articulators

181
1 RATIONALE OF USE OF DENTAL ARTICULATORS IN COMPLETE DENTURE TREATMENT PRESENTED BY : DR. SAKSHI SHARMA M.D.S. 1 ST YEAR

Upload: sakshi-sharma

Post on 30-Oct-2014

382 views

Category:

Documents


11 download

TRANSCRIPT

Page 1: ARTICULATORs

1

RATIONALE OF USE OF DENTAL ARTICULATORS IN COMPLETE DENTURE TREATMENT

PRESENTED BY :DR. SAKSHI SHARMAM.D.S. 1ST YEAR

Page 2: ARTICULATORs

2

CONTENTS Introduction Definition of an articulator Historical review Advantages of an articulator Classification systems Requirements of articulators Parts of an articulator Parts of Hanau articulator and its functions Types of Hanau articulators Whip Mix articulator Dentatus articulator Summary Bibliography.

Page 3: ARTICULATORs

3

INTRODUCTION

The primary function of the articulator is to

act as the patient in the absence of the

patient. An articulator is used to simulate

the patient’s temporomandibular joints,

mandible and maxilla, and the complex

neuromuscular mechanism that programs

mandibular movements.

Page 4: ARTICULATORs

4

Other purposes for which an articulator is used

are:

1. Mounting of dental casts for diagnosis, treatment

planning, and patient presentation.

2. Fabrication of occlusal surfaces for dental

restorations.

3. Arrangement of artificial teeth for complete and

removable partial dentures.

Page 5: ARTICULATORs

5

In 1858 Bonwill developed the first articulator with a serious effort to imitate the movements of the mandible in eccentric positions.

Page 6: ARTICULATORs

6

DEFINITION

An articulator may be defined as “a

mechanical instrument that represents

the temporomandibular joints and jaws, to

which maxillary and mandibular casts

may be attached to simulate some or all

mandibular movements”.

- (GPT-8).

Page 7: ARTICULATORs

7

HISTORICAL REVIEW The plaster articulator was first described by PHILLIP

PFAFF in 1756

Also known as SLAB ARTICULATORS

GARIOT designed the hinge joint articulator about 1805.

Page 8: ARTICULATORs

8

In 1899, Richmond S. Hayes received a patent for an articulator that was the first to incorporate a fixed descending condylar path.

In the same year, Hayes introduced the first example of a functional face-bow like device .He named this device as “articulating caliper”.

Page 9: ARTICULATORs

9

The New Century articulator was manufactured by the Snow Dental Co., Buffalo, New York, In 1906.

Page 10: ARTICULATORs

10

The Acme articulator, which also was made by George B. Snow.

Page 11: ARTICULATORs

11

The Gysi Adaptable articulator was introduced in 1910.

Gysi Simplex was introduced in 1914 as a mean value articulator.

Page 12: ARTICULATORs

12

The maxillomandibular instrument designed in 1918 by George Monsoon, was based on the spherical theory.

Page 13: ARTICULATORs

13

The Stephan articulator, developed in 1921.

Page 14: ARTICULATORs

14

The Hagman Balancer, developed in the 1920s.

Opens and closes on a hinge

Requires no face-bow or interocclusal records for mounting.

Page 15: ARTICULATORs

15

Rudolph L. Hanau,In 1921 developed a research model called Hanau Model C Articulator.

In 1923 he developed another research instrument, the Hanau Model M Kinoscope articulator.

Page 16: ARTICULATORs

16

In the Wardsworth articulator, developed in 1924.

Page 17: ARTICULATORs

17

The Hanau Model H110 manufactured in 1922 and 1923.

It has individual condylar guidance adjustments.

Page 18: ARTICULATORs

18

The Hanau Model H110 Modified was introduced in 1927.

Page 19: ARTICULATORs

19

In 1926, GYSI TRUBYTE ARTICULATOR was introduced. It is a non arcon instrument with a fixed intercondylar distance.

Page 20: ARTICULATORs

20

The Stansbery articulator was developed by C. J. Stansbery Of Los Angeles in 1929.

Page 21: ARTICULATORs

21

The HOUSE articulator was developed by M. M. HOUSE in the early 1930s

Page 22: ARTICULATORs

22

The PRECISION COORDINATOR was developed by W. H. TERRELL, in the early 1930s.

It is an ARCON type of instrument that has curvilinear condylar guides.

Page 23: ARTICULATORs

23

In 1944, DENTATUS was designed in SWEDEN.

This articulator is unique in that the relationship between the upper and lower members can be standardized with a gauge block, so that casts can be transferred from one articulator to another and still maintain the same relation.

Page 24: ARTICULATORs

24

BERGSTORM designed an instrument in 1950 called the ARCON, which is similar to HANAU H, except that the condyles are on the lower member of the instrument, and the condylar guides are curved and on the upper member.

Page 25: ARTICULATORs

25

The TRANSOGRAPH presented in 1952, is a split axis instrument designed to allow each condylar axis to function independently of the other .

Page 26: ARTICULATORs

26

The GNATHOSCOPE, designed by CHARLES STUART in 1955.

Page 27: ARTICULATORs

27

The NEY articulator was designed by DE PIETRO in 1960 and was an arcon instrument.

Page 28: ARTICULATORs

28

The HANAU 130 – 21 was designed by RICHARD BEU and JAMES JANIK IN 1964.

It has a split horizontal axis that can be adjusted vertically and horizontally, adjustable intercondylar distance, adjustable bennet guides, and adjustable horizontal condylar path guides.

It is utilised with protrusive and lateral interocclusal records.

Page 29: ARTICULATORs

29

The DUPLI-FUNCTIONAL articulator was designed by IRISH and presented in 1965.

It was primarily for use in complete denture construction.

It records each patient’s mandibular movements

It serves as a three dimensional tripod type of dental articulator on which denture may be constructed and their occlusion balanced.

Page 30: ARTICULATORs

30

In 1968, NILES GUICHET designed the DENAR fully adjustable articulator (D4A).

Page 31: ARTICULATORs

31

The TELEDYNE articulator was designed by RICHARD BEU IN 1975.

Page 32: ARTICULATORs

32

The HANAU ARCON H2 was introduced by HANAU in 1977.

It is quite similar to HANAU H2, except that it is an ARCON instrument.

The next articulator to be introduced in HANAU series was HANAU RADIAL SHIFT ARTICULATOR

The HANAU WIDE VUE and WIDE VUE II are the newest to be introduced by HANAU.

Page 33: ARTICULATORs

33

ADVANTAGES OF USING AN ARTICULATOR

Properly mounted casts allow the operator to better visualize the patients occlusion, especially from the lingual view. This is essential if a proper occlusal scheme is to be developed for complete dentures.

Patient cooperation is not a factor when using an articulator once the appropriate interocclusal records have been obtained.

Page 34: ARTICULATORs

34

Considerable more chair time and patient appointment is required when utilizing the mouth as an articulator.

