Page 1: Yuxiang Zhao, INFN Trieste For the COMPASS RICH -1 collaborationthgem.ts.infn.it/PapersTalks/YuxiangZhao_VCI2019.pdf · 2019. 4. 29. · COMPASS RICH-1 has been upgraded with 1.4

Yuxiang Zhao, INFN TriesteFor the COMPASS RICH-1 collaboration

1

Page 2: Yuxiang Zhao, INFN Trieste For the COMPASS RICH -1 collaborationthgem.ts.infn.it/PapersTalks/YuxiangZhao_VCI2019.pdf · 2019. 4. 29. · COMPASS RICH-1 has been upgraded with 1.4

Introduction: COMPASS RICH-1 at CERNPhoton detector upgrade using MPGD based Hybrid detectorsDetector commissioning and performanceOngoing activities and conclusions

2

Page 3: Yuxiang Zhao, INFN Trieste For the COMPASS RICH -1 collaborationthgem.ts.infn.it/PapersTalks/YuxiangZhao_VCI2019.pdf · 2019. 4. 29. · COMPASS RICH-1 has been upgraded with 1.4

3

NIM A 577 (2007) 455

NIM A 779 (2015) 69

COMPASS Spectrometer at CERN SPS

Hadron PID range: 3-60 GeVRICH-1

MAPMTs coupled to lens telescope

Top photon detectors

MWPCs+CsI(from RD26): successful but performance limitations, in particular for the 4 central chambers

n. of ph.s@ β= 1

JINST 9 (2014) P01006

MWPCs+CsI

Page 4: Yuxiang Zhao, INFN Trieste For the COMPASS RICH -1 collaborationthgem.ts.infn.it/PapersTalks/YuxiangZhao_VCI2019.pdf · 2019. 4. 29. · COMPASS RICH-1 has been upgraded with 1.4

4

NIM A 577 (2007) 455

NIM A 779 (2015) 69

COMPASS Spectrometer at CERN SPS

Hadron PID range: 3-60 GeVRICH-1

MAPMTs coupled to lens telescopes

Top photon detectors

MWPCs+CsI(from RD26): successful but performance limitations, in particular for the 4 central chambers

Page 5: Yuxiang Zhao, INFN Trieste For the COMPASS RICH -1 collaborationthgem.ts.infn.it/PapersTalks/YuxiangZhao_VCI2019.pdf · 2019. 4. 29. · COMPASS RICH-1 has been upgraded with 1.4

Introduction: COMPASS RICH-1 at CERNPhoton detector upgrade using MPGD based Hybrid detectorsDetector commissioning and performanceOngoing activities and conclusions

5

Page 6: Yuxiang Zhao, INFN Trieste For the COMPASS RICH -1 collaborationthgem.ts.infn.it/PapersTalks/YuxiangZhao_VCI2019.pdf · 2019. 4. 29. · COMPASS RICH-1 has been upgraded with 1.4

6

modular structure: one module = 600x300 mm2Hybrid PD scheme

Standard Bulk Micromegas produced at CERN

CsITH1

TH2

MM

quartz

Drift

8mm X 8mm padsat positive HV Capacitive coupling APV25

Page 7: Yuxiang Zhao, INFN Trieste For the COMPASS RICH -1 collaborationthgem.ts.infn.it/PapersTalks/YuxiangZhao_VCI2019.pdf · 2019. 4. 29. · COMPASS RICH-1 has been upgraded with 1.4

7

Measurement of the raw material thickness before the THGEM production, accepted:

± 15 µm ↔ gain uniformity σ < 7%

THGEM polishing with an “ad hoc” protocol setup by us including baking:

>90% break-down limit obtained

X-ray THGEM test to accessgain uniformity (<7%) and sparkbehaviour 55Fe source

spectra

X-ray MM test to access integrity and gain uniformity (<5%)

MICROMEGAS

Page 8: Yuxiang Zhao, INFN Trieste For the COMPASS RICH -1 collaborationthgem.ts.infn.it/PapersTalks/YuxiangZhao_VCI2019.pdf · 2019. 4. 29. · COMPASS RICH-1 has been upgraded with 1.4

8

Turbopump

THGEM

THGEM box

4 evaporators

piston

QE measurement

QE uniformity 3 % r.m.s. within a photocathode 10 % r.m.s. among photocathodes mean value: 93% of reference

Page 9: Yuxiang Zhao, INFN Trieste For the COMPASS RICH -1 collaborationthgem.ts.infn.it/PapersTalks/YuxiangZhao_VCI2019.pdf · 2019. 4. 29. · COMPASS RICH-1 has been upgraded with 1.4

9

mechanics

Wire planes

detector layers

Automatized glueing

Glueing the support pillars

Assembling CsI coated THGEM ina dedicated Glove box flushing withN2 gas

Page 10: Yuxiang Zhao, INFN Trieste For the COMPASS RICH -1 collaborationthgem.ts.infn.it/PapersTalks/YuxiangZhao_VCI2019.pdf · 2019. 4. 29. · COMPASS RICH-1 has been upgraded with 1.4

