to a common point within the

    sensor and the power supply common (power 0 V) should be connected to only one of them.

    The other signal/power ground conductor should be used as the signal output common (0V).

    Each signal is then measured between the signal output conductor and the signal output

    common. In this way, the signal output common carries no power supply currents.

    The minimum current in the power ground conductor is approximately 19mA. On long cables, this

    will give rise to an appreciable potential difference between the power supply end and the sensor

    end of the power ground conductor. The use of separate power and signal ground conductors will

    ensure that this voltage is not included in the voltage measured between the signal output and the

    signal common.

    Figure 3displays the voltages between sensor and power supply. In order to ensure that the

    power supply return current does not affect the analogue measurements in any way, the followingprecautions should be observed:

    A. A signal common line, separate from the power common line, should be connected between the

    Mag-03 magnetic field sensor and any measurement or data acquisition system.

    B. If the signal ground line is to constitute a system ground point then a fully floating power

    supply must be employed, e.g. a pair of batteries or a fully isolated power supply. A number of

    commercially available dc to dc converters fulfil the voltage isolation requirement adequately.

    For this arrangement, only single ended analogue inputs to the data acquisition system are

    required for the three axes.

    C. If the power supply is to constitute a system ground point then the data acquisition analogue

    inputs must be of the differential type. Each differential input can then be connected between

    the remote end of the signal common line and the individual analogue outputs.

    D. The above considerations also apply if more than one Mag-03 sensor is used.

    E. Any data acquisition system analogue inputs should ideally have a very high input impedance,

    but satisfactory performance can be obtained with impedances down to 50kW. Impedances

    below this should be avoided, particularly where very long cables are used.

    F. When using the low noise unit, to obtain optimum performance, additional care should be

    exercised to avoid ground currents in the signal leads.

  • 8/12/2019 Barington Three Axis Magnetometer

    11/22

    BARTINGTON INSTRUMENTS

    Page 11 of 22 OM1004/26

    When using a compatible power supply unit from Bartington Instruments as described in Section

    7, the above requirements will be met without further consideration by the user.

    11.5. Cabling

    Note: The connecting cable to the sensor should be an eight-core screened cable. Two cores

    will be used for positive and negative power supply lines, three cores for output signals, one

    core for signal common and one for power supply ground. The screen should be connected

    to supply ground at the supply end only. The capacitance between cores should be less than

    200pF per metre. A cable with individually shielded cores should be considered for long

    cable applications.

    Note: on flying lead versions, leads are susceptible to EM interference and should be

    screened wherever possible. This may also apply to the leads between the electronics andsensor head for the Mag-03MCTP, Mag-03MCUP, Mag-03IE, Mag-03IEv1, Mag-03IEv2 and

    Mag-03IEHV sensors.

    Note: The length of the cable is limited by the voltage drop in the power supply lines and the

    capacitance between the cores. For this reason it is recommended that the cable is limited

    to a maximum length of 600 metres.

    Bartington Instruments can supply cables for connection of the sensor to a compatible power

    supply unit from Bartington Instruments, as described in Section 7. If no cable is ordered with the

    sensor then a mating connector is provided.

    11.6. Connecting Power

    Caution:Check that the polarity of the supply is correct. Using the power supply provided

    by Bartington Instruments will guarantee that the polarity is correct, as it is impossible to

    connect the wrong way round.

    The power supply should be connected to the sensor before the supply is energised, as this

    prevents high surge currents which could cause damage. Apply the positive and negative supplies

    simultaneously and avoid leaving the sensor connected to one polarity only.

    11.7. Electromagnetic Compatibility

    Except for the Mag-03MCT, the Mag-03 range of sensors are not shielded for immunity from, or

    emission of, electromagnetic fields. Any shield placed around the sensor will limit the bandwidth

    of the sensor response. The emissions generated are at a low level with a primary frequency of

    15.625kHz, this being the frequency of the energising field of the sensor. The sensor is required to

    respond to magnetic fields within the specified frequency band.

