assessment of trabecular bone microstructure using dental ... · 7 table of contents chapter 1...

137
Assessment of Trabecular Bone Microstructure using Dental Cone Beam CT Norliza Ibrahim

Upload: others

Post on 10-Jun-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

  • Assessment of

    Trabecular Bone Microstructure

    using Dental Cone Beam CT

    Norliza Ibrahim

  • 2

    The studies in this thesis were conducted at the section of Oral Radiology of the Academic

    Centre for Dentistry Amsterdam (ACTA), Vrije University Amsterdam, The Netherlands.

    The author was financially supported with a scholarship by the University of Malaya, Kuala

    Lumpur, Malaysia.

    Financial support for printing:

    - ACTA Graduate School of Dentistry

    - Oral Radiology Foundation Amsterdam (ORFA)

    © Norliza Ibrahim, Amsterdam, 2014. All rights reserved.

    Cover design: Norliza Ibrahim

    Printed by Gildeprint Drukkerijen, Enschede.

    ISBN: 9789461086082

  • 3

    VRIJE UNIVERSITEIT

    Assessment of

    Trabecular Bone Microstructure

    using Dental Cone Beam CT

    ACADEMISCH PROEFSCHRIFT

    ter verkrijging van de graad Doctor aan

    de Vrije Universiteit Amsterdam,

    op gezag van de rector magnificus

    prof.dr. F.A. van der Duyn Schouten,

    in het openbaar te verdedigen

    ten overstaan van de promotiecommissie

    van de Faculteit der Tandheelkunde

    op vrijdag 21 maart 2014 om 9.45 uur

    in de aula van de universiteit,

    De Boelelaan 1105

    door

    Norliza Binti Ibrahim

    geboren te Selangor, Maleisië

  • 4

    promotor: prof.dr. P.F. van der Stelt

    co promotor: dr. B.A. Hassan

  • 5

    To my parents,

    Haji Ibrahim Sharif and

    Hajjah Ramlah Arifin.

  • 6

  • 7

    Table of Contents

    CHAPTER 1 INTRODUCTION 9

    CHAPTER 2 ACCURACY OF TRABECULAR BONE

    MICROSTRUCTURAL MEASUREMENT USING CONE

    BEAM CT DATASETS

    25

    CHAPTER 3 BONE QUALITY EVALUATION AT DENTAL IMPLANT

    SITE USING MULTI-SLICE CT, MICRO-CT AND CBCT

    41

    CHAPTER 4 INFLUENCE OF SCAN PARAMETERS ON CBCT

    TRABECULAR BONE MICROSTRUCTURAL

    MEASUREMENTS

    63

    CHAPTER 5 INFLUENCE OF OBJECT LOCATION ON CBCT

    MICROSTRUCTURAL ASSESSMENTS

    81

    CHAPTER 6 CBCT AND MICRO CT ASSESSMENTS OF TRABECULAR

    BONE MICROSTRUCTURE AT DIFFERENT

    MANDIBULAR REGIONS

    97

    CHAPTER 7 DISCUSSION 111

    CHAPTER 8 SUMMARY AND CONCLUSIONS 125

    CHAPTER 9 SAMENVATTING EN CONCLUSIES 129

    PUBLICATIONS 133

    CHAPTER 10

    ACKNOWLEDGMENTS

    135

  • 8

  • 9

    Chapter 1

    Introduction

    Part of this chapter has been published as:

    Ibrahim N, Parsa A, Hassan B, van der Stelt P, Wismeijer D. Diagnostic imaging of

    trabecular bone microstructure for oral implants: a literature review.

    Dentomaxillofacial Radiology 2013; 42: 20120075.

  • 10

  • 11

    Current and future trends of CBCT in implant dentistry

    Presurgical radiographic assessment is essential in planning dental implant treatment1

    and dental implant thread design.2 Currently, the role of cone beam CT (CBCT) for

    dental implant treatment is gradually increasing due to the wide accessibility and the

    advantages obtained from these systems.3,4

    CBCT assessments, however, are

    generally focused on bone density5,6

    and linear bone measurement.7 Structural

    properties of bone are much less involved in CBCT evaluations.

    The structural properties of trabecular bone are amongst the significant determinants

    for bone strength. The assessment of trabecular bone structures is recommended

    when predicting an implant success8,9

    owing to its significant role in the healing and

    osseointegration process at the implant-bone surface.10

    However, studies on trabecular

    microstructure assessment using CBCT are scarce because of the low scanning

    resolution11

    of the earlier generation of CBCT devices.

    The latest CBCT equipment with a resolution of 80µm may probably serve as a 3D

    imaging modality for the clinical assessment of the anisotropic trabecular

    microstructures.12

    Subsequently, this chapter reviews the imaging modalities for

    trabecular bone microstructure in oral implants and the potential of CBCT for

    microstructural assessments.

    Diagnostic imaging of trabecular bone microstructure for oral implants

    The term ‘bone quality’ has been extensively used in the literature to describe

    different aspects of bone characteristics with variable definitions depending on the

    utilized context. Among inseparable factors that influence bone quality is the

    trabecular bone.13-16

    The trabeculae or ‘trabecular’ is the primary anatomical and

    functional unit of cancellous bone. Cortical bone helps to attain primary implant

    stability, but the role of cancellous bone is also remarkable. This is because

    cancellous bone has a higher bone turnover rate than cortical bone17

    and has a direct

    contact with majority of the implant surface.8 Accordingly, it influences the healing

    and osseointegration process at the implant-bone surface.10

  • 12

    Bone strength has a significant role in determining implant success. To improve

    prediction of bone strength, the measurements of trabecular density and trabecular

    microstructure should be combined.18

    This is because those measurements do not

    always denote each other. For instance, a high bone density does not always

    correspond to high trabecular parameters such as trabecular number (Tb.N) and

    trabecular thickness (Tb.Th).19

    Therefore, estimating implant success by assessing

    the trabecular density alone is no longer suggested.9

    Precise clinical assessment of bone structural and mechanical properties is essential in

    planning dental implant treatment and implant thread design.2 The task can be

    performed on two-dimensional (2D) plain radiographs (e.g. intra oral radiograph) by

    calculating fractal dimensions of the trabecular bone.20

    In three-dimensional (3D)

    imaging modalities (e.g. HR-pQCT) the high resolution images are analyzed using

    dedicated imaging software (e.g. CT Analyser [CTAn]; Skyscan®, Kontich,

    Belgium). Computational techniques such as finite element methods (FEM)21,22

    are

    also utilized in analyzing 3D images to simulate the status of implant surface and the

    bone adjacent to the implant.2

    To date, bone quality assessments in oral implant studies have largely focused on

    trabecular bone density.5,23-25

    What follows is a review of the imaging techniques used

    in oral implant studies for assessing the trabecular microstructures as evidenced in the

    literature. Articles reported on trabecular microstructural imaging methods were

    searched in PubMed electronic database. Titles and abstracts of the related articles

    were reviewed base on keywords that had initially been set as inclusion criteria: bone

    quality, imaging, trabecular microstructure, cone beam CT and dental or oral implant.

    1. Dental Radiographs

    Periapical (PA) and panoramic (OPG) radiographs are the first-choice diagnostic

    clinical instruments in dentistry. PA radiographs with superior resolution and

    sharpness provide valuable information for evaluating the amount and pattern of

    trabecular bone structure.26,27

    Trabecular visibility was reported to be high on PA

    radiographs, 28

    thus enhancing its potential in trabecular imaging studies.29-34

  • 13

    Bone classification systems are used to study bone quality on PA images. Of the

    Lekholm & Zarb, Trisi & Rao and Misch systems, the first is largely adopted in oral

    implant studies on trabecular bone assessment.29-32

    A visual index was proposed in

    1996 to simplify trabecular classification on PA radiographs.30

    This index categorizes

    trabecular pattern according to the intertrabecular spaces (small or large) and degree

    of trabeculation (sparse or dense).32-34

    However, these subjective techniques remain

    partially validated.29

    On the other hand, panoramic radiographs have also been used to

    assess trabecular structure.35-36

    However, the technique applies the rotational

    principles that structures not centered in the focal trough are not sharply imaged. The

    formation of geometrical distortion, magnification and loss of information are thus

    commonly observed artifacts on panoramic radiographs. Moreover, the reduced

    resolution of panoramic images degrades its ability in identifying fine trabeculae.37

    Therefore, its applications in trabecular assessments are less favorable than PA

    radiographs.34

    Undeniably, utilizing dental radiographs for assessing trabecular microstructure is a

    rapid, relatively safe and convenient method to apply in the jaws. Although the nature

    of the 2D image could never provide information in the bucco-lingual direction,38

    dental radiographs are still largely employed in many countries for pre implant

    assessment due to availability and cost. 39

    The complex shapes and structure of trabecular bone can be calculated by performing

    fractal dimension (FD) analysis on 2D images such as periapical and panoramic

    radiographs.37

    Current studies on 2D FD analysis of trabecular microarchitecture

    parameters (porosity, connectivity and anisotropy) are reported to be adequately

    comparable to that of 3D FD method.40

    FD analyses and calculations of trabecular

    structures require several complex steps.32

    Nowadays, the FD applications are

    simplified by using personal computers and a simple JAVA software (Oracle®, Los

    Angeles, CA). However, the overall reproducibility of the projection techniques

    remains as contentious issue that requires further investigations.41

  • 14

    2. Magnetic Resonance Imaging (MRI)

    MRI is a non-invasive, non-ionizing system which applies high magnetic fields,

    transmission of radiofrequency waves and detection of radiofrequency signals from

    excited hydrogen protons. Trabecular bone is filled with bone marrow that contains

    free protons and generates a strong MR signal.42,43

    Fat and water protons in the

    marrow tissue are depicted as negative image. Because trabecular structure can not

    directly be visualized, this technique employs image processing to invert the negative

    image.44,45

    Using this technique, values for implant loading and bone healing time at

    trabecular alveolar bone were proposed to improve implant success.46

    Despite

    improving the trabecular structure assessment, the quality of the acquired MR images

    is largely influenced by the field strength, pulse sequence, echo time, and signal to

    noise ratio (SNR). Additionally, the measurements are affected by the selected

    threshold values, image processing algorithms, complex analysis and interpretation of

    the images.46-48

    Moreover, the availability and accessibility of MRI machines for the

    dental practitioners remains limited.

