Instruction Guide

VitaBreath

Know your Device

Charger Cover and PlugCharge Level Indicator (Green LED)On/Off ButtonStatus Indicator (White LED)Mouthpiece and CapCarry StrapAir Intake CoverInlet Screen

B

C

D

G

H

E

F

A

Device Features

Intended UseThe Philips VitaBreath dyspnea relief device is intended as an adjunct therapy for Chronic Obstructive Pulmonary Disease (COPD) patients to provide short-term relief of shortness of breath associated with physical exertion. It is for use in the home and hospital/institutional environments.

The device is designed for use by patients under the supervision of a physician.

AB

OU

T T

HE

DE

VIC

E

WA

RN

ING

S

WarningsA warning indicates the possibility of injury to the user or operator.

• This device is designed for use by patients under the supervision of a physician.

• This instruction guide serves as a reference. The instructions in this guide are not intended to supersede a health care professional’s instructions regarding use of the device.

• The VitaBreath is not intended to replace bronchodilators, anti-inflammatorymedicines,and/orsupplementaloxygen.Follow your doctor’s instruction for your currently prescribed treatmentregimenwhileusingVitaBreath.

• The operator should read and understand this entire guide before using the device.

• Neveroperatethedeviceifanypartsaredamagedorifitisnotworking properly.

• Donotopenordisassemblethedeviceexceptasinstructedbythisguide.RepairsandadjustmentsmustbeperformedbyPhilipsRespironics authorized service personnel only. Unauthorized service could cause injury, invalidate the warranty, or result incostlydevicedamage.

• Donotusethisdeviceoutsideoftheenvironmentalconditionsstatedinthisguide.Temperatureshigherthan40˚C(104˚F)orlowerthan0˚C(32˚F)couldcauseirritationorinjurytothe airway.

• Ifthereareanyunexplainedchangesintheperformanceofthedevice, discontinue use and contact Philips Respironics for service. Examplesinclude,butarenotlimitedto:unusualsounds,deviceisdropped, enclosure is cracked or broken, or water is spilled into the device enclosure.

• Whenchargingthedevice,routethepowercordfromthepowersupplyinawaythatwillpreventthecordfrombeingtrippedoveror interfered with by chairs or other furniture.

• Periodicallyinspectelectricalcordsandcablesfordamageorsigns of wear. Discontinue use and replace electrical cords andcablesifdamaged.

• Toavoidelectricalshock,alwaysunplugthepowersupplyfromthe wall outlet before cleaning the device.

• Donotplacethedeviceunderarunningfaucetorsubmergein liquid.

• When charging, always insure that the power supply is securely connected to both an AC outlet and the device’s power inlet to prevent arcing.

• OnlyusethepowersupplyrecommendedbyPhilipsRespironicswith this device. Use of a power supply not supplied by Philips Respironicsmaycauseoverheatingordamagetothedevice.

• Aproperlyinstalled,undamagedInletScreenandMouthpieceare required for proper operation. Wash prior to use and periodically as instructed by this guide and replace when damagedforproperoperation.Allowcleanedpartstodrycompletelybeforeusing.

• PortableandmobileRFcommunicationsequipmentcanaffectMedicalElectricalEquipment.

• MedicalElectricalEquipmentneedsspecialprecautionsregarding EMC and needs to be installed and put into service accordingtotheEMCinformationprovidedinthisguide.

• MedicalElectricalEquipmentmaybeinterferedwithbyotherequipment,evenifthatotherequipmentcomplieswithCISPREMISSIONrequirements.

• The use of a power supply or accessory other than those specifiedmayresultinincreasedemissionsordecreasedimmunityoftheMedicalElectricalEquipment.

• Thisproductcontainsalithiumionbattery.Itisimportanttodisposeofthisdeviceinasafemannerinaccordancewithlocal regulations.

• This device has not been tested for use on airplanes.• Donotusethedeviceinthepresenceofaflammableanaestheticmixtureincombinationwithoxygenorair,orinthepresenceofnitrousoxide.

