automated extraction system method note compendiumcem.com › ... › literature ›...

136
EDGE ® Automated Extraction System Method Note Compendium Updated February 20, 2019

Upload: others

Post on 29-May-2020

7 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

EDGE®

Automated Extraction System

Method Note Compendium

Updated February 20, 2019

Page 2: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Agriculture

Avocados 2

Beans 3

Brazil Nuts 4

Cacao 5

Cannabis Plant (Cannabinoids) 6

Cannabis Plant (Mycotoxins) 7

Cannabis Plant (Pesticides) 8

Dry Fish Samples 9

Dry Soy Meal & Fish Meal 10

Dry Spice Samples 11

Hazel Nuts 12

Hops 13

Pumpkin Seeds 14

Rat Poison Powder 15

Rice 16

Strawberries 17

Wet Feed/ Food Samples (PCBs) 18

Wet Feed/ Food Samples (Dioxins) 19

Wet Fish Tissue 20

Wet Food Sample 21

Total Fat From Feed/Food 22

Cannabis

Cannabis Plant (Cannabinoids) 24

Cannabis Plant (Mycotoxins) 25

Cannabis Plant (Pesticides) 26

Environmental & Regulatory

Environmental Samples 28

EPA 3545A 29

Large PUF Filters 30

Small PUF Filters 31

Wet Fish Tissue 32

PUF Filters 33

Soil, Loam or Clay (PAHs) 34

Soil, Loam, or Clay (PCBs) 35

Soil, Loam, or Clay (Dioxins) 36

Food Testing

Environmental Samples 38

EPA 3545A 39

Film Powder 40

Plastics 41

Avocados 42

Beans 43

Brazil Nuts 44

Cacao 45

Cannabis Plant (Cannabinoids) 46

Cannabis Plant (Mycotoxins) 47

Cannabis Plant (Pesticides) 48

Cashews 49

Dry Fish Samples 50

Dry Soy Meal & Fish Meal 51

Dry Spice Samples 52

Echinacea 53

Egg Powder 54

Hazel Nuts 55

Hops 56

Large PUF Filters 57

Mouthwash 58

PET 59

Contents

Page 3: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Pigment 60

Polyethylene Powder 61

Polyethylene Resin 62

Polypropylene Powder 63

Polyvinyl Chloride 64

Protein Powder 65

Pumpkin Seeds 66

Rat Poison Powder 67

Resin 68

Rice 69

Small PUF Filters 70

Soil 71

Strawberries 72

Wet Feed/Food Samples (PCBs) 73

Wet Feed/ Food Samples (Dioxins) 74

Wet Fish Tissue 75

Wet Food Sample 76

XAD Resin 77

Yerba Mate 78

Molybdenum 79

Total Fat From Feed/Food 80

Total Fat From High-Fat Meats 81

Total Fat From Hotdogs 82

Total Fat From Peanuts 83

Total Fat From Pepperoni 84

Total Fat From Potted Meat 85

Total Fat From Protein Powder 86

PUF Filters 87

Soil, Loam, or Clay (PAHs) 88

Soil, Loam, or Clay (PCBs) 89

Soil, Loam, or Clay (Dioxins) 90

Toothpaste 91

Total Fat From Chocolate 92

Total Fat From Cookies 93

Geoscience & Mining

Molybdenum 95

Industrial Hygiene

Large PUF Filters 97

Rat Poison Powder 98

Small PUF Filters 99

PUF Filters 100

Materials Science

Film Powder 102

Plastics 103

Large PUF Filters 104

Polypropylene Powder 105

Polyvinyl Chloride 106

Rat Poison Powder 107

Resin 108

Small PUF Filters 109

XAD Resin 110

PUF Filters 111

Metals & Alloys

Molybdenum 113

Nutraceuticals

Dry Fish Samples 115

Echinacea 116

Protein Powder 117

Wet Fish Tissue 118

Yerba Mate 119

Total Fat From Protein Powder 120

Page 4: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Personal Care Products

Mouthwash 122

Toothpaste 123

Plastics, Polymers, & Oils

Film Powder 125

Plastics 126

PET 127

Pigment 128

Polyethylene Powder 129

Polyethylene Resin 130

Polypropylene Powder 131

Polyvinyl Chloride 132

Page 5: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Agriculture

1

Page 6: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PesticidesFrom AvocadosFrom Avocados

Agriculture, Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight5 g

Q-DiscᵀᴹM2 + C9

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. Layer the M2 Q-Disc (textured side up) on top of the

C9 Q-Disc. 2. Weigh the sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup. Gently mix the sample and sand in the Q-Cup.

SolventsAcetonitrile

Sorbents- Sodium Sulfate (5 g)- Magnesium Sulfate (1,200 mg)- Primary Second Amine (50 mg)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 10020 01:00010

Stage Solvent Volume (mL)

1 Water 15

2 Acetonitrile 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

1152

Page 7: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PolyphenolsFrom BeansFrom Beans

Agriculture, Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight0.1 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

Solvents(50:50) Methanol: Water

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 8020 10:001010

Stage Solvent Volume (mL)

1 (50:50) Methanol: Water 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

1243

Page 8: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of FatFrom Brazil NutsFrom Brazil Nuts

Agriculture, Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight12 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup. Gently mix the sample and sand in the Q-Cup.

SolventsPetroleum Ether

SorbentsSand (12 g)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 17510 00:00010

Stage Solvent Volume (mL)

1 Petroleum Ether 20

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

1254

Page 9: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PolyphenolsFrom CacaoFrom Cacao

Agriculture, Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight0.1 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

Solvents(50:50) Methanol: Water

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 8020 10:001010

Stage Solvent Volume (mL)

1 (50:50) Methanol: Water 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

1075

Page 10: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of CannabinoidsFrom Cannabis PlantFrom Cannabis Plant

Agriculture, Cannabis, Food TesEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight1 g

Q-DiscᵀᴹM2 + C9

EquipmentEDGE40 mL collection vials - clear

NotesPossessing, using, distributing or selling marijuana or marijuana-based products constitute federal crimes in the United States, even where a state law decriminalizes or legalizes such activities. CEM Corporation produces instruments that are intended for use in testing laboratories and applications only where such use is permitted under applicable state/country law.

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. Layer the M2 Q-Disc (textured side up) on top of the C9 Q-Disc.2. Weigh the sample into the Q-Cup.

SolventsMethanol

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 10010 0:3055

Stage Solvent Volume (mL)

1 Methanol 20

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

686

Page 11: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of MycotoxinsFrom Cannabis PlantFrom Cannabis Plant

Agriculture, Cannabis, Food TesEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight1.5 g

Q-DiscᵀᴹC4

EquipmentEDGE40 mL collection vials - clear

NotesPossessing, using, distributing or selling marijuana or marijuana-based products constitute federal crimes in the United States, even where a state law decriminalizes or legalizes such activities. CEM Corporation produces instruments that are intended for use in testing laboratories and applications only where such use is permitted under applicable state/country law.

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

SolventsMethanol May also use a percentage of Water:Methanol as needed

SorbentsNone

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 5015 02:0055

Stage Solvent Volume (mL)

1 Methanol 10

2 Methanol 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

737

Page 12: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PesticidesFrom Cannabis PlantFrom Cannabis Plant

Agriculture, Cannabis, Food TesEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight1.5 g

Q-DiscᵀᴹM2 + C9

EquipmentEDGE40 mL collection vials - clear

NotesPossessing, using, distributing or selling marijuana or marijuana-based products constitute federal crimes in the United States, even where a state law decriminalizes or legalizes such activities. CEM Corporation produces instruments that are intended for use in testing laboratories and applications only where such use is permitted under applicable state/country law.

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. Layer the M2 Q-Disc on top of the C9 Q-Disc.2. Weigh the sorbent into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup layering the sample on top of the salt. No mixing is required.

Solvents1% Acetic Acid in Acetonitrile

SorbentsContents of Agilent QuEChERS SPE 15 mL Tube

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 4020 02:0055

Stage Solvent Volume (mL)

1 1% Acetic Acid in Acetonitrile 15

2 1% Acetic Acid in Acetonitrile 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

1118

Page 13: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PAHsFrom Dry Fish SamplesFrom Dry Fish Samples

Agriculture, Food Testing, NutraEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight5 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™.

2. Weight the sorbent in the Q-Cup. Tap lightly to spread evenly. 3. Weigh the sample into the Q-Cup layering the sample on top of the salt. No mixing is required.

SolventsDichloromethane

SorbentsC18 (1.7 g)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 12020 03:001010

Stage Solvent Volume (mL)

1 Dichloromethane 15

2 Hexane 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

859

Page 14: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PCBsFrom Dry Soy Meal & Fish MealFrom Dry Soy Meal & Fish Meal

Agriculture, Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight5 g

Q-DiscᵀᴹC1M1

EquipmentEDGE40 mL collection vials - clear

NotesLess than 5 grams of sample can be used with a ratio of 1:2 C18: Sample Either C1 or M1 Q-Disc can be used

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sorbent into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup layering the sample on top of the salt. No mixing is required.

