iERP: 126103D5617/6 [English]© BW Technologies 2008. All rights reserved.
Automatic Test and Calibration Station
User Manual
Limited Warranty and Limitation LiabilityBW Technologies LP (BW) warrants the product to be free from defects in material and workmanship under normal use and service for a period of two years, beginning on the date of shipment to the buyer. This warranty extends only to the sale of new and unused products to the original buyer. BW’s warranty obligation is limited, at BW’s option, to refund of the purchase price, repair or replacement of a defective product that is returned to a BW authorized service center within the warranty period. In no event shall BW’s liability hereunder exceed the purchase price actually paid by the buyer for the Product. This warranty does not include:
a) fuses, disposable batteries or the routine replacement of parts due to the normal wear and teat of the product arising from use;b) any product which in BW’s opinion, has been misused, altered, neglected or damaged, by accident or abnormal conditions of operation, handling or use;c) any damage or defects attributable to repair of the product by any person other than an authorized dealer, or the installation of unapproved parts on the product; or
The obligations set forth in this warranty are conditional on:a) property storage, installation, calibration, use, maintenance and compliance with the product manual instructions and any other applicable recommendations of BW;b) the buyer promptly notifying BW of any defect and, if required, promptly making the product available for correction. No goods shall be returned to BW until receipt by the buyer of
shipping instructions from BW; andc) the right of BW to require that the buyer provide proof of purchase such as the original invoice, bill of sale or packing slip to establish that the product is within the warranty period.
THE BUYER AGREES THAT THIS WARRANTY IS THE BUYER’S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. BW SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL, OR BASED ON CONTRACT, TORT OR RELIANCE OR ANY OTHER THEORY.Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or limitation of incidental or consequential damages, the limitations and exclusions of this warranty may not apply to every buyer. If any provision of this warranty is held invalid or unenforceable by a court of competent jurisdiction, such holding will not affect the validity or enforceability of any other provision.
BW Technologies by Honeywell BW Technologies by Honeywell BW Technologies by HoneywellCorporate Headquarters America Europe2840 - 2nd Ave. SE 3279 West Pioneer Parkway 5 Canada CloseCalgary, AB Arlington, TX Banbury, OxfordshireCanada T2A 7X9 USA 76013 United Kingdom OX16 2RT
Table of ContentsPage
..................................................... 1.................................................... 2..................................................... 2..................................................... 5.................................................... 6.................................................... 7..................................................... 8..................................................... 8..................................................... 8..................................................... 9................................................... 10................................................... 11.................................................. 12................................................... 14................................................... 16................................................... 19................................................... 21................................................... 23................................................... 27................................................... 27................................................... 28.................................................. 30................................................... 33
i
Table
Introduction ....................................................................................................................Contacting BW Technologies by Honeywell ................................................................Safety Information - Read First.....................................................................................Getting Started ...............................................................................................................MicroDock II Base Station and Docking Modules........................................................
Parts of the MicroDock II Base Station and Docking Module ......................................Display Elements ...........................................................................................................Pushbuttons ...................................................................................................................
Docking Module Pushbuttons .....................................................................................Base Station Pushbuttons ..........................................................................................
Installation ......................................................................................................................Battery Installation ......................................................................................................Inserting the GasAlertClip Extreme and the GasAlert Extreme ...................................Inserting the GasAlertMicro ........................................................................................Inserting the GasAlertMicro 5/PID/IR..........................................................................Inserting the GasAlertMicroClip ..................................................................................Inserting the GasAlertMax XT.....................................................................................
Adding Another Docking Module .................................................................................Initializing the New Docking Module ...........................................................................Defining Pump Speed.................................................................................................
Mounting the MicroDock II ............................................................................................Changing Dip Switch Settings (GasAlertClip Extreme only) ......................................Using the Base Station..................................................................................................
MicroDoUser Man
Title Page
Confir ................................................... 33Activa ................................................... 33Self-T ................................................... 33Deact ................................................... 34
User Op ................................................... 35Time/ ................................................... 35Inlet S ................................................... 36
Ga ................................................... 37Ga ................................................... 39Ga ................................................... 40
Pump ................................................... 41Contr ................................................... 41Backl ................................................... 42About ................................................... 42Forma ................................................... 43Inlet S ................................................... 44Pass ................................................... 45
En .................................................. 46Langu ................................................... 46
Results ................................................... 47Reconfig ................................................... 48Gas Con ................................................... 50
Abort ................................................... 51Confli ................................................... 51
Ba ................................................... 52Us ................................................... 52
Not F ................................................... 53
ck IIual
ii
ming Inlet Setup ................................................................................................ting the Base Station .........................................................................................est......................................................................................................................ivating the Base Station.....................................................................................tions Menu ........................................................................................................Date ...................................................................................................................etup...................................................................................................................s Type ...............................................................................................................s Concentration Level .......................................................................................s Cylinder Lot # Field ........................................................................................ Setup ................................................................................................................ast ......................................................................................................................ight ................................................................................................................................................................................................................................................tting the MMC/SD Card ....................................................................................elect ..................................................................................................................
Code ..................................................................................................................tering User Options When Pass Code Protected ...............................................age ....................................................................................................................
History...............................................................................................................uring the Detector (not applicable to the GasAlertClip Extreme)...............flicts ..................................................................................................................Option................................................................................................................cts ......................................................................................................................se Station Selects the Gas Inlet ........................................................................er Selects Gas Inlet ...........................................................................................ound ...................................................................................................................
MicroDock IIUser Manual
Title Page
Accessin ................................................... 54Order of .................................................. 55Bump Te ................................................... 58
Bump ................................................... 60Bu ................................................... 60Bu ................................................... 60Inc ................................................... 61Se ................................................... 61
Calibrati ................................................... 62Guide ................................................... 62Calibr .................................................. 63
Ga ................................................... 64Calibr ................................................... 65
Ca ................................................... 65Ca ................................................... 65Inc ................................................... 66Se ................................................... 66
Data Tra .................................................. 67Trans ................................................... 67Autom ................................................... 68
Us ................................................... 68Us ................................................... 68
Base Sta ................................................... 70Event ................................................... 70Insert ................................................... 70Acces ................................................... 71
PC ................................................... 71
iiig the MMC/SD Card During a Test ................................................................ Gases Applied for Bump Tests and Calibrations..........................................st....................................................................................................................... Test Results......................................................................................................mp Pass ............................................................................................................mp Fail...............................................................................................................orrect Inlet Setup...............................................................................................nsor Disabled ....................................................................................................on ......................................................................................................................lines ...................................................................................................................ation Procedure (All models excluding the GasAlertClip Extreme).....................sAlertMicro 5 IR with CO2 Sensor.....................................................................ation Results ......................................................................................................libration Pass.....................................................................................................libration Fail .......................................................................................................orrect Inlet Setup...............................................................................................nsor Disabled ....................................................................................................nsfer (GasAlert Extreme, GasAlertMicroClip, and GasAlertMax XT only)...ferring Datalogs .................................................................................................atic Datalog Download (GasAlertMicroClip and GasAlertMax XT only) ...........ing the IR Link ...................................................................................................ing the MicroDock II...........................................................................................tion MMC/SD Card........................................................................................... Logging .............................................................................................................ing/Replacing a MMC/SD Card..........................................................................sing Test Results............................................................................................... Requirements ..................................................................................................
MicroDoUser Man
Title Page
Charging e) .............................................. 71Charg ................................................... 71Charg ................................................... 72Charg ................................................... 72
Maintena ................................................... 73Batter ................................................... 73
Ga ................................................... 73Ga ................................................... 73
Troubles ................................................... 74Replacem ................................................... 79Specifica ................................................... 80
Charg ................................................... 81
ck IIual
iv
the Battery Pack (Excluding GasAlert Extreme and GasAlertClip Extremer/Battery Pack Guidelines................................................................................ing Different Detectors Together .......................................................................ing Procedure....................................................................................................nce ...................................................................................................................y Pack Storage .................................................................................................sAlertMicro and GasAlertMicro 5/PID/IR...........................................................sAlertMicroClip ..................................................................................................hooting.............................................................................................................ent Parts and Accessories ...........................................................................
tions.................................................................................................................er Specifications................................................................................................
List of TablesPage
..................................................... 4
..................................................... 7
..................................................... 8
..................................................... 8
..................................................... 9
................................................... 10
................................................... 12
................................................... 14
................................................... 17
................................................... 19
................................................... 21
................................................... 23
................................................... 23.................................................. 29................................................... 30................................................... 30................................................... 37................................................... 56................................................... 57................................................... 72................................................... 74................................................... 79
v
Title Figure1. International Symbols.......................................................................................2. The MicroDock II and Docking Module............................................................3. Display Elements ............................................................................................4. Docking Module Pushbuttons..........................................................................5. Base Station Pushbuttons ...............................................................................6. Base Station Connections ...............................................................................7. Inserting the GasAlertClip Extreme and the GasAlert Extreme.......................8. Inserting the GasAlertMicro.............................................................................9. Inserting the GasAlertMicro 5/PID/IR ..............................................................10. Inserting the GasAlertMicroClip.......................................................................11. Inserting the GasAlertMax XT .........................................................................12. Torque Specifications......................................................................................13. Adding Another Docking Module.....................................................................14. Mounting the MicroDock II................................................................................15. Dip Switch Settings .........................................................................................16. Changing Dip Switch Settings .........................................................................17. Inlets 2-5 – Available Gas Types.....................................................................18. Gas Type Application Table ............................................................................19. Gas Application Rules .....................................................................................20. Charger Status LED ........................................................................................21. Troubleshooting Tips ......................................................................................22. Replacement Parts and Accessories ..............................................................
List of FiguresPage
..................................................... 6
..................................................... 7
..................................................... 9
................................................... 10
................................................... 12
................................................... 14
................................................... 16
................................................... 17
................................................... 19
................................................... 21
................................................... 24
................................................... 25
................................................... 26
................................................... 28
................................................... 29
................................................... 31
................................................... 32.................................................. 43................................................... 68................................................... 69................................................... 69................................................... 70
vii
Figure Title
1. MicroDock II Base Station and Docking Modules............................................2. The MicroDock II Docking Module...................................................................3. Base Station Pushbuttons ...............................................................................4. Base Station Connections ...............................................................................5. Inserting the GasAlertClip Extreme and the GasAlert Extreme.......................6. Inserting the GasAlertMicro.............................................................................7. Inserting the GasAlertMicro 5/PID/IR ..............................................................8. Removing the Diffusion Adapter......................................................................9. Inserting the GasAlertMicroClip.......................................................................10. Inserting the GasAlertMax XT .........................................................................11. Adding Another Docking Module (Front View) ................................................12. Adding Another Docking Module (Back View).................................................13. Attaching Back Cover Plate (Back View).........................................................14. Attaching the Wall Mounting Plate ..................................................................15. Parts of the Wall Mounting Plate .....................................................................16. Detaching the GasAlertClip Extreme Module ..................................................17. Changing Dip Switch Settings .........................................................................18. Inserting the MMC/SD Card into the Base Station ...........................................19. Load Configuration File Dialog Box.................................................................20. Save Configuration Dialog Box .......................................................................21. Save to MicroDock(s) Dialog Box....................................................................22. Replacing a MMC/SD Card .............................................................................
1
MicroDock II
ides data transfer for
,, andetectors.
s can be connected to the station.
Note
of six charging modules.
four non-charging modules can be connected arging the Battery Pack (Excluding GasAlert Extreme).
Introductiona Warning
To ensure personal safety, read the Safety Information before using the MicroDock II Base Station.
The MicroDock II Automatic Test and Calibration Station (“the base sta-tion”) provides automated calibration and bump testing for
• GasAlertClip Extreme,• GasAlert Extreme,• GasAlertMicro,• GasAlertMicro 5,• GasAlertMicro 5 PID,• GasAlertMicro 5 IR, • GasAlertMicroClip, and• GasAlertMax XT detectors.
The base station also prov
• GasAlert Extreme• GasAlertMicroClip• GasAlertMax XT d
A maximum of 10 module
There is a maximum
Six charging modules plusto the station. Refer to ChExtreme and GasAlertClip
MicroDUser M
2
ContTo cont
USA: 1-CanadaEuropeOther c
Address
BW Tec2840 – CalgaryCANAD
Email uVisit BW
ISO 900
n - Read Firstas specified in this manual.
he base station and in this manual are
nts on the following pages before using the
ec Caution
s batteries. Do not mix with the solid waste should be disposed of by a qualified aterials handler.
ock IIanual
acting BW Technologies by Honeywellact BW Technologies by Honeywell, call:
888-749-8878 : 1-800-663-4164 : +44 (0) 1295 700300 ountries:+1-403-248-9226
correspondence to:
hnologies by Honeywell 2 Avenue S.E. , AB T2A 7X9 A
s at: [email protected] Technologies by Honeywell’s website at: www.gasmonitors.com
1
Safety InformatioUse the base station only
International symbols on texplained in Table 1.
