a closed-loop, non-linear, miniaturised capillary ...20draft... · a closed-loop, non-linear,...

1
A Closed-loop, Non-linear, Miniaturised Capillary Electrophoresis System Enabled by Zeta-potential Modification of Electroosmotic Flow Adam P. Lewis 1* , Andy Cranny 1 , Nicolas G. Green 1 , Nick R. Harris 1 , Keith Stokes 2 , Julian A. Wharton 3 and Robert J. K. Wood 3 1 Electronics and Computer Science, University of Southampton, UK. 2 Defence Science and Technology Laboratory, Porton Down, Salisbury, Wiltshire, UK. 3 National Centre for Advanced Tribology at Southampton, University of Southampton, UK. *Corresponding author: [email protected] Theory of Operation The figure on the right shows the normal mode of operation in which a sample is injected at the start of the channel and subjected to an E-field. After some time, the ions in the sample reach the end of the channel and have become slightly separated. At this point a potential is applied to the ZPM electrode to reverse the EOF. This will move all the ions in the sample back to the start of the channel; separation continues during this process. Once the ions reach the start of the channel, the potential applied to ZPM electrode is switched off and the sample migrates along the channel once again, still separating. This process is repeated until the desired separation is achieved. CCDs located at each end of the channel inform when the sample arrives at the channel ends, enabling a microprocessor based system to decide when to switch the ZPM voltage. This creates a novel robust closed-loop control system which is insensitive to sample migration time. Experimental Results COMSOL Simulation Results To illustrate the benefit of bi- directional transit along the channel length, a model was developed using COMSOL Multiphysics (v. 4.1). The results presented from this simulation, shown in the figure on the right, assume zero EOF during sample transit toward the end of the channel. At the start a sample mixture of three ions is injected, assuming a Gaussian distribution. After the first iteration the Fe 3+ ions are separated from the Cu 2+ and Fe 2+ ions. This separation improves on each further iteration. After the third iteration, the Cu 2+ and Fe 2+ peaks start to become noticeably separated. These simulation results indicate that further iterations will lead to a greater separation of Cu 2+ ions from Fe 2+ ions. Acknowledgements The authors would like to thank the Engineering and Physical Sciences Research Council UK (EPSRC) and Defence Science and Technology Laboratory (Dstl) for their financial support under grant number EP/F004362/1. Project Dissemination Results from the project have been presented at 3 national conferences, 5 international conferences and published in 12 peer reviewed proceedings and journals. For more information visit the project website: http://www.mdfshm.ecs.soton.ac.uk Ion Mobility (m 2 /Vs) [1] Al 3+ 6.2725 x 10 -8 Cr 3+ 6.8996 x 10 -8 Cu 2+ 5.5188 x 10 -8 Fe 2+ 5.5575 x 10 -8 Fe 3+ 7.0029 x 10 -8 Mn 2+ 5.5034 x 10 -8 Ni 2+ 5.1092 x 10 -8 [1] Values calculated with data taken from: Petr Vanýsek, “Ionic Conductivity and Diffusion at Infinite Dilution” CRC: Handbook of Chemistry and Physics (88 th Edition), 2007-2008 The figure above shows the effect of varying the ZPM voltage on the migration time of a fluorescein dye over a voltage range from 25 V to 250 V. At lower voltages it takes longer for the sample to get to the end of the channel, therefore the EOF can be significantly reduced. At high voltages it can be seen that the EOF is increased, up to a limiting level. 0 0.05 0.1 0.15 0.2 0.25 0.3 0.35 0.4 0.45 0.5 100 150 200 250 300 350 400 Detector Response Migration Time (seconds) VZPM: 250V VZPM: 150V VZPM: 75V CCD 1 ZPM Electrodes CCD 2 In-channel Electrochemical Detector Electrodes Electrokinetic Sample Injection Electrodes E-field Generation Electrodes At Start: Sample injected into channel. Mix of Cu 2+ , Fe 2+ and Fe 3+ . Cu 2+ and Fe 2+ After One Iteration: The sample arrives at end of channel but the Cu 2+ and Fe 2+ ions are not separated. Fe 3+ Cu 2+ and Fe 2+ After Two Iterations: The sample has effectively travelled further and thus is more separated. Fe 3+ Cu 2+ and Fe 2+ After Three Iterations: The separation between Cu 2+ and Fe 2+ is improved but more iterations are required to further increase the separation. Fe 3+ Start of channel End of channel Concentration (μmol/m 3 ) 0 50 40 30 20 10 60 70 80 90 Concentration (μmol/m 3 ) 0 50 40 30 20 10 60 70 80 90 Concentration (μmol/m 3 ) 0 50 40 30 20 10 60 70 80 90 Problem Outline Miniaturised capillary electrophoresis (CE) systems tend to