More procedures can be delegated to auxiliary personnel when utilizing an articulator for development of the patients occlusion.

The patient’s saliva, tongue and cheeks are not factors when using an articulator.

Page 35: ARTICULATORs

35

MINIMAL REQUIREMENTS OF ARTICULATORS

It should hold casts in the correct horizontal and vertical relationships.

It should provide a positive anterior vertical stop.

It should accept a face- bow transfer record.

It should open and close in hinge movement.

It should allow protrusive and lateral motion.

The moving parts should move freely and be accurately machined.

The non moving parts should be a rigid construction.

Page 36: ARTICULATORs

36

ADDITIONAL REQUIREMENTS

Adjustable horizontal and lateral condylar

guide elements.

The condylar elements as a part of the lower

frame and the condylar guides as a part of the

upper frame.

A mechanism to accept a third reference point

from a face-bow transfer record.

A terminal hinge position locking device.

Immovable mounting plates that can be

repositioned accurately.

An adjustable incisal guide table.

Adjustable intercondylar width of the condylar

elements when graphic tracings to be used to

set and select condylar guidances.

Page 37: ARTICULATORs

37

LIMITATIONS OF AN ARTICULATOR

An articulator can simulate but not duplicate jaw movements.

An articulator made of metal may show error due to tooling or resulting

from metal fatigue and wear.

The articulator may not exactly simulate the intraborder and functional

movements of the mandible.

Page 38: ARTICULATORs

38

Using the jaws as an articulator also has limitations:

Inability of humans to detect visually the finer

changes in the motion.

Making accurate marks in the presence of

saliva.

Exact location of the condyles.

The resiliency of the supporting structures.

The dentures are movable.

Page 39: ARTICULATORs

39

CLASSIFICATION OF ARTICULATORS

BASED ON INSTRUMENT FUNCTIONS

ABILITY TO SIMULATE JAW MOVEMENTS

ADJUSTABILITY OF THE

ARTICULATOR

THEORIES OF OCCLUSION

TYPE OF INTER- OCCLUSAL

RECORD USED

Page 40: ARTICULATORs

40

BASED ON INSTRUMENT FUNCTION At the International Prosthodontic Workshop on

Complete Denture Occlusion at the University

of Michigan in 1972, an articulator classification

was developed based on the instrument’s

function, instrument capability, intent,

registration procedure, and registration

acceptance were considered in developing the

classification. The classification is as follows-

Page 41: ARTICULATORs

41

BASED ON INSTRUMENT FUNCTION

CLASS I

CLASS II

CLASS IV

CLASS III

Page 42: ARTICULATORs

42

CLASS I:

Simple holding instruments capable of accepting a single

static registration. Vertical motion may or may not be

possible.

The first articulators were called “slab articulators” and

were formed by extending plaster indices from the rear of

the casts.

The hinge- joint articulator is representative of this class.

J.B. Gariot reportedly designed the first hinge articulator

about 1805.

The “barn- door” hinge with an anterior vertical stop

qualifies for this classification.

Page 43: ARTICULATORs

• CLASS II :Instruments that permit horizontal as well as vertical motion but do not orient the motion to the temporomandibular joint via a face-bow transfer.

43

CLASS II

CLASS II-A

CLASS II-B

CLASS II-C

Page 44: ARTICULATORs

44

CLASS II-A : Eccentric motion permitted is based on average or arbitrary values. Example :

a) Articulator by GRITTMAN (1899) ; condylar path inclined at 15 degree

b) GYSI SIMPLEX articulator (1924) ; condylar path inclined at 30 degree and the incisal guidance is fixed at 60 degree.

CLASS II-B : Eccentric motion permitted is based on theories of arbitrary motion. Example :

MONSOON’s articulator (1914)

CLASS II-C : Eccentric motion permitted is determined by the patient using engraving methods. Examples :

HOUSE articulator (1927)

Page 45: ARTICULATORs

45

•CLASS III :Instruments that simulate condylar pathways by using average or mechanical equivalents for all or part of the motion. These instruments allow for joint orientation of the casts via a facebow transfer.

CLASS III

CLASS III-A

CLASS III-B

Page 46: ARTICULATORs

46

CLASS III-A : Instruments that accept a static protrusive registration

and use equivalents for the rest of the motion. EXAMPLE :

A) Hanau Model H (1923)

B) Dentatus (1944)

C) Arcon (1950)

CLASS III-B : Instruments that accept a static lateral protrusive

registrations and use equivalents for the rest of the motion. EXAMPLE:

A) Gysi’s trubyte Articulator (1926)

B) Kinoscope (1927)

C) Tripod Type Articulator (1928)

D) Ney Articulator (1960)

E) Hanau 130-21 (1964)

F) Teledyne Articulator (1975)

G) Panadent Articulator

Page 47: ARTICULATORs

47

CLASS IV : Instruments that will accept three – dimensional dynamic registrations. These instruments allow for joint orientation of casts via a face – bow transfer.

CLASS IV

CLASS IV-A

CLASS IV-B

Page 48: ARTICULATORs

48

CLASS IV –A : The cams representing the condylar paths are formed by the registrations engraved by the patient. These instruments do not allow for discrminating capability.

EXAMPLE : TMJ Instrument (1965)

CLASS IV-B : Instruments that have condylar paths that can be angled and customized either by selection from a variety of curvatures, by modification, or both.

EXAMPLES : Gnathoscope (1955) Denar Fully Adjustable Simulator

Page 49: ARTICULATORs

49

BASED ON THE ABILITY TO SIMULATE JAW MOVEMENTS

CLASS I:

These are instruments that receive and reproduce

stereograms (pantograms).

These articulators can be adjusted to permit individual

condylar movement in each of the three planes.

They are capable of reproducing the timing of the side

shift of the balancing side and its direction on the

working side.

Examples: Mc Collum Gnathoscope, Hanau Kinescope,

Ney articulator, Denar D5A.

Page 50: ARTICULATORs

50

CLASS II:

Instruments that will not receive stereograms.

Some of the instruments have fixed controls

while others are adjustable, but usually in no more

than two planes.

This class is divided into four types:

1) Hinge

2) Arbitrary

3) Average

4) special

Page 51: ARTICULATORs

51

Type 1: HINGE: This type is capable of

opening and closing in a hinge

movement. Examples: Barn Door Hinge,

Gariot, Hageman Balancer, Gysi, Acme.

Type 2: ARBITRARY: This is designed to

adapt to specific theories of occlusion or

is oriented to a specific technique.

Examples: Monson, Transgraph, Gnathic

Relator.

Page 52: ARTICULATORs

52

Type 3: AVERAGE: This type is designed to provide

condylar element guidance by means of averages,

positional records or mini- recorder systems. Most

permit adjustments of both horizontal and lateral

guidance surfaces. Some type of facebow can be

used in maxillary cast orientation. Examples:

Dentatus, Whip- Mix, Denar, Panadent.

Type 4: SPECIAL: This type is designed and used

primarily for complete dentures. Example: Stansbery

Tripod, Kile Dentograph, Irish Dupli- Functional.

