assessment of trabecular bone microstructure using dental ... · 7 table of contents chapter 1...
TRANSCRIPT
-
Assessment of
Trabecular Bone Microstructure
using Dental Cone Beam CT
Norliza Ibrahim
-
2
The studies in this thesis were conducted at the section of Oral Radiology of the Academic
Centre for Dentistry Amsterdam (ACTA), Vrije University Amsterdam, The Netherlands.
The author was financially supported with a scholarship by the University of Malaya, Kuala
Lumpur, Malaysia.
Financial support for printing:
- ACTA Graduate School of Dentistry
- Oral Radiology Foundation Amsterdam (ORFA)
© Norliza Ibrahim, Amsterdam, 2014. All rights reserved.
Cover design: Norliza Ibrahim
Printed by Gildeprint Drukkerijen, Enschede.
ISBN: 9789461086082
-
3
VRIJE UNIVERSITEIT
Assessment of
Trabecular Bone Microstructure
using Dental Cone Beam CT
ACADEMISCH PROEFSCHRIFT
ter verkrijging van de graad Doctor aan
de Vrije Universiteit Amsterdam,
op gezag van de rector magnificus
prof.dr. F.A. van der Duyn Schouten,
in het openbaar te verdedigen
ten overstaan van de promotiecommissie
van de Faculteit der Tandheelkunde
op vrijdag 21 maart 2014 om 9.45 uur
in de aula van de universiteit,
De Boelelaan 1105
door
Norliza Binti Ibrahim
geboren te Selangor, Maleisië
-
4
promotor: prof.dr. P.F. van der Stelt
co promotor: dr. B.A. Hassan
-
5
To my parents,
Haji Ibrahim Sharif and
Hajjah Ramlah Arifin.
-
6
-
7
Table of Contents
CHAPTER 1 INTRODUCTION 9
CHAPTER 2 ACCURACY OF TRABECULAR BONE
MICROSTRUCTURAL MEASUREMENT USING CONE
BEAM CT DATASETS
25
CHAPTER 3 BONE QUALITY EVALUATION AT DENTAL IMPLANT
SITE USING MULTI-SLICE CT, MICRO-CT AND CBCT
41
CHAPTER 4 INFLUENCE OF SCAN PARAMETERS ON CBCT
TRABECULAR BONE MICROSTRUCTURAL
MEASUREMENTS
63
CHAPTER 5 INFLUENCE OF OBJECT LOCATION ON CBCT
MICROSTRUCTURAL ASSESSMENTS
81
CHAPTER 6 CBCT AND MICRO CT ASSESSMENTS OF TRABECULAR
BONE MICROSTRUCTURE AT DIFFERENT
MANDIBULAR REGIONS
97
CHAPTER 7 DISCUSSION 111
CHAPTER 8 SUMMARY AND CONCLUSIONS 125
CHAPTER 9 SAMENVATTING EN CONCLUSIES 129
PUBLICATIONS 133
CHAPTER 10
ACKNOWLEDGMENTS
135
-
8
-
9
Chapter 1
Introduction
Part of this chapter has been published as:
Ibrahim N, Parsa A, Hassan B, van der Stelt P, Wismeijer D. Diagnostic imaging of
trabecular bone microstructure for oral implants: a literature review.
Dentomaxillofacial Radiology 2013; 42: 20120075.
-
10
-
11
Current and future trends of CBCT in implant dentistry
Presurgical radiographic assessment is essential in planning dental implant treatment1
and dental implant thread design.2 Currently, the role of cone beam CT (CBCT) for
dental implant treatment is gradually increasing due to the wide accessibility and the
advantages obtained from these systems.3,4
CBCT assessments, however, are
generally focused on bone density5,6
and linear bone measurement.7 Structural
properties of bone are much less involved in CBCT evaluations.
The structural properties of trabecular bone are amongst the significant determinants
for bone strength. The assessment of trabecular bone structures is recommended
when predicting an implant success8,9
owing to its significant role in the healing and
osseointegration process at the implant-bone surface.10
However, studies on trabecular
microstructure assessment using CBCT are scarce because of the low scanning
resolution11
of the earlier generation of CBCT devices.
The latest CBCT equipment with a resolution of 80µm may probably serve as a 3D
imaging modality for the clinical assessment of the anisotropic trabecular
microstructures.12
Subsequently, this chapter reviews the imaging modalities for
trabecular bone microstructure in oral implants and the potential of CBCT for
microstructural assessments.
Diagnostic imaging of trabecular bone microstructure for oral implants
The term ‘bone quality’ has been extensively used in the literature to describe
different aspects of bone characteristics with variable definitions depending on the
utilized context. Among inseparable factors that influence bone quality is the
trabecular bone.13-16
The trabeculae or ‘trabecular’ is the primary anatomical and
functional unit of cancellous bone. Cortical bone helps to attain primary implant
stability, but the role of cancellous bone is also remarkable. This is because
cancellous bone has a higher bone turnover rate than cortical bone17
and has a direct
contact with majority of the implant surface.8 Accordingly, it influences the healing
and osseointegration process at the implant-bone surface.10
-
12
Bone strength has a significant role in determining implant success. To improve
prediction of bone strength, the measurements of trabecular density and trabecular
microstructure should be combined.18
This is because those measurements do not
always denote each other. For instance, a high bone density does not always
correspond to high trabecular parameters such as trabecular number (Tb.N) and
trabecular thickness (Tb.Th).19
Therefore, estimating implant success by assessing
the trabecular density alone is no longer suggested.9
Precise clinical assessment of bone structural and mechanical properties is essential in
planning dental implant treatment and implant thread design.2 The task can be
performed on two-dimensional (2D) plain radiographs (e.g. intra oral radiograph) by
calculating fractal dimensions of the trabecular bone.20
In three-dimensional (3D)
imaging modalities (e.g. HR-pQCT) the high resolution images are analyzed using
dedicated imaging software (e.g. CT Analyser [CTAn]; Skyscan®, Kontich,
Belgium). Computational techniques such as finite element methods (FEM)21,22
are
also utilized in analyzing 3D images to simulate the status of implant surface and the
bone adjacent to the implant.2
To date, bone quality assessments in oral implant studies have largely focused on
trabecular bone density.5,23-25
What follows is a review of the imaging techniques used
in oral implant studies for assessing the trabecular microstructures as evidenced in the
literature. Articles reported on trabecular microstructural imaging methods were
searched in PubMed electronic database. Titles and abstracts of the related articles
were reviewed base on keywords that had initially been set as inclusion criteria: bone
quality, imaging, trabecular microstructure, cone beam CT and dental or oral implant.
1. Dental Radiographs
Periapical (PA) and panoramic (OPG) radiographs are the first-choice diagnostic
clinical instruments in dentistry. PA radiographs with superior resolution and
sharpness provide valuable information for evaluating the amount and pattern of
trabecular bone structure.26,27
Trabecular visibility was reported to be high on PA
radiographs, 28
thus enhancing its potential in trabecular imaging studies.29-34
-
13
Bone classification systems are used to study bone quality on PA images. Of the
Lekholm & Zarb, Trisi & Rao and Misch systems, the first is largely adopted in oral
implant studies on trabecular bone assessment.29-32
A visual index was proposed in
1996 to simplify trabecular classification on PA radiographs.30
This index categorizes
trabecular pattern according to the intertrabecular spaces (small or large) and degree
of trabeculation (sparse or dense).32-34
However, these subjective techniques remain
partially validated.29
On the other hand, panoramic radiographs have also been used to
assess trabecular structure.35-36
However, the technique applies the rotational
principles that structures not centered in the focal trough are not sharply imaged. The
formation of geometrical distortion, magnification and loss of information are thus
commonly observed artifacts on panoramic radiographs. Moreover, the reduced
resolution of panoramic images degrades its ability in identifying fine trabeculae.37
Therefore, its applications in trabecular assessments are less favorable than PA
radiographs.34
Undeniably, utilizing dental radiographs for assessing trabecular microstructure is a
rapid, relatively safe and convenient method to apply in the jaws. Although the nature
of the 2D image could never provide information in the bucco-lingual direction,38
dental radiographs are still largely employed in many countries for pre implant
assessment due to availability and cost. 39
The complex shapes and structure of trabecular bone can be calculated by performing
fractal dimension (FD) analysis on 2D images such as periapical and panoramic
radiographs.37
Current studies on 2D FD analysis of trabecular microarchitecture
parameters (porosity, connectivity and anisotropy) are reported to be adequately
comparable to that of 3D FD method.40
FD analyses and calculations of trabecular
structures require several complex steps.32
Nowadays, the FD applications are
simplified by using personal computers and a simple JAVA software (Oracle®, Los
Angeles, CA). However, the overall reproducibility of the projection techniques
remains as contentious issue that requires further investigations.41
-
14
2. Magnetic Resonance Imaging (MRI)
MRI is a non-invasive, non-ionizing system which applies high magnetic fields,
transmission of radiofrequency waves and detection of radiofrequency signals from
excited hydrogen protons. Trabecular bone is filled with bone marrow that contains
free protons and generates a strong MR signal.42,43
Fat and water protons in the
marrow tissue are depicted as negative image. Because trabecular structure can not
directly be visualized, this technique employs image processing to invert the negative
image.44,45
Using this technique, values for implant loading and bone healing time at
trabecular alveolar bone were proposed to improve implant success.46
Despite
improving the trabecular structure assessment, the quality of the acquired MR images
is largely influenced by the field strength, pulse sequence, echo time, and signal to
noise ratio (SNR). Additionally, the measurements are affected by the selected
threshold values, image processing algorithms, complex analysis and interpretation of
the images.46-48
Moreover, the availability and accessibility of MRI machines for the
dental practitioners remains limited.
3. Computed Tomography (CT)
Computed tomography techniques are being progressively developed to meet the
clinical needs in assessing bone microstructure. Structural analysis of trabecular bone
requires scanners with contiguous isotropic pixel resolution of less than 300μm.49
High resolution CT systems that are commonly employed for trabecular
microstructural assessment in oral implant studies are discussed below.