10

Gain stability vs P, T: G = G(V, T/P) Enhanced in a multistage detector ΔT = 1 degree ΔG ≈ 12 % ΔP = 10 mbar ΔG ≈ 20 %

THE WAY OUT: Compensate T/P variations by V

Page 11: Yuxiang Zhao, INFN Trieste For the COMPASS RICH -1 collaborationthgem.ts.infn.it/PapersTalks/YuxiangZhao_VCI2019.pdf · 2019. 4. 29. · COMPASS RICH-1 has been upgraded with 1.4

Introduction: COMPASS RICH-1 at CERNPhoton detector upgrade using MPGD based Hybrid detectorsHybrid detector commissioning and performanceOngoing activities and conclusions

11

Page 12: Yuxiang Zhao, INFN Trieste For the COMPASS RICH -1 collaborationthgem.ts.infn.it/PapersTalks/YuxiangZhao_VCI2019.pdf · 2019. 4. 29. · COMPASS RICH-1 has been upgraded with 1.4

12

1 channel in a particular pedestal run 6144 channels for 1 detector in a particular pedestal run

Sigma

The niose levels are:• MWPC: ~ 600 e-

• Hybrid: ~ 900 e-

The noise levels are very stable in time

12 Gaseous Photon Detectors using APV based readout (4 hybrid, 12 MWPC), 6144 Channels/detector

39 pedestal runs in 2017 COMPASS run

Page 13: Yuxiang Zhao, INFN Trieste For the COMPASS RICH -1 collaborationthgem.ts.infn.it/PapersTalks/YuxiangZhao_VCI2019.pdf · 2019. 4. 29. · COMPASS RICH-1 has been upgraded with 1.4

13

6.4 GeV pionHybrid

PMTPMT

6.76 GeV pionHybrid

PMT

6.36 GeV pionHybrid

• Ring center calculated from particle trajectory• Detected photoelectrons : hits on the sensors

Ring centerfrom track

photons

For reference:

θ(β= 1) = 52.5 mrad

Page 14: Yuxiang Zhao, INFN Trieste For the COMPASS RICH -1 collaborationthgem.ts.infn.it/PapersTalks/YuxiangZhao_VCI2019.pdf · 2019. 4. 29. · COMPASS RICH-1 has been upgraded with 1.4

14

Resolution ~ 1.86 mrad

Page 15: Yuxiang Zhao, INFN Trieste For the COMPASS RICH -1 collaborationthgem.ts.infn.it/PapersTalks/YuxiangZhao_VCI2019.pdf · 2019. 4. 29. · COMPASS RICH-1 has been upgraded with 1.4

15

Temperature fluctuation

Pressure fluctuation

Gain fluctuation: RMS/MEAN ~ 6%

Gain 13511.4 +- 121.764

Fit Range: 50 to 160Formula: P0*exp(-P1*x)1 ADC Channel -> 300 electrons

A non-negligible achievement dueto P, T corrections on HV application

2000

hou

rs o

f run

ning

Page 16: Yuxiang Zhao, INFN Trieste For the COMPASS RICH -1 collaborationthgem.ts.infn.it/PapersTalks/YuxiangZhao_VCI2019.pdf · 2019. 4. 29. · COMPASS RICH-1 has been upgraded with 1.4

16

Extrapolate to saturation, number of photon= 12.9First part of the function = 10.3 +/- 0.4Second part of the function= 2.6 +/- 0.3

Page 17: Yuxiang Zhao, INFN Trieste For the COMPASS RICH -1 collaborationthgem.ts.infn.it/PapersTalks/YuxiangZhao_VCI2019.pdf · 2019. 4. 29. · COMPASS RICH-1 has been upgraded with 1.4

Introduction: COMPASS RICH-1 at CERNPhoton detector upgrade using MPGD based Hybrid detectorsHybrid detector commissioning and performanceOngoing activities and conclusions

17

Page 18: Yuxiang Zhao, INFN Trieste For the COMPASS RICH -1 collaborationthgem.ts.infn.it/PapersTalks/YuxiangZhao_VCI2019.pdf · 2019. 4. 29. · COMPASS RICH-1 has been upgraded with 1.4

18

• Development of an optimized detector for finer spatial resolution based on the hybrid THGEM + MM and “mini-pads” of size 3mm x 3mm

• Study the compatibility of these hybrid PDs with CF4 for a windowless RICH for the future Electron Ion Collider

• Exploring the possibility to use a more robust photocathode in the far UV: hydrogenated nano-diamond crystals

Beam test

Diamond & Related Materials 76 (2017) 1

Page 19: Yuxiang Zhao, INFN Trieste For the COMPASS RICH -1 collaborationthgem.ts.infn.it/PapersTalks/YuxiangZhao_VCI2019.pdf · 2019. 4. 29. · COMPASS RICH-1 has been upgraded with 1.4

COMPASS RICH-1 has been upgraded with 1.4 m2 of MPGD-based PDsThe Hybrid PD: 2 THGEMs (1 with CsI) + Micromegas are nicely operatingThey present good gain performance (>10k in a real experiment) and

stability, clean rings etc.1.85 mrad single photon angular resolution 10 detected photons per ring at saturationR&D for future RICH projects are considering the use of this technology

19