welch allyn microtymp - hillrom

118
4341 State Street Road, Skaneateles Falls, NY 13153 www.welchallyn.com WELCH ALLYN® MICROTYMP® 4 HAND HELD PORTABLE TYMPANOMETER USER MANUAL

Upload: others

Post on 05-Dec-2021

18 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: WELCH ALLYN MICROTYMP - Hillrom

4341 State Street Road, Skaneateles Falls, NY 13153 www.welchallyn.com

WELCH ALLYN® MICROTYMP® 4

HAND HELD PORTABLE TYMPANOMETER

USER MANUAL

Page 2: WELCH ALLYN MICROTYMP - Hillrom

Title: Welch Allyn MicroTymp® 4 Tympanometer User Manual

© 2021 Welch Allyn. All rights are reserved. To support the intended use of the product described in this publication, the

purchaser of the product is permitted to copy this publication, for internal distribution only, from the media provided by Welch

Allyn. No other use, reproduction, or distribution of this publication, or any part of it, is permitted without written permission

from Welch Allyn. Welch Allyn assumes no responsibility for any injury to anyone, or for any illegal or improper use of the

product, that may result from failure to use this product in accordance with the instructions, cautions, warnings, or statement

of intended use published in this manual.

For information about any Welch Allyn product, contact Welch Allyn Technical Support: www.welchallyn.com/about/

company/locations.htm.

93790 This manual applies to 901033 TYMPANOMETRIC INSTRUMENT. GSI D-0122381 Rev. C WA DIR 80024157 Ver. C Revision date: 2021-01 Distributed by Welch Allyn, Inc. 4341 State Street Road Skaneateles Falls, NY 13153-0220 USA www.welchallyn.com

Grason-Stadler 10395 West 70th Street Eden Prairie, MN 55344 USA GSI is an ISO 13485 certified corporation.

Grason-Stadler c/o DGS Diagnostics A/S Audiometer Alle 1 5500 Middelfart Denmark

0123

Caution: US Federal law restricts this device to sale by or on the order of a physician or licensed

hearing care professional.

Page 3: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page i

TABLE OF CONTENTS TABLE OF CONTENTS ....................................................................................................................................................... i

Preface ........................................................................................................................................................................... v

Manual Conventions ............................................................................................................................................................v

Regulatory Symbols ............................................................................................................................................................ vi

Device Symbols ................................................................................................................................................................. viii

Important Safety Instructions ............................................................................................................................................. ix

Precautions ......................................................................................................................................................................... ix

Electromagnetic compatibility (EMC) considerations .......................................................................................................... x

Warranty ............................................................................................................................................................................. xi

Recycling / Disposal ............................................................................................................................................................ xi

Introduction ................................................................................................................................................................... 2

Indication for Use ............................................................................................................................................................... 2

Intended Use ...................................................................................................................................................................... 2

Contraindications ................................................................................................................................................................ 3

Description and Operating Principles ................................................................................................................................. 3

Admittance measurement .............................................................................................................................................. 3

Tympanogram ................................................................................................................................................................. 3

Acoustic reflex measurement ......................................................................................................................................... 3

Installation ..................................................................................................................................................................... 5

External Inspection ............................................................................................................................................................. 5

Unpacking ........................................................................................................................................................................... 5

Standard contents .............................................................................................................................................................. 5

Initial Set Up ....................................................................................................................................................................... 6

Power supply ...................................................................................................................................................................... 6

Cradle Connections ............................................................................................................................................................. 6

Cradle LED Indicators .......................................................................................................................................................... 7

Page 4: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page ii

Handset


 ...................................................................................................................................................................... 8

Handset LED Indicators ....................................................................................................................................................... 9

Handset probe .................................................................................................................................................................... 9

Printer


. ........................................................................................................................................................................ 9

Operation and Configuration ........................................................................................................................................ 10

Start-up and menu displays .............................................................................................................................................. 10

Main Menu Options ..................................................................................................................................................... 10

Configuration .................................................................................................................................................................... 11

Test sequence ............................................................................................................................................................... 13

Ear seal check................................................................................................................................................................ 13

Reflex levels .................................................................................................................................................................. 13

Reflex frequencies ........................................................................................................................................................ 13

Reflex selection ............................................................................................................................................................ 13

Reflex threshold ........................................................................................................................................................... 13

Reflex auto-stop ........................................................................................................................................................... 13

Reflex polarity .............................................................................................................................................................. 14

Reflex filter ................................................................................................................................................................... 14

Set Time/Date ............................................................................................................................................................... 14

Power Off Delay ............................................................................................................................................................ 14

LCD Contrast ................................................................................................................................................................. 14

Report Cal Date............................................................................................................................................................. 14

Set Date Format ............................................................................................................................................................ 14

Hospital Name .............................................................................................................................................................. 14

Department .................................................................................................................................................................. 15

Reload Defaults ............................................................................................................................................................ 15

Language ....................................................................................................................................................................... 15

Data Collection ............................................................................................................................................................. 16

Prior to testing and Ambient conditions ........................................................................................................................... 16

Page 5: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page iii

Ear tips
.. ......................................................................................................................................................................... 16

Performing a test .............................................................................................................................................................. 17

Ear seal check ................................................................................................................................................................... 19

Error messages ................................................................................................................................................................. 23

Saving Results in the Database ..................................................................................................................................... 24

Data entry ......................................................................................................................................................................... 24

Database full ..................................................................................................................................................................... 25

Sending the Results to a Printer .................................................................................................................................... 26

Printing results .................................................................................................................................................................. 26

Data Management ........................................................................................................................................................ 27

List records........................................................................................................................................................................ 27

Delete records .................................................................................................................................................................. 28

Print records ..................................................................................................................................................................... 28

Performing Daily Checks ............................................................................................................................................... 29

Routine Maintenance ................................................................................................................................................... 31

Cleaning the MicroTymp 4 ................................................................................................................................................ 31

Eartip and Probe ............................................................................................................................................................... 31

Calibration and Repair of the Instrument ......................................................................................................................... 32

Error Messages & Fault Conditions ............................................................................................................................... 34

Ordering Consumables and Accessories ........................................................................................................................ 36

Ear Tips – Single Use ......................................................................................................................................................... 36

Appendix - Menu Summary .......................................................................................................................................... 37

Main menu........................................................................................................................................................................ 37

Sub-Menu selections ........................................................................................................................................................ 37

Appendix - Technical Specification ................................................................................................................................ 41

Equipment classification ................................................................................................................................................... 45

Audiometric Standards ..................................................................................................................................................... 45

Appendix - EMC Guidance & Manufacturer’s Declaration ............................................................................................. 46

Page 6: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page iv

Electromagnetic Compatibility ......................................................................................................................................... 46

Electrical Safety, EMC and Associated Standards ............................................................................................................. 46

Appendix - Use with Non-medical Electrical Equipment................................................................................................ 47

Page 7: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page v

PREFACE

This user manual provides information about the Welch Allyn MicroTymp 4 tympanometer. This

manual is intended for technically qualified personnel. Please note: This User Manual is not

intended as a training manual for tympanometry. The reader should consult standard audiology

texts for the theory and application of the screening tests provided by this instrument.

MANUAL CONVENTIONS

Throughout this manual, the following meaning of warnings, cautions and notices are used.

WARNING

The WARNING symbol identifies conditions or practices that may present danger to

the patient and/or user.

CAUTION

The CAUTION Symbol identifies conditions or practices that could result in damage

to the equipment

NOTE: Notes help you identify areas of possible confusion and avoid potential problems

during system operation

Page 8: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page vi

REGULATORY SYMBOLS

Symbol Description

Conforms to European Medical Device Directive 93/42/EEC

Symbol for "SERIAL NUMBER"

Regulatory Product Identifier (RPI) number

Welch Allyn Part Number

Return to Authorized Representative, Special disposal required

Symbol for “European Representative”

Symbol for “Manufacturer”

Symbol for “Date of Manufacture”

Symbol for “Caution”

Type B Applied Part according to IEC 60601-1

Consult Operating Instructions

On/Off - Next to power mains

Keep Dry

Page 9: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page vii

Symbol Description

This side up

Follow Instructions for Use

Consult the operating instructions/directions for use.

A copy of the operating manual is available on this website:

www.welchallyn.com/mt4

A printed copy of the operating instructions can be ordered from

Welch Allyn for shipment within 7 days

Page 10: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page viii

DEVICE SYMBOLS

The following symbols appear on the tympanometer, the instrument cradle or the mains adapter:

Definition: Consult operating instructions.

Definition: Type B applied part – an applied part providing protection against

electric shock, particularly regarding allowable patient leakage current and patient

auxiliary current.

The applied part is the ear tip.

Definition: The output from the mains AC adapter is Direct Current.

Definition: Class II equipment – equipment in which protection against electric

shock does not rely on basic insulation only, but in which additional safety

precautions such as double insulation or reinforced insulation are provided,

there being no provision for protective earth connection or reliance upon

installation conditions.

USB Definition: Industry-standard Type-B USB connection to a computer.

Definition: printer connection.

Page 11: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page ix

IMPORTANT SAFETY INSTRUCTIONS

WARNING

The Welch Allyn MicroTymp 4 instrument must be used only by medical

professionals including, but not limited to, Physicians, Physician Assistants, Nurse

Practitioners, Nurses, Audiologists and Medical Technologists knowledgeable in the

theory and application of the screening tests provided by this instrument. It is intended for

transient use as a screening and diagnostic tool; however no surgical or medical procedure should

be undertaken solely on the basis of results obtained from the instrument.

PRECAUTIONS

READ THIS USER MANUAL BEFORE ATTEMPTING TO USE THE

INSTRUMENT

Users should use their professional skills when interpreting the results and this should be done

in conjunction with other testing as deemed appropriate given their professional skills. Incorrect

use could lead to wrong results.

To comply with the standards IEC 60601-1 for safety and IEC 60601-1-2 for EMC the

tympanometer is designed to be used only with the medically-approved mains adapter supplied,

which is specified as part of the equipment. Do not use any other type of mains adapter with

this instrument.

The tympanometer is for indoor use only and should be used only as described in this manual.

Before the first use of the instrument each day, or if suspect or inconsistent results are apparent,

the checks specified in the Performing Daily Checks section should be carried out. If the system

is not functioning properly, do not operate it until all necessary repairs are made and the unit is

tested and calibrated for proper functioning.

Never insert the probe into a patient’s ear canal without a suitable ear tip fitted to the probe.

Use only the recommended disposable ear tips. These are for single use only - that is, each ear

tip is intended to be used once only for a single ear for a single patient. Do not reuse ear tips as

this will pose the risk of ear-to-ear or patient-to-patient cross infection.

Latex is not used anywhere in the manufacturing process. The base material for the ear tips is

made from silicone rubber.

Page 12: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page x

Do not immerse the unit in any fluids. See the Routine Maintenance Section of this manual for

the proper cleaning procedure for the instrument and its accessories and the function of single-

use parts.

Do not use the instrument in an oxygen-rich environment or in the presence of a flammable

anesthetic mixture or other flammable agents.

Thermal paper printouts fade with exposure to light or heat. Photocopying the patient record

test results will ensure a more permanent record is kept.

Do not drop or otherwise impact this instrument. If the instrument is dropped or damaged, return

it to the manufacturer for repair and/or calibration. Do not use the instrument if any damage is

suspected.

The instrument must be stored and used indoors within the specified temperature, pressure and

humidity ranges.

As with all instruments of this nature the measurements taken will be influenced by significant

changes in elevation and pressure. See Daily Check section for more information.

Do not attempt to open, modify or service the instrument. Return the instrument to the

manufacturer or distributor for all repair and servicing requirements. Opening the instrument

will void the warranty.

This instrument contains a rechargeable Nickel-Metal Hydride (NiMH) battery-pack. The battery

is not intended to be changed by the user. Batteries may explode or cause burns, if disassembled,

crushed or exposed to fire or high temperatures. Do not short-circuit.

ELECTROMAGNETIC COMPATIBILITY (EMC) CONSIDERATIONS

Medical electrical equipment needs special precautions regarding EMC and needs to be installed

and put into service according to the EMC information in the Appendix. This provides guidance

on the electromagnetic environment in which to operate the instrument.

Portable and mobile radio-frequency (RF) communications equipment can affect medical

electrical equipment. The instrument should not be used adjacent to or stacked with other

equipment; if this is unavoidable the instrument should be observed to verify normal operation.

Page 13: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page xi

WARRANTY

We, Welch Allyn, warrant that this product is free from defects in material and workmanship and,

when properly installed and used, will perform in accordance with applicable specifications. If

within one year after original shipment, it is found not to meet this standard; it will be repaired,

or at our option, replaced at no charge except for transportation costs, when returned to an

authorized Welch Allyn facility

NOTE: Changes in the product not approved in writing by Welch Allyn shall void this

warranty. Welch Allyn shall not be responsible for any indirect, special or consequential

damages, even if notice has been given in advance of the possibility of such damages. The

pressure pump and transducers may go out of calibration due to rough handling or impact

(dropping). The lifetime of probe, probe seals and eartips is dependent upon conditions of

use. These parts are only guaranteed against faulty materials or manufacture.

THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING

BUT NOT LIMITED TO, ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A

PARTICULAR PURPOSE.

RECYCLING / DISPOSAL

Directive 2002/96/EC-WEEE: Disposal of noncontaminated electrical and electronic equipment

Many local laws and regulations require special procedures to recycle or dispose of electrical

equipment-related waste including batteries, printed circuit boards, electronic components,

wiring and other elements of electronic devices. Follow all your respective local laws and

regulations for the proper disposal of batteries and any other parts of this system. Do not dispose

of this product as unsorted municipal waste. Prepare this product for reuse or separate collection

as specified by Directive 2002/96/ EC of the European Parliament and the Council of the

European Union on Waste Electronic and Electrical Equipment (WEEE). If this product is

contaminated, this directive does not apply.

For specific disposal or compliance information contact, contact Welch Allyn Technical Support.

Page 14: WELCH ALLYN MICROTYMP - Hillrom

4341 State Street Road, Skaneateles Falls, NY 13153 www.welchallyn.com

INTRODUCTION

Thank you for purchasing a Welch Allyn MicroTymp 4, a hand-held, portable tympanometer that

will give many years of reliable service if treated with care. The instrument performs two types

of measurement:

Tympanometry is used to measure the admittance of the tympanic membrane and middle ear

at a fixed frequency over a range of pressures.

Acoustic Reflex tests are used to measure stapedial reflexes. The MicroTymp 4 measures

ipsilateral reflexes and, when selected, reflex measurement is automatically carried out after a

tympanogram is taken.

Features

Automatic measurement of ear canal volume, tympanic admittance peak, placement of

the peak and the gradient

Automatic detection of stapedial reflexes

Up to 32, dual-ear patient tests can be stored in non-volatile memory

Configurable settings for user preferences, held in non-volatile memory

Printout of data to a printer

English, German, French, Spanish, Portuguese or Italian operating language (selectable by

the user)

INDICATION FOR USE

The Welch Allyn MicroTymp 4 is intended to be used for the measurement of acoustic

impedance/admittance within the human external ear canal. These measures are useful in the

evaluation, identification, documentation and diagnosis of ear disorders. The device is intended

to be used on patients of any age.

INTENDED USE

The Welch Allyn MicroTymp 4 is an auditory impedance tester intended to detect possible

otologic disorders associated with the functioning of the middle ear. It is intended to be used in

a hospital, clinic or other healthcare facility with a suitable quiet testing environment such as a

private exam room.

Page 15: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 3

CONTRAINDICATIONS

Ear canal examination with an illuminated otoscope is an essential prerequisite to successful

middle-ear testing. Make sure that the canal is free of any obstruction. If the canal is completely

plugged at the entrance or if fluid is running from the ear canal, tympanometry should not be

attempted until the condition is cleared. Testing should not be performed on patients with

conditions listed below without a medical doctor’s approval.

Recent stapedectomy or other middle ear surgery

Discharging ear

Acute external auditory canal trauma

Discomfort (e.g. severe otitis externa)

Presence of tinnitus, hyperacusis or other sensitivity to loud sounds may contraindicate testing when high intensity stimuli are used

DESCRIPTION AND OPERATING PRINCIPLES

The Welch Allyn MicroTymp 4 is clinical aural acoustic impedance/admittance instrument (Type

2). The main components of the instrument consist of a hand held unit with an LCD and a probe

assembly and a cradle. A printer, eartips and test cavity are included with the system.