3.1 Multi Detector Computed Tomography (MDCT)
The latest generation of MDCT system has improved its resolution to 150-300μm in
plane and 300-500μm in slice thickness.50
Trabecular microstructure parameters such
as trabecular number (Tb.N), trabecular thickness (Tb.Th) and trabecular separation
(Tb.Sp) were measured using MDCT and compared to high-resolution peripheral
quantitative CT (HR-pQCT).49
Although the resolution is still beyond trabecular
dimensions (50-200µm), the measurements from both techniques were highly
correlated. In a human cadaver study, trabecular microstructure parameters were
-
15
compared among MDCT and micro-CT and micro-CT FE modelling.51
The study
concluded that trabecular bone structure assessment using MDCT is overall feasible,
although still limited by its spatial resolution. These studies were conducted using a
high-resolution mode, which is not routinely used in clinical settings protocols.49-50
Consequently, although MDCT is largely employed in oral implant studies, its
applicability remains mostly confined to bone density measurements. 52-54
3.2 High-resolution peripheral Quantitative Computed Tomography (HR-
pQCT)
With a spatial resolution of 82μm, this device is used for trabecular microstructural
imaging. The measurements of microstructural parameters are reported to be
comparable with that of micro-CT (voxel size of 25μm).55
The technology has a
higher spatial resolution than MDCT; however, scanning sites are limited to
peripheral skeletal region (e.g. wrist and tibia) and accessibility is currently limited.50
Unlike MRI technique, microstructural assessment using high-resolution CT permits
direct visualization of the trabecular bone. However, the later technique involves a
relatively high radiation dose which is beyond the recommended clinical setting.45
Moreover, the results are also affected by the selected threshold, image analysis and
processing techniques.56
Thus, its application in oral implant imaging studies remains
restricted.
3.3 Micro Computed Tomography (Micro-CT)
Two-dimensional histomorphometric analysis was previously considered as the gold
standard for assessing trabecular size, shape, connectivity and orientation. As it is
time consuming and costly, micro-CT is now routinely employed for structural 2D or
3D evaluations of trabecular microstructure.50,57
This non-destructive high resolution
(approaching 10µm) method depicts trabecular network in different gray levels
according to its mineral content. It has been reported that the trabecular parameters
quantified by micro-CT are comparable to the traditional 2D histomorphometric
values.42,43
As it permits high resolution scans, in 2004 micro-CT has been
recommended as a gold standard imaging for ex-vivo bone studies at implant sites.57
-
16
However, only studies with small jaw specimens were conducted to observe
trabecular microstructure in oral implant research.5,8,19,58
3.4 Cone beam computed tomography (CBCT)
Cone beam computed tomography systems were developed in the1990s. In 2001,
CBCT was introduced as a 3D imaging modality. Since then it has largely replaced
both single- and multi- slice CT for diagnostic imaging in oral implants. 59
Owing to
the wide availability of the machines, rapid scan and processing time, high resolution
images and relatively reduced scan radiation dose and costs, the demand for CBCT
images preceding implant placement has increased exponentially.3,60-63
Although
many studies have been conducted on CBCT, the literature on its suitability in
measuring trabecular bone microstructural parameters at oral implant site remains
scarce. This may be due to the insufficient resolution of the past generations of CBCT
systems to depict bone microstructure. The applications of CBCT in evaluating bone
quality are still restricted on bone density assessment.52-54
Recently, however, a study
on assessing bone microstructure described CBCT as a promising modality for
analyzing trabecular bone.63
Bone parameters (Tb.Th, Tb.N and Tb.Sp) at mandibular
condyle were also successfully evaluated by CBCT at a resolution of 125µm coupled
with image processing.64
The visibility of small anatomical structures with CBCT is largely influenced by the
field of view (FOV) and scan setting selection.65
Visibility of trabecular
microstructure is mainly determined by the chosen voxel size and SNR plus image
artifacts.66
In CBCT, voxel size and slice thickness, spatial and contrast resolutions
vary with respect to machine type, FOV and scan settings. 65,66
Additionally, several
image artifacts specific to CBCT technology could influence the effective system
resolution, which could be lesser than the nominal system resolution expressed in
voxel size alone. It has been previously stated that the accuracy of 3D measurement of
anisotropic trabecular structure can be improved by performing in-vivo rather than in-
vitro investigation.16,18
In this respect, the use of CBCT could prove appealing. As the
need to evaluate the implant insertion sites prior to surgical placement has
dramatically increased, CBCT should be validated as a non-invasive procedure for
assessing bone microstructure.
-
17
Figure 1 Sagittal images of trabecular structure at the lingual foramen region derived
from A: MDCT (650μm), B: CBCT (80μm) and C: Micro CT (35μm).
Müller et al. has described that a CT scanner with a resolution up to 60μm can present
morphometric information comparable to that of 10 μm.67
Using the latest CBCT
system, the appearance of trabecular structure was observed using a 4x4cm FOV at a
nominal resolution of 80µm. The resultant image was compared with images derived
from MDCT and micro-CT (Figure 1). It is expected that this system could be useful
in measuring trabecular microstructures. However, thorough investigation and
validation are required prior to applying this technique in clinical practice.
Although there is a rapid progress in advanced bone imaging modalities, their routine
clinical employment remains limited due to the technical features, cost and complex
procedures. The current review recommends studies to validate CBCT as a clinical
imaging modality to evaluate trabecular microstructure at oral implant sites.
Aims of the thesis:
The primary aim of this thesis is to validate the applicability and accuracy of CBCT
for trabecular bone microstructural assessments. Additionally, the consistency of
structural analysis is assessed using various scanning protocols, different sites of the
mandible and different locations of the object in the CBCT field FOV. Therefore, the
current studies are conducted according to the following research questions:
-
18
1. What is the accuracy of CBCT trabecular bone microstructure measurements in
comparison with µCT?
2. What is the accuracy of CBCT trabecular bone density and bone volume fraction
in comparison with MSCT and µCT?
3. What is the effect of scan parameters (FOV, rotation steps and resolution) on
CBCT trabecular bone microstructural measurements?
4. What is the influence of the object location on CBCT trabecular bone
microstructure measurements?
5. What is the difference between CBCT and µCT in measuring trabecular bone
microstructure at different sites of edentulous mandibles?
-
19
References:
1. Jacobs R, Van Steenberghe D. Radiographic planning and assessment of endosseous
oral implants. Berlin: Springer, 1997.
2. DeTolla DH, Andreana S, Patra A, Buhite R, Comella B. Role of the finite element
model in dental implants. J Oral Implantol 2000; 26: 77-81.
3. De Vos W, Casselman J, Swennen GR. Cone-beam computerized tomography
(CBCT) imaging of the oral and maxillofacial region: a systematic review of the
literature. Int J Oral and Maxillofac Surg 2009; 38: 609-625.
4. Guerrero ME, Jacobs R, Loubele M, Schutyser F, Suetens P, van Steenberghe D.
State-of-the-art on cone beam CT imaging for preoperative planning of implant
placement. Clin Oral Investig 2006; 10: 1-7.
5. González-García R, Monje F. The reliability of cone-beam computed tomography to
assess bone density at dental implant recipient sites: a histomorphometric analysis by
micro-CT. Clin Oral Implants Res 2012 doi: 10.1111/j.1600-0501.2011.02390.x.
6. Monje A, Monje F, González-García R, Galindo-Moreno P, Rodriguez-Salvanes F,
Wang HL. Comparison between microcomputed tomography and cone-beam
computed tomography radiologic bone to assess atrophic posterior maxilla density
and microarchitecture. Clin Oral Implants Res. 2013 Feb 26. doi: 10.1111/clr.12133.
7. Quereshy FA, Savell TA, Palomo JM. Applications of cone beam computed
tomography in the practice of oral and maxillofacial surgery. Journal of Oral
Maxillofacial Surgery 2008; 66: 791-796.
8. Fanuscu MI, Chang TL. Three-dimensional morphometric analysis of human cadaver
bone: microstructural data from maxilla and mandible. Clin Oral Impl Res 2004; 15:
213-218.
9. Wirth AJ, Goldhahn J, Flaig C, Arbenz P, Müller R, van Lenthe GH. Implant stability
is affected by local bone microstructural quality. Bone 2011; 49: 473-478.
10. Minkin C, Marinho VC. Role of the osteoclast at the bone-implant interface. Adv Dent
Res 1999; 13: 49-56.
11. Kothari M, Keaveny TM, Lin JC, Newitt DC, Genant HK, Majumdar S. Impact of
spatial resolution on the prediction of trabecular architecture parameters. Bone 1998;
22: 437-443.
12. Müller R, Van Campenhout H, Van Damme B, Van Der Perre G, Dequeker J,
Hildebrand T, et al. Morphometric analysis of human bone biopsies: a quantitative
-
20
structural comparison of histological sections and micro-computed tomography. Bone
1998; 23:59–66.
13. Sievänen H, Kannus P, Järvinen TLN. Bone quality: An empty term. PLoS Med 2007;
4: e27.
14. Compston J. Bone Quality: What is it and how is it measured? Arq Bras Endocrinol
Metab 2006; 50: 579-585.
15. Licata A. Bone density vs bone quality: What’s a clinician to do? Cleve Clin J Med
2009; 76: 331-336.
16. Fyhrie DP. Summary - Measuring ‘‘bone quality’’. J Musculoskelet Neuronal Interact
2005; 5: 318-320.
17. Sakka S, Coulthard P. Bone quality: a reality for the process of osseointegration.
Implant Dent 2009; 18: 480-485.
18. Müller R. Bone microarchitecture assessment: current and future trends. Osteoporos
Int 2003; 14 Suppl 5:S89-S95.
