RESULTS ABSTRACT Introduction: Accurate identification of uropathogens in a timely manner is important to correctly understand urinary tract infections (UTI’s), which affects nearly 150 million people each year. The current standard approach for detecting the UTI pathogens is culture based. This method is time consuming, has low throughput, and can lack sensitivity and/or specificity. In addition, not all uropathogens grow equally well under standard culture conditions which can result in a failure to detect the species. To address these gaps, we have developed a unique workflow from sample preparation to target identification using the nanofluidic OpenArray™ platform for spatial multiplexing of target specific assays. In this study, we tested pre-determined blinded research samples and confirmed the subset of results with orthogonal Sanger sequences. Methods: The in-house solution allows for the detection of 17 uropathogens including 16 bacterial and 1 fungal target. All assays have been verified with different sample types including synthetic plasmid control and ATCC gDNA samples for sensitivity and specificity testing. More than 120 pre-determined blinded samples from relevant sources were processed using the MagMAX DNA Multi-Sample Ultra Kit on the KingFisher Flex platform and screened by a nanofluidic qPCR platform using target specific TaqMan® pathogen detection assays(Table 1). Results: More than one pathogens were detected simultaneously in most of the samples using our research assays and nanofluidic platform. We observed greater than 98% concordance with the result generated at different site using the different OpenArray™ build with most of the assays similar to our panel. Results were highly reproducible between two different geographical sites. To confirm the accuracy of the OpenArray™ plate results we further investigated the subset of samples with orthogonal testing using capillary electrophoresis DNA sequencing. We observed 100% concordance for the sample tested. This further testing demonstrated that our workflow for UTI related pathogens detection is accurate. Conclusions: Based on these study results, we concluded that our application produced highly concordant results that are more sensitive and accurate. In summary, we have developed highly efficient, cost-effective research application for urinary tract microbiota pathogen profiling using high performance verified assay for each microorganism. CONCLUSIONS § All assays for urinary tract research pathogen detection were verified for sensitivity, specificity, and reproducibility using plasmids , target specific genomic DNA controls and repository cultured samples § OpenArray™ data was verified as having a high degree of accuracy by qPCR and sequencing. § OpenArray™ qPCR using urinary tract microbiota research assays are accurate in identifying multiple pathogens in a single sample § Our results demonstrate that multiplex profiling on Nanofluidic TaqMan™ OpenArray™ platform along with MagMAX™ sample prep is a successful method for detection of pathogens associated with human urinary tract microbiota imbalance. ACKNOWLEDGEME NTS We greatly appreciate Miguel Peñaranda and his team from Pathnostic for their support in providing samples and qPCR data(Site 2) TRADEMARKS For Resear ch Use Only. Not for use in diagnostic pr ocedur es. © 2018 Thermo Fisher S cientific Inc. A ll rights reserved. A ll trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified. TaqMan is a trademark of Roche Molecular S ystems, Inc., used under permission and license Novel Spatial Multiplex Screening of Uropathogens Associated with Urinary Tract Microbiota Research using a Nanofluidic qPCR Platform Thermo Fisher S cientific • 5781 V an A llen Way • Carlsbad, CA 92008 • thermofisher.com Sunali Patel, Kelly Li, Jisheng Li, Bonnie Moy, Boli Huang, Ioanna Pagani, Emily Zeringer, Nicole Fantin, Nitin Puri, Evan Diamond, Lienchi Nguyen, and Kamini Varma Thermo Fisher Scientific, 