    3. Computed Tomography (CT)

    Computed tomography techniques are being progressively developed to meet the

    clinical needs in assessing bone microstructure. Structural analysis of trabecular bone

    requires scanners with contiguous isotropic pixel resolution of less than 300μm.49

    High resolution CT systems that are commonly employed for trabecular

    microstructural assessment in oral implant studies are discussed below.

    3.1 Multi Detector Computed Tomography (MDCT)

    The latest generation of MDCT system has improved its resolution to 150-300μm in

    plane and 300-500μm in slice thickness.50

    Trabecular microstructure parameters such

    as trabecular number (Tb.N), trabecular thickness (Tb.Th) and trabecular separation

    (Tb.Sp) were measured using MDCT and compared to high-resolution peripheral

    quantitative CT (HR-pQCT).49

    Although the resolution is still beyond trabecular

    dimensions (50-200µm), the measurements from both techniques were highly

    correlated. In a human cadaver study, trabecular microstructure parameters were

  • 15

    compared among MDCT and micro-CT and micro-CT FE modelling.51

    The study

    concluded that trabecular bone structure assessment using MDCT is overall feasible,

    although still limited by its spatial resolution. These studies were conducted using a

    high-resolution mode, which is not routinely used in clinical settings protocols.49-50

    Consequently, although MDCT is largely employed in oral implant studies, its

    applicability remains mostly confined to bone density measurements. 52-54

    3.2 High-resolution peripheral Quantitative Computed Tomography (HR-

    pQCT)

    With a spatial resolution of 82μm, this device is used for trabecular microstructural

    imaging. The measurements of microstructural parameters are reported to be

    comparable with that of micro-CT (voxel size of 25μm).55

    The technology has a

    higher spatial resolution than MDCT; however, scanning sites are limited to

    peripheral skeletal region (e.g. wrist and tibia) and accessibility is currently limited.50

    Unlike MRI technique, microstructural assessment using high-resolution CT permits

    direct visualization of the trabecular bone. However, the later technique involves a

    relatively high radiation dose which is beyond the recommended clinical setting.45

    Moreover, the results are also affected by the selected threshold, image analysis and

    processing techniques.56

    Thus, its application in oral implant imaging studies remains

    restricted.

    3.3 Micro Computed Tomography (Micro-CT)

    Two-dimensional histomorphometric analysis was previously considered as the gold

    standard for assessing trabecular size, shape, connectivity and orientation. As it is

    time consuming and costly, micro-CT is now routinely employed for structural 2D or

    3D evaluations of trabecular microstructure.50,57

    This non-destructive high resolution

    (approaching 10µm) method depicts trabecular network in different gray levels

    according to its mineral content. It has been reported that the trabecular parameters

    quantified by micro-CT are comparable to the traditional 2D histomorphometric

    values.42,43

    As it permits high resolution scans, in 2004 micro-CT has been

    recommended as a gold standard imaging for ex-vivo bone studies at implant sites.57

  • 16

    However, only studies with small jaw specimens were conducted to observe

    trabecular microstructure in oral implant research.5,8,19,58

    3.4 Cone beam computed tomography (CBCT)

    Cone beam computed tomography systems were developed in the1990s. In 2001,

    CBCT was introduced as a 3D imaging modality. Since then it has largely replaced

    both single- and multi- slice CT for diagnostic imaging in oral implants. 59

    Owing to

    the wide availability of the machines, rapid scan and processing time, high resolution

    images and relatively reduced scan radiation dose and costs, the demand for CBCT

    images preceding implant placement has increased exponentially.3,60-63

    Although

    many studies have been conducted on CBCT, the literature on its suitability in

    measuring trabecular bone microstructural parameters at oral implant site remains

    scarce. This may be due to the insufficient resolution of the past generations of CBCT

    systems to depict bone microstructure. The applications of CBCT in evaluating bone

    quality are still restricted on bone density assessment.52-54

    Recently, however, a study

    on assessing bone microstructure described CBCT as a promising modality for

    analyzing trabecular bone.63

    Bone parameters (Tb.Th, Tb.N and Tb.Sp) at mandibular

    condyle were also successfully evaluated by CBCT at a resolution of 125µm coupled

    with image processing.64

    The visibility of small anatomical structures with CBCT is largely influenced by the

    field of view (FOV) and scan setting selection.65

    Visibility of trabecular

    microstructure is mainly determined by the chosen voxel size and SNR plus image

    artifacts.66

    In CBCT, voxel size and slice thickness, spatial and contrast resolutions

    vary with respect to machine type, FOV and scan settings. 65,66

    Additionally, several

    image artifacts specific to CBCT technology could influence the effective system

    resolution, which could be lesser than the nominal system resolution expressed in

    voxel size alone. It has been previously stated that the accuracy of 3D measurement of

    anisotropic trabecular structure can be improved by performing in-vivo rather than in-

    vitro investigation.16,18

    In this respect, the use of CBCT could prove appealing. As the

    need to evaluate the implant insertion sites prior to surgical placement has

    dramatically increased, CBCT should be validated as a non-invasive procedure for

    assessing bone microstructure.

  • 17

    Figure 1 Sagittal images of trabecular structure at the lingual foramen region derived

    from A: MDCT (650μm), B: CBCT (80μm) and C: Micro CT (35μm).

    Müller et al. has described that a CT scanner with a resolution up to 60μm can present

    morphometric information comparable to that of 10 μm.67

    Using the latest CBCT

    system, the appearance of trabecular structure was observed using a 4x4cm FOV at a

    nominal resolution of 80µm. The resultant image was compared with images derived

    from MDCT and micro-CT (Figure 1). It is expected that this system could be useful

    in measuring trabecular microstructures. However, thorough investigation and

    validation are required prior to applying this technique in clinical practice.

    Although there is a rapid progress in advanced bone imaging modalities, their routine

    clinical employment remains limited due to the technical features, cost and complex

    procedures. The current review recommends studies to validate CBCT as a clinical

    imaging modality to evaluate trabecular microstructure at oral implant sites.

    Aims of the thesis:

    The primary aim of this thesis is to validate the applicability and accuracy of CBCT

    for trabecular bone microstructural assessments. Additionally, the consistency of

    structural analysis is assessed using various scanning protocols, different sites of the

    mandible and different locations of the object in the CBCT field FOV. Therefore, the

    current studies are conducted according to the following research questions:

  • 18

    1. What is the accuracy of CBCT trabecular bone microstructure measurements in

    comparison with µCT?

    2. What is the accuracy of CBCT trabecular bone density and bone volume fraction

    in comparison with MSCT and µCT?

    3. What is the effect of scan parameters (FOV, rotation steps and resolution) on

    CBCT trabecular bone microstructural measurements?

    4. What is the influence of the object location on CBCT trabecular bone

    microstructure measurements?

    5. What is the difference between CBCT and µCT in measuring trabecular bone

    microstructure at different sites of edentulous mandibles?

  • 19

    References:

    1. Jacobs R, Van Steenberghe D. Radiographic planning and assessment of endosseous

    oral implants. Berlin: Springer, 1997.

    2. DeTolla DH, Andreana S, Patra A, Buhite R, Comella B. Role of the finite element

    model in dental implants. J Oral Implantol 2000; 26: 77-81.

    3. De Vos W, Casselman J, Swennen GR. Cone-beam computerized tomography

    (CBCT) imaging of the oral and maxillofacial region: a systematic review of the

    literature. Int J Oral and Maxillofac Surg 2009; 38: 609-625.

    4. Guerrero ME, Jacobs R, Loubele M, Schutyser F, Suetens P, van Steenberghe D.

    State-of-the-art on cone beam CT imaging for preoperative planning of implant

    placement. Clin Oral Investig 2006; 10: 1-7.

    5. González-García R, Monje F. The reliability of cone-beam computed tomography to

    assess bone density at dental implant recipient sites: a histomorphometric analysis by

    micro-CT. Clin Oral Implants Res 2012 doi: 10.1111/j.1600-0501.2011.02390.x.

    6. Monje A, Monje F, González-García R, Galindo-Moreno P, Rodriguez-Salvanes F,

    Wang HL. Comparison between microcomputed tomography and cone-beam

    computed tomography radiologic bone to assess atrophic posterior maxilla density

    and microarchitecture. Clin Oral Implants Res. 2013 Feb 26. doi: 10.1111/clr.12133.

    7. Quereshy FA, Savell TA, Palomo JM. Applications of cone beam computed

    tomography in the practice of oral and maxillofacial surgery. Journal of Oral

    Maxillofacial Surgery 2008; 66: 791-796.

    8. Fanuscu MI, Chang TL. Three-dimensional morphometric analysis of human cadaver

    bone: microstructural data from maxilla and mandible. Clin Oral Impl Res 2004; 15:

    213-218.

    9. Wirth AJ, Goldhahn J, Flaig C, Arbenz P, Müller R, van Lenthe GH. Implant stability

    is affected by local bone microstructural quality. Bone 2011; 49: 473-478.

    10. Minkin C, Marinho VC. Role of the osteoclast at the bone-implant interface. Adv Dent

    Res 1999; 13: 49-56.

    11. Kothari M, Keaveny TM, Lin JC, Newitt DC, Genant HK, Majumdar S. Impact of

    spatial resolution on the prediction of trabecular architecture parameters. Bone 1998;

    22: 437-443.

    12. Müller R, Van Campenhout H, Van Damme B, Van Der Perre G, Dequeker J,

    Hildebrand T, et al. Morphometric analysis of human bone biopsies: a quantitative

  • 20

    structural comparison of histological sections and micro-computed tomography. Bone

    1998; 23:59–66.

    13. Sievänen H, Kannus P, Järvinen TLN. Bone quality: An empty term. PLoS Med 2007;

    4: e27.

    14. Compston J. Bone Quality: What is it and how is it measured? Arq Bras Endocrinol

    Metab 2006; 50: 579-585.

    15. Licata A. Bone density vs bone quality: What’s a clinician to do? Cleve Clin J Med

    2009; 76: 331-336.

    16. Fyhrie DP. Summary - Measuring ‘‘bone quality’’. J Musculoskelet Neuronal Interact

    2005; 5: 318-320.

    17. Sakka S, Coulthard P. Bone quality: a reality for the process of osseointegration.

    Implant Dent 2009; 18: 480-485.