19. Gomes de Oliveira RC, Leles CR, Lindh C, Ribeiro-Rotta RF. Bone tissue
microarchitectural characteristics at dental implant sites. Part 1: Identification of
clinical-related parameters. Clin Oral Implants Res 2011; 000-000. doi:
10.1111/j.1600-0501.2011.02243.x.
20. Yi WJ, Heo MS, Lee SS, Choi SC, Huh KH. Comparison of trabecular bone
anisotropies based on fractal dimensions and mean intercept length determined by
principal axes of inertia. Med Biol Eng Comput 2007; 45: 357-364.
21. Krug R, Burghardt AJ, Majumdar S, Link TM. High-resolution imaging techniques
for the assessment of osteoporosis. Radiol Clin North Am 2010; 48: 601-621.
22. Müller R, van Lenthe GH. Trabecular bone failure at the microstructural level. Curr
Osteoporos Rep 2006; 4: 80-86.
23. Turkyilmaz I, McGlumphy EA. Influence of bone density on implant stability
parameters and implant success: a retrospective clinical study. BMC Oral Health
2008; 8: 32.
24. Turkyilmaz I, Sennerby L, McGlumphy EA, Tözüm TF. Biomechanical aspects of
primary implant stability: a human cadaver study. Clin Implant Dent Relat Res 2009;
11: 113-119.
25. Molly L. Bone density and primary stability in implant therapy. Clin Oral Implants
Res 2006; 17: 124-135.
-
21
26. White SC, Pharoah MJ. Oral radiology: principles and interpretation (5th edn). St.
Louis: Mosby, 2004.
27. Whaites E. Essentials of dental radiography and radiology (3rd edn). London:
Churchill Livingstone, 2007.
28. Couture RA, Whiting BR, Hildebolt CF, Dixon DA. Visibility of trabecular structures
in oral radiographs. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2003; 96:
764–771.
29. Ribeiro-Rotta RF, Lindh C, Pereira AC, Rohlin M. Ambiguity in bone tissue
characteristics as presented in studies on dental implant planning and placement: a
systematic review. Clin Oral Implants Res 2011; 22: 789-801.
30. Lindh C, Petersson A, Rohlin M. Assessment of the trabecular pattern before
endosseous implant treatment: diagnostic outcome of periapical radiography in the
mandible. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 1996; 82: 335–343.
31. Aalam AA, Nowzari H, Krivitsky A. Functional restoration of implants on the day of
surgical placement in the fully edentulous mandible: a case series. Clin Implant Dent
Relat Res 2005; 7: 10-16.
32. Jonasson G, Jonasson G, Jonasson L, Kiliaridis S. Skeletal bone mineral density in
relation to thickness, bone mass, and structure of the mandibular alveolar process in
dentate men and women. Eur J Oral Sci 2007; 115: 117-123.
33. Lindh C, Horner K, Jonasson G, Olsson P, Rohlin M, Jacobs R, et al. The use of
visual assessment of dental radiographs for identifying women at risk of having
osteoporosis: the OSTEODENT project. Oral Surg Oral Med Oral Pathol Oral
Radiol Endod 2008; 106: 285–293.
34. Pham D, Jonasson G, Kiliaridis S. Assessment of trabecular pattern on periapical and
panoramic radiographs: A pilot study. Acta Odontol Scand 2010; 68: 91-97.
35. Taguchi A, Suei Y, Ohtsuka M, Otani K, Tanimoto K, Ohtaki M. Usefulness of
panoramic radiography in the diagnosis of postmenopausal osteoporosis in women.
Width and morphology of inferior cortex of the mandible. Dentomaxillofac Radiol
1996; 25: 263-267.
36. Watanabe PCA, Oliveira TM, Monteiro SAC, Iyomasa MM, Regalo SCH, Siessere S.
Morphodigital study of the mandibular trabecular bone in panoramic radiographs. Int
J Morphol 2007; 25: 875-880.
37. Bollen AM, Taguchi A, Hujoel PP, Hollender LG. Fractal dimension on dental
radiographs. Dentomaxillofac Radiol 2001; 30: 270-275.
-
22
38. Lofthag-Hansen S, Gröndahl K, Ekestubbe A. Cone-beam CT for preoperative
implant planning in the posterior mandible: visibility of anatomic landmarks. Clin
Implant Dent Relat Res 2009; 11: 246-255.
39. Sakakura CE, Morais JA, Loffredo LC, Scaf G.A survey of radiographic prescription
in dental implant assessment. Dentomaxillofac Radiol 2003; 32: 397-400.
40. Pothuaud L, Benhamou CL, Porion P, Lespessailles E, Harba R, Levitz P. Fractal
dimension of trabecular bone projection texture is related to three-dimensional
microarchitecture. J Bone Miner Res 2000; 15: 691-699.
41. Sánchez I, Uzcátegui G. Fractals in dentistry. J Dent 2011; 39: 273-292.
42. Genant HK, Jiang Y. Advanced imaging assessment of bone quality. Ann N Y Acad
Sci 2006; 1068: 410-428.
43. Lespessailles E, Chappard C, Bonnet N, Benhamou CL. Imaging techniques for
evaluating bone microarchitecture. Joint Bone Spine 2006; 73: 254-261.
44. Jiang Y, Zhao J, White DL, Genant HK. Micro CT and Micro MR imaging of 3D
architecture of animal skeleton. J Musculoskelet Neuronal Interact 2000; 1: 45-51.
45. Ito M. Recent progression in assessment of bone microstructure. Nihon Rinsho 2011;
691233-691238.
46. Choël L, Last D, Duboeuf F, Seurin MJ, Lissac M, Briquet A, et al. Trabecular
alveolar bone microarchitecture in the human mandible using high resolution
magnetic resonance imaging. Dentomaxillofac Radiol 2004; 33: 77–182.
47. Sell CA, Masi JN, Burghardt A, Newitt D, Link TM, Majumdar S. Quantification of
trabecular bone structure using magnetic resonance imaging at 3 Tesla-calibration
studies using microcomputed tomography as a standard of reference. Calcif Tissue Int
2005; 76: 355-364.
48. Celenk P, Celenk C. Evaluation by quantitative magnetic resonance imaging of
trabecular bone quality in mandible and cervical vertebrae. Clin Oral Implants Res
2010; 21: 409-413.
49. Issever AS, Link TM, Kentenich M, Rogalla P, Burghardt AJ, Kazakia GJ, et al.
Assessment of trabecular bone structure using MDCT: comparison of 64- and 320-
slice CT using HR-pQCT as the reference standard. Eur Radiol 2010; 20: 458-468.
50. Burghardt AJ, Link TM, Majumdar S. High-resolution computed tomography for
clinical imaging of bone microarchitecture. Clin Orthop Relat Res 2011; 469: 2179-
2193.
-
23
51. Issever AS, Link TM, Kentenich M, Rogalla P, Schwieger K, Huber MB, et al.
Trabecular bone structure analysis in the osteoporotic spine using a clinical in vivo
setup for 64-slice MDCT imaging: comparison to micro CT imaging and micro FE
modelling. J Bone Miner Res 2009; 24:1628-1637.
52. Naitoh M, Aimiya H, Hirukawa A, Ariji E. Morphometric analysis of mandibular
trabecular bone using cone beam computed tomography: an in vitro study. Int J Oral
Maxillofac Implants 2010; 25: 1093-1098.
53. Naitoh M, Hirukawa A, Katsumata A, Ariji E. Prospective study to estimate
mandibular cancellous bone density using large-volume cone-beam computed
tomography. Clin Oral Implants Res 2010; 21: 1309-1313.
54. Araki K, Okano T. The effect of surrounding conditions on pixel value of cone beam
computed tomography. Clin Oral Impl Res 2011; 00:1-4.
55. Liu XS, Zhang XH, Sekhon KK, Adams MF, McMahon DJ, Bilezikian JP, et al.
High-Resolution Peripheral Quantitative Computed Tomography can assess
microstructural and mechanical properties of human distal tibial bone. J Bone Miner
Res 2010; 25: 746–756.
56. MacNeil JA, Boyd SL. Accuracy of high-resolution peripheral quantitative computed
tomography for measurement of bone quality. Med Eng Phys 2007; 29: 1096-1105.
57. Yip G, Schneider P, Roberts EW. Micro-Computed Tomography: High resolution
imaging of bone and implants in three dimensions. Semin Orthod 2004; 10: 174-187.
58. Swain MV, Xue J. State of the art of micro CT applications in dental research. Int J
Oral Sci 2009; 1: 177-188.
59. Hatcher DC. Operational principles for cone-beam computed tomography. J Am Dent
Assoc 2010; 141 Suppl 3:3S-6S.
60. Al-Rawi B, Hassan B, Vandenberge B, Jacobs R. Accuracy assessment of three-
dimensional surface reconstructions of teeth from Cone Beam Computed Tomography
scans. J Oral Rehabil 2010; 37: 352-358.
61. Kau CH, Bozic M, English J, Lee R, Bussa H, Ellis RK. Cone-beam computed
tomography of the maxillofacial region - an update. Int J Med Robot 2009; 5: 366-
380.
62. Liang X, Jacobs R, Hassan B, Li L, Pauwels R, Corpas L, et al. A comparative
evaluation of Cone Beam Computed Tomography (CBCT) and Multi-slice CT
(MSCT) Part I. on subjective image quality. Eur J Radiol 2010; 75: 265-269.
-
24
63. Corpas LS, Jacobs R, Quirynen M, Huang Y, Naert I, Duyck J. Peri-implant bone
tissue assessment by comparing the outcome of intra-oral radiograph and cone beam
computed tomography analyses to the histological standard. Clin Oral Impl Res 2011;
22: 492-499.
64. Liu SM, Zhang ZY, Li JP, Liu DG, Ma XC. A study of trabecular bone structure in
the mandibular condyle of healthy young people by cone beam computed
tomography. Zhonghua Kou Qiang Yi Xue Za Zhi 2007; 42: 357-360.
