dosimetry for imrt
TRANSCRIPT
Dosimetry for IMRTDosimetry for IMRT
Thomas Rockwell Mackie
Department of Medical PhysicsUniversity of Wisconsiny
Madison [email protected]
Financial Disclosure
I am a founder and Chairman of TomoTherapy Inc. (Madison, WI) which is participating in the ( , ) p p gcommercial development of helical tomotherapy.
Outline
• Dosimetry Issues Relevant to IMRT• Charged Particle Equilibrium• Temporal Non-Constancy• Partial Volume EffectsPartial Volume Effects
• Non-Standard Fields • IMRT as a Large Collection of Small Non-Standard Fields• IAEA Calibration Initiative for Non Standard Fields• IAEA Calibration Initiative for Non-Standard Fields
• Patient-Specific Methods To Verify Delivery of IMRT• Independent Calculations • Delivery of Individual Radiation Fields• Delivery of All Fields into a Phantom • Metrics for Patient-Specific Dosimetry QAMetrics for Patient Specific Dosimetry QA
Dosimetry Issues Relevant to IMRT
• Charged particle equilibrium– Different spectrum for collection of small fields
N if d– Non-uniform dose
• Temporal non-constancyA very small effect for ion chambers– A very small effect for ion chambers
– May not be true for other dosimeters
• Partial volume effect• Partial volume effect– Most important effect especially when measuring
output factors for small fieldsp
Non-Uniform Intensity Changes the Energy Spectrumgy p
More High EnergygyElectrons
Photon Photon
More Low EnergyElectronsPhoton
BeamletPhotonBeamlet
Electrons
Change in Stopping Power RatioComparison of Spencer-Attix (∆=10 keV) restricted mass collisional stopping powers ratios (water/air) at 5 cm depth in water for various 6 MV beams with stereotactic and MLC beams.
6 MV beams Beam quality, TPR(20,10)
Andreo, (1994)
Sánchez-Doblado, (2003)
Column4/Column3
El kt SL 18 0 690 1 1187 1 1188 1 000Elekta SL-1810 x 10 cm2
0.690 1.1187 1.1188 1.000
1.0 cm diameter stereo field
1.1155 0.997
0.3 cm diameter stereo field
1.1153 0.997
Siemens Primus10 10 2
0.677 1.1213 1.1221 1.00110 x 10 cm2
2 x 2 cm2 irregularon-axis beamlet
1.1203 0.999
2 x 2 cm2 irregular8 cm off-axis
1.1250 1.003
Temporal Delivery of IMRT
IMRT Delivery Signatures
Delivery of Dose to a Single Voxel
12000
Step & Shoot HIART
8000
10000
4000
6000
Sign
al (f
C)
0
2000
0.16
46.9
93.7
141
188
235
282
329
376
423
470
517
564
613
658
702
747
792
836
881
926
971
015
060
105
149
194
239
-2000
0 46 93 1 1 2 2 3 3 4 4 5 5 6 6 7 7 7 8 8 9 9 10 10 11 11 11 12
Elapsed Time (sec)
From Tim Holmes, St. Agnes Baltimore
Temporal Delivery of IMRTD li f D t Si l V l
Step and Shoot Helical Tomo
Delivery of Dose to a Single Voxel
0.80.9
1
ve D
ose
0 40.50.60.7
Cum
ulat
i
0 10.20.30.4
rmal
ized
5 %
90 % 5 %
00.1
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21
No
Elapsed Time (minutes)
From Tim Holmes, St. Agnes Baltimore
Partial Volume Effect
Dose to Water For Small Fields
0
0.8
0.9
r M t C l
0 5
0.6
0.7
ut F
acto
r Monte CarloDiamondDiodeLAC+filmFilm
0 3
0.4
0.5
Out
pu
PTW 31014 (Pinpoint)PTW 31010 (Semiflex)PTW 30006 (Farmer)
0.5 1.0 1.5 2.0 2.5 3.00.2
0.3
Side of Square Field (cm)Side of Square Field (cm)
From Roberto Capote, IAEA
Partial Volume Effect
output factors relative to 4 cm x 4 cm
1 00
1.10
0.80
0.90
1.00
r rel
to 4
x4
0.60
0.70
outp
ut fa
cto
standard dataphoton diodedi d
0.40
0.50
o diamondpinpoint0.125 cc chamberelectron diode
0 1 2 3 4 5 6 7 8 9 10equivalent field lenght (cm)
From Karen Rosser, Royal Marsden
High Uncertainty in Output FactorsE l St ti ti f 45 O t t F t f 6 d 18 fi ldExample: Statistics of 45 Output Factors for 6 mm and 18 mm square fields
(Novalis, SSD = 1000 mm, depth = 50 mm, various detectors)
Factor of Two in Beam Calibration!