    18. Müller R. Bone microarchitecture assessment: current and future trends. Osteoporos

    Int 2003; 14 Suppl 5:S89-S95.

    19. Gomes de Oliveira RC, Leles CR, Lindh C, Ribeiro-Rotta RF. Bone tissue

    microarchitectural characteristics at dental implant sites. Part 1: Identification of

    clinical-related parameters. Clin Oral Implants Res 2011; 000-000. doi:

    10.1111/j.1600-0501.2011.02243.x.

    20. Yi WJ, Heo MS, Lee SS, Choi SC, Huh KH. Comparison of trabecular bone

    anisotropies based on fractal dimensions and mean intercept length determined by

    principal axes of inertia. Med Biol Eng Comput 2007; 45: 357-364.

    21. Krug R, Burghardt AJ, Majumdar S, Link TM. High-resolution imaging techniques

    for the assessment of osteoporosis. Radiol Clin North Am 2010; 48: 601-621.

    22. Müller R, van Lenthe GH. Trabecular bone failure at the microstructural level. Curr

    Osteoporos Rep 2006; 4: 80-86.

    23. Turkyilmaz I, McGlumphy EA. Influence of bone density on implant stability

    parameters and implant success: a retrospective clinical study. BMC Oral Health

    2008; 8: 32.

    24. Turkyilmaz I, Sennerby L, McGlumphy EA, Tözüm TF. Biomechanical aspects of

    primary implant stability: a human cadaver study. Clin Implant Dent Relat Res 2009;

    11: 113-119.

    25. Molly L. Bone density and primary stability in implant therapy. Clin Oral Implants

    Res 2006; 17: 124-135.

  • 21

    26. White SC, Pharoah MJ. Oral radiology: principles and interpretation (5th edn). St.

    Louis: Mosby, 2004.

    27. Whaites E. Essentials of dental radiography and radiology (3rd edn). London:

    Churchill Livingstone, 2007.

    28. Couture RA, Whiting BR, Hildebolt CF, Dixon DA. Visibility of trabecular structures

    in oral radiographs. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2003; 96:

    764–771.

    29. Ribeiro-Rotta RF, Lindh C, Pereira AC, Rohlin M. Ambiguity in bone tissue

    characteristics as presented in studies on dental implant planning and placement: a

    systematic review. Clin Oral Implants Res 2011; 22: 789-801.

    30. Lindh C, Petersson A, Rohlin M. Assessment of the trabecular pattern before

    endosseous implant treatment: diagnostic outcome of periapical radiography in the

    mandible. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 1996; 82: 335–343.

    31. Aalam AA, Nowzari H, Krivitsky A. Functional restoration of implants on the day of

    surgical placement in the fully edentulous mandible: a case series. Clin Implant Dent

    Relat Res 2005; 7: 10-16.

    32. Jonasson G, Jonasson G, Jonasson L, Kiliaridis S. Skeletal bone mineral density in

    relation to thickness, bone mass, and structure of the mandibular alveolar process in

    dentate men and women. Eur J Oral Sci 2007; 115: 117-123.

    33. Lindh C, Horner K, Jonasson G, Olsson P, Rohlin M, Jacobs R, et al. The use of

    visual assessment of dental radiographs for identifying women at risk of having

    osteoporosis: the OSTEODENT project. Oral Surg Oral Med Oral Pathol Oral

    Radiol Endod 2008; 106: 285–293.

    34. Pham D, Jonasson G, Kiliaridis S. Assessment of trabecular pattern on periapical and

    panoramic radiographs: A pilot study. Acta Odontol Scand 2010; 68: 91-97.

    35. Taguchi A, Suei Y, Ohtsuka M, Otani K, Tanimoto K, Ohtaki M. Usefulness of

    panoramic radiography in the diagnosis of postmenopausal osteoporosis in women.

    Width and morphology of inferior cortex of the mandible. Dentomaxillofac Radiol

    1996; 25: 263-267.

    36. Watanabe PCA, Oliveira TM, Monteiro SAC, Iyomasa MM, Regalo SCH, Siessere S.

    Morphodigital study of the mandibular trabecular bone in panoramic radiographs. Int

    J Morphol 2007; 25: 875-880.

    37. Bollen AM, Taguchi A, Hujoel PP, Hollender LG. Fractal dimension on dental

    radiographs. Dentomaxillofac Radiol 2001; 30: 270-275.

  • 22

    38. Lofthag-Hansen S, Gröndahl K, Ekestubbe A. Cone-beam CT for preoperative

    implant planning in the posterior mandible: visibility of anatomic landmarks. Clin

    Implant Dent Relat Res 2009; 11: 246-255.

    39. Sakakura CE, Morais JA, Loffredo LC, Scaf G.A survey of radiographic prescription

    in dental implant assessment. Dentomaxillofac Radiol 2003; 32: 397-400.

    40. Pothuaud L, Benhamou CL, Porion P, Lespessailles E, Harba R, Levitz P. Fractal

    dimension of trabecular bone projection texture is related to three-dimensional

    microarchitecture. J Bone Miner Res 2000; 15: 691-699.

    41. Sánchez I, Uzcátegui G. Fractals in dentistry. J Dent 2011; 39: 273-292.

    42. Genant HK, Jiang Y. Advanced imaging assessment of bone quality. Ann N Y Acad

    Sci 2006; 1068: 410-428.

    43. Lespessailles E, Chappard C, Bonnet N, Benhamou CL. Imaging techniques for

    evaluating bone microarchitecture. Joint Bone Spine 2006; 73: 254-261.

    44. Jiang Y, Zhao J, White DL, Genant HK. Micro CT and Micro MR imaging of 3D

    architecture of animal skeleton. J Musculoskelet Neuronal Interact 2000; 1: 45-51.

    45. Ito M. Recent progression in assessment of bone microstructure. Nihon Rinsho 2011;

    691233-691238.

    46. Choël L, Last D, Duboeuf F, Seurin MJ, Lissac M, Briquet A, et al. Trabecular

    alveolar bone microarchitecture in the human mandible using high resolution

    magnetic resonance imaging. Dentomaxillofac Radiol 2004; 33: 77–182.

    47. Sell CA, Masi JN, Burghardt A, Newitt D, Link TM, Majumdar S. Quantification of

    trabecular bone structure using magnetic resonance imaging at 3 Tesla-calibration

    studies using microcomputed tomography as a standard of reference. Calcif Tissue Int

    2005; 76: 355-364.

    48. Celenk P, Celenk C. Evaluation by quantitative magnetic resonance imaging of

    trabecular bone quality in mandible and cervical vertebrae. Clin Oral Implants Res

    2010; 21: 409-413.

    49. Issever AS, Link TM, Kentenich M, Rogalla P, Burghardt AJ, Kazakia GJ, et al.

    Assessment of trabecular bone structure using MDCT: comparison of 64- and 320-

    slice CT using HR-pQCT as the reference standard. Eur Radiol 2010; 20: 458-468.

    50. Burghardt AJ, Link TM, Majumdar S. High-resolution computed tomography for

    clinical imaging of bone microarchitecture. Clin Orthop Relat Res 2011; 469: 2179-

    2193.

  • 23

    51. Issever AS, Link TM, Kentenich M, Rogalla P, Schwieger K, Huber MB, et al.

    Trabecular bone structure analysis in the osteoporotic spine using a clinical in vivo

    setup for 64-slice MDCT imaging: comparison to micro CT imaging and micro FE

    modelling. J Bone Miner Res 2009; 24:1628-1637.

    52. Naitoh M, Aimiya H, Hirukawa A, Ariji E. Morphometric analysis of mandibular

    trabecular bone using cone beam computed tomography: an in vitro study. Int J Oral

    Maxillofac Implants 2010; 25: 1093-1098.

    53. Naitoh M, Hirukawa A, Katsumata A, Ariji E. Prospective study to estimate

    mandibular cancellous bone density using large-volume cone-beam computed

    tomography. Clin Oral Implants Res 2010; 21: 1309-1313.

    54. Araki K, Okano T. The effect of surrounding conditions on pixel value of cone beam

    computed tomography. Clin Oral Impl Res 2011; 00:1-4.

    55. Liu XS, Zhang XH, Sekhon KK, Adams MF, McMahon DJ, Bilezikian JP, et al.

    High-Resolution Peripheral Quantitative Computed Tomography can assess

    microstructural and mechanical properties of human distal tibial bone. J Bone Miner

    Res 2010; 25: 746–756.

    56. MacNeil JA, Boyd SL. Accuracy of high-resolution peripheral quantitative computed

    tomography for measurement of bone quality. Med Eng Phys 2007; 29: 1096-1105.

    57. Yip G, Schneider P, Roberts EW. Micro-Computed Tomography: High resolution

    imaging of bone and implants in three dimensions. Semin Orthod 2004; 10: 174-187.

    58. Swain MV, Xue J. State of the art of micro CT applications in dental research. Int J

    Oral Sci 2009; 1: 177-188.

    59. Hatcher DC. Operational principles for cone-beam computed tomography. J Am Dent

    Assoc 2010; 141 Suppl 3:3S-6S.

    60. Al-Rawi B, Hassan B, Vandenberge B, Jacobs R. Accuracy assessment of three-

    dimensional surface reconstructions of teeth from Cone Beam Computed Tomography

    scans. J Oral Rehabil 2010; 37: 352-358.

    61. Kau CH, Bozic M, English J, Lee R, Bussa H, Ellis RK. Cone-beam computed

    tomography of the maxillofacial region - an update. Int J Med Robot 2009; 5: 366-

    380.

    62. Liang X, Jacobs R, Hassan B, Li L, Pauwels R, Corpas L, et al. A comparative

    evaluation of Cone Beam Computed Tomography (CBCT) and Multi-slice CT

    (MSCT) Part I. on subjective image quality. Eur J Radiol 2010; 75: 265-269.

  • 24

    63. Corpas LS, Jacobs R, Quirynen M, Huang Y, Naert I, Duyck J. Peri-implant bone

    tissue assessment by comparing the outcome of intra-oral radiograph and cone beam

    computed tomography analyses to the histological standard. Clin Oral Impl Res 2011;

    22: 492-499.