The probe contains one microphone, two receivers and an air channel. One of the receivers is

used for probe tone signal. The second receiver is used for the acoustic reflex stimulus signal.

The microphone measures the response. The air channel is connected to the pump system which

makes it possible to supply the eardrum with air pressure

ADMITTANCE MEASUREMENT

The MicroTymp 4 measures the admittance of the tympanic membrane and middle ear by playing

a continuous 226Hz tone into the ear canal at a level calibrated to give 85dB SPL into a 2ml cavity.

The sound level this produces in the ear canal is measured using a microphone and the

admittance calculated from the result. In line with normal audiometric practice admittance is

displayed as an equivalent volume of air in ml.

TYMPANOGRAM

To record the tympanogram the admittance is measured while the air pressure in the ear canal

is varied from +200daPa to -400daPa by means of a small pump. The admittance peaks when the

air pressure is the same on both sides of the tympanic membrane. The changing admittance with

pressure is displayed as a graph.

ACOUSTIC REFLEX MEASUREMENT

Using the same principle, it is also possible to establish whether an acoustic reflex is present. In

this case, the 226Hz tone is used to measure the admittance of the ear, while a short tone at a

Page 16: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 4

different frequency is presented (the reflex stimulus). The sound pressure level (SPL) of this

stimulus is increased in steps until the middle ear muscles respond causing the tympanic

membrane to become stiffer, or a preset maximum SPL is reached. When the change in

admittance exceeds a predetermined threshold, this constitutes a reflex and the change in

admittance at that level when the stimulus is applied is displayed as a plot against time.

The acoustic reflex is measured at the static ear canal pressure that produces the maximum

membrane admittance, so reflex measurements are taken after the tympanogram is measured

when the peak admittance pressure has been established.

The MicroTymp 4 can measure an acoustic reflex at any combination of 500Hz, 1000Hz, 2000Hz

and 4000Hz. The maximum level for the reflex stimulus may be preset, along with the step size

in dB between the three preceding lower levels of stimulus.

Page 17: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 5

INSTALLATION

EXTERNAL INSPECTION

Although this Welch Allyn MicroTymp 4 was carefully tested, inspected, and packed for shipping,

it is good practice after receiving the instrument to immediately examine the outside of the

container for any signs of damage. Notify the carrier if any damage is observed.

UNPACKING

Please retain the carton and packaging as the tympanometer will need calibrating on an annual

basis and should be returned to the distributor or Welch Allyn in its original shipping carton.

Please check the contents of the shipping carton against the delivery note to make sure that all

items ordered have been included. If anything is missing, please contact the distributor who

supplied the tympanometer or Welch Allyn.

STANDARD CONTENTS

MicroTymp 4 handset (P/N 93701)

MicroTymp 4 charging cradle (P/N 93710)

Power supply (P/N 93715)

4 in 1 calibration test cavity (P/N 93750)

Eartip/Probe Tip Starter kit (P/N 93720)

Probe Floss Cleaning Kit (P/N 93730)

User Manuals (on USB Thumb Drive) (P/N 93790-X)

USB cable (A/B 2 meters) (P/N 39414)

Serial Printer Cable (P/N 39771)

Calibration certificate

MPT-II Printer Set (P/N 39410) Includes, MPT-II printer, battery, power supply/battery

charger and printer paper (Not included with 93700-NP)

Page 18: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 6

INITIAL SET UP

Place the cradle on a stable counter or table where it will be used. The location should be near a

properly grounded wall outlet. When placing the handset in the cradle make sure that the

connectors on the handset and cradle align.

POWER SUPPLY

The Welch Allyn MicroTymp 4 tympanometer is designed for continuous operation and is

powered by a rechargeable Nickel-Metal Hydride (NiMH) battery-pack which is fitted in the

instrument. If the instrument is placed onto its cradle the battery within it will be charged.

The mains adapter is supplied and specified as part of the equipment. Connect the output lead

from the adapter into the power socket on the rear of the instrument cradle. Switch on the mains

supply - the indicator on the adapter will illuminate green. The mains adapter is the mains

disconnect device and therefore the tympanometer should be positioned such that easy access

to the mains adapter is possible.

The output from the mains adapter is fitted with electronic circuit protection. In case of overload

the adapter will shut down and the indicator will be off. When the fault is cleared the adapter

will operate as normal.

The input to the mains adapter is protected with a non-replaceable fuse. If this fails, the adapter

will not operate and will need to be replaced. If a replacement mains adapter is required, please

contact your Welch Allyn distributor.

CRADLE CONNECTIONS

The cradle connections are labeled to ensure correct identification and connection as follows:

Socket Label Socket Type Connected Part

Page 19: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 7

RJ6 socket Supplied printer *

5V 0.2A

2.5mm power jack Mains AC/DC Adapter *

USB

USB connector

Type B

Computer (via USB port)

WARNING

For connected parts marked * only connect the parts or accessories supplied with

the instrument or supplied by Welch Allyn or a Welch Allyn distributor. These parts

have been tested for use with the Welch Allyn MicroTymp 4 tympanometer for

compliance with the standards IEC 60601-1 and IEC 60601-1-2. The use of accessories other than

those specified may compromise compliance with these standards.

CRADLE LED INDICATORS

The LED indicators on the instrument cradle show the status of the mains connection and the

battery charging.

LED displays green when power is applied to the cradle;

otherwise it will be off.

LED shows green when the handset is in the cradle and its

internal battery pack is charging; it will be off when the

handset is removed.

Page 20: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 8

HANDSET

Press the On/Off key momentarily to turn the Welch Allyn MicroTymp 4 on (refer to the diagram

above). No warm-up time is required, although a short diagnostic routine will run for a few

seconds. During this time the internal pump will operate. To switch off, again press and hold the

On/Off key for a few seconds.

Press the up ▲ and down â–Œ navigation keys to scroll through the menus or set values

Press the right navigation key ► to accept a menu choice or go to the next step.

Press the left navigation key ◄ to cancel an operation or go back to the previous step.

The function of the left and right keys is usually shown on the bottom line of the display.

When not located in the cradle and not performing a test the Welch Allyn MicroTymp 4 will switch

off automatically if no key is pressed for 90 seconds. This time may be extended to 180 seconds

in the CONFIGURATION menu.

LCD

Display

On/Off

Button

Navigation

Buttons

LED

Indicators

Probe

Page 21: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 9

HANDSET LED INDICATORS

The indicators on the instrument body show the status of the system. Typical indications during a

measurement sequence are as follows:

Green

Indicator

Yellow

Indicator

Status

Off Off MicroTymp 4 turned off

On Off Idle & ready to use

Off Slow flash Waiting for probe to be inserted

Slow flash Off Taking a measurement

HANDSET PROBE

The probe tip must be fitted with a new ear tip before it is presented to a patient’s ear canal. The

ear tip must be fitted completely to the probe tip and must not occlude any of the four holes in

the probe tip

PRINTER

The Welch Allyn MicroTymp 4 can be supplied with portable thermal printer for printing

tympanometric test results. Upon receipt of the printer it must be initially charged prior to use.

Refer to the printer instructions for further details. Printing is from the cradle connected to the

printer via the supplied serial cable.

WARNING

Please refer to Appendix - Use with Non-Medical Electrical Equipment for

important information regarding the connection of non-medical electrical

equipment to medical electrical equipment.

Page 22: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 10

OPERATION AND CONFIGURATION

Prior to performing tests with the Welch Allyn MicroTymp 4, the system should be properly

configured. Set the values for the time and date to ensure that test data and calibration status

are correctly identified. These values along with the instrument language and preferences for

the parameters used in testing are set in the CONFIGURATION menu.

START-UP AND MENU DISPLAYS

When the Welch Allyn MicroTymp 4 is turned on, the start-up screen is shown while internal tests

are performed, and the pump is initialized. When the start-up sequence is complete the MAIN

MENU is displayed. The LCD display shows the first 3 menu items with the highlight on the first

item in the menu.

MAIN MENU

NEW TEST

CONFIGURATION

VIEW THE LAST TEST

Select

A battery state indicator is shown in the top right corner of the display (except when showing

test results). This shows the battery state as a progressively emptying battery. The battery-pack

should be recharged when the symbol has a “!” in front of it, or when advised to do so when the

instrument is switched on.

Press the down â–Œ and up ▲ navigation keys to scroll through the menu.

MAIN MENU OPTIONS

NEW TEST

CONFIGURATION

VIEW THE LAST TEST

DAILY CHECK

DATA MANAGEMENT

SYSTEM INFORMATION

Page 23: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 11

Press the down ▌ navigation keys to scroll through the menu until CONFIGURATION is

highlighted and then press the right navigation key ► to select.

CONFIGURATION

The configuration menu contains 17 items with the values and defaults indicated in the table

below. Select and change the items as necessary to set up your device before you begin testing.

The settings are retained in memory after the unit is turned off.

Configuration Item

(Sweep Settings)

Value Options Default Value

Test Sequence Both: L, R

Both: R, L

Both: R, L

Ear Seal Check Standard or Extended Standard

Reload Defaults

(Sweep Settings)

Yes or No No

Configuration Item

(Reflex Settings)

Value Options Default Value

Reflex Levels 100 dB/10 dB Steps

100 dB/5 dB Steps

95 dB/5 dB Steps

90 dB/5 dB Steps

85 dB/5 dB Steps

95 dB/5 dB steps

Reflex Frequencies 500 Hz, 1k, 2k, & 4kHz (individually

selectable)

1 kHz

Reflex Selection Always Measure

Never Measure

Only If Peak Found

Prompt To Measure

Only if Peak Found

Reflex Threshold 0.01 to 0.5 ml 0.03 (ml)

Reflex Auto Stop Yes or No Yes

Page 24: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 12

Reflex Polarity Up or Down Down

Reflex Filter 2 Hz or 1.5 Hz 2 Hz

Reload Defaults

(Reflex Settings)

Yes or No No

Configuration Item

(System Settings)

Value Options Default Value

Set Time/Date Date and time formatted selections

– individual values for MM/DD/YY

and HH:MM:SS

Date currently set

Power Off Delay 90 or 180 seconds 90 seconds

LCD Contrast (Change using Up & Down keys) Mid-range

Report Cal. Dates Print or Hide Print

Date Format DD/MM/YY or MM/DD/YY DD/MM/YY

Hospital Name A-Z, -, 0-9 (max length of 19) Blank

Department A-Z, -, 0-9 (max length of 19) Blank

Reload Defaults

(System Settings)

Yes or No No

Language English, German, French, Spanish,

Portuguese, Italian

English

Configuration Item

(Reload Defaults)

Value Options Default Value

Reload Defaults (All

Configuration

Settings)

Yes or No No

Page 25: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 13

TEST SEQUENCE

Use the ▲ and â–Œ keys to choose the order to be used for a both-ear test. Select either L, R (left

then right) or R, L (right then left). Press the ► key to confirm the selection or the ◄ key to

cancel.

EAR SEAL CHECK

Use the ▲ and â–Œ keys to choose the type of ear seal check employed at the start of a test. The

default STANDARD option is adequate for most circumstances, and this checks that an adequate

pressure can be created in the ear canal before starting the test.

However, if difficulty is experienced in using the eartips to create a seal the alternative EXTENDED

option may be helpful. This checks that a range of pressures will be available before starting a

test by means of a visual indication of the quality of the seal. Press the ► key to confirm the

selection or the ◄ key to cancel.

REFLEX LEVELS

Use the ▲ and â–Œ keys to choose the maximum level of reflex stimulus to apply and the step size

between the levels of the preceding stimuli. The maximum level of stimulus may be set between

85dBHL & 100dBHL with a step size of 5dB (plus the option for 10dB step size at 100dBHL). Press

the ► key to confirm the selection or the ◄ key to cancel.

REFLEX FREQUENCIES

Use the ▌ key to scroll through the frequencies available for the ipsilateral reflex stimulus

(500Hz, 1000Hz, 2000Hz & 4000Hz), and then the ▲ key to select or deselect the frequencies at

which this stimulus is to be applied. Press the ► key to confirm the selection.

REFLEX SELECTION

Use the ▲ and â–Œ keys to choose the circumstances when a reflex measurement is to be made

(always, never, only if an admittance peak is found, or only after confirmation is made at the start

of the test sequence). In cases where an admittance peak has not been established a pressure

of 0daPa is used. Press the ► key to confirm the selection or the ◄ key to cancel.

REFLEX THRESHOLD

Use the keys to choose the change in admittance that determines that a reflex has been detected

(0.01ml to 0.5ml). Use the ▲ and â–Œ keys to change the values and press the ► key to confirm

and save the selection or the ◄ key to cancel.

REFLEX AUTO-STOP

By default, the reflex test at each frequency will stop at the lowest level of stimulus that produces

a response. By setting REFLEX AUTO-STOP to NO the MicroTymp 4 will test for a reflex at all

Page 26: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 14

selected levels. Press the ► key to confirm the selection or the ◄ key to cancel. (Note that

100dBHL at 4000Hz is not available).

REFLEX POLARITY

Use the ▲ and â–Œ keys to choose whether the reflex traces are displayed as ascending (UP) or

descending (DOWN). Press the ► key to confirm the selection or the ◄ key to cancel.

REFLEX FILTER

Use the keys to choose either 2Hz or 1.5Hz. The default of 2Hz is suitable for most circumstances.

However, if a smoother reflex plot is required for better interpretation 1.5Hz may be chosen.

Press the ► key to confirm the selection or the ◄ key to cancel.

SET TIME/DATE

Use the keys to enter the values for the date and time. Use the ▲ and â–Œ keys to change the

values. Press the ► key to confirm and save the selection or the ◄ key to cancel.

POWER OFF DELAY

The Welch Allyn MicroTymp 4 will switch off automatically if no key is pressed for a specified

duration. Use the ▲ and â–Œ keys to change this duration between 90 and 180 seconds and press

the ► key to confirm and save the selection or the ◄ key to cancel.

LCD CONTRAST

Use the ▲ and â–Œ keys to change the contrast of the LCD screen; press the ► key to confirm and

save the selection or the ◄ key to cancel.

REPORT CAL DATE

The printout of the test results may include date of the instrument’s calibration. Use the ▲ and

â–Œ keys to select if the calibration date is printed or hidden. Press the ► key to confirm and save

the selection or the ◄ key to cancel.

SET DATE FORMAT

The Welch Allyn MicroTymp 4 supports two different date formats. Use the ▲ and â–Œ keys to

select either DD/MM/YY or MM/DD/YY and press the ► key to confirm and save the selection or

the ◄ key to cancel.

HOSPITAL NAME

The printout of the test results may include the hospital name (up to 19 characters). To enter

the hospital name use the ▲ and â–Œ and ◄ and ► keys to select the letter then press and briefly

hold the ► key to confirm. To delete the last letter briefly hold the ◄ key. Once the name has

been entered highlight the # key then press and briefly hold the ► key to save the name.

Highlight the # key then press and briefly hold the ◄ key to cancel.

Page 27: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 15

DEPARTMENT

The printout of the test results may include the department name (up to 19 characters). To enter

the department name, use the ▲ and â–Œ and ◄ and ► keys to select the letter then press and

briefly hold the ► key to confirm. To delete the last letter briefly hold the ◄ key. Once the

name has been entered highlight the # key then press and briefly hold the ► key to save the

name. Highlight the # key then press and briefly hold the ◄ key to cancel.

RELOAD DEFAULTS

The settings for the device may be returned to the factory defaults. The Sweep, Reflex or System

settings may be returned separately to the factory defaults or all the configurations settings at

once. Use the ▲ and â–Œ keys to select either YES (reloads defaults) or NO (keep existing settings).

Press the ► key to confirm and save the selection or the ◄ key to cancel.

LANGUAGE

The Welch Allyn MicroTymp 4 supports multiple languages. To set the operating language

(English, German, French, Spanish, Portuguese or Italian) use the ▲ and â–Œ keys to select the

language. Press the ► key to confirm and save the selection or the ◄ key to cancel.