• It can be unsafe to use accessories, detachable parts or othermaterialswiththisdevicethatarenotdescribedintheinstructions for use or are approved by Philips Respironics.

• This device is designed for non-continuous use only and shouldonlybeoperatedforlessthan10minutesatatime.Aftersuchtime,thedeviceshouldbeturnedoffforatleast30minutes.Ifitisnecessarytousethedevicelongerthan10minutes,consultwithaphysician.

CA

UT

ION

S, C

ON

TR

AIN

DIC

AT

ION

S, C

HA

RG

ING

TH

E D

EV

ICE

CautionsA caution indicates the possibility of damage to the device.

• Whenchargingornotusingthedevice,placeitonafirm,flatsurfacetopreventdamage.Do not place the device in a location whereitcanbeaccidentallydroppedfromanelevatedpositionsuchas the edge of a table or countertop.

• Donotstorethisdeviceoutsideoftheenvironmentalstoragespecificationsstatedinthisguide.

• Moistureandcondensationmaydamagethedevice.Ifthisdevicehasbeenexposedtoeitherveryhotorverycoldtemperatures,allowittoadjusttoroomtemperature(operatingtemperature)before starting. Do not operate the device outside of the operating temperaturerangeshowninthespecificationssection.

• Donotuseextensioncordswiththisdevice.• Do not place the device directly onto carpet, fabric, or other flammablematerials.

• WhenremovingtheInletScreenforcleaning,donotinsertanythinginto the interior of the device.

Shutdown• Press the power On/Off Button.• Cleanthemouthpiece(see“CleaningtheDevice”section).• Replace the cap.Note: The device will automatically shut down after 10 minutes of continuous use.Note: If the device is not used for 30 seconds, it will automatically turn off. U

SIN

G T

HE

DE

VIC

E

Before Use• Unplugthechargerfromdevice.• Removethecapfrommouthpiece.Note: Clean the mouthpiece before using (see “Cleaning the Device” section).Note: If the Charge Level Indicator (green LEDs) are flashing, let the device sit at room temperature for a few minutes. Then press the On/Off Button to activate the device.

How to reach Philips Respironics:www.philips.com/healthcarehealthcare@philips.com

ContactPhilipsRespironicsCustomerServiceat:Philips Respironics

1010MurryRidgeLane Murrysville,PA156681-724-387-4000

DeutschlandGewerbestrasse17

82211HerrschingGermany+49815293060

Pleasevisitwww.philips.com/healthcaretofindoutmore.

Respironics Inc.1001 Murry Ridge LaneMurrysville, PA 15668 USA

Respironics DeutschlandGewerbestrasse 1782211 Herrsching, Germany 1116762R00

JJB04/27/2015

Shield_CMYK_2013

Version 1.1 – 25 October 2013

Charging the DeviceRead and understand this Instruction Guide before using thedeviceforthefirsttime.

Charge the device battery before use.

• The Charge Level Indicator will cycle green LEDs while charging.

• Solidgreenbarsaredisplayedwhencompletelycharged.• Thenumberofsolidgreenbarswilldecreaseasthe

battery charge is depleted.

Note: Device functionality is disabled while charging.Note: The device will turn off 30 seconds after the battery is fully charged. The Charge Level Indicator will not stay lit after the charging is completed.

LED Battery Capacity4LEDsarelit* >75%3LEDsarelit 51-75%capacity2LEDsarelit 26-50%capacity1LEDislit 0-25%capacity1LEDblinks Low

*If4LEDsblink,thebatterytemperatureisoutofrange and the device will not operate. Let the device returntoroomtemperature.

When the battery is not charging the green LEDs will indicatethebatterycapacityasfollows:

Using the Device

Use

Before Use

ChargingB

C

D

GH

E F

A Use• Press the On/Off Button.

The Status Indicator will cycle (White LEDs) during power up. When the LEDs stop spinning, the device is ready for use.Note: When functioning, it is normal to hear the sound of the device as it is cycling.