SolventsOptions:Dichloromethane(50:50) Hexane: Acetone

SorbentsC18 (2.5 g)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 12020 03:001010

Stage Solvent Volume (mL)

1 Dichloromethane 15

2 Hexane 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

7010

Page 15: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PesticidesFrom Dry Spice SamplesFrom Dry Spice Samples

Agriculture, Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight1 g

Q-DiscᵀᴹC9 + G1 + C9

EquipmentEDGE40 mL collection vials - clear

NotesMatrix matched calibration curve required without the use of sorbents.

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. Sandwich the G1 disc between the two C9 discs.2. Weigh the sample into the Q-Cup.

SolventsAcetonitrile with 1% Acetic Acid

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 4015 03:00510

Stage Solvent Volume (mL)

1 Acetonitrile with 1% Acetic Acid 10

2 Acetonitrile with 1% Acetic Acid 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

8011

Page 16: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of FatFrom Hazel NutsFrom Hazel Nuts

Agriculture, Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight12 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup. Gently mix the sample and sand in the Q-Cup.

SolventsPetroleum Ether

SorbentsSand (12 g)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 17510 00:00010

Stage Solvent Volume (mL)

1 Petroleum Ether 20

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

12712

Page 17: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PesticidesFrom HopsFrom Hops

Agriculture, Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight0.5 g

Q-DiscᵀᴹM2 + C9

EquipmentEDGE40 mL collection vials - clear

NotesCanadian Pesticide Mix 4 in LCMS grade obtained from SPEX CertiPrep.

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. Layer the M2 Q-Disc (textured side up) on top of the

C9 Q-Disc.2. Add sorbents to bottom of Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping. No mixing is required.

3. Weigh the sample into the Q-Cup. Add 5.0 g hops on top of sorbent bed. 4. Spike with 100 ug/mL Canadian Pesticide Mix 4 and leave in hood for approximately 30 minutes to evaporate solvent.

Solvents1% Acetic Acid in Acetonitrile

Sorbents-ACS grade magnesium sulfate anhydrous > 99.5% (1.25 g)-Supelᵀᴹ QuE Citrate (1.25 g)-Primary secondary amine (1.25 g)-Siliabond C18 (1.25 g)*Different sorbents can be used

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 10015 00:30510

Stage Solvent Volume (mL)

1 1% Acetic Acid in Acetonitrile 20

2 1% Acetic Acid in Acetonitrile 20

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

6213

Page 18: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of FatFrom Pumpkin SeedsFrom Pumpkin Seeds

Agriculture, Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight12 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup. Gently mix the sample and sand in the Q-Cup.

SolventsPetroleum Ether

SorbentsSand (12 g)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 17510 00:00010

Stage Solvent Volume (mL)

1 Petroleum Ether 20

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

12614

Page 19: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of BrodifacoumFrom Rat Poison PowderFrom Rat Poison Powder

Agriculture, Industrial Hygiene, EDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight2 g

Q-DiscᵀᴹC4

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

SolventsMethanol

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 13015 02:001010

2 13015 02:001010

Stage Solvent Volume (mL)

1 Methanol 15

2 Methanol 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

9815

Page 20: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PesticidesFrom RiceFrom Rice

Agriculture, Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight5 g

Q-DiscᵀᴹM2 + C9

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. Layer the M2 Q-Disc (textured side up) on top of the C9 Q-Disc. 2. Weigh the sorbents into the Q-Cup. Spread evenly across the sample by gentle tapping.3. Weigh the sample into the Q-Cup layering the sample on top of the salt. No mixing is required.

SolventsAcetonitile

Sorbents- Sodium Sulfate (5 g)- Magnesium Sulfate (1,200 mg)- Primary Second Amine (50 mg)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 10020 01:00010

Stage Solvent Volume (mL)

1 Acetonitile 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

11616

Page 21: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PesticidesFrom StrawberriesFrom Strawberries

Agriculture, Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight5 g

Q-DiscᵀᴹM2 + C9

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. Layer the M2 Q-Disc (textured side up) on top of the C9 Q-Disc. 2. Weigh the sorbents into the Q-Cup. Spread evenly across the sample by gentle tapping.3. Weigh the sample into the Q-Cup layering the sample on top of the salt. No mixing is required.

SolventsAcetonitrile

Sorbents- Sodium Sulfate (5 g)- Magnesium Sulfate (1,200 mg)- Primary Second Amine (50 mg)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 10020 01:00010

Stage Solvent Volume (mL)

1 Water 15

2 Acetonitrile 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

11417

Page 22: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PCBsFrom Wet Feed/ Food SamplesFrom Wet Feed/ Food Samples

Agriculture, Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight5 g

Q-DiscᵀᴹC9 + G1 + C9

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. Sandwich the G1 disc between the two C9 discs.2. Weigh the sorbent into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup layering on top of the sorbent. No mixing is needed.

SolventsDichloromethane

Sorbents-Sodium Sulfate (2.5 g)-C18 (2 g)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 13520 02:00510

Stage Solvent Volume (mL)

1 Water 15

2 Hexane 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

6918

Page 23: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of DioxinsFrom Wet Feed/ Food SamplesFrom Wet Feed/ Food Samples

Agriculture, Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight5 g

Q-DiscᵀᴹC4

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sorbent into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup layering on top of the sorbent. No mixing is needed.

SolventsDichloromethane

Sorbents-Sodium Sulfate (2.5 g)-C18 (2 g)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 15020 02:00510

Stage Solvent Volume (mL)

1 Water 15

2 Hexane 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

7219

Page 24: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PAHsFrom Wet Fish TissueFrom Wet Fish Tissue

Agriculture, Environmental and EDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight2 g

Q-DiscᵀᴹC9 + G1 + C9

EquipmentEDGE40 mL collection vials - clear

NotesIf there is moisture in collection vial after the extraction:-Using a glass funnel with glass wool and about 20 g of Sodium Sulfate. Pour over into new collection vial and rinse thoroughly with fresh solvent.

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. Sandwich the G1 disc between the two C9 discs.2. Weigh the sorbent into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup layering on top of the sorbent. No mixing is needed.

SolventsOptions:-Dichloromethane-(50:50) Hexane: Acetone

Sorbents- Sodium Sulfate (1 g)- C18 (0.8 g)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 12020 03:001010

Stage Solvent Volume (mL)

1 Water 15

2 Extraction Solvent 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

8420

Page 25: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PesticidesFrom Wet Food SampleFrom Wet Food Sample

Agriculture, Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight5 g

Q-DiscᵀᴹC9 + G1 + C9

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. Layer the M2 Q-Disc (textured side up) on top of the C9 Q-Disc. 2. Weigh the sorbents into the Q-Cup. Spread evenly across the sample by gentle tapping.3. Weigh the sample into the Q-Cup layering the sample on top of the salt. No mixing is required.

SolventsAcetonitrile

Sorbents- Sodium Sulfate (5 g)- Magnesium Sulfate (1,200 mg)- Primary Second Amine (50 mg)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 10020 02:00010

Stage Solvent Volume (mL)

1 Water 30

2 Acetonitrile 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

7921

Page 26: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extractionof Total Fat From Feed/Foodof Total Fat From Feed/Food

Agriculture, Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight≤ 5 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup. Gently mix the sample and sand in the Q-Cup.4. Dry Q-Cups with sample and sand in the oven for 1 hour at 100° C while tilted at an angle in ~2" aluminum pans or equivalent. This provides stability and enables easier visual identification that no sample has been lost.

SolventsPetroleum Ether

SorbentsWet Samples Only: Sand (use equal amount to sample size)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 14010 01:00010

2 13010 01:00010

Stage Solvent Volume (mL)

1 Water 15

2 Petroleum Ether 20

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

8622

Page 27: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Cannabis

23

Page 28: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of CannabinoidsFrom Cannabis PlantFrom Cannabis Plant

Agriculture, Cannabis, Food TesEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight1 g

Q-DiscᵀᴹM2 + C9

EquipmentEDGE40 mL collection vials - clear

NotesPossessing, using, distributing or selling marijuana or marijuana-based products constitute federal crimes in the United States, even where a state law decriminalizes or legalizes such activities. CEM Corporation produces instruments that are intended for use in testing laboratories and applications only where such use is permitted under applicable state/country law.

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. Layer the M2 Q-Disc (textured side up) on top of the C9 Q-Disc.2. Weigh the sample into the Q-Cup.