Page 28: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 16

DATA COLLECTION

WARNING

Ensure that the appropriate settings have been made before carrying out a test.

See the information below and the CONFIGURATION options in the previous

section.

PRIOR TO TESTING AND AMBIENT CONDITIONS

A qualified health care professional should perform a thorough otoscopic examination to

establish that the condition of the ear is suitable for the test options selected and that no

contraindications are present. The latter would include obstruction of the external ear canal due

to excessive wax and/or hairs, both of which would need to be removed.

Tympanometric and reflex testing should always be performed in a quiet room or in an acoustic

booth.

EAR TIPS

These must be selected and fitted by a practitioner qualified to perform tympanometric tests.

WARNING

The probe tip must be fitted with a new ear tip before it is presented to a patient’s

ear canal. The ear tip must be fitted completely to the probe tip and must not

occlude any of the four holes in the probe tip. The ear tip size is chosen to suit the

patient’s ear and provide a comfortable pressure seal.

Page 29: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 17

PERFORMING A TEST

Testing should be conducted in a quiet environment such as a private examination room. No

specific action is required by the patient during the automatic test. However, the patient must

be advised to remain still and avoid speaking or swallowing while the probe is applied to the ear.

A typical tympanogram measurement and reflex test is carried out as follows.

From the MAIN MENU select NEW TEST:

MAIN MENU

NEW TEST

CONFIGURATION

VIEW THE LAST TEST

Select

Select the ear(s) required for test:

SELECT EAR

BOTH: L, R

LEFT

RIGHT

Back Select

The message “Deleting last test” will be displayed momentarily and a message displayed to insert

the probe into the ear to be tested:

Page 30: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 18

TESTING LEFT EAR

INSERT PROBE

Cancel

Place the ear tip into the ear canal to obtain a seal and the following messages will be displayed:

TESTING LEFT EAR

Equalizing Pressure

Cancel

TESTING LEFT EAR

Pressure Settling

Cancel

Page 31: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 19

EAR SEAL CHECK

The type of ear seal check employed at the start of a test may be set in the CONFIGURATION

menu. The default STANDARD option is adequate for most circumstances, and this checks that

an adequate pressure can be created in the ear canal before starting the test.

However, if difficulty is experienced in using the eartips to create a seal the alternative EXTENDED

option may be helpful. This checks that a range of pressures will be available before starting a

test by means of a visual indication of the quality of the seal:

TESTING LEFT EAR

Obtaining ear seal

Low :

High :

Cancel

The number of bars shown indicates the robustness of the seal. The probe should be adjusted in

the ear until two or more bars are shown for Low and High.

Once an adequate seal is detected the following message will be seen and a tympanogram

measurement is made.

TESTING LEFT EAR

Seal Obtained

Taking Tympanogram

Cancel

Page 32: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 20

Taking a tympanogram takes about 3 seconds. It is important not to move the probe and to ask

the patient to remain very still during the test.

When the tympanogram is complete the instrument will perform the reflex test(s), if selected.

By default, this test is only performed if a peak is found in the tympanogram. This and other

reflex test options may be changed in the CONFIGURATION menu.

Before starting the reflex test the ear canal pressure will be set to the value that gave the peak

admittance during the tympanogram test. The instrument will then step through the tone

frequencies and levels set in the CONFIGURATION menu searching for a reflex response:

TESTING LEFT EAR

Seeking Ipsi Reflex

1000 Hz 80 dB

Cancel

When the measurement is complete withdraw the probe and the tympanogram will be displayed:

The display shows:

The peak admittance, in ml (Pk)

The pressure which gave the peak admittance in daPa

Page 33: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 21

The Gradient, in daPa (Gr)

The Ear Canal Volume (ECV) in ml measured at 200 daPa

A plot of admittance against pressure

The normalized rectangle showing the ideal location for the tympanogram peak

Review the tympanogram to ensure that the peak admittance point selected by the MicroTymp

4 is suitable. If required, it is possible to select an alternative peak using the ▲ and â–Œ keys. The

figures displayed will change to reflect the peak selected and will be saved with the

tympanogram.

To repeat the test, press ◄.

When satisfied with the tympanogram press ►.

If the reflex test was carried out the results will now be displayed:

The display shows:

The frequency of the reflex stimulus

“” if a reflex was found, otherwise “X”

The lowest level of tone (dBHL) at which a reflex was found

A trace of the admittance change against time

If the reflex test was performed at a single frequency use the ▲ and â–Œ keys to view the results

for each of the reflex tone levels used. If the reflex test was performed at more than one

frequency use the ▲ and â–Œ keys to view the results for the other frequencies.

If the MicroTymp 4 was set to test for a reflex at all levels of the stimulus (see Reflex Autostop)

press ► to view an additional display following the reflex graphs. This shows a summary of the

levels and frequencies at which a reflex was detected. The dash symbol “-” is shown if a reflex

tone was not presented at the level indicated.

Page 34: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 22

REFLEX SUMMARY

dB

100 x -

90 x

80 x

70 x x x

Hz 500 1k 2k 4k

Press ◄ to return and view the tympanogram, reflex results or to repeat the test. When satisfied

with the results press ►.

The message “Saving as last test” will be displayed briefly and the results will be saved in the “last

test” memory. The results will remain available until a new test is started, even if the MicroTymp

4 is turned off.

If both ears were chosen for test the entire sequence will now be repeated for the right ear:

TESTING RIGHT EAR

INSERT PROBE

Cancel Skip

Press ► to skip testing of the right ear and view results for the left ear. Press ◄ to return to the

main menu.

When the selected ears have been tested and the results saved the PROCESS RESULTS menu will

be displayed. This accesses the following functions:

PRINT (Print the results)

Page 35: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 23

SAVE RESULTS (Save the results in the internal database)

VIEW TEST (Review the results as described above)

MAIN MENU (Return to the main menu)

The results of the last test performed remain available even if the MicroTymp 4 has been turned

off. To view these results select VIEW THE LAST TEST from the main menu. After selecting the

required ear the tympanogram will be displayed. It will then be possible to view the results and

select the PROCESS RESULTS menu as if the test had just been completed.

NOTE: Results of the last test will be erased as soon as a new test is started. Test results

should be saved to the internal database or printed to ensure that data is not lost.

ERROR MESSAGES

The following error messages may be seen during the test sequence.

Message Displayed Indicator

Status Likely Cause(s)

WITHDRAW PROBE Yellow

Flashing

The probe has been moved during measurement.

Re-insert the probe to repeat the test.

Volume outside range

WITHDRAW PROBE

Yellow

Flashing

The ear canal volume is above the 5ml. This

message can also occur when the probe is not

properly inserted into the ear.

Blocked ear

WITHDRAW PROBE

Green

Flashing

The ear canal volume is below 0.1ml. Check that

the probe is correctly inserted into the ear. Also

check that the probe is not blocked.

INSERT PROBE Yellow

Flashing

The seal was lost. Reinsert the probe to repeat

the test.

Page 36: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 24

SAVING RESULTS IN THE DATABASE

To save the results of a test select SAVE RESULTS from the PROCESS RESULTS menu that is

displayed on completion of a test. This option can also be accessed by selecting VIEW THE LAST

TEST from the main menu and scrolling through the results using the ► key as long as the test

results have not already been saved or deleted (e.g. by starting and then aborting a new test).

A three-character identifier is used for the record. This is also used as the reference for the

patient’s name on the printed record and for data transferred to a computer. The identifier

would typically be the patient’s initials, and as the tympanometer uses a combination of this

identifier and the date/time of a test to refer to stored records this same identifier may be used

for different tests for the same patient.

DATA ENTRY

PATIENT INITIALS

________

ABCDEFGHIJKLM

NOPQRSTUVWXYZ

- 0123456789

Hold to Enter / Cancel

To enter the identifier:

Use the ▲, â–Œ, ◄ and ► keys to select a character.

Press and hold the ► key to enter the selected character.

Press and hold the ◄ key to delete the last character.

To save the test results:

Enter all three characters for the identifier.

Press and hold the ► key to save the record.

Page 37: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 25

To cancel saving the last test:

Delete any characters that have been entered.

Press and hold the ◄ key.

DATABASE FULL

A warning will be displayed if the database is full when attempting to save a test:

MEMORY IS FULL

MANAGE RECORDS

OVERWRITE OLDEST

Back Select

Selecting MANAGE RECORDS will display the DATA MANAGEMENT menu which provides options

for printing or transferring data to a computer prior to deleting records to make space for the

new test.

OVERWRITE OLDEST will overwrite the oldest record in memory with the results being stored.

Back will return to the previous menu.

Page 38: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 26

SENDING THE RESULTS TO A PRINTER

The MPT-II printer is available as an option for use with the MicroTymp 4. Printing is by a cable

connecting the printer to the instrument cradle. Before attempting to print ensure the printer is

fully charged, switched on, loaded with paper and ready to print. If the MicroTymp 4 is in the

cradle the data will be sent via the connecting cable. This operation is carried out automatically,

although reference should be made to the appropriate guidance notes below.

Connect the printer to the MicroTymp 4 cradle using the supplied cable. With the device located

in the cradle print the required data.

PRINTING RESULTS

To print the results of the last test select SEND TO PRINTER from the PROCESS RESULTS menu on

completion of the test. (Similar facilities for printing are available from the VIEW THE LAST TEST

and DATA MANAGEMENT options in the MAIN MENU.)

The following display is then presented:

PROCESS RESULTS

PRESS KEY WHEN READY

Back Select

Press ► when the printer is ready.

Once the print operation has been carried out the PROCESS RESULTS menu is displayed.

Page 39: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 27

DATA MANAGEMENT

Up to 32 patient records can be stored in the database of the Welch Allyn MicroTymp 4. Records

can be listed, viewed, deleted, printed or sent to a computer using the DATA MANAGEMENT

option of the main menu.

DATA MANAGEMENT

LIST RECORDS

DELETE RECORDS

PRINT RECORDS

Back Select

LIST RECORDS is used to work with the record of an individual test. All other options operate on

groups of records.

LIST RECORDS

LIST RECORDS shows the number of records stored and maximum number of records that can be

stored and shows the saved tests, 6 at a time, most recent first.

RECORDS STORED: 15/32

ABC 09/29/16 09:43 L

123 09/28/16 15:05 2

KSM 09/28/16 14:22 2

BEN 09/28/16 12:11 2

KAM 09/28/16 10:15 2

LOL 09/27/16 16:03 2

Back Select

Page 40: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 28

Each entry shows:

Three-letter patient identifier entered when the test was stored;

Date and time of the test

Whether the test has been printed ( )

Whether the test has been sent to a computer ( )

Whether the test is for the Left (L), Right (R) or both (2) ears

Press ▲ or â–Œ to scroll through the records. Press ► to select the highlighted record

Press ◄ to return to the previous menu.

When a record is selected the PROCESS RECORD menu will be displayed. This accesses the

following functions.

View the selected record

Print the selected record

Delete the selected record

DELETE RECORDS

DELETE RECORDS allows a group of records to be deleted. It is possible to delete all records, all

records that have been printed or all records that have been sent to a computer. Confirmation

of the deletion is required.

PRINT RECORDS

PRINT RECORDS allows a group of records to be sent to the printer. It is possible to print all stored

records or just those records that have not already been printed. If printing the entire database,

it is recommended that a full roll of paper is loaded into the printer.

Page 41: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 29

PERFORMING DAILY CHECKS

The operation of the MicroTymp 4 should be checked daily using the 4 in 1 test cavity assembly

supplied with the instrument.

Select the DAILY CHECK option in the main menu:

DAILY CHECK

INSERT PROBE

Cancel

Wait until “INSERT PROBE” is displayed.

Insert the probe, without an ear tip, into the hole at the 2ml end of the test cavity. Make sure

that the probe is pushed fully home and is held tight against the stop. The probe must be square

to the end of the test cavity.

When measured at elevations below 1,000 ft, the display should show the volume of the 2ml test

cavity to within ± 0.1ml.

DAILY CHECK

Volume: 2.0 ml

Cancel

Remove the probe and repeat the test with the three remaining test cavities. When tested at

elevations below 1,000 ft, the display should show the volume of the 0.2ml, & 0.5ml test cavities

Page 42: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 30

to within ± 0.1ml. The volume of the 5.0ml test cavity should be shown within ± 0.25ml. When

the checks have been completed press ◄ to return to main menu.

TEST CAVITY READINGS AT ELEVATIONS GREATER THAN 1,000FT ABOVE SEA LEVEL

The instrument is a pressure sensitive device that makes measurements relative to ambient air

pressure. Changes in air pressure due to weather or elevation will affect the ECV readout of the

instrument. Slight weather related barometric pressure changes will usually yield volume

readouts with ± 0.1 ml of the expected cavity value, but barometric pressure changes due to

elevation can be more significant. These changes in pressure do not affect the accuracy of the

compliance measurement system in any way. However, it will affect the ECV and Test Cavity

values.

If the MicroTymp 4 is being used at an elevation greater than 1,000 ft, when new, and after

each recalibration, perform a Daily Check on all 4 test cavity volumes (0.2ml, 0.5ml, 2ml & 5ml).

Record the displayed values for each cavity and use these values are “normal” when

performing Daily Checks each day thereafter.

Elevation (above sea level) Resulting 2.0 ml reading

Feet Meters

0 0 2.0 ±0.1

1000 304.8 2.1 ±0.1

2000 609.6 2.2 ±0.1

3000 914.4 2.2 ±0.1

4000 1219.2 2.3 ±0.1

5000 1524.0 2.4 ±0.1

6000 1828.8 2.5 ±0.1

7000 2133.6 2.6 ±0.1

8000 2438.4 2.7 ±0.1

Page 43: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 31

9000 2743.2 2.8 ±0.1

10,000 3048.0 2.9 ±0.1

ROUTINE MAINTENANCE

CLEANING THE MICROTYMP 4

WARNING Electric shock hazard. Before cleaning the device, disconnect the power cord

from the power source and the device.

WARNING Take care to prevent water or other fluid from entering any connectors on

the device. Should this occur, dry the connectors and check the accuracy of all operating

functions.

CAUTION The device is not heat-resistant. Do not autoclave.

The MicroTymp 4 is a precision instrument. Handle it carefully to ensure its continued accuracy

and service. Use a soft damp cloth and mild detergent to clean the instrument panel and case

when required. Ensure no moisture enters the instrument.

If low-level disinfection is required, Oxivir Tb (Diversey, Inc.) a hydrogen-peroxide based solution,

was tested and found to be compatible with the plastic device housing.

EARTIP AND PROBE

WARNING

Handle the probe and accessories with care. Do not allow moisture, condensation,

fluids or debris to enter the probe.

Page 44: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 32

Ear tips should be replaced after a single use.

The probe tip and its associated sealing washer are disposable devices. The probe tip should be

checked before each ear insertion to ensure it is undamaged and that none of the tubes through

it are blocked. It should be replaced if necessary.

The small holes through the probe tip must be kept clear. If these become blocked a warning

message will be displayed. The tip must be removed and cleaned or replaced.

To remove the tip, unscrew the nose cone and pull the tip off the probe boss. A small seal will

be found in the base of the probe tip. This should be examined and replaced if it is damaged. Do

not remove the nut securing the boss to the body of the instrument.

CAUTION

The sealing washer should be replaced when the probe tip is replaced if it shows

signs of wear, or if a pressure leak is suspected. When replacing the probe tip,

ensure that the seal is correctly located with the flat side aligned with the flat side

within the base of the probe tip. Push the probe tip over the boss and replace the nose cone.

Make sure that the nose cone is screwed home firmly but do not over-tighten. Do not use any

tools to tighten the nose cone.

After replacing the tip, a Daily Check should be carried out.

CALIBRATION AND REPAIR OF THE INSTRUMENT

Welch Allyn recommends that the MicroTymp 4 is calibrated annually. Please contact your Welch

Allyn distributor for details.

Page 45: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 33

WARNING

The instrument should be returned to the Welch Allyn distributor for service and

repair. There are no user-serviceable parts within it.

When packing the instrument for shipping, please use the original shipping carton and packing

materials. Place the instrument in a plastic bag before packing to prevent dirt and dust getting

into the probe.