• Seallipsaroundthemouthpiece.• Breathethroughthemouththroughthedeviceuntilrelief.Note: The device should be in hand positioned upright when in use.

Contraindications

• Recentpneumothorax• Recentbarotraumatothechestregion

If the patient has had any of the following conditions, consult their health care professional before using the device.

Note: Use only the power supply recommended by Philips Respironics with the following VitaBreath device model numbers.

Power Supply Reorder Numbers

1121743(Model1122023-Australia)

1116862(Model1122089-India)

1121742(Model1116847-UnitedKingdom)

Single Patient UseUseofabacteriafilterisoptional,butcanhelpmaintaindevicecleanliness.Acompatiblebacteriafilterandmouthpieceareavailable separately.

Caution: The duration of filter use will vary by patient and usage. Monitor for blockage by secretions or other particulate and replace as necessary.

Note: This filter is not intended to be disinfected or cleaned. Philips Respironics cannot guarantee the performance specifications once this filter has been disinfected or cleaned after patient use.

Multi-Patient UsePhilipsRespironicsPatientFilterKit(ReorderNumber1122088)mustbeusedwhenmultiplepatientsareusingthesamedevice. The Patient Filter Kit is sold separately. Contact PhilipsRespironicsfororderinginformation.

Thefollowingactivitiesmustbecompletedbetweenusers:• Replacebacteriafilterandmouthpiece.• Inspectthedeviceandallpartsfordamageorwear.• CleantheInletFilter(andperformadditionalcleaning

where needed).• Charge the device.Note: Refer to the Cleaning the Device section for instructions on how to clean the outside of the device.

Assembly• Removethedevicecapandmouthpiece.• ReplacethemouthpiecewiththePatientFilterKit,attachingthebacteriafilterbetweenthenewmouthpieceanddevice.

Note: A protective cap is provided with every Patient Filter Kit. The cap should be placed on the device when it is not being used.

Using the Patient Filter Kit

ACB

Assembly KeyA = MouthpieceB = Bacteria FilterC = Philips VitaBreath Device

CLE

AN

ING

TH

E D

EV

ICE

Respironics, Inc. warrants that the device and replaceable components(includingthemouthpieceassembly,inletscreen,andcap)shallbefreefromdefectsinworkmanshipandmaterialsandwillperforminaccordancewiththeproductspecificationsforaperiodoftwo(2)yearsforthedeviceandsixty(60)daysforthereplaceablecomponentsfromthedateofsalebyRespironics,Inc.tothecustomer.Iftheproductfailstoperforminaccordancewiththeproductspecifications,Respironics,Inc.willrepairorreplace,atitsoption,thedefectivematerialorpart.Thiswarrantydoesnotcoverdamagecausedbyaccident,misuse,abuse,alterationandotherdefectsnotrelatedtomaterialorworkmanship.

RESPIRONICS, INC. DISCLAIMS ALL LIABILITY FOR INDIRECT, SPECIAL,PUNITIVEORCONSEQUENTIALDAMAGESWHICHMAYBECLAIMEDTOARISEFROMANYUSEOFTHISPRODUCT.SOMESTATESDONOTALLOWTHEEXCLUSIONOR LIMITATION OF INCIDENTAL OR CONSEQUENTIAL DAMAGES,SOTHEABOVELIMITATIONOREXCLUSIONMAYNOT APPLY TO YOU.

THISWARRANTYISGIVENINLIEUOFALLOTHEREXPRESSORIMPLIEDWARRANTIES,INCLUDING,WITHOUTLIMITATION, ANY WARRANTY OF NON-INFRINGEMENT, MERCHANTABILITYORFITNESSFORAPARTICULARPURPOSE.SOME STATES DO NOT ALLOW LIMITATIONS ON IMPLIED WARRANTIES,SOTHEABOVELIMITATIONMAYNOTAPPLYTOYOU.THISWARRANTYGIVESYOUSPECIFICLEGALRIGHTS,ANDYOUMAYALSOHAVEOTHERRIGHTSWHICHVARY FROM STATE TO STATE.

