Page 1: 3.0 GHz BITG Replacement Kit Installation Noteliterature.cdn.keysight.com/litweb/pdf/E4403-90056.pdfThis kit contains the parts and instructions to install the replacement T.G. Module

Installation Note

Agilent Technologies 3.0 GHz BITG Replacement Kit

Part Number E4403-90056 Supersedes: E4403-90055Printed in USA November 2007

Page 2: 3.0 GHz BITG Replacement Kit Installation Noteliterature.cdn.keysight.com/litweb/pdf/E4403-90056.pdfThis kit contains the parts and instructions to install the replacement T.G. Module

Notice.

The information contained in this document is subject to change without notice.

Agilent Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.

Microsoft® is a U.S. registered trademark of Microsoft Corp.

© Copyright 2005 - 2007 Agilent Technologies, Inc.

Page 3: 3.0 GHz BITG Replacement Kit Installation Noteliterature.cdn.keysight.com/litweb/pdf/E4403-90056.pdfThis kit contains the parts and instructions to install the replacement T.G. Module

IntroductionProducts Affected: E4402B Spectrum Analyzer

E4403B Spectrum AnalyzerE4404B Spectrum AnalyzerE4405B Spectrum AnalyzerE4407B Spectrum AnalyzerE4408B Spectrum AnalyzerE7402A EMC AnalyzerE7403A EMC AnalyzerE7404A EMC AnalyzerE7405A EMC Analyzer

Serial Numbers: US00000000/US99999999MY00000000/MY99999999SG00000000/MY99999999

To Be Performed By: (X) Agilent Technologies Service Center

(X) Personnel Qualified by Agilent

(X) CustomerEstimated Installation Time: 1.5 Hours Estimated Verification Time: 2.0 Hours

Installation Note E4403-90056 3

Page 4: 3.0 GHz BITG Replacement Kit Installation Noteliterature.cdn.keysight.com/litweb/pdf/E4403-90056.pdfThis kit contains the parts and instructions to install the replacement T.G. Module

Introduction

The Tracking Generator replacement is now a kit due to design changes that were made to the T.G. assembly. This kit contains the parts and instructions to install the replacement T.G. Module in the ESA or EMC Analyzer. Analyzers built between March, 1999 to May, 2005 used a 3-port design (See Figure 1). Analyzers built after June, 2005 use a newer 2-port design (See Figure 2). The new 2-port T.G. Module and replacement semi-rigid cables supplied in this kit will function in any ESA or EMC Analyzer.

Figure 1 3-Port T.G. Module

Figure 2 2-Port T.G. Module

Installation Note E4403-900564

Page 5: 3.0 GHz BITG Replacement Kit Installation Noteliterature.cdn.keysight.com/litweb/pdf/E4403-90056.pdfThis kit contains the parts and instructions to install the replacement T.G. Module

Installation Kit Parts ListTable 1

Qty. Description Item Number

1 Tracking Generator Module 5087-7709

1 Cable, Control board to T.G. Module E4403-20073

1 Cable, Attenuator to T.G. Module E4403-20072

1 Cable, Attenuator to T.G. Module E4403-20083

5 Screw, M3X8 mm (T-10 TORX) 0515-0372

3 Screw, M3X10 mm (T-10 TORX) 0515-0374

1 Installation Note This Document

Installation Note E4403-90056 5

Page 6: 3.0 GHz BITG Replacement Kit Installation Noteliterature.cdn.keysight.com/litweb/pdf/E4403-90056.pdfThis kit contains the parts and instructions to install the replacement T.G. Module

Tools Required

• T-10 TORX screwdriver

• T-15 TORX screwdriver

• Spectrum Analyzer, microwave

• Power Meter

• Power Sensor

• Digital Voltmeter

• BNC (f) to dual banana plug

• Cable, BNC 91 cm (36 in.)

• Adapter, Type-N tee, (m,f,m)

• Adapter, Type-N (m) to BNC (f)

• Cable, Type-N, 62 cm (24 in.)

• Cable, BNC, 23 cm (9 in.)

• Torque Driver (Torque driver requirements are shown in the following table)

• Torque Wrench (5/16-inch) (Torque driver requirements are shown in the following table)

Torque Settings

To avoid potential RFI leakage and prevent connector damage, tighten screws and RF coax cable connectors to the following torque limits:

WARNING Before you disassemble the instrument, turn the power switch OFF and unplug the instrument. Failure to unplug the instrument can result in personal injury.

CAUTION Electrostatic discharge (ESD) can damage or destroy electronic components. All work on electronic assemblies should be performed at a static-safe workstation. Refer to the documentation that pertains to your instrument for information about static-safe workstations and ordering static safe accessories.