SolventsMethanol

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 10010 0:3055

Stage Solvent Volume (mL)

1 Methanol 20

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

6824

Page 29: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of MycotoxinsFrom Cannabis PlantFrom Cannabis Plant

Agriculture, Cannabis, Food TesEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight1.5 g

Q-DiscᵀᴹC4

EquipmentEDGE40 mL collection vials - clear

NotesPossessing, using, distributing or selling marijuana or marijuana-based products constitute federal crimes in the United States, even where a state law decriminalizes or legalizes such activities. CEM Corporation produces instruments that are intended for use in testing laboratories and applications only where such use is permitted under applicable state/country law.

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

SolventsMethanol May also use a percentage of Water:Methanol as needed

SorbentsNone

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 5015 02:0055

Stage Solvent Volume (mL)

1 Methanol 10

2 Methanol 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

7325

Page 30: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PesticidesFrom Cannabis PlantFrom Cannabis Plant

Agriculture, Cannabis, Food TesEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight1.5 g

Q-DiscᵀᴹM2 + C9

EquipmentEDGE40 mL collection vials - clear

NotesPossessing, using, distributing or selling marijuana or marijuana-based products constitute federal crimes in the United States, even where a state law decriminalizes or legalizes such activities. CEM Corporation produces instruments that are intended for use in testing laboratories and applications only where such use is permitted under applicable state/country law.

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. Layer the M2 Q-Disc on top of the C9 Q-Disc.2. Weigh the sorbent into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup layering the sample on top of the salt. No mixing is required.

Solvents1% Acetic Acid in Acetonitrile

SorbentsContents of Agilent QuEChERS SPE 15 mL Tube

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 4020 02:0055

Stage Solvent Volume (mL)

1 1% Acetic Acid in Acetonitrile 15

2 1% Acetic Acid in Acetonitrile 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

11126

Page 31: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Environmental & Regulatory

27

Page 32: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of SVOCsEnvironmental SamplesEnvironmental Samples

Environmental and RegulatoryEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight30 g

Q-DiscᵀᴹC4

EquipmentEDGE40 mL collection vials - clear

NotesThis is a general method for the miscellaneous extraction of environmental samples.

Procedure 1.Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2.Weigh the sample into the Q-Cup.

SolventsOptions:AcetoneDichloromethaneHexaneToluene

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 12015 00:001010

Stage Solvent Volume (mL)

1 Dichloromethane 10

2 Extraction Solvent 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

6728

Page 33: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of SVOCsEPA 3545AEPA 3545A

Environmental and RegulatoryEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight30 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesSee the CEM Application note: Extraction of Semi-Volatile Organic Compounds from Soil in Accordance with EPA 3545A if needed.

Procedure 1.Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2.Weigh the sample into the Q-Cup.

Solvents(50:50) Acetone: Hexane

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 12015 01:001010

Stage Solvent Volume (mL)

1 Acetone 10

2 (50:50) Acetone: Hexane 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

11329

Page 34: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PCB'sFrom Large PUF FiltersFrom Large PUF Filters

Environmental and Regulatory, IEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample WeightEntire Filter

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Assemble the Q-Cup™ with the Q-Disc™. 2. Compress and push the filter into the base of the Q-Cup.

SolventsHexane

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 11520 03:00155

Stage Solvent Volume (mL)

1 Hexane 10

2 Hexane 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

11030

Page 35: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PCB'sFrom Small PUF FiltersFrom Small PUF Filters

Environmental and Regulatory, IEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample WeightEntire Filter

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Assemble the Q-Cup™ with the Q-Disc™. 2. Compress and push the filter into the base of the Q-Cup.

SolventsHexane

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 10015 00:10150

Stage Solvent Volume (mL)

1 Hexane 10

2 Hexane 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

10931

Page 36: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PAHsFrom Wet Fish TissueFrom Wet Fish Tissue

Agriculture, Environmental and EDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight2 g

Q-DiscᵀᴹC9 + G1 + C9

EquipmentEDGE40 mL collection vials - clear

NotesIf there is moisture in collection vial after the extraction:-Using a glass funnel with glass wool and about 20 g of Sodium Sulfate. Pour over into new collection vial and rinse thoroughly with fresh solvent.

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. Sandwich the G1 disc between the two C9 discs.2. Weigh the sorbent into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup layering on top of the sorbent. No mixing is needed.

SolventsOptions:-Dichloromethane-(50:50) Hexane: Acetone

Sorbents- Sodium Sulfate (1 g)- C18 (0.8 g)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 12020 03:001010

Stage Solvent Volume (mL)

1 Water 15

2 Extraction Solvent 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

8432

Page 37: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Cleaning ofPUF FiltersPUF Filters

Environmental and Regulatory, IEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample WeightEntire Filter

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Assemble the Q-Cup™ with the Q-Disc™. 2. Compress and push the filter into the base of the Q-Cup.

SolventsStage 1: AcetoneStage 2: Hexane

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 11515 03:00150

2 11523 03:00200

Stage Solvent Volume (mL)

1 Acetone 15

2 Hexane 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

10833

Page 38: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PAHs FromSoil, Loam, or Clay

Environmental and RegulatoryEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight30 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

Solvents(50:50) Acetone: Hexane

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 12015 01:001010

Stage Solvent Volume (mL)

1 Acetone 10

2 Acetone 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtainrequired results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position andload method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

12134

Page 39: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PCBs FromSoil, Loam, or Clay

Environmental and RegulatoryEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight30 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

Solvents(50:50) Acetone: Hexane

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 12015 01:001010

Stage Solvent Volume (mL)

1 Acetone 10

2 Acetone 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtainrequired results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position andload method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

12235

Page 40: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of Dioxins FromSoil, Loam, or Clay

Environmental and RegulatoryEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight30 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

SolventsToluene

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 15015 03:001010

Stage Solvent Volume (mL)

1 Acetone 10

2 Acetone 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtainrequired results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position andload method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

12336

Page 41: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Food Testing

37

Page 42: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of SVOCsEnvironmental SamplesEnvironmental Samples

Environmental and RegulatoryEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight30 g

Q-DiscᵀᴹC4

EquipmentEDGE40 mL collection vials - clear

NotesThis is a general method for the miscellaneous extraction of environmental samples.

Procedure 1.Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2.Weigh the sample into the Q-Cup.

SolventsOptions:AcetoneDichloromethaneHexaneToluene

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 12015 00:001010

Stage Solvent Volume (mL)

1 Dichloromethane 10

2 Extraction Solvent 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

6738

Page 43: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of SVOCsEPA 3545AEPA 3545A

Environmental and RegulatoryEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight30 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesSee the CEM Application note: Extraction of Semi-Volatile Organic Compounds from Soil in Accordance with EPA 3545A if needed.

Procedure 1.Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2.Weigh the sample into the Q-Cup.

Solvents(50:50) Acetone: Hexane

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 12015 01:001010

Stage Solvent Volume (mL)

1 Acetone 10

2 (50:50) Acetone: Hexane 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

11339

Page 44: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction ofFilm PowderFilm Powder

Materials Science, Plastics, PolyEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight1 g

Q-DiscᵀᴹM1

EquipmentEDGE40 mL collection vials - clear

Notes*Plastics and polymers should only be extracted at temperatures below melting point range.

Expansion of the sample during extraction is common during polymer based extraction.

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

SolventsDichloromethane

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 *40-7015 10:0005

Stage Solvent Volume (mL)

1 Dichloromethane 10

2 Cyclohexane 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

9340

Page 45: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PigmentsFrom PlasticsFrom Plastics

Materials Science, Plastics, PolyEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight1 g

Q-DiscᵀᴹM1

EquipmentEDGE40 mL collection vials - clear

Notes*Plastics and polymers should only be extracted at temperatures below melting point range.

Expansion of the sample during extraction is common during polymer based extraction.

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

SolventsDichloromethane

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 *40-8015 10:0005

Stage Solvent Volume (mL)

1 Dichloromethane 10

2 Cyclohexane 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

9441

Page 46: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PesticidesFrom AvocadosFrom Avocados

Agriculture, Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight5 g

Q-DiscᵀᴹM2 + C9

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. Layer the M2 Q-Disc (textured side up) on top of the

C9 Q-Disc. 2. Weigh the sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup. Gently mix the sample and sand in the Q-Cup.

SolventsAcetonitrile

Sorbents- Sodium Sulfate (5 g)- Magnesium Sulfate (1,200 mg)- Primary Second Amine (50 mg)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 10020 01:00010

Stage Solvent Volume (mL)

1 Water 15

2 Acetonitrile 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

11542

Page 47: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PolyphenolsFrom BeansFrom Beans

Agriculture, Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight0.1 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

Solvents(50:50) Methanol: Water

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 8020 10:001010

Stage Solvent Volume (mL)

1 (50:50) Methanol: Water 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

12443

Page 48: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of FatFrom Brazil NutsFrom Brazil Nuts

Agriculture, Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight12 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup. Gently mix the sample and sand in the Q-Cup.