Page 53: ARTICULATORs

53

BASED ON THE ADJUSTABILITY OF THE ARTICULATOR

1. NON- ADJUSTABLE2. SEMI- ADJUSTABLE3. FULLY- ADJUSTABLE

Page 54: ARTICULATORs

54

Non- adjustable articulators:

They can open and close in a fixed horizontal axis.

They have a fixed condylar path along which the condylar ball can be moved to simulate lateral and protrusive jaw movement.

The incisal guide pins ride on an inclined plate in a fixed inclination.

The distance between the teeth and the axis of rotation on this small instrument is considerably shorter than it is in the skull, with a resultant loss of accuracy.

Page 55: ARTICULATORs

55

Advantages

Inexpensive

Less time required to

mount the casts on the articulator

Disadvantages

It reproduces only one contact

position

It demands increased intraoral

adjustment time.

Page 56: ARTICULATORs

56

Semi adjustable articulators:

They have :

adjustable horizontal condylar paths

adjustable lateral condylar paths

adjustable incisal guide tables

adjustable intercondylar distances.

There are two types of semi- adjustable

articulators:

1. Arcon articulators

2. Non- arcon articulators.

Page 57: ARTICULATORs

57

Comparison between ARCON and NON ARCON articulators

Arcon instrument has its condyles on the lower member and the condylar guides on the upper member.

e.g. Whipmix articulator, hanau university series.

Non arcon or condylar instruments are those that have the condyles on the upper member and condylar guides on the lower member.

e.g. Hanau Model H, dentatus, gysi.

Page 58: ARTICULATORs

58

Advantages of Arcon articulators:

The face- bow transfer, occlusal plane and the

relationship of the opposing casts are preserved when the

articulator is opened and closed.

Condyles move in a relationship to their condylar housing that is similar to the

way the condyles move in the relationship to the

glenoid fossa in the skull. This makes visualization and understanding of condylar

movements easier.

Page 59: ARTICULATORs

59

Fully adjustable articulators: It accepts three dimensional dynamic

registrations. They are capable of being adjusted to

follow the mandibular movement in all directions.

The articulator is capable of repeating most of the precise condylar movements depicted in any individual patient. These adjustments include:

a) Adjustable horizontal condylar guidanceb) Adjustable lateral condylar guidancec) Adjustable incisal guide tabled) Adjustable intercondylar distance.

Page 60: ARTICULATORs

60

Condylar inclination:- As on the semiadjustable articulator , the

angle at which the condyle descends on the fully adjustable articulator during protrusive & laterotrusive movements can be altered.

Although the semi adjustable articulator can usually provide a condylar movement only in a straight pathway , the fully adjustable articulator is capable of adjusting the condylar pathway to duplicate the angle & curvature of the patient’s condylar movement.

Page 61: ARTICULATORs

61

Bennett angle & lateral translation movement:-

The fully adjustable articulator has adjustments that permit duplication of both the patient’s bennett angle & the immediate lateral translation movement of the patient condyle .

Many semiadjustable articulator cannot duplicate this exact pathway because only flat surfaces are available to guide the condyle.

Page 62: ARTICULATORs

62

Rotating condylar movements

During a laterotrusive movemement ,the rotating (i.e ., working) condyle does not purely rotate around a fixed point ; however it can move slightly laterally .

This lateral shift can also have a superior , inferior, forward , or backward component , which can influence the fossa depth & cusp height , as well as the ridge & groove direction developed in the posterior teeth.

Page 63: ARTICULATORs

63

The rotating condylar movements affects both the working & non working sides but has its greatest effect on the working side.

Semiadjustable articulators do not have the ability to compensate for this movement.

The fully adjustable articulators can be set so that the pathway of the rotating condyle on the articulator will duplicate that of the patient.

Page 64: ARTICULATORs

64

Intercondylar distance The distance between the rotating

centers of the condyle on the fully adjustable articulator can be modified to match that in the patient .

Often three general settings are available on the semiadjustable articulator:› Small › Medium › large

Page 65: ARTICULATORs

65

With the fully adjustable articulator , a complete range of intercondylar distance can be selected .

Therefore the intercondylar adjustment is set at the precise milimeter distance as determined from the patient.

This then allows more accurate duplication of intercondylar distance & thus minimize errors .

Page 66: ARTICULATORs

66

Advantages

Most accurate instrument to

reproduce restorations that precisely fit the

occlusal requirements of the patient.

Disadvantages

Demands high degree of skill

Expensive

Time consuming

its ability to duplicate mandibular movement.

its ability to duplicate mandibular movement.

Page 67: ARTICULATORs

67

ARTICULATORS BASED ON THEORIES OF OCCLUSION

THE BONWILL THEORY OF OCCLUSION proposed that the teeth move in relation to each

other as guided by the condylar controls and the incisal point.

It was known as the theory of the equilateral triangle, in which there

was a 4-inch (10 cm) distance between the condyles and

between each condyle and the incisor point.

Page 68: ARTICULATORs

68

The articulator designed by W.G.A. Bonwill allows

lateral movement, but since the condylar guidances

are not adjustable, they permit movement only in the

horizontal plane.

Page 69: ARTICULATORs

69

THE CONICAL THEORY OF OCCLUSION proposed that the lower teeth move over the surfaces of upper teeth as over the

surface of a cone, generating an angle of 45 degrees with the central axis of the cone tipped 45 degrees to the occlusal

plane.

The Hall Automatic articulator, designed by R.E.Hall confirms to conical theory of

occlusion.

Page 70: ARTICULATORs

70

THE SPHERICAL THEORY OF OCCLUSION showed the lower teeth

moving over the surface of the upper teeth as

over the surface of a sphere with a diameter of 8 inches (20 cm).

The center of the sphere was located in the

region of glabella, and the surface of

the sphere passed through

the glenoid fossae along or concentric with the articulating

eminences.

The theory was

proposed by G.S. Monson

in 1918.

Page 71: ARTICULATORs

71

The articulators that function on theories

of occlusion have one common fault:

they make no provision for variations

from the theoretical relationships that

occur in different persons. When the

varying inclinations of condylar paths of

the two sides of many patients are

recognized, the need for modification

becomes apparent.

Page 72: ARTICULATORs

72

ARTICULATORS BASED ON TYPE OF RECORD USED FOR THEIR ADJUSTMENT

- Interocclusal records

- Graphic records

- Hinge- axis records.

The general classes of records are used for transferring maxillomandibular relationships from the patient to the articulator:

Page 73: ARTICULATORs

73

Interocclusal record adjustment-

Interocclusal records may be made in wax,

plaster of Paris, ZOE paste, or cold- cure

acrylic resin.

Each record is of only one positional

relationship of the lower jaw to the upper jaw.

Page 74: ARTICULATORs

74

• Individually adjustable horizontal condylar guidances

• Variable controls for the Bennett shift

• Variable intercondylar distances• Split-axis condylar guidance

controls • Adjustable incisal guidance

controls.