Read the Caution statemebase station.
This instrument containstream. Spent batteriesrecycler or hazardous m
3
MicroDock IISafety Information - Read First
oneywell immediately.not use the base station in a hazard-
st be attached to a venting system or
ackground gas.
cal shock.viced by non-BW Technologies per-
ersywell immediately.rger in a hazardous location. Failure
arger. Failure to do so can result in
a CautionsMicroDock ll base station
• If the base station is damaged or parts are missing, contact BW Technologies by H• The base station must be used only in an area that is free of background gas. Do
ous area. Failure to adhere to this caution can lead to fire and/or explosion.• This equipment uses potentially harmful gas for calibrations. The base station mu
be used in a well-ventilated area.• Perform calibrations and bump checks only in a clean atmosphere that is free of b• The maximum recommended exhaust line length is 50 ft. (15.24 m).• Ensure that the inlet filter is clean.• Ensure that all gas cylinders contain enough gas.• Ensure the exhaust line is not connected to a negative pressure system.• A demand flow regulator must be used with all gas cylinder connections.• Do not expose the base station to electrical shock or severe continuous mechani• The base station warranty will be void if the unit is disassembled, adjusted, or ser
sonnel.• Do not immerse the base station in liquids.
GasAlertMicro, GasAlertMicro 5/PID/IR, GasAlertMicroClip, and GasAlertMax XT Charg• If the charger is damaged or parts are missing, contact BW Technologies by Hone• Do not change or charge the batteries in a hazardous location. Do not use the cha
to adhere to these precautions can lead to fire and/or explosion.• Read and adhere to all instructions and precautions that are provided with the ch
fire, electric shock, or other personal injury and/or property damage.
4
MicroDUser M
es with this charger.
ock.
en cleaning or performing mainte-
ge.
are provided to perform a procedure, gies replacement parts. Refer to
e the charger during repair attempts. ranty.
ock IIanual
a Cautions
Table 1.International Symbols
• Use only BW approved batteries; do not use alkaline or other rechargeable batteri• For indoor use only.• Do not immerse the charger in liquids.• Do not expose the charger to electrical shock or severe continuous mechanical sh• Ensure the detector battery pack is locked in place before operating the detector.• To eliminate the risk of electrical shock, disconnect and deactivate the charger wh
nance.• Avoid touching the charger and detector contact pins.• Substitution of components may impair Intrinsic Safety of the detector under char• Do not charge the battery pack with any other charger.• Do not attempt to disassemble, adjust, or service the charger unless instructions
or a part is listed as a replacement part in the user manual. Use only BW TechnoloReplacement Parts and Accessories.
• The charger warranty will be void if a customer, personnel, or third parties damagAny non-BW Technologies by Honeywell service/repair attempts will void this war
Symbol Description
n Approved to both U.S. and Canadian Standards by CSA International
X Conforms to European Union Directives
5
MicroDock IIGetting Started
GettiConfirmthe baspurchas
••••
••••
••
Note
k II base station is shipped with two inlets. alibration gas inlets can be included if before purchasing.
s, refer to Replacement Parts and Accesso-
the operations and functions of the base sta-figures and tables.
ck II Base Station and Docking Modulese 2. The MicroDock II and Docking Module e station)lements e station LCD icons)odule Pushbuttons
e 5. Base Station Pushbuttonse 6. Base Station Connections
ng Started that the items listed below are included with the base station. If e station is damaged or parts are missing, contact the place of e immediately.
Batteries (four replaceable C-cell alkaline batteries)128 MB Delkin MultiMediaCard (MMC) insertedUSB cableTwo to four calibration gas hoses (depending upon order at the time of purchase) with quick connect fittingsInlet filter assemblyPower adapterCharger adapter (charger models only)CD: MicroDock II User Manual and MicroDock II Quick Refer-ence Guide translationsMicroDock II Quick Reference GuideFleet Manager II CD ROM
A standard MicroDocA maximum of four cspecified by the user
To order replacement partries.
For information regarding tion, refer to the following
• Figure 1. MicroDo• Figure 2. and Tabl
(describes the bas• Table 3. Display E
(describes the bas• Table 4. Docking M• Figure 3. and Tabl• Figure 4. and Tabl
6
MicroDUser M
Micro
ules
To conn Installation.
ock IIanual
Dock II Base Station and Docking Modules
Figure 1. MicroDock II Base Station and Docking Mod
ect additional docking modules and to correctly insert detectors into docking modules, refer to
7
MicroDock IIk II Base Station and Docking Modules
Parts Modu
icroDock II and Docking Module
Description
ocking module lid
etector bay
harger status LED (optional)
ocking module
ocking module pushbuttons and status EDs
attery cover
harger port (optional)
SB port
ower port
ase station
xhaust outlet
tation pushbuttons
alibration gas inlet (inlet 2)
ero air/purge inlet (inlet 1)
iquid crystal display (LCD)
MicroDoc
of the MicroDock II Base Station and Docking le
Figure 2. The MicroDock II Docking Module
Table 2. The M
Item1 D
2 D
3 C
4 D
5 DL
6 B
7 C
8 U
9 P
10 B
11 E
12 S
13 C
14 Z
15 L
MicroDUser M
8
Displ
hbuttonsocking Module Pushbuttons
DescriptionTo bump test a detector, press C
BUMP CHECK.After connecting a new docking mod-ule, press and hold C BUMP CHECK to send a confirmation signal back to the base station.
To calibrate a detector, press C
CALIBRATION (all models exclud-ing GasAlertClip Extreme).
To transfer datalog/event log informa-tion from a detector, press C DATA TRANSFER. (GasAlert Extreme, GasAlertMicroClip, and GasAlertMax XT only).The Automatic Datalog Download option is available for the GasAlertMi-croClip and GasAlertMax XT docking modules only. For more information, refer to Automatic Datalog Download (GasAlertMicroClip and GasAlert-Max XT only)
ock IIanual
ay ElementsTable 3. Display Elements
Pushbuttons
Docking Module PusTable 4. DIcon Description
AC power
Batteries charged
Batteries half-charged
Batteries at low level
MMC or SD card
MMC or SD card not inserted
Test pass
Test fail
Cursor and sensor disabled
Scroll up
Scroll down
Selection arrow
Selected to be modified
Pass code protected
Pushbutton
C BUMP CHECK
C CALIBRATION
C DATA TRANSFER
9
MicroDock IIPushbuttons
Base S
The basand per
Base Station Pushbuttons
Descriptionivate the base stationect menu to access the user optionsect to scroll to different user options or ther functions/selections within a user ionect OKect log to view the results history logect sel to select an optioness a modifiable field
ect exit to exit from a modifiable optionect exit from the user options menu to er normal operation
activate the base stationect to scroll to different user options or ther functions/selections within a user ion
tation Pushbuttons
Figure 3. Base Station Pushbuttons
e station pushbuttons activate, deactivate, scroll, select options, form functions.
Table 5.
Item
1
• Act• Sel• Sel
to oopt
• Sel
2• Sel• Sel• Acc
3• Sel• Sel
ent
4
• De• Sel
to oopt
MicroDUser M
10
Insta Base Station Connections
a Warninge free of background gas. Do not use the ous area.
ectrical Codes (NEC) and safety standards
Description
lter assembly (PURGE 1)
batteries (4)
y cover
s pan head retaining screws (2)
or SD card
y compartment
er port
ort
r port
st outlet
ylinder
nd flow regulator
ation gas hose
ock IIanual
llation
Figure 4. Base Station Connections
Table 6.
The atmosphere must bbase station in a hazardAll required National Elmust be followed.
Item1 Inlet fi
2 C-cell
3 Batter
4 Phillip
5 MMC
6 Batter
7 Charg
8 USB p
9 Powe
10 Exhau
11 Gas c
12 Dema
13 Calibr
11
MicroDock IIInstallation
Thsouba
1.
2.
3.
4.
5.
For AC ing insta
The disstation tion.
a Warningn an atmosphere that is free of hazardous o this warning can result in personal injury e.ase station, refer to Figure 4. and complete the
ng screws from the battery cover. Do not ng screws from the battery cover.
ry cover.
atteries into the battery compartment.
y cover and tighten the retaining screws. Do e screws. Refer to Table 12.
Note
e base station can operate from either an electrical power rce or batteries. The batteries will provide automatic
ckup power if the main power fails.
Connect the power cord to the POWER port on the base station and then plug the cord into an AC outlet.
To install the batteries, refer to Battery Installation.
Connect the charger cord to the CHARGER port on the base sta-tion and then plug the cord into an AC outlet (if applicable).
Attach all gas connections. Inlet 1 (PURGE) is configured for ambient air and inlets 2-5 are configured for calibration/test gases. Refer to Confirming Inlet Setup.
Connect the demand flow regulator. A demand flow regulator must be used with all gas cylinder connections.
Ensure the exhaust line is not connected to a negative pressure system.
main installation, a circuit breaker must be integrated in the build-llation as a disconnect device for the base station.
connect device must be installed in close proximity to the base and must be marked as a disconnecting method for the base sta-
Battery Installation
Only change batteries igas. Failure to adhere tand/or property damag
To install batteries in the bfollowing:
1. Loosen the retainiremove the retaini
2. Remove the batte
3. Insert four C-cell b
4. Replace the batternot over tighten th
MicroDUser M
12
InsertExtrem
Infrarewith th
Figure
GasAlertClip Extreme and the GasAlert Extreme
tor and wait until it is in normal operation.
r clip is closed and the ring is resting flat to s with the data transmission.
ase tabs on the docking module and raise the
(serial number up) into the detector bay. Push the top of the detector connects with the top of
press down until the release tabs click.
has been inserted correctly, the RUN LED(s) dule lights yellow and the base station LCD
g module number,ted,pe, andrial number.
Description
Alligator clip
Docking module lid
Release tabs
Detector bay
ock IIanual
ing the GasAlertClip Extreme and the GasAlert e
a Cautiond or intense ambient light (sun or halogen) may interfere e base station/detector communication.
5. Inserting the GasAlertClip Extreme and the GasAlert Extreme
Table 7. Inserting the
1. Activate the detec
2. Ensure the alligatoprevent disruption
3. Press the two relelid.
4. Lower the detectorforward to ensure the bay.
5. Lower the lid and
When the detectoron the docking modisplays the
• the dockin• Unit Inser• detector ty• detector se
Item1
2
3
4
13
MicroDock IIInstallation
6.
To bypass the base station identification screen, press any C button on the base station.MicroDUser M
14
Insert
Infrarewith th
nserting the GasAlertMicro
tor and wait until it is in normal operation.
r clip is closed and the ring is resting flat on vent disruptions with data transmission.
ase tabs on the docking module and raise the
f the detector into the detector bay (serial en lower the top into place.
Description
Docking module lid
Release tabs
Detector bay
Charge status indicator (charging model only)
Connector pins (charging model only)
Alligator clip
ock IIanual
ing the GasAlertMicroa Caution
d or intense ambient light (sun or halogen) may interfere e base station/detector communication.
Figure 6. Inserting the GasAlertMicro
Table 8. I
1. Activate the detec
2. Ensure the alligatothe detector to pre
3. Press the two relelid.
4. Insert the bottom onumber up) and th
Item1
2
3
4
5
6
15
MicroDock IIInstallation
5.
6.
Lower the lid and press down until the release tabs click.
When the detector has been inserted correctly, the RUN LED(s) on the docking module lights yellow and the base station LCD displays the
• the docking module number,• Unit Inserted,• detector type, and• detector serial number.
To bypass the base station identification screen, press any C button on the base station.
MicroDUser M
16
Insert
Th5 I
Infrarewith th
rting the GasAlertMicro 5/PID/IR
Note
gure 7 and Figure 8.
ock IIanual
ing the GasAlertMicro 5/PID/IRNote
e GasAlertMicro 5, GasAlertMicro 5 PID, and GasAlertMicro R are referenced as the GasAlertMicro 5/PID/IR.
a Cautiond or intense ambient light (sun or halogen) may interfere e base station/detector communication.
Figure 7. Inse
Table 9 references Fi
17
MicroDock IIInstallation
1.
2.
ase tabs on the docking module and raise the
asAlertMicro 5/PID/IR detector is fitted with a adapter must be removed from the docking
moving the Diffusion Adapter
ead screwdriver, press the release tab down-apter forward slightly and then lift upwards to
Table 9. Inserting the GasAlertMicro 5/PID/IR
Activate the detector and wait until it is in normal operation.