65. Loubele M, Jacobs R, Maes F, Denis K, White S, Coudyzer W, et al. Image quality vs
radiation dose of four cone beam computed tomography scanners. Dentomaxillofac
Radiol 2008; 37: 309-318.
66. Schulze R, Heil U, Gross D, Bruellmann DD, Dranischnikow E, Schwanecke U, et al.
Artefacts in CBCT: a review. Dentomaxillofac Radiol 2011; 40: 265-273.
67. Müller R, Koller B, Hildebrand T, Laib A, Gianolini S, Rüegsegger P. Resolution
dependency of microstructural properties of cancellous bone based on three-
dimensional μ-tomography. Technol Health Care 1996; 4: 113–119.
-
25
Chapter 2
Accuracy of trabecular bone
microstructural measurement using
cone-beam CT datasets
This chapter has been published as:
Ibrahim N, Parsa A, Hassan B, van der Stelt P, Aartman IHA,Wismeijer D. Accuracy of
trabecular bone microstructural measurement at planned dental implant sites using cone-beam
CT datasets. Clinical Oral Implants Research. doi: 10.1111/clr.12163.
-
26
Accuracy of trabecular bone microstructural
measurement at planned dental implant sites using
cone-beam CT datasets
Norliza Ibrahim, Azin Parsa, Bassam Hassan, Paul van der Stelt, Irene
HA Aartman, Daniel Wismeijer.
Clinical Oral Implants Research. doi: 10.1111/clr.12163
Summary
Objective: Cone-beam CT (CBCT) images are infrequently utilized for trabecular
bone microstructural measurement due to the system's limited resolution. The aim of
this study was to determine the accuracy of CBCT for measuring trabecular bone
microstructure in comparison with micro CT (μCT).
Materials and methods: Twenty-four human mandibular cadavers were scanned
using a CBCT system (80μm) and a μCT system (35μm). Three bone microstructural
parameters trabecular number (Tb.N), thickness (Tb.Th) and separation (Tb.Sp) were
assessed using CTAn imaging software.
Results: Intraclass correlation coefficients (ICC) showed a high intra-observer
reliability (≥ 0.996) in all parameters for both systems. The Pearson correlation
coefficients between the measurements of the two systems were for Tb.Th 0.82, for
Tb.Sp 0.94 and for Tb.N 0.85 (all P's
-
27
Introduction
The influence of bone quality on implant success is well acknowledged. Information of bone
quality is best gained by combining bone mineral density (BMD) and trabecular structure
assessments (Felsenberg & Boonen 2005). Trabecular bone is the source of osteoblasts and
osteoclasts which are largely responsible for the physiological changes which take place
subsequent to implant placement (Minkin & Marinho 1999).
The role of trabecular
microstructure in dental implants is significant since the implant is surrounded by trabecular
bone which directly contributes to implant stability (Fanuscu & Chang 2004; Sakka &
Coulthard 2009).
Radiographic information of the trabecular microstructure can be achieved by using high
resolution imaging modalities that approach the trabecular dimensions (50-300μm).
However, most of the utilized modalities have limitations in clinical practice. High resolution
radiographic systems’ limitations rest on excessive radiation exposure to the patient (MDCT,
HR-pQCT), complex image analysis (hr-MRI), restricted accessibility (MDCT, hr MRI) and
limited scanning sites (HR-pQCT, µCT) (Genant et al. 2000; Issever et al. 2010).
Since 2001, cone-beam CT (CBCT) has been used to provide three-dimensional images for
diagnosis and treatment planning in dentistry (Hatcher 2010). It provides comparable images
at reduced scan costs and dose. Additionally, CBCT systems are largely accessible to dental
professionals and the scan time is relatively short (Patcas et al. 2012). Numerous CBCT
studies were conducted using CBCT for dental implants. However, up to date the focus has
mainly been on bone quantity (alveolar bone width and height) measurement accuracy, bone
density, visibility of anatomical landmarks and virtual guided surgery (Quereshy et al. 2008;
Tahmaseb et al. 2011). Recently, CBCT has been suggested as a modality for analysing
trabecular microstructure at implant sites (Corpas et al. 2011; Ibrahim et al. 2012). To the best
of our knowledge, there has been no report validating the ability of this system in measuring
trabecular bone microstructural parameters. Therefore, the aim of this study was to determine
the accuracy of CBCT in trabecular bone microstructure measurement in comparison with
µCT, after first assessing the intraobserver reliability.
-
28
Materials and methods
Image acquisition
Twenty-four human mandibular cadavers were obtained from the Department of Functional
Anatomy. Approval was obtained from the department to use this material for research
purposes. All cadavers were scanned using a CBCT system (3D Accuitomo 170, Morita,
Japan). To obtain the highest spatial resolution possible of an isotropic 80µm, the smallest
field of view (FOV), that is, 4x4cm with the high-resolution scan mode and 360º arm
rotation, were the selected scan settings. Images were acquired at 90kv and 5.0mA. The
cadavers were then scanned using a µCT system (SkyScan 1173, Kontich, Belgium). This
system allows scanning of large size specimens (140mm in diameter, 200mm in height).
Hence the mandibles were not sectioned into smaller sizes. To reduce any possible
movements of the samples during scanning, the samples were fixed in a cylindrical shape of
Styrofoam, fit and mounted into the holder before scanning. The resultant images were again
checked for motion artifacts. The settings were set by the manufacturer operating at 130kVp,
61mA with nominal isotropic resolution of 35µm.
Figure 1: A mandible scanned using cone beam CT (CBCT) was cropped into a smaller
block in Amira software.
-
29
Image processing
The datasets from both systems were exported as DICOM 3 files and imported into image
analysis software (Amira v4.1, Visage Imaging Inc., Carlsbad, CA). To select the exact
region of interest (ROI) from both systems, the datasets were processed as follows. The three-
dimensional (3D) isosurfaces datasets from both µCT and CBCT were created and saved as
STL files. Subsequently, the mandibles were cropped (confined) to the edentulous posterior
region (Fig. 1). The z-axis of the CBCT images was flipped to match that of µCT because the
scan orientation differs between the two systems. The isosurfaces were used to provide
coordinates for matching the original volumes (voxel data) of both systems. The bone blocks
were then manually superimposed on each other to provide maximum alignment (Fig. 2). The
µCT was set as the reference standard while CBCT was considered as the experimental
modality. For each matching set, a smaller ROI was selected confined in the vertical and
horizontal slice directions to the cortical bone margins. For µCT ten and for CBCT five
consecutive and contiguous slices were chosen for evaluation. Since µCT isotropic voxel size
(35µm3) is approximately half of that for CBCT (80µm
3), a double amount of slices was
selected to ensure that the measurements are from the same region. The selected slices were
exported as BMP images and imported into trabecular bone analysis software CTAn (v 1.11,
SkyScan, Kontich, Belgium). Through histogram analysis, the images were thresholded and
binarized (Fig. 3). The selected CBCT and micro CT image slices were then averaged to
reduce the bias that may occurred during the manual registration procedure in Amira
software. Twenty-four ROI were carefully matched and compared prior to the measurement
process (Fig. 4). The regions disturbed by metal artifacts or cut off during the µCT scanning
process were excluded. All measurements were performed by one trained observer twice for
both systems independently with at least one week interval between the first and the second
measurement to assess the observer’s bias.
-
30
Figure 2: A block of cropped
CBCT image was manually
transformed, superimposed
and matched onto µCT to
select the region of interest.
(ROI).
Figure 3: Binarized images of
CBCT (a) and µCT (b).
Figure 4: Trabecular bone
microstructure parameters
were measured and
compared using CBCT (A)
and µCT (B) datasets.
-
31
Statistical analysis
The bone microstructural parameters that were considered for evaluations were the trabecular
number (Tb.N), thickness (Tb.Th) and separation (Tb.Sp). The intraclass correlation
coefficient (ICC) was used to determine the intraobserver reliability in measuring trabecular
microstructure using CBCT and µCT. Paired t-tests were used to assess the difference in
means between the two systems and the linear relation between both systems was assessed
using the Pearson correlation coefficient (r). Finally a Bland-Altman plot was used to assess
the accuracy of CBCT in measuring the microstructural parameters by plotting the difference
between the measurements of CBCT and µCT against the means of those measurements (Fig.
5).
Figure 5: Bland-Altman plot of three bone microstructural parameters trabecular number
(Tb.N), thickness (Tb.Th) and separation (Tb.Sp). Tb.Th (a), Tb.Sp (b) and Tb.N (c)
measurements between CBCT and µCT.
-
32
Results
The ICC’s revealed excellent intraobserver reliability for all parameters and both systems
separately (Table 1). Therefore the first measurement was used for the following analyses.
Paired t-tests showed significant differences between the CBCT and µCT for all parameters.
Tb.Th and Tb.Sp were higher on CBCT images than on µCT images, but Tb.N was lower by
CBCT (Table 2). The Pearson correlation coefficients for the relation between CBCT and
µCT were all significant at p< 0.001 (Table 2). Bland-Altman analysis showed the smallest
bias of measurements in Tb.N (-0.37 µm-1
) followed by Tb.Th (1.6µm) and Tb.Sp (8.8µm).
The confidence interval for the measurement differences between the two systems was
smallest for Tb.N (-0.79 to +0.06), followed by Tb.Th (-2.1 to +5.3) and Tb.Sp (-21.4 to
+43.1).
Table 1. Intraobserver reliability for CBCT and µCT using the intraclass correlation
coefficient (ICC).
N=24
ICC
CBCT µCT
Tb.Th 0.998 0.999
Tb.Sp 0.997 0.999
Tb.N 0.999 0.996
Three bone microstructural parameters: trabecular number (Tb.N), thickness (Tb.Th) and
separation (Tb.Sp).
-
33
Table 2: Structural parameters, paired t-test, and correlation between measurements obtained
by CBCT and µCT.