Toexerciseyourrightsunderthiswarranty,contactRespironics,Inc.at:

1001 Murry Ridge LaneMurrysville, Pennsylvania

15668-85501-724-387-4000

Deutschland Gewerbestrasse 17

82211 Herrsching Germany+49 8152 93060

Note: For Australian and New Zealand customers this warranty replaces the warranty contained above.

1.Respironics,Inc.,aPhilipsHealthcarecompanywarrantsthattheProductsshallbefreefromdefectsofworkmanshipandmaterialsandwillperforminaccordancewiththeProductspecifications.2.Thiswarrantyisvalidforaperiodoftwo(2)yearsforthedeviceandsixty(60)daysforthereplaceablecomponentsfromthedateofpurchasefromanauthorisedRespironicsdealer.3.IftheProductisfoundtocontainadefectofworkmanshipormaterialsorfailstoperforminaccordancewiththeProductspecificationsduring the applicable warranty period, Respironics will repair orreplace,atitsoption,thedefectivematerialorpart.

LimitedWarranty

Warning: Do not place the device under a running faucet or submerge in liquid.

Inlet Screen TheInletScreenshouldbecleanedeverymonth.

• RemovetheAirIntakeCover.

• Loosen the screw holding the Inlet Screen until the screencanberemoved.

Note: The screw is designed to be loosened, but not removed.

• InspecttheInletScreenfordamageorwear.Discardand replace if necessary.

• HandwashtheInletScreen.Thefollowingcleaningagentsmaybeused:

• Mild Detergent• Water• 70%IsopropylAlcohol

• Rinsethoroughlyandairdryawayfromdirectsunlight.

• Reattach the Inlet Screen and retighten the screw.

• Replace Air Intake Cover.

Note: The Inlet Screen should be cleaned before first use and daily.

Transport and StorageWhen transporting or storing the device, these steps will preventdebrisfromenteringthedevice.

• Replacethecapoverthemouthpiece.

• Make certain the charger cover is in place.

Note: Refer to the “Specifications” section for environmental information regarding storage and operation.

EnvironmentalOperating Storage

Temperature 0°Cto40°C(32°Fto104°F)

-25°Cto60°C(-13°Fto140°F)

Relative Humidity

15to95%(non-condensing)

15to95%(non-condensing)

Atmospheric Pressure

101kPato57kPa(approximately0-15,000feet)

N/A

ElectricalAC/DCPowerSupply:100to240VACat50to60HzLi-ionRechargeableBattery:

Voltage:14.8VDCCurrent:2.3AhPower:34Wh

TypeofProtectionAgainstElectricShock:ClassIIDegreeProtectionAgainstElectricShock:

Type BF Applied PartDegreeProtectionAgainstIngressofWater:IP22ModeofOperation:Non-continuous

Note: AC/DC Power Supply is used for charging battery only.

Maximum Activation (ON) Time

Minimum Deactivation (OFF) Time

Duty Cycle <10minutes ≥30minutes

ExpectedServiceLife• HandheldRecoveryDevice(includingpowersupplyandstrap):3years

• Mouthpiece(includingcap):6months• InletScreen:6months

DisposalSeparatecollectionforelectricalandelectronicequipmentperECDirective2012/19/EU.Disposeofthisdeviceinaccordance with local regulations.

Specifications EMCInformationGuidance and Manufacturer’s Declaration - Electromagnetic EmissionsThisdeviceintherapymodewhileoperatingfromarechargeablebatteryisintendedforuseintheelectromagneticenvironmentspecifiedbelow.Theuserofthisdeviceshouldassureitisusedinsuchanenvironment.

Emissions Test Compliance Electromagnetic Environment - Guidance

RFemissions

CISPR11

Group1 The device uses RF energy only for its internal function. Therefore,itsRFemissionsareverylowandarenotlikelytocauseanyinterferenceinnearbyelectronicequipment.