SolventsPetroleum Ether

SorbentsSand (12 g)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 17510 00:00010

Stage Solvent Volume (mL)

1 Petroleum Ether 20

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

12544

Page 49: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PolyphenolsFrom CacaoFrom Cacao

Agriculture, Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight0.1 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

Solvents(50:50) Methanol: Water

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 8020 10:001010

Stage Solvent Volume (mL)

1 (50:50) Methanol: Water 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

10745

Page 50: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of CannabinoidsFrom Cannabis PlantFrom Cannabis Plant

Agriculture, Cannabis, Food TesEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight1 g

Q-DiscᵀᴹM2 + C9

EquipmentEDGE40 mL collection vials - clear

NotesPossessing, using, distributing or selling marijuana or marijuana-based products constitute federal crimes in the United States, even where a state law decriminalizes or legalizes such activities. CEM Corporation produces instruments that are intended for use in testing laboratories and applications only where such use is permitted under applicable state/country law.

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. Layer the M2 Q-Disc (textured side up) on top of the C9 Q-Disc.2. Weigh the sample into the Q-Cup.

SolventsMethanol

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 10010 0:3055

Stage Solvent Volume (mL)

1 Methanol 20

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

6846

Page 51: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of MycotoxinsFrom Cannabis PlantFrom Cannabis Plant

Agriculture, Cannabis, Food TesEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight1.5 g

Q-DiscᵀᴹC4

EquipmentEDGE40 mL collection vials - clear

NotesPossessing, using, distributing or selling marijuana or marijuana-based products constitute federal crimes in the United States, even where a state law decriminalizes or legalizes such activities. CEM Corporation produces instruments that are intended for use in testing laboratories and applications only where such use is permitted under applicable state/country law.

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

SolventsMethanol May also use a percentage of Water:Methanol as needed

SorbentsNone

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 5015 02:0055

Stage Solvent Volume (mL)

1 Methanol 10

2 Methanol 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

7347

Page 52: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PesticidesFrom Cannabis PlantFrom Cannabis Plant

Agriculture, Cannabis, Food TesEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight1.5 g

Q-DiscᵀᴹM2 + C9

EquipmentEDGE40 mL collection vials - clear

NotesPossessing, using, distributing or selling marijuana or marijuana-based products constitute federal crimes in the United States, even where a state law decriminalizes or legalizes such activities. CEM Corporation produces instruments that are intended for use in testing laboratories and applications only where such use is permitted under applicable state/country law.

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. Layer the M2 Q-Disc on top of the C9 Q-Disc.2. Weigh the sorbent into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup layering the sample on top of the salt. No mixing is required.

Solvents1% Acetic Acid in Acetonitrile

SorbentsContents of Agilent QuEChERS SPE 15 mL Tube

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 4020 02:0055

Stage Solvent Volume (mL)

1 1% Acetic Acid in Acetonitrile 15

2 1% Acetic Acid in Acetonitrile 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

11148

Page 53: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of FatFrom CashewsFrom Cashews

Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight12 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup. Gently mix the sample and sand in the Q-Cup.

SolventsPetroleum Ether

SorbentsSand (12 g)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 17510 00:00010

Stage Solvent Volume (mL)

1 Petroleum Ether 20

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

7749

Page 54: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PAHsFrom Dry Fish SamplesFrom Dry Fish Samples

Agriculture, Food Testing, NutraEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight5 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™.

2. Weight the sorbent in the Q-Cup. Tap lightly to spread evenly. 3. Weigh the sample into the Q-Cup layering the sample on top of the salt. No mixing is required.

SolventsDichloromethane

SorbentsC18 (1.7 g)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 12020 03:001010

Stage Solvent Volume (mL)

1 Dichloromethane 15

2 Hexane 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

8550

Page 55: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PCBsFrom Dry Soy Meal & Fish MealFrom Dry Soy Meal & Fish Meal

Agriculture, Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight5 g

Q-DiscᵀᴹC1M1

EquipmentEDGE40 mL collection vials - clear

NotesLess than 5 grams of sample can be used with a ratio of 1:2 C18: Sample Either C1 or M1 Q-Disc can be used

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sorbent into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup layering the sample on top of the salt. No mixing is required.

SolventsOptions:Dichloromethane(50:50) Hexane: Acetone

SorbentsC18 (2.5 g)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 12020 03:001010

Stage Solvent Volume (mL)

1 Dichloromethane 15

2 Hexane 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

7051

Page 56: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PesticidesFrom Dry Spice SamplesFrom Dry Spice Samples

Agriculture, Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight1 g

Q-DiscᵀᴹC9 + G1 + C9

EquipmentEDGE40 mL collection vials - clear

NotesMatrix matched calibration curve required without the use of sorbents.

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. Sandwich the G1 disc between the two C9 discs.2. Weigh the sample into the Q-Cup.

SolventsAcetonitrile with 1% Acetic Acid

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 4015 03:00510

Stage Solvent Volume (mL)

1 Acetonitrile with 1% Acetic Acid 10

2 Acetonitrile with 1% Acetic Acid 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

8052

Page 57: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of FlavonoidsFrom EchinaceaFrom Echinacea

Food Testing, NutraceuticalsEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight2 g

Q-DiscᵀᴹM2 + C9

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. Layer the M2 Q-Disc (textured side up) on top of the C9 Q-Disc.2. Weigh the sample into the Q-Cup.

Solvents(70:30) Methanol: Water

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 6020 01:00020

Stage Solvent Volume (mL)

1 (70:30) Methanol: Water 10

2 (70:30) Methanol: Water 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

8753

Page 58: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of FatFrom Egg PowderFrom Egg Powder

Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight5 g

Q-DiscᵀᴹC4

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

SolventsToluene

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 17515 01:001010

Stage Solvent Volume (mL)

1 Toluene 15

2 Toluene 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

8354

Page 59: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of FatFrom Hazel NutsFrom Hazel Nuts

Agriculture, Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight12 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup. Gently mix the sample and sand in the Q-Cup.

SolventsPetroleum Ether

SorbentsSand (12 g)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 17510 00:00010

Stage Solvent Volume (mL)

1 Petroleum Ether 20

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

12755

Page 60: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PesticidesFrom HopsFrom Hops

Agriculture, Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight0.5 g

Q-DiscᵀᴹM2 + C9

EquipmentEDGE40 mL collection vials - clear

NotesCanadian Pesticide Mix 4 in LCMS grade obtained from SPEX CertiPrep.

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. Layer the M2 Q-Disc (textured side up) on top of the

C9 Q-Disc.2. Add sorbents to bottom of Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping. No mixing is required.

3. Weigh the sample into the Q-Cup. Add 5.0 g hops on top of sorbent bed. 4. Spike with 100 ug/mL Canadian Pesticide Mix 4 and leave in hood for approximately 30 minutes to evaporate solvent.

Solvents1% Acetic Acid in Acetonitrile

Sorbents-ACS grade magnesium sulfate anhydrous > 99.5% (1.25 g)-Supelᵀᴹ QuE Citrate (1.25 g)-Primary secondary amine (1.25 g)-Siliabond C18 (1.25 g)*Different sorbents can be used

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 10015 00:30510

Stage Solvent Volume (mL)

1 1% Acetic Acid in Acetonitrile 20

2 1% Acetic Acid in Acetonitrile 20

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

6256

Page 61: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PCBsFrom Large PUF Filters

Environmental and Regulatory, IEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample WeightEntire Filter

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Assemble the Q-Cup™ with the Q-Disc™. 2. Compress and push the filter into the base of the Q-Cup.

SolventsHexane

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 11520 03:00155

Stage Solvent Volume (mL)

1 Hexane 10

2 Hexane 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtainrequired results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position andload method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

11057

Page 62: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

ExtractionFrom MouthwashFrom Mouthwash

Personal Care (Health and BeauEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight5 g

Q-DiscᵀᴹC9 + G1 + C9

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure1. Assemble the Q-Cup™ with the Q-Disc™. Sandwich the G1 disc between the two C9 discs.

2. Weigh the sorbent into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping. 3. Weigh the sample into the Q-Cup layering on top of the sorbent. No mixing is needed.

SolventsHexane

SorbentsOptions:-Sodium Polyacrylate (5 g)-Diatomaceous Earth (5 g)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 4020 03:00010

Stage Solvent Volume (mL)

1 Hexane 10

2 Hexane 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

7658

Page 63: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of AdditivesFrom PETFrom PET

Plastics, Polymers, and OilsEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight1 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

Notes*Plastics and polymers should only be extracted at temperatures below melting point range.

Expansion of the sample during extraction is common during polymer based extraction.