Ensure the alligator clip is closed and the ring is resting flat on the detector to prevent disruptions with data transmission.
3. Press the two relelid.
Important: If the Gpump, the diffusionmodule lid.
Figure 8. Re
Using a small flathwards. Pull the adremove.
Item Description1 Docking module lid
2 Release tabs
3 Diffusion adapter
4 Detector bay
5 Connector pins (charging model only)
6 Charge status indicator (charging model only)
7 Connector outlets
8 Diffusion adapter release tab
9 Tool used to push down on diffusion adapter release tab
MicroDUser M
18
4.
5.
6.
ock IIanual
Hold the detector (serial number down) at a 45° angle and insert the bottom into the detector bay.
Ensure that the connector outlets on the bottom of the detector lock into place over the connector pins in the docking bay, and then lower the top into place.
Lower the lid and press down until the release tabs click.
When the detector has been inserted correctly, the RUN LED(s) on the docking module lights yellow and the base station LCD displays the
• the docking module number,• Unit Inserted,• detector type, and• detector serial number.
The GasAlertMicro 5/PID/IR detector LCD displays Microdock.
To bypass the base station identification screen, press any C button on the base station.
19
MicroDock IIInstallation
Insert
Infrarewith th
serting the GasAlertMicroClip
Note
lip detector is fitted with a calibration cap st be removed prior to inserting it into the
tor and wait until it is in normal operation.
ase tabs on the docking module and raise the
a Warning only raises upward 30° angle. Do not force
f the detector (serial number up) at a 30° ctor bay and then lower the top into place.
Description
ule lid
ease tabs
ector bay
ator clip
rger status indicator
ing the GasAlertMicroClipa Caution
d or intense ambient light (sun or halogen) may interfere e base station/detector communication.
Figure 9. Inserting the GasAlertMicroClip
Table 10. In
If the GasAlertMicroCor auxiliary filter, it mudocking module.
1. Activate the detec
2. Press the two relelid.
The docking module lidthe lid beyond its limit.
3. Insert the bottom oangle into the dete
Item1 Mod
2 Rel
3 Det
4 Allig
5 Cha
MicroDUser M
20
4.
5.
ock IIanual
Lower the lid and press down until the release tabs click.
When the detector has been inserted correctly, the RUN LED(s) on the docking module lights yellow and the base station LCD displays the
• the docking module number,• Unit Inserted,• detector type, and• detector serial number.
To bypass the base station identification screen, press any C button on the base station.
21
MicroDock IIInstallation
Insert
Infrarewith th
serting the GasAlertMax XT
tor and wait until it is in normal operation.
r clip is closed and the ring is resting flat on vent disruptions with data transmission.
ase tabs on the docking module and raise the
a Warning only raises upward 30° angle. Do not force
f the detector (serial number up) at a 30° ctor bay and then lower the top into place.
nnector towards the detector pump.
Description
ector pump
ator clip
rge status
ule lid
ease tabs
ector bay
p connector
cator (closed)
cator (open)
ing the GasAlertMax XTa Caution
d or intense ambient light (sun or halogen) may interfere e base station/detector communication.
Figure 10. Inserting the GasAlertMax XT
Table 11. In
1. Activate the detec
2. Ensure the alligatothe detector to pre
3. Press the two relelid.
The docking module lidthe lid beyond its limit.
4. Insert the bottom oangle into the dete
5. Push the pump co
Item1 Det
2 Allig
3 Cha
4 Mod
5 Rel
6 Det
7 Pum
8 Indi
9 Indi
MicroDUser M
22
Thinsde
6.
7.
ock IIanual
Note
e detector will briefly alarm when the pump connector is erted. This is normal. The alarm will deactivate when the tector is properly inserted into the docking module.
Ensure green is visible in the closed indicator. If it is not visible, the base station will not recognize the detector.
Lower the lid and press down until the release tabs click.
When the detector has been inserted correctly, the RUN LED(s) on the docking module lights yellow and the base station LCD displays the
• the docking module number,• Unit Inserted,• detector type, and• detector serial number.
To bypass the base station identification screen, press any C button on the base station.
23
MicroDock IIAdding Another Docking Module
AddinIf requirule.
Only o20 for
To add 11., Fig
ding Another Docking Module
Description
ips pan-head screw (3)
plate
ed fitting ports
ket block
ips flat-head screw
connectors
m cover plate
g Another Docking Moduleed, refer to Using the Base Station before adding a docking mod-
a Warningne module can be connected at a time. Complete steps #1-each docking module that is added.another docking module, refer to Table 12. and Table 13., Figure ure 12., and Figure 13., and complete the following procedures:
Table 12. Torque Specifications
Table 13. Ad
Housing Assembly Torque
Alligator clip 4-5 in lbs
Wall mounting plate 9-10 in lbs
Item1 Phill
2 End
3 Barb
4 Gas
5 Phill
6 PCB
7 Botto
27
MicroDock IIAdding Another Docking Module
1.
2.
3.
4.
InitialiWhen acommu
5.
6.
7.
8.
a Cautione deactivated after each module has been
d defined for each new docking module that is
a Warningp procedures in a clean atmosphere only.e station and wait until it enters normal opera-
a flow meter and to inlet 1 (PURGE) on the
er sufficient time to stabilize (approximately 30 ing it to ensure accurate flow readings.
tion, access the user options menu.
to scroll to the Pump Setup option.
tivate the pump setup field ( changes to ).
mediately begins pumping the ambient air.
Deactivate the base station.
Remove the power cord from the POWER port. If applicable, remove the charger cord from the CHARGER port.
Remove the end plate. There are three Phillips screws on the front and three Phillips screws on the back.
Attach the new docking module.
Ensure the barbed fitting ports mate correctly with the gasket block. Ensure that the male and female PCB connectors mate correctly.
zing the New Docking Module new docking module is connected, it must be initialized (setup to nicate with the base station).
Simultaneously press and hold C BUMP CHECK on the new docking module while pressing C (leftmost button) on the base station.
All of the LEDs on the new docking module light. On the base station LCD, the following screen displays to confirm that the new docking module has been successfully added.
Press C OK.
Deactivate the base station.
The base station must badded.
Defining Pump SpeeThe pump speed must beconnected.
Perform the pump setu9. Reactivate the bas
tion.
10. Connect a hose tobase station.
Allow the flow metminutes) before us
11. From the base sta
12. Press C or C
13. Press C sel to ac
The base station im
MicroDUser M
28
Thspe
14.
15.
16.
17.
18.
19.
20.
Whan
roDock II
aching the Wall Mounting Plate
Note
r more docking modules, each docking dividual set of wall mounting plates.
y of surfaces that the base station can be screws are not provided.
ock IIanual
The factory default pump speed displays beside Pump Setup. The base station pump speed is measured as a percentage (%) and the flow meter measures speed in ml/min.
Note
e base station is shipped with the factory default pump ed set to 350 ml/min. (40-45%).
Monitor the flow meter until it stabilizes (5-20 minutes depending upon the type of flow meter).
From the base station, press C or C until the flow meter displays 350 ml/m.
From the base station, press C exit to save the new value and exit the Pump Setup field ( changes to ).
Press C exit again to return to normal operation.
Repeat steps # 1-17 for each docking module that is added. Con-tinue to step #18 and 19 for the last module added.
After all of the modules have been added, attach the end plate. Refer to Figure 11.
Fasten the bottom cover plate to the space between the two docking modules using six Phillips flat-head screws. Refer to Fig-ure 13.
Note
en assembling or reassembling parts of the base station d docking modules, refer to the Table 12.
Mounting the Mic
Figure 14. Att
When mounting two omodule requires an in
Because of the varietmounted to, mounting
29
MicroDock IIMounting the MicroDock II
The bassecure plete th
e location to mount the base station.
that are provided, attach the wall mounting station and docking modules.
re mounted on the base station, measure hori-l a distance of 2.38 in. (60.32 mm) for the base
more docking modules that are attached on horizontally on the wall a distance of 1.64 in. e between each docking module.
attach the base station to the secure surface.
It
Figure 15. Parts of the Wall Mounting Plate
Table 14. Mounting the MicroDock II
e station and docking modules can be easily mounted to a surface. To mount the base station and docking modules, com-e following:
1. Determine a secur
Using the screws plates to the base
2. When the plates azontally on the walstation width.
If mounting two orthe wall, measure (41.7 mm) of spac
3. Use four screws to
em Description1 Wall mounting plate (2)
2 Grommet (4)
3 Phillips self-tapping screw (8)
MicroDUser M
30
Chanonly)
The gasGasAle
etting on a GasAlertClip Extreme docking and 16, Figures 16 and 17, and complete the
e station.
Phillips pan-head screws from the front and -head screws from the back cover plate.
e docking module from the base station or ules. Refer to Figure 16.
type by setting the dip switches to the desired 15. for gas type and dip switch settings.
sAlertClip Extreme docking module. Refer to
for torque specifications.
cover plate using the six Phillips flat-head remaining three Phillips pan-head screws to ing module.
hanging Dip Switch Settings
S
O
H
C
Description
ver plate
t-head screw (6)
n-head screw (3)
ting ports
ock IIanual
ging Dip Switch Settings (GasAlertClip Extreme
type for bump tests is defined by setting the dipswitch within the rtClip Extreme docking module.
Table 15. Dip Switch Settings
To change the dipswitch smodule refer to Tables 15following:
1. Deactivate the bas
2. Remove the threethe six Phillips flat
3. Gently separate thother docking mod
4. Configure the gasgas. Refer to Table
5. Reconnect the GaFigure 17.
Refer to Table 12.
6. Fasten the bottomscrews. Fasten thethe top of the dock
Table 16. C
Gas Type Dip Switch Setting
ulphur dioxide (SO2)
xygen (O2)
ydrogen sulfide (H2S)
arbon monoxide (CO)
Item1 Bottom co
2 Phillips fla
3 Phillips pa
4 Barbed fit
5 Dip switch
31
MicroDock IIh Settings (GasAlertClip Extreme only)
Wh orrectly label the gas type on the docking mo
ule
Changing Dip Switc
Note
en reattaching the docking module, ensure the barbed tubing is inserted correctly. Ensure to cdule.
Figure 16. Detaching the GasAlertClip Extreme Mod
33
MicroDock IIUsing the Base Station
Using
To prebase sEnsurused i
The basated bythat dis
Confir
FailedBefore cylinderconfirmOptions
ActivaTo activ
1.
(the left most button) until the initializing
nters normal operation.
cally performs a self-test during startup.
r: The base station checks for connected rted detectors.
tion about the docking modules, the detector ial number.
the Base Stationa Warning
vent possible injury and/or property damage, only use the tation in a safe atmosphere that is free of hazardous gas.e that the base station is attached to a venting system or n a well ventilated area.e station pushbuttons are not labelled. The base station is oper-
pressing the C pushbutton that is located directly below the text plays on the LCD.
ming Inlet Setupa Warning
tests can result if the inlets are not setup correctly.activating the base station for the first time, ensure that the gas s are connected to the inlets correctly. Refer to Installation. To that the inlets are setup correctly, refer to Inlet Setup in the User Menu section.
ting the Base Stationate the base station, complete the following:
Connect power to the base station. Refer to Installation.
2. Press and hold Cscreen displays.
The base station e
Self-TestThe base station automati
Docking Module/Detectodocking modules and inse
The LCD displays informatype, and the detector ser
MicroDUser M
34
If a detemessag
Press C
Batteriethereaft
MultiMe
icon (operatiorecords
Pass Cpass comal ope
e Stationtion, it must be in normal operation. From the press and hold C (the rightmost button) until .
n displays briefly before the base station deac-
ock IIanual
ctor is inserted but not activated, the LCD displays the following e.
OK to acknowledge the error message.
s: The batteries are tested during activation and continuously
er. If battery power is extremely low, low battery icon flashes.
diaCard (MMC) / Secure Digital (SD) Card: The MMC/SD card
) displays continuously on the base station LCD during normal n when the MMC/SD card is inserted. The MMC/SD card
a variety of data.
ode Protection: If the base station is pass code protected, the de protect icon displays on the base station LCD during nor-ration.
Deactivating the BasTo deactivate the base stanormal operation screen, Powering Down displays
The powering down screetivates.
35
MicroDock IIUser Options Menu
User The useuser opLCD.
Buingor cal
To acce
1.
2.