RFemissions

CISPR11

Class B Thedeviceissuitableforuseinallestablishments,includingdomesticestablishmentsandthosedirectlyconnectedtothepublic low-voltage power supply network that supplies building usedfordomesticpurpose.HarmonicEmissions

IEC61000-3-2*

Class A

Voltage Fluctuations /FlickerEmissions

IEC61000-3-3*

Complies

*TestrequirementareapplicabletoAC-DCadapteronly.ConformitytoteststandardsisshowninOEMEMCtest reports.

Guidance and Manufacturer’s Declaration - Electromagnetic ImmunityThisdeviceisintendedforuseintheelectromagneticenvironmentspecifiedbelow.Theuserofthisdeviceshouldassureitisusedinsuchanenvironment.

Immunity Test IEC 60601 Test Level

Compliance Level

Electromagnetic Environment -

GuidanceElectrostatic Discharge (ESD)

IEC61000-4-2

±6kVcontact

±8kVair

±6kVcontact

±8kVair

Floors should be wood, concrete or ceramictile.Iffloorsarecoveredwithsyntheticmaterial,therelativehumidityshouldbeatleast30%.

Electrical Fast Transient / Burst

IEC61000-4-4*

±2kVforPowerSupply Lines

±1kVforInput/Output Lines

N/A for Battery Operated Device

N/A, Device does not have user I/O Lines that are longer than3minlength.

Mains power quality should be that of a typicalhomeorhospitalenvironment.

Surge

IEC61000-4-5*

±1kVLinetoLine

±2kVLinetoGround

N/A for Battery Operated Device

N/A for Battery Operated Device

Mains power quality should be that of a typicalhomeorhospitalenvironment.

Voltage Dips, Short Interruptions and Voltage Variations on Power Supply Input Lines

IEC61000-4-11*

<5%UT(>95%dipinUT) for 0.5cycle40%UT(60%dipinUT) for 5cycles70%UT(30%dipinUT)for25cycles <5%UT(>95%dipin UT)for5sec

N/A for Battery Operated Device

Mains power quality should be that of a typicalhomeorhospitalenvironment.If the user of the device requires continued operation during power mainsinterruptions,itisrecommendedthatthedevicebepoweredfromanuninterruptible power supply or a battery.

Power frequency (50/60Hz)magneticfield

IEC61000-4-8

3A/m 3A/m Powerfrequencymagneticfieldsshouldbe at levels characteristic of a typical homeorhospitalenvironment.

NOTE:UTisthea.c.mainsvoltagepriortoapplicationofthetestlevel.*TestrequirementareapplicabletoAC-DCadapteronly.ConformitytoteststandardsisshowninOEMEMCtest reports.

Guidance and Manufacturer’s Declaration - Electromagnetic ImmunityThisdeviceisintendedforuseintheelectromagneticenvironmentspecifiedbelow.Theuserofthisdeviceshouldmakesureitisusedinsuchanenvironment.

Immunity Test

IEC 60601 Test Level

Compliance Level

Electromagnetic Environment -Guidance

Conducted RFIEC61000-4-6*

Radiated RFIEC61000-4-3

3Vrms150kHzto80MHz

3V/m80MHzto2.5GHz

3Vrms

3V/m

PortableandmobileRFcommunicationsequipmentshouldbeusednoclosertoanypart of the device, including cables, than the recommendedseparationdistancecalculatedfromtheequationapplicabletothefrequencyofthetransmitter.Recommendedseparationdistance:d=1.2√Pd=1.2√P80MHzto800MHzd=2.3√P 800MHzto2.5GHzwherePisthemaximumoutputpowerratingofthetransmitterinwatts(W)accordingtothetransmittermanufactureranddistherecommendedseparationdistanceinmeters(m).