Page 46: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 34

ERROR MESSAGES & FAULT CONDITIONS

CAUTION

If a fault condition cannot be cleared, the operator is cautioned against repeatedly

starting the instrument. In some fault conditions the internal pump may

progressively advance towards the end of its travel to clear the fault. If the end of

travel is reached in such conditions the instrument may lock up and become un-usable.

If difficulties resolving fault conditions occur the equipment distributor should be consulted.

Message Meaning / Action

PROBE NOT CLEAR

Please ensure the probe is not

blocked or obstructed

Examine the probe tip for blockages. If necessary,

remove it and clean or replace it. If the problem

persists, contact your Welch Allyn service center.

AIRFLOW ERROR

Unknown pump fault. Restart the

unit. If problem persists, contact

Welch Allyn

WARNING! CALIBRATION EXPIRED.

Recalibration needed before further

tests are performed

The current date is later than the next calibration

date. Check that the clock is set to the correct date. If

so, arrange for the instrument to be recalibrated.

Tests can still be performed.

“WARNING! BATTERIES LOW.

Recharge the batteries before

performing tests

Recharge the batteries immediately

Powering down Other than after the specified power off delay, the

MicroTymp 4 may turn off because the internal

batteries are spent. To replace the batteries, contact

your Welch Allyn service center.

AIRFLOW ERROR.

Cannot determine pump direction. If

problem persists, contact Welch

Allyn

Pump fault. If the fault persists, contact your Welch

Allyn service center.

Page 47: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 35

“WARNING! DEVICE UNCALIBRATED.

One or more default values require

recalibration before further tests are

performed

This message should never normally be seen. If it

persists, contact your Welch Allyn service center.

WARNING! DEFAULTS RELOADED.

Default configuration settings

reloaded. Check before making new

tests

This message should never be seen. Check all the

CONFIGURATION settings before taking any

measurements. If the error persists, contact your

Welch Allyn service center.

WITHDRAW PROBE The probe has been moved during measurement. Re-

insert the probe to repeat the test.

Volume outside range

WITHDRAW PROBE

The ear canal volume is above the 5ml. This message

also occurs when the probe is not properly inserted

into the ear.

Blocked probe

WITHDRAW PROBE

The ear canal volume is below 0.1ml. This message

also occurs when the probe tip is blocked. Check that

the probe is correctly inserted into the ear. Check

that the probe is not blocked.

INSERT PROBE The seal was lost. Reinsert the probe to repeat the

test.

Page 48: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 36

ORDERING CONSUMABLES AND ACCESSORIES

To order consumables, additional accessories and to replace detachable parts that have been

damaged, please contact Welch Allyn or your Welch Allyn distributor for current prices and

delivery charges. Some of the items available are listed below:

Part Number Description

93710 MICROTYMP 4 CHARGING CRADLE

93715 MICROTYMP 4 CHARGING CRADLE POWER SUPPLY

93720 MICROTYMP 4 EARTIP/PROBE TIP STARTER KIT

93730 MICROTYMP 4 PROBE FLOSS CLEANING KIT

93740 MICROTYMP 4 PROBE TIP AND GASKET KIT

93750 MICROTYMP 4 TEST CAVITY

93760 MICROTYMP 4 CARRY CASE

93790-X MICROTYMP 4 DFU, QUICK START GUIDE & SOFTWARE (USB THUMB DRIVE)

X = latest revision number.

39414 OAE & MICROTYMP 4 USB CABLE

39410 MPT-II PRINTER SET

39407 MPT-II PRINTER POWER SUPPLY

39416 MPT-II REPLACEMENT BATTERY

39412 MPT-II PRINTER PAPER, SINGLE ROLL

EAR TIPS – SINGLE USE

Part Number

25/Box

Part Number

100/Box Description

39422-07-025 39422-07-100 7 mm mushroom style disposable ear tips

39422-08-025 39422-08-100 8 mm mushroom style disposable ear tips

39422-09-025 39422-09-100 9 mm mushroom style disposable ear tips

39422-10-025 39422-10-100 10 mm mushroom style disposable ear tips

Page 49: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 37

39422-11-025 39422-11-100 11 mm mushroom style disposable ear tips

39422-12-025 39422-12-100 12 mm mushroom style disposable ear tips

39422-13-025 39422-13-100 13 mm mushroom style disposable ear tips

39422-14-025 39422-14-100 14 mm mushroom style disposable ear tips

39422-15-025 39422-15-100 15 mm mushroom style disposable ear tips

39422-19-025 39422-19-100 19 mm mushroom style disposable ear tips

APPENDIX - MENU SUMMARY

Default values are shown in bold where appropriate.

MAIN MENU

Menu Sub-menu

MAIN MENU NEW TEST

CONFIGURATION

VIEW THE LAST TEST

DAILY CHECK

DATA MANAGEMENT

SYSTEM INFORMATION

SUB-MENU SELECTIONS

Sub-menu Option Choices / Description

NEW TEST SELECT EAR Choose which ear(s) to test and start the

test. A tympanogram is taken followed by

reflex measurements, if selected. On-

screen messages and indicators show

progress. Graphical displays are shown

automatically at the end.

Page 50: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 38

CONFIGURATION

(SWEEP SETTINGS)

TEST SEQUENCE Select the test order for a both-ear test -

left then right or right then left.

EAR SEAL CHECK Select STANDARD or EXTENDED.

RELOAD DEFAULTS The options in this group are reset to their

default values

CONFIGURATION

(REFLEX SETTINGS)

REFLEX LEVELS Select the maximum tone level and step

size to be used for the reflex test. Default

is 95dBHL with 5dB steps.

REFLEX FREQUENCIES Selectable from 500, 1000, 2000 and 4000

Hz.

REFLEX SELECTION ALWAYS MEASURE

NEVER MEASURE

ONLY IF PEAK FOUND

PROMPT TO MEASURE

REFLEX THRESHOLD Default is 0.03 ml

REFLEX AUTO-STOP Default is YES.

REFLEX POLARITY Choose whether a reflex trace is shown

ascending (UP) or descending (DOWN).

REFLEX FILTER Select either 2 Hz or 1.5 Hz.

RELOAD DEFAULTS The options in this group are reset to their

default values.

CONFIGURATION

(SYSTEM SETTINGS)

SET DATE/TIME Set the internal clock date and time.

POWER-OFF DELAY The time before the unit turns off

automatically if no key is pressed. Select 90

or 180 seconds.

LCD CONTRAST Use the UP/DOWN arrow keys to change

the display contrast.

Page 51: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 39

REPORT CAL. DATES Select PRINT CAL. DATES or HIDE

CAL.DATES

SET DATE FORMAT Select DD/MM/YY or MM/DD/YY

HOSPITAL NAME Allows the Hospital name to be entered

(this will appear at the top of the print out).

DEPARTMENT Allows the Department name to be entered

(this will appear at the top of the print out).

RELOAD DEFAULTS The options in this group are reset to their

default values.

SELECT LANGUAGE Select ENGLISH, GERMAN, FRENCH,

SPANISH, PORTUGUESE or ITALIAN for

operating language.

CONFIGURATION

(RELOAD DEFAULTS)

All configuration options are reset to their

default values

VIEW THE LAST TEST SELECT EAR Recalls the last stored test for the selected

ear. Shows the tympanogram and reflex

responses, if available. Also allows the last

test to be printed or saved in the internal

database

DAILY CHECK Shows the volume in ml measured by the

probe.

DATA MANAGEMENT LIST RECORDS Lists the test results stored in the internal

database. Allows individual records to be

viewed, printed or deleted.

Page 52: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 40

DELETE RECORDS Delete stored records. Select:

ALL PRINTED RECORDS – Delete all records

that have been printed.

ALL SENT RECORDS – Delete all records that

have been sent to a computer.

ALL RECORDS – Delete all records

PRINT RECORDS Print stored records. Select:

UNPRINTED RECORDS – Print all records

not previously printed.

ALL RECORDS – Print all records

SYSTEM INFORMATION Displays: Battery voltage

Date calibrated

Date of next calibration

Instrument serial number

Software version

Current date and time

Page 53: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 41

APPENDIX - TECHNICAL SPECIFICATION

Tympanometry

Instrument type Meatus compensated tympanometer

Analysis performed Admittance peak level (in ml); Pressure of same;

Gradient (in daPa);

Ear Canal Volume (ECV) @ 200 daPa

Probe tone levels and accuracy 226Hz +/- 2%; 85dB SPL +/-2dB over range 0.2ml

to 5ml

Pressure levels and accuracy +200daPa to -400daPa +/-10daPa or +/-10%

(whichever is larger) over range

Ear volume measurement range and

accuracy

0.2ml to 5ml +/- 0.1ml or +/-5% (whichever is

larger) over entire range

Sweep speed Typically 200daPa/sec; dependent on ear/cavity

volume

Pressure limits (safety cutout) +600 to -800 daPa

Number of samples stored 100 per tympanogram

Reflex measurements

Measurement modes Ipsilateral

Reflex tone levels and accuracy 500Hz, 1kHz, 2kHz, 4kHz (+/-2%)

Configurable over range 70dB to 100dBHL (4kHz

restricted to 95dBHL) +/-3dB, referenced to 2ml

calibration volume; Compensates for measured

ear volume

Reflex detection threshold and accuracy 0.01ml to 0.5ml +/-0.01ml configurable in 0.01ml

steps

Page 54: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 42

Number of reflex levels (see Acoustic

Reflex Measurement)

Four: 100dB with 5dB or 10 dB steps; 95dB, 90dB

or 85dB with 5 dB steps

Reflex analysis Reflex pass/fail at each level tested; maximum

amplitude of each reflex (seen on printed report

& computer report); pressure at which reflex was

performed

Pressure used for reflex measurement Pressure at Tympanogram peak, or 0 daPa

Reflex level cut-off Optionally, Auto-stop when reflex found

Reflex tone duration 0.6 seconds

Data Management

Number of records stored in Patient

Database

32

Data storage Any recording can be stored once the

tympanogram is viewed. Patient Initials (A-Z, 0-9,

“-“) must be entered before storage.

Data held Patient Initials, Tympanogram and Reflex graphs

and analysis for Left Ear and/or Right Ear, Time

and Date of recording, which ears were tested,

whether or not the record has been printed

and/or sent to a computer, parameters used for

analysis, 128 bit Globally Unique Identifier (GUID)

Display mode Records listed in reverse chronological order

(latest first), with indication of data stored as

described above

Real Time Clock

Time stamps Time and date stamp applied to all recordings,

and to the last calibration date

Page 55: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 43

Languages

Operating Languages English, German, French, Spanish, Portuguese or

Italian

Printing

Supported printer MPT-II

Interface Wired connection to cradle

Information printed Space for patient & clinician’s details,

Tympanogram analysis parameters,

Tympanogram, Reflex analysis parameters, Reflex

graph, Serial Number of device, Last and Next

Due Calibration dates

Interface to computer

Interface USB Version 1.1

Information sent Patient header, left and right ear data

Power Supply

Battery NiMH rechargeable battery pack.

Mains power (to cradle) 100-240Vac; 50/60Hz; 0.2A

Warm-up period None at room temperature

Number of recordings with full charge Up to 100

Auto power-off delay 90 or 180 seconds

Idle current 70mA

Current while testing 230mA

Page 56: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 44

Physical

Display 128 x 64 pixels / 8 lines of 21 characters

Dimensions 230mm (L) x 115mm (W) x 70mm (H)

Total Weight (handset and cradle) 650g

Environmental

Operating temperature range +15oC to +35oC

Operating humidity range 30% to 90% RH, non-condensing

Operating atmospheric pressure range 980 to 1040 mb

Transport and storage temperature

range

+5oC to +40oC

Transport and storage humidity range 30% to 90% RH, non-condensing

Transport and storage atmospheric

pressure range

900 to 1100 mb

Standards conformance

Safety IEC 60601-1 (plus UL, CSA & EN deviations)

EMC IEC 60601-1-2

Performance IEC 60645-5, Type 2 Tympanometer

CE mark To the EU Medical Device Directive

Reflex HL RETSPL

500 Hz 5.5 dB

1000 Hz 0 dB

2000 Hz 3 dB

4000 Hz 5.5 dB

Page 57: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 45

EQUIPMENT CLASSIFICATION

The Welch Allyn MicroTymp 4 Tympanometer is classified as a Class IIa device under Annex IX

(Section 1) of the EU Medical Devices Directive.

Type of protection against electric shock Internally Powered

Degree of protection against electric shock Type B applied part

Degree of protection against ingress of water Not protected

Mode of operation Continuous operation

Equipment mobility Portable

AUDIOMETRIC STANDARDS

The Welch Allyn MicroTymp 4 Tympanometer is designed to meet or exceed the Aural

Impedance/Admittance Instrument Standard Requirements - Type 2 listed below.

ANSI S3.39 Specification for Instruments to measure Aural Acoustic Impedance and Admittance

(Aural Acoustic Immittance)

IEC 60645-5 Electroacoustics - Audiometric Equipment – Instruments for the measurement of

aural acoustic impedance/admittance

ISO 389-2 Reference Equivalent Threshold SPLS for Pure Tones and Insert Earphones

Page 58: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 46

APPENDIX - EMC GUIDANCE & MANUFACTURER’S DECLARATION

Portable and Mobile RF communications equipment can affect the Welch Allyn MicroTymp 4.

Install and operate the Welch Allyn MicroTymp 4 according to the EMC information presented in

this appendix and in EMC Tables available at www.welchallyn.com/emc-mt4.

The Welch Allyn MicroTymp 4 has been tested for EMC emissions and immunity as a standalone

instrument. Do not use the device adjacent to or stacked with other electronic equipment. If

adjacent or stacked use is necessary, the user should verify normal operation in the configuration.

Portable RF communications equipment (including peripherals such as antenna cables and

external antennas) should be used no closer than 30 cm (12 inches) to any part of the MicroTymp

4, including cables specified by the manufacturer. Otherwise, degradation of the performance of

this equipment could result.

The use of accessories, transducers and cables other than those specified, with the exception of

servicing parts sold by Welch Allyn as replacement parts for internal components, may result in

increased EMISSIONS or decreased IMMUNITY of the device. Anyone connecting additional

equipment is responsible for making sure the system complies with the IEC 60601-1-2 standard.

ELECTROMAGNETIC COMPATIBILITY

Although the instrument fulfils the relevant EMC requirements precautions should be taken to

avoid unnecessary exposure to electromagnetic fields, e.g. from mobile phones, etc. If the device

is used adjacent to other equipment it must be observed that no mutual disturbance appears.

ELECTRICAL SAFETY, EMC AND ASSOCIATED STANDARDS

UL 60601-1: Medical Electrical Equipment, Part 1 General Requirements for Safety

IEC/EN 60601-1: Medical Electrical Equipment, Part 1 General Requirements for Safety

CAN/CSA-C22.2 No. 60601-1: Medical Electrical Equipment, Part 1 General Requirements for

Safety Electrical Equipment for Laboratory Use

IEC/EN 60601-1-1: Collateral Standard, Safety Requirements for Medical Electrical Systems

IEC/EN 60601-1-2: Medical Electrical Equipment, Part 1 - Electromagnetic Compatibility -

Requirements and Tests

Essential Requirements of the current European Union Medical Device Directive 93/42/EEC

RoHS (Restriction of the use of certain Hazardous Substance)

WEEE (Waste Electrical & Electronic Equipment) Legislation

Page 59: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 47

APPENDIX - USE WITH NON-MEDICAL ELECTRICAL EQUIPMENT

Any person who connects external equipment to signal input, signal output or other connectors

has created a medical electrical system and is therefore responsible for the system complying

with the requirements of clause 16 of IEC 60601-1(General requirements for basic safety and

essential performance).

If connections are made to standard equipment such as printers and computers, special

precautions must be taken in order to maintain medical safety. The following notes are provided

for guidance in making such connections to ensure that the general requirements of clause 16 of

IEC 60601- are met.

The following signal inputs and outputs on the Welch Allyn MicroTymp 4 tympanometer are

electrically isolated to the requirements of IEC 60601-1:

Socket Label Socket Type Typical Connection

USB USB connector

Type B

Computer

RJ6 socket Supplied printer

These measures are incorporated to reduce any potential hazard associated with the use of

mains-powered equipment connecting to these interfaces.