From Wolfgang Ullrich, BrainLab
Reasons for Drop in Output with Small Field Size
• Backscatter into monitor unit from beam defining jaws
• Reduced scatter (phantom and head)• Reduced scatter (phantom and head)• Electronic disequilibrium• Obscuration of the source
Problems with Conventional Output Factors
• Small field openings obscure
Factors
p gthe source which is difficult to measure and error prone.
• Amount of phantom scatter changes as well as lateral disequilibriumdisequilibrium.
• Partial volume effects can mask machine outputmachine output factor.
Standard Reference Conditions
Small Field Conditions
Scan Phantom Across a Narrow BeamIf the source were a point source the• If the source were a point source, the measured primary dose would be directly proportional to the jaw field width and inversely proportional to thewidth and inversely proportional to the scan speed.
• Nearly equal phantom scatter conditions for all jaw settings.for all jaw settings.
• No partial volume effects once scanning is complete.
Beam’s eye viewJaw field width (FW)
Animation of Scanning PathFrom the Point of View of the PhantomFrom the Point of View of the Phantom
FW i the f be idthThe fan beam
Th f b
FW is the fan beam widthThe fan beam field 10 cm long by FW wide b i The fan beam
field is always scanned the
begins at one side of the field and is scanned
same distance (10 cm) at the same speed
across the phantom. FW is defined by the
Beam’s eye view
same speed. defined by the 50% to 50% dose.
Scanning at Velocity vAssume that the source is a point source
Jaw field width (FW)
Assume that the source is a point source
r( )
D r( )( )
10 cm
Animation of Scanning PathFrom the Point of View of the PhantomFrom the Point of View of the Phantom
FW i the f be idthThe fan beam
Th f b
FW is the fan beam widthThe fan beam field 10 cm long by FW wide b i The fan beam
field is always scanned the
begins at one side of the field and is scanned
same distance (10 cm) at the same speed
across the phantom. FW is defined by the
Beam’s eye view
same speed. defined by the 50% to 50% dose.
Scanning at Velocity v
Jaw field width (FW)
r( )
FWD
D r( )
Dv
D r( )
10 cm
Impact of Source Occlusion for Variable Jaw WidthsJaw Widths
Dashed line is t dWater Tank Dose Ratios
16
18
20
expected result for a point source
8
10
12
14
Dose Ra
tio
• Only the field width, FW for each se
2
4
6
8D each measurement was altered.
• The scan distance
Dos
0
0 1 2 3 4 5
Measured Field Width, FW (cm)
(10 cm) and scan velocity were identical.Measured Field Width, FW (cm)
Mainly Due to Source Obscuration
Dose and Dose/FW as a Function of Field Width (FW)Function of Field Width (FW)
Divide by Field Width and Normalize
Water Tank Dose/FW Ratios
1.2
Water Tank Dose Ratios
20
0.8
1
W Ratio
Basically a measure of the clipping the photon so rce seen
12
14
16
18
atio
ose/
FW0.2
0.4
0.6
Dose/FW photon source seen
for the smaller field widths.