have short channel lengths which makes it difficult to achieve good separation of complex mixtures of ions. In CE, ionic species migrate along a capillary (or channel) at a migration velocity which is dependant on their electrophoretic mobility and the electroosmotic flow (EOF). Migration Velocity = (Electrophoretic Mobility + EOF Mobility) × E-field The effect on migration time caused by varying the ZPM voltage is illustrated in the graph above where a fluorescein dye was used as a sample with a phosphate buffer (pH 7.2). A separation voltage of 300 V was applied over the 80 mm long channel. With negative ZPM voltages (not shown) we were able to visually confirm reversal of the EOF. Conclusions and Future Work This work has discussed a novel miniaturised CE system for ion separation using EOF modification to effectively increase the channel length. The concept of modification of EOF by control of zeta-potential has been experimentally verified, and the microfluidic device has been fabricated. The use of contactless conductivity detectors at either end of the channel along with associated microprocessor circuitry creates a closed-loop control system. The use of a closed-loop control system means that it: has application in a variety of different scenarios; is not restricted by the range of mobilities within a sample; requires minimal sample preparation; is insensitive to environmental variations; does not require calibration; has the potential to be an important component in autonomous field measurement systems. Future work will involve refining the device in order to progress towards a fully integrated system. Photograph of microfluidic device (above); the channel is highlighted with the dashed line. The channel width, depth and length are 200 μm, 35 μm and 80 mm respectively. PDMS Layer SU-8 Insulation Glass wafer Microfluidic channel Solution Reservoirs Electrode Layer Schematic (below) to illustrate the device layers. The metal electrodes (such as the ZPM and CCD electrodes) on the glass wafer are insulated from the channel by a thin layer (3-4 μm) of SU-8. Simulation results from COMSOL showing the separation of three metal ions increasing for every repeated transit along the channel length. Step 1: Start Sample of mixed ions injected into channel. Step 2: Sample Approaches End of Channel Sample starts to separate but many ions overlap. Step 3: Voltage Applied to ZPM Electrode EOF reversed sending ions back to start of channel whilst maintaining their separation. Step 4: Sample Approaches End of Channel Again Separation of sample is improved compared to that at the end of step 2. If further separation is required then the process from step 3 can be repeated. Extension to Technique Greater separation resolution can be obtained by taking this technique to the extreme, where the zeta-potential is modified such that the EOF is equal and opposite to the electrophoretic velocity of the slowest species of interest. Under this condition maximum separation resolution with respect to the slowest species is obtained. However, this requires very precise control of the EOF. A robust solution would be to use a method which modifies the EOF to give a long sample transit time as it travels along the channel. 0 200 400 600 800 1000 1200 1400 1600 0 50 100 150 200 250 Average Migration Time (seconds) ZPM Voltage (V) 80 mm An application area we are interested in is the identification of corrosion by-products. As shown in the table on the left, the mobilities of key metal ions are very similar making their separation in a conventional laboratory CE system difficult, and challenging in a miniaturised system. To address this issue we have developed a novel separation technique where we cycle the ions up and down the channel by periodically changing the direction of the EOF, thus realising a greater effective channel length for separation. Control of the EOF is achieved by the modification of the channel wall zeta-potential by placing an electrode close to the channel wall - the zeta-potential modification (ZPM) electrode. Microfluidic Device The device was fabricated in our state-of-the-art cleanroom facility using photolithographic techniques. Gold electrodes are patterned on a glass wafer; the microfluidic channels is formed in a layer of PDMS which is reversibly sealed to a layer of SU-8 that selectively insulates some of the underlying electrodes. A closed-loop system is enabled by the inclusion of contactless conductivity detectors (CCDs) at both ends of the channel which indicate when the ZPM voltage controlling the EOF needs to be switched. EOF E-field E-field E-field E-field EOF EOF EOF

