xn series - sysmex · » the ssc-fsc scattergram of the wdf channel showed two neutrophil...

41
XN series Case interpretation Gebruikersdag Vlaanderen- 6 oktober 2016

Upload: others

Post on 08-Apr-2020

21 views

Category:

Documents


3 download

TRANSCRIPT

Page 1: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

XN seriesCase interpretation

Gebruikersdag Vlaanderen- 6 oktober 2016

Page 2: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

RET channel PLT-F channel

WDF channel WPC channel WNR channel

Fluorescence flow cytometry

Page 3: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

Case 1

Page 4: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

Case 1: Initial measurement

• 402.04.2015

Patient details:69-year-old male with myocardial infarction 10 years ago;

now obvious spleno-megaly and pal-pable lymph nodes

Conv. units:HGB: 11.3 g/dLMCH: 48.6 pgMCHC: 40.6 g/dL

Page 5: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

Case 1: Reflex measurement

Patient details:69-year-old male with myocardial infarction 10 years ago;

now obvious spleno-megaly and pal-pable lymph nodes

Conv. units:HGB: 11.3 g/dLMCH: 48.6 pgMCHC: 40.6 g/dL

Page 6: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

Smear results

Differential ResultsSeg neutrophilsLymphocytesMonocytesEosinophilsBasophilsBandsMetamyelocytesMyelocytesPromyelocytesMyeloblasts

Nucleated red blood cells

199

Comments» 12 smudge cells (excl. from the differential count

as per recommendation)

Page 7: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

Case 1B-CLL with extreme leukocytosis

presence of uniformly round to somewhat irregular CD5 and CD23 positive

B-cells in the peripheral blood

Page 8: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

In addition: B-CLL with extreme leukocytosis

» The RBC histogram is interfered bylymphocytes and the count falselyelevated.

» Correct the RBC, MCV, HCT and MCHC

Parameter correction:

RBCcorr. = S-RBC = 1.88 x1012/LMCVcorr. = S-MCV = 113.1 fL

HCTcorr. = (S-RBC x S-MCV) :10 = (1.88 x 113.1) : 10 = 21.3 %

MCHCcorr. = (HGB : HCTcorr.) x 100 = (7.0 : 21.3) x 100 = 32.9 g/dL

Conv. units:HGB: 11.3 g/dLMCH: 48.6 pgMCHC: 40.6 g/dL

RBC= S-RBC

WBC= L-RBC

XN Service data

Page 9: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

In addition: ALL patient with „typical“ shark vin

Patient details:19-year-old boy diagnosed with ALL

Page 10: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

Case 2

Page 11: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

Case 2: Initial measurement

Patient details:19-year old man with fever, tired-ness and frequent headaches.

Conv. units:HGB: 13.0 g/dLMCH: 32.38 pgMCHC: 33.0 g/dL

Page 12: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

Case 2: Reflex WPC channel

Patient details:19-year old man with fever, tired-ness and frequent headaches.

Conv. units:HGB: 12.9 g/dLMCH: 32.38 pgMCHC: 31.9 g/dL

Page 13: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

Case 2:Smear results

Differential ResultsSeg neutrophilsLymphocytesMonocytesEosinophilsBasophilsBandsMetamyelocytesMyelocytesPromyelocytesMyeloblasts

Nucleated red blood cells

603134

1

Comments» 60% atypical lymphocytes suspected reactive

Page 14: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

Case 2:Infectious mononucleosis

Comments» 60% atypical lymphocytes suspected reactive

EB VCA G rat +++EB VCA GM rat +++EB NA IgG +++EB VCA IgG +++

Alk phos 100GGT 110ASAT 334ALAT 458LD 622ASLO 233CRP 9.4

Other lab tests

Page 15: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

In addition: European consensus nomenclature

15

Atypicallymphocytes

Atypical cells, suspect reactive

Atypical cells, suspect

neoplastic

Conspicuous lymphatic cells

‘Abnormal lympho?’

Atypical cells, suspect

neoplastic

‘Atypical lympho?’

Atypical cells, suspectreactive

European consensus nomenclature

XN-Series flagging

Analyser flagging

Meaning

Conspicuous lymphatic cells

Page 16: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

In addition: Cell-mediated vs humoral immune response

» Within the adaptive immune response, the cell-mediated immune response is activated during the early stages of an infection while the humoral immune response is typically the sign of a recovering infection.

» Since different cell types are involved in these different stages, this is reflected in the WDF scattergram by two new parameters.

• 16

Th2CD4+ T-cells Plasma cells

Th1CD4+ T-cells

CD8+ T-cells

NK-cells

Adaptive immune response – time line

Cell-mediated immune response

Humoralimmune response

RE-LYMP#/%

AS-LYMP#/%

Page 17: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

• 17

In addition: Extended Inflammation Parameters – launch end 2016*

Parameter Without license With license

RE-LYMP#/% Diagnostic

AS-LYMP#/% Diagnostic

NE-SSC Research NEUT-GI (diagnostic)

NE-SFL Research NEUT-RI (diagnostic)

In case of a lymphocytosis:→ RE-LYMP > 6% without AS-LYMP indicates a cellular immune response→ AS-LYMP presence indicates a humoral immune response

* in combination with WPC channel

Page 18: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

Cell-mediated immune response

Page 19: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

Humoral immune response

Page 20: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

Case 3

Page 21: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

Case 3: Initial measurement

Patient details:A 34-year old man who frequently travels to India presents with fever, nausea, headache and painful joints.