2
4
6
8
10
Dose R
Dos
e
orm
aliz
ed D
o
0
0 1 2 3 4 5
Measured Field Width, FW (cm)
Plot of dose D(FW) Plot of D/FW
0
2
0 1 2 3 4 5
Measured Field Width, FW (cm)Measured Field Width, FW (cm) Measured Field Width, FW (cm)
N
Plot of dose D(FW) Plot of D/FW
From Brian Hundertmark
No Difference Between Water and Solid Water
From Brian Hundertmark
Little Difference With Depth
From Brian Hundertmark
Small Vs. Farmer Chambers
A1SL: 0 057 ccA1SL: 0.057 ccPTW: 0.6 cc
From Brian Hundertmark
Helical Delivery
From Brian Hundertmark
The Same Technique Could Be Applied for Other Small BeamsOther Small Beams
Now for a point source, the measured primary dose would be proportional to the area of the beam andwould be proportional to the area of the beam and inversely proportional to the scan velocity.
You wouldA circular beam is scanned in a You would
actually scan the phantom instead of
raster pattern to create a broad beam using the
Beam’s eye view
instead of scanning the beam.
beam using the same overall irradiation time.
Avoids Problems with Measuring Output Factors forMeasuring Output Factors for
Small Fields
• Charged particle equilibriumg p q– Nearly same charged particle spectrum
• Temporal non-constancyTemporal non constancy– Deliberately uses temporal non-constancy– Mimics how composite small fields become– Mimics how composite small fields become
larger irradiated fields
• Partial volume effect• Partial volume effect– No partial volume effects
Measurements Would Benchmark Source Model
Model the source distribution
Calculate a delivery to a large area using a
superposition of small bbeams
Measure the dose in the center of the fieldcenter of the field
Repeat for another set of ll b til fsmall beams until a range of beam sizes are done.
Agreement
no
Enough? between calculation and measurement?yes yes
Doneno
0.05 0.30 0.55 0.80
IMRT Is Far from Reference Conditions0.05 0.30 0.55 0.80
0.10 0.35 0.60 0.85
0.15 0.40 0.65 0.90
0.20 0.45 0.70 0.95
0.25 0.50 0.75 1.00
From Art Boyer, Stanford
Audit for TRS 398 Reference Dosimetry
10
468
on (%
) Farmer Semiflex PinPoint
024
Dev
iatio
-6-4-2
ativ
e D
-10-86
Rel
CASE NUMBER1 2 3 4 5 6 7
CASE NUMBER
From Roberto Capote, IAEA
IMRT Audit
810
Farmer Semiflex PinPoint
%)
2468
DynamicStep-and-shoot
atio
n (%
-202
e de
via
-6-42
Rel
ativ
e
1 2 3 4 5 6 7 8 9 10 11 12 13
-10-8R
CASE NUMBER1 2 3 4 5 6 7 8 9 10 11 12 13
Sánchez-Doblado F, Hartmann G., Pena J., Capote R. et alIJROBP 68 (2007) 301-310
IMRT Is About Using Small FieldsIMRT Is About Using Small FieldsIntensity pattern
possibly unconstrained p yintensity levels
What is delivered
Intensity limited to a few discrete intensity levelsy
Adapted from Michael Sharpe, U. of Toronto
Leaf Penumbra is Important
Lines DefiningOne Half Value Layer of Beam
Attenuation
Lines DefiningLight Field Boundary
CentralAxis
Boundary
• It is important to have an accurate model of the curved leaf endsthe curved leaf ends.
• More important for summation of small fields.