Upload: others

Post on 22-Aug-2020

11 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: A Closed-loop, Non-linear, Miniaturised Capillary ...20Draft... · A Closed-loop, Non-linear, Miniaturised Capillary Electrophoresis System Enabled by Zeta-potential Modification

A Closed-loop, Non-linear, Miniaturised Capillary Electrophoresis System Enabled by Zeta-potential Modification of Electroosmotic Flow

Adam P. Lewis1*, Andy Cranny1, Nicolas G. Green1, Nick R. Harris1, Keith Stokes2, Julian A. Wharton3 and Robert J. K. Wood3 1Electronics and Computer Science, University of Southampton, UK. 2Defence Science and Technology Laboratory, Porton Down, Salisbury, Wiltshire, UK.

3National Centre for Advanced Tribology at Southampton, University of Southampton, UK. *Corresponding author: [email protected]

Theory of Operation The figure on the right shows the normal mode of operation in which a sample is injected at the start of the channel and subjected to an E-field. After some time, the ions in the sample reach the end of the channel and have become slightly separated. At this point a potential is applied to the ZPM electrode to reverse the EOF. This will move all the ions in the sample back to the start of the channel; separation continues during this process. Once the ions reach the start of the channel, the potential applied to ZPM electrode is switched off and the sample migrates along the channel once again, still separating. This process is repeated until the desired separation is achieved. CCDs located at each end of the channel inform when the sample arrives at the channel ends, enabling a microprocessor based system to decide when to switch the ZPM voltage. This creates a novel robust closed-loop control system which is insensitive to sample migration time.

Experimental Results

COMSOL Simulation Results To illustrate the benefit of bi-directional transit along the channel length, a model was developed using COMSOL Multiphysics (v. 4.1). The results presented from this simulation, shown in the figure on the right, assume zero EOF during sample transit toward the end of the channel. At the start a sample mixture of three ions is injected, assuming a Gaussian distribution. After the first iteration the Fe3+ ions are separated from the Cu2+ and Fe2+ ions. This separation improves on each further iteration. After the third iteration, the Cu2+ and Fe2+ peaks start to become noticeably separated. These simulation results indicate that further iterations will lead to a greater separation of Cu2+ ions from Fe2+ ions.

Acknowledgements The authors would like to thank the Engineering and Physical Sciences Research Council UK (EPSRC) and Defence Science and Technology Laboratory (Dstl) for their financial support under grant number EP/F004362/1.

Project Dissemination Results from the project have been presented at 3 national conferences, 5 international conferences and published in 12 peer reviewed proceedings and journals. For more information visit the project website: http://www.mdfshm.ecs.soton.ac.uk

Ion Mobility (m2/Vs)[1]

Al3+ 6.2725 x 10-8

Cr3+ 6.8996 x 10-8

Cu2+ 5.5188 x 10-8

Fe2+ 5.5575 x 10-8

Fe3+ 7.0029 x 10-8

Mn2+ 5.5034 x 10-8

Ni2+ 5.1092 x 10-8

[1] Values calculated with data taken from: Petr Vanýsek, “Ionic Conductivity and Diffusion at Infinite Dilution” – CRC: Handbook of Chemistry and Physics (88th Edition), 2007-2008

The figure above shows the effect of varying the ZPM voltage on the migration time of a fluorescein dye over a voltage range from 25 V to 250 V. At lower voltages it takes longer for the sample to get to the end of the channel, therefore the EOF can be significantly reduced. At high voltages it can be seen that the EOF is increased, up to a limiting level.

0

0.05

0.1

0.15

0.2

0.25

0.3

0.35

0.4

0.45

0.5

100 150 200 250 300 350 400

Det

ecto

r R

esp

on

se

Migration Time (seconds)

VZPM: 250V

VZPM: 150V

VZPM: 75V

CCD 1

ZPM Electrodes

CCD 2 In-channel Electrochemical Detector Electrodes

Electrokinetic Sample Injection Electrodes

E-field Generation Electrodes

Distance Along Channel

At Start: Sample injected into channel. Mix of Cu2+, Fe2+ and Fe3+.

Cu2+ and Fe2+

After One Iteration: The sample arrives at end of channel but the Cu2+ and Fe2+ ions are not separated.

Fe3+

Cu2+ and Fe2+

After Two Iterations: The sample has effectively travelled further and thus is more separated.

Fe3+

Cu2+ and Fe2+

After Three Iterations: The separation between Cu2+ and Fe2+ is improved but more iterations are required to further increase the separation.