ItemTorque

In-lb N-cm

SMA Connectors 8.5 95

3-mm, T-10 TORX screws 9.0 101

3.5-mm, T-10 TORX screws 14 157

4-mm, T-15 TORX screws 21 236

Installation Note E4403-900566

Page 7: 3.0 GHz BITG Replacement Kit Installation Noteliterature.cdn.keysight.com/litweb/pdf/E4403-90056.pdfThis kit contains the parts and instructions to install the replacement T.G. Module

Procedure

Remove the Rear Frame and Outer Cover

1. Refer to Figure 3 and disconnect the analyzer from ac power (1).

2. Remove any adapters or cables (2) connected to the front frame of the instrument.

3. Carefully place the analyzer on the work surface with the front frame (3) facing down.

4. Position the handle (4) as shown.

5. Remove the four screws (5) that hold the rear frame (6) and outer cover (7) in place.

6. Remove the rear frame (6).

7. Lift the instrument rear frame (6) and the outer cover (7) up and off the instrument and set them aside.

Figure 3 Outer Cover and Rear Frame Removal

Installation Note E4403-90056 7

Page 8: 3.0 GHz BITG Replacement Kit Installation Noteliterature.cdn.keysight.com/litweb/pdf/E4403-90056.pdfThis kit contains the parts and instructions to install the replacement T.G. Module

Remove Chassis Cover

8. Refer to Figure 4 and lay the analyzer flat.

9. Remove the screws (2) and (3) attaching the chassis cover (1) to the chassis. Note that the number of screws attaching the chassis cover may vary depending upon which options are installed.

10.Remove the chassis cover (1) from the chassis.

Figure 4 Chassis Cover Removal

Installation Note E4403-900568

Page 9: 3.0 GHz BITG Replacement Kit Installation Noteliterature.cdn.keysight.com/litweb/pdf/E4403-90056.pdfThis kit contains the parts and instructions to install the replacement T.G. Module

Remove Front Frame

CAUTION Use ESD precautions when performing this replacement procedure.

The A1 front frame assembly can be extended from the instrument without detaching any connections.

11. Refer to Figure 5. Remove the two screws (1), on the bottom side of the instrument that secure the front frame to the RF assembly.

12.Place the instrument with the top side facing up and remove the remaining two screws (2) that secure the front frame subpanel to the chassis.

Figure 5 A1 Front Frame Assembly Removal

Installation Note E4403-90056 9

Page 10: 3.0 GHz BITG Replacement Kit Installation Noteliterature.cdn.keysight.com/litweb/pdf/E4403-90056.pdfThis kit contains the parts and instructions to install the replacement T.G. Module

13.Refer to Figure 6. Slide the front frame forward until it latches. Disengage the latches on each side of the instrument by sliding the front frame back toward the instrument a small amount and then gently pulling out the latch tabs.

14.Slide the front frame forward just far enough to reach the ribbon connector attached to J2 on the front panel keyboard printed circuit board. Disconnect the ribbon cable from J2.

NOTE Instruments with option B7B, TV Picture on Screen will have an additional cable connected to the front panel interface board that will need to be disconnected before the front frame assembly can be removed.

Figure 6 Remove Front Panel Connector(s)

15.Remove the front frame and put it aside in a safe location.

Installation Note E4403-9005610

Page 11: 3.0 GHz BITG Replacement Kit Installation Noteliterature.cdn.keysight.com/litweb/pdf/E4403-90056.pdfThis kit contains the parts and instructions to install the replacement T.G. Module

16.Refer to Figure 7. Disconnect the SMB cable (1) and the SMA cable (2) from the tracking generator assembly.

NOTE The figure refers to a 3.0 GHz analyzer. Other ESA or EMC models will be different, however items (1) & (2) are located in the same location on the T.G. Assembly.

Figure 7 A2 Tracking Generator Cables

Installation Note E4403-90056 11

Page 12: 3.0 GHz BITG Replacement Kit Installation Noteliterature.cdn.keysight.com/litweb/pdf/E4403-90056.pdfThis kit contains the parts and instructions to install the replacement T.G. Module

17.Refer to Figure 8. Remove the five screws (1), (four from the side and one from the back), securing the A2 Tracking Generator Assembly (2) top the chassis.

18.Carefully pull up on the Tracking Generator Assembly to remove it from the motherboard connector.

Figure 8 3.0 GHz Tracking Generator Assembly Hardware

19.Remove the two semi-rigid cables completely and unplug the power control harness.

Installation Note E4403-9005612

Page 13: 3.0 GHz BITG Replacement Kit Installation Noteliterature.cdn.keysight.com/litweb/pdf/E4403-90056.pdfThis kit contains the parts and instructions to install the replacement T.G. Module

NOTE Figure 9 illustrates the older style 3-port T.G. Module and Figure 10 illustrates the newer style 2-port T.G. Module. Regardless of T.G. Module version, remove the semi-rigid cables completely. New semi-rigid cables provided in this kit should be used.

