ACCURATE WORKING LENGTH DETERMINATION

Solaiman Al-Hadlaq B.D.S., M.S., Ph.D

Page 3: Accurate Working Length Detemination Final en Do Club

Where to end your root canal preparation?

“ The proper point to which root canals should be filled is the junction of the dentin and the cementum and that the pulp should be severed at the point of its union with the periodontal membrane”

Groove C., JADA 1930

Page 4: Accurate Working Length Detemination Final en Do Club

Methods of Working Length Determination

• Predetermined normal tooth length

• Radiographs

• Tactile sense

• Paper points

• Patient response

• Electronic Apex Locators

Page 5: Accurate Working Length Detemination Final en Do Club

1. Tooth apex (radiographic apex)

2. Apical foramen (major foramen)

3. Apical constriction (minor foramen)

12

3

Anatomy of the Root Apex

Page 6: Accurate Working Length Detemination Final en Do Club

12

3

Johnson WT, Color atlas of Endodontics, 2002

Anatomy of the Root Apex

Page 7: Accurate Working Length Detemination Final en Do Club

1. Tooth apex (radiographic apex)

2. Apical foramen (major foramen)

3. Apical constriction (minor foramen)

1

2

3

Anatomy of the Root Apex

Page 8: Accurate Working Length Detemination Final en Do Club

1

2

3

• Distance between 1 and 2:

The apical foramen deviates from the apex in 50-98% of the teeth.

This deviation averages 0.3 to 0.6 mm but could be as much as 3 mm.

Kuttler, JADA, 1955; Green, OOO, 1960; Pineda and Kuttler, Oral Surg, 1972; Burch and Hulen, Oral Surg, 1972; Dummer et al., Int Endo J, 1984

1. Tooth apex (radiographic apex)

2. Apical foramen (major foramen)

3. Apical constriction (minor foramen)

Anatomy of the Root Apex

Page 9: Accurate Working Length Detemination Final en Do Club

1

2

3

• Distance between 2 and 3: 0.5 mm in 18-25 y old, and 0.7

in 55+ y old.Kuttler, JADA,

1955,

• Distance between 1 and 3: 0.89 mm with a range of 0.1 to

2.7 mm. Dummer et al., Int Endo J,

1984

1. Tooth apex (radiographic apex)

2. Apical foramen (major foramen)

3. Apical constriction (minor foramen)

Anatomy of the Root Apex

Page 10: Accurate Working Length Detemination Final en Do Club

Anatomy of the Root Apex

Johnson WT, Color atlas of Endodontics, 2002

Page 11: Accurate Working Length Detemination Final en Do Club

Johnson WT, Color atlas of Endodontics, 2002

Anatomy of the Root Apex

Page 12: Accurate Working Length Detemination Final en Do Club

Anatomy of the Root Apex

Johnson WT, Color atlas of Endodontics, 2002

Page 13: Accurate Working Length Detemination Final en Do Club

46 % 30 %

19 % 5 %

Dummer et al., Int Endo J, 1984

Page 14: Accurate Working Length Detemination Final en Do Club

Bergenholtz G et al., Textboook of Endodontology, 2002

Page 15: Accurate Working Length Detemination Final en Do Club

Bergenholtz G et al., Textboook of Endodontology, 2002

Page 16: Accurate Working Length Detemination Final en Do Club
Page 17: Accurate Working Length Detemination Final en Do Club

• When radiographic working length in vitro was set to be 0.5 – 2.0 mm short of the apex: 25% of the time the file was beyond the apical

foramen. 51% of premolars and 22 % of molars were

beyond the apical foramen. 66% of files were longer than a point 0.5 mm

shorter than the apical foramen.

