AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT PREPARATIVE AND ANALYTICAL SCALES KATIE SCHAEFER, PH.D. AND PETER WINSHIP, PH.D.

Upload: others

Post on 02-Aug-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT PREPARATIVE AND ANALYTICAL SCALESKATIE SCHAEFER, PH.D. AND PETER WINSHIP, PH.D.

Page 2: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

PART 1.TELEDYNE CETAC TECHNOLOGIES AND BIO-RADLABORATORIES:THE COMPANIES AND THE PRESENTERS

Page 3: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

KATIE SCHAEFER, PH.D.BIO-RAD LABORATORIES

Based in Northern California, Katie Schaefer is the Global Product Manager forlaboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsiblefor the NGC chromatography system and the accompanying ChromLab software. Katiebegan her research career as an undergraduate student and has over 12 years ofexperience in chromatographic techniques. Katie received her Ph.D. in Biochemistryand Molecular Biophysics from the California Institute of Technology in 2015, studyingthe biological implications of DNA-mediated charge transport. She began her careerwith Bio-Rad as a Proteomics Field Applications Scientist, training and supportingcustomers in chromatography, western blotting, digital imaging, and Bio-Plex assays.She now supports the NGC chromatography system and ChromLab software at Bio-Radheadquarters as a product manager and a technical application expert.

Page 4: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

BIO-RAD LABORATORIESFounded in 1952

David and Alice Schwartz began Bio-Rad in a Quonset hut

Goal was to advance scientific discovery

Went public in 1966 with $1 million in sales

Global leader in clinical diagnostics and life science research

Located in Hercules, CA

Page 5: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

PETER WINSHIP, PH.D.TELEDYNE CETAC TECHNOLOGIES

Based in the United Kingdom, Peter Winship is a Technical Product Manager atTeledyne CETAC Technologies and is responsible for the ASX-7000 Series automationproduct line. He joined the company in 2011. Pete has a degree in Chemistry, an MScin Analytical and Pharmaceutical Science and a Ph.D. in Analytical Chemistry. Pete hasworked with ICP-MS and ICP-OES instrumentation since 2001 and he has furtherexperience in the oil industry and biological and geological testing. Pete spent 5 yearsas a Senior Analyst at the Medical Research Council in the UK where he operated ICP-MS and ICP-OES instrumentation to study clinical and biological samples in support ofa number of research projects. He has been an author of, or contributor to, a numberof peer reviewed publications and has presented widely at international conferences.

Page 6: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

TELEDYNE CETAC TECHNOLOGIESFounded in 1985, Ames, Iowa, USA

Began with introduction of the U5000 ultrasonic nebulizer.

First autosampler introduced in the early 90’s, this laid the foundation for the industry’s leader in autosamplers with a reputation for quality and reliability.

Sold to Steve Dwyer in 2000.

Acquired by Teledyne Technologies Inc. in 2013.

Today Teledyne CETAC Technologies manufactures automation and sampling handling/introduction systems, that support a gamut of analytical techniques, in Omaha, Nebraska, USA.

Page 7: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

PART 2.PROTEIN/ANTIBODY PURIFICATION, AUTOMATION AND INSTRUMENTATION

K

Page 8: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

PROTEIN PURIFICATION: AN INTRODUCTION

Goal 1: Isolate one protein of interest from a biological expression platform

o Protein expression: E. coli, yeast, cell cultures, hybridoma cell lines

Goal 2: Purify large quantities of high purity proteins

Goal 3: Utilize these proteins in the research, development, and therapeutic treatment

Chromatographic protein purification workflows typically utilize multiple column chemistries

Preparative – small to large scale production

Analytical – high resolution separation and purity confirmation

Capture•Tagged Protein•Native Protein

Intermediate•Ion exchange•Desalting•Hydrophobic

Interaction

Polishing•High resolution

resins•Size Exclusion•Mixed Mode

ConfirmPurity

Purit

y

Procedure time

K

Page 9: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

MONOCLONAL IGG ANTIBODIES

Highly specific biomolecule from the immune response to an antigen

Highly favorable for scientific applications such as western blotting, ELISA, and flow cytometry

Monoclonal IgG antibodies are produced via hybridoma cell lines, phage display technologies and single-sorted human B cells

The high specificity of IgG antibodies makes them exceptional for therapeutic purposes - targeting a very specific antigen representative of a diseased state and eliciting an immune response

o Humira (adalimumab): Treatment for arthritis, Crohn’s disease, ulcerative colitis

o Rituxan (rituximab): Treatment for certain autoimmune diseases and cancers

K

Page 10: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

MONOCOLONAL ANTIBODY DEVELOPMENT (TWO WORKFLOWS)

Screening of Antibody Candidates – Express IgG antibody candidates on a small scale for efficacy testing

o ASX-500 Series autosampler + NGC with Sample Pump + Inlet Valve + Fraction Collector

Confirm Purity of Production – Analytical analysis for sample purity and integrity

o MVX-7100 µL Workstation + NGC with analytical flow cell

K

Page 11: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

BIO-RAD AND THE NGC CHROMATOGRAPHY SYSTEM

Customizable chromatography platform to meet your current and future purification needs

The system can change with your changing application needs

Medium-pressure system allows some for reversed-phase applications

Flexible fraction collection options

Intuitive ChromLab software

NGC Discover Pro

NGC Quest

K

Page 12: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

TELEDYNE CETAC TECHNOLOGIES AND AUTOMATION: WHY AUTOMATE PROTEIN PURIFICATION?