External equipment intended for connection to signal input, signal output or other connectors,

shall comply with the relevant IEC or international standards (e.g. IEC 60950, CISPR 22 & CISPR

24 for IT equipment, and the IEC 60601 series for medical electrical equipment).

Equipment not complying with IEC 60601 shall be kept outside the patient environment, as

defined in IEC 60601-1:(at least 1.5m from the patient).

The operator must not touch the connected equipment and the patient at the same time as this

would result in an unacceptable hazard.

Refer to Welch Allyn at the address given on the front of this user manual if advice is required

regarding the use of peripheral equipment.

Page 60: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 48

WELCH ALLYN® MICROTYMP® 4

手持ち型携垯甚ティンパノメヌタ聎力枬

定噚

取扱説明曞

Page 61: WELCH ALLYN MICROTYMP - Hillrom

DIR 80024157 Ver C Page 49

タむトルWelch Allyn MicroTymp® 4 Tympanometer User Manual

© 2021 Welch Allyn.著䜜暩所有。本補品の賌入者には、本曞に蚘茉された補品本来の甚途通りの䜿甚を支揎するために、内郚で配垃する堎合に限り、Welch

Allynが提䟛する媒䜓から本曞をコピヌするこずが蚱可されおいたす。Welch Allynからの曞面による蚱可なく、本曞たたはその䞀郚を他の甚途で䜿甚するこず、たたは耇補や配垃を行うこずを犁じたす。Welch

Allynは、この取扱説明曞に蚘茉されおいる指瀺、泚意、譊告、たたは蚘茉された䜿甚目的に埓っお本補品を䜿甚しなかったために生じた任意の怪我、たたは補品の違法たたは䞍適切な䜿甚に぀いおの責任を䞀切負いたせん。

Welch Allyn補品に぀いおの情報は、Welch Allynテクニカルサポヌトたでお問い合わせください: www.welchallyn.com/about/ company/locations.htm.

93790 この取扱説明曞は、 901033 ティンパノメヌタ機噚に適甚されたす。 GSI D-0122381 Rev. C WA DIR 80024157 Ver. C 改蚂日2021-01 発行元Welch Allyn, Inc. 4341 State Street Road Skaneateles Falls, NY 13153-0220 USA www.welchallyn.com

Grason-Stadler 10395 West 70th Street Eden Prairie, MN 55344 USA GSI is an ISO 13485 certified corporation.

Grason-Stadler c/o DGS Diagnostics A/S Audiometer Alle 1 5500 Middelfart Denmark

0123

泚意米囜連邊法においお、この装眮は、医垫たたは公認の聎芚専門家の指瀺による販売に限定されおいたす。

Page 62: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page i

目次 目次 .................................................................................................................................................................................... i

序文 ................................................................................................................................................................................... v

説明曞䞊の衚蚘 v

芏制甚暙識 vi

装眮䞊の暙識 viii

安党に関する指瀺 ............................................................................................................................................................................ ix

䜿甚前の泚意事項 ............................................................................................................................................................................ ix

電磁䞡立性 (EMC) に関する 考慮事項 ...................................................................................................................................................... xi

保蚌 ............................................................................................................................................................................ xii

リサむクル廃棄 ........................................................................................................................................................................... xii

はじめに ............................................................................................................................................................................... 1

䜿甚の適応 1

䜿甚目的 1

犁忌 ............................................................................................................................................................................. 2

説明および操䜜原理 .......................................................................................................................................................................... 2

アドミッタンス枬定 ...................................................................................................................................................................... 2

ティンパノグラム聎力図 .............................................................................................................................................................. 2

音響反射枬定 2

むンストヌル ........................................................................................................................................................................... 4

倖芳怜査 4

開梱 ............................................................................................................................................................................. 4

梱包物䞀芧 4

I初期蚭定 5

電源 ............................................................................................................................................................................. 5

クレヌドル の接続 ........................................................................................................................................................................... 5

クレヌドルの接続には、次のように正しい識別ず接続を保蚌するためのラベルが付いおいたす。 ...................................................................................................... 5

Page 63: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page ii

クレヌドルの LED むンゞケヌタヌ .......................................................................................................................................................... 6

ハンドセット 7

ハンドセットのLED むンゞケヌタヌ ......................................................................................................................................................... 8

ハンドセットのプロヌブ ...................................................................................................................................................................... 8

プリンタヌ 8

操䜜ず蚭定 ............................................................................................................................................................................. 9

起動およびメニュヌ衚瀺 ...................................................................................................................................................................... 9

メむンメニュヌのオプション .............................................................................................................................................................. 9

蚭定 ............................................................................................................................................................................. 9

怜査手順 12

むダヌシヌルの確認 .................................................................................................................................................................... 12

反射レベル 12

反射の呚波数 12

反射の遞択 12

反射の閟倀 12

反射の自動停止 12

反射の極性 12

反射フィルタヌ 12

時間日付の蚭定 ...................................................................................................................................................................... 13

電源オフ遅延 13

LCDコントラスト .................................................................................................................................................................... 13

范正日の報告 13

日付フォヌマットを蚭定する ............................................................................................................................................................ 13

病院名 ....................................................................................................................................................................... 13

蚺療科名 13

デフォルトのリロヌド .................................................................................................................................................................. 13

蚀語 ....................................................................................................................................................................... 13

デヌタ収集 ........................................................................................................................................................................... 15

Page 64: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page iii

怜査前の環境条件 .......................................................................................................................................................................... 15

むダヌチップ 15

怜査を実斜する 16

むダヌシヌルの確認 ........................................................................................................................................................................ 17

゚ラヌメッセヌゞ .......................................................................................................................................................................... 22

デヌタベヌスに結果を保存する .......................................................................................................................................................... 22

デヌタ入力 23

デヌタベヌスが満杯になった時 .............................................................................................................................................................. 24

結果をプリンタヌに送信する ............................................................................................................................................................ 25

結果を印刷する 25

デヌタ管理 ........................................................................................................................................................................... 26

蚘録を䞀芧にする .......................................................................................................................................................................... 26

蚘録を削陀する 27

蚘録を印刷する 27

日垞点怜を実斜する .................................................................................................................................................................... 28

海抜1,000FT以䞊の暙高での怜査容噚の読み取り ..................................................................................................................................... 28

定期メンテナンス ...................................................................................................................................................................... 29

MICROTYMP 4の掃陀 ................................................................................................................................................................ 29

むダヌチップずプロヌブ .................................................................................................................................................................... 30

機噚の范正および修理 ...................................................................................................................................................................... 31

゚ラヌメッセヌゞず゚ラヌ状態 .......................................................................................................................................................... 33

消耗品および付属品の泚文 .............................................................................................................................................................. 35

むダヌチップス 䜿い捚お ................................................................................................................................................................. 35

付録 - メニュヌのたずめ ............................................................................................................................................................... 36

[メむンメニュヌ].......................................................................................................................................................................... 36

サブメニュヌ遞択 .......................................................................................................................................................................... 36

付録 - 技術仕様 ....................................................................................................................................................................... 39

蚭備の分類 42

Page 65: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page iv

聎力枬定基準 42

付録 - EMCガむダンスず補造元の公衚事項 ............................................................................................................................................ 44

電磁䞡立性 44

電気の安党性、EMCおよび関連芏栌 ...................................................................................................................................................... 44

付録 - 医療甚以倖の電気機噚ずの䜵甚 .................................................................................................................................................... 44

Page 66: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page v

序文

この取扱説明曞には、Welch Allyn MicroTymp 4ティンパノメヌタに぀いおの情報が蚘茉されおいたす。

この取扱説明曞は、技術的な資栌保持者を察象に䜜成されおいたす。 ご泚意この取扱説明曞は、ティンパノメトリヌのトレヌニングマニュアルではありたせん。

この機噚が提䟛するスクリヌニング怜査の理論ず応甚に぀いおは、暙準的な聎芚孊のテキストを確認しおください。

説明曞䞊の衚蚘

本説明曞党般に、譊告、危険事項、泚意事項の以䞋の意味が䜿甚されおいたす

è­Šå‘Š

[è­Šå‘Š] の暙識は、患者および/たたはナヌザヌに危険をおよがす可胜性がある状態や行為を瀺しおいたす。

泚意

[泚意] の暙識は、機噚に損傷を䞎える可胜性のある状態や行いを瀺しおいたす。

泚蚘: [泚蚘] は、システム操䜜䞭に起こる可胜性のある取り違えの領域を瀺し、朜

圚的な問題を避けるのに圹立ちたす。

Page 67: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page vi

芏制甚暙識

暙識 内容

欧州医療機噚指什93/42/EECに準拠。

「シリアル番号」の暙識。

WELCH ALLYN 郚品番号。

認可を受けた担圓者に返华。特別廃棄が必芁。

「欧州担圓者」の暙識。

「補造者」の暙識。

「補造幎月日」の暙識。

「泚意」の暙識。

IEC 60601-1に準拠するタむプB適甚郚品。

操䜜指瀺を調べおください。

オンオフ – 電源に隣接

氎濡れ厳犁。

こちら偎が䞊です。

Page 68: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page vii

暙識 内容

䜿甚説明曞に埓っおください

操䜜説明曞/䜿甚説明曞を参照しおください。

操䜜説明曞のコピヌは、次のWebサむトで入手できたすwww.welchallyn.com/mt4

操䜜説明曞の印刷版は、Welch Allynに発泚し7日以内に発送しおもらうこずができたす。

Page 69: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page viii

装眮䞊の暙識

以䞋の暙識が、ティンパノメヌタ、機噚クレヌドル、電源アダプタヌ䞊に衚瀺されおいたす。

定矩 操䜜指瀺を調べおください。

定矩タむプB適甚郚品  特に蚱容可胜な患者挏れ電流および患者枬定電流に関する電気ショックに察しお、保護を提䟛する適甚郚品。

適甚郚品はむダヌチップです。

定矩電源ACアダプタヌからの出力は盎流です。

定矩クラス II機噚 

電気ショックに察する保護が基本の絶瞁に頌らず、二重絶瞁や匷化絶瞁等、远加の安党察策がなされおる機噚で、保護アヌスたたは蚭眮条件ぞの

䟝存の芏定はありたせん。

USB 定矩コンピュヌタヌぞの産業基準タむプBUSB 接続。

定矩 プリンタヌ接続。

.

Page 70: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page ix

安党に関する指瀺

è­Šå‘Š

Welch Allyn MicroTymp

4機噚は、本機噚により提䟛されるスクリヌニング怜査の理論ず応甚の知識が豊富な医垫、医垫助手、正看護垫、看護垫、聎芚蚓緎士、医療技術者を含むがこれらに限定

されない医療の専門家によっおのみ、䜿甚されなければなりたせん。

本機噚はスクリヌニングおよび蚺断ツヌルずしお䞀時的な䜿甚を目的ずしおいたすが、本機噚から埗られた結果のみに基づいた倖科的たたは内科的凊眮を実斜しおはいけ

たせん。

䜿甚前の泚意事項

本機噚の䜿甚を詊す前に、本取扱説明曞を読んでください。

ナヌザヌは、結果の解釈の際にその専門的な技術を䜿う必芁があり、その専門的な技術によっお適切であるず芋なされる他の怜査ず合わせお行う必芁がありたす。䞍適切

な䜿甚は間違った結果に぀ながる可胜性がありたす。

安党面で IEC 60601-1 を、たた EMC (電磁䞡立性) 面では IEC 60601-1-

2を順守するために、ティンパノメヌタは機噚の䞀郚ずしお指定されおいる、医療承認電源アダプタヌ䟛絊のみず䜵せお䜿甚されるよう蚭蚈されおいたす。他のいかなる

タむプの電源アダプタヌも、本機噚ず䞀緒に䜿甚しないでください。

ティンパノメヌタは屋内のみの䜿甚のためのものであり、本説明曞に説明されおいる通りにのみ䜿甚しなければなりたせん。

毎日、機噚を最初に䜿甚する前に、たたは疑わしいか䞀貫性のない結果が明らかな堎合は、「日々の確認を実斜する」セクションに指定の確認を行わなければなりたせん

。 システムが正しく機胜しおいない堎合は、必芁な修理をすべお行い、 Welch

Allyn瀟発衚の仕様に埓っお適切に機胜するようにナニットをテストし范正するたでは、システムを操䜜しないでください。

プロヌブに合った適切なむダヌチップなしでは、絶察にプロヌブを患者の倖耳道に挿入しないでください。

掚奚の䜿い捚おむダヌチップのみを䜿甚しおください。単発䜿甚のみのものであり、各むダヌチップは䞀人の患者の片方の耳のみの䜿甚を察象ずしおいたす。むダヌチッ

プの再䜿甚は、耳から耳ぞの感染たたは患者間の亀差感染のリスクを高めるため、行わないでください。

ゎム補品は、補造過皋のどこにも䜿甚されおいたせん。むダヌチップの基瀎資材はシリコンゎムでできおいたす。

ナニットはいかなる液䜓にも浞さないでください。本機噚および付属品の適切な掗浄凊眮、たた単発䜿甚郚品の機胜に぀いおは、本説明曞の「定期メむンテナンス」セク

ションを参照しおください。

酞玠が豊富な環境、たたは可燃性麻酔薬混合物あるいは他の可燃性薬剀がある堎所では、本機噚を䜿甚しないでください。

感熱玙の印刷物は、光や熱にさらすず色萜ちしたす。患者蚘録テスト結果をコピヌするず、より氞久的な蚘録を確実に保぀こずができたす。

Page 71: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page x

本機噚を萜ずしたり、その他の衝撃を䞎えないでください。機噚が萜ちたり砎損した堎合、補造者に返华し、修理およびたたは范正を行っおください。なんらかの損傷

が疑われる堎合は、本機噚を䜿甚しないでください。

本機噚は宀内で、特定の枩床・気圧・湿床の範囲内で保管および䜿甚しなければなりたせん。

このような特性の党おの機噚ず同様に、枬定結果は暙高および気圧の倧きな倉化に圱響されたす。Welch Allyn MicroTymp

4ティンパノメヌタは意図した操䜜暙高にお再范正しなければなりたせん。

本機噚の䞭を開けたり、倉曎したり修理したりしようずしないでください。本機噚の党おの修理および点怜芁件には、補造者たたは販売業者に返华しおください。機噚の

䞭を開けるず、保蚌が無効になりたす。

本機噚には充電可胜のニッケル氎玠 (NiMH) 電池パックが含たれおいたす。

この電池はナヌザヌが倉曎するように䜜られおいるものではありたせん。電池は、分解したり壊したり、火や高枩にさらすず、爆発したり火傷を招く堎合がありたす。シ

ョヌトさせないでください。

Page 72: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page xi

電磁䞡立性 (EMC) に関する 考慮事項

医療電子機噚はEMCに関しお特別な泚意が必芁であり、付録のEMC情報に埓っお蚭眮し皌働させる必芁がありたす。これは、本機噚を操䜜する電磁環境䞊でのガむ

ダンスを提䟛したす。

携垯および移動タむプの無線呚波数 (RF)