ElAyouti A et al., J Endodon, 2001

Page 18: Accurate Working Length Detemination Final en Do Club

Walton and Torabinejad., Principles and practice of Endodontics, 1996

Page 19: Accurate Working Length Detemination Final en Do Club
Page 20: Accurate Working Length Detemination Final en Do Club
Page 21: Accurate Working Length Detemination Final en Do Club

Electronic Apex Locators

• The idea of using electrical conductance to measure root length was suggested first by by Cluster in 1918.

Cluster LE, J Natl Dent Assoc, 1918

Page 22: Accurate Working Length Detemination Final en Do Club

Electronic Apex Locators

• In 1942 Suzuki noticed a constant value of electrical resistance between an instrument in the root canal and an electrode on the oral mucous membrane and speculated that this would measure canal length.

Suzuki K, J Jpn Stomatol, 1942

Page 23: Accurate Working Length Detemination Final en Do Club

Electronic Apex Locators

• In 1962 Sunada took the principle of Suzuki and constructed a simple device that used direct current to measure canal length. This was based on the concept that the resistance between the periodontium and the mucous membrane was constant and equal 6 k .

Sunada I, J Dent Res, 1962

Page 24: Accurate Working Length Detemination Final en Do Club

Electronic Apex Locators

First generation• Resistance type based on the principle of Suzuki

and Sunada.• Root canal meter, Endodontic meter, Dentometer,

and Endo Radar.• Pain was felt during earlier models due to high

currents.• Unreliable, electrolytes, exudate, hemorrhage,

vital pulp tissue, and excessive moisture caused inaccurate results.

Page 25: Accurate Working Length Detemination Final en Do Club

Electronic Apex Locators

Second generation• Single frequency impedance.• Highest impedance at the apical constriction.• Need to be calibrated before each use.• Sono-Explorer, Endocator (sheath over probe).• Incorrect readings with electrolytes in the canal.

Page 26: Accurate Working Length Detemination Final en Do Club

Electronic Apex Locators

Third generation

• Multiple frequencies to measure impedance.

• Calculate difference or ratio of impedance.

Page 27: Accurate Working Length Detemination Final en Do Club

Electronic Apex Locators

Third generation

• Endex (Osada Electric, Tokyo, Japan) measures the difference between two impdances.

• It works in the presence of pus and electrolytes.

• Needs to be calibrated for each canal.

Page 28: Accurate Working Length Detemination Final en Do Club

Electronic Apex Locators

Third generation

• Endex has an average accuracy of 81% to within + 0.5 mm of the apical foramen.

Investigator Accuracy (%) Test condition Compared with

Fouad et al. (1993) 73 (±0.5 mm) In vitro - NaOCI Tooth lengthMayeda et al. (1993) 66 (±0.75 mm) In vivo Tooth lengthFrank & Torabinejad (1993) 90 (±0.5 mm) In vivo RMFelippe 8< Scares (1994) 96.5 (±0.5 mm) In vitro Tooth lengthArora & Gulabivala (1995) 72 (±0.5 mm) In vivo Tooth lengthPratten & McDonald (1996) 82 (±0.5 mm) In vitro RM and tooth lengthLauper etal. (1996) 93 (±0.5 mm) In vivo Tooth lengthOunsi & Haddad (1998) 85 (±0.5 mm) In vitro Tooth lengthWeiger et al. (1999) 59 (±0.5 mm) In vitro - NaOCI Tooth lengthDe Moor et al. (1999) 100 (±0.5 mm) In vitro Tooth lengthMartinez-Lozano et al. (2001) 68 (±0.5 mm) In vitro Tooth length  Accuracy studies for the Endex/ Apit apex locator (Gordon and Chandler, 2004)

Page 29: Accurate Working Length Detemination Final en Do Club

Electronic Apex Locators

Third generation

• Root ZX (J. Morita, Tokyo, Japan) measures the ratio of two impedance frequencies.

• Self calibrating.

• Was not affected by electrolytes in the canal.