Flexible automation to meet the requirements of the protein purification workflows

o mL sample introduction scale (ASX-500 Series autosampler)

o µL sample introduction scale (MVX-7100 µL Workstation)

Configurable, robust and reliable

Undertake unattended chromatography instrument operation

o 24 hours a day, 7 days per week operation (improve efficiency)

o Allow purification of 84 samples in two days

Accurate and precise sample uptake and introduction

o Minimize sample introduction error

MVX-7100 µL Workstation

ASX-500 Series Autosampler

P

Page 13: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

PART 3.PREPARATIVE PROTEIN PURIFICATION WORKFLOW

K

Page 14: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

MONOCOLONAL ANTIBODY DEVELOPMENT (TWO WORKFLOWS)

K

Screening of Antibody Candidates – Express IgG antibody candidates on a small scale for efficacy testing

o ASX-500 Series autosampler + NGC with Sample Pump+ Inlet Valve + Fraction Collector

Confirm Purity of Production – Analytical analysis for sample purity and integrity

o MVX-7100 µL Workstation + NGC with analytical flow cell

Page 15: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

PREPARATIVE PROTEIN PURIFICATION WORKFLOW

K

Sample preparation/pre-

treatment

Transfer to ASX-500 Series automation

Sample uptake and introduction

to the NGC instrument

Tandem chromatography

(Protein A to Desalting)

Fraction collection

Repeated tandem

chromatography

Data generation Purity confirmation

Page 16: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

SCREENING OF ANTIBODY CANDIDATES -- SAMPLE PRODUCTION

K

Page 17: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

PREPARATIVE PURIFICATION: SAMPLES TRANSFERRED TO ASX-500 SERIES AUTOMATION

ASX-280 autosampler (2 sample rack capacity)

ASX-560 autosampler (4 sample rack capacity)

Robust and reliable XYZ automation for 10’s and 100’s mL sample volumes

Numerous sample tube/bottle options: 7 mL to 250 mL sizes

Rinse station (peristaltic pump filled/drained)

Resistant to chaotropic agents, organic solvents, and detergents

Sample probe connected to the NGC sample inlet valve

ASX-560 Autosampler

ASX-280 Autosampler

P

Page 18: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

ASX-500 SERIES AUTOMATION AND NGC INSTRUMENT: INTERFACE

AScript Operating Software

P

Page 19: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

SAMPLE UPTAKE AND INTRODUCTION PROCESS

Method begins for column equilibration and to zero baseline

In method, sample loading phase sends trigger signal received by ASX-500 Series autosampler

Sample probe moves to the appropriate sample/standard position

NGC Sample pump turns ON

Sample pump draws up sample and loads directly to column

Air sensor triggers and terminates the loading process

Resume remainder of purification method

ASX-500 Series probe washed in preparation for next sample

Air Sensor

P

Page 20: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

TANDEM CHROMATOGRAPHY: PROTEIN A TO DESALTING

K

IgG Protein

Protein A Elution Buffer

Protein ACaptures IgG

Elutes at pH 3.0

DesaltingSize Separation

pH 7.3

FractionspH 7.3

Page 21: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

NGC AND ASX FOR AUTOMATED TANDEM CHROMATOGRAPHY

Desalting column equilibration

Protein A column equilibration

Clean system for next sample

from ASX

Protein A column wash

Sample Application

ASX + Inlet Valve + Sample Pump

Protein A elution+ desalting

column sample application

Desalting ColumnElution and

Fraction Collection

K

Repeat for all samples on ASX

Page 22: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

OVERLAY OF AUTOMATED IGG PURIFICATION SERIES WITH ASX

“Method is easy to use and truly hands-off”A

280

nm (m

AU)

Conductivity mS/cm

Volume

pH 7.3

Conductivity 22 mS/cm

KA

280

nm (m

AU)

Page 23: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

NGC FRACTION COLLECTORCollects into multiple sized vessels within a method from 96 well plates to 250ml bottles

Peltier Cooling option

Supports flow rates up to 200 ml/min

High quality racks

Factory calibrated

Triggers for collection including slope, pH, & conductivity

NGC will support up to two fraction collectors per machine

Designed to fit into both swing door and sliding door deli-fridges

Tube Type Tube/Rack FC Capacity

13 mm 96 384

16 mm 75 300

15 ml (18 mm) 70 280

50 ml (30 mm) 27 108

96-well 2 8

2 ml microtube 96 384

250 ml bottle 4 16

Prep-Rack 20 40

K

Page 24: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

SAMPLE PURITY WITHOUT CROSS-CONTAMINATION

Purity is >95% No cross contamination detected by LC-MS

Ab1

Ab2

Ab3

Typical yield is 2-6 mg recovered which is sufficient to enable secondary screening procedure (functional assay)