3.
e (hour/minute), the date (month/day/year) -7) of the base station.
and/or day of the week, complete the follow-
ions menu, press C or C to scroll to ion.
nfirm the selection and access the time/date
mm: month
dd: day
yy: year
hh: hour
mm: minute
d: day of the week Monday = 1, Tuesday = 2, Wednesday = 3, etc.
Note
e values can only be changed in the esented in this table. To bypass any press C sel. The base station auto-the current value and proceeds to the lue.
Options Menur options menu provides access to 10 options. The following
tions are listed in the order they are displayed on the base station
Note
mp tests or calibrations cannot be performed while access- the user options menu. However, if a test is initiated before while accessing the user options menu, the test automati-ly begins after exiting the user options menu.
ss the user options menu, complete the following:
Press and hold C menu to access the first screen.
If the base station is pass code protected, refer to Entering User Options When Pass Code Protected.
Press C or C to scroll to different options. When dis-plays beside the desired option, press C sel to select.
To exit the options menu and return to the normal operation, press C exit.
Time/DateTime/Date adjusts the timand the day of the week (1
To change the date, time, ing:
1. From the user optthe Time/Date opt
Press C sel to coscreen.
The time and datorder they are prtime/date setting,matically retains next date/time va
MicroDUser M
36
2.
3.
4.
5.
Inlet SThe Inle
•••
a Warningf the inlets are not setup correctly.ivation, ensure the inlets are installed correctly
lt connection for ambient air.lt connection for four-gas mix, unless other-urchase.gned connections for additional gases. How-fied at the time of purchase, inlets 3-5 are con-t air.
re setup correctly, complete the following:
ions menu, press C or C to scroll to tion.
nfirm the selection and access the inlet 1 n.
Note
for ambient air only and cannot be
reen, press C to access the inlet 2 screen.
ock IIanual
The cursor automatically displays below the first value of the month. Each value is selected, changed, and bypassed individu-ally.
Press C to toggle between 0 or 1. When the correct value displays, press C sel to confirm the change. The cursor then automatically moves to the second value of the month.
Or
Press C sel to bypass the first value. The base station automat-ically retains the current value and proceeds to the second value of the month.
Repeat step #2 for the remaining time and date values.
Press C exit to return to the user options menu.
Press C exit again to return to normal operation.
etupt Setup option defines the
the gas type(s)the corresponding gas concentrations level(s), andthe gas cylinder lot # of each selected gas.
Failed tests can result iFor initial base station actas follows:
• Inlet 1 is the defau• Inlet 2 is the defau
wise specified at p• Inlets 3-5 are desi
ever, unless specifigured for ambien
To confirm that the inlets a
1. From the user optthe Inlet Setup op
2. Press C sel to co(ambient air) scree
Inlet 1 is designated changed.
3. From the inlet 1 sc
37
MicroDock IIUser Options Menu
e, proceed to Gas Type. Refer to Table 16. for s.
select from. Refer to the following table.
lets 2-5 – Available Gas Types
e Gas ConcentrationNote
three indicates that the gas concen- be modified on the base station.
ir) %
%
ppm *
xide) ppm *
ide) ppm *
ppm *
Continue to press C to access the inlet 3, 4, and 5 screens. O2 displays as the default gas type when each of the inlet screens are accessed for the first time.
The inlet screens (2-5) have three options to modify:
• Gas Type• Gas Concentration Level• Gas Cylinder Lot # Field (field initially displays blank until
data is entered).
There are two types of screens:
• Single gas
• Multi-gas Type Screen
4. To select a gas typavailable gas type
Gas TypeThere are 33 gas types to
Table 17. In
Gas Typ
The * in columntration level can
Purge (ambient a
O2 (oxygen)
NH3 (ammonia)
CO (carbon mono
ETO (ethylene ox
H2 (hydrogen)
MicroDUser M
38
Note
s and functions when displays. The is selected and can be modified.
plete the following:
inlet screen, displays.
type is displayed, press C to move down ration level. Proceed to the Gas Concentration
e Gas Concentration
ppm and %LEL *
ppm and %LEL *
ppm and %LEL *
ppm and %LEL
ppm *
%LEL *
ppm *
trogen) ppm *
ock IIanual
Scroll through optiondisplays when a field
To select a gas type, com
5. From the required
6. If the required gasto the gas concentLevel section.
Gas Type Gas Concentration
Cl2 (chlorine) bump only ppm *
HCN (hydrogen cyanide) ppm *
H2S (hydrogen sulfide) ppm *
NO (nitric oxide) ppm *
NO2 (nitrogen dioxide) ppm *
PH3 (phosphine) ppm *
SO2 (sulfur dioxide) ppm *
Ethanol % LEL *
Ethanol Equiv. % LEL *
Hexane % LEL *
Hexane Equiv. % LEL *
Methane % LEL *
Methane Equiv. % LEL *
Pentane % LEL *
Pentane Equiv. % LEL *
Propane % LEL *
Propane Equi % LEL *
4 – Gas 2.5%vCH4 (H2S, CO, LEL, O2) ppm and %LEL
4 – Gas Equiv. ppm and %LEL
Gas Typ
Custom 4-Gas
Custom 3-gas
Custom 2-Gas
3 – Gas SO2 Mi (3-gas mix)
Isobutylene
LEL
CO2
CO2 zero / N2 (ni
39
MicroDock IIUser Options Menu
7.
8.
9.
10.
Gas CDependbeside
Thcan
•
••
•
ve down to the gas concentration level.
e gas concentration level can be modified, and the cursor displays below the first value.
ulti-gas: All of the applicable gases display to scroll to the remaining gas types cur-
d).
or
To select a different gas type, press C sel. changes to to indicate that the field is activated. Press C or C to scroll through the list of gas types.
When the desired gas type displays, press C sel. changes to to confirm the new selection and to deactivate the field.
To enter the gas concentration level for the selected gas, pro-ceed to the following section Gas Concentration Level.
Or
Press C exit to return to the user options menu.
Press C exit again to return to normal operation.
oncentration Leveling upon the gas type selected, ppm, %LEL, or % displays
the gas concentration level.
Note
e following gas concentrations are factory configured and not be changed:
Four-gas-mix (4-Gas 2.5% vCH4)
Four-gas equivalent (4-Gas Equiv.)Three-gas SO2 mix (3-Gas SO2 Mi)The Custom 4-Gas, Custom 3-Gas, and Custom 2-Gas can be modified. However, only the gas concentration levels can be changed.
11. Press C to mo
12. Press C sel. If thchanges to the
For custom and m(for 4-Gas press Crently not displaye
MicroDUser M
40
13.
14.
15.
16.
ldned to enter the lot number of the correspond-es of data can be entered. A maximum of 14 numbers) can be selected.
ve down to the lot number field. Initially this
changes to and the cursor displays for
to scroll to the first desired letter/number. cept the value and automatically move the space.
ock IIanual
Press C sel to select the required gas. changes to and the cursor automatically displays below the first value.
Press C or C to scroll to the required value. Press C sel to save the new value.
Or
Press C sel to retain the current value. The cursor automatically moves to the next value.
Repeat step #13 for the remaining values, then press C exit. The then changes back to .
To enter a lot number for the corresponding gas cylinder, pro-ceed to Gas Cylinder Lot # Field.
Or
Press C exit to return to the user options menu.
Press C exit again to return to normal operation.
Gas Cylinder Lot # FieAlthough this field is desiging gas cylinder, other typcharacters (letters and/or
17. Press C to mofield is blank.
18. Press C sel. Thethe first value.
19. Press C or CPress C sel to accursor to the next
41
MicroDock IIUser Options Menu
20.
21.
22.
23.
24.
PumpRefer tosection.
tens or darkens the LCD. To adjust the con-llowing:
ions menu, press C or C to scroll to n.
nfirm the selection and access the field
s range from 2 (brightest) to 10 (darkest).
Note
hipped with the contrast level defined
to select the desired contrast level.
ve the new value and exit the field to ).
Repeat step #19 for the remaining values.
The following screen displays the corresponding lot number for the attached gas cylinder.
Press C exit. The changes back to .
To setup another inlet, repeat the procedures listed in the Gas Type, Gas Concentration Level, and the Gas Cylinder Lot # Field sections.
When all of the entries are completed, press C exit to return to the user options menu.
Press C exit again to return to normal operation.
Setup Defining Pump Speed in the Adding Another Docking Module
ContrastThe Contrast option brightrast level, complete the fo
1. From the user optthe Contrast optio
2. Press C sel to co( changes to ).
The contrast level
The base station is sas 5.
3. Press C or C
4. Press C sel to sa( changes back
MicroDUser M
42
5.
BackliThis BaLCD ligthe bas
Thenmo
To enab
1.
2.
to scroll through the options.
plays beside Backlight and the backlight acti-tes, depending upon the selection.
ve the selection and exit the field
to ).
to scroll to another user option.
turn to normal operation.
the firmware revision for the base station and is connected. To view the firmware revision for ing module(s), complete the following:
to scroll to the About option.
cess the about base station screen.
Enabled
Disabled
Automatic
ock IIanual
Press C or C to scroll to another user option.
Or
Press C exit to return to the normal operation.
ghtcklight option enables, disables, or selects auto mode for the hting. The auto mode automatically disables the backlight when e station is not in use.
Note
e base station is shipped with the auto backlight option abled. While operating from battery power only, select auto de or (off) to conserve the batteries.
le/disable or select auto backlight, complete the following:
From the user options menu, press C or C to scroll to the Backlight option.
Press C sel to select the option and access the field ( changes to ).
3. Press C or C
Each selection disvates and deactiva
4. Press C sel to sa( changes back
5. Press C or C
Or
Press C exit to re
AboutThe About option displayseach docking module thatthe base station and dock
1. Press C or C
Press C sel to ac
auto
43
MicroDock IIUser Options Menu
MaThmothe
2.
FormaFormat(SD) ca
FormaTo form
1.
2.
card into the slot above the battery compart-
he MMC/SD Card into the Base Station
Note
as never been formatted, the base sta- prompt to format the card (refer to steps
to scroll to the Format MMC option.
If more than one docking module is connected to the base sta-tion, press C or C to scroll to the additional docking module firmware revisions (M1 - M10).
Note
ximum 10 docking modules can be connected to a station. ere is a maximum limit of six charging modules. Six charging dules plus four non-charging modules can be connected to station.
Press C exit to return to normal operation.
tting the MMC/SD Card MMC formats the MultiMediaCard (MMC) or the Secure Digital rd.
a Cautiontting a MMC/SD card will erase all data on the card.at an MMC/SD card, complete the following:
Activate the base station if required.
On the base station, remove the cover from the battery compart-ment. Refer to Battery Installation.
3. Insert the MMC/SDment.
Figure 18. Inserting t
If the MMC/SD card htion will automatically#5-6).
4. Press C or C
MicroDUser M
44
5.
6.
to scroll to another user option.
turn to normal operation.
about the MMC/SD card, refer to Base Station
Note
ly to custom and multi-gas types.
les either the automatic (auto) or manual ing an inlet.
ed, the base station automatically selects the mp test or calibration.
the applicable test gas must be selected each tion is performed. If the manu option is Gases Applied for Bump Tests and Calibra-tion.
or manu option, complete the following:
ions menu, press C or C to scroll to n.
ock IIanual
Press C sel to confirm the selection and to access the format-ting confirmation screen.
Press C Yes to initiate formatting. The following screen dis-plays.
When formatting is complete, the LCD automatically returns to the user options menu.
If No is selected, the LCD automatically returns to the user options menu.
If there is a problem with the card, the following screen displays.
Press C OK to acknowledge and insert a new card and repeat steps #4-6.
7. Press C or C
Or
Press C exit to re
For additional information MMC/SD Card.
Inlet Select
Applicable on
The Inlet Sel option enab(manu) function for select
If the auto option is enablcorrect gas inlet for the bu
If manu option is enabled,time a bump test or calibraenabled, refer to Order of tions for important informa
To enable either the auto
1. From the user optthe Inlet Sel optio
45
MicroDock IIUser Options Menu
2.
3.
4.
5.
Pass CThe Pas
menu. Wnormal
Th
To enab
ions menu, press C or C to scroll to tion.
lect the option and access the field ( changes
displays beside the Pass Code option.
to toggle between the options.
option displays, press C sel to confirm the tivate the field ( changes back to ).
to scroll to another user option.
turn to the normal operation.
ass code protect enabled
ass code protect disabled
Press C sel to confirm the selection and access the field ( changes to ).
Press C or C to toggle between the auto and manu option.
When the desired option displays, press C sel to confirm the selection and exit the field ( changes back to ).
Press C or C to scroll to another user option.
Or
Press C exit to return to normal operation.
odes Code option prevents unauthorized access to the user options
hen the base station is pass code protected, displays during operation.