Page 30: Accurate Working Length Detemination Final en Do Club

Investigator Variable tested Accuracy (%) Compared with Sample (n)

Clinical accuracy permanent teeth in vitroCzerw et al. (1995) Accuracy in vitro 100 (±0.5 mm) Tooth length 30White et al. (1996) Accuracy in vitro 84 (±0.5 mm) Tooth length 51 Ounsi & Naaman (1999) Accuracy in vitro 85 (±0.5 mm) Tooth length 39 Accuracy in the presence of irrigantsShabahang et al. (1996) Accuracy in vivo 96 (±0.5 mm) Extracted tooth length 26 McGinty et al. (1996) Irrigants and accuracy in vitro No difference Tooth length 16

between irrigantsWeiger et al, (1999) Irrigants and accuracy in vitro - NaOCI 85 (±0.5 mm) Tooth length 41 Jenkins era/. (2001) Various irrigants and accuracy in vitro No difference Tooth length 30 Meares & Steiman (2002) Accuracy with NaOCI in vitro 83 (±0.5 mm) Tooth length 40

No difference Clinical accuracy permanent teeth in vivoVajrabhaya & Accuracy in vivo 100 (±0.5 mm) Extracted tooth length 20 Tepmongkol (1997) Pagavino et al. (1998) Accuracy in vivo - SEM 83 (±0.5 mm) Extracted tooth length 29

100 (±1.0 mm)Dunlap et al. (1998) Accuracy vital versus necrotic in vivo 82 (±0.5 mm) Extracted tooth length 34 McDonald et a/. (1999) Accuracy in vivo 95 (±0.5 mm) Extracted tooth length 20 Welk et al. (2003) Accuracy in vivo 91 (±0.5 mm) Extracted tooth length 32

Minor diameter Clinical accuracy in primary teethKatz et al. (1996) Accuracy in primary teeth - in vitro 100 (±0.5 mm) Extracted tooth length 20 Mente et al. (2002) Accuracy in resorbed primary teeth – 98 (±1.0 mm) Tooth length 24

In vitro Kielbassa et al. (2003) Accuracy in primary teeth - in vivo 64 (±1.0 mm) Extracted tooth length 71  The properties of the Root ZX (Gordon and Chandler, 2004)

•Root ZX has an average accuracy of 89 % to within + 0.5 mm of the apical foramen.

Page 31: Accurate Working Length Detemination Final en Do Club
Page 32: Accurate Working Length Detemination Final en Do Club

Electronic Apex Locators• Other third generation apex locators• Apex Finder AFA (EIE Analytic Endodontics, CA)

Five frequencies 86 % accuracy at –1 + 0.5 mm from radiographic apex.

Pommer et al., JOE, 2002

• Bingo 1020 (Forum Engineering Technologies, Lishon Lezion, Israel) also marketed as Ray-Pex 4 (Dentsplay). Two separate frequencies No difference between Root ZX or Bingo under

different irrigants. Kaufman et al., Int Endo J, 2002

Page 33: Accurate Working Length Detemination Final en Do Club

Electronic Apex Locators

• Other third generation apex locators• Justy II (Yoshida Co., Tokyo, Japan)

82 % accuracy in determining the area between minor and major foramen. Hoer and Attin, Int Endo J, 2004.

• Endy 5000 (Loser, Leverkusen, Germany)

81 % accuracy in determining the area between minor and major foramen. Hoer and Attin, Int Endo J, 2004.

• Neosono Ultima EZ (Satelec Inc., NJ)

100 % accuracy within 0.5 mm range. De moor et al,

Endodon Dent Traumatol, 1999. • Mark V plus (Moyco/Union Broach, Bethpage, NY)

Page 34: Accurate Working Length Detemination Final en Do Club

Electronic Apex Locators

Apex locators with other functions

• Tri Auto ZX (J. Morita Corp) an apex locator and a rechargeable electric handpiece for NiTi rotary files.