Parameters Means and standard
deviation
Paired t-test Pearson ( r2 )
CBCT µCT means sd SEM t df p
Tb.Th (µm) 0.41 + 0.15 0.28 +0.10 0.13 0.08 0.02 7.48 23 0.001 0.67*
Tb.Sp (µm) 0.85 +0.44 0.71 +0.39 0.15 0.15 0.03 4.83 23 0.001 0.89*
Tb.N (µm-1) 0.87 +0.26 1.11 +0.29 -0.24 0.16 0.03 -7.35 23 0.001 0.72*
Three bone microstructural parameters: trabecular number (Tb.N), thickness (Tb.Th) and
separation (Tb.Sp).
* Correlation is significant at the 0.001 level
Discussion
Bone quality evaluations using CBCT are usually concentrated on bone density (Corpas et al.
2011; Gonzalez-Garcia & Monje 2012; Ibrahim et al. 2012; Parsa et al. 2012). However, in
order to understand its influence on implant-bone integration, bone quality should also be
assessed from the microstructural aspect of trabecular bone (Fanuscu & Chang 2004;
Gonzalez-Garcia & Monje 2012). Apart from its role in bone healing and implant retention
(Minkin & Marinho 1999), trabecular bone microstructure also contributes to bone strength
(Felsenberg & Boonen 2005; Manske et al. 2010). Amongst the recommended
microstructural parameters to estimate bone strength are trabecular thickness (Tb.Th),
number (Tb.N), and spacing between each other (Tb.Sp). Bone density and trabecular
microstructure measurements do not always correlate with each other (Gomes de Oliveira et
al. 2012). For instance, a low-density bone may be associated with low Tb.Th (Hildebrand et
al. 1999; Ranjanomennahary et al. 2011) and Tb.N, but high Tb.Sp (Hans et al. 2011;
Ranjanomennahary et al. 2011). But, after a series of medications in osteoporotic patients, the
increased bone density did not improve its bone strength (Riggs et al. 1990). Therefore, in
selecting the best bone quality prior to dental implant treatment, the trabecular microstructure
should also be included as part of the pre-surgical bone assessment.
-
34
µCT is a gold standard modality for trabecular microstructural assessment. However, it can
only evaluate small bone samples and it cannot be employed to scan patients in a clinical
setting (Burghardt et al. 2011). The use of CBCT in dental implants has increased, but its
ability in analyzing the bone microstructural parameters at implant receptor site remains
unverified. At present, only few studies reported that CBCT has potentials in measuring
trabecular microstructure (Corpas et al. 2011; Liu et al. 2007; Ibrahim et al. 2013). In this
study, Tb.N, Tb.Th and Tb.Sp at posterior mandibular regions were directly assessed,
analyzed and compared between CBCT and µCT datasets using human mandibular cadavers
at implant receptor sites. A strong correlation was observed when comparing the trabecular
bone microstructure measurements with µCT. Bland-Altman plots show strong agreements
between CBCT and µCT when measuring Tb.N (-0.37µm), Tb.Th (1.6µm) and Tb.Sp
(8.8µm). Only Tb.N was measured smaller by CBCT (indicated by a negative bias value)
while Tb.Th and Tb.Sp were measured greater than µCT (indicated by positive bias values).
This discrepancy is in accordance to the voxel size of CBCT (80µm) which is twice that of
µCT (35µm). Since the voxel in CBCT is significantly larger than its counterpart in µCT, the
system can only depict significantly thick trabeculae, which in turn results in a smaller
trabecular number measurement. This is in accordance with a study on trabecular
microstructural parameters which also concluded that small trabeculae are poorly depicted by
low resolution systems when comparing MDCT (274µm) with µCT (16µm) (Issever et al.
2010) and HR-pQCT (82µm) with µCT (18µm) (Tjong et al. 2012). The latter study
described that partial volume effects (PVE) were increased and a rough estimation of Tb.Th
was observed when bigger voxel size images were utilized when measuring microstructural
parameters. However, it has to be emphasized that while spatial resolution expressed in voxel
size is a limiting factor for image quality, it’s not the only contributing factor. Contrast to
noise ratio (CNR) plays an equally important role in determining the capability of an imaging
system for depicting delicate anatomical structures. A sufficiently high CNR can possibly
compensate for a relatively low spatial resolution (Bechara et al. 2012). Therefore, thin
trabeculae can be made visible even when the trabecular thickness itself is beyond the
nominal voxel size. Image quality is also largely affected by micro-movement of the jaw. The
stationary situation of the cadavers in the present study might have thus enhanced the
visibility of the trabecular structures because the scans did not suffer from patient’s
movement.
-
35
One difficulty in this study was triggered during identifying the ROI. The projections in
CBCT were not completely similar to µCT because the mandibles were positioned vertically
in µCT to meet the space constraints in the scanner. Through using 3D virtual
reconstructions, CBCT and µCT were brought into maximum alignment. The measurements
could still somewhat vary due to the limitation of the non-automated registration. However,
when an automated registration is not attainable, Diederichs et al. (2008) had suggested that
both scans should be performed at a similar position to improve the accuracy of the selected
ROIs. Further, prior to microstructural assessment in CTAn software, the visualized
trabecular bone was again compared slice by slice to ascertain that the ROI was
correspondingly selected.
The size of the ROI and density of the bone strongly influence trabecular bone assessments.
A diameter of 5mm or larger bone specimen is recommended to adequately analyze bone
parameters (Vigorita 1984). In this study we used the full height and width of the bone
sections sampled at 5 and 10 consecutive and contiguous coronal slices for CBCT and µCT,
respectively (Fig. 3). Nevertheless, this careful sampling still does not guarantee that all
measurements were from anatomically identical regions. This is due to the manual nature of
aligning the two datasets and the possibility for observer error. An automated, observer-
independent 3D volumetric matching using software tool is recommended to improve the
alignment accuracy (Li et al. 2008). However, this was not possible in this study due to
technical constrains. At any rate, the measurements were repeated twice and the ICC results
show strong agreements between the first and the second measurement (Table 1).
Apart from voxel size (Tjong et al. 2012); the accuracy of structural measurement is also
influenced by the threshold selection (Parkinson et al. 2008). Bone segmentation with CBCT
remains difficult as image quality is affected by artifacts specific to the scanning technology
(Parsa et al. 2013; Schulze et al. 2011). CBCT has a lower signal to noise ratio than µCT, a
larger amount of scatter radiation, reduced contrast and is largely susceptible to beam
hardening and edge aliasing artifacts (Schulze et al. 2011). All these combined factors
introduce errors in the grey level values in CBCT which results in underestimation of thin
trabecular bone. Therefore, it was recommended to use denser bone specimens to overcome
this problem (Wirth et al. 2012). In this study all 24 mandibles were used regardless of their
density. Owing to its extremely high resolution, µCT was able to image bone specimens even
for poor bone density blocks. However, the same could not be stated for CBCT since the
-
36
partial volume effect is accentuated for sparse trabecular patterns. Therefore, trabecular
spacing in particular was overestimated since many small trabeculae were thresholded out
when binarizing the images. This could possibly explain the outliers depicted in the Bland-
Altman plots (Fig. 5). Only small discrepancies were observed (Tb.N
-
37
References:
1. Bechara, B., McMahan, C.A., Moore, W.S., Noujeim, M., Geha, H. & Teixeira, F.B.
(2012) Contrast- to-noise ratio difference in small field of view cone beam computed
tomography machines. Journal of Oral Science 54: 227–232.
2. Burghardt, A.J., Link, T.M. & Majumdar, S. (2011) High-resolution computed
tomography for clinical imaging of bone microarchitecture. Clinical Orthopaedics
and Related Research 469: 2179–2193.
3. Corpas, L.S., Jacobs, R., Quirynen, M., Huang, Y., Naert, I. & Duyck, J. (2011) Peri-
implant bone tissue assessment by comparing the outcome of intra-oral radiograph
and cone beam computed tomography analyses to the histological standard. Clinical
Oral Implants Research 22: 492-499.
4. Diederichs, G., Link, T., Marie, K., Huber, M., Rogalla, P., Burghardt, A., Majumdar,
S. & Issever, A. (2008) Feasibility of measuring trabecular bone structure of the
proximal femur using 64-slice multidetector computed tomography in a clinical
setting. Calcified Tissue International 83: 332–341.
5. Fanuscu, M.I. & Chang, T.L. (2004) Three-dimensional morphometric analysis of
human cadaver bone: microstructural data from maxilla and mandible. Clinical Oral
Implants Research 15: 213-218.
6. Felsenberg, D. & Boonen, S. (2005) The bone quality framework: determinants of
bone strength and their interrelationships, and implications for osteoporosis
management. Clinical Therapeutics 27: 1–11.
7. Genant, H.K., Gordon, C., Jiang, Y., Link, T.M., Hans, D., Majumdar, S. & Lang,
T.F. (2000) Advanced imaging of the macrostructure and microstructure of bone.
Hormone Research 54 (suppl 1): 24-30.
8. Gomes de Oliveira, R.C., Leles, C.R., Lindh, C. & Ribeiro-Rotta, R.F. (2012) Bone
tissue microarchitectural characteristics at dental implant sites. Part 1: identification
of clinical-related parameters. Clinical Oral Implants Research 23: 981–986.
9. Gonzalez-Garcia, R. & Monje, F. (2012) The reliability of cone-beam computed
tomography to assess bone density at dental implant recipient sites: a
histomorphometric analysis by micro-CT. Clinical Oral Implants Research doi:
10.1111/j.1600-0501.2011.02390.x.
10. Hans, D., Barthe, N., Boutroy, S., Pothuaud, L., Winzenrieth, R. & Krieg, M.A.
(2011) Correlations between trabecular bone score, measured using anteroposterior
-
38
dual-energy X-ray absorptiometry acquisition, and 3-dimensional parameters of bone
microarchitecture: an experimental study on human cadaver vertebrae. Journal of
Clinical Densitometry 14: 302–312.