    64. Liu SM, Zhang ZY, Li JP, Liu DG, Ma XC. A study of trabecular bone structure in

    the mandibular condyle of healthy young people by cone beam computed

    tomography. Zhonghua Kou Qiang Yi Xue Za Zhi 2007; 42: 357-360.

    65. Loubele M, Jacobs R, Maes F, Denis K, White S, Coudyzer W, et al. Image quality vs

    radiation dose of four cone beam computed tomography scanners. Dentomaxillofac

    Radiol 2008; 37: 309-318.

    66. Schulze R, Heil U, Gross D, Bruellmann DD, Dranischnikow E, Schwanecke U, et al.

    Artefacts in CBCT: a review. Dentomaxillofac Radiol 2011; 40: 265-273.

    67. Müller R, Koller B, Hildebrand T, Laib A, Gianolini S, Rüegsegger P. Resolution

    dependency of microstructural properties of cancellous bone based on three-

    dimensional μ-tomography. Technol Health Care 1996; 4: 113–119.

  • 25

    Chapter 2

    Accuracy of trabecular bone

    microstructural measurement using

    cone-beam CT datasets

    This chapter has been published as:

    Ibrahim N, Parsa A, Hassan B, van der Stelt P, Aartman IHA,Wismeijer D. Accuracy of

    trabecular bone microstructural measurement at planned dental implant sites using cone-beam

    CT datasets. Clinical Oral Implants Research. doi: 10.1111/clr.12163.

  • 26

    Accuracy of trabecular bone microstructural

    measurement at planned dental implant sites using

    cone-beam CT datasets

    Norliza Ibrahim, Azin Parsa, Bassam Hassan, Paul van der Stelt, Irene

    HA Aartman, Daniel Wismeijer.

    Clinical Oral Implants Research. doi: 10.1111/clr.12163

    Summary

    Objective: Cone-beam CT (CBCT) images are infrequently utilized for trabecular

    bone microstructural measurement due to the system's limited resolution. The aim of

    this study was to determine the accuracy of CBCT for measuring trabecular bone

    microstructure in comparison with micro CT (μCT).

    Materials and methods: Twenty-four human mandibular cadavers were scanned

    using a CBCT system (80μm) and a μCT system (35μm). Three bone microstructural

    parameters trabecular number (Tb.N), thickness (Tb.Th) and separation (Tb.Sp) were

    assessed using CTAn imaging software.

    Results: Intraclass correlation coefficients (ICC) showed a high intra-observer

    reliability (≥ 0.996) in all parameters for both systems. The Pearson correlation

    coefficients between the measurements of the two systems were for Tb.Th 0.82, for

    Tb.Sp 0.94 and for Tb.N 0.85 (all P's

  • 27

    Introduction

    The influence of bone quality on implant success is well acknowledged. Information of bone

    quality is best gained by combining bone mineral density (BMD) and trabecular structure

    assessments (Felsenberg & Boonen 2005). Trabecular bone is the source of osteoblasts and

    osteoclasts which are largely responsible for the physiological changes which take place

    subsequent to implant placement (Minkin & Marinho 1999).

    The role of trabecular

    microstructure in dental implants is significant since the implant is surrounded by trabecular

    bone which directly contributes to implant stability (Fanuscu & Chang 2004; Sakka &

    Coulthard 2009).

    Radiographic information of the trabecular microstructure can be achieved by using high

    resolution imaging modalities that approach the trabecular dimensions (50-300μm).

    However, most of the utilized modalities have limitations in clinical practice. High resolution

    radiographic systems’ limitations rest on excessive radiation exposure to the patient (MDCT,

    HR-pQCT), complex image analysis (hr-MRI), restricted accessibility (MDCT, hr MRI) and

    limited scanning sites (HR-pQCT, µCT) (Genant et al. 2000; Issever et al. 2010).

    Since 2001, cone-beam CT (CBCT) has been used to provide three-dimensional images for

    diagnosis and treatment planning in dentistry (Hatcher 2010). It provides comparable images

    at reduced scan costs and dose. Additionally, CBCT systems are largely accessible to dental

    professionals and the scan time is relatively short (Patcas et al. 2012). Numerous CBCT

    studies were conducted using CBCT for dental implants. However, up to date the focus has

    mainly been on bone quantity (alveolar bone width and height) measurement accuracy, bone

    density, visibility of anatomical landmarks and virtual guided surgery (Quereshy et al. 2008;

    Tahmaseb et al. 2011). Recently, CBCT has been suggested as a modality for analysing

    trabecular microstructure at implant sites (Corpas et al. 2011; Ibrahim et al. 2012). To the best

    of our knowledge, there has been no report validating the ability of this system in measuring

    trabecular bone microstructural parameters. Therefore, the aim of this study was to determine

    the accuracy of CBCT in trabecular bone microstructure measurement in comparison with

    µCT, after first assessing the intraobserver reliability.

  • 28

    Materials and methods

    Image acquisition

    Twenty-four human mandibular cadavers were obtained from the Department of Functional

    Anatomy. Approval was obtained from the department to use this material for research

    purposes. All cadavers were scanned using a CBCT system (3D Accuitomo 170, Morita,

    Japan). To obtain the highest spatial resolution possible of an isotropic 80µm, the smallest

    field of view (FOV), that is, 4x4cm with the high-resolution scan mode and 360º arm

    rotation, were the selected scan settings. Images were acquired at 90kv and 5.0mA. The

    cadavers were then scanned using a µCT system (SkyScan 1173, Kontich, Belgium). This

    system allows scanning of large size specimens (140mm in diameter, 200mm in height).

    Hence the mandibles were not sectioned into smaller sizes. To reduce any possible

    movements of the samples during scanning, the samples were fixed in a cylindrical shape of

    Styrofoam, fit and mounted into the holder before scanning. The resultant images were again

    checked for motion artifacts. The settings were set by the manufacturer operating at 130kVp,

    61mA with nominal isotropic resolution of 35µm.

    Figure 1: A mandible scanned using cone beam CT (CBCT) was cropped into a smaller

    block in Amira software.

  • 29

    Image processing

    The datasets from both systems were exported as DICOM 3 files and imported into image

    analysis software (Amira v4.1, Visage Imaging Inc., Carlsbad, CA). To select the exact

    region of interest (ROI) from both systems, the datasets were processed as follows. The three-

    dimensional (3D) isosurfaces datasets from both µCT and CBCT were created and saved as

    STL files. Subsequently, the mandibles were cropped (confined) to the edentulous posterior

    region (Fig. 1). The z-axis of the CBCT images was flipped to match that of µCT because the

    scan orientation differs between the two systems. The isosurfaces were used to provide

    coordinates for matching the original volumes (voxel data) of both systems. The bone blocks

    were then manually superimposed on each other to provide maximum alignment (Fig. 2). The

    µCT was set as the reference standard while CBCT was considered as the experimental

    modality. For each matching set, a smaller ROI was selected confined in the vertical and

    horizontal slice directions to the cortical bone margins. For µCT ten and for CBCT five

    consecutive and contiguous slices were chosen for evaluation. Since µCT isotropic voxel size

    (35µm3) is approximately half of that for CBCT (80µm

    3), a double amount of slices was

    selected to ensure that the measurements are from the same region. The selected slices were

    exported as BMP images and imported into trabecular bone analysis software CTAn (v 1.11,

    SkyScan, Kontich, Belgium). Through histogram analysis, the images were thresholded and

    binarized (Fig. 3). The selected CBCT and micro CT image slices were then averaged to

    reduce the bias that may occurred during the manual registration procedure in Amira

    software. Twenty-four ROI were carefully matched and compared prior to the measurement

    process (Fig. 4). The regions disturbed by metal artifacts or cut off during the µCT scanning

    process were excluded. All measurements were performed by one trained observer twice for

    both systems independently with at least one week interval between the first and the second

    measurement to assess the observer’s bias.

  • 30

    Figure 2: A block of cropped

    CBCT image was manually

    transformed, superimposed

    and matched onto µCT to

    select the region of interest.

    (ROI).

    Figure 3: Binarized images of

    CBCT (a) and µCT (b).

    Figure 4: Trabecular bone

    microstructure parameters

    were measured and

    compared using CBCT (A)

    and µCT (B) datasets.

  • 31

    Statistical analysis

    The bone microstructural parameters that were considered for evaluations were the trabecular

    number (Tb.N), thickness (Tb.Th) and separation (Tb.Sp). The intraclass correlation

    coefficient (ICC) was used to determine the intraobserver reliability in measuring trabecular

    microstructure using CBCT and µCT. Paired t-tests were used to assess the difference in

    means between the two systems and the linear relation between both systems was assessed

    using the Pearson correlation coefficient (r). Finally a Bland-Altman plot was used to assess

    the accuracy of CBCT in measuring the microstructural parameters by plotting the difference

    between the measurements of CBCT and µCT against the means of those measurements (Fig.

    5).

    Figure 5: Bland-Altman plot of three bone microstructural parameters trabecular number

    (Tb.N), thickness (Tb.Th) and separation (Tb.Sp). Tb.Th (a), Tb.Sp (b) and Tb.N (c)

    measurements between CBCT and µCT.

  • 32

    Results

    The ICC’s revealed excellent intraobserver reliability for all parameters and both systems

    separately (Table 1). Therefore the first measurement was used for the following analyses.

    Paired t-tests showed significant differences between the CBCT and µCT for all parameters.

    Tb.Th and Tb.Sp were higher on CBCT images than on µCT images, but Tb.N was lower by

    CBCT (Table 2). The Pearson correlation coefficients for the relation between CBCT and

    µCT were all significant at p< 0.001 (Table 2). Bland-Altman analysis showed the smallest

    bias of measurements in Tb.N (-0.37 µm-1

    ) followed by Tb.Th (1.6µm) and Tb.Sp (8.8µm).

    The confidence interval for the measurement differences between the two systems was

    smallest for Tb.N (-0.79 to +0.06), followed by Tb.Th (-2.1 to +5.3) and Tb.Sp (-21.4 to

    +43.1).

    Table 1. Intraobserver reliability for CBCT and µCT using the intraclass correlation

    coefficient (ICC).

    N=24

    ICC

    CBCT µCT

    Tb.Th 0.998 0.999

    Tb.Sp 0.997 0.999

    Tb.N 0.999 0.996

    Three bone microstructural parameters: trabecular number (Tb.N), thickness (Tb.Th) and

    separation (Tb.Sp).