  • 8/12/2019 Barington Three Axis Magnetometer

    12/22

    BARTINGTON INSTRUMENTS

    Page 12 of 22 OM1004/26

    Note: The user should ensure that the sensor is not operated in areas where a high

    electromagnetic field exists, even if the frequency is above the bandwidth of the sensor,

    as false information may appear due to aliasing. This effect is seen in data acquisition

    systems when the frequency of sampling is lower than the frequency of the signal which is

    being sampled. It may produce apparent signals at lower frequencies than the noise, which

    may be within the frequency band of the sensor. Similarly, the user should not place the

    sensor near to any equipment which may be affected by the fields produced by the sensor

    excitation.

    12. Performance

    12.1. Frequency Response

    The typical amplitude and phase response for the Mag-03 range of sensors is shown in Figure 4.

    The sensors provide a bandwidth of 3kHz with a flat response from DC to 1kHz.

    12.2. Noise

    A typical noise plot for the standard version is shown in Figure 5and for the low noise version in

    Figure 6.

    The output signal for each axis will also contain signals at the power line frequency, other

    interference and the excitation frequency of 15.625kHz. For many measurements these

    components will be outside the response of the readout or recording system. For applications

    where low field levels or measurements of the highest resolution are required it will be necessary

    to provide a filter to select only the frequency bands of interest.

    12.3. Over Range

    Sensors are available with ranges from 70T, which corresponds to the maximum value of the

    Earths magnetic field, to 1000T. As the field in any axis approaches the full scale value, the

    output will rise in proportion until it reaches a value of approximately 1V less than the relevant

    supply line. The output will then saturate and remain at this level regardless of any further rise in

    the field. Very high fields in the hundreds of mT should be avoided as they may give rise to a few

    nT shift in offset, measured at zero field.

  • 8/12/2019 Barington Three Axis Magnetometer

    13/22

    BARTINGTON INSTRUMENTS

    Page 13 of 22 OM1004/26

    13. Signal Processing

    All Bartington Instruments power supply and data acquisition products feature a degree of

    signal processing, though the most comprehensive processing is provided by the SCU1 signal

    conditioning unit.

    For different applications, it may be necessary to process the signal from the sensor in different

    ways.

    A. In order to increase the sensitivity of the recording system, it may be necessary to back-off the

    Earths field and amplify only the changes in the field from the current value. This requires a

    high-pass filter, which could be a simple capacitively coupled arrangement, or a multi-pole

    filter to provide a steep frequency roll off characteristic. These features are all present in the

    SCU1 signal conditioning unit.

    B. To monitor small signals within the bandwidth of the sensor, it may be necessary to remove

    the higher frequency noise which is outside the band of frequencies of interest. It may also

    be necessary, when using sampling data acquisition systems, to provide an anti-alias filter to

    prevent the appearance of apparent lower frequency components in the recorded signals, due

    to undersampling of the high frequency components. The filter should be a low-pass type, with

    the top of the pass band as far below the sampling frequency as practical for the application.

    Note: Bartington Instruments data acquisition systems are already equipped with an anti-

    aliasing filter. Users of other systems should consult the relevant manufacturers.

    C. In applications such as surveillance and magnetic signature monitoring, it may be required

    to remove both the DC standing field and all AC noise and pick-up above a set frequency. The

    band of interest will be, say, 0.01 to 10Hz, and a band pass filter can be used to provide the

    required signal.

    The output from all fluxgate sensors includes a low level of the excitation frequency. For the

    Mag-03 range this noise is at 15.625kHz, which is well above the bandwidth of the sensors. It

    is particularly important to attenuate the breakthrough before digitising the signal to prevent

    aliasing. This can generate spurious signals or noise in the wanted frequency band (03kHz) byintermodulation between excitation and data sampling frequencies.

    The SCU1 signal conditioning unit provides filters with independent control of the low and high

    pass filter sections, together with offset and gain control for the output of each axis. The PSU1,

    Magmeter, Spectramag-6 and Mag-03DAM all contain a single low pass and high pass filter.