11. Hatcher, D.C. (2010) Operational principles for cone-beam computed tomography.
Journal of American Dental Association 141(suppl 3): 3–6.
12. Hildebrand, T., Laib, A., Müller, R., Dequeker, J. & Ruegsegger, P. (1999) Direct
three-dimensional morphometric analysis of human cancellous bone: microstructural
data from spine, femur, iliac crest, and calcaneus. Journal of Bone Mineral Research
14: 1167–1174.
13. Ibrahim, N., Parsa, A., Hassan, B., van der Stelt, P. & Wismeijer, D. (2012)
Diagnostic imaging of trabecular bone microstructure for oral implants: a literature
review. Dentomaxillofacial Radiology 42:20120075.
14. Issever, A.S., Link, T.M., Kentenich, M., Rogalla, P., Burghardt, A.J., Kazakia, G.J.,
Majumdar, S. & Diederichs, G. (2010) Assessment of trabecular bone structure using
MDCT: comparison of 64- and 320-slice CT using HR-pQCT as the reference
standard. European Radiology 20: 458-468.
15. Li, G., Xie, H., Ning, H., Citrin, D., Capala, J., Maass-Moreno, R., Guion, P., Arora,
B., Coleman, C., Camphausen, K. & Miller, R. (2008) Accuracy of 3D volumetric
image registration based on CT, MR and PET/CT phantom experiments. Journal of
Applied Clinical Medical Physics 9: 17-36.
16. Liu, S.M., Zhang, Z.Y., Li, J.P., Liu, D.G. & Ma, X.C. (2007) A study of trabecular
bone structure in the mandibular condyle of healthy young people by cone beam
computed tomography. Zhonghua Kou Qiang Yi Xue Za Zhi 42: 357-360.
17. Manoglas, S.C. & Jilka, R.L. (1995) Bone marrow, cytokines, and bone modelling.
Emerging insights into the pathophysiology of osteoporosis. New England Journal of
Medicine 332: 305-311.
18. Manske, S.L., Macdonald, H.M., Nishiyama, K.K., Boyd, S.K. & McKay, H.A.
(2010) Clinical tools to evaluate bone strength. Clinical Reviews in Bone and Mineral
Metabolism 8: 122–134.
19. Minkin, C. & Marinho, V.C. (1999) Role of the osteoclast at the bone-implant
interface. Advances in Dental Research 13: 49–56.
20. Parkinson, I.H., Badiei, A. & Fazzalari, N.L. (2008) Variation in segmentation of
bone from micro-CT imaging: implications for quantitative morphometric analysis.
Australasian Physical & Engineering Sciences in Medicine 31: 160-164.
-
39
21. Parsa, A., Ibrahim, N., Hassan, B., Motroni, A., van der Stelt, P. & Wismeijer, D.
(2013) Influence of cone beam CT scanning parameters on grey value measurements
at implant site. Dentomaxillofacial Radiology 42:79884780.
22. Patcas, R., Markic, G., Müller, L., Ullrich, O., Peltomäki, T., Kellenberger, C.J &
Karlo C.A. (2012) Accuracy of linear intraoral measurements using cone beam CT
and multidetector CT: A tale of two CTs. Dentomaxillofacial Radiology 41: 637–644.
23. Quereshy, F.A., Savell, T.A. & Palomo, J.M. (2008) Applications of cone beam
computed tomography in the practice of oral and maxillofacial surgery. Journal of
Oral Maxillofacial Surgery 66: 791-796.
24. Ranjanomennahary, P., Ghalila, S.S., Malouche, D., Marchadier, A., Rachidi, M.,
Benhamou, C. & Chappard, C. (2011) Comparison of radiograph-based texture
analysis and bone mineral density with three-dimensional microarchitecture of
trabecular bone. Medical Physics 38: 420–428.
25. Riggs, B.L., Hodgson, S.F., O’Fallon, W.M., Chao, E.Y., Wahner, H.W., Muhs, J.M.,
Cedel, S.L. & Melton, J.M. (1990) Effect of fluoride treatment on the fracture rate in
postmenopausal women with osteoporosis. New England Journal of Medicine 332:
802–809.
26. Sakka, S. & Coulthard, P. (2009) Bone quality: a reality for the process of
osseointegration. Implant Dentistry 18: 480-485.
27. Schulze, R., Heil, U., Gross, D., Bruellmann, D.D., Dranischnikow, E., Schwanecke,
U. & Schoemer, E. (2011) Artefacts in CBCT: a review. Dentomaxillofacial
Radiology 40: 265-273.
28. Tahmaseb, A., De Clerck, R., Eckert, S. & Wismeijer, D. (2011) Reference-based
digital concept to restore partially edentulous patients following an immediate loading
protocol: a pilot study. International Journal of Oral Maxillofacial Implants 26: 707-
717.
29. Tjong, W., Kazakia, G.J., Burghardt, A.J. & Majumdar, S. (2012) The effect of voxel
size on high-resolution peripheral computed tomography measurements of trabecular
and cortical bone microstructure. Medical Physics 39: 1893-1903.
30. Vigorita, V.J. (1984) The bone biopsy protocol for evaluating osteoporosis and
osteomalacia. The American Journal of Surgical Pathology 8: 925-930.
31. Wirth, A.J., Müller, R., van Lenthe, G.H. (2012) The discrete nature of trabecular
bone microarchitecture affects implant stability. Journal of Biomechanics 45: 1060-
1067.
-
40
-
41
Chapter 3
Bone quality evaluation at dental
implant sites using Multi-slice CT,
Micro-CT and CBCT
This chapter has been published as:
Parsa A, Ibrahim N, Hassan B, van der Stelt P, Wismeijer D. Bone quality evaluation
at dental implant site using Multi-slice CT, Micro-CT and Cone Beam CT. Clinical
Oral Implants Research. doi: 10.1111/clr.12315.
-
42
Bone quality evaluation at dental implant site using
Multi-slice CT, Micro-CT and Cone Beam CT
Azin Parsa, Norliza Ibrahim, Bassam Hassan, Paul van der Stelt,
Daniel Wismeijer.
Clinical Oral Implants Research. doi: 10.1111/clr.12315.
Summary
Objectives: The first purpose of this study was to analyze the correlation between
bone volume fraction (BV/TV) and calibrated radiographic bone density (HU) in
human jaws, derived from micro-CT and MSCT respectively. The second aim was to
assess the accuracy of CBCT in evaluating trabecular bone density and microstructure
using MSCT and micro-CT, respectively, as reference gold standards.
Material and methods: Twenty partially edentulous human mandibular cadavers
were scanned by three types of CT modalities: MSCT (Philips, Best, the Netherlands),
CBCT (3D Accuitomo 170, J.Morita, Kyoto, Japan), and micro-CT (SkyScan 1173,
Kontich, Belgium). Image analysis was performed using Amira (v4.1, Visage Imaging
Inc., Carlsbad, CA), 3Diagnosis (v5.3.1, 3diemme, Italy), Geomagic (studio® 2012,
Morrisville, NC), and CTAn (v1.11, SkyScan, Kontich, Belgium). MSCT, CBCT, and
micro-CT scans of each mandible were matched to select the exact region of interest
(ROI). MSCT HU, micro-CT BV/TV, and CBCT grey value and bone volume
fraction of each ROI were derived. Statistical analysis was performed to assess the
correlations between corresponding measurement parameters.
Results: Strong correlations were observed between CBCT & MSCT density (r=0.89)
and between CBCT and micro-CT BV/TV measurements (r=0.82). Excellent
correlation was observed between MSCT HU and micro-CT BV/TV (r=0.91).
However significant differences were found between all comparisons pairs (p
-
43
Conclusions: An excellent correlation exists between bone volume fraction and bone
density as assessed on micro-CT and MSCT, respectively. This suggests that bone
density measurements could be used to estimate bone microstructural parameters. A
strong correlation was also found between CBCT grey values and BV/TV and their
gold standards, suggesting the potential of this modality in bone quality assessment at
implant site.
-
44
Introduction
Primary implant stability is the key factor for the long term success of an implant treatment
by improving osseointegration (Fuh et al. 2010). Primary instability of an implant induces
movements during healing. This micro-motion leads to fibroplasia as a biological response at
bone tissue surrounding the implant. The replacement of bone by fibrous tissue and loss of
osseointegration cause implant failure (Lioubavina-Hack et al 2006). Bone quality, which
refers to the combination of all bone characteristics that influence bone resistance to fracture
(Fyhrie 2005), is one of the most important factors influencing primary implant stability
(Ozan et al. 2007; Tolstunov 2007). Among the bone characteristics, bone mineral density
(BMD) and trabecular microstructure are the strongest predictors for bone strength (Muller
2003). However, these two parameters need to be simultaneously assessed to provide better
estimation of bone strength (Teo et al. 2007; Diederichs et al. 2009).
Several radiographic modalities have been used for bone quality assessment. For bone
microstructure, micro-computed tomography (micro-CT) was recommended as gold standard
for assessing bone morphology and micro architecture (Burghardt et al. 2011; Ibrahim et al.
2013a). However, it is limited to ex-vivo small bone samples and cannot be employed for
patients. Multiple X-ray projections with different angles in micro-CT allow a precise three-
dimensional (3D) reconstruction of the bone samples and assessment of bone trabeculae
(Martin-Badosa et al. 2003). Micro-CT is used to measure several histomorphometric
variables including bone volume (BV), total volume (TV), bone volume fraction (BV/TV),
trabecular thickness (Tb.Th), trabecular number (Tb.N) and trabecular separation (Tb.Sp)
(Odgaard 1997).
For bone density, multislice computed tomography (MSCT) is an established clinical
modality in which calibrated Hounsfield units (HU) can accurately be converted to BMD
measurements (Shahlaie et al. 2003; Shapurian et al. 2006). However, higher radiation
exposure risk to patients in comparison with other modalities remains a main concern for
applying MSCT for assessing bone quality (Dula et al. 1996; Ekestubbe et al. 1992;
Ekestubbe et al. 1993; Frederiksen et al. 1995). Cone Beam Computed Tomography (CBCT),
due to increased accessibility to dental practitioners, more compact equipment and reduced
cost and radiation dose, has widely replaced medical CT for oral and maxillofacial imaging.