• 95 % accuracy to within 0.5 mm of the constriction. Gimberg et al., Int Endo J, 2002

Page 35: Accurate Working Length Detemination Final en Do Club

Electronic Apex Locators

Apex locators with other functions

• Dentaport ZX (J. Morita Corp) is Root ZX with attached endodontic hand piece for rotary NiTi files.

Page 36: Accurate Working Length Detemination Final en Do Club

Electronic Apex Locators

Apex locators with other functions

• Element Diagnostic Unit (SybronEndo, CA) and Apex Locator and EPT.

Page 37: Accurate Working Length Detemination Final en Do Club

• When radiographic working length in vitro was set to be 0.5 – 2.0 mm short of the apex and using the Root ZX in premolars: The proportion of overestimation was reduced

from 51% for radiographic method to 21% when the Root ZX was used.

ElAyouti A et al., J Endodon, 2002

Page 38: Accurate Working Length Detemination Final en Do Club

• Using an apex locator reduces exposure to radiation by minimizing the number of radiographs required to get an acceptable working length radiograph and also produces more accurate working length determination.

Brunton P.A. et al., J Endodon, 2002

Page 39: Accurate Working Length Detemination Final en Do Club

• Third generation Electronic Apex Locators work well in the presence of different irrigants including NaOCl, EDTA, and RC Prep.

Jenkins et al., J Endodon, 2001; Kaufman et al., Int Endo J, 2002;

Weiger et al., J Endodon, 1999

Page 40: Accurate Working Length Detemination Final en Do Club

• Preflaring the canal increases the accuracy of Electronic Apex Locator reading (Ibarrola et

al., J Endodon, 1999) and tactile sensation (Stabholtz et

al., J Endodon, 1995). It also reduces the potential reduction in canal length due to the elimination of curvatures.

Cohen and Burns., Pathways of the Pulp, 2002

Page 41: Accurate Working Length Detemination Final en Do Club

• Apex locators are not affected with the use of stainless steel files or NiTi files. (Thomas et al., J Endodon, 2003; Nekoofar et al., J Cali Dent Assoc, 2002)

• Apex locators are not affected by the size of the file used (Nguyen et al., Int Endo J, 1996).

Page 42: Accurate Working Length Detemination Final en Do Club

Investigator Country Use %

Whitten et al. (1996) USA GDPs 10 Saunders et al. (1999) Scotland GDPs 7.7 Yoshikawa et al. (2001) Japan GDPs 90 Chandler & Koshy (2002) New Zealand GDPs 27.5

Specialists 60 Hommez et al. (2003) Belgium GDPs 16 Electronic apex locator use (Gordon and Chandler, 2004)

Acceptance of Electronic Apex Locators

Page 43: Accurate Working Length Detemination Final en Do Club

• Although a recent study have shown no effect of 4 types of Electronic apex locators in vitro on Cardiac Pace Maker (Garofalo et al.

J Endodon, 2002). Use in patients with Cardiac Pace Maker should be consulted with their cardiologists.

Page 44: Accurate Working Length Detemination Final en Do Club

Limitations of Electronic Apex Locators

• Touching a metallic restoration will affect performance of Electronic Apex Locators.

• Leakage of saliva through cervical caries or open margin will also affect Electronic Apex Locators.

• Immature “Blunderbuss” apex will give short readings on Electronic Apex Locators.

Page 45: Accurate Working Length Detemination Final en Do Club

Accurate Working Length Determination

• A recent preoperative diagnostic quality radiograph is a must.

• Do not rely on a single method for accurate working length determination.

• Preflare the canal to increase your tactile sensation.

• Use an Electronic Apex Locator to determine you working length.

• Confirm you working length with a radiograph.

Page 46: Accurate Working Length Detemination Final en Do Club

Accurate Working Length Determination

The key to accurate working length determination is clear understanding of apex anatomy, knowledge of limitations of current techniques, and corroboration of information gained by different methods for working length determination.

Page 47: Accurate Working Length Detemination Final en Do Club
Page 48: Accurate Working Length Detemination Final en Do Club

Thank you