通信機噚は、医療電子機噚に圱響を䞎える可胜性がありたす。本機噚は他の機噚に隣接させたり積み重ねたりしお䜿甚しおはいけたせん。これが避けられない堎合は、本

機噚を芳察し、正垞に操䜜できるよう確認しおください。

Page 73: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page xii

保蚌

Welch

Allyn瀟は、本補品が資材および仕䞊がりにおいお党く欠陥がないこず、たた適切に蚭眮し䜿甚した際は該圓する仕様の通りに䜜動するこずを保蚌いたしたす。

最初の発送から幎以内にこの暙準に満たないこずが分かった堎合は、正芏の Welch Allyn 瀟

の斜蚭に返华いただければ、修理を行うか、匊瀟のオプションにより、茞送費を陀き無料で亀換いたしたす。

珟地でのサヌビスをご垌望の堎合は、人件費や資材費は無料ですが、サヌビスセンタヌでの珟時点のレヌトでの出匵費がかかりたす。

泚蚘Welch Allyn 瀟による曞面での承認がない補品の倉曎は、この保蚌を無効ずしたす。Welch Allyn

瀟は、いかなる間接的たたは特別な、たたは掟生的な損害にも、たずえ圓該の損害の可胜性に぀いお事前に通知されおいた堎合でも、責任を負わないものずしたす

。 圧力ポンプおよびトランスデュヌサヌは乱暎な扱いたたは衝撃 (萜䞋) により范正から倖れる堎合がありたす。

プロヌブ、プロヌブシヌル、むダヌチップの耐甚期間は䜿甚の状態によりたす。 このような郚品は、資材たたは補造の欠陥に察しおのみ保蚌されたす。

この保蚌は、明瀺か黙瀺かを問わず、商品性、特定目的ぞの適合性の保蚌を含め、他の党おの保蚌に代わり、それらを排陀するものです。

リサむクル廃棄

WEEE (電気電子機噚廃棄物) 指什 非汚染の電気および電子機噚の廃棄

バッテリヌ、プリント回路基板、電子郚品、配線、その他の電子機噚゚レメントなどの電子機噚関連廃棄物のリサむクル、たたは凊分に぀いおは、各地方自治䜓が専甚の

手順芏定を蚭けおいたすバッテリヌやこのシステムの任意の郚品は、各地方自治䜓の定める法芏に埓っお適切に廃棄しおください。

本補品を、分別されおいない䞀般廃棄物ずしお凊分しないでください。電気電子機噚廃棄物 (WEEE) に関する欧州議䌚および欧州連合理事䌚指什

2002/96/ECに指定されおいるように、本補品を再利甚するか、分別収集甚に準備しおください。

この補品が汚染されおいる堎合、この指什は適甚されたせん。

特定の廃棄手順やコンプラむアンス情報に぀いおは、Welch Allynテクニカルサポヌトにお問い合わせください。

Page 74: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 1

はじめに

Welch Allyn MicroTymp

4瀟の手持ち型携垯甚ティンパノメヌタ聎力枬定噚賌入いただきありがずうございたす。本装眮は、䞁寧な扱いをしおいただければ、䜕幎もの信頌できるサヌビスを

ご提䟛したす。 本機噚は以䞋の2タむプの枬定を実斜したす。:

ティンパノメトリヌ聎力枬定 は、䞀定範囲の圧力に枡っお䞀定の呚波数で錓膜および䞭耳のアドミッタンスを枬定するために䜿甚されたす。

音響反射怜査 は、アブミ骚筋反射を枬定するために䜿甚されたす。 MicroTymp

4は同偎性反射を枬定し、遞択されおいる堎合、反射枬定はティンパノグラム聎力図の実斜埌に自動的に実斜されたす。

特城

倖耳道の容積、錓膜のアドミッタンスピヌク、ピヌクおよび募配の配眮

アブミ骚筋反射の自動怜出

最倧32の患者の䞡耳怜査を䞍揮発性メモリに保存可胜

䞍揮発性メモリに保持された、ナヌザ環境構成可胜な蚭定

プリンタヌでのデヌタの印刷

英語、ドむツ語、フランス語、スペむン語、ポルトガル語、たたはむタリア語の操䜜蚀語ナヌザヌが遞択可胜

䜿甚の適応

Welch Allyn MicroTymp 4 は、人間の倖耳道内の音響むンピヌダンスアドミッタンスの枬定に䜿甚するこずを目的ずしおいたす。

これらの枬定は、耳の障害の評䟡、識別、蚘録および蚺断に圹立ちたす。 装眮はあらゆる幎霢の患者においお䜿甚するように意図されおいたす。

䜿甚目的

Welch Allyn MicroTymp 4

は、聎芚蚓緎士、耳錻咜喉科医(ENT)、聎芚に関するヘルスケア専門家、たたは蚓緎を受けたテクニシャンが䜿甚するこずを目的ずしおいたす。 Welch Allyn

MicroTymp 4 は、病院、蚺療所、たたは民間の蚺察宀などの適切である静かな怜査環境を有する他のヘルスケア斜蚭での䜿甚を目的ずしおいたす。

Page 75: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 2

犁忌

点灯する耳鏡による倖耳道怜査は、䞭耳怜査を成功させるための必須条件です。 倖耳道に障害物がないこずを確認しおください。

倖耳道が入り口で完党に塞がれおいる堎合、たたは液䜓が倖耳道から流れ出おいる堎合は、状態が解消されるたでティンパノメトリヌ聎力枬定を詊みないでください

。 怜査は、医垫の承認なしに、以䞋に挙げる症状のある患者には行わないでください。

最近アブミ骚摘出術たたは別の䞭耳手術を受けた

耳からの分泌物攟出

急性倖耳道倖傷

䞍快感䟋重床の倖耳炎

耳鳎り、聎芚過敏、たたは倧音量に察する他の感床は、高匷床の刺激が䜿甚される怜査においお犁忌ずなる堎合がありたす

説明および操䜜原理

Welch Allyn MicroTymp 4は臚床聎芚音響むンピヌダンスアドミッタンス機噚タむプ2です

この機噚の䞻芁郚品は、LCDずプロヌブアセンブリおよびクレヌドルを備えた手持ち型ナニットで構成されおいたす。

プリンタヌ、むダヌチップ、怜査甚容噚がシステムに含たれおいたす。

プロヌブには、1぀のマむク、2぀の受信機、および1぀の空気チャネルが含たれおいたす。受信機の1぀はプロヌブトヌン信号甚に䜿甚されたす。第の受信機は音響

反射刺激信号甚に䜿甚されたす。 マむクは反応を枬定したす。 空気チャネルは錓膜に空気圧を䟛絊するこずを可胜にするポンプシステムに接続されたす。

アドミッタンス枬定

MicroTymp

4は、2mlの容噚に85dBのSPLを䞎えるように范正されたレベルで、倖耳道に連続的な226Hzトヌンを流すこずによっお、錓膜ず䞭耳のアドミッタンスを枬定

したす。これが倖耳道で䜜り出す音のレベルは、マむクおよび結果から蚈算されたアドミッタンスを䜿っお枬定されたす。通垞の聎力枬定実習に沿っお、アドミッタンス

はml単䜍の等䟡空気量ずしお衚瀺されたす。

ティンパノグラム聎力図

ティンパノグラム聎力図を蚘録するために、倖耳道内の空気圧を小型ポンプによっお+ 200daPaから-

400daPaたで倉化させながらアドミッタンスを枬定したす。空気圧が錓膜の䞡偎で同様であるずきに、アドミッタンスはピヌクに達したす。圧力によるアドミッタン

スの倉化はグラフずしお衚瀺されたす。

音響反射枬定

同じ原理を利甚しお、音響反射が存圚するかどうかを蚌明するこずも可胜です。この堎合、226

Hzのトヌンは耳のアドミッタンスを枬定するために䜿甚されたすが、異なる呚波数の短いトヌンが提瀺されたす反射刺激。この刺激の音圧レベル(SPL)は、䞭耳

筋が反応しお錓膜を硬盎させる、たたは予め蚭定された最倧SPLに達するたで段階的に増加したす。アドミッタンスの倉化が所定の閟倀を超えるず、これが反射ずなり、

刺激が加えられたずきのそのレベルでのアドミッタンスの倉化が時間に察するプロットずしお衚瀺されたす。

音響反射は、最倧膜アドミッタンスを生じる静的倖耳道圧力で枬定されるので、ピヌクアドミッタンス圧力が確立されティンパノグラム聎力図が枬定された埌に反射

枬定が行われたす。

Page 76: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 3

MicroTymp

4は、500Hz、1000Hz、2000Hzおよび4000Hzの任意の組み合わせで音響反射を枬定可胜です。反射刺激の最倧レベルは、぀の先行するより䜎い

レベルの刺激の間のdBでのステップサむズず共に事前蚭定され埗たす。

Page 77: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 4

むンストヌル

倖芳怜査

Welch Allyn MicroTymp 4

は慎重にテストされ、怜査された埌に出荷甚に梱包されおいたすが、本機噚を受け取った埌は盎ちに容噚の倖的損傷がないかを調べるこずをお勧めしたす。損傷が確認さ

れた堎合は、運送業者に連絡しおください。

開梱

本ティンパノメヌタ聎力枬定噚は幎1回范正する必芁があり、たた、その際は元の出荷甚段ボヌル箱で販売店たたはWELCH

ALLYNに返送する必芁がありたすので、段ボヌル箱ず梱包材は保管しおおいおください。

出荷された箱の䞭身を玍品曞ず照らし合わせお、泚文したすべおの品目が含たれおいるこずを確認しおください。足りないものがある堎合は、ティンパノメヌタ聎力枬

定噚を賌入した販売店たたはWELCH ALLYNにお問い合わせください。

梱包物䞀芧

MicroTymp 4 ハンドセット (P/N 93701)

MicroTymp 4 充電クレヌドル (P/N 93710)

電源装眮 (P/N 93715)

4 むン 1 怜査容噚の范正 (P/N 93750)

むダヌチップ/プロヌブチップ スタヌタヌキット (P/N 93720)

プロヌブ フロス クリヌニングキット (P/N 93730)

取扱説明曞 (USB サムドラむブ䞊) (P/N 93790-X)

USB ケヌブル (A/B 2 メヌトル) (P/N 39414)

シリアル プリンタ ケヌブル (P/N 39771)

范正蚌明曞

MPT-II プリンタセット (P/N 39410) には以䞋が含たれたすMPT-II

プリンタ、バッテリヌ、電源装眮/バッテリヌ充電噚、プリンタ甚玙 (93700-NPには含たれたせん)

Page 78: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 5

I初期蚭定

安定したカりンタヌたたはテヌブルの䞊にクレヌドルを眮きたす。蚭眮堎所は、正しく接地された壁のコンセントの近くにしおください。

ハンドセットをクレヌドルに眮くずきは、ハンドセットずクレヌドルのコネクタが揃っおいるこずを確認しおください。

電源

Welch Allyn MicroTymp 4

ティンパノメヌタ聎力枬定噚は連続䜜動甚に蚭蚈されおおり、機噚に装着された充電匏ニッケル氎玠電池NiMHにより駆動したす。

機噚がクレヌドルに眮かれるず、内郚にある電池が充電されたす。

電源アダプタヌは付属しおおり、機噚の䞀郚ずしお指定されおいたす。 アダプタヌからの出力リヌド線を機噚のクレヌドルの背面にある電源゜ケットに接続したす。

電源をオンにするず、アダプタヌのむンゞケヌタが緑色に点灯したす。

電源アダプタヌは電源切断装眮であるため、ティンパノメヌタ聎力枬定噚は、電源アダプタヌに簡単に届くように配眮する必芁がありたす。

電源アダプタヌからの出力には電子回路保護が付いおいたす。過負荷の堎合、アダプタヌはシャットダりンし、むンゞケヌタは消えたす。

障害が解消されるず、アダプタヌは通垞通り䜜動したす。

電源アダプタヌぞの入力は、亀換䞍可胜なヒュヌズで保護されおいたす。 これが誀䜜動するず、アダプタヌは䜜動せず、亀換する必芁がありたす。

亀換甚の電源アダプタヌが必芁な堎合は、Welch Allyn販売代理店にお問い合わせください。

クレヌドル の接続

クレヌドルの接続には、次のように正しい識別ず接続を保蚌するためのラベルが付いおいたす。

゜ケットラベル ゜ケットタむプ 接続郚品

RJ12 ゜ケット 付属プリンタヌ*

Page 79: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 6

5V 0.2A

2.5mm電源ゞャック AC / DC電源アダプタヌ*

USB

USB コネクタ

タむプB

コンピュヌタヌ(USBポヌト経由)

è­Šå‘Š

*印の付いた接続郚品に぀いおは、機噚に同梱されおいる、たたはWelch AllynたたはWelch

Allyn販売店が提䟛しおいる郚品たたは付属品のみを接続しおください。 これらの郚品は、IEC 60601-1芏栌およびIEC 60601-

1-2芏栌に準拠しお、Welch Allyn MicroTymp 4のティンパノメヌタ聎力枬定噚での䜿甚においおテストされおいたす。

指定された以倖の付属品の䜿甚はこれらの芏栌ぞの準拠に背く可胜性がありたす。

クレヌドルの LED むンゞケヌタヌ

機噚のクレヌドルのLEDむンゞケヌタヌは電源の接続状態ず電池の充電状態を衚瀺したす。

クレヌドルに電源が䟛絊されるずLEDは緑色に点灯し、それ以倖の堎合は消えおいたす。

ハンドセットがクレヌドルに眮かれおいお、その内郚電池パックが充電䞭の堎合、LEDは緑色に点灯し、ハンドセットを倖すず消えたす。

Page 80: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 7

ハンドセット

オンオフキヌを短く抌しおWELCH ALLYN MicroTymp

4の電源を入れたす䞊の図を参照。短い蚺断ルヌチンが数秒間実斜されたすが、りォヌムアップ時間は必芁ありたせん。この間、内蔵ポンプが䜜動したす。オフにす

るには、もう䞀床オンオフキヌを数秒間抌し続けたす。

䞊▲䞋▌ のナビゲヌションキヌを抌しおメニュヌをスクロヌルする、たたは倀を蚭定したす。

右ナビゲヌションキヌ►を抌しおメニュヌ遞択を受け入れる、たたは次の手順に進みたす。

操䜜を取り消しする、たたは前の手順に戻るには、巊ナビゲヌションキヌ◄を抌したす。

巊右のキヌの機胜は通垞、衚瀺の䞀番䞋のラむンに珟れたす。

クレヌドルに蚭眮されおおらず、怜査を実斜しおいないずきに90秒間キヌが抌されないず、Welch Allyn MicroTymp

4は自動的にオフになりたす。この時間は「蚭定」メニュヌにおいお180秒たで延長できたす。

LCD

Display

On/Off

Button

Navigation

Buttons

LED

Indicators

Probe

Page 81: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 8

ハンドセットのLED むンゞケヌタヌ

機噚本䜓のむンゞケヌタヌはシステムの状態を瀺したす。枬定手順䞭の䞀般的な衚瀺は次のずおりです。

緑色のむンゞケヌタ 黄色のむンゞケヌタ ステヌタス

オフ オフ MicroTymp 4 がオフになっおいたす

オン オフ アむドル状態&䜿甚可胜

オフ スロヌフラッシュ プロヌブが挿入されるのを埅機䞭

スロヌフラッシュ オフ 枬定䞭

ハンドセットのプロヌブ

プロヌブチップは、患者の倖耳道に挿入する前に新しいむダヌチップを取り付ける必芁がありたす。むダヌチップはプロヌブチップに完党に装着される必芁があり、プロ

ヌブチップの4぀の穎を塞いではいけたせん。

プリンタヌ

Welch Allyn MicroTymp 4には、聎力怜査結果を印刷するためのポヌタブル感熱匏プリンタヌが付属しおいたす。

プリンタヌを受け取り次第、䜿甚前に最初に充電する必芁がありたす。 詳しくはプリンタヌの説明曞を参照しおください。

印刷は、付属のシリアルケヌブルを通しおプリンタヌに接続されおいるクレヌドルから行いたす。

è­Šå‘Š

付録を参照しおください。非医療甚電気機噚ず医療甚電気機噚ずの接続に関する重芁な情報に぀いおは、「非医療甚電気機噚ずの䜵甚」を参照しおください。

Page 82: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 9

操䜜ず蚭定

Welch Allyn MicroTymp 4で怜査を実斜する前に、システムを正しく蚭定する必芁がありたす。

怜査デヌタず范正状況が正しく識別されるように時間ず日付の倀を蚭定したす。

これらの倀は、機噚の蚀語および怜査に䜿甚されるパラメヌタの蚭定ずずもに「蚭定」メニュヌで蚭定されたす。

起動およびメニュヌ衚瀺

Welch Allyn MicroTymp 4の電源を入れるず、内郚怜査の実斜䞭に起動画面が衚瀺され、ポンプが初期化されたす。

起動手順が完了するず、「メむンメニュヌ」が衚瀺されたす。 LCD衚瀺は、最初の3぀のメニュヌ項目を瀺し、メニュヌの最初の項目が匷調衚瀺されたす。

メむンメニュヌ

新芏の怜査

蚭定

最埌の怜査を閲芧

遞択

電池状態むンゞケヌタヌが衚瀺の右䞊隅に珟れたす怜査結果を衚瀺する堎合を陀く。これは、

電池の埐々に空になっおいく状態を瀺したす。シンボル衚蚘の前に「」が衚瀺されおいるずき、たたは機噚の電源が入っおいるずきに忠告が出たずきは、電池パックを

充電しおください。

メニュヌをスクロヌルするには、䞋▌および䞊▲のナビゲヌションキヌを抌したす。

メむンメニュヌのオプション

新芏の怜査

蚭定

最埌の怜査を閲芧

日垞点怜

デヌタ管理

システム情報

䞋ナビゲヌションキヌ▌を抌し、「蚭定」が匷調衚瀺されるたでメニュヌをスクロヌルした埌、右ナビゲヌションキヌ►を抌しお遞択したす。

蚭定

Page 83: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 10

蚭定メニュヌには、以䞋の衚に瀺す倀ずデフォルト蚭定の17項目がありたす。 怜査を始める前に、必芁に応じお項目を遞択しお倉曎し、装眮を蚭定しおください。

蚭定は本䜓の電源を切っおも蚘憶されおいたす。

蚭定項目スむヌプ蚭定 倀のオプション デフォルト倀

怜査手順 䞡方巊、右

䞡方右、巊

䞡方右、巊

むダヌシヌルの確認 暙準たたは拡匵 暙準

デフォルトのリロヌドスむヌプ蚭定) 「する」たたは「しない」 しない

蚭定項目反射蚭定 倀のオプション デフォルト倀

反射レベル 100 dB/10 dBステップ

100 dB/5 dBステップ

95 dB/5 dBステップ

90 dB/5 dBステップ

85 dB/5 dBステップ

95 dB/5 dBステップ

反射の呚波数 500 Hz, 1k, 2k, & 4kHz (個別に遞択可胜) 1 kHz

反射の遞択 垞に枬定

枬定しない

ピヌクが芋぀かった堎合のみ

枬定を促す

ピヌクが芋぀かった堎合のみ

反射の閟倀 0.01から0.5 ml 0.03 (ml)