Note
e pass code is provided separately.
le/disable pass code protection, complete the following:
1. From the user optthe Pass Code op
2. Press C sel to seto ).
The current mode
3. Press C or C
4. When the desired selection and deac
5. Press C or C
Or
Press C exit to re
P
P
MicroDUser M
46
EnterinTo acceprotecte
1.
2.
3.
4.
five language options and can display all text g languages:
age, complete the following:
ions menu, press C or C to scroll to ion.
Note
hipped with language defined as English
nfirm the selection and activate the field
ock IIanual
g User Options When Pass Code Protectedss the user options menu when the base station is pass code d, complete the following:
From the normal operating screen, press C menu to access the user options. The enter pass code screen displays.
The cursor displays below the first value.
Press C or C to scroll to the first number of the pass code.
When the correct value displays, press C sel to enter the first value. The cursor automatically moves to the next value.
Repeat step #2 for the remaining values.
Press C OK. If the correct pass code is entered, the user options menu displays.
Incorrect Pass Code Entered: If an incorrect pass code is entered, the base station automatically returns to normal opera-tion.
Repeat steps #1-4 to access the user options menu to try again.
LanguageThe base station provideson the LCD in the followin
• English• French• German• Portuguese• Spanish
To select a different langu
1. From the user optthe Language opt
The base station is s(Eng).
2. Press C sel to co( changes to ).
47
MicroDock IIResults History
3.
4.
5.
s the results of the last 10 bump tests or cali-
e newest result first. The following test errors
Note
station erases the results history log.
log, complete the following:
perating screen, press C log.
istory log (1) displays.
t log, press C . Continue to press C to logs.
turn to normal operation.
Press C or C to scroll to the different language options. The available options are
• Eng (English),• Fran (French),• Deut (German),• Port (Portuguese), and• Espa (Spanish).
When the desired language displays, press C sel to confirm the selection and exit the field ( changes back to ).
The LCD now displays all of the text in the selected language.
Press C or C to scroll to another user option
Or
Press C exit to return to normal operation.
Note
exit will now displays in the selected language.
Results HistoryThe results history displaybrations.
The results display from thcan occur:
• Unit removed• MMC error• IR error
Deactivating hte base
To view the results history
1. From the normal o
The most recent h
2. To access the nexview the remaining
3. Press C OK to re
MicroDUser M
48
RecoGasAle
Fleet Mstation:
•••••
The detbration.
To reco
1.
2.
3.
4.
5.
6.
7.
8.
9.
ollowing options:
ting configuration file from a MicroDockting configuration file from hard driveult configuration fileuration dialog box displays.
e to configure by clicking its tab.
ice.
able checkbox is checked.
option that will reprogram the detector. Each ation screen has different reprogram options. g Detectors in the Fleet Manager II Operator’s
ock IIanual
nfiguring the Detector (not applicable to the rtClip Extreme)
anager II can reconfigure the following detectors with the base
GasAlert ExtremeGasAlertMicroGasAlertMicro 5/PID/IRGasAlertMicroClipGasAlertMax XT
ector can be reconfigured prior to performing a bump test or cali-
nfigure a detector, complete the following:
Activate the base station. Ensure it is in normal operation.
Ensure that a correctly formatted MMC/SD card is inserted in the base station. Refer to Inserting/Replacing a MMC/SD Card.
Connect the USB cable to the computer and to the USB port on the base station.
Activate the detector and wait until it is in normal operation.
Insert the detector into the detector bay.
From the computer, open Fleet Manager II.
Login to the Administration functions.
Under the Devices tab, click Configure Devices.
The Load Configuration File dialog box displays.
Select one of the f
• Load exis• Load exis• Load defa
10. The Device Config
11. Select which devic
12. Configure the dev
Ensure the Updat
13. Select a program detector’s configurRefer to ConfigurinManual.
49
MicroDock IIpplicable to the GasAlertClip Extreme)
14.
15.
16.
17.
guration will be updated when the detector is ocking module, and a bump test or calibration
detector, from he base station press C C CALIBRATION.
user’s consent was selected under Program ure detector? displays on the base station
configure the detector and proceed to the ration.
pass reconfiguration and proceed to the bump
tion is reconfiguring the detector, the following the base station LCD.
Reconfiguring the Detector (not a
When the configuration is complete, click Save. The Save Con-figuration dialog box displays.
Select Save to MicroDock(s)
The Save to MicroDock(s) dialog box displays.
Select a MicroDock II to save to and click OK.
A dialog box displays the message Successfully saved MicroDock Configuration File. Click OK.
The detector confiinserted into the dis initiated.
18. To reconfigure theBUMP CHECK or
If Reprogram withoptions, ReconfigLCD.
Press C Yes to rebump test or calib
Press C No to bytest or calibration.
19. When the base stascreen displays on
MicroDUser M
50
20.
Resolvsolution
•
••
test or calibration, gas must be applied in a as conflicts that can damage the detector’s
Note
nlets are configured correctly. Refer to
utomatically displays if a gas conflict occurs libration is initiated.
s conflicts:
an one inlet is configured for a specific gas
nlet 2 is configured for SO2 and inlet 3 is con-3-gas SO2 mix.ase station is unable to locate the required gas sensor on the detector.
additional information regarding the
flicts,ot found,.g., Bay 1),s),
ock IIanual
Successful Reconfiguring: If reconfiguration is successful, the following screen displays.
Unsuccessful Reconfiguration: If reconfiguration is not suc-cessful, the following screen displays.
Press C ok. The LCD then returns to normal operation.
ing IR Errors: Refer to the following solutions. If the following s do not resolve the IR error, refer to Troubleshooting.
Check the lighting conditions. Infrared or intense light (sunlight or halogen) can cause IR disruptions.Remove and reposition the detector in the docking module.Communication between the detector and base station may have been temporarily disrupted. Complete the procedures again.
Gas ConflictsWhen performing a bumpspecific order to prevent gsensor(s).
Ensure that the gas iInlet Select.
The gas conflicts screen awhen a bump check or ca
There are two types of ga
• Conflicts: More thtype.
Example: Ifigured for the
• Not Found: The btype for a specific
The base station displays
• number of gas con• number of gases n• docking module (e• detector gas type(• inlet, and• inlet gas type(s).
51
MicroDock IIGas Conflicts
If a gasautoma
Refer toceedingOption.
Abort The Abexits theoperatio
Procee
To abor
1.
und screen displays.
to scroll to the Abort Test option.
beside Abort Test, press C sel or press C ion automatically exits the gas conflict and al operation.
be resolved to continue the bump test or cali-ect the gas inlet or accept the gas inlet sug-.
e gas conflict screen and return to normal calibration will not be performed.
conflict is detected by the base station, the gas conflicts screen tically displays.
the following sections, Conflicts and Not Found. Before pro- to resolve gas conflicts, refer to the following section Abort
Optionort option is located in the gas conflicts screen. Selecting Abort gas conflict function and the base station returns to normal n. A bump test or calibration will not be performed.
d to the user options menu and select the Inlet Setup option.
t a gas conflict, complete the following:
From the gas conflicts screen, press C more.
2. The conflicts/not fo
3. Press C or C
4. When displays OK. The base statreturns to the norm
ConflictsIf a conflict occurs, it mustbration. The user may selgested by the base station
Press C cancel to exit thoperation. A bump test or
MicroDUser M
52
Base STo reso
1.
2.
ed within 30 seconds, the base station to the inlet it has selected.
t the gas conflicts screen to access the gas view which gas is configured for more than
.
ock IIanual
tation Selects the Gas Inletlve a conflict automatically, complete the following steps.
Press C more in the gas conflicts screen to access the gas type/inlet screen to view which gas is configured for more than one inlet.
The gases display.
displays on the far right side on the row of the gas that is
setup for more than one inlet. The also indicates the inlet that the base station has selected as the optimum inlet.
Press C OK to accept the inlet that is selected and automati-cally resume the bump test or calibration.
Note
If C sel is not pressautomatically defaults
User Selects Gas Inle1. Press C more in
type/inlet screen toone inlet.
The gases display
53
MicroDock IIGas Conflicts
2.
3.
4.
5.
If Cau
esolution: If the conflict is not resolved by refer to the following:
lected inlet is correct.linder that is attached to the selected inlet is
cept the base station’s default selection.
plays before Not Found, the base station is ed gas type for a specific sensor on the detec-
e gas conflict screen and return to normal calibration will not be performed.
mplete the following.
the gas conflicts screen to access the gas view which gas type(s) cannot be found.
Press C or C to move to the gas that is in conflict.
Press C sel to access the inlet field. The automatically dis-plays on the left side of the inlet.
Press C or C to scroll to the desired inlet (1-5).
When the required inlet value displays, press C sel to confirm the selection and exit the field. The no longer displays to the right of the inlet indicating the conflict has been resolved.
Note
sel is not pressed within 30 seconds, the base station tomatically defaults to the inlet it has selected.
Unsuccessful Conflict Rselecting a different inlet,
• Ensure that the se• Ensure the gas cy
correct.• Press C OK to ac
Not FoundIf a value of 1 or more disunable to locate the requirtor.
Press C cancel to exit thoperation. A bump test or
To define the gas inlets, co
1. Press C more intype/inlet screen to
MicroDUser M
54
2.
Foon
3.
C/SD Card During a Testccess the MMC/SD card on the base station st or calibration is being performed, the follow-
Note
t bump tests and calibrations be com-ting to access test results from the PC.
ock IIanual
The gases display.
displays on the far right side of the gas that cannot be found. The base station automatically defaults to the Purge inlet.
Press C OK to accept the default Purge inlet.
Or
Press C sel to select a different inlet.
Note
r gases not found: If C sel is not pressed within 30 sec-ds, the base station automatically defaults to the Purge inlet.
If required, repeat steps #1-2 for any additional gases not found and confirm the following:
• Ensure that the selected inlet is correct.• Ensure that the gas cylinder attached to the selected inlet
is correct.• Press C OK to accept the base station’s default selec-
tion.
Accessing the MMIf an attempt is made to afrom a PC while a bump teing screen displays.
BW recommends thapleted before attemp
55
MicroDock IIpplied for Bump Tests and Calibrations
OrdeCalib
It is exspecifFailuremanuaperson
Macro
To manfollowin
1.
2.
3.
4.
cate the corresponding rule number. Ensure der to apply gases corresponds to the gas
nceTable 18. and Table 19. to ensure the ied in the correct order.
rder is determined, proceed to the Bump Test ion.
Order of Gases A
r of Gases Applied for Bump Tests and rations
a Warningtremely important that gases are applied in the order that is ied in this manual. to adhere to the rules in Table 18. and Table 19. of this l will result in incorrect sensor readings and possible al injury and/or property damage.
Note
nual application of gas types is available for the GasAlertMi- 5/PID/IR detectors only.
ually apply gases for bump tests and calibrations, complete the g:
Review Table 18. and Table 19.
Determine which gases will be applied.
Using Table 18., locate the first gas type (sensor) to be bump tested or calibrated on the top row.
Within the far left column, locate the next gas type (sensor) to be bumped or calibrated.
The square where the column and row intersect will indicate which gas should be applied first and a rule number (if applica-ble).
5. Using Table 19., lothat the desired orapplication rules.
6. Continue to referegases will be appl
7. When the correct oor Calibration sect
56
MicroDUser M
Gas ) SO2 PH3 PID
H2S ne on
SO2 PH3 Rule 11
H2S
CO ne on
SO2 PH3 Rule 11
CO
CI2 Bump
CI2 Rule 9
PH3 Rule 9
CI2
NH3NH3 Rule 9
PH3 Rules 9 & 10
NH3
HCNHCN Rules 8 & 10
PH3 Rules 8 & 10
HCN
NO2NO2 NO2
Rule 1NO2
COSHH2S/C
SO2 Rule 6
PH3 Rules 10 & 11
COSH H2S/CO
SO2PH3 Rules 9 & 10
SO2
PH3 PH3
PID
ock IIanual
Table 18. Gas Type Application Table
es H2S CO CI2(Bump only)
NH3 HCN NO2COSH
(H2S/CO
Quad gas CI2 Rule 4
NH3 HCN NO2 quad gas - ostep calibrati
CI2 Rule 5
NH3 HCN NO2 quad gas - ostep calibrati
only
NH3 Rule 9
HCN NO2 Rules 8 & 10
CI2 Rule 6
HCN Rule 9
NO2 NH3 Rule 6
HCN HCN Rule 6
NO2 Rule 6
O
57
MicroDock IIpplied for Bump Tests and Calibrations
– wait 5 minutes (minimum) between
is applied before verifying calibration. ferent cylinder to ensure the calibration
t calibration from cross gas
nsor prior to applying any quad gas
Order of Gases A
Table 19. Gas Application Rules
Rule # Gas Type Apply Exception(s)
1 PH3 First Two exceptions: NO2 and O3
2 HCN First Three exceptions: PH3 and ClO2
3 NO2 First Three exceptions: HCN
4 H2S Last One exception: PID
5 CO Last One exception: PID
6 COSH (H2S/CO) Last One exception: PID
7 PID Last No exceptions
8 Toxic gas Both toxic sensors are cross sensitive to each othercalibrations before verifying sensors.
9 2nd Toxic gasWait 5 minutes (minimum) after the second toxic gasVerify calibration: Apply the same test gas from a difis successful.