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

Solvents(40:60) Acetone: Hexane

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 115*25 05:00510

Stage Solvent Volume (mL)

1 Acetone 15

2 Hexane 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

10259

Page 64: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of AdditivesFrom PigmentFrom Pigment

Plastics, Polymers, and OilsEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight1 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

SolventsDimethylformamide

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 10020 01:001010

Stage Solvent Volume (mL)

1 Dimethylformamide 10

2 Dimethylformamide 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

10560

Page 65: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of AdditivesFrom Polyethylene Powder

Plastics, Polymers, and OilsEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight1 g

Q-DiscᵀᴹM1 + C9

EquipmentEDGE40 mL collection vials - clear

Notes*Plastics and polymers should only be extracted at temperatures below melting point range.

Expansion of the sample during extraction is common during polymer based extraction.

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. Layer the C9 Q-Disc on top of the M1 Q-Disc.

2. Weigh the sample into the Q-Cup.

SolventsIsopropanol

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 *9030 15:00010

Stage Solvent Volume (mL)

1 Isopropanol 15

2 Isopropanol 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtainrequired results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position andload method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

9761

Page 66: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of Chemical ModifierFrom Polyethylene ResinFrom Polyethylene Resin

Plastics, Polymers, and OilsEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight0.1 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

Notes*Plastics and polymers should only be extracted at temperatures below melting point range.

Expansion of the sample during extraction is common during polymer based extraction.

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

Solvents(40:60) Acetone:Hexane

SorbentsNone

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 *6020 15:001010

Stage Solvent Volume (mL)

1 (40:60) Acetone:Hexane 10

2 (40:60) Acetone:Hexane 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

10362

Page 67: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of AdditivesFrom Polypropylene Powder

Materials Science, Plastics, PolyEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight1 g

Q-DiscᵀᴹC9 + G1 + C9

EquipmentEDGE40 mL collection vials - clear

Notes*Plastics and polymers should only be extracted at temperatures below melting point range.

Expansion of the sample during extraction is common during polymer based extraction.

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. Sandwich the G1 disc between the two C9 discs.

2. Weigh the sample into the Q-Cup.

SolventsIsopropanol

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 12030 15:00010

Stage Solvent Volume (mL)

1 Isopropanol 15

2 Isopropanol 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtainrequired results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position andload method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

9663

Page 68: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PhthalatesFrom Polyvinyl ChlorideFrom Polyvinyl Chloride

Materials Science, Plastics, PolyEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight1 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

Notes*Plastics and polymers should only be extracted at temperatures below melting point range.

Expansion of the sample during extraction is common during polymer based extraction.

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

Solvents(50:50) Isopropanol: Cyclohexane

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 *8020 10:00010

Stage Solvent Volume (mL)

1 (50:50) Isopropanol: Cyclohexane 10

2 (50:50) Isopropanol: Cyclohexane 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

9064

Page 69: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of NutrientsFrom Protein PowderFrom Protein Powder

Food Testing, NutraceuticalsEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight1 g

Q-DiscᵀᴹC3

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup. Gently mix the sample and sand in the Q-Cup.

Solvents(80:20) Methanol:Water

SorbentsSand (3 g)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 15015 00:001010

Stage Solvent Volume (mL)

1 (80:20) Methanol:Water 15

2 (80:20) Methanol:Water 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

8165

Page 70: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of FatFrom Pumpkin SeedsFrom Pumpkin Seeds

Agriculture, Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight12 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup. Gently mix the sample and sand in the Q-Cup.

SolventsPetroleum Ether

SorbentsSand (12 g)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 17510 00:00010

Stage Solvent Volume (mL)

1 Petroleum Ether 20

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

12666

Page 71: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of BrodifacoumFrom Rat Poison PowderFrom Rat Poison Powder

Agriculture, Industrial Hygiene, EDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight2 g

Q-DiscᵀᴹC4

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

SolventsMethanol

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 13015 02:001010

2 13015 02:001010

Stage Solvent Volume (mL)

1 Methanol 15

2 Methanol 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

9867

Page 72: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of AntioxidantsFrom ResinFrom Resin

Materials ScienceEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight1 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

SolventsMethanol

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 10015 02:00510

Stage Solvent Volume (mL)

1 Methanol 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

10668

Page 73: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PesticidesFrom RiceFrom Rice

Agriculture, Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight5 g

Q-DiscᵀᴹM2 + C9

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. Layer the M2 Q-Disc (textured side up) on top of the C9 Q-Disc. 2. Weigh the sorbents into the Q-Cup. Spread evenly across the sample by gentle tapping.3. Weigh the sample into the Q-Cup layering the sample on top of the salt. No mixing is required.

SolventsAcetonitile

Sorbents- Sodium Sulfate (5 g)- Magnesium Sulfate (1,200 mg)- Primary Second Amine (50 mg)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 10020 01:00010

Stage Solvent Volume (mL)

1 Acetonitile 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

11669

Page 74: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PCB'sFrom Small PUF FiltersFrom Small PUF Filters

Environmental and Regulatory, IEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample WeightEntire Filter

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Assemble the Q-Cup™ with the Q-Disc™. 2. Compress and push the filter into the base of the Q-Cup.

SolventsHexane

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 10015 00:10150

Stage Solvent Volume (mL)

1 Hexane 10

2 Hexane 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

10970

Page 75: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of DioxinsFrom SoilFrom Soil

EDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight10 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™.

2. Weigh the sample into the Q-Cup.

SolventsToluene

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 17515 01:001010

Stage Solvent Volume (mL)

1 Toluene 10

2 Toluene 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Bring to volume or evaporate as necessary for analysis.

10471

Page 76: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PesticidesFrom StrawberriesFrom Strawberries

Agriculture, Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight5 g

Q-DiscᵀᴹM2 + C9

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. Layer the M2 Q-Disc (textured side up) on top of the C9 Q-Disc. 2. Weigh the sorbents into the Q-Cup. Spread evenly across the sample by gentle tapping.3. Weigh the sample into the Q-Cup layering the sample on top of the salt. No mixing is required.

SolventsAcetonitrile

Sorbents- Sodium Sulfate (5 g)- Magnesium Sulfate (1,200 mg)- Primary Second Amine (50 mg)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 10020 01:00010

Stage Solvent Volume (mL)

1 Water 15

2 Acetonitrile 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

11472

Page 77: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PCBsFrom Wet Feed/ Food SamplesFrom Wet Feed/ Food Samples

Agriculture, Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight5 g

Q-DiscᵀᴹC9 + G1 + C9

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. Sandwich the G1 disc between the two C9 discs.2. Weigh the sorbent into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup layering on top of the sorbent. No mixing is needed.

SolventsDichloromethane

Sorbents-Sodium Sulfate (2.5 g)-C18 (2 g)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 13520 02:00510

Stage Solvent Volume (mL)

1 Water 15

2 Hexane 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

6973

Page 78: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of DioxinsFrom Wet Feed/ Food SamplesFrom Wet Feed/ Food Samples

Agriculture, Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight5 g

Q-DiscᵀᴹC4

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sorbent into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup layering on top of the sorbent. No mixing is needed.

SolventsDichloromethane

Sorbents-Sodium Sulfate (2.5 g)-C18 (2 g)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 15020 02:00510

Stage Solvent Volume (mL)

1 Water 15

2 Hexane 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

7274

Page 79: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PAHsFrom Wet Fish TissueFrom Wet Fish Tissue

Agriculture, Environmental and EDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight2 g

Q-DiscᵀᴹC9 + G1 + C9

EquipmentEDGE40 mL collection vials - clear

NotesIf there is moisture in collection vial after the extraction:-Using a glass funnel with glass wool and about 20 g of Sodium Sulfate. Pour over into new collection vial and rinse thoroughly with fresh solvent.

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. Sandwich the G1 disc between the two C9 discs.2. Weigh the sorbent into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup layering on top of the sorbent. No mixing is needed.

SolventsOptions:-Dichloromethane-(50:50) Hexane: Acetone

Sorbents- Sodium Sulfate (1 g)- C18 (0.8 g)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 12020 03:001010

Stage Solvent Volume (mL)

1 Water 15

2 Extraction Solvent 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

8475

Page 80: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PesticidesFrom Wet Food SampleFrom Wet Food Sample

Agriculture, Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight5 g

Q-DiscᵀᴹC9 + G1 + C9

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. Layer the M2 Q-Disc (textured side up) on top of the C9 Q-Disc. 2. Weigh the sorbents into the Q-Cup. Spread evenly across the sample by gentle tapping.3. Weigh the sample into the Q-Cup layering the sample on top of the salt. No mixing is required.