180 Oyster Point Blvd, South San Francisco, CA , USA, 94080 Table 1: Urinary Tract Research Pathogen Detection Assay Collection : The panel consist of 17 species specific assays which are closely associated with urinary tract infection including 13 Gram negative bacteria, 3 Gram positives bacteria and 1 fungal target. Organism Type Targe ts Gram Positive /Ne gative Y e ast Candidaalbicans Bacteria Acinetobacter baumannii Gram Ne gative Bacteria Citrobacter freundii Gram Ne gative Bacteria Klebsiellaaerogenes Gram Ne gative Bacteria Enterobacter cloacae Gram Ne gative Bacteria Enterococcus faecalis Gram Positive Bacteria Enterococcus faecium Gram Positive Bacteria Escherichia coli Gram Ne gative Bacteria Klebsiellaoxytoca Gram Ne gative Bacteria Klebsiellapneumoniae Gram Ne gative Bacteria Morganella morganii Gram Ne gative Bacteria Proteus mirabilis Gram Ne gative Bacteria Proteus vulgaris Gram Ne gative Bacteria Providenciastuartii Gram Ne gative Bacteria Pseudomonas aeruginosa Gram Ne gative Bacteria Staphylococcus saprophyticus Gram Positive Bacteria Stre ptococcus agalactiae Gram Positive Sample Preparation Liquid Handling for OpenArray Setup QuantStudio™ 12K Flex Real Time qPCR Detection Data Analysis Software <2 hrs 0.5 hrs 2 hrs 1 hrs Figure 1: CompleteWorkflow Solution: Sample to answer in less than 5.5 hours. Each OpenArray™ plate consists of 48 subarrays each containing 56 TaqMan™ assays spotted according to customer’s specifications. Up to 4 OpenArray™ plates can be included in a single QuantStudio™ 12K Flex run allowing for a throughput of 192 samples. Figure 2: Number of Sample Identified: Each urine samples underw ent nucleic acid extraction using MagMAX™ Kit and the extracted DNA samples were run us ing OpenArray™ plates at two different geogra phica l sites. qPCR results were highly reproducib le between two sites with > 98% concordance *E.coli assay was different in both OpenA rray™ panel Table 2: Multiplex Research Screening of Urinary Tract Microbiota on the OpenArray™ Platform. Subset of sample results for OpenArray™ are shown as presence(+) or as absence (blank cell) for respective uropathogens. The OpenArray results from 2 di fferent geographical sites were identica l for this subset of samples an d pathoge ns detected in b oth sites were highlighted i n green. Both sites were able to detect multiple pathogens includ ing fungal and bacteria l targets which were hard to culture Research Samples Assays for CE Sequencing Confirmed by both site qPCR Confirmed by Sanger Sequencing PUX17-000566 S.agalactiae YES YES PUX17-000574 K.pneumoniae YES YES P.mirabilis YES YES E.faecalis YES YES PUX17-000577 E.faecium YES YES C.albicans YES YES PUX17-000580 K.oxytoca YES YES K.pneumoniae NO NO E.coli YES YES PUX17-000764 E.faecalis YES YES S.agalactiae YES YES PUX17-000777 E.coli YES YES E.faecalis YES YES PUX17-000375 S.aureus YES YES Table 3: Orthogonal Testing of Samples by Sanger Sequencing to Confirm OpenArray™ Results: Sequencing was used to confirm results obtained usi ng OpenArray™ qPCR for 7 different samples tested. These results indicate that OpenArray™ qPCR can be used to accurately identify the correct pathogens. Sample Name C.albicans C.freundii E.coli E.faecalis K.oxytoca K.pneumo niae M.morgan ii P.mirabilis S.agalactia e PU X17-000345 + PU X17-000346 + + + PU X17-000348 + + PU X17-000453 + + + + PU X17-000452 + PU X17-000442 + + PU X17-000563 + + PU X17-000362 + + + PU X17-000562 + + PU X17-000349 PU X17-000363 + + PU X17-000365 + + PU X17-000451 + + + + PU X17-000574 + + + + + PU X17-000577 + + PU X17-000580 + + + PU X17-000764 + + + C .al b i can s C .fre u n di i E.co l i * E.fae cal i s K .o xyto c a K.pneum o n iae M .mo rga nii P .ae ru gi n o sa P .mi rab i l i s S.agal acti ae C .p arap si l o si s C .tro p i cal is Site 1 qPCR 3 1 48 38 7 7 5 4 5 12 1 1 Site 2 qPCR 3 1 46 38 6 7 5 4 5 13 1 1 0 5 10 15 20 25 30 35 40 45 50 # of samples Number of Samples Identified MATERIALS AND METHODS