10 Cross gas Must perform individual calibrations to avoid incorrec
11 H2S/PH3 H2S contaminates PH3 – calibrate and verify PH3 se
MicroDUser M
58
BumpA bumpgas, an
The follduring a
••••
Ensura bum
If tClimuReEx
To perfo
1.
2.
station and wait until it is in normal operation.
module, press C BUMP CHECK. The RUN
nitiated while in the user menu: Bump tests xiting the user options menu.
tion enters normal operation, the base station ns the bump test.
cking module is queued: If more than one dock-ed for a bump test, the first docking module bump test.
hes yellow on the remaining docking modules r a bump test. When the first module is com-dule in the queue begins the bump test.
e type of detector, the option to reconfigure is
reconfigured without user’s consent, the fol-lays.
reconfigured with user’s consent the following
ock IIanual
Test test is performed to confirm that the detector is responding to d that the audio and visual alarms are operational.
owing detectors transfer their event logs to the base station MMC bump test:
GasAlertClip ExtremeGasAlert ExtremeGasAlertMicroClipGasAlertMax XT
a Cautione the detector is not in a low battery state prior to performing p test.
Note
he GasAlertMicro, GasAlertMicro 5/PID/IR, GasAlertMicro-p, or GasAlertMax XT will be reconfigured, reconfiguration st be completed prior to performing a bump test. Refer to configuring the Detector (not applicable to the GasAlertClip treme).
rm a bump test, complete the following:
Ensure the MMC/SD card is inserted if datalogging is required.
If the MMC/SD card is not inserted, displays on the LCD when the base station attempts to write results to the MMC/SD card. If required, refer to Inserting/Replacing a MMC/SD Card.
Activate the detector and wait until it is in normal operation. Insert it into the docking module.
3. Activate the base
4. From the docking LED lights yellow.
If a bump test is iare initiated after e
When the base staautomatically begi
If more than one doing module is queuqueued begins the
The RUN LED flasthat are queued foplete, the next mo
5. Depending upon thprovided.
If the detector waslowing screen disp
If the detector wasscreen displays.
59
MicroDock IIBump Test
If aauthe
6.
Thpla
Ththe
nflict or a selected gas cannot be found, the n displays.
be resolved to continue the bump test. Refer
conflicts, the base station then applies the
To reconfigure the detector, press C Yes. The MMC/SD card then transfers the new configuration data to the detector. If required, refer to Reconfiguring the Detector (not applicable to the GasAlertClip Extreme).
Or
Press C No to continue with the bump test.
Note
selection is not made within 15 seconds, the base station tomatically defaults to the No selection and proceeds with bump test.
The base station begins the bump test and displays the
• docking module number,• test being performed,• detector type, and• detector serial number.
Note
e serial number of the GasAlertClip Extreme does not dis-y on the base station when performing a bump test.
e GasAlertMicro 5/PID/IR beeps and flashes twice to test sensors when the bump check begins.
7. If there is a gas cogas conflicts scree
Gas conflicts mustto Gas Conflicts.
If there are no gasgas.
MicroDUser M
60
If the AGasAlautomuntil thwill fa
8.
plete, the base station displays the results of
tor passed the bump tests. The detector is C OK to return to normal operation.
p test, displays beside the applicable gas
type not detected (H2S is used as an ple, this applies to all gases)
or is disabled
ock IIanual
If more than one gas is being applied, the base station performs a purge between each gas. Depending upon the gas type, the purge time(s) will vary.
a Importantutomatic Datalog Download (GasAlertMicroClip and
ertMax XT only) is enabled in Fleet Manager II, data transfer atically begins during the purge. Do not remove the detector e DATA TRANSFER PASS LED lights, otherwise the transfer
il.For more information, refer to Automatic Datalog Download (GasAlertMicroClip and GasAlertMax XT only).
After the bump test is complete, the base station displays the results of the bump test and performs a final purge to clear any remaining gas.
The countdown of seconds remaining displays beside Purge.
Note
After a bump check is performed, the GasAlertMicro, GasAlert-Micro 5/PID/IR, GasAlertMicroClip, and GasAlertMax XT deacti-vate in 5 minutes if no activity is detected.
Bump Test ResultsAfter the bump test is comthe test.
Bump Pass
indicates that the detecnow ready for use. Press
Bump Fail
If the detector failed a bum
Example: H2S .
Pass
Fail
H2S Gas exam
Sens
61
MicroDock IIBump Test
If the decable teTrouble
If a
Copo5/P
IncorreIf an inltype.
Examp
If a test
••••
a Warningury and/or property damage, replace the fer to the corresponding detector user erence guide for sensor replacement
ults from the sensor do not display on the base here the gas type typically displays on the
Note
, it is not recorded in the results log.
tector fails an audible or visual test, displays beside the appli-st (AUD or VIS). For information regarding solutions, refer to shooting.
Note
gas type fails, calibrate the detector prior to use.
nfirm that the gas concentration exceeds the low alarm set-int of the detector being tested (applicable to GasAlertMicro ID/IR, GasAlertMicroClip, and GasAlertMax XT only).
ct Inlet Setupet is not setup correctly, a hyphen ( ) displays beside the gas
le: SO2, CO, LEL, and O2 gases were not detected.
gas fails, verify that the
gas cylinder is not empty,connections are attached correctly, inlets are setup correctly, andgas concentration exceeds the low alarm setpoint (applicable to GasAlertMicro, GasAlertMicro 5/PID/IR, GasAlertMicroClip, and GasAlertMax XT detectors).
Sensor Disabled
To prevent personal injsensor immediately. Remanual or technical refprocedures.
If a sensor is disabled, resstation LCD. displays wLCD.
If a sensor is disabled
MicroDUser M
62
Calib
Guide
If adjudate w
•••••
RecaliWhen c
•
balance N2 balance N2
or bump checks only) butylene lance air or 0-5.0 % v/v CO2 EL or 0.5 to 5% by vol. methane balance air clean atmosphere that is free of background e the base station in a hazardous area. Failure uideline can result in possible personal injury mage.using premium grade calibration gases and certified to National Standards. The calibration the accuracy of the detector.ylinder beyond its expiration date.ders must be used with a demand flow regula- the following maximum inlet pressure specifi-
cylinders 0-1000 psig/70 barylinders 0-3000 psig/207 bar
e detector during or immediately after charging
before each work shift) allow the detector to temperature that it will be operated in, and
ctor.eriodically re-zero the ETO sensor.or if the ambient gas display varies during
ock IIanual
ration
linesa Caution
stments are made to the real-time clock, the calibration due ill be invalid for theGasAlert Extreme,GasAlertMicro,GasAlertMicro 5/PID/IR,GasAlertMicroClip, GasAlertMax XT detectors.brate the detector immediately.alibrating the sensors, adhere to the following guidelines:
Recommended gas mixture: O2: clean air, 18% vol. CO: (low H2 sensitivity): 50 to 500 ppm balance N2 CO: 50 to 500 ppm balance N2 H2S (high range): 10 to 100 ppm balance N2 H2S (low methanol): 10 to 100 ppm balance N2 H2S: 10 to 100 ppm balance N2 PH3: 1 to 5 ppm balance N2 SO2: 10 to 50 ppm balance N2 NH3: 20 to 100 ppm balance N2 NH3: (high range) 20 to 100 ppm balance N2 NO2: 5 to 50 ppm balance N2 HCN: 5 to 20 ppm balance N2
ETO: 5 to 50 ppm NO: 10 to 250 ppmCl2: 3 to 25 ppm (fVOC: 100 ppm isoCO2: 5000 ppm baLEL: 10 to 100% L
• Calibrate only in agas. Do not operatto adhere to this gand/or property da
• BW recommends cylinders that are gases must meet
• Do not use a gas c• All calibration cylin
tor and must meetcations:
• Disposable• Refillable c
• Do not calibrate this complete.
• For ETO sensors (fully stabilize in thethen zero the dete
• It is necessary to p• Calibrate the sens
start-up.
63
MicroDock IICalibration
•
•
•
••
•
CalibrExtreme
A calibrrate res
Ensura calib
The basment noMicro 5
nsfer their event logs to the base station MMC
omplete the following:
D card is inserted if datalogging is required.
rd is not inserted, displays on the LCD tion attempts to log the results of the calibra-fer to Inserting/Replacing a MMC/SD Card.
tor and wait until it is in normal operation. cking module.
station and wait until it is in normal oepration.
ATION on the corresponding docking module.
Note
ures for the GasAlertMicro 5 IR with CO2 lertMicro 5 IR with CO2 Sensor
itiated while in the user options menu: Cal-ed after exiting the user options menu.
docking module is queued: If more than one queued for calibration, the first module calibration.
hes yellow on the remaining docking modules r calibration. When the first docking module is t module in the queue begins calibration.
Calibrate a new sensor before use. Allow the sensor to stabilize before starting calibration:
• Used sensor: 60 seconds• New sensor: 5 minutes.
Before calibrating a new NO or ETO sensor, allow the sensor to stabilize for 2 hours in a safe area that is free of gas.Calibrate the sensors at least once every 180 days, depending upon use and sensor exposure to poisons and contaminants. For HCN sensors, calibrate at least once every 90 days.Calibrate the sensor before changing the alarm setpoints.When calibrating multiple times, wait 10 minutes between cali-brations to allow the sensor to stabilize.If a certified calibration is required, contact BW Technologies by Honeywell.
ation Procedure (All models excluding the GasAlertClip )
ation adjusts the sensitivity level of a sensor to ensure an accu-ponse to gas.
a Cautione the detector is not in a low battery state prior to performing ration.e station automatically accepts an unusually large span adjust-tification for the GasAlertMicro 5/PID/IR. Refer to the GasAlert-
/PID/IR User Manual.
The following detectors traduring a or calibration:
• GasAlert Extreme• GasAlertMicroClip• GasAlertMax XT
To perform a calibration, c
1. Ensure the MMC/S
If the MMC/SD cawhen the base station. If required, re
2. Activate the detecInsert it into the do
3. Activate the base
4. Press C CALIBR
For calibration procedsensor, refer to GasA
If calibration is inibrations are initiat
If more than one docking module isqueued begins the
The RUN LED flasthat are queued focomplete, the nex
MicroDUser M
64
ThcalMa
If the AGasAlautomuntil thwill fa
5.
6.
as is being applied, the base station performs ach gas. Depending upon the gas type, the ary.
Note
Download is enabled in Fleet Manager lip and GasAlertMax XT automatically
ansfer is complete. To bypass the shut-tector within 15 seconds after the DATA D lights.
n Results.
th CO2 Sensor
Note
ertMicro 5 IR detector, the base station and then calibrates the CO2 sensor first.
eroes the CO2 sensor. The CO2 sensor can ing nitrogen (N2).
en configured to an inlet. If required, refer to
pplies the gas.
ock IIanual
The base station begins the calibration and displays the
• docking module,• test being performed,• detector type, and• detector serial number.
Note
e GasAlertMicro 5/PID/IR beeps and flashes two times when ibration begins, and the GasAlertMicroClip and GasAlert-x XT beeps and flashes once.
a Importantutomatic Datalog Download (GasAlertMicroClip and
ertMax XT only) is enabled in Fleet Manager II, data transfer atically begins during the purge. Do not remove the detector e DATA TRANSFER PASS LED lights, otherwise the transfer
il.If there is a gas conflict or a selected gas cannot be found, refer to Gas Conflicts.
If there are no gas conflicts, the base station then applies the gas.
If more than one ga purge between epurge time(s) will v
If Automatic DatalogII, the GasAlertMicroCdeactivate after the trdown, remove the deTRANSFER PASS LE
7. Refer to Calibratio
GasAlertMicro 5 IR wi
If calibrating a GasAlautomatically zeroes
1. The base station zonly be zeroed us
Ensure N2 has beInlet Setup.
2. The base station a
65
MicroDock IICalibration
3.
4.
5.
6.