Meeting for the Dosimetry Code of Practice:Meeting for the Dosimetry Code of Practice:Small Fields and Novel Beams 3-5/12/07
• Jan Seuntjens
IAEA• Pedro Andreo
Outside Participantsj
• Hugo Palmans• Karen Rosser
• Ken R. Shortt• Stanislav VatnitskiyKaren Rosser
• Saiful Huq• Wolfgang Ullrich
y• Roberto Capote• Joanna Izewska• Wolfgang Ullrich
• Warren Kilby• Rock Mackie
Joa a e s a• Ahmed Meghzifene• Rodolfo Alfonso• Rock Mackie Rodolfo Alfonso
• My Pictures\IAEA_2008\IAEA_2008 006.jpgjpg
Examples of Small and Novel Fields
• GammaKnife (1 8 cm diameter collimator is the• GammaKnife (1.8 cm diameter collimator is the largest collimator – intrinsically composite field –not 100 cm SSD)
• Linac SRS beams (extrapolate to small field conditions)
(6 di lli i h l• Accuray (6 cm diameter collimator is the largest collimator)
• TomoTherapy (5 cm is the largest slice width• TomoTherapy (5 cm is the largest slice width –no field flattening filter)
• IMRT (made up of numerous small fields)IMRT (made up of numerous small fields)
Static Field CalibrationUses a machine-specific reference field, fmsr
msr
msr
Q Cow D ,w Q QD M N k k 60
msrQ( D / M )k
msrQDfmsr
msr
wQ Q
w
( D / M )k
( D / M )wD
Corrects for the differences between the conditions of field size, geometry, and beam quality of the conventional reference field g y, q yand the machine-specific reference field.
Calculate Using MCUsing method of Sempau et al 2004 PMB 49;4427-44
msrQw aD D
k( / )
Using method of Sempau et al 2004 PMB 49;4427-44
msrQ Qw a
kD D( / )
is the dose at the reference point in water
D
wD
is the average dose in the ion chamberaDaD
DwD
Adapted from Edmond Sterpin
Composite Field CalibrationpUses a plan-class specific reference field, fpcsr
pcsr
pcsr
Q Cow D ,w Q QD M N k k 60
pcsrQ( D / M ) fpcsr
pcsr
wQ Q
w
( D / M )k( D / M )w
Corrects for the differences between the conditions of field size, t d b lit f th ti l f fi ldgeometry, and beam quality of the conventional reference field
and the plan-class specific reference field.
Dose Calibration Correction Factor to G t D i Cli i l Fi ldGet Dose in a Clinical Field
M clinclin msr clin msr
msr clin
QQ Q Q Qw w Q w Q
Q
MD D D k
MmsrQM
clinQ( D / M )
clin msr
wQ Q
w
( D / M )k
( D / M )
In principle, the dose calibration correction factor would be different for each clinical measurement. Eventually, Monte Carlo treatment planning systems could compute for both the patient and a QA phantom.
clinQk
Overview of Static Field Dosimetry
REFERENCE DOSIMETRY
msr
msr
Q Cow D ,w Q QD M N k k 60 clin m sr clin
m sr
Q Q Qw w QD D
10 cm x 10 cm Reference Field
msrQk
Linac Stereotactic Field
CoD ,w QN k 60
BrainLab MiniMLC (10 cm x 10 cm)
Hypothetical 10 cm x 10 cmReference Field clin
m sr
CyberKnife (6 cm Diameter)
Qk
Reference Field
msrQ
TomoTherapy(5 cm x 10 cm )
Ion Chamber =
“Effective kQ” for Tomo
1. Determine the %dd(10)x[HT Ref] for a tomotherapy unit for a 5 cm x 10 ( ) [ ] pycm field at 85 cm SSD.
2. Using the graph at the right look up the value of %dd(10)x[HT TG-51]. 3. Using the graph on the left and the derived value %dd(10)x[HT TG-51]
determine the abscissa and this effective kQ value is then called in the IAEA protocol.