Upload: others

Post on 16-Oct-2021

1 views

Category:

Documents


0 download

TRANSCRIPT

RESULTS ABSTRACT Introduction: Accurate identification of uropathogens in a timely manner is important to correctly understand urinary tract infections (UTI’s), which affects nearly 150 million people each year. The current standard approach for detecting the UTI pathogens is culture based. This method is time consuming, has low throughput, and can lack sensitivity and/or specificity. In addition, not all uropathogens grow equally well under standard culture conditions which can result in a failure to detect the species. To address these gaps, we have developed a unique workflow from sample preparation to target identification using the nanofluidic OpenArray™ platform for spatial multiplexing of target specific assays. In this study, we tested pre-determined blinded research samples and confirmed the subset of results with orthogonal Sanger sequences.

Methods: The in-house solution allows for the detection of 17 uropathogens including 16 bacterial and 1 fungal target. All assays have been verified with different sample types including synthetic plasmid control and ATCC gDNA samples for sensitivity and specificity testing. More than 120 pre-determined blinded samples from relevant sources were processed using the MagMAX DNA Multi-Sample Ultra Kit on the KingFisher Flex platform and screened by a nanofluidic qPCR platform using target specific TaqMan® pathogen detection assays(Table 1).

Results: More than one pathogens were detected simultaneously in most of the samples using our research assays and nanofluidic platform. We observed greater than 98% concordance with the result generated at different site using the different OpenArray™ build with most of the assays similar to our panel. Results were highly reproducible between two different geographical sites. To confirm the accuracy of the OpenArray™ plate results we further investigated the subset of samples with orthogonal testing using capillary electrophoresis DNA sequencing. We observed 100% concordance for the sample tested. This further testing demonstrated that our workflow for UTI related pathogens detection is accurate.

Conclusions: Based on these study results, we concluded that our application produced highly concordant results that are more sensitive and accurate. In summary, we have developed highly efficient, cost-effective research application for urinary tract microbiota pathogen profiling using high performance verified assay for each microorganism.

CONCLUSIONS § All assays for urinary tract research pathogen detection were

verified for sensitivity, specificity, and reproducibility using plasmids , target specific genomic DNA controls and repository cultured samples

§ OpenArray™ data was verified as having a high degree of accuracy by qPCR and sequencing.

§ OpenArray™ qPCR using urinary tract microbiota research assays are accurate in identifying multiple pathogens in a single sample

§ Our results demonstrate that multiplex profiling on Nanofluidic TaqMan™ OpenArray™ platform along with MagMAX™ sample prep is a successful method for detection of pathogens associated with human urinary tract microbiota imbalance.

ACKNOWLEDGEME NTSWe greatly appreciate Miguel Peñaranda and his team from Pathnostic for their support in providing samples and qPCR data(Site 2)

TRADEMARKSFor Research Use Only. Not for use in diagnostic procedures. © 2018 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified. TaqMan is a trademark of Roche Molecular Systems, Inc., used under permission and license

Novel Spatial Multiplex Screening of Uropathogens Associated with Urinary Tract Microbiota Research using a Nanofluidic qPCR Platform

Thermo Fisher Scientific • 5781 Van Allen Way • Carlsbad, CA 92008 • thermofisher.com

Sunali Patel, Kelly Li, Jisheng Li, Bonnie Moy, Boli Huang, Ioanna Pagani, Emily Zeringer, Nicole Fantin, Nitin Puri, Evan Diamond, Lienchi Nguyen, and Kamini VarmaThermo Fisher Scientific, 180 Oyster Point Blvd, South San Francisco, CA , USA, 94080

Table 1: Urinary Tract Research Pathogen Detection Assay Collection : The panel consist of 17 species specific assays which are closely associated with urinary tract infection including 13 Gram negative bacteria, 3 Gram positives bacteria and 1 fungal target.