Figure 9 Older Style 3-Port Tracking Generator Module

Figure 10 Newer Style 2-Port Tracking Generator Module

Installation Note E4403-90056 13

Page 14: 3.0 GHz BITG Replacement Kit Installation Noteliterature.cdn.keysight.com/litweb/pdf/E4403-90056.pdfThis kit contains the parts and instructions to install the replacement T.G. Module

20.Refer to Figure 11. Remove and discard the (3) screws that hold the T.G. Module to the bracket using a T-10 TORX.

Figure 11 Tracking Generator Module Screws

21.Install the replacement T.G. Module onto the T.G. bracket.

22.Refer to Figure 11. Install (3) 0515-0374 new screws provided in the kit that attach the T.G. Module to the T.G. bracket.

NOTE Based on the date of shipment of the TG option in your analyzer, you could have 2 different TG output attenuator part numbers. The attenuator can be found on the TG board and is connected to the output of the BITG/TG module. The 2 attenuator part numbers are 0955-0679 or 0955-2022. The part number can be found on the attenuator. Based on the part number of your attenuator, it will use different cables for the input and output of the attenuator.

If the part number of the attenuator is 0955-0679, you will need semi-rigid cable E4403-20072 supplied in this kit to perform the next step. If the part number of the attenuator is 0955-2022, you will need semi-rigid cable E4403-20083 supplied in this kit to perform the next step.

Refer to Figure 11-2 to see where the part number is on the attenuator.

Installation Note E4403-9005614

Page 15: 3.0 GHz BITG Replacement Kit Installation Noteliterature.cdn.keysight.com/litweb/pdf/E4403-90056.pdfThis kit contains the parts and instructions to install the replacement T.G. Module

Figure 11-2 Attenuator Part Number Location

23.Refer to Figure 12. Install the E4403-20072 semi-rigid cable if the attenuator part number is 0955-0679 or install the E4403-20083 semi-rigid cable if the attenuator part number is 0955-2022

Figure 12 E4403-20072/E4403-20083 Semi-Rigid Cable

Installation Note E4403-90056 15

Page 16: 3.0 GHz BITG Replacement Kit Installation Noteliterature.cdn.keysight.com/litweb/pdf/E4403-90056.pdfThis kit contains the parts and instructions to install the replacement T.G. Module

24.Refer to Figure 13. Install the E4403-20073 semi-rigid cable.

Figure 13 E4403-20073 Semi-Rigid Cable

25.Refer to Figure 14. Install the power control harness.

Figure 14 Power Control Harness

26.Refer to Figure 8. Install the completed T.G. assembly into the analyzer being careful as you plug the assembly into the motherboard. Install the (5) screws that hold the T.G. assembly to the chassis.

27.Refer to Figure 7. Reconnect the SMB cable (1) and the SMA cable (2) to the T.G. module.

Installation Note E4403-9005616

Page 17: 3.0 GHz BITG Replacement Kit Installation Noteliterature.cdn.keysight.com/litweb/pdf/E4403-90056.pdfThis kit contains the parts and instructions to install the replacement T.G. Module

Replace Front Frame Assembly

28.Refer to Figure 6. Carefully slide the front frame onto the chassis and attach the ribbon cable to connector J2 on the front panel keyboard printed circuit board. Make sure that the ribbon cable is not pinched between assemblies. If option B7B exists in the analyzer, be sure to reconnect that cable as well. Be sure to use the outermost connector on the B7B cable as the inner connector is used for grounding purposes. Align the Input connector and RF Out connector with their respective openings in the front frame.

29.Refer to Figure 5. Replace the four screws that secure the front frame to the chassis at locations (1) and (3). Torque these screws to 101 N-cm (9 in-lb.).

Replace Chassis Cover

30.Refer to Figure 4. Carefully position the chassis cover on the instrument as shown.

31.Replace the 15 pan head screws (3) and three flathead screws (2) as shown. Torque these screws to 101 N-cm (9 in-lb.).

Replace Outer Cover and Rear Frame

32.Refer to Figure 3. Carefully place the spectrum analyzer on the work surface with the front frame facing down.

33.Replace the instrument outer cover by matching the grill on the side of the case to the fan on the A5 power supply assembly.

34.Fit the leading edge of the case completely into the slot on the back of the front frame assembly. The rear EMI contact grid that clips onto the rear frame can come loose. When replacing the rear frame, make sure the gasket is in the correct position.

35.Replace the rear frame assembly (6) using the four screws (5) to fasten the rear frame to the instrument. Torque these screws to 236 N-cm (21 in-lb.).

Verify Analyzer Performance

Adjustments Required

Align Now, TG (Press System, Alignments, Align Now, TG on the analyzer).

Performance Testing Required

The performance test procedures can be found in the ESA or EMC Calibration Guide or by using the Calibration Software. The latest software information and downloads are available at:http://www.agilent.com/find/calibrationsoftware

• Absolute Amplitude and Vernier Sweep Accuracy

• Tracking Generator Level Flatness

• Harmonic Spurious Outputs

• Non-Harmonic Spurious Outputs

Installation Note E4403-90056 17