  • 33

    Table 2: Structural parameters, paired t-test, and correlation between measurements obtained

    by CBCT and µCT.

    Parameters Means and standard

    deviation

    Paired t-test Pearson ( r2 )

    CBCT µCT means sd SEM t df p

    Tb.Th (µm) 0.41 + 0.15 0.28 +0.10 0.13 0.08 0.02 7.48 23 0.001 0.67*

    Tb.Sp (µm) 0.85 +0.44 0.71 +0.39 0.15 0.15 0.03 4.83 23 0.001 0.89*

    Tb.N (µm-1) 0.87 +0.26 1.11 +0.29 -0.24 0.16 0.03 -7.35 23 0.001 0.72*

    Three bone microstructural parameters: trabecular number (Tb.N), thickness (Tb.Th) and

    separation (Tb.Sp).

    * Correlation is significant at the 0.001 level

    Discussion

    Bone quality evaluations using CBCT are usually concentrated on bone density (Corpas et al.

    2011; Gonzalez-Garcia & Monje 2012; Ibrahim et al. 2012; Parsa et al. 2012). However, in

    order to understand its influence on implant-bone integration, bone quality should also be

    assessed from the microstructural aspect of trabecular bone (Fanuscu & Chang 2004;

    Gonzalez-Garcia & Monje 2012). Apart from its role in bone healing and implant retention

    (Minkin & Marinho 1999), trabecular bone microstructure also contributes to bone strength

    (Felsenberg & Boonen 2005; Manske et al. 2010). Amongst the recommended

    microstructural parameters to estimate bone strength are trabecular thickness (Tb.Th),

    number (Tb.N), and spacing between each other (Tb.Sp). Bone density and trabecular

    microstructure measurements do not always correlate with each other (Gomes de Oliveira et

    al. 2012). For instance, a low-density bone may be associated with low Tb.Th (Hildebrand et

    al. 1999; Ranjanomennahary et al. 2011) and Tb.N, but high Tb.Sp (Hans et al. 2011;

    Ranjanomennahary et al. 2011). But, after a series of medications in osteoporotic patients, the

    increased bone density did not improve its bone strength (Riggs et al. 1990). Therefore, in

    selecting the best bone quality prior to dental implant treatment, the trabecular microstructure

    should also be included as part of the pre-surgical bone assessment.

  • 34

    µCT is a gold standard modality for trabecular microstructural assessment. However, it can

    only evaluate small bone samples and it cannot be employed to scan patients in a clinical

    setting (Burghardt et al. 2011). The use of CBCT in dental implants has increased, but its

    ability in analyzing the bone microstructural parameters at implant receptor site remains

    unverified. At present, only few studies reported that CBCT has potentials in measuring

    trabecular microstructure (Corpas et al. 2011; Liu et al. 2007; Ibrahim et al. 2013). In this

    study, Tb.N, Tb.Th and Tb.Sp at posterior mandibular regions were directly assessed,

    analyzed and compared between CBCT and µCT datasets using human mandibular cadavers

    at implant receptor sites. A strong correlation was observed when comparing the trabecular

    bone microstructure measurements with µCT. Bland-Altman plots show strong agreements

    between CBCT and µCT when measuring Tb.N (-0.37µm), Tb.Th (1.6µm) and Tb.Sp

    (8.8µm). Only Tb.N was measured smaller by CBCT (indicated by a negative bias value)

    while Tb.Th and Tb.Sp were measured greater than µCT (indicated by positive bias values).

    This discrepancy is in accordance to the voxel size of CBCT (80µm) which is twice that of

    µCT (35µm). Since the voxel in CBCT is significantly larger than its counterpart in µCT, the

    system can only depict significantly thick trabeculae, which in turn results in a smaller

    trabecular number measurement. This is in accordance with a study on trabecular

    microstructural parameters which also concluded that small trabeculae are poorly depicted by

    low resolution systems when comparing MDCT (274µm) with µCT (16µm) (Issever et al.

    2010) and HR-pQCT (82µm) with µCT (18µm) (Tjong et al. 2012). The latter study

    described that partial volume effects (PVE) were increased and a rough estimation of Tb.Th

    was observed when bigger voxel size images were utilized when measuring microstructural

    parameters. However, it has to be emphasized that while spatial resolution expressed in voxel

    size is a limiting factor for image quality, it’s not the only contributing factor. Contrast to

    noise ratio (CNR) plays an equally important role in determining the capability of an imaging

    system for depicting delicate anatomical structures. A sufficiently high CNR can possibly

    compensate for a relatively low spatial resolution (Bechara et al. 2012). Therefore, thin

    trabeculae can be made visible even when the trabecular thickness itself is beyond the

    nominal voxel size. Image quality is also largely affected by micro-movement of the jaw. The

    stationary situation of the cadavers in the present study might have thus enhanced the

    visibility of the trabecular structures because the scans did not suffer from patient’s

    movement.

  • 35

    One difficulty in this study was triggered during identifying the ROI. The projections in

    CBCT were not completely similar to µCT because the mandibles were positioned vertically

    in µCT to meet the space constraints in the scanner. Through using 3D virtual

    reconstructions, CBCT and µCT were brought into maximum alignment. The measurements

    could still somewhat vary due to the limitation of the non-automated registration. However,

    when an automated registration is not attainable, Diederichs et al. (2008) had suggested that

    both scans should be performed at a similar position to improve the accuracy of the selected

    ROIs. Further, prior to microstructural assessment in CTAn software, the visualized

    trabecular bone was again compared slice by slice to ascertain that the ROI was

    correspondingly selected.

    The size of the ROI and density of the bone strongly influence trabecular bone assessments.

    A diameter of 5mm or larger bone specimen is recommended to adequately analyze bone

    parameters (Vigorita 1984). In this study we used the full height and width of the bone

    sections sampled at 5 and 10 consecutive and contiguous coronal slices for CBCT and µCT,

    respectively (Fig. 3). Nevertheless, this careful sampling still does not guarantee that all

    measurements were from anatomically identical regions. This is due to the manual nature of

    aligning the two datasets and the possibility for observer error. An automated, observer-

    independent 3D volumetric matching using software tool is recommended to improve the

    alignment accuracy (Li et al. 2008). However, this was not possible in this study due to

    technical constrains. At any rate, the measurements were repeated twice and the ICC results

    show strong agreements between the first and the second measurement (Table 1).

    Apart from voxel size (Tjong et al. 2012); the accuracy of structural measurement is also

    influenced by the threshold selection (Parkinson et al. 2008). Bone segmentation with CBCT

    remains difficult as image quality is affected by artifacts specific to the scanning technology

    (Parsa et al. 2013; Schulze et al. 2011). CBCT has a lower signal to noise ratio than µCT, a

    larger amount of scatter radiation, reduced contrast and is largely susceptible to beam

    hardening and edge aliasing artifacts (Schulze et al. 2011). All these combined factors

    introduce errors in the grey level values in CBCT which results in underestimation of thin

    trabecular bone. Therefore, it was recommended to use denser bone specimens to overcome

    this problem (Wirth et al. 2012). In this study all 24 mandibles were used regardless of their

    density. Owing to its extremely high resolution, µCT was able to image bone specimens even

    for poor bone density blocks. However, the same could not be stated for CBCT since the

  • 36

    partial volume effect is accentuated for sparse trabecular patterns. Therefore, trabecular

    spacing in particular was overestimated since many small trabeculae were thresholded out

    when binarizing the images. This could possibly explain the outliers depicted in the Bland-

    Altman plots (Fig. 5). Only small discrepancies were observed (Tb.N

  • 37

    References:

    1. Bechara, B., McMahan, C.A., Moore, W.S., Noujeim, M., Geha, H. & Teixeira, F.B.

    (2012) Contrast- to-noise ratio difference in small field of view cone beam computed

    tomography machines. Journal of Oral Science 54: 227–232.

    2. Burghardt, A.J., Link, T.M. & Majumdar, S. (2011) High-resolution computed

    tomography for clinical imaging of bone microarchitecture. Clinical Orthopaedics

    and Related Research 469: 2179–2193.

    3. Corpas, L.S., Jacobs, R., Quirynen, M., Huang, Y., Naert, I. & Duyck, J. (2011) Peri-

    implant bone tissue assessment by comparing the outcome of intra-oral radiograph

    and cone beam computed tomography analyses to the histological standard. Clinical

    Oral Implants Research 22: 492-499.

    4. Diederichs, G., Link, T., Marie, K., Huber, M., Rogalla, P., Burghardt, A., Majumdar,

    S. & Issever, A. (2008) Feasibility of measuring trabecular bone structure of the

    proximal femur using 64-slice multidetector computed tomography in a clinical

    setting. Calcified Tissue International 83: 332–341.

    5. Fanuscu, M.I. & Chang, T.L. (2004) Three-dimensional morphometric analysis of

    human cadaver bone: microstructural data from maxilla and mandible. Clinical Oral

    Implants Research 15: 213-218.

    6. Felsenberg, D. & Boonen, S. (2005) The bone quality framework: determinants of

    bone strength and their interrelationships, and implications for osteoporosis

    management. Clinical Therapeutics 27: 1–11.

    7. Genant, H.K., Gordon, C., Jiang, Y., Link, T.M., Hans, D., Majumdar, S. & Lang,

    T.F. (2000) Advanced imaging of the macrostructure and microstructure of bone.

    Hormone Research 54 (suppl 1): 24-30.

    8. Gomes de Oliveira, R.C., Leles, C.R., Lindh, C. & Ribeiro-Rotta, R.F. (2012) Bone

    tissue microarchitectural characteristics at dental implant sites. Part 1: identification

    of clinical-related parameters. Clinical Oral Implants Research 23: 981–986.

    9. Gonzalez-Garcia, R. & Monje, F. (2012) The reliability of cone-beam computed

    tomography to assess bone density at dental implant recipient sites: a

    histomorphometric analysis by micro-CT. Clinical Oral Implants Research doi:

    10.1111/j.1600-0501.2011.02390.x.

    10. Hans, D., Barthe, N., Boutroy, S., Pothuaud, L., Winzenrieth, R. & Krieg, M.A.

    (2011) Correlations between trabecular bone score, measured using anteroposterior

  • 38

    dual-energy X-ray absorptiometry acquisition, and 3-dimensional parameters of bone

    microarchitecture: an experimental study on human cadaver vertebrae. Journal of

    Clinical Densitometry 14: 302–312.