-
45
Several studies reported high geometric accuracy of CBCT for linear measurement
(Lagravère et al. 2008; Lou et al. 2007; Naitoh et al. 2004), while its reliability in bone
quality evaluation remains controversial. Only few studies suggested that CBCT could be
applied to assess trabecular bone microstructure (Corpas Ldos et al. 2011; Liu et al. 2007).
Additionally, CBCT does not represent calibrated voxel grey values expressed in HU (Hua et
al. 2009).
Yet, many attempts have been conducted to assess the feasibility of converting
CBCT grey values to actual density measurements. High correlation between HU derived
from MSCT and CBCT voxel grey values has been demonstrated , hinting at the potential of
CBCT in bone density assessment (Aranyarachkul et al. 2005; Cassetta et al. 2013; Lagravère
et al. 2006; Naitoh et al. 2009; Naitoh et al. 2010b; Nomura et al. 2010; Parsa et al. 2012;
Reeves et al. 2012). However, the excessive scattering and technology-specific artifacts
produced in CBCT have been denoted as the perpetrator for the unreliable BMD
measurements (Araki et al. 2011; Hua et al. 2009; Nackaerts et al. 2011; Schulze et al. 2011;
Yoo & Yin 2006).
High correlation between bone volume fraction (BV/TV) provided by micro-CT and voxel
grey value from CBCT, and also between bone volume fraction derived from CBCT and CT
numbers from MSCT has been reported (González-García & Monje 2012; Naitoh et al.
2010a). However the relation between bone volume fraction and radiographic bone density in
human jaws remains controversial (Aksoy et al. 2009; Stoppie et al. 2006). Therefore, the
first purpose of this study was to analyze the correlation between bone volume fraction
(BV/TV) and calibrated radiographic bone density (HU) in human jaws, derived from micro-
CT and MSCT respectively. The second aim was to assess the accuracy of CBCT in
evaluating trabecular bone density and microstructure using MSCT and micro-CT,
respectively, as reference gold standards.
Material and methods
Sample preparation and radiographic evaluation
Twenty partially edentulous human mandibular cadavers not identified by age, sex or ethnic
group were obtained from the functional anatomy department. The cadavers were sectioned
at the mid-ramus level and fixed in formaldehyde (formaldehyde 74.79%, Glycerol 16.7%,
Alcohol8.3 %, and Phenol 0.21%) and stored. A declaration was obtained from the
Functional Anatomy department to use this human remains material for research purposes.
-
46
The restorative materials which can produce artifact such as amalgam filling and metal
crowns were removed from dentitions. The mandibles were scanned by three types of CT
modalities: MSCT (Philips, 120 kVp, 222 mA, 1.128 S, 0.67mm isotropic voxel size, Best,
the Netherlands), CBCT (3D Accuitomo 170, 90 kVp, 5 mA, 30.8 S, 4x4 cm FOV, 0.08 mm
isotropic voxel size, J.Morita, Kyoto, Japan), and micro-CT (SkyScan 1173,130 kVp, 61 mA,
35 min, 35μm isotropic voxel size, Kontich, Belgium). In MSCT scans, the occlusal plane of
each mandible was set perpendicular to the floor with zero gantry tilt, whereas in CBCT
scans it was set parallel to the floor according to manufacturer’s recommended protocol. The
edentulous region of each mandible was located at the center of FOV in CBCT scans. Owing
to the large gantry of applied micro-CT (140mm in diameter, 200mm in height), mandibles
were not sectioned to smaller samples. To prevent the possible micro movements during the
scanning due to the large size of the samples, a cylindrical shape Styrofoam was used to fix
and mount the sample into the holder.
Image processing
All CT data sets were converted to Digital Imaging and Communication in Medicine
(DICOM3) format. As the scan orientation differs between micro-CT and the other two
systems, the z-axis of CBCT and MSCT images were flipped to match that of micro-CT for
further procedures. Micro-CT data sets were large in size, therefore was not possible to be
flipped by our workstation. Image analysis was performed using Amira (v4.1, Visage
Imaging Inc., Carlsbad, CA), 3Diagnosis (v5.3.1, 3diemme, Italy), Geomagic (studio® 2012,
Morrisville, NC), and CTAn (v1.11, SkyScan, Kontich, Belgium). MSCT images were
imported to 3Diagnosis software. Two cylindrical shape virtual probes (with diameter and
height of 0.7 and 8 mm, respectively) were inserted at the edentulous region within the
cancellous bone, with 3 mm bucco-lingual distance between them (Figure1a, b). These
probes were used as indicators to facilitate the selection of exact region of interest (ROI)
from MSCT, CBCT and micro-CT. For MSCT, the probes were visible in a single cross-
sectional slice since the voxel size of MSCT scans was 0.67 mm, which is thick enough to
allow the probes to be visible in one slice. Subsequently, a rectangular area was drawn
between the two probes from the slice of interest to define the ROI for density measurements.
-
47
Figure 1. (a) Three-dimensional reconstruction of a mandible MSCT scan with inserted
probes. (b) close-up view of probes (C) three-dimensional reconstruction of a mandible
CBCT scan with transferred probes.
The mean HU value from each ROI was calculated. All ROIs were totally within the
cancellous bone, excluding cortical bone, inferior dental canal and any large bone defect.
For CBCT, a volume-based 3D registration algorithm using Geomagic software was applied
to transform the inserted probes from the MSCT datasets to the CBCT scans. A standard
triangulation language (STL) surface file of the MSCT and CBCT scans were matched and
the probes were transferred from MSCT scans to the exact region on CBCT’s (Figure1c). As
a result, new CBCT datasets which include the probes were obtained. Using 3Diagnosis eight
consecutive slices passing through the probes were selected from each CBCT dataset to
calculate the mean grey values (radiological density). This is because slice thickness in
CBCT is 0.08mm which approximately amounts to 8 times thinner than the equivalent slice
thickness in MSCT. A rectangular region was also drawn between the two probes similar to
MSCT and grey values from corresponding anatomical locations were derived.
CBCT radiological density of each mandible’s ROI was considered as the mean of eight
calculated grey values. Subsequently, the selected ROIs were saved as a bitmap (BMP) image
files to allow the trabecular micro-structure evaluation.
Using Amira, each micro-CT scan was cropped to have a smaller sample including the ROI.
Due to large micro-CT data sets, the superimposition of CBCT and micro-CT scans was done
as follows: maximum alignment of both datasets was obtained by manually matching and
superimposition of isosurfaces generated in Amira software. Subsequently, sixteen micro-CT
slices (correspondence to the eight CBCT slices) were selected and saved as a 16-bitmap
(BMP) image file (65536 gray values). Then, these bmp files were exported to CTAn
software for trabecular microstructure evaluations (Figure 2a). A rectangular ROI for
-
48
trabecular was selected on each dataset slice by slice (Figure 2b). All images were
thresholded using an automated histogram analysis and binarized (Figure 2c) to allow the
measurement process. On micro-CT datasets, the ROI was again verified by carefully
comparing slices with CBCT’s (as reference). This was performed to reduce bias which may
have been introduced during the manual superimposition of the two datasets. All
measurements were performed twice with one month interval by a trained maxillofacial
radiologist.
Figure 2. (a) Images of micro-CT and CBCT were compared slice by slice from the same
anatomical region. (b) A rectangular region of interest (ROI) was selected for each dataset.
(c) Images were binarized and (d) processed to allow the microstructural measurements.
-
49
Data analysis
Statistical analysis was performed using SPSS (v17.0, SPSS Inc., Chicago, IL). To determine
the intraobserver reliability of the radiological and microstructural density measurement,
intraclass correlation coefficient (ICC) was used. The Shapiro-Wilk test was used to verify
the normality of the data. Paired t-test was used to assess the mean difference between MSCT
and CBCT density measurements and between CBCT BV/TV and micro-CT BV/TV, while
Pearson correlation coefficient was used to assess the linear relation between corresponding
measurement parameters. Finally a Bland-Altman plot was used to assess the accuracy of
CBCT in measuring trabecular BMD and bone microstructural density by plotting the
difference between the measurements of CBCT against MSCT density and microCT BV/TV
against the means of the compared measurements.
Results
Excellent intraobserver reliability (ICC ≥ 0.97) was revealed for repeated measurements in
the three systems. Therefore, the mean of two measurements was calculated for further
analysis. The mean HU of the selected ROI ranged from -60 to 507.6 (mean 222.85 &
standard deviation [SD] 140.5) while CBCT grey values ranged from161.6 to 665.6 (mean
377.49 & SD 127.4). The negative HU derived from MSCT for case 4, 16 and 20 (Table 1)
may indicate fat in trabecular spaces (Parsa et al. 2012). Calculated BV/TV of the same ROI
ranged from 2.24 to 75.83 (mean 32.35 & SD 18.81) for micro-CT and from 3.73 to 68.72
(mean 36.79 & SD 23.17) for CBCT (Table1). Paired t-test showed significant differences (p
< 0.001) between all comparison pairs except for mean measurement between CBCT BV/TV
and micro CT BV/TV (p=0.147) . In all selected ROIs, CBCT showed a higher density than
MSCT HU and a higher BV/TV than that of micro-CT. The normal distribution of
measurements was confirmed by visually inspecting the histogram and the result of the
Shapiro-Wilk test (p > 0.05). Therefore, the use of the t-test and Bland-Altman test is
justified. Strong correlations were observed between CBCT and MSCT density
measurements (r=0.89) and between CBCT and micro-CT BV/TV measurements (r=0.82).
Excellent correlation was observed between MSCT HU and micro-CT BV/TV (r=0.91).