The mechanical features that determine whether an articulator can be adjusted to accommodate interocclusal records include:

Page 75: ARTICULATORs

75

Graphic record adjustment-

Articulators designed for use with graphic records

are generally more complicated than those designed

for interocclusal records.

Since graphic records consist of records of the

extreme border positions of mandibular movements,

the articulators must be capable of producing at

least the equivalent of curved movements. The

reason is that the border movements of the mandible

are in curves.

Page 76: ARTICULATORs

76

Hinge- axis location for adjusting

articulators-

All the instruments that can be adjusted

to graphic records have one feature in

common: the necessity for correct

location of the opening axis of the

mandible.

Page 77: ARTICULATORs

77

CLASSIFICATION ACCORDING TO VARIOUS AUTHORS

According to Gillis (J Am Dent Assoc 13:3, 1926)

Adjustable Fixed

According to Boucher (J Dent Res 14:39, 1934) Non- adjustable Adjustable a) A Two- dimensional instrument.

b) A Three- dimensional instrument.

Page 78: ARTICULATORs

78

According to Kingery (The American Textbook of

prosthetic Dentistry, ed 7. Philadelphia, 1942)

Simple articulatorsAdjustable or adaptable articulators

According to Beck (J Am Dent Assoc 64:468, 1962)

Suspension instrument Axis instrument Tripod instrument

Page 79: ARTICULATORs

79

According to Weinberg (J Prosthet Dent 13:645, 1963)

Arbitrary – e.g. – Monson articulator, Hageman balancer

Positional – e.g. – Stansbery Tripod articulator

Semi adjustable – e.g. – Hanau Model H Fully adjustable – e.g. – Hanau kinescope,

Gysi Trubyte

Page 80: ARTICULATORs

80

According to Posselt (Posselt, U: Physiology of occlusion, 1968)

Plain line Mean value Adjustable

According to Thomas (J Prosthet Dent 30:11, 1973)

Arbitrary Positional Functional

Page 81: ARTICULATORs

81

According to Heartwell and Rahn (1974)

Instruments that will receive and reproduce pantographs and graphic tracings

Instruments that will not receive pantographs, which are sub divided into four types:

a) hinge type b) arbitrary c) adjustable d) instruments designed and used for complete

denture construction.

Page 82: ARTICULATORs

82

According to Sharry (Sharry, J.J.: Complete Denture Prosthodontics, ed 3, 1974)

Simple Hinge type Fixed guide type Adjustable

According to Bergstrom Arcon - e.g. – Whipmix articulator Non arcon - e.g. – Hanau model H

Page 83: ARTICULATORs

83

According to Rihani (1980)

Fully adjustable-

examples: Hanau Kinescope, Mc Collum Gnathoscope,

Simulator, Denar D4-A

Semi adjustable-

examples: Snow Acme, Gysi Adaptable, Gysi trubyte,

Dentatus, Hanau 130-28, Whipmix.

Non adjustable- examples: Gariot, Barn door hinge, Walker, Gritman, Snow,

Transograph

Page 84: ARTICULATORs

84

FULLY ADJUSTABLECan accept the following five records:Face bow recordCentric jaw relation recordProtrusive recordLateral recordIntercondylar distance record

SEMI-ADJUSTABLECan accept the following three records:Face bow recordCentric jaw relation recordProtrusive record

NON-ADJUSTABLECan accept one or two of following three records:Face bow recordCentric jaw relation recordProtrusive record

Page 85: ARTICULATORs

85

ADJUSTABLE VS. SEMIADJUSTABLE ARTICULATORS

(J Prosthet Dent 1973; 29(3): 269-75)

The semi-adjustable articulator guides only the lateral component of the rotating condylar movement, whereas the fully adjustable articulator may be set to simulate all components of mandibular movement.

The purpose of this study was to measure the discrepancies that may exist in articulator capability due to incomplete movement simulation.

Page 86: ARTICULATORs

86

The effects of following on cusp positions were ascertained:

a) Intercondylar widthb) The shape of the condylar housingc) The timing and direction of the side shift.

To preclude the necessity of transferring the casts from one articulator to the other, only the fully adjustable articulator was used.

The capabilities of semi- adjustable articulator was used as control, while the capabilities of fully adjustble articulator was used as experimental.

Page 87: ARTICULATORs

87

A semiadjustable articulator was compared with a fully adjustable articulator to determine the discrepancy of eccentric pathways of cusp motion.

The effects of intercondylar width, mediotrusion timing, superior wall shape, laterotrusion direction, and immediate side shift were measured.

All produced a significant variation.

Page 88: ARTICULATORs

88

PARTS OF AN ARTICULATORIt consists of an upper member and a lower member. The upper member represents the

maxilla and the lower member represents the mandible.

The upper member is a triangular metal plate and the lower member is a L -shaped frame. The upper and lower members articulate around the condylar

guidance. The condylar guidance represents the glenoid fossa of the temporomandibular joint.

A vertical rod (incisal pin) separates the upper and lower triangular components in the anterior end. The vertical rod rests on the incisal table of the lower member also known as the incisal guide

table.

Page 89: ARTICULATORs

89

PARTS OF HANAU ARTICULATOR AND THEIR FUNCTIONS

Page 90: ARTICULATORs

90

CONDYLAR GUIDANCE

The condylar guidances are the control

centers of the Articulator and they

adjustably assimilate the multiple

function of the glenoid fossa.

Page 91: ARTICULATORs

91

The condylar Track may be adjustibly inclined on the horizontal transverse axis from a “zero” to a plus 60 degree or to a minus 20 degree. These inclinations are

termed the protrusive inclination and simulate the patients superior

wall of the fossa.

The condylar track may also be adjusted on the vertical axis from a “zero” sagittal to 30 degree. This angle is termed the progressive

Bennett angle and corresponds to the medial wall of the patients

fossa.

Page 92: ARTICULATORs

92

CLOSED CONDYLAR TRACK

The closed condylar guidance track

rotates in an enclosed housing which

stops the Condylar Element ,preventing

the accidental disengagement of the

upper member.

Page 93: ARTICULATORs

93

FIXED CENTRIC STOP

Some products have a centric stop at

the posterior end of the Track to limit

anterior movement of the Condylar

Element. When the Element rests

against the Stop it is at centric, the

point from which the protrusive or

Bennett angles both emanate.

Page 94: ARTICULATORs

94

ADJUSTABLE PROTRUSIVE-RETRUSIVE

This micrometer adjustment permits the Condylar Element to be protruded 6mm from centric or to be retruded 3mm from centric.

An axial reference line transcribes the one mm. spaced lines on the P-R Screw and a like line appears on the Sleeve of the Guidance.

Page 95: ARTICULATORs

95

Loosen the Thumbscrew at the medial side of the Guidance and rotate the P-R screw to abut the wide “zero” centric line with the Sleeve end .