  • 8/12/2019 Barington Three Axis Magnetometer

    14/22

    BARTINGTON INSTRUMENTS

    Page 14 of 22 OM1004/26

    14. Troubleshooting, Care and Maintenance

    14.1. Troubleshooting

    Special equipment is required for the diagnosis of faults within the unit. Much of this equipment is

    beyond the scope of normal service facilities. Therefore, in the event of any apparent malfunction,

    please email [email protected] telephone the Bartington Instruments service team

    on +44 (0)1993 706565. Attempted repair or opening of the casing by users may invalidate the

    warranty.

    A re-calibration service is available which is traceable to international standards.

    14.2. Care and Maintenance

    Note: Surface or dirt contamination should be removed using a mild detergent solution only.

    If the connector pins become contaminated then they should be lightly cleaned with a swab

    of isopropyl alcohol.

    15. Storage & Transport

    Your sensor is a precision electronic instrument and should be treated as such.

    Note: Avoid exposing this intrument to shocks or continuous vibration.

    Note: Store only within the temperature range specified in the product brochure.

    Note: Do not expose this instrument to strong magnetic fields while being stored.

    16. Disposal

    This product should not be disposed of in domestic or municipal waste. For information about

    disposing of your sensor safely, check local regulations for disposal of electrical / electronic

    products.

    15.1. Waste Electrical and Electronic Equipment (WEEE) Regulations

    Bartington Instruments Mag-03 sensors comply fully with Restriction of the Use of

    Certain Hazardous Substances in Electrical and Electronic Equipment (RoHS) and WEEE

    Regulations current at the time of printing.

  • 8/12/2019 Barington Three Axis Magnetometer

    15/22

    BARTINGTON INSTRUMENTS

    Page 15 of 22 OM1004/26

    Appendix 1: Mag03MSS-TC addendum

    -TC versions of Mag-03 MSS feature a test coil which applies a magnetic field to all axes when

    activated. The corresponding changes in X, Y and Z output voltages confirm that all axes of the

    sensor are operational.

    Variations from the standard Mag-03 MSS are as follows.

    1. Connector is a 9-way type IE-XSJ-9-BCR with pin 8 used as the TEST ENABLE pin. See

    drawing DR3000available on the Bartington Instruments website.

    2. Cable features a 9-way IE-XSJ-9-CCP connector at sensor end: see drawing DR2982. A version

    can be supplied with a Hirose RM15TPD10S connector at the user end: see drawing DR2980.

    3. TEST ENABLE is activated by connecting pin 8 of IE-XSJ-9-BCR connector to PG (pin 4).

  • 8/12/2019 Barington Three Axis Magnetometer

    16/22

    BARTINGTON INSTRUMENTS

    Page 16 of 22 OM1004/26

    Appendix 2: Magnetic Units and Measurements

    A1.1. Conventions

    The Mag-03 analogue output is positive for conventional flux direction South to North in the

    direction of the arrow given for each axis. The measurement axes are designated X, Y and Z in the

    Cartesian co-ordinate system, when viewed from the top or non-connector end of the sensor.

    A1.2. Measurement Units

    Magnetic flux density = B

    Magnetic field strength = H

    SI (Systme Internationale) is the preferred system of measurement in this manual. However,

    measurements are still frequently expressed in units based on the CGS (centimetregram

    second) system. For clarity the following relationships may be useful. SI units are shown below

    with their CGS numerical (but not dimensional) equivalents.

    SI = CGS

    B Wbm-2(Weber per metre2) 104G (Gauss)

    or T (Tesla)

    H Am-1(Amperes per metre) 4px 10-3(Oe)

    The fundamental equation describing the relationship between H, B and the permeability of free

    space m0is:

    B = m0H

    It will be seen that the term 4poccurs in the CGS units shown above. The SI units, however, are

    rationalised indirectly by incorporating this term in m0. Thus in the SI system:

    m0= 4px 10-7Hm-1(Henries per metre).

    Example: For free space If H = 80 Am-1

    Then B = 4px 10-7x 80 1 x 10-4T

    Tesla is the preferred unit for flux density in the SI system. A magnetic field sensor can only be

    said to measure flux density.

