Antti Meriläinen1, Ivan Kassamakov1,2, Kenneth Österberg1,2 and Edward Hæggström1

1) Department of Physics, University of Helsinki

2) Helsinki Institute of Physics

Rapid Ultralow Pressure Manometry by Light

Absorption Spectroscopy

4.2.2014 1 of 10

Page 2: Www.helsinki.fi/yliopisto Antti Meriläinen 1, Ivan Kassamakov 1,2, Kenneth Österberg 1,2 and Edward Hæggström 1 1) Department of Physics, University of

www.helsinki.fi/yliopisto

Motivation

• CLIC main beam vacuum requirements from beam dynamics: H20, CO, CO2 partial pressures < 3*109 mbar

• The RF causes outgassing in CLIC accelerator structures mainly to two mechanism: breakdown and dark current

• Expected pressure rise from simulations: several orders of magnitude on 100 ns time scale. To be verified experimentally.

• Aim: develop method to measure partial pressure changes of ~108 mbar on ~100 ns time scale at a RF test stand (”Xbox”). For the moment focusing on absorption spectroscopy and Cu.

4.2.2014 2 of 10

Page 3: Www.helsinki.fi/yliopisto Antti Meriläinen 1, Ivan Kassamakov 1,2, Kenneth Österberg 1,2 and Edward Hæggström 1 1) Department of Physics, University of

www.helsinki.fi/yliopisto

Test setup

Quartz crystal microbalance (QCM) as reference

l = 5 ± 0.5 cm

R = 25 mm

RQCM = 6.5 mm

4.2.2014 3 of 10

Photo multiplier tube (PMT)

Page 4: Www.helsinki.fi/yliopisto Antti Meriläinen 1, Ivan Kassamakov 1,2, Kenneth Österberg 1,2 and Edward Hæggström 1 1) Department of Physics, University of

www.helsinki.fi/yliopisto

Simulated Visualization

Light Source

QCM

CU source

PMT

4.2.2014 4 of 10

Page 5: Www.helsinki.fi/yliopisto Antti Meriläinen 1, Ivan Kassamakov 1,2, Kenneth Österberg 1,2 and Edward Hæggström 1 1) Department of Physics, University of

www.helsinki.fi/yliopisto

Setup inside vacuum

QCM

Cu vapour source:Cu wire around W wire

1 cm

1 cm

1 mm

l = 5 ± 0.5 cm

4.2.2014 5 of 10

Page 6: Www.helsinki.fi/yliopisto Antti Meriläinen 1, Ivan Kassamakov 1,2, Kenneth Österberg 1,2 and Edward Hæggström 1 1) Department of Physics, University of

www.helsinki.fi/yliopisto

Results

Measurement 1 Measurement 2

Sample rate: 100 ks/s6 kHz low pass filter for PMT1 Hz low pass filter for QCM

Change in intensity ~5%≈ 2 * 1011 absorbances≈ 10-4 mbar

4.2.2014 6 of 10

Page 7: Www.helsinki.fi/yliopisto Antti Meriläinen 1, Ivan Kassamakov 1,2, Kenneth Österberg 1,2 and Edward Hæggström 1 1) Department of Physics, University of

www.helsinki.fi/yliopisto 4.2.2014 7 of 10

Calibration (Relate intensity change to number of atoms)

Cu-source

QCM → ΔM → N1

Light beam ΔI

• atoms was measured by QCM

- Geometrical correction gives

- Calibration gives

N2

∫∆ 𝐼 ⅆ𝑡∫∆𝑀 ⅆ𝑡

Page 8: Www.helsinki.fi/yliopisto Antti Meriläinen 1, Ivan Kassamakov 1,2, Kenneth Österberg 1,2 and Edward Hæggström 1 1) Department of Physics, University of

www.helsinki.fi/yliopisto

Progresst – dP – P Diagram

Target

Measurement 1 & 2

Faster PMT

1 ng/s = 1013 Cu atoms/s ≈ 5∙10-5 mbar 4.2.2014 8 of 10

Page 9: Www.helsinki.fi/yliopisto Antti Meriläinen 1, Ivan Kassamakov 1,2, Kenneth Österberg 1,2 and Edward Hæggström 1 1) Department of Physics, University of

www.helsinki.fi/yliopisto

Achieved:

Measured few ng/s Cu at 10-5 mbar base pressure

Challenges:

Faster response time

Absolute calibration

Interpretation of measurement:

Can one atom have multiple absorptions?

Conclusion

4.2.2014 9 of 10

Page 10: Www.helsinki.fi/yliopisto Antti Meriläinen 1, Ivan Kassamakov 1,2, Kenneth Österberg 1,2 and Edward Hæggström 1 1) Department of Physics, University of

www.helsinki.fi/yliopisto

Thank You!

Questions and Comments

4.2.201410 of 10

Page 11: Www.helsinki.fi/yliopisto Antti Meriläinen 1, Ivan Kassamakov 1,2, Kenneth Österberg 1,2 and Edward Hæggström 1 1) Department of Physics, University of

www.helsinki.fi/yliopisto 11 of 10

Light Source

Setup & Cross section dimension

QCM

Cu source

RLight = 25 mm

RQCM = 6.5 mm

l = 5 ± 0.5 cm

Photo multiplier tube (PMT)

4.2.2014

Page 12: Www.helsinki.fi/yliopisto Antti Meriläinen 1, Ivan Kassamakov 1,2, Kenneth Österberg 1,2 and Edward Hæggström 1 1) Department of Physics, University of

www.helsinki.fi/yliopisto 12 of 10

Absorption spectroscopy

- 324.8 nm and 327.4 nm wavelengths are strongly absorbed in copper vapor

- Light source must be stable- Copper hollow cathode lamp

is suitable low cost light source

- Light intensity changes is measured with Photomultiplier tube (PMT), bandwidth 20 kHz

4.2.2014

Page 13: Www.helsinki.fi/yliopisto Antti Meriläinen 1, Ivan Kassamakov 1,2, Kenneth Österberg 1,2 and Edward Hæggström 1 1) Department of Physics, University of

www.helsinki.fi/yliopisto 13 of 10

Beer-Lamberts law(Light absorbtion in copper vapour)

Cross section for 324.8 nm and 327.4 nm wavelengths

Emitted intensity of 324.8 nm and 327.4 nm wavelengths

length of light beam in Cu vapornumber of absorbing Cu atoms

Current operation area (Signal > |Noise| )I = 90% - 10% lN = 4.9 * 1011 - 1.2 * 1013 Absorbances/meter

Operation point sensitivity: 9 * 1011 Absorbances/meter

4.2.2014