Aftdede
plete, the base station LCD displays the
at the calibration has passed. The detector is C OK to return to normal operation.
ays beside the applicable gas.
ass
ail
o gas found (H2S is used as an exam-le, this applies to all gases)
ensor is disabled
The base station then purges the N2 before applying the CO2 gas.
The base station then applies the gas to the CO2 sensor.
The base station then begins calibrating the remaining sensors. For more information, refer to Automatic Datalog Download (GasAlertMicroClip and GasAlertMax XT only).
After calibration is complete, the base station displays the results and performs a final purge to clear any remaining gas.
The countdown of seconds remaining displays beside Purge.
Note
er calibration, the GasAlertMicro 5/PID/IR automatically activates. To bypass the automatic shutdown, remove the tector within 15 seconds after the final purge begins.
Calibration ResultsAfter the calibration is comresults.
Calibration Pass
and Pass indicates thnow ready for use. Press
Calibration Fail
If calibration fails, displ
Example: H2S
P
F
H2S Np
S
MicroDUser M
66
IncorreIf an inlplays F
Examp
If a test
•••
If the deapplicabTrouble
a Warningury and/or property damage, replace the fer to the corresponding detector user cement procedures.
ults from the sensor do not display on the base where the gas type typically displays on the
Note
, it will not be recorded in the results log.
ock IIanual
ct Inlet Setupet is not setup correctly during a calibration, the base station dis-ail and a hyphen ( ) beside the gas type.
le: H2S, CO, and LEL gases were not detected.
gas fails, verify that the
gas cylinder is not empty,connections are attached correctly, andinlets are setup correctly.
tector fails an audible or visual test fails, displays beside the le test (AUD or VIS). For information regarding solutions, refer to
shooting.
Sensor Disabled
To prevent personal injsensor immediately. Remanual for sensor repla
If a sensor is disabled, resstation LCD. displays LCD.
If a sensor is disabled
67
MicroDock IIrtMicroClip, and GasAlertMax XT only)
Data GasAle
TransfThe datbase stManage
DependtransfercroClip
To transthe bas
1.
2.
3.
during the transfer of a datalog, typically some rs to the MMC/SD card. To determine if any sferred, access the MMC data using Fleet to the Fleet Manager II Operator’s Manual.
step #3 to transfer the datalog.
transfer to the base station, the base station og to the MMC/SD card. The RUN and the n the docking module and the following screen se station LCD.
has successfully transferred from the base sta-D card, the following screen displays on the
Data Transfer (GasAlert Extreme, GasAle
Transfer (GasAlert Extreme, GasAlertMicroClip, and rtMax XT only)
erring Datalogsa transfer function transfers datalogs from the detector to the ation’s MMC/SD card. The data can then be imported into Fleet r II.
ing upon the amount of information stored in the detector, the of data may require 12-14 minutes (1-2 minutes for GasAlertMi-and GasAlertMax XT) to complete.
fer datalogs from the detector to the docking module and then to e station MMC/SD card, complete the following:
Activate the detector and wait until it is in normal operation. Insert it into the docking module.
Activate the base station and wait until it is in normal operation.
Press C DATA TRANSFER on the docking module. The yellow RUN LED flashes rapidly and the docking module begins to transfer the datalogs to the base station.
When the transfer process is complete, the docking module indi-cates the status of the transfer as PASS or FAIL.
Successful Transfer: If the datalog transfer is successful, the PASS LED lights green, the RUN LED flashes yellow slowly, and the detector beeps.
Unsuccessful Transfer: If the datalog transfer is unsuccessful, the FAIL LED lights red and the RUN LED flashes slowly.
If a failure occurs of the data transfedata has been tranManager II. Refer
If required, repeat
4. After a successful transfers the datalPASS LEDs light odisplays on the ba
When the datalog tion to the MMC/Sbase station LCD.
MicroDUser M
68
5.
A mstoCli
Whreplog
AutomMax XT
The dettime a b
Using To confrefer to Tools M
To confGasAle
Using To enabManage
1.
2.
toolbar, click Configure Devices via
ration File dialog box displays.
d Configuration File Dialog Box
nd click OK.
etector tab.
tions, click Automatic Datalog Download.
tions, select an option to reprogram the detec-
with user’s consent (but do not change ): When selected, the base station asks if the ill be reconfigured each time a bump check or is initiated. The detector’s clock is not match the MicroDock II’s clock.ally reprogram (but do not change Date/en selected, the detector is automatically d when a bump test or calibration is initiated.
or’s clock is not changed to match the II’s clock.
ock IIanual
On the base station, press C Pass to return to the normal oper-ating screen.
Note
aximum of ten datalogs from all modules combined can be red on an MMC/SD card (100 datalogs for GasAlertMicro-p and GasAlertMax XT).
en the maximum storage is reached, the base station laces the oldest datalogs with the newest datalogs. Data-s are not recorded in the Results History.
atic Datalog Download (GasAlertMicroClip and GasAlert- only)
ector can be configured to automatically transfer datalogs every ump test or calibration is performed.
the IR Linkigure the GasAlertMicroClip using the IR Link and Soft Tools, the GasAlertMicroClip User Manual and GasAlertMicroClip Soft anual.
igure the GasAlertMax XT using the IR Link, refer to the rtMax XT Technical Reference Guide.
the MicroDock IIle Automatic Datalog Download using MicroDock II and Fleet r II, complete the following:
From the PC, open Fleet Manager II.
Log in to the Administration functions.
3. From the DevicesMicroDock.
The Load Configu
Figure 19. Loa
4. Select an option a
5. Click the desired d
6. Under Testing Op
7. Under program optor.
• ReprogramDate/Timedetector wcalibrationchanged to
• AutomaticTime): WhreconfigureThe detectMicroDock
69
MicroDock IIrtMicroClip, and GasAlertMax XT only)
8.
9.
10.
MicroDock(s) checkbox and click OK.
Dock(s) dialog box displays.
ve to MicroDock(s) Dialog Box
k II to save the configuration file and click OK.
g box displays.
ring the Detector (not applicable to the me) to download the new configuration to the
Data Transfer (GasAlert Extreme, GasAle
• Reprogram with user’s consent (including Date/Time): When selected, the base station asks if the detec-tor will be reconfigured each time a bump check or cali-bration is initiated. The detector’s clock will be changed to match the MicroDock II’s clock.
• Automatically reprogram (including Date/Time): When selected, the detector is automatically reconfig-ured when a bump test or calibration is initiated. The detector’s clock is changed to match the MicroDock II’s clock.
Ensure the base station is connected to the computer using a USB cable.
Click Save in the lower left-hand corner.
The Save Configuration dialog box displays.
Figure 20. Save Configuration Dialog Box
11. Click the Save to
The Save to Micro
Figure 21. Sa
12. Select a MicroDoc
The following dialo
13. Click OK.
14. Refer to ReconfiguGasAlertClip Extredetector.
MicroDUser M
70
Base
EventBump tea Securtion insidownloa
The folling a bu
••••
MMC/SD Cardrted in the base station inside the battery com-ace the MMC/SD card, refer to Figure 22 and cedures:
. Replacing a MMC/SD Card
e station.
ng screws from the battery cover. Do not s from the battery cover.
ry cover and set it aside.
nt MMC/SD card (if applicable).
C/SD card into the memory card slot.
y cover and tighten the retaining screws. Do screws
d, refer to Formatting the MMC/SD Card in the n.
ock IIanual
Station MMC/SD Card
Loggingsts and calibrations are recorded on a MultiMediaCard (MMC) or e Digital (SD) card. The MMC/SD card is located in the base sta-de the battery compartment. It stores test records that are then ded from the base station to a PC.
owing detectors transfer event logs to the base station MMC dur-mp test or calibration:
GasAlertClip Extreme (bump test only)GasAlert ExtremeGasAlertMicroClipGasAlertMax XT
Inserting/Replacing aThe MMC/SD card is insepartment. To insert or replcomplete the following pro
Figure 22
1. Deactivate the bas
2. Loosen the retainiremove the screw
3. Remove the batte
4. Remove the curre
5. Insert the new MM
6. Replace the batternot overtighten the
To format an MMC/SD carUser Options Menu sectio
71
MicroDock IIrt Extreme and GasAlertClip Extreme)
AccesTo acceManage
Data thMicroDo
PC ReThe follports Fl
• W
The com
• 80• 10• 51
The com
• A
BWple
tery Pack (Excluding GasAlert Extreme )
a Warningsix charging docking modules that can be ation (six charging modules plus four non- maximum of ten docking modules per
, the temperature must be between 50°F to arge the battery when the detector emits a
a Cautionrger is for indoor use only.
t time, refer to the detector’s user manual or
Note
charge while bump tests or calibrations If the battery is low, charge the detector egin a test. When the test is complete battery.
k Guidelinesector before charging.complete, the detector can remain in the ut wear or damage to the battery.ry pack in temperatures above 86°F (30°C)
e number of possible charges to the battery
Charging the Battery Pack (Excluding GasAle
sing Test Resultsss and view test results using Fleet Manager II, refer to the Fleet r II Operator’s Manual.
at is saved on the MMC/SD card can be transferred to a PC. ck II test results are saved in the .CSV format.
quirementsowing versions of the Microsoft Windows operating system sup-eet Manager II:
indows XP
puter must have the following minimum system requirements:
0 MHz processor0 MB of free disk space2 MB of RAM
puter must also have the following software:
PDF reader
Note
recommends that bump tests and calibrations be com-ted before attempting to access results on the PC.
Charging the Batand GasAlertClip Extreme
There is a maximum of installed on the base stcharging modules for astation). To charge successfully95°F (10°C to 35°C). Chlow battery alarm.
The chaWhen charging for the firsoperator’s manual.
The detector will not are being performed.for 30 minutes then bresume charging the
Charger/Battery Pac• Deactivate the det• When charging is
detector bay witho• Charging the batte
greatly reduces thpack.
MicroDUser M
72
••
••
•
•
Do nois com
•
ChargUp to si
For anydockingchargin
•
tMicro 5/PID/IR or GasAlertMicro charging added, the added module will begin charging icroClip or GasAlertMax XT charging docking
harging. The GasAlertMicroClip or GasAlert-ocking module will begin charging when a
ID/IR or GasAlertMicro charging docking mod- charging.
20. Charger Status LED
charging docking module, complete the follow-
dapter of the charger to the CHARGE port on d then plug the cord into the AC outlet. The LED briefly lights red then green during the
then deactivates.
ector and insert it into the charging cradle. The lights red.
r is inserted correctly onto the contact pins.
lights red (all charging modules).
Description
Charging normally
Charge complete
Charge or temperature fault
ock IIanual
When charging is complete, the battery pack may be warm.The battery pack requires approximately three charge cycles to achieve full charge capacity.The base station only charges using DC power.Charging more than four units simultaneously increases the total charge time.When charging an extremely depleted battery, the battery requires a longer charging time.The detector cannot charge during bump tests or calibrations. If battery power is low, charge the detector for 30 minutes and then initiate the bump/calibration.
a Cautiont calibrate or bump the detector immediately after charging plete. Allow 10 minutes before initiating.When a fully charged battery is inserted into the detector bay, the charger LED lights red for 6 to 10 minutes before lighting green. This action will not wear or damage the battery.
ing Different Detectors Togetherx charging docking modules can be attached to the base station.
combination of GasAlertMicroClip or GasAlertMax XT charging modules with GasAlertMicro 5/PID/IR or GasAlertMicro docking g modules, refer to the following:
If another GasAlertMicroClip or GasAlertMax XT charging dock-ing module is added, the added module will wait in queue until a GasAlertMicro 5/PID/IR or GasAlertMicro charging docking mod-ule has completed charging before beginning to charge.
• If another GasAlerdocking module isand one GasAlertMmodule will stop cMax XT charging dGasAlertMicro 5/Pule has completed
Charging ProcedureTable
To charge a detector in a ing:
1. Connect the AC athe base station anCHARGE STATUSself-test. The LED
2. Deactivate the detcharging LED then
Ensure the detecto
3. The CHARGE LED
Charger StatusRed
Green
Off
73
MicroDock IIMaintenance
4.
a Warningr-serviceable parts inside.
on in good operating condition, perform the fol- as required.
roduct operation, maintain a log of all mainte-rmed.
with a soft, damp cloth. Do not use solvents, .let filter is free of dirt and replace it if required. ent parts, refer to Replacement Parts and
e base station in liquids.
sAlertMicro 5/PID/IR periods of time, ensure that the detector is
ed every 30 days.
ill be stored for an extended period of time, charged before storage and then recharged due dates may be lost if the detector is not
Allow the battery to obtain a full charge (2 to 6 hours, depending upon detector type and number of docking modules connected to the base station).