msrQ Qk k
Thomas et al (2005)
Static and Composite Field C l l ti f TCalculations for Tomo
Static Field Calibration Composite Field CalibrationStatic Field Calibration(Section III.A.1)
Composite Field Calibration(Section III.A.2)
QkkJeraj et al 2005 Duane et al 2006
5 cm x 10 cm 0.997 Unmodulated Helical Delivery5 cm Slice Width
1.000pcsrQk
msrQk
2 cm x 10 cm 0.993 Unmodulated Helical Delivery2.5 cm Slice Width
1.000
2 cm x 2 cm 0.990 Unmodulated Helical Delivery 0.9971 cm Slice Width
ICRU Committee on IMRTSponsors
• André Wambersie MD, PhD, Brussels, Belgium• Paul DeLuca PhD, Madison, USA
R i h d G hb MD Ph D Cl l d USA• Reinhard Gahbaue MD, Ph.D, Cleveland, USA• Gordon Whitmore Ph.D, Toronto, Canada
ChairmenVi t G é i MD PhD B l B l i• Vincent Grégoire MD PhD, Brussels, Belgium
• T. Rock Mackie PhD, Madison, USA
MembersWilfried De Neve MD PhD Gent Belgium• Wilfried De Neve MD PhD, Gent, Belgium
• Mary Gospodarowicz MD, Toronto, Canada• Andrzej Niemierko PhD, Boston, USA• James A. Purdy PhD, Davis, USAy• Marcel van Herk PhD, Amsterdam, The Netherla• Nilendu Gupta PhD, Colombus, USA
Consultants• Anders Ahnesjö PhD, Uppsala, Sweden• Michael Goiten PhD, Windisch, Switzerland
Supplement to ICRU-50 and ICRU-62
• More emphasis on statistics.
• Prescription and reporting with dose-volume specifications.p
• No longer use ICRU-Reference Point.
• Want median dose D50% reported.
• Use model-based dose calculations.
• Include the effect of tissue heterogeneities.
R t d t ll f t t d• Report dose to small mass of water, not dose to tissue.
QA for IMRTQ f S• Appropriate QA of TPS and delivery
equipment
• Patient-specific QA:• Delivery of individual fields into a• Delivery of individual fields into a
dosimeter
D li f ll f th fi ld i t h t• Delivery of all of the fields into a phantom
• Independent dose calculation algorithms with similar of better dose calculation accuracy
• In-vivo dosimetry not limited to a single point.
G DGap Error
Gap error Dose error20.0
15.0
rror
Range of gap width
10.0
Dos
e er
2.0
Gap error (mm)
5.0% D 1.0
0.50.2
0.00 1 2 3 4 5
Nominal gap (cm) g p ( )
From Tom Losasso, Memorial Sloan Kettering
Errors Introduced
Leaf Positioning Errors1 mm Bands
- 0.5 mm
- 0.2 mm
+ 0.2 mm
+ 0.5 mm
Mechanical Alignment of gStatic MLC Approach
No offset 0.6 mm offset 1.0 mm offset
From Jatinder Palta, University of Florida
Radiation Field Edge Offset
0.7mm offset 0.6mm offset
0.3mm offset0.5mm offset
Segmental Multileaf Collimator (SMLC) D li S t(SMLC) Delivery System
T l A tiTolerance Limit
Action Level
MLC*Leaf position accuracyLeaf position reproducibilityGap width reproducibility
1 mm0.2 mm0 2 mm
2 mm0.5 mm0 5 mmGap width reproducibility 0.2 mm 0.5 mm
Gantry, MLC, and Table 0.75 mm di
1.00 mm di
* Measured at all four cardinal gantry angles
Isocenter radius radius
Beam StabilityLow MU Output (<2MU) 2% 3%Low MU Output (<2MU)Low MU Symmetry (<2MU)
2%2%
3%3%
Jatinder Palta, AAPM Summer School 2003