Protrusion or retrusion of the Condylar Element can be adjusted by rotation of the P-r Screw. One full turn of the Screw equals 1mm protrusion or retrusion .

Page 96: ARTICULATORs

96

CENTRIC LOCK

Engagement of a centric lock depresses a Centric Pin, causing it to arrest the Condylar Element at the centric position.

When locked, the Upper Member is restricted to an opening and closing movement only.

Releasing the Centric Lock two full turns will disengage the Centric Pin and return the Elements freedom of movement in the Condylar track.

Page 97: ARTICULATORs

97

CONDYLAR SHAFTS

The Condylar Shafts adjustibly slide in the “wings” of the Lower Member.

They have been factory fixed by Setscrews.

A resilient Bumper will protectively stop the Upper Member and against the “wing” of the Lower member when fully opening the Articulator.

Page 98: ARTICULATORs

98

DUAL-END INCISAL PIN

Coinciding with these Condylar Shaft adjustments is an alignment of the chisel edge of the incisal pin with the central table of the incisal guide.

The incisal pin serves as the forward control of the Aticulator.

It cooperatively maintains a vertical stop.

Page 99: ARTICULATORs

99

It provides a stylus contact for the excursive movements of the Articulator.

A midline groove is cut in the Incisal Pin about one inch from the spherical tip. Five additional lines calibrated in mm extend on either side thereof. These lines are used for recording or altering the vertical dimension.

Page 100: ARTICULATORs

100

Two annular grooves appear on the Incisal Pin at 37 and 54mm below the Frankfort Horizontal Plane. These grooves form arbitrary vertical landmarks for alignment of the incisal edge of the maxillary centrals when making a Facebow transfer.

Page 101: ARTICULATORs

101

The incisal pin extends beyond the top of the Upper Member and provides a third point of stability when inverting the Articulator for mandibular cast mounting.

The spherical tip of this incisal pin serves as the Dual-End and is useful for fabricating customized acrylic anterior guide tables.

Page 102: ARTICULATORs

102

ADJUSTABLE INCISAL GUIDE

The Adjustable Incisal Guide provides an independent adjustment of anterior guidance. It cooperates with the Incisal Pin and Condylar Guidances to present a stable ,three dimensional programmed guide pattern for the mounted casts.

Page 103: ARTICULATORs

103

The Incisal Guide rotates antero posteriorly from a horizontal “zero” degree to a 60 degree positive inclination of protrusion which is then secured by the small Locknut. The central guiding table is 5.56mm wide and forms the inclined surface for the protrusive guidance of the incisal pin.

Page 104: ARTICULATORs

104

Seperately adjustable Lateral Wings

elevate by a Thumbscrew from a “zero’

horizontal to a 45degree inclined and

are fixed by a Thumbnut. The

calibrations are very small and serve

only as a reference

Page 105: ARTICULATORs

105

An anterior slot in the lower member allows repositioning of the incisal guide. Adjust and lock the Guide at “zero” horizontal and slightly loosen the platform Lockscrew.

This adjustment will place the incisal pin contact on the rotational center of the guide, thereby maintaining the vertical dimension when adjusting the inclination for protrusion.

Page 106: ARTICULATORs

106

ORBITALE INDICATOR

This crescent represents the patient’s infra- orbitale notch and is the anterior reference landmark of the Frankfort horizontal plane.

When used with an Orbitale Pointer on a face bow, it provides an anatomical vertical orientation for the upper arch, obviating the use of any average reference lines on the incisal pin.

Page 107: ARTICULATORs

107

MOUNTING PLATES

Mounting plates are used to lute the upper and lower casts to the upper and lower members of articulator by means of a gypsum material.

Page 108: ARTICULATORs

108

OPERATORY PROCEDURE

INTEROCCLUSAL RELATION RECORDS:

The maxillary and mandibular occlusal rims have been prepared and patient adjusted to the correct vertical dimension, the occlusal plane and the desired horizontal overlap.

The top of the Bitefork or Biteplane, the stern at the patient’s left, is covered with a triple layer of base plate wax. Heat seal the periphery and soften the wax throughout in a water bath.

Page 109: ARTICULATORs

109

The softened wax impression

material on the Bitefork is

seated against the occlusal

surface of the upper rim and is

hand molded into and around

the notches.

Chill the Bitefork index and

check to assure removal and

accurate replacement of the

bite rim.

The Bitefork may alternately

be attached directly to the

upper occlusal rim by heating

the forks and piercing them

fully into the wax rim.

Page 110: ARTICULATORs

110

Two interocclusal positional records of a terminal relation are required:

1. A centric interocclusal relation record is patient recorded at the vertical dimension of occlusion.

2. A protrusive interocclusal relation record is patient recorded at approximately 6 mm anterior of centric. Closure is made to a near anterior, non- piercing occlusal contact, being certain that the bite rims remain securely seated on their ridges.

Page 111: ARTICULATORs

111

LABORATORY PROCEDURE ARTICULATOR PREPARATION: Adjust the protrusive inclination of both condylar

guidances to 30 degrees and tighten the Thumbnuts.

Adjust the Bennett angles of both condylar guidances at 30 degrees.

Adjust the incisal pin to align the midline calibration to the top edge of the upper member.

Adjust the incisal guide to a “zero” degree.

Page 112: ARTICULATORs

112

Slide the platform to align the incisal pin contact over the “zero” indicating line on guidance.

Adjust both the P-R screws to about “zero” centric line.

Secure the “zero” centric adjustment by tightening the thumbscrews at the medial side.

Apply a thin coating of petroleum jelly to all surfaces of the articulator that will be exposed to the stone mounting media.

Firmly attach a mounting plate to the upper member.

Page 113: ARTICULATORs

113

MOUNTING MAXILLARY CAST Swing the upper member

of the articulator open to the centric bumper. Securely attach the mounting platform to the lower member.

Insert the lower end of the Transfer Rod/ Bitefork index assembly into the socket of the mounting guide.

Raise the pivot to contact the underside of the Bitefork index and lock in position by the thumbscrew.

Page 114: ARTICULATORs

114

Securely seat and accurately lute the upper

occlusal rim into the occlusal wax index on the

bitefork.

Mixture of stone is placed on the wetted mounting surface of the cast. The upper member is then

swung forward to embed the mounting plate and to bring the incisal pin into contact with the incisal

guide.

Upon complete set of mounting, the sticky wax luting is carefully broken

from the occlusal rim. The mounting guide and transfer record are

removed from the lower member.

Page 115: ARTICULATORs

115

MOUNTING MANDIBULAR CAST

Invert the articulator and swing the lower member back to the bench.

Apply a thin coating of petroleum jelly to all surfaces of lower member that may be contacted by the gypsum material.

Firmly attach a mounting plate to the lower member.

Page 116: ARTICULATORs

116

Securely seat the mandibular cast in the bite rim and lute together with sticky wax.

A mix of stone is placed on the wetted mounting surface of the lower cast and mounting is completed, securing the central locks.