Conv. units:HGB: 7.2 g/dLMCH: 31.7 pgMCHC: 33.7 g/dLRET-He: 32.5 pg

Page 22: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

Case 3Smear results

Differential ResultsSeg neutrophilsLymphocytesMonocytesEosinophilsBasophilsBandsMetamyelocytesMyelocytesPromyelocytesMyeloblasts

Nucleated red blood cells

731552

41

Comments» Schizonts and early schizont stages ++

» 3% atypical lymphocytes suspected reactive

Page 23: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

Case 3Malaria Plasmodium vivax

Page 24: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

In addition:Malaria Plasmodium vivax

» The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count from the WNR channel (WBC-N).

» This is indicative of nucleic acid-containing cellular inclusions in RBC, which do not interfere with the WBC count in the WNR channel due to the strong lysis reagent.

» Since RBC are not completely lysed in the WDF channel, they are visible as an additional neutrophil population.

» ‘IG present’ was triggered due to the wrong clustering and the automated differential is incorrect.

Unusual cell clustering

Page 25: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

In additionMalaria Plasmodium vivax ‚iRBC?‘ flag

» Without the activated ‘ìRBC?’ flag, the reported values can be incorrect, such as false high neutrophil, IG or eosinophil counts, when schizonts, gametocytes and/or trophozoites are present and overlap with these cells.

» Even a small number of infected RBC with these larger forms may have a significant impact on the differential count, as the concentration of RBC is a thousand-fold higher than WBC and infected RBC are not completely lysed in the WDF channel.

» After flag activation:‒ ‘iRBC?’ is triggered and infected RBC are shown in

purple in the WDF scattergram‒ WBC differential is corrected

‚iRBC?‘ not activated ‚iRBC?‘ activated

Page 26: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

Case 4

Page 27: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

Case 4: Patient 1

SFL

Heinz bodies ++++Patient had a spenectomyLab checked HGB by HPLCResult: HGB variant! Name: HGB KölnThe mutated HGB seems to interact (bind?) Fluorocell WDF

Page 28: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

Case 4: Patient 2

Patient details:A blood collection of

13-year boy was initiated by GP

HPLC was done: diagnosed with Hgb Koeln

Conv. units:HGB: 18,1 g/dLMCH: 32.3 pgMCHC: 34.8 g/dL

Page 29: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

Case 4: Patient 3

Patient details:woman * 1944send by the GPno furtherinformation

Conv. units:HGB: 14,5 g/dLMCH: 28,4 pgMCHC: 28,4 g/dL

Page 30: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

Case 4: Patient 3

» 80 year old patient without anaemia or other reasonto think about Hgb variant

» Reason for low SFL inknown

Page 31: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

Similar cases described…» Published about HGB variant Ferrara and variant Leiden

» Both also caused by genetic mutation in HGB gene

» Also strongly reduce SFL in DIFF channel

M. Rosetti, et al., Serendipitous detection of Hemoglobin G-Ferrara variant with Sysmex DIFF channel, Clin Biochem(2015), http://dx.doi.org/10.1016/j.clinbiochem.2015.09.009

“We hypothesize that the polymethine fluorescent dye used for leukocyte differentiation binds some Hbvariants with affinity greater than nucleic acids or reduces theWBC permeability to the dye. In conclusion, the presence of unstable Hbvariants can interfere with the DIFF channel of Sysmex XE-2100 hematology analyzer.”

Page 32: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

Case 5

Page 33: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

Case 5: Initial measurement

Patient details:73-year-old male with known CLL,

diagnosed 12 years ago. Now obvious spleno

megaly and pal-pable lymph nodes

Conv. units:HGB: 6,0 g/dLMCH: 26,2 pgMCHC: 22,4 g/dL

Page 34: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

Case 5Smear results

Differential ResultsSeg neutrophilsLymphocytesMonocytesEosinophilsBasophilsBandsMetamyelocytesMyelocytesPromyelocytesBlasts

Nucleated red blood cells

90

10

Comments10 smudge cells (excl. from the differential count as per recommendation)

Page 35: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

Case 5CLL in combination with AML/acute monoblastic leukaemia

30 % CLL population, CD20+, CD79a+, CD23+ en CD5+2/3 of BM blast CD15+ (partly) CD4+ (weak) CD68+, CD34- and CD117-

BM immunhistology

Page 36: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

Case 6

Page 37: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

Case 5: Initial measurement

Patient details:22-year-old male visited his family in

Nigeria. Back home he has fever andheadache

Conv. units:HGB: 10,5 g/dLMCH: 27,7 pgMCHC: 31,9 g/dL

Page 38: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

Case 4Smear results

Differential ResultsSeg neutrophilsLymphocytesMonocytesEosinophilsBasophilsBandsMetamyelocytesMyelocytesPromyelocytesBlasts

Nucleated red blood cells

6824521

Comments6% atypical lymphocytes suspected reactive19% plasmodium falciparum !

Page 39: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

Case 6Malaria falciparum infection

Page 40: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count

In additionMalaria falciparum infection

» The high amount of plasmodia trigger the flag “Ret abn scattergram.” The reticulocyte count is interfered by the plasmodia and is not possible to use the automated reticulocyte count

» The reticulocyte count can be used for monitoring the therapy. Already after 12 hours therapy is the reticulocyte count (including plasmodia containing RBC) about factor 3 fallen.

» The total amount of PLT is 40 10/3ul. The result was checked by PLT-f. 12.7% of PLT are reticulated (IPF)

Page 41: XN series - Sysmex · » The SSC-FSC scattergram of the WDF channel showed two neutrophil populations and the WBC count from the WDF channel (WBC-D) was much higher than the count