Organism Type Targets Gram Positive/NegativeYeast CandidaalbicansBacteria Acinetobacterbaumannii Gram NegativeBacteria Citrobacterfreundii Gram NegativeBacteria Klebsiellaaerogenes Gram NegativeBacteria Enterobactercloacae Gram NegativeBacteria Enterococcusfaecalis Gram PositiveBacteria Enterococcusfaecium Gram PositiveBacteria Escherichia coli Gram NegativeBacteria Klebsiellaoxytoca Gram NegativeBacteria Klebsiellapneumoniae Gram NegativeBacteria Morganella morganii Gram NegativeBacteria Proteusmirabilis Gram NegativeBacteria Proteusvulgaris Gram NegativeBacteria Providenciastuartii Gram NegativeBacteria Pseudomonasaeruginosa Gram NegativeBacteria Staphylococcussaprophyticus Gram PositiveBacteria Streptococcusagalactiae Gram Positive

Sample Preparation

Liquid Handling for OpenArray Setup

QuantStudio™ 12K Flex Real Time qPCR Detection

Data Analysis Software

<2 hrs 0.5 hrs

2 hrs1 hrs

Figure 1: CompleteWorkflow Solution: Sample to answer in less than5.5 hours. Each OpenArray™ plate consists of 48 subarrays eachcontaining 56 TaqMan™ assays spotted according to customer’sspecifications. Up to 4 OpenArray™ plates can be included in a singleQuantStudio™ 12K Flex run allowing for a throughput of 192 samples.

Figure 2: Number of Sample Identified: Each urine samples underwentnucleic acid extraction using MagMAX™ Kit and the extracted DNAsamples were run us ing OpenArray™ plates at two different geographica lsites. qPCR results were highly reproducib le between two sites with >98%concordance*E.coli assay was different in both OpenArray™ panel

Table 2: Multiplex Research Screening of UrinaryTract Microbiotaonthe OpenArray™ Platform. Subset of sample results for OpenArray™are shown as presence(+) or as absence (blank cell) for respectiveuropathogens. The OpenArray results from 2 different geographical siteswere identica l for this subset of samples and pathogens detected in bothsites were highlighted in green. Both sites were able to detect multiplepathogens includ ing fungal and bacteria l targets which were hard toculture

Research Samples

Assays for CE Sequencing

Confirmed by both site qPCR

Confirmed by Sanger Sequencing

PUX17-000566 S.agalactiae YES YES

PUX17-000574K.pneumoniae YES YES

P.mirabilis YES YESE.faecalis YES YES

PUX17-000577 E.faecium YES YESC.albicans YES YES

PUX17-000580K.oxytoca YES YES

K.pneumoniae NO NOE.coli YES YES

PUX17-000764 E.faecalis YES YESS.agalactiae YES YES

PUX17-000777 E.coli YES YESE.faecalis YES YES

PUX17-000375 S.aureus YES YES

Table 3: Orthogonal Testing of Samples by Sanger Sequencing toConfirm OpenArray™ Results: Sequencing was used to confirmresults obtained using OpenArray™ qPCR for 7 different samplestested. These results indicate that OpenArray™ qPCR can be used toaccurately identify the correct pathogens.

SampleName C.albicans C.freundii E.coli E.faecalis K.oxytoca K.pneumoniae

M.morganii

P.mirabilis S.agalactiae

PUX17-000345 +

PUX17-000346 + + +

PUX17-000348 + +

PUX17-000453 + + + +

PUX17-000452 +

PUX17-000442 + +

PUX17-000563 + +

PUX17-000362 + + +

PUX17-000562 + +

PUX17-000349

PUX17-000363 + +

PUX17-000365 + +

PUX17-000451 + + + +

PUX17-000574 + + + + +

PUX17-000577 + +

PUX17-000580 + + +

PUX17-000764 + + +

C .alb icans

C .fre undii

E.co li* E.fae calisK .oxytoc

aK .pne umon iae

M.morgan ii

P .ae ruginosa

P .mirab il is

S.agalactiae

C .parapsilo sis

C .trop icalis

Site 1 qPCR 3 1 4 8 3 8 7 7 5 4 5 1 2 1 1

Site 2 qPCR 3 1 4 6 3 8 6 7 5 4 5 1 3 1 1

0

5

1 0

1 5

2 0

2 5

3 0

3 5

4 0

4 5

5 0

#ofsam

ples

NumberofSamplesIdentified

MATERIALS AND METHODS