Background: IMRT has become the choice of treatment for disease sites that require critical structure sparing such as head and neck cancer. It has evolved into a standard practice in many radiation therapy centers. However, due to the cost and knowledge required with the implementation of IMRT, many centers worldwide still use cobalt-60 machines or early versions of LINAC without multi-leaf collimators, and do not have the capability of providing IMRT to the needed patients. DAO enabled JO-IMRT has been proven to be effective for simpler cases such as breast and prostate. But it was suspected that Head and Neck cancer might be too complex for the JO-IMRT. It was believed that to reach a clinically acceptable JO-IMRT plan for a relatively complex case, 20 segments per beam angle was required. Objective: 1.To investigate the feasibility of DAO enabled Jaws-only IMRT (JO-IMRT) treatment planning for complex Head and Neck cancer patients. 2.To determine the proper number of segments needed for a clinically acceptable plan for JO-IMRT. Can Direct Aperture Optimized IMRT using only Conventional Jaws produce satisfactory treatment plans for Head and Neck Cancer? INTRODUCTION . METHODS AND MATERIALS CONCLUSIONS Using a commercial Treatment Planning System Prowess Panther, which utilizes Direct Aperture Optimization (DAO) algorithm, a set of randomly selected six head and neck patients previously treated at our institution were planned for JO- IMRT and MLC based IMRT (MLC-IMRT). The identical clinical acceptance criteria, identical set of beam orientations as the clinical plans were used. Number of segments per beam from 5 to 9 was experimented to try to achieve acceptable plans. For each patient, the best efforts were recorded and analyzed to verify the validity of possible clinical application. The plan acceptance criteria were established according to the RTOG-0225 protocol. Guangwei Mu, Ping Xia Department of Radiation Oncology, University of California San Francisco, San Francisco, CA, USA UCSF RESULTS With a randomly selected patient, 8 or 9 segments per beam enabled the successful reaching a clinically satisfactory JO-IMRT plan for head and neck patients within reasonable time frame. Six randomly selected HNC patients were re-planned with inverse-planned JO-IMRT and MLC- IMRT. The JO-IMRT plans were shown to be effective. Of the six HNC patients, all had at least one JOIMRT plan, with 9 segments per beam, to meets the acceptance criteria. Four acceptable MLC-IMRT plans with 9 segments per beam were achieved, with another one very close. The amount of time to reach an acceptable JO-IMRT varied widely but on average required around 15 times of iterations as for other simpler anatomic sites. The number of segments for quality plan did not have to be high. Nine segments per beam was a good choice to arrive at quality plan in most cases. Among the JO-IMRT and MLC-IMRT plans with identical number of apertures per beam angle, no significant difference was observed in the average DVHs, and the plan conformal index. The averaged plan quality evaluation metrics: JO-IMRT is feasible for head and neck tumors. The plan quality of JO-IMRT plans was comparable to that of traditional IMRT or DAO optimized MLC-IMRT plans with regards to plan conformity and uniformity as well as sensitive structure sparing. JO-IMRT plans require nearly twice numbers of MUs than the MLC IMRT plans. Nine segments per beam were sufficient to obtain acceptable jaw-only IMRT plans. (a)Exam ple ofdose distribution on the slice 1 for M LC IM R T (left)and JO IM R T (right). (b)Exam ple ofdose distribution on the slice 2 for M LC IM R T (left)and JO IM R T (right). 