    11. Hatcher, D.C. (2010) Operational principles for cone-beam computed tomography.

    Journal of American Dental Association 141(suppl 3): 3–6.

    12. Hildebrand, T., Laib, A., Müller, R., Dequeker, J. & Ruegsegger, P. (1999) Direct

    three-dimensional morphometric analysis of human cancellous bone: microstructural

    data from spine, femur, iliac crest, and calcaneus. Journal of Bone Mineral Research

    14: 1167–1174.

    13. Ibrahim, N., Parsa, A., Hassan, B., van der Stelt, P. & Wismeijer, D. (2012)

    Diagnostic imaging of trabecular bone microstructure for oral implants: a literature

    review. Dentomaxillofacial Radiology 42:20120075.

    14. Issever, A.S., Link, T.M., Kentenich, M., Rogalla, P., Burghardt, A.J., Kazakia, G.J.,

    Majumdar, S. & Diederichs, G. (2010) Assessment of trabecular bone structure using

    MDCT: comparison of 64- and 320-slice CT using HR-pQCT as the reference

    standard. European Radiology 20: 458-468.

    15. Li, G., Xie, H., Ning, H., Citrin, D., Capala, J., Maass-Moreno, R., Guion, P., Arora,

    B., Coleman, C., Camphausen, K. & Miller, R. (2008) Accuracy of 3D volumetric

    image registration based on CT, MR and PET/CT phantom experiments. Journal of

    Applied Clinical Medical Physics 9: 17-36.

    16. Liu, S.M., Zhang, Z.Y., Li, J.P., Liu, D.G. & Ma, X.C. (2007) A study of trabecular

    bone structure in the mandibular condyle of healthy young people by cone beam

    computed tomography. Zhonghua Kou Qiang Yi Xue Za Zhi 42: 357-360.

    17. Manoglas, S.C. & Jilka, R.L. (1995) Bone marrow, cytokines, and bone modelling.

    Emerging insights into the pathophysiology of osteoporosis. New England Journal of

    Medicine 332: 305-311.

    18. Manske, S.L., Macdonald, H.M., Nishiyama, K.K., Boyd, S.K. & McKay, H.A.

    (2010) Clinical tools to evaluate bone strength. Clinical Reviews in Bone and Mineral

    Metabolism 8: 122–134.

    19. Minkin, C. & Marinho, V.C. (1999) Role of the osteoclast at the bone-implant

    interface. Advances in Dental Research 13: 49–56.

    20. Parkinson, I.H., Badiei, A. & Fazzalari, N.L. (2008) Variation in segmentation of

    bone from micro-CT imaging: implications for quantitative morphometric analysis.

    Australasian Physical & Engineering Sciences in Medicine 31: 160-164.

  • 39

    21. Parsa, A., Ibrahim, N., Hassan, B., Motroni, A., van der Stelt, P. & Wismeijer, D.

    (2013) Influence of cone beam CT scanning parameters on grey value measurements

    at implant site. Dentomaxillofacial Radiology 42:79884780.

    22. Patcas, R., Markic, G., Müller, L., Ullrich, O., Peltomäki, T., Kellenberger, C.J &

    Karlo C.A. (2012) Accuracy of linear intraoral measurements using cone beam CT

    and multidetector CT: A tale of two CTs. Dentomaxillofacial Radiology 41: 637–644.

    23. Quereshy, F.A., Savell, T.A. & Palomo, J.M. (2008) Applications of cone beam

    computed tomography in the practice of oral and maxillofacial surgery. Journal of

    Oral Maxillofacial Surgery 66: 791-796.

    24. Ranjanomennahary, P., Ghalila, S.S., Malouche, D., Marchadier, A., Rachidi, M.,

    Benhamou, C. & Chappard, C. (2011) Comparison of radiograph-based texture

    analysis and bone mineral density with three-dimensional microarchitecture of

    trabecular bone. Medical Physics 38: 420–428.

    25. Riggs, B.L., Hodgson, S.F., O’Fallon, W.M., Chao, E.Y., Wahner, H.W., Muhs, J.M.,

    Cedel, S.L. & Melton, J.M. (1990) Effect of fluoride treatment on the fracture rate in

    postmenopausal women with osteoporosis. New England Journal of Medicine 332:

    802–809.

    26. Sakka, S. & Coulthard, P. (2009) Bone quality: a reality for the process of

    osseointegration. Implant Dentistry 18: 480-485.

    27. Schulze, R., Heil, U., Gross, D., Bruellmann, D.D., Dranischnikow, E., Schwanecke,

    U. & Schoemer, E. (2011) Artefacts in CBCT: a review. Dentomaxillofacial

    Radiology 40: 265-273.

    28. Tahmaseb, A., De Clerck, R., Eckert, S. & Wismeijer, D. (2011) Reference-based

    digital concept to restore partially edentulous patients following an immediate loading

    protocol: a pilot study. International Journal of Oral Maxillofacial Implants 26: 707-

    717.

    29. Tjong, W., Kazakia, G.J., Burghardt, A.J. & Majumdar, S. (2012) The effect of voxel

    size on high-resolution peripheral computed tomography measurements of trabecular

    and cortical bone microstructure. Medical Physics 39: 1893-1903.

    30. Vigorita, V.J. (1984) The bone biopsy protocol for evaluating osteoporosis and

    osteomalacia. The American Journal of Surgical Pathology 8: 925-930.

    31. Wirth, A.J., Müller, R., van Lenthe, G.H. (2012) The discrete nature of trabecular

    bone microarchitecture affects implant stability. Journal of Biomechanics 45: 1060-

    1067.

  • 40

  • 41

    Chapter 3

    Bone quality evaluation at dental

    implant sites using Multi-slice CT,

    Micro-CT and CBCT

    This chapter has been published as:

    Parsa A, Ibrahim N, Hassan B, van der Stelt P, Wismeijer D. Bone quality evaluation

    at dental implant site using Multi-slice CT, Micro-CT and Cone Beam CT. Clinical

    Oral Implants Research. doi: 10.1111/clr.12315.

  • 42

    Bone quality evaluation at dental implant site using

    Multi-slice CT, Micro-CT and Cone Beam CT

    Azin Parsa, Norliza Ibrahim, Bassam Hassan, Paul van der Stelt,

    Daniel Wismeijer.

    Clinical Oral Implants Research. doi: 10.1111/clr.12315.

    Summary

    Objectives: The first purpose of this study was to analyze the correlation between

    bone volume fraction (BV/TV) and calibrated radiographic bone density (HU) in

    human jaws, derived from micro-CT and MSCT respectively. The second aim was to

    assess the accuracy of CBCT in evaluating trabecular bone density and microstructure

    using MSCT and micro-CT, respectively, as reference gold standards.

    Material and methods: Twenty partially edentulous human mandibular cadavers

    were scanned by three types of CT modalities: MSCT (Philips, Best, the Netherlands),

    CBCT (3D Accuitomo 170, J.Morita, Kyoto, Japan), and micro-CT (SkyScan 1173,

    Kontich, Belgium). Image analysis was performed using Amira (v4.1, Visage Imaging

    Inc., Carlsbad, CA), 3Diagnosis (v5.3.1, 3diemme, Italy), Geomagic (studio® 2012,

    Morrisville, NC), and CTAn (v1.11, SkyScan, Kontich, Belgium). MSCT, CBCT, and

    micro-CT scans of each mandible were matched to select the exact region of interest

    (ROI). MSCT HU, micro-CT BV/TV, and CBCT grey value and bone volume

    fraction of each ROI were derived. Statistical analysis was performed to assess the

    correlations between corresponding measurement parameters.

    Results: Strong correlations were observed between CBCT & MSCT density (r=0.89)

    and between CBCT and micro-CT BV/TV measurements (r=0.82). Excellent

    correlation was observed between MSCT HU and micro-CT BV/TV (r=0.91).

    However significant differences were found between all comparisons pairs (p

  • 43

    Conclusions: An excellent correlation exists between bone volume fraction and bone

    density as assessed on micro-CT and MSCT, respectively. This suggests that bone

    density measurements could be used to estimate bone microstructural parameters. A

    strong correlation was also found between CBCT grey values and BV/TV and their

    gold standards, suggesting the potential of this modality in bone quality assessment at

    implant site.

  • 44

    Introduction

    Primary implant stability is the key factor for the long term success of an implant treatment

    by improving osseointegration (Fuh et al. 2010). Primary instability of an implant induces

    movements during healing. This micro-motion leads to fibroplasia as a biological response at

    bone tissue surrounding the implant. The replacement of bone by fibrous tissue and loss of

    osseointegration cause implant failure (Lioubavina-Hack et al 2006). Bone quality, which

    refers to the combination of all bone characteristics that influence bone resistance to fracture

    (Fyhrie 2005), is one of the most important factors influencing primary implant stability

    (Ozan et al. 2007; Tolstunov 2007). Among the bone characteristics, bone mineral density

    (BMD) and trabecular microstructure are the strongest predictors for bone strength (Muller

    2003). However, these two parameters need to be simultaneously assessed to provide better

    estimation of bone strength (Teo et al. 2007; Diederichs et al. 2009).

    Several radiographic modalities have been used for bone quality assessment. For bone

    microstructure, micro-computed tomography (micro-CT) was recommended as gold standard

    for assessing bone morphology and micro architecture (Burghardt et al. 2011; Ibrahim et al.

    2013a). However, it is limited to ex-vivo small bone samples and cannot be employed for

    patients. Multiple X-ray projections with different angles in micro-CT allow a precise three-

    dimensional (3D) reconstruction of the bone samples and assessment of bone trabeculae

    (Martin-Badosa et al. 2003). Micro-CT is used to measure several histomorphometric

    variables including bone volume (BV), total volume (TV), bone volume fraction (BV/TV),

    trabecular thickness (Tb.Th), trabecular number (Tb.N) and trabecular separation (Tb.Sp)

    (Odgaard 1997).