Bland-Altman analysis showed the bias in measuring BV/TV between CBCT and micro-CT
is smaller (4.44µm-1
) than measuring the density between CBCT and MSCT (154.65HU)
(Figure 3a, b). The 95% measurement errors are between -21.31 to 30.19 for BV/TV and
-
50
29.74 to 279.56 for density measurement. The differences of CBCT and micro-CT BV/TV
measurements were minimal (4.44µm-1
), suggesting strong agreement.
Table1. Mean results of MSCT, Micro-CT and CBCT density (grey value) and bone volume
fraction (BV/TV) measurements.
Mandible
No.
MSCT HU CBCT grey value Micro-CT
BV/TV (%)
CBCT BV/TV
(%)
1 390,9 513,3 37,11 44,43
2 205,1 444,6 41,40 46,17
3 507,6 665,6 75,83 63,58
4 -16,6 204,7 3,91 3,73
5 136 324,4 17,44 9,98
6 202,4 270,8 23,99 24,87
7 245,2 368,5 31,68 24,68
8 327,1 417,5 47,62 50,41
9 341 442,9 49,34 37,94
10 316 474,3 48,68 62,94
11 270 379,3 50,86 48,58
12 210,2 278 44,64 68,72
13 320 559,1 44,12 63,28
14 204,4 344,9 22,65 26,31
15 285,7 357 34,57 66,29
16 -27,6 161,6 2,24 4,33
17 220 483,7 24,87 55,30
18 240,4 376,7 26,97 23,21
19 139,2 285,3 15,97 6,68
20 -60 197,7 3,16 4,45
-
51
Figure 3. Bland-Altman plots of (a) BV/TV measurements between CBCT and µCT, and (b)
density measurements between CBCT and MSCT.
Discussion
It has been proven that the success of an inserted implant strongly depends on the quality,
beside the quantity, of the surrounded bone (Jaffin & Berman 1991; Jemt et al. 1992). In
jawbones, density measurements derived from MSCT HU are highly reliable (Schwarz et al.
1987; Shapurian et al. 2006). However, bone density alone does not fully represent bone
quality, and should be considered together with bone microarchitecture to estimate bone
strength and fracture resistance (Diederichs et al. 2009). Histomorphometrically, bone
volume fraction, which is the trabecular bone volume (BV) per tissue volume (TV) expressed
in %, is the most important parameter (Parfitt et al. 1987). Micro-CT is accepted as a gold
standard modality for trabecular microstructure assessment, but it cannot be employed in the
clinic (Burghardt et al. 2011). In this study our aim was to investigate the possible correlation
between bone quality measurements of clinically applicable scanners in comparison with
micro-CT.
A study on porcine vertebral cancellous bone revealed a high correlation between HU derived
from CT images and BV/TV from micro-CT and suggested the use of HU from medical CT
for the prediction of microarchitecture (Teo et al. 2006). Our results support these findings
that correlation between MSCT HU and Micro-CT BV/TV is high (r=0.91). However, the
mean of calculated BV/TV in mentioned study deviated from our findings in human
-
52
mandibles. This could be due to different samples, ROI selections and different scanner
systems. In similar studies using human zygomatic and jawbones, a high correlation was
found only in female subjects (Aksoy et al. 2009; Nkenke et al. 2003). Thus, they suggested
that only female trabecular BV/TV can be predicted from bone mineral density. In contrast,
another study found a high correlation between BV/TV and HU in trabecular bone
surrounded by a thin layer of cortical bone regardless to gender (Stoppie et al. 2006). This
study suggested that with the development of MSCT scanners and imaging software, more
precise HU measurement would be achievable (Stoppie et al. 2006). Our results showed a
strong correlation between BV/TV and HU in human mandibular trabecular bone, regardless
to gender and thickness of surrounding cortical bone (r=0.91). This confirms the possibility
of prediction of bone volume fraction from MSCT bone density measurement. The usefulness
of this prediction can be emphasized by the limitation of micro-CT in clinical settings.
CBCT has several advantages over MSCT in terms of more compact equipment, small
footprint for the clinic, and relatively reduced scan costs. Additionally, lower radiation dose
levels to the main organs of the head and neck region have been cited as one of the most
important advantages of CBCT over MSCT (Carrafiello et al. 2010; Kau et al. 2005; White
2008). Due to these advantages, the use of this modality in dental implant planning is
growing so fast and it is more accessible to the dental practitioners than before. Therefore, the
validity of CBCT in bone quality assessment has been studied broadly. The majority of these
studies have focused on the bone density measurement and found CBCT a reliable modality
for bone density measurement (Aranyarachkul et al. 2005; Cassetta et al. 2013; Lagravère et
al. 2006; Naitoh et al. 2009; Naitoh et al. 2010b; Nomura et al. 2010; Parsa et al. 2012;
Reeves et al. 2012). The high correlation between measured CBCT grey values and CT
numbers in our study (r=0.89) may confirms the possible potential of CBCT in radiographic
density measurement. However, the limit of agreement in Bland and Altman plot (Figure 3b)
is huge (29.74 to 279.56) with a high bias value (mean = 154.64). This indicates
an unfavorable strength of agreement. Thus, although the measurements is reliable (ICC
>0.97) and validated between two compared systems (r = 0.82), the density measurement
using CBCT is less accurate when compared to its gold standard system (MSCT). It should
be considered that CBCT density measurement can be effected by scanning parameters and
the location of the ROI within the scanner (Nackaerts et al. 2011; Parsa et al. 2013).
-
53
Using micro-CT as gold standard, the reliability of CBCT in trabecular microstructure
assessment has been validated in human mandibles, but BV/TV was not among the assessed
microstructural parameters (Ibrahim et al. 2013b). Our results also confirm the reliability of
CBCT in trabecular microstructure assessment, based on a high correlation between BV/TV
measured by CBCT and micro-CT (r=0.82). The positive bias value (4.44µm-1
) in the Bland
and Altman plot (Figure 3a) indicates that BV/TV was measured higher by CBCT. The small
range between the confidence interval for the measurement differences between the two
systems was small (-21.31 to 30.19) indicates a strong agreement between CBCT and micro-
CT in measuring BV/TV. In present study, smallest available FOV (40x40 mm) and high
resolution scan mode were applied in CBCT scans in order to achieve the highest possible
spatial resolution (0.08 mm isotropic voxel size). Therefore using different CBCT scanning
parameters the results may differ.
It should be emphasized that the CBCT bone quality measurements in our study deviated
from those of gold standards. This deviation arises from increased scattering, noise level and
artefacts specific to the scanner technology which operates at lower peak kilovoltage and tube
loading setting than MSCT and micro-CT, resulting in a reduced signal-to-noise ratio
(Schulze et al. 2011). A higher noise level in comparison to MSCT can cause more
inconsistencies in voxel grey values (Araki & Okano 2011; Aranyarachkul et al. 2005).
Additionally, as the acquired volume in CBCT is larger than collimated fan beam in MSCT,
the influence of these artifacts is excessively exacerbated (Nackaerts et al. 2011; Schulze et
al. 2011).
Unlike the majority of other studies on bone volume fraction, our bone samples were not
harvested for micro-CT scans. As such, in our sample the possible deviation between the
planned and excised ROI, which might arise during the trepanation procedure, was eliminated
(Stoppie et al. 2006). Additionally, in the present study, a fully automated and observer
independent 3D matching algorithm was employed for MSCT and CBCT scans registration
to ensure that all measurements are exactly from the same site up to voxel accuracy.
However, due to the manual alignment of CBCT and micro-CT datasets, there is a possibility
for observer error and selection of not identical regions. Since micro-CT datasets are large
and therefore computationally expensive, technical limitations prohibited applying the 3D
registration algorithm for automated alignment. Technical advancements in the future might
resolve this issue. Finally, the difference in voxel size of CBCT (0.080 mm), micro-CT
-
54
(0.035mm) and MSCT (0.67 mm) can also contribute to the observed discrepancy in
calculating BV/TV and bone density.
Voxel size in CBCT influences image quality among other factors including the unit itself,
tube voltage and FOV selection (Kamburoglu et al. 2011). Generally, the smaller the voxels
the higher the spatial resolution and therefore the sharper the images appear to be. However,
small voxels result in decreased contrast to noise ratio levels and they require higher exposure
dose to the patient (Davies et al. 2012). The higher the spatial resolution the more technical
demands are imposed on the imaging system as a whole and on the imaging detector in
specific to attempt to suppress noise and increase signal levels. CBCT suffers from increased
noise levels especially at smaller voxel sizes due to low tube voltage, cone beam divergence
phenomena and inferior detector efficiency when compared to MSCT and micro-CT (Hassan
et al. 2010). However, the potential influence of varying voxel size on visibility of hard tissue
structures such as bone remains largely unknown. A recent systematic review of the literature
concluded that there is a systematic lack of evidence regarding the impact of varying voxel
size in CBCT on diagnostic performance and that possibly different voxel sizes might lead to
comparable diagnostic outcomes (Spin-Neto et al. 2013). Only one study could be identified
which demonstrated a possible effect of varying voxel size on cancellous bone measurements
in micro-CT (Yeni et al. 2005). However, it remains unknown whether the same applies to
CBCT.
In this study, a conscious effort was made to optimize image quality through selecting the
scan protocols and voxels sizes as recommended by the manufacturer for the chosen FOV’s.
Our results are limited to one CBCT system (Accuitomo 170) and results may vary on other
systems. The design specifications of different systems still vary (De Vos et al. 2009). The
lack of a technical standard for the development of CBCT systems has led to a wide disparity
in the physical parameters of each model. Developing such a standard for manufacturing
CBCT systems may help in generalizing research findings in the future. The study was also
limited as surrounding anatomical structures including the tongue and vertebra were absent.
As a result, in CBCT scans partial object artefacts resulting from structures placed outside the
scan field were not simulated. It has been previously noted that artefacts resulting from
partial sampling of objects outside the scan field could result in