70 G y 59.4 Gy 54 G y 45 G y Patient #1 Patient #2 Patient #3 MLC JO MLC JO MLC JO Actual MUs per fraction 443 842 411 1623 564 1027 pGTV D 95% (cGy) 7080 7053 7087 7003 7003 7007 pCTV D 95% (cGy) 6008 5988 5955 5965 6180 6008 SpinalCord D 0.1cc (cGy) 3840 4050 5065 4470 4225 3846 Brainstem D 0.1cc (cGy) 4640 4680 4380 5390 3860 5080 LT-Parotid D mean (cGy) 2503 1791 N/A N/A 2316 1928 RT-Parotid D mean (cGy) 2520 2565 2518 2516 4471 2621 pGTV COIN 19.2% 17.3% 47.1% 38.2% 42.1% 56.4% pCTV COIN 62.2% 56.7% 65.3% 59.5% 65.5% 69.5% Uniformity 86% 87.1% 81.7% 77.8% 82.6% 83.7% DVH pG TV (JO IM R T and M LCIM RT) 0 20 40 60 80 100 0 2000 4000 6000 8000 D ose (cG y) Volum e (% ) 9beam _5seg 9beam _6seg 9beam _7seg 9beam _8seg 9beam _9seg 9beam _M LC D VH pC TV (JO IM R T and M LCIM RT) 0 20 40 60 80 100 0 2000 4000 6000 8000 D ose (cG y) V olum e (% ) 9beam _5seg 9beam _6seg 9beam _7seg 9beam _8seg 9beam _9seg 9beam _M LC D VH B rain Stem (JO IM R T and M LCIM RT) 0 20 40 60 80 100 0 1000 2000 3000 4000 5000 6000 D ose (cG y) Volum e(% ) 9beam _5seg 9beam _6seg 9beam _7seg 9beam _8seg 9beam _9seg 9beam _M LC D VH SpinalC ord (JO IM R T and M LCIM RT) 0 20 40 60 80 100 0 1000 2000 3000 4000 D ose (cG y) Volum e(% ) 9beam _5seg 9beam _6seg 9beam _7seg 9beam _8seg 9beam _9seg 9beam _M LC (a)pG TV DVHs for various num berofsegm ents per beam ,and M LC plan. (b)pC TV DVHs forvarious num berofsegm ents perbeam ,and M LC plan. (c)B rain Stem DVHs for various num berofsegm ents per beam ,and M LC plan. (d)SpinalC ord D V H s forvarious num ber ofsegm ents per beam ,and M LC plan. D VH R -Parotid (Jaw O nly) 0 20 40 60 80 100 0 1000 2000 3000 4000 5000 6000 7000 D ose (cG y) Volum e (% ) 9beam _5seg 9beam _6seg 9beam _7seg 9beam _8seg 9beam _9seg 9bean_M LC D VH L-Parotid (Jaw O nly) 0 20 40 60 80 100 0 1000 2000 3000 4000 5000 6000 7000 D ose (cG y) Volum e (% ) 9beam _5seg 9beam _6seg 9beam _7seg 9beam _8seg 9beam _9seg 9beam _M LC Figure 1. Dose distribution of a specific patient at two locations. On left is the IMRT plan delivered with MLC and on the right with conventional jaws only. Figure 2. DVHs of the targets and some key sensitive structures for the JO-IMRT plans with 9 to 5 segments per beam angle. For comparison reason, the MLC-IMRT plan with 9 segments per beam was also shown. Figure 3. Part of an actual JO- IMRT delivery sequence, the intensity map and the monitor units (MU) associated with each aperture. Figure 4. Part of an actual MLC-IMRT delivery sequence, the intensity map and the monitor units (MU) associated with each aperture. Table 1. Example of the key endpoint doses and plan quality evaluation metrics for three nasopharyngeal cancer patient planned with DAO JO-IMRT and MLC-IMRT, with 9 segments per beam angle. Plan Metric MLC- IMRT JO-IMRT COIN (pGTV) 37.70% ± 18.67% 37.95% ± 18.06% COIN (pCTV) 63.72% ± 4.44% 61.32 ± 6.39% Uniformi ty 82.82% ± 2.46% 83.55% ± 3.11% MU 453 ± 69 982 ± 333