SolventsAcetonitrile

Sorbents- Sodium Sulfate (5 g)- Magnesium Sulfate (1,200 mg)- Primary Second Amine (50 mg)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 10020 02:00010

Stage Solvent Volume (mL)

1 Water 30

2 Acetonitrile 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

7976

Page 81: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PAH'sFrom XAD ResinFrom XAD Resin

Materials ScienceEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight15 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesBe sure that sample or sand is present in the Q-Cup when running this method. Sample size does matter, as you decrease the sample size you will build up more pressure within the cell.

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™.

2. Weigh the sample into the Q-Cup.

SolventsDichloromethane

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 18020 00:001010

Stage Solvent Volume (mL)

1 Dichloromethane 15

2 Dichloromethane 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

10177

Page 82: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PAH'sFrom Yerba MateFrom Yerba Mate

Food Testing, NutraceuticalsEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight0.5 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

SolventsOptions:(50:50) Hexane: AcetoneDichloromethaneAcetone

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 120-15015 01:00105

Stage Solvent Volume (mL)

1 Solvent used in extraction 10

2 Solvent used in extraction 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

8978

Page 83: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction ofMolybdenumMolybdenum

Geoscience and Mining, Metals EDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight2 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

SolventsAcetone

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 14015 01:00105

Stage Solvent Volume (mL)

1 Acetone 10

2 Acetone 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

9979

Page 84: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extractionof Total Fat From Feed/Foodof Total Fat From Feed/Food

Agriculture, Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight≤ 5 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup. Gently mix the sample and sand in the Q-Cup.4. Dry Q-Cups with sample and sand in the oven for 1 hour at 100° C while tilted at an angle in ~2" aluminum pans or equivalent. This provides stability and enables easier visual identification that no sample has been lost.

SolventsPetroleum Ether

SorbentsWet Samples Only: Sand (use equal amount to sample size)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 14010 01:00010

2 13010 01:00010

Stage Solvent Volume (mL)

1 Water 15

2 Petroleum Ether 20

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

8680

Page 85: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extractionof Total Fat From High-Fat Meatsof Total Fat From High-Fat Meats

Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight2 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup. Gently mix the sample and sand in the Q-Cup.4. Dry Q-Cups with sample and sand in the oven for 1 hour at 100° C while tilted at an angle in ~2" aluminum pans or equivalent. This provides stability and enables easier visual identification that no sample has been lost.

SolventsPetroleum Ether

SorbentsSand (2 g)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 14010 01:00010

2 13010 01:00010

Stage Solvent Volume (mL)

1 Water 10

2 Petroleum Ether 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

11281

Page 86: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extractionof Total Fat From Hotdogsof Total Fat From Hotdogs

Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight2 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup. Gently mix the sample and sand in the Q-Cup.4. Dry Q-Cups with sample and sand in the oven for 1 hour at 100° C while tilted at an angle in ~2" aluminum pans or equivalent. This provides stability and enables easier visual identification that no sample has been lost.

SolventsPetroleum Ether

SorbentsSand (2 g)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 14010 01:00010

2 13010 01:00010

Stage Solvent Volume (mL)

1 Water 10

2 Petroleum Ether 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

12082

Page 87: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extractionof Total Fat From Peanutsof Total Fat From Peanuts

Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight12 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup. Gently mix the sample and sand in the Q-Cup.4. Dry Q-Cups with sample and sand in the oven for 1 hour at 100° C while tilted at an angle in ~2" aluminum pans or equivalent. This provides stability and enables easier visual identification that no sample has been lost.

SolventsPetroleum Ether

SorbentsSand (12 g)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 15015 00:00010

Stage Solvent Volume (mL)

1 Petroleum Ether 20

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

7883

Page 88: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extractionof Total Fat From Pepperoniof Total Fat From Pepperoni

Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight2 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup. Gently mix the sample and sand in the Q-Cup.4. Dry Q-Cups with sample and sand in the oven for 1 hour at 100° C while tilted at an angle in ~2" aluminum pans or equivalent. This provides stability and enables easier visual identification that no sample has been lost.

SolventsPetroleum Ether

SorbentsSand (2 g)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 14010 01:00010

2 13010 01:00010

Stage Solvent Volume (mL)

1 Water 10

2 Petroleum Ether 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

11884

Page 89: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extractionof Total Fat From Potted Meatof Total Fat From Potted Meat

Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight2 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup. Gently mix the sample and sand in the Q-Cup.4. Dry Q-Cups with sample and sand in the oven for 1 hour at 100° C while tilted at an angle in ~2" aluminum pans or equivalent. This provides stability and enables easier visual identification that no sample has been lost.

SolventsPetroleum Ether

SorbentsSand (2 g)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 14010 01:00010

2 13010 01:00010

Stage Solvent Volume (mL)

1 Water 10

2 Petroleum Ether 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

11985

Page 90: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extractionof Total Fat From Protein Powderof Total Fat From Protein Powder

Food Testing, NutraceuticalsEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight2 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup. Gently mix the sample and sand in the Q-Cup.

SolventsPetroleum Ether

SorbentsSand (6 g)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 12015 02:00105

Stage Solvent Volume (mL)

1 Petroleum Ether 10

2 Petroleum Ether 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

8886

Page 91: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Cleaning ofPUF FiltersPUF Filters

Environmental and Regulatory, IEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample WeightEntire Filter

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Assemble the Q-Cup™ with the Q-Disc™. 2. Compress and push the filter into the base of the Q-Cup.

SolventsStage 1: AcetoneStage 2: Hexane

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 11515 03:00150

2 11523 03:00200

Stage Solvent Volume (mL)

1 Acetone 15

2 Hexane 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

10887

Page 92: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PAHs FromSoil, Loam, or Clay

Environmental and RegulatoryEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight30 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

Solvents(50:50) Acetone: Hexane

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 12015 01:001010

Stage Solvent Volume (mL)

1 Acetone 10

2 Acetone 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtainrequired results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position andload method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

12188

Page 93: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PCBs FromSoil, Loam, or Clay

Environmental and RegulatoryEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight30 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

Solvents(50:50) Acetone: Hexane

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 12015 01:001010

Stage Solvent Volume (mL)

1 Acetone 10

2 Acetone 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtainrequired results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position andload method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

12289

Page 94: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of Dioxins FromSoil, Loam, or Clay

Environmental and RegulatoryEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight30 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

SolventsToluene

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 15015 03:001010

Stage Solvent Volume (mL)

1 Acetone 10

2 Acetone 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtainrequired results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position andload method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

12390

Page 95: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction ofToothpasteToothpaste

Personal Care (Health and BeauEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight5 g

Q-DiscᵀᴹC9 + G1 + C9

EquipmentEDGE40 mL collection vials - clear

Notes* for sand- mix into sample* for Diatamaceous Earth- layer under sample, do not mix.

Procedure1. Assemble the Q-Cup™ with the Q-Disc™. Sandwich the G1 disc between the two C9 discs. 2. Weigh the sorbent/sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup*.

SolventsHexane

SorbentsOptions:Sand (15 g)Diatamaceous Earth (5 g)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 4020 03:00010

Stage Solvent Volume (mL)

1 Hexane 10

2 Hexane 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

12891

Page 96: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Hydrolysis & Extraction ofTotal Fat From ChocolateTotal Fat From Chocolate

Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight1 g

Q-DiscᵀᴹC9 + G1 + C9

EquipmentEDGE40 mL collection vials - clear

NotesStep 2- EDGE Method1. Assemble the Q-Cup™ with the Q-Disc™. Sandwich the G1 disc between the two C9 discs. 2. Place the Filter Paper into the Q-Cup.

ProcedureStep 1- Hydrolysis Method 1. Weigh out 1 g of milled or ground sample. 2. Add 1 mL Ethanol and 5 mL (25:11) Concentrated Hydrochloric Acid: Water. Heat for 1 hr at 100° C in a covered dish. 3. Add 5 mL of Ethanol. Filter the sample using Whatman 3 Filter Paper with a filter funnel, a vacuum pump, and filter flask. Wash any residual material from the beaker in the filter funnel using Ethanol. 4. Dry the filter paper in an oven at 100° C for 30 minutes. After the drying step, place Filter Paper into the Q-Cup and extract.

SolventsPetroleum Ether(25:11) Concentrated Hydrochloric Acid: Water

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 14010 01:00010

2 13010 01:00010

Stage Solvent Volume (mL)

1 Water 15

2 Petroleum Ether 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

13292

Page 97: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Hydrolysis & Extraction ofTotal Fat From CookiesTotal Fat From Cookies

Food TestingEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight1 g

Q-DiscᵀᴹC9 + G1 + C9

EquipmentEDGE40 mL collection vials - clear

NotesStep 2- EDGE Method1. Assemble the Q-Cup™ with the Q-Disc™. Sandwich the G1 disc between the two C9 discs. 2. Place the Filter Paper into the Q-Cup.