Dynamic Multileaf Collimator (DMLC) Delivery System(DMLC) Delivery System
Tolerance Li it
Action L lLimit Level
MLCLeaf position accuracy 0 5 1Leaf position accuracyLeaf position reproducibilityGap width reproducibility
0.5 mm0.2 mm0.2 mm
1 mm0.5 mm0.5 mm
Leaf speed 0.1 mm/s 0.2 mm/s
Gantry, MLC, and Table Isocenter
0.75 mm radius
1.00 mm radiusIsocenter radius radius
Beam StabilityLow MU Output (<2MU) 3% 5%p ( )Low MU Symmetry (<2MU)
3%2%
5%3%
Jatinder Palta, AAPM Summer School 2003
Fail FailFail Pass
Leaf open time histograms (15 sec gantry)Higher Avg. Leaf
Opening Time
Leaf UsageLeaf opening time
Leaf #
TomoTherapy Leaf Latencypy y
• TomoTherapy uses pylinear fit of measured data to model leaf latencmodel leaf latency
• Based on small lead opening times lead opening times we expect leaf latency as the cause
replannedoriginal plan
pitch = 0.143 pitch = 0.287
patient 1
PTV 70PTV 64PTV 54PTV 50larynxbrainstemr. parotid
pitch = 0.287pitch = 0.215
patient 2
PTV 54esophagusp glungscordl. ventricle
patient 3
pitch = 0.287pitch = 0.215PTV 60PTV 50esophaguslarynxcordr. parotidb i tbrainstem
End-to-End QA
QA Measurements
Measure “Cheese”Phantom
d f Q
plane and point dose at the sameused for QA
measurements
at the same time
Film PlanePhantom can be rotatedPhantom can be rotated or turned to acquire any
orthogonal plane
On and Off-Axis Results
Delivered Dose: 2 5cm Treatment Beam
Film and Ion Chamber Absolute Dose
Delivered Dose: 2.5cm Treatment Beam
2.5
On-Axis Tumor Off-Axis Tumor
1.5
2.0
Gy)
1.0Dos
e (G
0.0
0.5
-20 -15 -10 -5 0 5 10 15 20
Distance (cm)
QA for AllQA for All of the Fields
TomotherapypyExample
Delivery QA Panel
Delivery QA Panel
Comparison of Phantom Plan and Verification Film
Plan
FilmN t thNote theHighGradients
Film
Plan
From Chet Ramsey, Thompson Cancer Survival Center
QA of Individual FieldsQ
External diode/ion-chamber arrays• MapCheck• PTW Octavius phantom• IBA Matrix• IBA Matrix
Integrated detector systems• EPID portal dosimetryp y
Independent Calculation
Dose Accuracy and Distance to AgreementAgreement
Dose accuracy for low gradients Dose
m c m c 50r r D r D r D %( , ) ( ) ( ) /
Dose
Low gradients < 20%/cmMeasured
m c m cr r r r r( , )
Distance to agreement for high gradients
ComputedComputedRadius
Gamma Function2 2 1 2
m c m c M m c Mr r r r D r r r d /( , ) {[ ( , ) / ] [ ( , ) / ] }
MDDose accuracy axis
m c m c M m c M
m cr r( , )
M
Passm c
Distance to agreement
m cr r r( , )
d
agreement axis
Md
Gamma Function2 2 1 2
m c m c M m c Mr r r r D r r r d /( , ) {[ ( , ) / ] [ ( , ) / ] } m c m c M m c M
MDDose accuracy axis
m cr r( , )
M
Failm c
Distance to agreement
m cr r r( , )
d
agreement axis
Md
IMRT Evaluation using Anthropomorphic PhantomsAnthropomorphic Phantoms
For H&N, using a criteria of 5% or 4mm, the i t d f 75% t 58%Molineu et al IJROBP 2005
Ibott et al Tech in Ca RT 2006Followill et al Med Phys 2007
Courtesy Ibott, RPC
passing rate drops from 75% to 58%
Gortec IMRT Test PhantomTLDs are placed at seven locations.