If a docking module is in queue waiting charge, the CHARGE LED flashes red (all charging modules).
When the docking module begins charging, the CHARGE LED stops flashing and lights solid red.
The CHARGE LED lights green when the charging process is complete. If there are other docking modules in queue, waiting to charge, the docking modules will begin charging.
The CHARGE LED deactivates when the detector is removed.
Maintenance
No useTo maintain the base statilowing basic maintenance
• To ensure quality pnance that is perfo
• Clean the exteriorsoaps, or polishes
• Confirm that the inTo order replacemAccessories.
• Do not immerse th
Battery Pack StorageGasAlertMicro and GaWhen storing for extendedfully charged and recharg
GasAlertMicroClipIf the GasAlertMicroClip wensure the detector is fullyevery 30 days. Calibrationrecharged every 30 days.
74
MicroDUser M
TroubIf a prob be resolved, contact BW Technologies by Hon-eywell.
Solution
nnect the power adapter.
sert the batteries.
place the batteries.
ntact BW Technologies by Honeywell.
nfirm connection. Refer to Adding other Docking Module.
nnect a gas cylinder that has a higher s concentration level than the alarm set-ints.
nfirm that all gas connections are tached correctly.
fer to Inlet Setup and Installation.
e a new gas cylinder.
ntact BW Technologies by Honeywell.
ock IIanual
leshootinglem is encountered, refer to the solutions provided in the following table. If the problem cannot
[*Applicable to GasAlertMicro and GasAlertMicro 5/PID/IR docking modules only].
Table 21. Troubleshooting Tips
Problem Possible Cause
The base station does not activate.
There is no power connection. Co
Batteries are not inserted. In
The batteries are depleted. Re
The base station is damaged or defective. Co
The detector fails a bump test.
Barbed fittings are not connected correctly with the gasket.
CoAn
Detector alarm setpoints are set higher than the gas concentration levels of the gas cyl-inder.
Cogapo
Gas connections are not attached correctly or are blocked.
Coat
Gas inlets are not setup correctly. Re
Gas cylinder is empty. Us
Detector and/or base station are damaged or defective. Co
75
MicroDock IITroubleshooting
Solution
nnect a quad-gas cylinder that s 18% O2.
move the blockage or replace the inlet er.
e a new zero gas cylinder.
e a new gas cylinder.
nfirm connection. Refer to Adding other Docking Module.
nfirm that all gas connections are tached correctly
ntact BW Technologies by Honeywell.
place the batteries.
activate the base station, wait 5 seconds, en reactivate.
ntact BW Technologies by Honeywell.
Problem Possible Cause
The oxygen sensor fails a bump test.
Oxygen alarm setpoints of the detector are set to the same concentration level as the gas cylinder.
Coha
Ambient air (inlet 1) is blocked or the inlet filter is contaminated.
Refilt
Zero gas cylinder is empty. Us
The detector fails a calibration.
Gas cylinder is empty. Us
Barbed fittings are not connected correctly to the gasket.
CoAn
Gas connections are not attached correctly or are blocked.
Coat
Detector and/or base station are damaged or defective. Co
Flashing battery icon . Batteries are extremely depleted. Re
The base station is activated but does not respond to button presses.
Irregular power (not continuous or reliable) electrostatic discharge.
Deth
There is no audible beep (GasAlertMicro and GasAlertMicro 5/PID/IR).
Internal damage to battery or docking mod-ule. Co
76
MicroDUser M
Solution
tivate the detector.
fer to the label located on the back of the tector for IR capabilities.
ntact BW Technologies by Honeywell.
ait approximately 30 seconds.
-initialize the docking module—refer to ding Another Docking Module.
ntact BW Technologies by Honeywell.
low the battery pack time to adjust to the ecified operating temperature (approxi-ately 60 minutes).
arge (2–4 hours) or replace old battery th new battery. If problem persists, con-ct BW Technologies by Honeywell.
ntact BW Technologies by Honeywell.
ock IIanual
Problem Possible Cause
The docking module does not recognize the detector.
Detector is deactivated. Ac
Detector does not have IR communication capabilities.
Rede
Firmware of the detector requires an update. Co
Base station is currently attempting to establish communication with the detector. W
Docking module is not initialized to the cor-rect number.
ReAd
Detector and/or base station are damaged or defective. Co
There is an audible beep but no charge (GasAlertMicro and GasAlertMicro 5/PID/IR docking modules).
Battery is above or below the operating temperature 50°F–95°F (10°C–35°C).
Alspm
Severely depleted battery.Chwita
Damaged or defective battery pack. Co
77
MicroDock IITroubleshooting
Solution
rmly insert the detector into the detector cking bay.
low the battery pack time to adjust to the ecified operating temperature (approxi-ately 60 minutes).
low the detector to charge (2–4 hours).
sure Windows XP assigned a drive letter the MicroDock II.lution:
Right-click My Computer and then click anage. Under Computer Management (Local) ck Disk Management. In the list of drives in the right pane, right-ck the new drive and then click Change ive Letter and Path(s).
Click Change and in the drop-down box, lect a drive letter for the new drive (cannot assigned to a mapped network drive). Click OK and then click OK again.
peat the test.
sure the MMC/SD card is inserted.
Problem Possible Cause
The charger LED does not light when the detector is inserted (For GasAlertMicro, GasAlertMicro 5/PID/IR, GasAlertMicroClip, and GasAlertMax XT docking modules).
Detector is not inserted into the detector bay correctly.
Fido
The battery pack is above or below the operating temperature 50°F–95°F (10°C–35°C).
Alspm
The battery pack is severely depleted. Al
No drive letter is created for the MicroDock II (Windows XP) on the PC.
Windows XP is not correctly mapping the drive.
EntoSo1)M2)cli3)cliDr4)sebe5)
Problem transferring to the MMC/SD card during a calibration or bump test.
Re
En
78
MicroDUser M
Solution
sert a 128 MB Delkin MMC
fer to Formatting the MMC/SD Card.
place with a new MMC/SD card.
fer to Formatting the MMC/SD Card.
place with a new MMC/SD card.
eck for dirt on the IR port.
nfirm the detector is activated.
arge the detector battery if too low.
sert only datalogger models.
ebug code displays on the LCD, record it d contact BW Technologies by Honey-ll.
ock IIanual
Problem Possible Cause
Size of MMC is not supported. In
MMC is not correctly formatted.
Re
Re
Log file has exceeded maximum size.Re
Re
IR not working when detector is inserted into the module.
Ch
Co
Ch
Non-datalogger detector is inserted in the docking module. In
Software exception.If danwe
79
MicroDock IIReplacement Parts and Accessories
Repla
To avothe sp
To orde
GDDGDDGDGD
D
D
D
Description Docking Module-N Charging (power supply included)ocking Module
Charging (power supply included)ication Devices and Accessories
MMC USB Reader (USB port) for MMCNetwork USB communication deviceDelkin 128 MB MultiMediaCard (MMC)
ersFor MicroDock II and charging mod-ules (kit of 2)For 34 l gas cylinder (black)For 58 l gas cylinder (red)
s and AccessoriesDiffusion adapter
Hard-sided (1 module system)Heavy duty (1 to 3 module system)
ulatorsDemand flow regulatorFor refillable cylinders CGA con-nector
cement Parts and Accessoriesa Warning
id personal injury or damage to the base station, use only ecified replacement parts.r parts or accessories, contact BW Technologies by Honeywell.
Table 22. Replacement Parts and Accessories
Model Number DescriptionasAlertMicro Docking ModuleOCK2-0-1A-00-N Non-chargingOCK2-0-1B1C-00-N Charging (power supply included)asAlertMicro 5/PID/IR Docking ModuleOCK2-0-1J-00-N Non-chargingOCK2-0-1C1K-00-N Charging (power supply included)asAlert Extreme Docking ModuleOCK2-0-1D-00-N With dataloggingasAlertClip Extreme Docking ModuleOCK2-0-1E-00-N H2S (hydrogen sulfide)
OCK2-0-1H-00-N O2 (oxygen)
OCK2-0-1G-00-N CO (carbon monoxide)
OCK2-0-1F-00-N SO2 (sulfur dioxide)
Model NumberGasAlertMicroClipDOCK2-0-1C1L-00GasAlertMax XT DDOCK2-0-1C1MMMC/SD Commun
CR-MMC-USB1
DOCK2-LAN1
MMC128-D
Wall Mount Adapt
WMA-DOCK
MK-CG2-34BMK-CG2-58Replacement PartDOCK2-DA-1Carrying CasesDOCK2-CC1DOCK2-CC2Demand Flow RegREG-DF1
REG-DF2
MicroDUser M
80
SpecThe MicInstrum(8.3 x 1WeightBase sEnclosOperatHumiditAltitudeStoragePower (be adv10% of PollutioInstallaReal-timData st128 MBExternaPump: Flow raCalibra2-gas in4-gas +Automable alar
n: Automatic (instrument and sensor)eters: User-settableions: Built-in (base station)C connect sub-miniature couplingg modules) docking module)
d: Infrared (two-way) between docking module le to the GasAlertClip Extreme – one way
to a:mputer (PC), or
P HUB
rs, wide viewing angle, user selectable back-
enu navigationne touch bump test initiationne touch calibration initiation ot applicable to the GasAlertClip Extreme)ne touch data transfer initiation asAlert Extreme, GasAlertMicroClip, and
asAlertMax XT only)
ock IIanual
ificationsroDock II is for indoor use only.ent dimensions: 21.2 x 26.3 x 8.2 cm
0.4 x 3.2 in.) base station and one docking module: 0.98 kg (2.15 lb)ystem ingress protection: IP20ure: Impact resistant PC/ABS (polycarbonate)ing temperature: +10°C to +35°C (+50°F to +95°F)y: 0 to 50%: 2000 m (6561.66 ft.) temperature: -10°C to 60°C (14°F to 140 ºF)
supply: 6 Vdc , 1.5 A wall adapter or four C-cell batteries ised that the main supply voltage fluctuations are not to exceed the nominal supply)n degree: 2tion category: Ie clock: Provides time and date stamp
orage: Automatic (instrument and base station) Delkin MMC data storage systeml interface: USB interface for PCDC motor, micro-diaphragm; 3V PCB mountte: Maximum recommended 350 ml/min.tion gas cylinder inputs:lets (standard) air inlets (maximum)tic tests: Functional bump test, calibration, data transfer, audi-
m, visual alarm
Configuration recognitioAlarm/calibration paramCalibration gas connectGas connection: 1/8" SMSolenoid: Built-in (dockinLED indicators: (on eachYellow – Test Green – Pass Red – FailCommand keys:
Communications methoand detector (not applicabcommunication only)USB port for connection
Personal coUSB over I
Sensors: AudioOptical
LCD: 4 line x 16 charactelightingWarranty: 2 years
Base station M
Docking module
OO(nO(GG
81
MicroDock IISpecifications
ChargSize: 8.WeightChargeOperatHumidiAltitudePower:Charginand chaChargePollutioInstallaThis derequirem
••
tested and found to comply with the limits for a suant to Part 15 of the FCC Rules and ICES-ments.
to provide reasonable protection against the equipment is operated in a commercial
, uses, and can radiate radio frequency and used in accordance with the instruction ul interference to radio communications.
nt in a residential area is likely to cause harm-ase the user will be required to correct the expense.
a Warningd for installation in an indoor location only. ectrical Codes and Safety Standards must
n, a circuit breaker should be included in as a disconnect device for the equipment.
should be installed in close proximity to the ice should be marked as a disconnecting nt.
er Specifications6 x 8.2 x 7.8 cm (3.4 x 3.2 x 3.1 in.): 97 g (3.4 oz)r system ingress protection: IP20ing temperature: 10°C to 35°C (50°F to 95°F)ty: 0% to 50%: 2000 m (6561.66 ft.)
6 Vdc , 2.5 A g LED: Color-coded LED advises: charging, charge complete, rger fault. time: Typically 2-6 hoursn degree: 2tion category: Ivice complies with the FCC Part 15 and ICES-003 Canadian EMI
ents. Operation is subject to the following two conditions:
This device may not cause harmful interference, andThis device must accept any interference received, including interference that may cause undesired operation.
This equipment has been Class A digital device, pur003 Canadian EMI require
These limits are designedharmful interference whenenvironment.
This equipment generatesenergy, and if not installedmanual, may cause harmf
Operation of this equipmeful interference in which cinterference at his/her own
This product is designeAll required National Elbe followed.For AC main installatiothe building installationThe disconnect device equipment and the devmeans for the equipme