ProcedureStep 1- Hydrolysis Method 1. Weigh out 1 g of milled or ground sample. 2. Add 1 mL Ethanol and 5 mL (25:11) Concentrated Hydrochloric Acid: Water. Heat for 1 hr at 100° C in a covered dish. 3. Add 5 mL of Ethanol. Filter the sample using Whatman 3 Filter Paper with a filter funnel, a vacuum pump, and filter flask. Wash any residual material from the beaker in the filter funnel using Ethanol. 4. Dry the filter paper in an oven at 100° C for 30 minutes. After the drying step, place Filter Paper into the Q-Cup and extract.

SolventsPetroleum Ether(25:11) Concentrated Hyrdochloric Acid: Water

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 14010 01:00010

2 13010 01:00010

Stage Solvent Volume (mL)

1 Water 15

2 Petroleum Ether 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

13193

Page 98: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Geoscience & Mining

94

Page 99: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction ofMolybdenumMolybdenum

Geoscience and Mining, Metals EDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight2 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

SolventsAcetone

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 14015 01:00105

Stage Solvent Volume (mL)

1 Acetone 10

2 Acetone 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

9995

Page 100: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Industrial Hygiene

96

Page 101: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PCB'sFrom Large PUF FiltersFrom Large PUF Filters

Environmental and Regulatory, IEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample WeightEntire Filter

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Assemble the Q-Cup™ with the Q-Disc™. 2. Compress and push the filter into the base of the Q-Cup.

SolventsHexane

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 11520 03:00155

Stage Solvent Volume (mL)

1 Hexane 10

2 Hexane 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtainrequired results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position andload method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

11097

Page 102: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of BrodifacoumFrom Rat Poison PowderFrom Rat Poison Powder

Agriculture, Industrial Hygiene, EDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight2 g

Q-DiscᵀᴹC4

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

SolventsMethanol

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 13015 02:001010

2 13015 02:001010

Stage Solvent Volume (mL)

1 Methanol 15

2 Methanol 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

9898

Page 103: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PCB'sFrom Small PUF FiltersFrom Small PUF Filters

Environmental and Regulatory, IEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample WeightEntire Filter

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Assemble the Q-Cup™ with the Q-Disc™. 2. Compress and push the filter into the base of the Q-Cup.

SolventsHexane

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 10015 00:10150

Stage Solvent Volume (mL)

1 Hexane 10

2 Hexane 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

10999

Page 104: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Cleaning ofPUF FiltersPUF Filters

Environmental and Regulatory, IEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample WeightEntire Filter

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Assemble the Q-Cup™ with the Q-Disc™. 2. Compress and push the filter into the base of the Q-Cup.

SolventsStage 1: AcetoneStage 2: Hexane

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 11515 03:00150

2 11523 03:00200

Stage Solvent Volume (mL)

1 Acetone 15

2 Hexane 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

108100

Page 105: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Materials Science

101

Page 106: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction ofFilm PowderFilm Powder

Materials Science, Plastics, PolyEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight1 g

Q-DiscᵀᴹM1

EquipmentEDGE40 mL collection vials - clear

Notes*Plastics and polymers should only be extracted at temperatures below melting point range.

Expansion of the sample during extraction is common during polymer based extraction.

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

SolventsDichloromethane

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 *40-7015 10:0005

Stage Solvent Volume (mL)

1 Dichloromethane 10

2 Cyclohexane 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

93102

Page 107: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PigmentsFrom PlasticsFrom Plastics

Materials Science, Plastics, PolyEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight1 g

Q-DiscᵀᴹM1

EquipmentEDGE40 mL collection vials - clear

Notes*Plastics and polymers should only be extracted at temperatures below melting point range.

Expansion of the sample during extraction is common during polymer based extraction.

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

SolventsDichloromethane

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 *40-8015 10:0005

Stage Solvent Volume (mL)

1 Dichloromethane 10

2 Cyclohexane 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

94103

Page 108: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PCB'sFrom Large PUF FiltersFrom Large PUF Filters

Environmental and Regulatory, IEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample WeightEntire Filter

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Assemble the Q-Cup™ with the Q-Disc™. 2. Compress and push the filter into the base of the Q-Cup.

SolventsHexane

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 11520 03:00155

Stage Solvent Volume (mL)

1 Hexane 10

2 Hexane 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

110104

Page 109: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of AdditivesPolypropylene Powder

Materials Science, Plastics, PolyEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight1 g

Q-DiscᵀᴹC9 + G1 + C9

EquipmentEDGE40 mL collection vials - clear

Notes*Plastics and polymers should only be extracted at temperatures below melting point range.

Expansion of the sample during extraction is common during polymer based extraction.

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. Sandwich the G1 disc between the two C9 discs.

2. Weigh the sample into the Q-Cup.

SolventsIsopropanol

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 12030 15:00010

Stage Solvent Volume (mL)

1 Isopropanol 15

2 Isopropanol 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtainrequired results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position andload method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

96105

Page 110: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PhthalatesFrom Polyvinyl ChlorideFrom Polyvinyl Chloride

Materials Science, Plastics, PolyEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight1 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

Notes*Plastics and polymers should only be extracted at temperatures below melting point range.

Expansion of the sample during extraction is common during polymer based extraction.

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

Solvents(50:50) Isopropanol: Cyclohexane

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 *8020 10:00010

Stage Solvent Volume (mL)

1 (50:50) Isopropanol: Cyclohexane 10

2 (50:50) Isopropanol: Cyclohexane 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

90106

Page 111: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of BrodifacoumFrom Rat Poison PowderFrom Rat Poison Powder

Agriculture, Industrial Hygiene, EDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight2 g

Q-DiscᵀᴹC4

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

SolventsMethanol

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 13015 02:001010

2 13015 02:001010

Stage Solvent Volume (mL)

1 Methanol 15

2 Methanol 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

98107

Page 112: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of AntioxidantsFrom ResinFrom Resin

Materials ScienceEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight1 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

SolventsMethanol

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 10015 02:00510

Stage Solvent Volume (mL)

1 Methanol 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

106108

Page 113: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PCB'sFrom Small PUF FiltersFrom Small PUF Filters

Environmental and Regulatory, IEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample WeightEntire Filter

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Assemble the Q-Cup™ with the Q-Disc™. 2. Compress and push the filter into the base of the Q-Cup.

SolventsHexane

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 10015 00:10150

Stage Solvent Volume (mL)

1 Hexane 10

2 Hexane 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

109109

Page 114: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PAH'sFrom XAD ResinFrom XAD Resin

Materials ScienceEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight15 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesBe sure that sample or sand is present in the Q-Cup when running this method. Sample size does matter, as you decrease the sample size you will build up more pressure within the cell.

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™.

2. Weigh the sample into the Q-Cup.

SolventsDichloromethane

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 18020 00:001010

Stage Solvent Volume (mL)

1 Dichloromethane 15

2 Dichloromethane 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

101110

Page 115: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Cleaning ofPUF FiltersPUF Filters

Environmental and Regulatory, IEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample WeightEntire Filter

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Assemble the Q-Cup™ with the Q-Disc™. 2. Compress and push the filter into the base of the Q-Cup.

SolventsStage 1: AcetoneStage 2: Hexane

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 11515 03:00150

2 11523 03:00200

Stage Solvent Volume (mL)

1 Acetone 15

2 Hexane 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

108111

Page 116: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Metals & Alloys

112

Page 117: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction ofMolybdenumMolybdenum

Geoscience and Mining, Metals EDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight2 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

SolventsAcetone

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 14015 01:00105

Stage Solvent Volume (mL)

1 Acetone 10

2 Acetone 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

99113

Page 118: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Nutraceuticals

114

Page 119: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PAHsFrom Dry Fish SamplesFrom Dry Fish Samples

Agriculture, Food Testing, NutraEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight5 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™.

2. Weight the sorbent in the Q-Cup. Tap lightly to spread evenly. 3. Weigh the sample into the Q-Cup layering the sample on top of the salt. No mixing is required.

SolventsDichloromethane

SorbentsC18 (1.7 g)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 12020 03:001010

Stage Solvent Volume (mL)

1 Dichloromethane 15

2 Hexane 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

85115

Page 120: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of FlavonoidsFrom EchinaceaFrom Echinacea

Food Testing, NutraceuticalsEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight2 g

Q-DiscᵀᴹM2 + C9

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. Layer the M2 Q-Disc (textured side up) on top of the C9 Q-Disc.2. Weigh the sample into the Q-Cup.

Solvents(70:30) Methanol: Water

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 6020 01:00020

Stage Solvent Volume (mL)

1 (70:30) Methanol: Water 10

2 (70:30) Methanol: Water 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

87116

Page 121: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of NutrientsFrom Protein PowderFrom Protein Powder

Food Testing, NutraceuticalsEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight1 g

Q-DiscᵀᴹC3

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup. Gently mix the sample and sand in the Q-Cup.

