ASSESSMENT OF SUPPORT SURFACES (STATIC & MIXED) FOR NEONATES. FP-50002 EWMA/GNEAUPP 2014. M. M. Sánchez-Lorente 1 ; P. García-Molina 2-3 ; C. Vergara-Hernández 2 ; E. Sanchis-Sánchez 2 ; J. Martos-Torres 2 ; A. Álvarez-Ordiales 4 ; & E. Balaguer-López 2-3. - PowerPoint PPT Presentation

TRANSCRIPT

ASSESSMENT OF SUPPORT SURFACES (STATIC & MIXED)

FOR NEONATESM. M. Sánchez-Lorente1; P. García-Molina2-3; C. Vergara-Hernández2;

E. Sanchis-Sánchez2; J. Martos-Torres2; A. Álvarez-Ordiales4; & E. Balaguer-López2-3

FP-5

0002

EW

MA/

GNEA

UPP

201

4

1. Conselleria de Sanitat, Valencia (SPA).

2. University of Valencia, Valencia (SPA).

3. Hospital Clínico Universitario, Valencia (SPA).

4. Hospital Universitari i Politècnic “La Fe”, Valencia (SPA).

Presenting author: Mr. Pablo García-Molina

Background

FP-5

0002

EW

MA/

GNEA

UPP

201

4

Existence of several neonatal support surfaces

Lack of evaluation

Interface pressure & pressure redistribution Efficacy in the prevention or treatment of PU

however…

regarding…

Aim

FP-5

0002

EW

MA/

GNEA

UPP

201

4

• To assess the interface pressure and pressure redistribution of five devices of constant low pressure (static and mixed support surfaces).

152

43

Methods

FP-5

0002

EW

MA/

GNEA

UPP

201

4

Study design• Laboratory study with repeated measures design.

Sample• Eight neonatal simulation dummies with different characteristics

regarding head circumference, weight and height.

Methods

FP-5

0002

EW

MA/

GNEA

UPP

201

4

Support devicesVisc-T Visc-White Visc-Gray Visc-Blue Pol-3D

Structure Viscoelastic“T” units

ViscoelasticBlock

ViscoelasticBlock

ViscoelasticBlock

Polyester filaments

Layers 1 2 2 2 1

Density (Kg/m3) 50 50/50 50/50 40 NA

Hardness (Kpa) 1.6 1.01/5 1.01/3.6 2.5 NA

Compression factor 2 2/2 2/2 - NA

Methods

FP-5

0002

EW

MA/

GNEA

UPP

201

4

Variables• Mean pressure (mmHg).• Maximum pressure (mmHg).• Mean pressure in head region (mmHg).• Area of contact (cm2).

Equipment• XSENSOR X3 PX100:40.40.002 device.

Procedure• See: E-poster no. 50006.

Methods

FP-5

0002

EW

MA/

GNEA

UPP

201

4

Statistical analysis• Descriptive statistics.

• Due to a small sample size, a non-parametric test (Friedman test) was carried out to assess differences among groups. If a significant difference was found, post hoc comparisons among groups were performed using the Wilcoxon-Nemenyi-McDonald-Thompson test.

• The analysis was carried out by means of the R environment (The R Foundation for Statistical Computing).

• Data set and code employed for the analysis can be found at www.figshare.com (*).*http://dx.doi.org/10.6084/m9.figshare.977866

Results

FP-5

0002

EW

MA/

GNEA

UPP

201

4

Visc-T Visc-White Visc-Gray Visc-Blue Pol-3D

Mean pres., x̅ (SD) 16.26 (2.77) 17.51 (3.32) 18.12 (3.32) 17.11 (3.07) 17.82 (2.58)

Max. pres., x̅ (SD) 24.58 (7.39) 26.79 (8.88) 28.15 (7.91) 25.49 (7.55) 27.53 (6.12)

Head pres., x̅ (SD) 17.25 (3.37) 18.58 (4.17) 19.53 (4.45) 17.95 (3.71) 18.7 (2.92)

Area, x̅ (SD) 19.29 (11.29) 22.11 (12.13) 22 (9.75) 23.52 (13.21) 29.23 (12.46)

Lowest values!

Highest values!

Results

FP-5

0002

EW

MA/

GNEA

UPP

201

4

Friedman test χ2 Df P-value

Mean pres., x̅ (SD) 14,3 4 0,006

Max. pres., x̅ (SD) 11,2 4 0,02

Head pres., x̅ (SD) 8,7 4 0,07

Area, x̅ (SD) 18,51 4 < 0,001

Results

FP-5

0002

EW

MA/

GNEA

UPP

201

4

Row-Column Visc-T Visc-White Visc-Gray Visc-Blue Pol-3D

Visc-T - - - - -

Visc-White 0,24 - - - -

Visc-Gray 0,005 0,61 - - -

Visc-Blue 0,71 0,93 0,17 - -

Pol-3D 0,056 0,97 0,93 0,61 -

Post hoc analysis: mean pressure (p-value)

Results

Results

FP-5

0002

EW

MA/

GNEA

UPP

201

4

Row-Column Visc-T Visc-White Visc-Gray Visc-Blue Pol-3D

Visc-T - - - - -

Visc-White 0,32 - - - -

Visc-Gray 0,014 0,71 - - -

Visc-Blue 0,88 0,88 0,17 - -

Pol-3D 0,32 1 0,71 0,88 -

Post hoc analysis: maximum pressure (p-value)

Results

Results

FP-5

0002

EW

MA/

GNEA

UPP

201

4

Row-Column Visc-T Visc-White Visc-Gray Visc-Blue Pol-3D

Visc-T - - - - -

Visc-White 0,71 - - - -

Visc-Gray 0,95 0,98 - - -

Visc-Blue 0,27 0,95 0,71 - -

Pol-3D < 0,001 0,053 0,01 0,27 -

Post hoc analysis: pressure area (p-value)

Results

Conclusion

The “T” device presents the best values of pressure relief.

The polyester fiber device presents the best values of pressure redistribution.

Adding an upper layer of polyester fiber to the “T” device could result in a better support surface.

Next step: studying the properties of the new support surface in healthy subjects under clinical conditions.