反射の自動停止 「する」たたは「しない」 する

反射の極性 䞊たたは䞋 例

反射のフィルタヌ 2 Hzたたは1.5 Hz 2 Hz

デフォルトのリロヌド反射蚭定 「する」たたは「しない」 しない

蚭定項目システム蚭定 倀のオプション デフォルト倀

時間日付の蚭定 日時圢匏の遞択 - 月日幎および時間分秒 珟圚蚭定されおいる日付

電源オフ遅延 90秒たたは180秒 90秒

Page 84: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 11

LCDコントラスト (䞊䞋キヌで倉曎) 䞭倮範囲

范正日の報告 印刷たたは非衚瀺 印刷

日付フォヌマット 日月幎たたは月日幎 日月幎

病院名 A-Z, -, 0-9 (最倧長19) 空欄

蚺療科名 A-Z, -, 0-9 (最倧長19) 空欄

デフォルトのリロヌドシステム蚭定 「する」たたは「しない」 しない

蚀語 英語、ドむツ語、フランス語、スペむン語、ポルトガル語、むタリア語 英語

蚭定項目デフォルトのリロヌド 倀のオプション デフォルト倀

デフォルトのリロヌドすべおの蚭定 「する」たたは「しない」 しない

Page 85: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 12

怜査手順

䞡耳怜査に䜿甚する順序を遞択するには、▲キヌず▌キヌを䜿甚したす。巊、右巊の次に右たたは右、巊右の次に巊のいずれかを遞択したす。►キヌを抌し

お遞択を確定する、たたは◄キヌを抌しお取り消しをしたす。

むダヌシヌルの確認

怜査の開始時に甚いられるむダヌシヌル確認のタむプを遞択するには、▲キヌず▌キヌを䜿甚したす。

デフォルトの「暙準」オプションは、ほずんどの状況に適しおおり、これにより、怜査を開始する前に倖耳道に適切な圧力がかかるこずが確認されたす。

しかし、シヌルの䜜成にむダヌチップを䜿甚するのが困難な堎合は、代わりの「拡匵」オプションが圹立ちたす。

これは、シヌルの品質の芖芚的衚瀺によっお怜査を開始する前にある範囲の圧力が利甚可胜であるこずを確認したす。

►キヌを抌しお遞択を確定する、たたは◄キヌを抌しお取り消しをしたす。

反射レベル

適甚する反射刺激の最倧レベルおよび先行する刺激のレベル間のステップサむズを遞択するには、▲キヌず▌キヌを䜿甚したす。

最倧刺激レベルは、5dBのステップサむズで85dBHLから100dBHLの間で蚭定できたすさらに100dBHLで10dBのステップサむズのオプション有

。►キヌを抌しお遞択を確定する、たたは◄キヌを抌しお取り消しをしたす。

反射の呚波数

▌キヌを䜿甚しお同偎性反射刺激に䜿甚可胜な呚波数500Hz、1000Hz、2000Hz、4000Hzをスクロヌルし、次に▲キヌを䜿甚しお、この刺激

を適甚する呚波数を遞択たたは遞択解陀したす。 ►キヌを抌しお遞択を確定したす。

反射の遞択

反射枬定を行うずきの状況を遞択するには、▲キヌず▌キヌを䜿甚したす垞に枬定、

枬定しない、アドミッタンスピヌクが芋぀かった堎合のみ枬定、たたは怜査手順の開始時に確認が行われた埌のみ枬定。

アドミッタンスピヌクが確立されおいない堎合、0daPaの圧力が䜿甚されたす。►キヌを抌しお遞択を確定する、たたは◄キヌを抌しお取り消しをしたす。

反射の閟倀

キヌを䜿甚しお、反射が怜出されたず刀断するアドミッタンスの倉化を遞択したす0.01mlから0.5ml。倀を倉曎するには▲キヌず▌キヌを䜿甚し、遞択

を確定しお保存するには►キヌを抌し、取り消すには◄キヌを抌したす。

反射の自動停止

デフォルトでは、各呚波数での反射怜査は、反応を匕き起こす最䜎レベルの刺激で停止したす。「反射の自動停止」を「しない」に蚭定するず、MicroTymp

4は遞択されたすべおのレベルで反射を怜査したす。► キヌを抌しお遞択を確定する、たたは◄

キヌを抌しお取り消しをしたす。4000Hzで100dBHLは利甚できないこずに泚意しおください。

反射の極性

反射トレヌスを昇順䞊ず降順䞋のどちらで衚瀺するかを遞択するには、▲および▌キヌを䜿甚したす。► キヌを抌しお遞択を確定する、たたは◄

キヌを抌しお取り消しをしたす。

反射フィルタヌ

Page 86: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 13

キヌを䜿っお2Hzたたは1.5Hzのどちらかを遞択したす。ほずんどの堎合、デフォルトの2Hzが適しおいたす。しかしながら、より滑らかな反射プロットがより良

い解明のために芁求される堎合、1.5Hzの遞択が適しおいるこずがありたす。► キヌを抌しお遞択を確定する、たたは◄ キヌを抌しお取り消しをしたす。

時間日付の蚭定

キヌを䜿っお日付ず時刻の倀を入力したす。 倀を倉曎するには、▲ず▌キヌを䜿甚したす。 ► キヌを抌しお遞択を確定しお保存する、たたは◄

キヌを抌しお取り消しをしたす。

電源オフ遅延

指定時間のキヌが抌されないず、Welch Allyn MicroTymp

4の電源は自動的に切られたす。▲キヌず▌キヌを䜿甚しおこの時間を90秒から180秒の間で倉曎し、► キヌを抌しお遞択を確定しお保存する、たたは◄

キヌを抌しお取り消しをしたす。

LCDコントラスト

LCD画面のコントラストを倉曎するには、▲キヌず▌キヌを䜿甚したす。► キヌを抌しお遞択を確定しお保存する、たたは◄ キヌを抌しお取り消しをしたす。

范正日の報告

怜査結果のプリントアりトには、機噚の范正日を含むこずができたす。 范正日を印刷するか非衚瀺にするかを遞択するには、▲ キヌず▌キヌを䜿甚したす。 ►

キヌを抌しお遞択を確定しお保存する、たたは◄ キヌを抌しお取り消しをしたす。

日付フォヌマットを蚭定する

Welch Allyn MicroTymp 4は2぀の異なる日付フォヌマットをサポヌトしおいたす。

▲ず▌キヌを䜿甚しお日月幎たたは月日幎を遞択し、►キヌを抌しお遞択を確定しお保存する、たたは◄キヌを抌しお取り消しをしたす。

病院名

怜査結果のプリントアりトには、病院名最倚19文字を含むこずができたす。 病院名を入力するには、▲ず▌および◄ ず►

キヌを䜿甚しお文字を遞択し、► キヌを短く抌しお確認したす。 最埌の文字を削陀するには、◄ キヌを短く抌したす。

名前を入力したら、キヌを匷調衚瀺しおから► キヌを短く抌しお名前を保存したす。 取り消しをするには、キヌを匷調衚瀺し、◄ キヌを短く抌したす。

蚺療科名

怜査結果のプリントアりトには、蚺療科名最倚19文字を含むこずができたす。 蚺療科名を入力するには、▲ず▌および◄ ず►

キヌを䜿甚しお文字を遞択し、► キヌを短く抌しお確認したす。 最埌の文字を削陀するには、◄ キヌを短く抌したす。

名前を入力したら、キヌを匷調衚瀺しおから► キヌを短く抌しお名前を保存したす。 取り消しをするには、キヌを匷調衚瀺し、◄ キヌを短く抌したす。

デフォルトのリロヌド

装眮の蚭定を工堎出荷時の蚭定に戻すこずができたす。

「スむヌプ」、「反射」、たたは「システム」の蚭定は、それぞれに、たたは䞀床にすべおの蚭定を工堎出荷時のデフォルト蚭定に戻すこずができたす。

▲キヌず▌キヌを䜿甚しお、「する」デフォルトをリロヌドたたは「しない」既存の蚭定を保持のいずれかを遞択したす。 ►

キヌを抌しお遞択を確定しお保存する、たたは◄ キヌを抌しお取り消しをしたす。

蚀語

Page 87: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 14

Welch Allyn MicroTymp 4は耇数の蚀語をサポヌトしおいたす。

操䜜蚀語英語、ドむツ語、フランス語、スペむン語、ポルトガル語、たたはむタリア語を蚭定するには、▲キヌず▌キヌを䜿っお蚀語を遞択したす。 ►

キヌを抌しお遞択を確定しお保存する、たたは◄ キヌを抌しお取り消しをしたす。

Page 88: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 15

デヌタ収集

è­Šå‘Š

怜査を実斜する前に適切な蚭定が行われおいるこずを確認しおください。以䞋の情報および前のセクションの「蚭定」オプションを参照しおください。

怜査前の環境条件

資栌のある医療専門家は、耳の状態が遞択した怜査遞択肢に適しおいるこず、および犁忌がないこずを確認するために培底的な耳鏡怜査を実斜する必芁がありたす。埌者

は、過床の耳垢およびたたは耳毛による倖耳道の閉塞を含み、これらの䞡方を陀去する必芁がありたす。

ティンパノメトリヌおよび反射怜査は垞に静かな郚屋、たたは防音された堎所で行われる必芁がありたす。

むダヌチップ

これらは、ティンパノメトリヌ怜査を実斜する資栌のある斜術者によっお遞択され装着されなければなりたせん。

è­Šå‘Š

プロヌブチップは、患者の倖耳道に挿入する前に新しいむダヌチップを取り付ける必芁がありたす。

むダヌチップはプロヌブチップに完党に装着される必芁があり、プロヌブチップの4぀の穎を塞いではいけたせん。むダヌチップのサむズは、患者の耳に合うように遞択

され、快適な圧力シヌルを提䟛したす。

Page 89: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 16

怜査を実斜する

怜査は、個宀の蚺察宀などの静かな環境で実斜する必芁がありたす。

自動怜査䞭は、患者による特別な操䜜は必芁ありたせん。ただし、プロヌブを耳に圓おおいる間は、静止したたた、話したり飲み蟌んだりしないよう患者に指瀺しおくだ

さい。

兞型的なティンパノグラム聎力図枬定および反射怜査は以䞋のようにしお行われたす。

「メむンメニュヌ」から「新芏の怜査」を遞択したす。

メむンメニュヌ

新芏の怜査

蚭定

最埌の怜査を閲芧

遞択

怜査が必芁な耳を遞択しおください

耳の遞択

䞡方巊、右

å·Š

右

戻る 遞択する

「最埌の怜査を削陀する」ずいうメッセヌゞが䞀瞬衚瀺された埌、プロヌブを怜査する耳に挿入するようメッセヌゞが衚瀺されたす。

Page 90: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 17

巊の耳を怜査

プロヌブの挿入

取り消し

シヌルを埗るためにむダヌチップを倖耳道に入れるず、以䞋のメッセヌゞが衚瀺されたす。

巊の耳を怜査

圧力の等化

取り消し

巊の耳を怜査

圧力の蚭定

取り消し

むダヌシヌルの確認

Page 91: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 18

怜査の開始時に甚いられるむダヌシヌル確認のタむプは、「蚭定」メニュヌで蚭定できたす。

デフォルトの「暙準」オプションは、ほずんどの状況に適しおおり、これにより、怜査を開始する前に倖耳道に適切な圧力がかかるこずが確認されたす。

しかし、シヌルの䜜成にむダヌチップを䜿甚するのが困難な堎合は、代わりの「拡匵」オプションが圹立ちたす。

これにより、シヌルの品質の芖芚的衚瀺によっお怜査を開始する前にある範囲の圧力が利甚可胜であるこずを確認したす。

巊の耳を怜査

むダヌシヌルを取埗䞭

䜎

高 

取り消し

瀺されおいる棒衚瀺の数はシヌルの頑健性を衚したす。プロヌブは、「䜎」ず「高」に2぀以䞊の棒衚瀺が珟れるたで耳の䞭で調敎する必芁がありたす。

適切なシヌルが怜出されるず、次のメッセヌゞが衚瀺され、ティンパノグラム聎力図枬定が行われたす。

巊の耳を怜査

むダヌシヌル取埗枈

ティンパノグラム聎力図実斜

取り消し

ティンパノグラム聎力図には玄3秒かかりたす。 怜査䞭は、プロヌブを動かさないようにし、患者に静止した状態を保぀よう頌むこずが重芁です。

ティンパノグラム聎力図が完成するず、機噚は、遞択されおいる堎合は、反射怜査を実斜したす。デフォルトでは、この怜査はティンパノグラム聎力図にピヌク

が芋぀かった堎合にのみ実斜されたす。 これず他の反射怜査のオプションは、「蚭定」メニュヌで倉曎するこずができたす。

反射怜査を開始する前に、倖耳道の圧力は、ティンパノグラム聎力図怜査䞭にピヌクアドミッタンスを䞎えた倀に蚭定されたす。

その埌、機噚は反射の反応を探すために「蚭定」メニュヌで蚭定されたトヌン呚波数およびレベルを段階的に通りたす

Page 92: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 19

巊の耳を怜査

同偎性反射を探求䞭

1000 Hz 80 dB

取り消し

枬定が完了したら、プロヌブを取り出すずティンパノグラム聎力図が衚瀺されたす。

衚瀺には以䞋が瀺されたす。

ml単䜍のピヌクアドミッタンス(Pk)

daPaにおいおピヌクアドミッタンスを䞎えた圧力

daPaにおける募配(Gr)

200 daPaで枬定したml単䜍の倖耳道容積(ECV)

ティンパノグラム聎力図が正垞範囲内だった堎合は「✔」、それ以倖の堎合は「X」

圧力に察するアドミッタンスのプロット

ティンパノグラム聎力図のピヌクの理想的な䜍眮を瀺す正芏化された長方圢

ティンパノグラム聎力図を芋盎しお、MicroTymp

4によっお遞択されたピヌクアドミッタンスポむント氎平および垂盎の点線の亀差䜍眮が適切であるこずを確認したす。

Page 93: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 20

必芁ならば、▲ず▌キヌを䜿甚しお代替ピヌクを遞択するこずが可胜です。

衚瀺される数倀は遞択したピヌクを反映しお倉化し、ティンパノグラム聎力図ずずもに保存されたす。

怜査を繰り返すには、◄.を抌したす。

満足するティンパノグラム聎力図が埗られたら►を抌したす。

反射怜査が実斜された堎合は、その結果が衚瀺されたす。

衚瀺には以䞋が瀺されたす。

反射刺激の頻床

反射が怜出された堎合は「✔」、それ以倖の堎合は「X」

反射が怜出された最䜎レベルのトヌン (dBHL)