    For bone density, multislice computed tomography (MSCT) is an established clinical

    modality in which calibrated Hounsfield units (HU) can accurately be converted to BMD

    measurements (Shahlaie et al. 2003; Shapurian et al. 2006). However, higher radiation

    exposure risk to patients in comparison with other modalities remains a main concern for

    applying MSCT for assessing bone quality (Dula et al. 1996; Ekestubbe et al. 1992;

    Ekestubbe et al. 1993; Frederiksen et al. 1995). Cone Beam Computed Tomography (CBCT),

    due to increased accessibility to dental practitioners, more compact equipment and reduced

    cost and radiation dose, has widely replaced medical CT for oral and maxillofacial imaging.

  • 45

    Several studies reported high geometric accuracy of CBCT for linear measurement

    (Lagravère et al. 2008; Lou et al. 2007; Naitoh et al. 2004), while its reliability in bone

    quality evaluation remains controversial. Only few studies suggested that CBCT could be

    applied to assess trabecular bone microstructure (Corpas Ldos et al. 2011; Liu et al. 2007).

    Additionally, CBCT does not represent calibrated voxel grey values expressed in HU (Hua et

    al. 2009).

    Yet, many attempts have been conducted to assess the feasibility of converting

    CBCT grey values to actual density measurements. High correlation between HU derived

    from MSCT and CBCT voxel grey values has been demonstrated , hinting at the potential of

    CBCT in bone density assessment (Aranyarachkul et al. 2005; Cassetta et al. 2013; Lagravère

    et al. 2006; Naitoh et al. 2009; Naitoh et al. 2010b; Nomura et al. 2010; Parsa et al. 2012;

    Reeves et al. 2012). However, the excessive scattering and technology-specific artifacts

    produced in CBCT have been denoted as the perpetrator for the unreliable BMD

    measurements (Araki et al. 2011; Hua et al. 2009; Nackaerts et al. 2011; Schulze et al. 2011;

    Yoo & Yin 2006).

    High correlation between bone volume fraction (BV/TV) provided by micro-CT and voxel

    grey value from CBCT, and also between bone volume fraction derived from CBCT and CT

    numbers from MSCT has been reported (González-García & Monje 2012; Naitoh et al.

    2010a). However the relation between bone volume fraction and radiographic bone density in

    human jaws remains controversial (Aksoy et al. 2009; Stoppie et al. 2006). Therefore, the

    first purpose of this study was to analyze the correlation between bone volume fraction

    (BV/TV) and calibrated radiographic bone density (HU) in human jaws, derived from micro-

    CT and MSCT respectively. The second aim was to assess the accuracy of CBCT in

    evaluating trabecular bone density and microstructure using MSCT and micro-CT,

    respectively, as reference gold standards.

    Material and methods

    Sample preparation and radiographic evaluation

    Twenty partially edentulous human mandibular cadavers not identified by age, sex or ethnic

    group were obtained from the functional anatomy department. The cadavers were sectioned

    at the mid-ramus level and fixed in formaldehyde (formaldehyde 74.79%, Glycerol 16.7%,

    Alcohol8.3 %, and Phenol 0.21%) and stored. A declaration was obtained from the

    Functional Anatomy department to use this human remains material for research purposes.

  • 46

    The restorative materials which can produce artifact such as amalgam filling and metal

    crowns were removed from dentitions. The mandibles were scanned by three types of CT

    modalities: MSCT (Philips, 120 kVp, 222 mA, 1.128 S, 0.67mm isotropic voxel size, Best,

    the Netherlands), CBCT (3D Accuitomo 170, 90 kVp, 5 mA, 30.8 S, 4x4 cm FOV, 0.08 mm

    isotropic voxel size, J.Morita, Kyoto, Japan), and micro-CT (SkyScan 1173,130 kVp, 61 mA,

    35 min, 35μm isotropic voxel size, Kontich, Belgium). In MSCT scans, the occlusal plane of

    each mandible was set perpendicular to the floor with zero gantry tilt, whereas in CBCT

    scans it was set parallel to the floor according to manufacturer’s recommended protocol. The

    edentulous region of each mandible was located at the center of FOV in CBCT scans. Owing

    to the large gantry of applied micro-CT (140mm in diameter, 200mm in height), mandibles

    were not sectioned to smaller samples. To prevent the possible micro movements during the

    scanning due to the large size of the samples, a cylindrical shape Styrofoam was used to fix

    and mount the sample into the holder.

    Image processing

    All CT data sets were converted to Digital Imaging and Communication in Medicine

    (DICOM3) format. As the scan orientation differs between micro-CT and the other two

    systems, the z-axis of CBCT and MSCT images were flipped to match that of micro-CT for

    further procedures. Micro-CT data sets were large in size, therefore was not possible to be

    flipped by our workstation. Image analysis was performed using Amira (v4.1, Visage

    Imaging Inc., Carlsbad, CA), 3Diagnosis (v5.3.1, 3diemme, Italy), Geomagic (studio® 2012,

    Morrisville, NC), and CTAn (v1.11, SkyScan, Kontich, Belgium). MSCT images were

    imported to 3Diagnosis software. Two cylindrical shape virtual probes (with diameter and

    height of 0.7 and 8 mm, respectively) were inserted at the edentulous region within the

    cancellous bone, with 3 mm bucco-lingual distance between them (Figure1a, b). These

    probes were used as indicators to facilitate the selection of exact region of interest (ROI)

    from MSCT, CBCT and micro-CT. For MSCT, the probes were visible in a single cross-

    sectional slice since the voxel size of MSCT scans was 0.67 mm, which is thick enough to

    allow the probes to be visible in one slice. Subsequently, a rectangular area was drawn

    between the two probes from the slice of interest to define the ROI for density measurements.

  • 47

    Figure 1. (a) Three-dimensional reconstruction of a mandible MSCT scan with inserted

    probes. (b) close-up view of probes (C) three-dimensional reconstruction of a mandible

    CBCT scan with transferred probes.

    The mean HU value from each ROI was calculated. All ROIs were totally within the

    cancellous bone, excluding cortical bone, inferior dental canal and any large bone defect.

    For CBCT, a volume-based 3D registration algorithm using Geomagic software was applied

    to transform the inserted probes from the MSCT datasets to the CBCT scans. A standard

    triangulation language (STL) surface file of the MSCT and CBCT scans were matched and

    the probes were transferred from MSCT scans to the exact region on CBCT’s (Figure1c). As

    a result, new CBCT datasets which include the probes were obtained. Using 3Diagnosis eight

    consecutive slices passing through the probes were selected from each CBCT dataset to

    calculate the mean grey values (radiological density). This is because slice thickness in

    CBCT is 0.08mm which approximately amounts to 8 times thinner than the equivalent slice

    thickness in MSCT. A rectangular region was also drawn between the two probes similar to

    MSCT and grey values from corresponding anatomical locations were derived.

    CBCT radiological density of each mandible’s ROI was considered as the mean of eight

    calculated grey values. Subsequently, the selected ROIs were saved as a bitmap (BMP) image

    files to allow the trabecular micro-structure evaluation.

    Using Amira, each micro-CT scan was cropped to have a smaller sample including the ROI.

    Due to large micro-CT data sets, the superimposition of CBCT and micro-CT scans was done

    as follows: maximum alignment of both datasets was obtained by manually matching and

    superimposition of isosurfaces generated in Amira software. Subsequently, sixteen micro-CT

    slices (correspondence to the eight CBCT slices) were selected and saved as a 16-bitmap

    (BMP) image file (65536 gray values). Then, these bmp files were exported to CTAn

    software for trabecular microstructure evaluations (Figure 2a). A rectangular ROI for

  • 48

    trabecular was selected on each dataset slice by slice (Figure 2b). All images were

    thresholded using an automated histogram analysis and binarized (Figure 2c) to allow the

    measurement process. On micro-CT datasets, the ROI was again verified by carefully

    comparing slices with CBCT’s (as reference). This was performed to reduce bias which may

    have been introduced during the manual superimposition of the two datasets. All

    measurements were performed twice with one month interval by a trained maxillofacial

    radiologist.

    Figure 2. (a) Images of micro-CT and CBCT were compared slice by slice from the same

    anatomical region. (b) A rectangular region of interest (ROI) was selected for each dataset.

    (c) Images were binarized and (d) processed to allow the microstructural measurements.

  • 49

    Data analysis

    Statistical analysis was performed using SPSS (v17.0, SPSS Inc., Chicago, IL). To determine

    the intraobserver reliability of the radiological and microstructural density measurement,

    intraclass correlation coefficient (ICC) was used. The Shapiro-Wilk test was used to verify

    the normality of the data. Paired t-test was used to assess the mean difference between MSCT

    and CBCT density measurements and between CBCT BV/TV and micro-CT BV/TV, while

    Pearson correlation coefficient was used to assess the linear relation between corresponding

    measurement parameters. Finally a Bland-Altman plot was used to assess the accuracy of

    CBCT in measuring trabecular BMD and bone microstructural density by plotting the

    difference between the measurements of CBCT against MSCT density and microCT BV/TV

    against the means of the compared measurements.

    Results

    Excellent intraobserver reliability (ICC ≥ 0.97) was revealed for repeated measurements in

    the three systems. Therefore, the mean of two measurements was calculated for further

    analysis. The mean HU of the selected ROI ranged from -60 to 507.6 (mean 222.85 &

    standard deviation [SD] 140.5) while CBCT grey values ranged from161.6 to 665.6 (mean

    377.49 & SD 127.4). The negative HU derived from MSCT for case 4, 16 and 20 (Table 1)

    may indicate fat in trabecular spaces (Parsa et al. 2012). Calculated BV/TV of the same ROI

    ranged from 2.24 to 75.83 (mean 32.35 & SD 18.81) for micro-CT and from 3.73 to 68.72

    (mean 36.79 & SD 23.17) for CBCT (Table1). Paired t-test showed significant differences (p

    < 0.001) between all comparison pairs except for mean measurement between CBCT BV/TV

    and micro CT BV/TV (p=0.147) . In all selected ROIs, CBCT showed a higher density than

    MSCT HU and a higher BV/TV than that of micro-CT. The normal distribution of

    measurements was confirmed by visually inspecting the histogram and the result of the

    Shapiro-Wilk test (p > 0.05). Therefore, the use of the t-test and Bland-Altman test is

    justified. Strong correlations were observed between CBCT and MSCT density

    measurements (r=0.89) and between CBCT and micro-CT BV/TV measurements (r=0.82).

    Excellent correlation was observed between MSCT HU and micro-CT BV/TV (r=0.91).