Page 117: ARTICULATORs

117

ADJUSTMENT OF HORIZONTAL CONDYLAR GUIDANCE

Upon complete set of the mount, the Articulator is placed into an upright position

Carefully cut the tacking at the occlusal surface and remove the centric interocclusal relation record

Page 118: ARTICULATORs

118

Loosen the centric locks and the Thumbnuts for horizontal inclination of the condylar guidances. Raise the Incisal Pin to remove the possibility of mechanical interference with the Incisal Guide

Seat the protrusive interocclusal relation record onto the lower occlusal rim. Carefully guide the Upper Member into the imprint of the protrusive relation record

Page 119: ARTICULATORs

119

Grasp the upper cast to maintain a

tactile feel of the casts at the protrusive

record and-

Rotate the Right and Left Guidances

back and forth toaccurately seat the

upper and lower rims into the protrusive

relation record

Tighten the Thumbnuts for horizontal

inclination. The protrusive relation

record is then removed.

Page 120: ARTICULATORs

120

ADJUSTMENT OF LATERAL CONDYLAR GUIDANCE (BENNETT ANGLE) HANAU Formula

L = H/8 + 12

is used for adjusting the lateral condylar guidance

In this formula, ”H” is the Horizontal Condylar (Protrusive) inclination and

“L” is the Lateral Condylar Guidance (Bennett Angle)

Page 121: ARTICULATORs

121

ARRANGEMENT OF ANTERIOR TEETH

The six upper and six lower teeth are

set up in the arrangement dictated by

the patient’s esthetic and phonetic

requirements.

Page 122: ARTICULATORs

122

INCISAL GUIDE PREPARATION

Lock the Articulator into centric. The occlusal rims are then seated onto their casts at the established vertical and centric relation

Lower the incisal pin into contact with the “zero” horizontal incisal guide, the chisel end of the pin resting crosswise on the center table and lock in the position by the Thumbscrew

Page 123: ARTICULATORs

123

ADJUSTMENT OF INCISAL GUIDE Gently guide the upper

cast into a straight protrusion

The lingual edges of the upper central incisors are brought into contact with the incisal edges of the lower incisors

The incisal guide is then rotated antero- posteriorly to make contact with incisal pin, tightening the small locknut to maintain the angulation

Page 124: ARTICULATORs

124

The upper cast is then guided into a right lateral cuspid to cuspid guidance relation by thumb pressure at the right side of the upper cast to assure the Bennett Shift

The Lateral wing is then elevated to contact the corner of the incisal pin and the Locknut is tightend to maintain this adjustment

Page 125: ARTICULATORs

125

Apply thumb pressure at the left side of the upper cast and guide it into a left lateral cuspid to cuspid excursion

Adjust the remaining Lateral Wing to contact the incisal pin and secure the adjustment by tightening the Locknut.

Page 126: ARTICULATORs

126

ARRANGEMENT OF POSTERIOR TEETH

The remaining teeth are set into centric occlusion and checked in working, balancing in protrusive excursions.

The incisal pin acts as the vertical stop and must remain in contact with the incisal guide surfaces from centric throughout all excusive movements

The waxed occlusal rims may be tried in the patient for esthetics, occlusal function.

Page 127: ARTICULATORs

127

Reset the completed occlusal rims onto their master casts. Seal the rims to the casts to preserve the tissue surface from plaster or stone seepage during the Flasking procedure.

Page 128: ARTICULATORs

128

PROCESSING

Before removing the master cast/waxed dentures, record the articulator calibrations for later use

Record the –Serial Number, R & L Horizontal Inclination, R& L Bennett Angle and Vertical of incisal pin

Page 129: ARTICULATORs

129

THE HANAU ARTICULATOR

HANAU H2

HANAU ARCON H2

HANAU WIDE VUE HANAU RADIAL SHIFT

HANAU ARTICULATOR

Page 130: ARTICULATORs

130

THE HANAU H2 ARTICULATOR

The Hanau model H was designed by Rudolf

Hanau a mechanical engineer in 1923. This

articulator is the forerunner of the present-day

Hanau H2.

Non- arcon type.

Hanau felt that as the “realeff” (resiliency and like

effect” of the denture supporting tissues)

decreased, the instrument would more closely

simulate mandibular movement.

Page 131: ARTICULATORs

131

The maxillary cast may be mounted on the

Hanau H2 articulator by means of a face-

bow transfer.

The following face-bows can be utilized

with the Hanau H2. These are:

1. Facia face bow

2. Earpiece face bow

3. Twirl bow

Page 132: ARTICULATORs

132

When transferring the ear piece face bow to the

articulator, the earpieces are seated on the

auditory pins of the centric locks.

The facia and earpiece face bows may use either

the orbitale or the incisal plane as the anterior

reference point for making the face bow transfer.

The anterior reference point along with the two

posterior reference points form the axis orbital

plane, which is transferred to the articulator

along with the maxillary cast.

Page 133: ARTICULATORs

133

The axis-orbital plane transfer allows the maxillary cast to be transferred to the

articulator so that the occlusal plane has a relationship in the

articulator that is similar to that in the patients skull.

When the orbitale is used, the face bow should be

equipped with an orbitale pointer that is related to the orbitale indicator on the upper member of the

articulator.

Without the orbitale pointer, the incisal plane is adjusted so that it is level with the upper notch on

the incisal pin when making the face bow

transfer.

Page 134: ARTICULATORs

134

Following the relationship of the mandibular casts to the maxillary cast by means of a centric relation interocclusal record, the horizontal condylar inclinations must be determined.

This is accomplished by means of a protrusive interocclusal record. The patient should protrude approx. 6 mm.

Page 135: ARTICULATORs

135

The horizontal condylar guidances are adjusted until the maxillary cast seats into the protrusive interocclusal record.

Page 136: ARTICULATORs

136

The amount of Bennett

movement is calculated from

the horizontal condylar

settings by Hanau’s

equation. Bennett angle L = H + 12 8

H= Horizontal condylar

inclination determined by

protrusive interocclusal

record

Page 137: ARTICULATORs

137

The incisal guide table is

adjustable in both sagittal and

frontal planes.

There is a single sagittal and

right and left frontal

adjustments.

The adjustments are made to

compensate the amount of

horizontal and vertical overlap

incorporated in the anterior

arrangement of denture teeth.

Page 138: ARTICULATORs

138

ARCON PRINCIPLE IN THE CONDYLAR MECHANISM OF ADJUSTABLE ARTICULATORS

(J Prosthet Dent 1963; 13(2): 263-68)

An articulator of the Arcon type has the condylar slot fixed to the upper member and the ball attached to the lower member. This positioning of the condylar elements is the reverse of the usual arrangement. Many articulators, including the Gnathologic type, use the arcon principle.

Page 139: ARTICULATORs

139

It is claimed that this principle truly represents the condition found physiologically where the temporomandibular fossae are located in the skull and condyles are a part of mandible.