Upload: douglas-quinn

Post on 18-Dec-2015

215 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Background:  IMRT has become the choice of treatment for disease sites that require critical structure sparing such as head and neck cancer.  It has

Background:

IMRT has become the choice of treatment for disease sites that require critical structure sparing such as head and neck cancer.

It has evolved into a standard practice in many radiation therapy centers. However, due to the cost and knowledge required with the implementation of IMRT, many centers worldwide still use cobalt-60 machines or early versions of LINAC without multi-leaf collimators, and do not have the capability of providing IMRT to the needed patients.

DAO enabled JO-IMRT has been proven to be effective for simpler cases such as breast and prostate. But it was suspected that Head and Neck cancer might be too complex for the JO-IMRT.

It was believed that to reach a clinically acceptable JO-IMRT plan for a relatively complex case, 20 segments per beam angle was required.

Objective:1. To investigate the feasibility of DAO enabled Jaws-only IMRT

(JO-IMRT) treatment planning for complex Head and Neck cancer patients.

2. To determine the proper number of segments needed for a clinically acceptable plan for JO-IMRT.

Can Direct Aperture Optimized IMRT using only Conventional Jaws produce satisfactory treatment plans for Head and Neck Cancer?

Can Direct Aperture Optimized IMRT using only Conventional Jaws produce satisfactory treatment plans for Head and Neck Cancer?

INTRODUCTION

.

METHODS AND MATERIALS

CONCLUSIONS

Using a commercial Treatment Planning System Prowess Panther, which utilizes Direct Aperture Optimization (DAO) algorithm, a set of randomly selected six head and neck patients previously treated at our institution were planned for JO-IMRT and MLC based IMRT (MLC-IMRT). The identical clinical acceptance criteria, identical set of beam orientations as the clinical plans were used.

Number of segments per beam from 5 to 9 was experimented to try to achieve acceptable plans. For each patient, the best efforts were recorded and analyzed to verify the validity of possible clinical application.

The plan acceptance criteria were established according to the RTOG-0225 protocol.

Guangwei Mu, Ping Xia

Department of Radiation Oncology, University of California San Francisco, San Francisco, CA, USA

UCSF

RESULTS

With a randomly selected patient, 8 or 9 segments per beam enabled the successful reaching a clinically satisfactory JO-IMRT plan for head and neck patients within reasonable time frame. Six randomly selected HNC patients were re-planned with inverse-planned JO-IMRT and MLC-IMRT. The JO-IMRT plans were shown to be effective. Of the six HNC patients, all had at least one JOIMRT plan, with 9 segments per beam, to meets the acceptance criteria. Four acceptable MLC-IMRT plans with 9 segments per beam were achieved, with another one very close. The amount of time to reach an acceptable JO-IMRT varied widely but on average required around 15 times of iterations as for other simpler anatomic sites. The number of segments for quality plan did not have to be high. Nine segments per beam was a good choice to arrive at quality plan in most cases. Among the JO-IMRT and MLC-IMRT plans with identical number of apertures per beam angle, no significant difference was observed in the average DVHs, and the plan conformal index. The averaged plan quality evaluation metrics:

JO-IMRT is feasible for head and neck tumors. The plan quality of JO-IMRT plans was comparable to that of traditional IMRT or DAO optimized MLC-IMRT plans with regards to plan conformity and uniformity as well as sensitive structure sparing. JO-IMRT plans require nearly twice numbers of MUs than the MLC IMRT plans. Nine segments per beam were sufficient to obtain acceptable jaw-only IMRT plans.

(a) Example of dose distribution on the slice 1 for MLCIMRT (left) and JOIMRT (right).

(b) Example of dose distribution on the slice 2 for MLCIMRT (left) and JOIMRT (right).

