Page 1: Yury Gavrikov 25/03/2009 1. 2  I met with two very nice g…oniometers 3 Because we decided to adapt their for a new experiment

1

THE GONIOMETERS

Yury Gavrikov25/03/2009

Page 2: Yury Gavrikov 25/03/2009 1. 2  I met with two very nice g…oniometers 3 Because we decided to adapt their for a new experiment

2

Looking back

Page 3: Yury Gavrikov 25/03/2009 1. 2  I met with two very nice g…oniometers 3 Because we decided to adapt their for a new experiment

1 year ago… I met with two very nice g…oniometers

3

Because we decided to adapt their for a new experiment

Page 4: Yury Gavrikov 25/03/2009 1. 2  I met with two very nice g…oniometers 3 Because we decided to adapt their for a new experiment

The idea of mechanicsLinear movement

Gear boxBelt drive + “virtual”

screw

Engine DC-servo motor

Carriage slides on the 2 guides

4

Page 5: Yury Gavrikov 25/03/2009 1. 2  I met with two very nice g…oniometers 3 Because we decided to adapt their for a new experiment

Linear motion

(full range = 90mm)(full range = 90mm)

Inside of beam

Outside of beam

5

Page 6: Yury Gavrikov 25/03/2009 1. 2  I met with two very nice g…oniometers 3 Because we decided to adapt their for a new experiment

The idea of mechanics Angular

movement

1mm~

63m

m

AxisAxis

PusherPusher

Angle Stepper motor

&Belt driveMicroscrew

6

Page 7: Yury Gavrikov 25/03/2009 1. 2  I met with two very nice g…oniometers 3 Because we decided to adapt their for a new experiment

ModernizationUnificated solution for angular & linear

motion:

• Stepper motor for driving

• LVDT for position measurement

• Common Control system for all devices

It was needed the mashining of original parts

7

Page 8: Yury Gavrikov 25/03/2009 1. 2  I met with two very nice g…oniometers 3 Because we decided to adapt their for a new experiment

Mechanical revision

After modernization during last month before installation :

- 4 revisions and full disassembling- A lot of motion tests and measurements- Fixing of sources of the problems

8

Page 9: Yury Gavrikov 25/03/2009 1. 2  I met with two very nice g…oniometers 3 Because we decided to adapt their for a new experiment

9

Measurement & Calibration

Page 10: Yury Gavrikov 25/03/2009 1. 2  I met with two very nice g…oniometers 3 Because we decided to adapt their for a new experiment

Angle measurement Digital Autocollimator specially

designed for the experiment

Range: ~ 8000 mrad (horizontal)~ 6500 mrad (vertical)

Resolution: < 0.5mrad10

Page 11: Yury Gavrikov 25/03/2009 1. 2  I met with two very nice g…oniometers 3 Because we decided to adapt their for a new experiment

Angle measurement

The general Idea of angle measurement

Mirror

11

Page 12: Yury Gavrikov 25/03/2009 1. 2  I met with two very nice g…oniometers 3 Because we decided to adapt their for a new experiment

Angle calibration The Idea of crystal angle calibration

Reference mirror

Crystal

12

Page 13: Yury Gavrikov 25/03/2009 1. 2  I met with two very nice g…oniometers 3 Because we decided to adapt their for a new experiment

Angle calibration Angle vs. LVDT position

-50000 0 50000

-7500

-5000

-2500

0

2500

5000

7500

An

gle

, u

rad

LVDT, uV

Goniometer 2Goniometer 1

-50000 0 50000

-7500

-5000

-2500

0

2500

5000

7500

Ang

le, u

rad

LVDT, uV

13

Page 14: Yury Gavrikov 25/03/2009 1. 2  I met with two very nice g…oniometers 3 Because we decided to adapt their for a new experiment

Nonlinearity of goniometers

Periodical structure corresponds to the motor revolutionNonlinearity : about +/- 50mrad

-30000 -20000 -10000 0 10000 20000 30000-50

-40

-30

-20

-10

0

10

20

30

40

Non

linea

rity,

ura

d

LVDT, V

-30000 -20000 -10000 0 10000 20000-60

-50

-40

-30

-20

-10

0

10

20

30

No

nlin

ear

ity,

rad

LVDT, V

14

Page 15: Yury Gavrikov 25/03/2009 1. 2  I met with two very nice g…oniometers 3 Because we decided to adapt their for a new experiment

Stability

15

The goniometers are very SENSITIVE to any mechanical disturbance

The design is SOFT. “Don’t touch” should be placed everywhere

We can move it only in work direction to avoid all gaps

Mechanical stability

Page 16: Yury Gavrikov 25/03/2009 1. 2  I met with two very nice g…oniometers 3 Because we decided to adapt their for a new experiment

16

Stability Stability of LVDT during 14 hours looks quite good (*)

Gonio2, LVDT

3-5 mrad

(*) with averaging to reduce the noise

Page 17: Yury Gavrikov 25/03/2009 1. 2  I met with two very nice g…oniometers 3 Because we decided to adapt their for a new experiment

Repeatability Preliminary estimation for angular motion

17

-30000 -20000 -10000 0 10000 20000 30000

-70-60-50-40-30-20-10

0102030405060708090

No

nlin

ea

rity

, ra

d

LVDT, V

Page 18: Yury Gavrikov 25/03/2009 1. 2  I met with two very nice g…oniometers 3 Because we decided to adapt their for a new experiment

18

Alignment system

Page 19: Yury Gavrikov 25/03/2009 1. 2  I met with two very nice g…oniometers 3 Because we decided to adapt their for a new experiment

Crystal align andAngle measurement on the tank

T ank

D i g i tal Auto c o l l i m ato r

P e ntapr i zm 1

P e ntapr i zm 2

C r ys tal 1 C r ys tal 2

R e fe r e nc e m i r r o r

19

Crystal 1

Crystal 2

Ref. mirror

Page 20: Yury Gavrikov 25/03/2009 1. 2  I met with two very nice g…oniometers 3 Because we decided to adapt their for a new experiment

Photos of alignment system

Digital autocollimator

Reference mirror

Laser

20

Penta prizm

Viewport

Page 21: Yury Gavrikov 25/03/2009 1. 2  I met with two very nice g…oniometers 3 Because we decided to adapt their for a new experiment

Sources of uncertainty• Initial orientation of the alignment rail to

the tank geometry +/- 350mrad• Nonlinearity of goniometers +/- 50 mrad

21

Page 22: Yury Gavrikov 25/03/2009 1. 2  I met with two very nice g…oniometers 3 Because we decided to adapt their for a new experiment

Conclusions

We can use this goniometers with some magic gestures right now

BUT

To reach better results in the future we have to develop a new goniometer

22