Page 140: ARTICULATORs

140

MECHANISM OF THE ARTICULATORS

The mechanism of the articulators are reversed in function. In the condylar type of articulator , the angulation of the condylar slot is constant to the lower member.

During protrusive movements, the angulation changes between the upper member and the condylar slot of the articulator.

In the Arcon articulator, reverse is true.

Page 141: ARTICULATORs

141

PROTRUSIVE CONDYLAR READINGS The protrusive condylar

inclinations were measured in relation to the horizontal plane of the instrument.

When the condylar instrument (Hanau) returned to the centric position, the condylar slot remained at 40 degree angulation to that plane.

However, as the Arcon articulator returned to centric position, there was a changed angulation of the condylar slot at 39 degrees.

The difference in readings occurs because of the mechanical method of producing motion.

Page 142: ARTICULATORs

142

HANAU ARCON H2

The Hanau Arcon H2 was introduced by Hanau in 1977. It is a semiadjustable arcon type of instrument.

It consists of an upper member containing the condylar guidance elements and a lower

member to which the condylar spheres are attached.

The upper and lower members are mechanically connected.

The Hanau articulator is classed as a modified two- dimensional instrument.

Page 143: ARTICULATORs

143

The upper cast is oriented to the upper member

( which represents the skull) by face- bow transfer

record. The arbitrary face- bow is routinely used for

complete dentures.

The Hanau face- bow consists of a U- shaped frame or

assembly that is large enough to extend from the

region of the TMJs to a position 2 to 3 inches (5 to 7.5

cm) in front of the face and wide enough to avoid

contact with the sides of the face.

Page 144: ARTICULATORs

144

The parts that contact the skin over the TMJs are the

condyle rods, and the part that attaches to the

occlusion rims is the fork.

The condyle rods are positioned on a line extending

from the outer canthus to the top of the tragus and

approximately 13 mm in front of the external auditory

meatus.

This placement of the rods will locate them within 2

mm of the true center of the opening axis of the jaws.

Page 145: ARTICULATORs

145

The lower cast is oriented to the lower member of the

articulator, representing the mandible, by relating the

lower to the upper cast through an interocclusal CR

record.

The fork of the face- bow is attached to the maxillary

occlusion rim, thus the record is a simple

measurement from the jaws to the approximate axis

of the jaws.

Page 146: ARTICULATORs

146

The horizontal condylar guidances are adjusted by

an interocclusal protrusive record. The lateral

condylar guidances may be set arbitrarily or may be

adjusted by right and left lateral interocclusal

records.

The lateral condylar guidances on this articulator do

not allow upward, downward, forward, or backward

movement of the working condylar sphere.

Page 147: ARTICULATORs

147

The articulator is provided with an adjustable incisal

guide table that is routinely used for removable

prosthodontic restorations.

It has a straight incisal guide pin with a flat end, which

permits movements on the guide table.

The pin on the Hanau articulator is adjustable and

allows for vertical changes without changes in pin

position relative to the middle of the incisal guide

table.

Page 148: ARTICULATORs

148

HANAU RADIAL SHIFT : It was first produced in 1981.

Arcon instrument with fixed intercondylar distance of 110 mm.

The condylar guidance of the articulator is designed to incorporate a curved immediate side shift (radial shift) with an adjustable progressive Bennett angle.

Three incisal guide tables are available :- A mechanical table- A flat table- A pantacrylic table

Page 149: ARTICULATORs

149

HANAU WIDE VUE : These are the newest articulators introduced by HANAU.

HANAU WIDE VUE and HANAU WIDE VUE II, both are arcon instruments with a fixed intercondylar distance of 110 mm.

The only difference between the two instruments :

HANAU WIDE VUE has a closed condylar track and WIDE VUE II has an open condylar track.

Page 150: ARTICULATORs

150

THE WHIP MIX ARTICULATOR

The basic whip-mix articulator was designed by Charles Stuart (1955) and manufactured by Whip-Mix Corp.

It is an arcon articulator as the condylar controls are attached to the upper member of the articulator.

Page 151: ARTICULATORs

151

The intercondylar distance is adjustable to three positions by means of removable condylar guidance spacers along the instruments horizontal axis. :

Small (S) - 96 mm

Medium (M)- 110 mm

Large (L) - 124 mm

Page 152: ARTICULATORs

152

A face bow transfer may be utilized for mounting the maxillary cast. The horizontal condylar inclinations are set by means of protrusive interocclusal records.

The amount of Bennett movement is set by means of a lateral interocclusal record.

Page 153: ARTICULATORs

153

Spring latch assembly

The upper and lower members are mechanically attached by means of a spring latch assembly.

Page 154: ARTICULATORs

154

There are two different face bows that can be

utilized with the Whip Mix articulator. They

are:

1. Quick mount or earpiece face bow and

2. Adjustable axis or kinematic face bow

Page 155: ARTICULATORs

155

The earpieces are placed in the

external auditory canals when

adjusting the face bow to the patient.

The patient’s approximate intercondylar

distance is determined from the scale on the front of the face bow as S, M or L indicating small, medium or large

intercondylar distance.

The earpieces are seated on pins on

the condylar housings.

The bridge of the nose is utilized as

the anterior reference point.

Page 156: ARTICULATORs

156

Whip mix earpiece facebow attached to skull

Page 157: ARTICULATORs

157

Following the relationship of the

mandibular cast to the maxillary cast by

means of a centric relation interocclusal record, the horizontal condylar inclinations must be determined.

This is accomplished by means of protrusive interocclusal records. These records should be made at approx. 6

mm of protrusion.

The casts are seated in the interocclusal records and the

horizontal condylar guidance adjusted until it contacts the condyle.

The Bennett angulation is set with lateral

interocclusal records. The casts are seated in the lateral interocclusal records and the medial walls of the condylar

housings adjusted until they contact the

condyles.

Page 158: ARTICULATORs

158

The Whip-Mix articulator is equipped either with a mechanical incisal guide table or a plastic incisal guide table that can be individually customized with autopolymerizing resin.

The incisal guide pin is straight and one end is flat and the other end rounded. The flat end is used with the mechanical incisal guide table, while the rounded end is used with the plastic table.

Page 159: ARTICULATORs

159

The basic whip mix has numerous modifications that are available.

Condylar thumb- lock screws may be added to assure proper seating of condyles when making hinge articulator movements.

Page 160: ARTICULATORs

160

Optional immediate side shift guides are available from 0.25mm to 1mm.

When using the shift guides, the articulator must be equipped with detent mechanism to return the upper member of articulator to centric position.

Page 161: ARTICULATORs

161

The Accumount mounting system for articulator interchangeability utilizes a special table that is precisely attached to the lower member with a low-fusing alloy during manufacturing. The mounting plate is attached to this table when mounting the mandibular cast.

Page 162: ARTICULATORs

162

THE DENTATUS ARTICULATOR

The Dentatus is similar to the other articulators

described.

It is employed quite extensively in European

dental schools, particularly the Scandinavian ones.