    Bland-Altman analysis showed the bias in measuring BV/TV between CBCT and micro-CT

    is smaller (4.44µm-1

    ) than measuring the density between CBCT and MSCT (154.65HU)

    (Figure 3a, b). The 95% measurement errors are between -21.31 to 30.19 for BV/TV and

  • 50

    29.74 to 279.56 for density measurement. The differences of CBCT and micro-CT BV/TV

    measurements were minimal (4.44µm-1

    ), suggesting strong agreement.

    Table1. Mean results of MSCT, Micro-CT and CBCT density (grey value) and bone volume

    fraction (BV/TV) measurements.

    Mandible

    No.

    MSCT HU CBCT grey value Micro-CT

    BV/TV (%)

    CBCT BV/TV

    (%)

    1 390,9 513,3 37,11 44,43

    2 205,1 444,6 41,40 46,17

    3 507,6 665,6 75,83 63,58

    4 -16,6 204,7 3,91 3,73

    5 136 324,4 17,44 9,98

    6 202,4 270,8 23,99 24,87

    7 245,2 368,5 31,68 24,68

    8 327,1 417,5 47,62 50,41

    9 341 442,9 49,34 37,94

    10 316 474,3 48,68 62,94

    11 270 379,3 50,86 48,58

    12 210,2 278 44,64 68,72

    13 320 559,1 44,12 63,28

    14 204,4 344,9 22,65 26,31

    15 285,7 357 34,57 66,29

    16 -27,6 161,6 2,24 4,33

    17 220 483,7 24,87 55,30

    18 240,4 376,7 26,97 23,21

    19 139,2 285,3 15,97 6,68

    20 -60 197,7 3,16 4,45

  • 51

    Figure 3. Bland-Altman plots of (a) BV/TV measurements between CBCT and µCT, and (b)

    density measurements between CBCT and MSCT.

    Discussion

    It has been proven that the success of an inserted implant strongly depends on the quality,

    beside the quantity, of the surrounded bone (Jaffin & Berman 1991; Jemt et al. 1992). In

    jawbones, density measurements derived from MSCT HU are highly reliable (Schwarz et al.

    1987; Shapurian et al. 2006). However, bone density alone does not fully represent bone

    quality, and should be considered together with bone microarchitecture to estimate bone

    strength and fracture resistance (Diederichs et al. 2009). Histomorphometrically, bone

    volume fraction, which is the trabecular bone volume (BV) per tissue volume (TV) expressed

    in %, is the most important parameter (Parfitt et al. 1987). Micro-CT is accepted as a gold

    standard modality for trabecular microstructure assessment, but it cannot be employed in the

    clinic (Burghardt et al. 2011). In this study our aim was to investigate the possible correlation

    between bone quality measurements of clinically applicable scanners in comparison with

    micro-CT.

    A study on porcine vertebral cancellous bone revealed a high correlation between HU derived

    from CT images and BV/TV from micro-CT and suggested the use of HU from medical CT

    for the prediction of microarchitecture (Teo et al. 2006). Our results support these findings

    that correlation between MSCT HU and Micro-CT BV/TV is high (r=0.91). However, the

    mean of calculated BV/TV in mentioned study deviated from our findings in human

  • 52

    mandibles. This could be due to different samples, ROI selections and different scanner

    systems. In similar studies using human zygomatic and jawbones, a high correlation was

    found only in female subjects (Aksoy et al. 2009; Nkenke et al. 2003). Thus, they suggested

    that only female trabecular BV/TV can be predicted from bone mineral density. In contrast,

    another study found a high correlation between BV/TV and HU in trabecular bone

    surrounded by a thin layer of cortical bone regardless to gender (Stoppie et al. 2006). This

    study suggested that with the development of MSCT scanners and imaging software, more

    precise HU measurement would be achievable (Stoppie et al. 2006). Our results showed a

    strong correlation between BV/TV and HU in human mandibular trabecular bone, regardless

    to gender and thickness of surrounding cortical bone (r=0.91). This confirms the possibility

    of prediction of bone volume fraction from MSCT bone density measurement. The usefulness

    of this prediction can be emphasized by the limitation of micro-CT in clinical settings.

    CBCT has several advantages over MSCT in terms of more compact equipment, small

    footprint for the clinic, and relatively reduced scan costs. Additionally, lower radiation dose

    levels to the main organs of the head and neck region have been cited as one of the most

    important advantages of CBCT over MSCT (Carrafiello et al. 2010; Kau et al. 2005; White

    2008). Due to these advantages, the use of this modality in dental implant planning is

    growing so fast and it is more accessible to the dental practitioners than before. Therefore, the

    validity of CBCT in bone quality assessment has been studied broadly. The majority of these

    studies have focused on the bone density measurement and found CBCT a reliable modality

    for bone density measurement (Aranyarachkul et al. 2005; Cassetta et al. 2013; Lagravère et

    al. 2006; Naitoh et al. 2009; Naitoh et al. 2010b; Nomura et al. 2010; Parsa et al. 2012;

    Reeves et al. 2012). The high correlation between measured CBCT grey values and CT

    numbers in our study (r=0.89) may confirms the possible potential of CBCT in radiographic

    density measurement. However, the limit of agreement in Bland and Altman plot (Figure 3b)

    is huge (29.74 to 279.56) with a high bias value (mean = 154.64). This indicates

    an unfavorable strength of agreement. Thus, although the measurements is reliable (ICC

    >0.97) and validated between two compared systems (r = 0.82), the density measurement

    using CBCT is less accurate when compared to its gold standard system (MSCT). It should

    be considered that CBCT density measurement can be effected by scanning parameters and

    the location of the ROI within the scanner (Nackaerts et al. 2011; Parsa et al. 2013).

  • 53

    Using micro-CT as gold standard, the reliability of CBCT in trabecular microstructure

    assessment has been validated in human mandibles, but BV/TV was not among the assessed

    microstructural parameters (Ibrahim et al. 2013b). Our results also confirm the reliability of

    CBCT in trabecular microstructure assessment, based on a high correlation between BV/TV

    measured by CBCT and micro-CT (r=0.82). The positive bias value (4.44µm-1

    ) in the Bland

    and Altman plot (Figure 3a) indicates that BV/TV was measured higher by CBCT. The small

    range between the confidence interval for the measurement differences between the two

    systems was small (-21.31 to 30.19) indicates a strong agreement between CBCT and micro-

    CT in measuring BV/TV. In present study, smallest available FOV (40x40 mm) and high

    resolution scan mode were applied in CBCT scans in order to achieve the highest possible

    spatial resolution (0.08 mm isotropic voxel size). Therefore using different CBCT scanning

    parameters the results may differ.

    It should be emphasized that the CBCT bone quality measurements in our study deviated

    from those of gold standards. This deviation arises from increased scattering, noise level and

    artefacts specific to the scanner technology which operates at lower peak kilovoltage and tube

    loading setting than MSCT and micro-CT, resulting in a reduced signal-to-noise ratio

    (Schulze et al. 2011). A higher noise level in comparison to MSCT can cause more

    inconsistencies in voxel grey values (Araki & Okano 2011; Aranyarachkul et al. 2005).

    Additionally, as the acquired volume in CBCT is larger than collimated fan beam in MSCT,

    the influence of these artifacts is excessively exacerbated (Nackaerts et al. 2011; Schulze et

    al. 2011).

    Unlike the majority of other studies on bone volume fraction, our bone samples were not

    harvested for micro-CT scans. As such, in our sample the possible deviation between the

    planned and excised ROI, which might arise during the trepanation procedure, was eliminated

    (Stoppie et al. 2006). Additionally, in the present study, a fully automated and observer

    independent 3D matching algorithm was employed for MSCT and CBCT scans registration

    to ensure that all measurements are exactly from the same site up to voxel accuracy.

    However, due to the manual alignment of CBCT and micro-CT datasets, there is a possibility

    for observer error and selection of not identical regions. Since micro-CT datasets are large

    and therefore computationally expensive, technical limitations prohibited applying the 3D

    registration algorithm for automated alignment. Technical advancements in the future might

    resolve this issue. Finally, the difference in voxel size of CBCT (0.080 mm), micro-CT

  • 54

    (0.035mm) and MSCT (0.67 mm) can also contribute to the observed discrepancy in

    calculating BV/TV and bone density.

    Voxel size in CBCT influences image quality among other factors including the unit itself,

    tube voltage and FOV selection (Kamburoglu et al. 2011). Generally, the smaller the voxels

    the higher the spatial resolution and therefore the sharper the images appear to be. However,

    small voxels result in decreased contrast to noise ratio levels and they require higher exposure

    dose to the patient (Davies et al. 2012). The higher the spatial resolution the more technical

    demands are imposed on the imaging system as a whole and on the imaging detector in

    specific to attempt to suppress noise and increase signal levels. CBCT suffers from increased

    noise levels especially at smaller voxel sizes due to low tube voltage, cone beam divergence

    phenomena and inferior detector efficiency when compared to MSCT and micro-CT (Hassan

    et al. 2010). However, the potential influence of varying voxel size on visibility of hard tissue

    structures such as bone remains largely unknown. A recent systematic review of the literature

    concluded that there is a systematic lack of evidence regarding the impact of varying voxel

    size in CBCT on diagnostic performance and that possibly different voxel sizes might lead to

    comparable diagnostic outcomes (Spin-Neto et al. 2013). Only one study could be identified

    which demonstrated a possible effect of varying voxel size on cancellous bone measurements

    in micro-CT (Yeni et al. 2005). However, it remains unknown whether the same applies to

    CBCT.

    In this study, a conscious effort was made to optimize image quality through selecting the

    scan protocols and voxels sizes as recommended by the manufacturer for the chosen FOV’s.

    Our results are limited to one CBCT system (Accuitomo 170) and results may vary on other

    systems. The design specifications of different systems still vary (De Vos et al. 2009). The

    lack of a technical standard for the development of CBCT systems has led to a wide disparity

    in the physical parameters of each model. Developing such a standard for manufacturing

    CBCT systems may help in generalizing research findings in the future. The study was also

    limited as surrounding anatomical structures including the tongue and vertebra were absent.

    As a result, in CBCT scans partial object artefacts resulting from structures placed outside the

    scan field were not simulated. It has been previously noted that artefacts resulting from

    partial sampling of objects outside the scan field could result in