  • 8/12/2019 Barington Three Axis Magnetometer

    17/22

    BARTINGTON INSTRUMENTS

    Page 17 of 22 OM1004/26

    A1.3. Conversion Table

    The most common conversion performed will be from Tesla to Gauss, and vice-versa. The

    following table may be helpful.

    SI CGS CGS SI

    1 Tesla 10 kGauss 1 kGauss 100 mTesla

    1 mT 10 G 1 G 100 mT

    1 mT 10 mG 1 mG 100 nT

    1 nT 10 mG 1 mG 100 pT

    Table 2. Conversion of SI and CGS Units

    A1.4. Vector Measurements

    Each axis produces an analogue output Va in response to flux density B in the relationship:

    Va = B cos q

    where qis the angle between the flux direction and the direction of the individual sensing element.

    The scalar value of a magnetic field may be computed from the individual X, Y and Z vector

    components, using the RSS (Root of Sum of the Squares) where:

    B = (Vx2+ Vy2+ Vz2)

    Note: there will be a small error in the result of the calculation of the total field, due to the

    small error in the orthogonality between the sensing elements. This will be particularly

    noticeable when the total field is computed from the values measured with several

    orientations of the sensor. The sensor is extremely sensitive in the measurement of small

    variations in the total field, provided that the orientation is constant, i.e. the detector is

    stationary. The sensor is therefore limited in applications requiring total field measurement

    while moving, as in a towed ferrous metal detector, by the orthogonality error within the

    specified tolerance.

  • 8/12/2019 Barington Three Axis Magnetometer

    18/22

    BARTINGTON INSTRUMENTS

    Page 18 of 22 OM1004/26

    Appendix 3: Figures

    Figure 1 Mag-03MC-BR mounting bracket

    18

    18

    40

    55

    40

    55

    All dimensions mm

    6 x screwsM5 x 25 long nylon

  • 8/12/2019 Barington Three Axis Magnetometer

    19/22

    BARTINGTON INSTRUMENTS

    Page 19 of 22 OM1004/26

    Figure 2 Mag-03 interface schematic

    Figure 3 Mag-03 electrical schematic

    47F(20V)

    I-

    0.1F

    Vx Vy Vz

    47F(20V)

    I+

    0.1F

    OUTPUT

    -12V SUPPLY

    POWERGROUND

    SIGNALGROUND

    +12V SUPPLY

    X OUT

    Y OUT

    Z OUT

    CONNECTOR

    *

    *

    * Schottky diode protects against 30V reverse polarity

    (Typical values)10R

    10R

    10R

    SENSOR

    +V

    0V

    -V

    signal common

    30mA

    19mA

    11mA

    1, 2, 3

    no current 4

    5

    7

    POWER SUPPLY(e.g. PSU)

    6 +V supply

    >50k

    supply common (0V)

    -V supply

    high impedence

    buffer

    cable resistance

  • 8/12/2019 Barington Three Axis Magnetometer

    20/22

    BARTINGTON INSTRUMENTS

    Page 20 of 22 OM1004/26

    Figure 4 Mag-03 Typical frequency dependent amplitude & phase response

    AMPLITUDE

    (V RMS)

    (FULL POWER)

    7

    6

    -3dB

    4

    3

    2

    1

    1

    1002 3 4 5 6 7 8910

    1012 3 4 5 6 7 8910

    1022 3 4 5 6 7 8910

    1032 3 4 5 6 7 8910

    FREQUENCY Hz

    MAXIMUM RIPPLE +5%

    PHASE LAG

    AMPLITUDE

    PHASE LAG

    (DEGREES)

    0

    -20

    -40

    -60

    -80

    -100

    -120

    -140

    -160

    -180

    -200

  • 8/12/2019 Barington Three Axis Magnetometer

    21/22

    BARTINGTON INSTRUMENTS

    Page 21 of 22 OM1004/26

    Figure 5 Mag-03 noise plot standard version

    Figure 6 Mag-03 noise plot low noise version

  • 8/12/2019 Barington Three Axis Magnetometer

    22/22

    B ti t I t t Li it d T 44 (0)1993 706565 Th cific ti f th r d ct d crib d i thi br ch r

    OM1004/26