70 Gy

59.4 Gy

54 Gy

45 Gy

Patient #1 Patient #2 Patient #3

MLC JO MLC JO MLC JO

Actual MUs per fraction

443 842 411 1623 564 1027

pGTV D95%

(cGy)

7080 7053 7087 7003 7003 7007

pCTV D95%

(cGy)

6008 5988 5955 5965 6180 6008

SpinalCord D0.1cc (cGy)

3840 4050 5065 4470 4225 3846

Brainstem D0.1cc (cGy)

4640 4680 4380 5390 3860 5080

LT-Parotid Dmean (cGy)

2503 1791 N/A N/A 2316 1928

RT-Parotid Dmean (cGy)

2520 2565 2518 2516 4471 2621

pGTV COIN 19.2% 17.3% 47.1% 38.2% 42.1% 56.4%

pCTV COIN 62.2% 56.7% 65.3% 59.5% 65.5% 69.5%

Uniformity 86% 87.1% 81.7% 77.8% 82.6% 83.7%

DVH pGTV (JOIMRT and MLCIMRT)

0

20

40

60

80

100

0 2000 4000 6000 8000

Dose (cGy)

Vo

lum

e (%

)

9beam_5seg

9beam_6seg

9beam_7seg

9beam_8seg

9beam_9seg

9beam_MLC

DVH pCTV (JOIMRT and MLCIMRT)

0

20

40

60

80

100

0 2000 4000 6000 8000

Dose (cGy)

Vo

lum

e (%

)

9beam_5seg

9beam_6seg

9beam_7seg

9beam_8seg

9beam_9seg

9beam_MLC

DVH Brain Stem (JOIMRT and MLCIMRT)

0

20

40

60

80

100

0 1000 2000 3000 4000 5000 6000

Dose (cGy)

Vol

ume

(%)

9beam_5seg

9beam_6seg

9beam_7seg

9beam_8seg

9beam_9seg

9beam_MLC

DVH Spinal Cord (JOIMRT and MLCIMRT)

0

20

40

60

80

100

0 1000 2000 3000 4000

Dose (cGy)

Vol

ume

(%)

9beam_5seg

9beam_6seg

9beam_7seg

9beam_8seg

9beam_9seg

9beam_MLC

(a) pGTV DVHs for various number of segments per beam, and MLC plan. (b) pCTV DVHs for various number of segments per beam, and MLC plan.

(c)Brain Stem DVHs for various number of segments per beam, and MLC plan. (d)Spinal Cord DVHs for various number of segments per beam, and MLC plan.

DVH R-Parotid (Jaw Only)

0

20

40

60

80

100

0 1000 2000 3000 4000 5000 6000 7000

Dose (cGy)

Vol

ume

(%)

9beam_5seg

9beam_6seg

9beam_7seg

9beam_8seg

9beam_9seg

9bean_MLC

DVH L-Parotid (Jaw Only)

0

20

40

60

80

100

0 1000 2000 3000 4000 5000 6000 7000

Dose (cGy)

Vol

ume

(%)

9beam_5seg

9beam_6seg

9beam_7seg

9beam_8seg

9beam_9seg

9beam_MLC

Figure 1. Dose distribution of a specific patient at two locations. On left is the IMRT plan delivered with MLC and on the right with conventional jaws only.

Figure 2. DVHs of the targets and some key sensitive structures for the JO-IMRT plans with 9 to 5 segments per beam angle. For comparison reason, the MLC-IMRT plan with 9 segments per beam was also shown.

Figure 3. Part of an actual JO-IMRT delivery sequence, the intensity map and the monitor units (MU) associated with each aperture.

Figure 4. Part of an actual MLC-IMRT delivery sequence, the intensity map and the monitor units (MU) associated with each aperture.

Table 1. Example of the key endpoint doses and plan quality evaluation metrics for three nasopharyngeal cancer patient planned with DAO JO-IMRT and MLC-IMRT, with 9 segments per beam angle.

Plan Metric

MLC-IMRT

JO-IMRT

COIN (pGTV)

37.70% ± 18.67%

37.95% ± 18.06%

COIN (pCTV)

63.72% ± 4.44%

61.32 ± 6.39%

Uniformity 82.82% ± 2.46%

83.55% ± 3.11%

MU 453 ± 69 982 ± 333