FieldstrengthsfromfixedRFtransmitters,asdeterminedbyanelectromagneticsitesurveya, shouldbelessthanthecompliancelevelineachfrequency range.b

Interferencemayoccurinthevicinityofequipmentmarkedwiththefollowingsymbol:

NOTE1At80MHzand800MHz,thehigherfrequencyrangeapplies.NOTE2Theseguidelinesmaynotapplyinallsituations.Electromagneticpropagationisaffectedbyabsorptionandreflectionfromstructures,objects,andpeople.a Fieldstrengthsfromfixedtransmitters,suchasbasestationsforradio(cellular/cordless)

telephonesandlandmobileradios,amateurradio,AMandFMradiobroadcastandTVbroadcastcannotbepredictedtheoreticallywithaccuracy.ToassesstheelectromagneticenvironmentduetofixedRFtransmitters,anelectromagneticsitesurveyshouldbeconsidered.IfthemeasuredfieldstrengthinthelocationinwhichthedeviceisusedexceedstheapplicableRFcompliancelevelabove,thedeviceshouldbeobservedtoverifynormaloperation.Ifabnormalperformanceisobserved,additionalmeasuresmaybenecessary,suchasre-orientingorrelocatingthedevice.

b Overthefrequencyrange150kHzto80MHz,thefieldstrengthsshouldbelessthan3V/m.*TestrequirementareapplicabletoAC-DCadapteronly.ConformitytoteststandardsisshowninOEM EMC test reports.

Recommended Separation Distances between Portable and Mobile RF Communications Equipment and This DeviceThedeviceisintendedforuseinanelectromagneticenvironmentinwhichradiatedRFdisturbancesarecontrolled.ThecustomerortheuserofthisdevicecanhelppreventelectromagneticinterferencebymaintainingaminimumdistancebetweenportableandmobileRFcommunicationsequipment(transmitters)andthisdeviceasrecommendedbelow,accordingtothemaximumoutputpowerofthecommunicationsequipment.

Rated Maximum

Power Output of Transmitter

(Watts)

Separation Distance According to Frequency of Transmitter

(meters)150kHzto80MHz

d=1.2√P80MHzto800MHz

d=1.2√P800MHzto2.5GHz

d=2.3√P

0.01 0.12 0.12 0.23

0.1 0.38 0.38 0.73

1 1.2 1.2 2.3

10 3.8 3.8 7.27

100 12 12 23

Fortransmittersratedatamaximumoutputpowernotlistedabove,therecommendedseparationdistancedinmeters(m)canbeestimatedusingtheequationapplicabletothefrequencyofthetransmitter,wherePisthemaximumoutputpowerofthetransmittermanufacturer.NOTE1At80MHzand800MHz,thehigherfrequencyrangeapplies.NOTE2Theseguidelinesmaynotapplyinallsituations.Electromagneticpropagationisaffectedbyabsorptionandreflectionfromstructures,objectsandpeople.

TR

AN

SPO

RT A

ND

ST

OR

AG

E, T

RO

UB

LESH

OO

TIN

G, T

IPS

FO

R U

SE

SYM

BO

LS,

SPE

CIF

ICA

TIO

NS

EM

C I

NFO

RM

AT

ION

Warning: Do not insert anything into the interior of the device when the Inlet Screen is removed.

ExternalSurfaceTheexternalsurfacemaybecleanedasneeded.

• Use a cloth to wipe down the enclosure. The following cleaningagentsmaybeused:

• Mild Detergent• Water• 70%IsopropylAlcohol• DisCide Towelettes• 10%BleachSolution

• Allowthedevicetodrycompletelybeforeuse.

Carry Strap Inlet Screen

4.Thecustomerisresponsibleforreturningtheproduct to an authorised Philips Respironics dealer, andcollectingtheproductfromtheauthorisedPhilipsRespironicsdealerafterrepairorreplacement,atits own cost. Philips Respironics is responsible only for the freight cost of transporting the product between the authorised Philips Respironics dealer and Respironics. Respironics reserves the right to charge an evaluation and postage fee for any returned Productastowhichnoproblemisfoundfollowinginvestigation.5.Thiswarrantydoesnotcoverdamagecausedbyaccident,misuse,abuse,alteration,andotherdefectsnotrelatedtomaterialsorworkmanship.6.ThewarrantyprovidedbyRespironicshereinisnottransferrable by the Buyer in the event of any sale ortransferofProductspurchasedbytheBuyerfromanauthorisedRespironicsdealer.7.Totheextentpermittedbylaw,wheretheBuyerhasthebenefitofanimpliedguaranteeundertheAustralianConsumerLaw, but the Product is not of a kind ordinarily acquiredforpersonal,domesticorhouseholduseorconsumptionRespironics’liabilityshallbelimited,attheoptionofRespironics,tothereplacementorrepairof the Product or the supply of an equivalent Product. 8.Toexerciseyourrightsunderthiswarranty,contactyour local authorised Philips Respironics dealer. A list of all authorised dealers is available at the following link:http://www.philips.com.au/healthcare.