Fe3+

Start of channel End of channel

Co

nce

ntr

atio

n (

µm

ol/

m3)

0

50

40

30

20

10

60

70

80

90

Co

nce

ntr

atio

n (

µm

ol/

m3)

0

50

40

30

20

10

60

70

80

90

Co

nce

ntr

atio

n (

µm

ol/

m3)

0

50

40

30

20

10

60

70

80

90

Problem Outline Miniaturised capillary electrophoresis (CE) systems tend to have short channel lengths which makes it difficult to achieve good separation of complex mixtures of ions. In CE, ionic species migrate along a capillary (or channel) at a migration velocity which is dependant on their electrophoretic mobility and the electroosmotic flow (EOF).

Migration Velocity = (Electrophoretic Mobility + EOF Mobility) × E-field

The effect on migration time caused by varying the ZPM voltage is illustrated in the graph above where a fluorescein dye was used as a sample with a phosphate buffer (pH 7.2). A separation voltage of 300 V was applied over the 80 mm long channel. With negative ZPM voltages (not shown) we were able to visually confirm reversal of the EOF.

Conclusions and Future Work This work has discussed a novel miniaturised CE system for ion separation using EOF modification to effectively increase the channel length. The concept of modification of EOF by control of zeta-potential has been experimentally verified, and the microfluidic device has been fabricated. The use of contactless conductivity detectors at either end of the channel along with associated microprocessor circuitry creates a closed-loop control system. The use of a closed-loop control system means that it:

● has application in a variety of different scenarios; ● is not restricted by the range of mobilities within a sample; ● requires minimal sample preparation; ● is insensitive to environmental variations; ● does not require calibration; ● has the potential to be an important component in autonomous

field measurement systems.

Future work will involve refining the device in order to progress towards a fully integrated system.

Photograph of microfluidic device (above); the channel is highlighted with the dashed line. The channel width, depth and length are 200 µm, 35 µm and 80 mm respectively.

PDMS Layer

SU-8 Insulation

Glass wafer

Microfluidic channel

Solution Reservoirs

Electrode Layer

Schematic (below) to illustrate the device layers. The metal electrodes (such as the ZPM and CCD electrodes) on the glass wafer are insulated from the channel by a thin layer (3-4 µm) of SU-8.

Simulation results from COMSOL showing the separation of three metal ions increasing for every repeated transit along the channel length.

Step 1: Start Sample of mixed ions injected into channel.

Step 2: Sample Approaches End of Channel Sample starts to separate but many ions overlap.

Step 3: Voltage Applied to ZPM Electrode EOF reversed sending ions back to start of channel whilst maintaining their separation.

Step 4: Sample Approaches End of Channel Again Separation of sample is improved compared to that at the end of step 2. If further separation is required then the process from step 3 can be repeated.

Extension to Technique Greater separation resolution can be obtained by taking this technique to the extreme, where the zeta-potential is modified such that the EOF is equal and opposite to the electrophoretic velocity of the slowest species of interest. Under this condition maximum separation resolution with respect to the slowest species is obtained. However, this requires very precise control of the EOF. A robust solution would be to use a method which modifies the EOF to give a long sample transit time as it travels along the channel.

0

200

400

600

800

1000

1200

1400

1600

0 50 100 150 200 250

Ave

rage

Mig

rati

on

Tim

e (s

eco

nd

s)

ZPM Voltage (V)

80 mm

An application area we are interested in is the identification of corrosion by-products. As shown in the table on the left, the mobilities of key metal ions are very similar making their separation in a conventional laboratory CE system difficult, and challenging in a miniaturised system. To address this issue we have developed a novel separation technique where we cycle the ions up and down the channel by periodically changing the direction of the

EOF, thus realising a greater effective channel length for separation. Control of the EOF is achieved by the modification of the channel wall zeta-potential by placing an electrode close to the channel wall - the zeta-potential modification (ZPM) electrode.

Microfluidic Device The device was fabricated in our state-of-the-art cleanroom facility using photolithographic techniques. Gold electrodes are patterned on a glass wafer; the microfluidic channels is formed in a layer of PDMS which is reversibly sealed to a layer of SU-8 that selectively insulates some of the underlying electrodes. A closed-loop system is enabled by the inclusion of contactless conductivity detectors (CCDs) at both ends of the channel which indicate when the ZPM voltage controlling the EOF needs to be switched.

EOF

E-field

E-field

E-field

E-field

EOF

EOF

EOF