K

Page 25: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

AUTOMATED SCREENING SOLUTION

We have implemented a robust purification method for Antibodies candidate

The method is fully automated and can handle 84 Ab over a weekend

The described method is easy to implement

Can easily be adapted to more elaborate purification scheme (multi-dimensional)

P

Page 26: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

PART 4.ANALYTICAL SEPARATION/PURIFICATION WORKFLOW

P

Page 27: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

Screening of Antibody Candidates – Express IgG antibody candidates on a small scale for efficacy testing

o ASX-500 Series autosampler + NGC with Sample Pump + Inlet Valve + Fraction Collector

Confirm Purity of Production – Analytical analysis for sample purity and integrity

o MVX-7100 µL Workstation + NGC with analytical flow cell

MONOCOLONAL ANTIBODY DEVELOPMENT (TWO WORKFLOWS)

K

Page 28: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

ANALYTICAL SEPARATION/PURIFICATION WORKFLOW

K

Sample preparation/pre-

treatment

Transfer to MVX-7100 µL

Workstation

Sample uptake and introduction

to the NGC instrument

High-resolution purification on

NGC

Small-scale fraction

collection

Repeated High-resolution

purification on NGC

Data generation Purity confirmation

Page 29: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

SAMPLES TRANSFERRED TO MVX-7100 AUTOMATION

Accurate, precise and robust XYZ automationo < 5 µL to 950 µL sample volumes

Different sample vial and well plate optionso VT54, 96 well plates and 384 well plates

Fine control of sample aliquot uptake

Temperature control (4°C to 40°C)

Rinse station

Resistant to chaotropic agents, organic solvents, and detergents

Plates can be foil covered to prevent sample evaporationMVX-7100 µL Workstation

P

Page 30: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

MVX-7100 AUTOMATION AND NGC INSTRUMENT: INTERFACE

CETAC Workstation Operating Software

P

Page 31: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

SAMPLE UPTAKE AND INTRODUCTION PROCESS

Trigger signal received by MVX-7100

Sample probe moves to the sample/standard position

Syringe pump pulls sample onto MVX-7100 sample loop

MVX-7100 is connected to the inlet valve of NGC

Sample loaded into mobile phase through NGC loop port and sent to column

Remaining method resumed

P

Page 32: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

CETAC WORKSTATION SOFTWARE

Operating software for the MVX-7100 µL Workstation

Select sample locations, sample injection volume and uptake flow rates

Use the CETAC Workstation software to control the temperature of the samples and to configure the rinsing processes

Use the pre-configured operating methods for full sample loop and partial sample loop sample injections

P

Page 33: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

CETAC WORKSTATION SOFTWARE

Full sample loop injection

o Completely fill a sample loop with sample (e.g. 100 µL, 500 µL)

o Slight excess of sample on either side of the sample loop sent to waste

o Example: 27 µL of sample used to fill 25 µL sample loop with 2 µL sample excess

Partial sample loop injection

o Partially fill a sample loop with sample (e.g. 10 µL of sample in a 500 µL loop)

o Buffer solution on either side of the sample to ensure the loop is full of liquid

o Particularly applied to very low sample volumes

Full Sample Loop Uptake

Partial Sample Loop Uptake

P

Page 34: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

DATA FROM MVX-7100 : 48 RUNS

K

Full Sample Loop Uptake

Varied concentrations of a standard25 µl full loop injection

Page 35: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

MVX-7100 10 UL INJECTIONS: SEC STANDARD ON ENRICH 650

K

Partial Sample Loop Uptake

A, thyroglobulin; B, gamma globulin; C, ovalbumin; D, myoglobin; E, vitamin B12.

A

BC

D

E

Gel Filtration Standard10 µl partial loop uptake

Page 36: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

PART 5.CONCLUSIONS

Page 37: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

CONCLUDING COMMENTS SLIDE

Teledyne CETAC Technologies automation can easily streamline the protein/antibody purification workflows of the Bio-Rad NGC instrument.

o 24 hour, 7 days per week unattended operation

o Maintain accuracy and precision of sample introduction

o Automation to meet the full sample volume range requirements

Use the Bio-Rad NGC chromatography system for separation and purification of protein/antibody species.

o Intuitive system that adapts and grows with your purification needs

o Automate your purification for increased sample consistency and reproducibility

o NGC platform can be used for preparative as well as analytical separations

Page 38: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

THANK YOU FOR YOUR ATTENTION

Katie Schaefer, Ph.D.

o Global Product Manager: Laboratory Scale Chromatography Systems

o [email protected]

o www.bio-rad.com/ngc

Peter Winship, Ph.D.

o Technical Product Manager: ASX-7000 Series Automation

o [email protected]

o www.teledynecetac.com

Page 39: AUTOMATE YOUR ANTIBODY PURIFICATION WORKFLOW AT ... · laboratory-scale chromatography systems at Bio-Rad Laboratories. She is responsible for the NGC chromatography system and the

QUESTIONS??????