Point 1: Isocenter
P i t 2 S i l d i t Point 2: Spinal cord isocenter
Point 3: Spinal cord cranial
Point 4: PTV T R
Point 5: PTV T R cranial
Point 6: PTV N L
Point 7: PTV N L caudalCourtesy M. Tomsej,Brussels
Audit Results
Dm/Dc=f(CENTER) per meas. pt
1.15
1.2
1.05
1.1
isocenter
spinal cord iso
i l d i l
0.95
1
Dm/D
c spinal cord cranial
PTV T D
PTV T D cranial
PTV N G
PTV N G caudal
0.85
0.9
Sample Result
0.80 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20
CENTER
Sample Tomotherapy Results
Inter-Institution Dose AccuracyAccuracy Distribution
600
300400
500
uenc
y
0100
200Freq
00.885 0.905 0.925 0.945 0.965 0.985 1.005 1.025 1.045 1.065 1.085 1.105 1.125
Measured Dose/Computed Dose
Number of Measurements = 2679Mean = 0.995Standard Deviation = 0 025Standard Deviation = 0.025
(Updated from Zefkili et al 2004)
Intra-Institution Dose AccuracyAccuracy Distribution
300
350
200
250
300
uenc
y
50
100
150
Freq
0-10 -8 -6 -4 -2 0 2 4 6 8 10
Relative Difference Between Measured and Calculated Dose (%)
Number of Measurements = 1591Mean = 0.45%Standard Deviation = 2.5%
(Updated from Dong et al 2003)
QA Accuracy for IMRTy• Previous ICRU 5% point-dose accuracy
ifi ti l d b l t i dspecification replaced by a volumetric dose accuracy specification.
• Proposed new ICRU volumetric dose accuracy specification:
- High gradient (≥ 20%/cm): 85% of points within 5 mm (3.5 mm SD),
- Low gradient (< 20%/cm): 85% of points within 5% of predicted dose normalized to the prescribed dose (3.5% SD).
Monitor Units Calculations for M d l B d D C l l tiModel-Based Dose Calculation
D DA r A r A r b AMU MU
10
0
( , ) ( , ) ( , ) (1 ( ))
0
ComputedDirectly
Beam Output
Correction toAccount forBackscatterBackscatterInto MonitorChamberP
d A
Monitor Units Calculations for M d l B d D C l l tiModel-Based Dose Calculation
D cal cal
Measuredcal
D A dM b A
MU D0
( , )(1 ( ))
dcal Acal cal
Calculated
U D A d0
( , ) Ref
Not including the effect of backscatter into the it h b ill lt i b t 2% tmonitor chamber will result in about a 2% error at
worst.
Backscatter into Monitor ChamberThe effect is due to backscattered photonsThe effect is due to backscattered photons entering the monitor and resulting in feedback to the linac to lower its output
Varian 2100 – 10 MV. Results with other jaw completely open
Liu et al., Med. Phys 2000;27:737-744
Monitor Backscatter for Square On Axis FieldsOn-Axis FieldsVarian 2100 – 10 MV
Liu et al., Med. Phys 2000;27:737-744
Conclusion• IMRT uses complex field boundaries and many small
fields to achieve more conformal dose distributions. • Partial volume effects can result in severe error
especially when measuring the dose in high gradients.• A new method for determining the effect of field
obscuring is proposed to eliminate partial volume g p p peffects.
• IMRT deliveries are far from the measurement conditions of calibration.
• IAEA has developed a formalism to account for small and novel beams.
• Independent calculations or measurements for patient-Independent calculations or measurements for patientspecific QA are required for IMRT.
• IAEA will recommend that 85% of measurement values (dose or distance to agreement) are within 5%.(dose or distance to agreement) are within 5%.
• Model-based monitor unit calculations are simple but should include scatter into monitor chamber.