Alternatively,youcanmakeaclaimunderthiswarrantybycontactingRespironicsdirectlyat:PhilipsElectronicsAustraliaLimited,65EppingRoad,NorthRydeNSW2113,Australia.Tel:1300766488,Email:prcontact@philips.com.9.ThefollowingstatementisprovidedtoaBuyerwhoisa“consumer”undertheAustralianConsumerLaw:Our goods come with guarantees that cannot be excluded under the Australian Consumer Law. You are entitled to a replacement or refund for a major failure and for compensation for any other reasonably foreseeable loss or damage. You are also entitled to have the good repaired or replaced if the goods fail to be of acceptable quality and the failure does not amount to a major failure.10.ThefollowingstatementisprovidedtoaBuyerwhoisa“consumer”undertheConsumerGuaranteesAct1993,NewZealand:Our goods come with guarantees that cannot be excluded under the Consumer Guarantees Act 1993. This guarantee applies in addition to the conditions and guarantees implied by that legislation.

Mouthpiece and CapAssurethedeviceisunpluggedfromthechargerbeforecompletingthefollowingsteps.

• Removethemouthpieceandcap.

• Handwashthemouthpieceandcap.The following cleaning agentsmaybeused:

• Mild Detergent• Water• 70%IsopropylAlcohol

• Rinse thoroughly and airdryawayfromdirect sunlight.

Note: The Mouthpiece and Cap should be cleaned before the first use and after each use.

Aftercleaning,inspectthedeviceandallpartsfordamageor wear (cracking, crazing, tears, punctures, etc.).

Carry StrapTheCarryStrapmaybecleanedasneeded.

• RemovetheAirIntakeCover.

• Slide strap out of its port.

• Thefollowingcleaningagentsmaybeused:

• Mild Detergent• Water• 70%IsopropylAlcohol

• Airdrythestrap.Allowthestraptodrycompletelybefore replacing.

Cleanroutinelytoextendthelifeofthedevice.

Mouthpiece and Cap

Cleaning the Device

Tips for Use• Charge the device battery before use.

Note: Device functionality is disabled while charging.

• Practice using the device.

• Inhalethroughthedevicemouthpiece.

• Exhalethroughthemouthpiece.

• Becomecomfortablewithusingthedevice.

Note: Breathe through the mouth as breathing through the nose decreases the benefit.

Note: Contact Philips Respironics if assistance is needed to setup, use, or maintain this device.

TroubleshootingThe following tips will help to assure that your device will provide uninterupted therapy.

• Assure the connector end of the power supply is properly attached to the charger plug.

• If the Charge Level Indicator continues to cycle, the device is not fully charged and needs additional charg-ingtime.

• AssuretheInletscreen,mouthpieceandcap,arerou-tinely cleaned and properly installed.

MaintenanceThere are no user servicable parts in this device. Service mustbeperformedbyPhilipsRespironicsauthorizedservice personnel only.

SymbolsReorderNumber

DegreeofProtectionagainstIngressofWater:DripProof,IP22

SerialNumber

AuthorizedrepresentativeintheEuropeanCommunity.

Manufacturer

Date of Manufacture

Read the instructions for use.

Type BF Applied Part

DC Power

Li-ion Battery Recycling

CE Marking for Class II products

CompliantwiththeWasteElectricalandElectronicEquipment/RestrictionoftheUseofElectronicEquipment(WEEE/RoHS)recyclingdirectives