It too is a semi- adjustable instrument that, like the others, can be modified

according to several patient variables.

Page 163: ARTICULATORs

163

AN EVALUATION OF BASIC ARTICULATORS AND THEIR CONCEPTS

PART I : BASIC CONCEPTS

PART II : ARBITRARY, POSITIONAL, SEMIADJUSTABLE ARTICULATORS

(Lawrence A. Weinberg ;J Prosthet Dent, 1963, vol 13)

Page 164: ARTICULATORs

164

The objective of all articulators is to serve as a laboratory aid in imitating physiological motion by substituting mechanical equivalents for anatomic parts.

Human motion of the mandible are 3-D curved motions because the skeletal framework is moved by muscles and hinged at TMJ.

Page 165: ARTICULATORs

165

Reproduction of any such motion requires the establishment of 3 fixed points on or attached to the object.

Recording the starting position, the path , and the end position of these 3 points is obtained to reproduce mandibular motion.

2 of the required guidances of mandibular motion are the condylar paths, the third point is measured at incisors known as incisal guidance.

Page 166: ARTICULATORs

166

Establishment of fixed base

Maxilla is fixed base from which mandibular motion is measured.

Face bow serve to transfer this relationship b/w maxilla and starting position of paths of mandibular movement.

Page 167: ARTICULATORs

167

Basic elements of mandibular motion

Described as pure or border movements of protrusive and right and left lateral excursions.

Protrusive excursion. Incisal guidance. Balancing condylar path. Balancing cusp inclines Working condylar motion bennett lateral shift.

Page 168: ARTICULATORs

168

ARBITRARY, POSITIONAL, SEMIADJUSTABLE ARTICULATORS

Articulators have been classified as arbitrary, positional, semi- adjustable, and fully adjustable.

ARBITRARY (Based on Monson Spherical Theory)

Examples: Monson Articulator, Hagman Balancer.

POSITIONAL ARTICULATOR ( Stansbery Tripod Concept)

Page 169: ARTICULATORs

169

The Stansbery tripod is a “positional” articulator. The objective of this concept is to obtain the static, or positional relationships of the mandible in centric relation, protrusive, and each lateral position

Page 170: ARTICULATORs

170

SEMIADJUSTABLE ARTICULATOR (Hanau Model H Concept)

The Hanau model H is one of the most commonly used semi- adjustable articulators. It was primarily designed for complete denture construction.

The objective of the technique is to closely produce mechanical equivalents of mandibular movements on the instrument.

Page 171: ARTICULATORs

171

RATIONALE FOR MATHEMATICAL STUDY:

The conditions of the evaluation are such that the occlusal errors produced by the clinical procedure identified with each articulator are cumulative rather than self- correcting.

An error can be defined as “deviation from the truth”, and truth as “agreement with reality”.

When applied to the study of articulators, an error produced in the occlusion will be considered as deviation from the known hypothetical

The second molar cusp height has been selected for the calculations because variations in condylar movement have a greater effect on the posterior teeth than on the anterior teeth.

Page 172: ARTICULATORs

172

When applied to the study of articulators, an error produced in the occlusion will be considered as deviation from the known hypothetical

The second molar cusp height has been selected for the calculations because variations in condylar movement have a greater effect on the posterior teeth than on the anterior teeth.

Page 173: ARTICULATORs

173

THE EFFECT OF THE CHANGE IN CONDYLAR INCLINATION ON OCCLUSION

Approximate error at the second molarBalancing cusp height (mm)

Approximate error at the second molarWorking cusp height (mm)

Average anatomic location of the hinge axis

0.2 0.2

Arbitrary location of the anterior point of orientation

0.2 No error

Straight condylar path 0.2 0.2

Fischer angle 0.1 No error

Individual working condylar motion

No error 0.8

Maximum total error 0.7 1.2

Page 174: ARTICULATORs

174

SELECTION OF AN ARTICULATOR

If occlusal contacts are to be perfected in centric occlusion only, a simple, sturdy, hinge-

type articulator without provision for lateral or

protrusive movements can be selected.

This type of instrument has been called one- dimensional

because only one interocclusal record is necessary for its

adjustment and use.

Page 175: ARTICULATORs

175

If denture teeth are to have a cross-arch and cross-tooth

balanced occlusion, the minimum requirement is a semi-adjustable

articulator.

This may be an instrument with individually adjustable condylar

guidances in both the vertical and horizontal plane, such as the

simple instruments in the hanau university series, the whip mix

articulator, or the dentatus articulator.

Page 176: ARTICULATORs

176

If more control of the occlusion is desired, a completely adjustable

three dimensional articulator is of value.

A three dimensional articulator requires a CR

record, at least two lateral records, and

some means for controlling the height and inclinations of the

cusps.

The means for their adjustment may be

interocclusal records or three-dimensional

graphic tracings made by a kinematic face-bow

apparatus.

Page 177: ARTICULATORs

177

The more complicated articulators pose some

problems in making complete dentures because of

the resiliency of the soft tissues of the basal seat

on which the recording bases must rest. Because

this resiliency permits some movement of the

bases in relation to the bone, the records made are

not necessarily records of the true path of

movement of the bone.

Page 178: ARTICULATORs

178

SUMMARY

The trend in articulators is being influenced primarily by prosthodontists working in the fields of fixed restorations and full mouth rehabilitation. They have suggested and devised instruments that are much more complicated than those traditionally used by dentists providing removable prosthodontic service.

There are a number of very fine and fully adjustable articulators, but most of these are not easily adapted for use for complete denture patients.

Page 179: ARTICULATORs

179

The reason for this lack of practical application for complete denture patients is the soft tissue foundation upon which the recording bases must rest. As long as the recording bases that support the recording instruments can move in relation to the underlying bone, the highly sensitive articulators cannot be accurately adjusted.

Page 180: ARTICULATORs

180

REFERENCES

PROSTHODONTIC TREATMENTFOR EDENTULOUS PATIENTS; TENTH EDITION

SHELDON WINKLER; ESSENTIALS OF COMPLETE DENTURE PROSTHODONTICS; SECOND EDITION

L.A. WEINBERG; AN EVALUATION OF BASIC ARTICULATORS AND THEIR CONCEPTS PART 1;JPD 1963; 13(4)

L.A. WEINBERG; AN EVALUATION OF BASIC ARTICULATORS AND THEIR CONCEPTS PART 2; JPD 1963; 13(4)

Page 181: ARTICULATORs

181

MITCHELL; ARTICULATORS THROUGH THE YEARS. PART 1. UPTO 1940; JPD; 1978; 39(3)

A. RAHINI; CLASSIFICATION OF ARTICULATORS;JPD 1980; 43(3)

LAWRENCE A WEINBERG: ARCON PRINCIPLE IN CONDYLAR MECHANISM OF ADJUSTABLE ARTIVULATORS; JPD 1963; 13(2)

NEAL D BELLANTI: SIGNIFICANCE OF ARTICULATOR CAPABILITIES; JPD 1973; 29(3).