Solvents(80:20) Methanol:Water

SorbentsSand (3 g)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 15015 00:001010

Stage Solvent Volume (mL)

1 (80:20) Methanol:Water 15

2 (80:20) Methanol:Water 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

81117

Page 122: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PAHsFrom Wet Fish TissueFrom Wet Fish Tissue

Agriculture, Environmental and EDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight2 g

Q-DiscᵀᴹC9 + G1 + C9

EquipmentEDGE40 mL collection vials - clear

NotesIf there is moisture in collection vial after the extraction:-Using a glass funnel with glass wool and about 20 g of Sodium Sulfate. Pour over into new collection vial and rinse thoroughly with fresh solvent.

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. Sandwich the G1 disc between the two C9 discs.2. Weigh the sorbent into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup layering on top of the sorbent. No mixing is needed.

SolventsOptions:-Dichloromethane-(50:50) Hexane: Acetone

Sorbents- Sodium Sulfate (1 g)- C18 (0.8 g)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 12020 03:001010

Stage Solvent Volume (mL)

1 Water 15

2 Extraction Solvent 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

84118

Page 123: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PAH'sFrom Yerba MateFrom Yerba Mate

Food Testing, NutraceuticalsEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight0.5 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

SolventsOptions:(50:50) Hexane: AcetoneDichloromethaneAcetone

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 120-15015 01:00105

Stage Solvent Volume (mL)

1 Solvent used in extraction 10

2 Solvent used in extraction 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

89119

Page 124: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extractionof Total Fat From Protein Powderof Total Fat From Protein Powder

Food Testing, NutraceuticalsEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight2 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup. Gently mix the sample and sand in the Q-Cup.

SolventsPetroleum Ether

SorbentsSand (6 g)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 12015 02:00105

Stage Solvent Volume (mL)

1 Petroleum Ether 10

2 Petroleum Ether 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

88120

Page 125: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Personal Care Products

121

Page 126: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

ExtractionFrom MouthwashFrom Mouthwash

Personal Care (Health and BeauEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight5 g

Q-DiscᵀᴹC9 + G1 + C9

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure1. Assemble the Q-Cup™ with the Q-Disc™. Sandwich the G1 disc between the two C9 discs.

2. Weigh the sorbent into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping. 3. Weigh the sample into the Q-Cup layering on top of the sorbent. No mixing is needed.

SolventsHexane

SorbentsOptions:-Sodium Polyacrylate (5 g)-Diatomaceous Earth (5 g)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 4020 03:00010

Stage Solvent Volume (mL)

1 Hexane 10

2 Hexane 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

76122

Page 127: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction ofToothpasteToothpaste

Personal Care (Health and BeauEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight5 g

Q-DiscᵀᴹC9 + G1 + C9

EquipmentEDGE40 mL collection vials - clear

Notes* for sand- mix into sample* for Diatamaceous Earth- layer under sample, do not mix.

Procedure1. Assemble the Q-Cup™ with the Q-Disc™. Sandwich the G1 disc between the two C9 discs. 2. Weigh the sorbent/sand into the Q-Cup. Spread evenly across the base of the Q-Cup by gentle tapping.3. Weigh the sample into the Q-Cup*.

SolventsHexane

SorbentsOptions:Sand (15 g)Diatamaceous Earth (5 g)

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 4020 03:00010

Stage Solvent Volume (mL)

1 Hexane 10

2 Hexane 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

128123

Page 128: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Plastics, Polymers, & Oils

124

Page 129: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction ofFilm PowderFilm Powder

Materials Science, Plastics, PolyEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight1 g

Q-DiscᵀᴹM1

EquipmentEDGE40 mL collection vials - clear

Notes*Plastics and polymers should only be extracted at temperatures below melting point range.

Expansion of the sample during extraction is common during polymer based extraction.

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

SolventsDichloromethane

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 *40-7015 10:0005

Stage Solvent Volume (mL)

1 Dichloromethane 10

2 Cyclohexane 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

93125

Page 130: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PigmentsFrom PlasticsFrom Plastics

Materials Science, Plastics, PolyEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight1 g

Q-DiscᵀᴹM1

EquipmentEDGE40 mL collection vials - clear

Notes*Plastics and polymers should only be extracted at temperatures below melting point range.

Expansion of the sample during extraction is common during polymer based extraction.

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

SolventsDichloromethane

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 *40-8015 10:0005

Stage Solvent Volume (mL)

1 Dichloromethane 10

2 Cyclohexane 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

94126

Page 131: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of AdditivesFrom PETFrom PET

Plastics, Polymers, and OilsEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight1 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

Notes*Plastics and polymers should only be extracted at temperatures below melting point range.

Expansion of the sample during extraction is common during polymer based extraction.

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

Solvents(40:60) Acetone: Hexane

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 115*25 05:00510

Stage Solvent Volume (mL)

1 Acetone 15

2 Hexane 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

102127

Page 132: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of AdditivesFrom PigmentFrom Pigment

Plastics, Polymers, and OilsEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight1 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

NotesNone

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

SolventsDimethylformamide

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 10020 01:001010

Stage Solvent Volume (mL)

1 Dimethylformamide 10

2 Dimethylformamide 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

105128

Page 133: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of AdditivesFrom Polyethylene Powder

Plastics, Polymers, and OilsEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight1 g

Q-DiscᵀᴹM1 + C9

EquipmentEDGE40 mL collection vials - clear

Notes*Plastics and polymers should only be extracted at temperatures below melting point range.

Expansion of the sample during extraction is common during polymer based extraction.

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. Layer the C9 Q-Disc on top of the M1 Q-Disc.

2. Weigh the sample into the Q-Cup.

SolventsIsopropanol

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 *9030 15:00010

Stage Solvent Volume (mL)

1 Isopropanol 15

2 Isopropanol 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtainrequired results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position andload method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

97129

Page 134: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of Chemical ModifierFrom Polyethylene ResinFrom Polyethylene Resin

Plastics, Polymers, and OilsEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight0.1 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

Notes*Plastics and polymers should only be extracted at temperatures below melting point range.

Expansion of the sample during extraction is common during polymer based extraction.

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

Solvents(40:60) Acetone:Hexane

SorbentsNone

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 *6020 15:001010

Stage Solvent Volume (mL)

1 (40:60) Acetone:Hexane 10

2 (40:60) Acetone:Hexane 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

103130

Page 135: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of AdditivesFrom Polypropylene Powder

Materials Science, Plastics, PolyEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight1 g

Q-DiscᵀᴹC9 + G1 + C9

EquipmentEDGE40 mL collection vials - clear

Notes*Plastics and polymers should only be extracted at temperatures below melting point range.

Expansion of the sample during extraction is common during polymer based extraction.

Procedure1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. Sandwich the G1 disc between the two C9 discs.

2. Weigh the sample into the Q-Cup.

SolventsIsopropanol

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 12030 15:00010

Stage Solvent Volume (mL)

1 Isopropanol 15

2 Isopropanol 15

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtainrequired results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position andload method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

96131

Page 136: Automated Extraction System Method Note Compendiumcem.com › ... › literature › MetNote_EDGE_Compendium.pdf · Compendium Updated February 20, 2019. Agriculture Avocados 2 Beans

Extraction of PhthalatesFrom Polyvinyl ChlorideFrom Polyvinyl Chloride

Materials Science, Plastics, PolyEDGE

Note: Temperature and hold times may vary depending on the sample and analytes of interest.

Heating Program

Sample Weight1 g

Q-DiscᵀᴹC1

EquipmentEDGE40 mL collection vials - clear

Notes*Plastics and polymers should only be extracted at temperatures below melting point range.

Expansion of the sample during extraction is common during polymer based extraction.

Procedure 1. Homogenize, mill or cryomill the sample. Assemble the Q-Cup™ with the Q-Disc™. 2. Weigh the sample into the Q-Cup.

Solvents(50:50) Isopropanol: Cyclohexane

SorbentsN/A

Wash Program

Stage Temp (°C)Top Add (mL) Hold (mm:ss)Rinse (mL)Bottom Add (mL)

1 *8020 10:00010

Stage Solvent Volume (mL)

1 (50:50) Isopropanol: Cyclohexane 10

2 (50:50) Isopropanol: Cyclohexane 10

General Guidelinesa) This procedure is a reference point for sample extraction using a CEM system and may need to be modified or changed to obtain required results on your sample.b) Wear hand, eye, and body protection when handling organic solvents.c) Any of CEM's collection vials/centrifuge tubes in the correct size for the rack owned may be used.d) Verify needed solvents are loaded onto the system with sufficient volume. Load the sample rack into the EDGE. Select position and load method. Press play and add any any sample ID's, etc. as needed.e) Bring to volume or evaporate as necessary for analysis.

90132