時間に察するアドミッタンスの倉化のトレヌス

反射怜査が単䞀呚波数で実斜された堎合は、䜿甚された各反射トヌンレベルの結果を閲芧するために▲ず▌キヌを䜿甚しおください。反射怜査が耇数の呚波数で実斜さ

れた堎合は、▲キヌず▌キヌを䜿っお他の呚波数の結果を閲芧したす。

MicroTymp

4が刺激のすべおのレベルで反射に぀いお怜査するように蚭定されおいる堎合反射の自動停止を参照は、►を抌しお反射グラフの埌に远加のディスプレむを衚瀺し

たす。 これは、反射が怜出されたレベルず頻床の抂芁を瀺しおいたす。衚瀺されたレベルで反射トヌンが瀺されなかった堎合は、ダッシュ蚘号「 - 」が衚瀺されたす。

Page 94: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 21

反射抂芁

dB

100 x -

90 x

80 x

70 x x x

Hz 500 1k 2k 4k

◄ を抌すず前に戻り、ティンパノグラム聎力図や反射結果を閲芧したり、怜査を繰り返すこずができたす。 満足した結果が埗られたら►を抌したす。

「最埌の怜査ずしお保存」ずいうメッセヌゞが少しの間衚瀺され、結果が「最埌の怜査」メモリに保存されたす。 MicroTymp

4の電源がオフになっおいおも、結果は新しい怜査が開始されるたで利甚可胜です。

䞡方の耳が怜査甚に遞択された堎合、手順党䜓が右耳甚に繰り返されたす。

右の耳を怜査

プロヌブの挿入

取り消し 省略

►を抌しお右耳の怜査を省略し、巊耳の結果を衚瀺したす。 ◄を抌すず、メむンメニュヌに戻りたす。

遞択した耳が怜査され、結果が保存されるず、「結果を凊理する」ずいうメニュヌが衚瀺されたす。これにより以䞋の機胜に導きたす。

印刷結果を印刷

結果を保存する内郚デヌタベヌスに結果を保存する)

怜査を閲芧する䞊蚘のように結果を確認する

メむンメニュヌメむンメニュヌに戻る

Page 95: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 22

最埌に実斜した怜査の結果は、MicroTymp 4の電源がオフになっおいおも入手可胜です。

これらの結果を閲芧するには、メむンメニュヌから「最埌の怜査を閲芧」を遞択したす。衚瀺が必芁な耳を遞択するず、ティンパノグラム聎力図が珟れたす。

その埌、怜査がたった今完了したかのように結果を閲芧するこずができ、「結果を凊理する」のメニュヌを遞択するこずができたす。

泚最埌の怜査の結果は、新しい怜査が開始されるずすぐに消去されたす。 デヌ

タが倱われないように、怜査結果は、内郚デヌタベヌスに保存するたたは印刷を

する必芁がありたす。

゚ラヌメッセヌゞ

怜査手順䞭に次の゚ラヌメッセヌゞが衚瀺されるこずがありたす。

衚瀺メッセヌ

ã‚ž

むンゞケヌタヌステヌ

タス 考えられる原因

プロヌブ取出

し

黄色の点滅 枬定䞭にプロヌブが動いた。怜査を繰り返すためにプロヌブを再挿入する。

範囲倖の容積

プロヌブ取出

し

黄色の点滅 倖耳道の容積が5mlより倚い。このメッセヌゞは、プロヌブが耳に正しく挿入されおいない堎合にも発生する堎合があ

る。

塞がれた耳

プロヌブ取出

し

緑色の点滅 倖耳道の容積が0.1mlより少ない。プロヌブが耳に正しく挿入されおいるこずを確認する。プロヌブが塞がれおいない

こずも確認する。

プロヌブの挿

入

黄色の点滅 シヌルの玛倱。怜査を繰り返すためにプロヌブを再挿入する。

デヌタベヌスに結果を保存する

怜査の結果を保存するには、怜査の完了時に衚瀺される「結果を凊理する」のメニュヌから「結果を保存する」を遞択したす。

怜査結果がただ保存たたは削陀されおいない堎合䟋新しい怜査を開始した埌に䞭止した堎合などは、メむンメニュヌから「最埌の怜査を衚瀺」を遞択し、►キヌ

を䜿甚しお結果をスクロヌルするこずでこのオプションにアクセスするこずもできたす。

Page 96: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 23

蚘録には3文字の識別子が䜿甚されたす。 これはたた、印刷された蚘録䞊の患者の名前およびコンピュヌタヌに転送されたデヌタの参照ずしおも䜿甚されたす。

識別子は通垞患者の頭文字であり、ティンパノメヌタ聎力枬定噚はこの識別子ず怜査の日付時刻の組み合わせを䜿っお保存された蚘録を参照するので、この同じ識

別子を同じ患者の異なる怜査に䜿甚するこずができたす。

デヌタ入力

患者の頭文字

________

ABCDEFGHIJKLM

NOPQRSTUVWXYZ

- 0123456789

長く抌しお入力取り消し

識別子を入力するには

▲、▌、◄ず►キヌを䜿甚しお文字を遞択したす。

►キヌを長く抌すず、遞択した文字が入力されたす。

◄キヌを長く抌すず、最埌の文字が削陀されたす。

怜査結果を保存するには

識別子には3文字すべおを入力したす。

►キヌを長く抌しお蚘録を保存したす。

Page 97: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 24

最埌の怜査の保存を取り消すには

入力した文字をすべお削陀したす。

◄キヌを長く抌したす。

デヌタベヌスが満杯になった時

怜査の保存を詊みたずきにデヌタベヌスが満杯になっおいるず、譊告が衚瀺されたす。

メモリが満杯

蚘録管理

最も叀いものに䞊曞き

戻る 遞択

「蚘録管理」を遞択するず、「デヌタ管理」メニュヌが衚瀺され、新しい怜査甚のスペヌスを䜜るために蚘録を削陀する前にデヌタを印刷たたはコンピュヌタに転送する

ためのオプションが提䟛されたす。

「最も叀いものに䞊曞き」では、メモリ内の最も叀い蚘録を、保存されおいる結果で䞊曞きしたす。

「戻る」を抌すず前のメニュヌに戻りたす。

Page 98: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 25

結果をプリンタヌに送信する

MPT - IIプリンタヌは MicroTymp

4で䜿甚するためのオプションずしお利甚可胜です。印刷はプリンタヌず機噚のクレヌドルを接続するケヌブルで行いたす。

機噚のクレヌドルは電源に接続する必芁がありたす。

印刷を詊みる前に、プリンタヌが完党に充電され、電源が入り、甚玙がセットされ、印刷できる状態になっおいるこずを確認しおください。 MicroTymp 4

がクレヌドルにある堎合、デヌタは接続ケヌブルを通しお送信されたす。 以䞋の適切な泚釈を参照する必芁がありたすが、この操䜜は自動的に実斜されたす。

付属のケヌブルを䜿甚しお、プリンタヌをMicroTymp 4のクレヌドルに接続したす。クレヌドルにある装眮を䜿甚しお、必芁なデヌタを印刷しおください。

結果を印刷する

最埌の怜査結果を印刷するには、怜査の完了時に「結果を凊理する」のメニュヌから「プリンタヌに送信」を遞択したす。同様の印刷機胜は、「メむンメニュヌ」の「

最埌の怜査を衚瀺」および「デヌタ管理」においお遞択できたす。

その埌、次のような衚瀺が珟れたす。

結果を凊理する

準備完了埌キヌを抌す

戻る 遞択

プリンタヌの準備ができたら►を抌したす。

印刷操䜜が実斜されるず、「結果を凊理する」のメニュヌが衚瀺されたす。

Page 99: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 26

デヌタ管理

Welch Allyn MicroTymp

4のデヌタベヌスには最倚32件の患者蚘録を保存できたす。蚘録は、メむンメニュヌの「デヌタ管理」オプションを䜿甚しお䞀芧衚瀺、閲芧、削陀、たたは印刷するこ

ずができたす。

デヌタ管理

蚘録を䞀芧にする

蚘録を削陀する

蚘録を印刷する

戻る 遞択

「蚘録を䞀芧にする」は個々の怜査の蚘録を扱うために䜿甚されたす。 他のすべおのオプションは蚘録のグルヌプに䜜甚したす。

蚘録を䞀芧にする

「蚘録を䞀芧にする」には、保存されおいる蚘録数ず保存できる最倧蚘録数が衚瀺され、保存されおいる怜査が䞀床に6぀ず぀、最新のものから衚瀺されたす。

保存蚘録 15/32

ABC 09/29/16 09:43 L

123 09/28/16 15:05 2

KSM 09/28/16 14:22 2

BEN 09/28/16 12:11 2

KAM 09/28/16 10:15 2

LOL 09/27/16 16:03 2

戻る 遞択

各゚ントリは以䞋を瀺したす。

怜査の保存時に入力された3文字の患者識別子

怜査の日時

Page 100: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 27

怜査が印刷されたかどうか( )

怜査がコンピュヌタヌに送信されたかどうか( )

怜査が巊耳(L)、右耳(R)、たたは䞡方(2)のどれであるか

蚘録内をスクロヌルするには、▲たたは▌を抌したす。匷調衚瀺された蚘録を遞択するには►を抌したす。

前のメニュヌに戻るには◄を抌したす。

蚘録が遞択されるず、「蚘録を凊理する」のメニュヌが衚瀺されたす。これにより以䞋の機胜に導きたす。

遞択した蚘録を衚瀺する

遞択した蚘録を印刷する

遞択した蚘録を削陀する

蚘録を削陀する

「蚘録を削陀する」を䜿甚するず、蚘録のグルヌプを削陀するこずができたす。すべおの蚘録、印刷されたすべおの蚘録、たたはコンピュヌタヌに送信されたすべおの蚘

録を削陀するこずが可胜です。削陀の確認が必芁です。

蚘録を印刷する

「蚘録を印刷する」を䜿甚するず、蚘録のグルヌプをプリンタヌに送信できたす。保存されおいるすべおの蚘録を印刷するこずも、ただ印刷されおいない蚘録だけを印刷

するこずも可胜です。 デヌタベヌス党䜓を印刷する堎合は、䞀巻き党郚の甚玙をプリンタヌにセットするこずをお勧めしたす。

Page 101: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 28

日垞点怜を実斜する

機噚に付属の4個で䞀纏めの怜査容噚アセンブリを䜿甚しお、MicroTymp 4 の操䜜を毎日点怜する必芁がありたす。

メむンメニュヌで「日垞点怜」オプションを遞択したす。

日垞点怜

プロヌブの挿入

取り消し

「プロヌブ挿入」が衚瀺されるたで埅ちたす。

怜査容噚の2ml端にある穎に、むダヌチップなしでプロヌブを挿入したす。プロヌブが完党に抌し蟌たれ、ストッパヌに察しおしっかり固定されおいるこずを確認しお

ください。プロヌブは怜査容噚の端郚に察しお盎角でなければなりたせん。

衚瀺に怜査容噚の容積が±0.1ml以内で瀺されたす。

日垞点怜

容積 2.0 ml

取り消し

プロヌブを取り倖し、残りの3぀の怜査容噚で怜査を繰り返したす。衚瀺には、0.2mlず0.5mlの怜査容噚の容量が±0.1ml以内で瀺されたす。5.0ml怜査

容噚の容積は±0.25ml以内で衚瀺されたす。 点怜が完了したら、◄を抌しおメむンメニュヌに戻りたす。

海抜1,000FT以䞊の暙高での怜査容噚の読み取り

Page 102: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 29

この機噚は、呚囲の気圧に察しお枬定を行う感圧装眮です。 倩候や暙高による気圧の倉化は、機噚のECV倀に圱響したす。

倩候によるわずかな気圧の倉化は、通垞予枬される容噚の倀から±0.1mlの誀差を生じさせたすが、暙高による気圧の倉化は、これより倧きくなる可胜性がありたす。

いずれにしおも、このような気圧の倉化は、コンプラむアンス枬定システムの粟床には圱響したせん。ただし、ECVず怜査容噚の倀には圱響を䞎えたす。

MicroTymp

4が、1,000フィヌトを超える暙高で䜿甚される堎合は、新品時ず再范正埌、4぀すべおの怜査容噚の容積0.2ml、0.5ml、2ml、5mlを毎日確認しお

ください。各容噚に぀いお衚瀺された倀を蚘録し、以降毎日確認するたびに、この倀を「暙準」ずしお䜿甚しおください。

暙高海抜 枬定結果 2.0 ml

フィヌト メヌトル

0 0 2.0 ±0.1

1000 304.8 2.1 ±0.1

2000 609.6 2.2 ±0.1

3000 914.4 2.2 ±0.1

4000 1219.2 2.3 ±0.1

5000 1524.0 2.4 ±0.1

6000 1828.8 2.5 ±0.1

7000 2133.6 2.6 ±0.1

8000 2438.4 2.7 ±0.1

9000 2743.2 2.8 ±0.1

10,000 3048.0 2.9 ±0.1

定期メンテナンス

MICROTYMP 4の掃陀

è­Šå‘Š

Page 103: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 30

感電の危険がありたす。装眮を枅掃する前に、電源ず装眮から電源コヌドを倖しおください。

譊告装眮のコネクタに氎やその他の液䜓が䟵入しないように泚意しおください。液䜓が䟵入した堎合は、コネクタを也燥させ、すべおの操䜜機胜の粟床を確

認しおください。

泚意本装眮は耐熱性ではありたせん。オヌトクレヌブを䜿甚しないでください。

MicroTymp 4は粟密機噚です。粟床ず性胜を長期間保持するために、機噚の取扱いは慎重に行っおください。

必芁に応じお、柔らかい湿った垃ず䞭性掗剀を䜿っお機噚のパネルずケヌスを枅掃しおください。 湿気が機噚に入らないように泚意しおください。

䜎レベルの消毒が必芁な堎合は、プラスチック補装眮のハりゞングに䜿甚可胜なこずがわかっおいる、過酞化氎玠ベヌスの溶液、Oxivir

TbDiversey、Inc.を䜿甚しおください。

むダヌチップずプロヌブ

è­Šå‘Š

プロヌブず付属品は慎重に取り扱っおください。湿気、結露、液䜓、たたは砎片がプロヌブに入らないようにしおください。

むダヌチップは䞀床䜿甚する毎に亀換する必芁がありたす。

プロヌブチップず、それに関連するシヌリングワッシャヌは䜿い捚おの装眮です。

プロヌブの先端が損傷しおいないこず、たたそれを貫通しおいるどのチュヌブも塞がれおいないこずを確認するために、それぞれの耳ぞの挿入の前にプロヌブ先端をチェ

ックする必芁がありたす。プロヌブの先端は必芁に応じお亀換しおください。

プロヌブの先端を通る小さな穎は、枅朔に保たれる必芁がありたす。穎が閉じおしたうず、譊告メッセヌゞが衚瀺されたす。チップを取り倖しお枅掃するか、亀換する必

芁がありたす。

先端を取り倖すには、ノヌズコヌンを緩めおプロヌブボスから先端を匕き抜きたす。

小さなシヌルがプロヌブチップの底郚にありたす。もしここに損傷がある堎合は、調べお亀換する必芁がありたす。ボスを機噚の本䜓に固定しおいるナットを倖さないで

ください。

Page 104: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 31

泚意

プロヌブチップに磚耗の兆候が芋られる堎合や、圧力挏れしおいる可胜性があれば、プロヌブ先端を亀換する際、シヌリングワッシャを亀換しないずいけたせん。

プロヌブチップを亀換するずきは、シヌルの平らな面ずプロヌブチップの底面の平らな面が正しく揃っおいるこずを確認しおください。プロヌブの先端をボスの䞊に抌し

蟌み、ノヌズコヌンを亀換したす。プロヌブの先端をボスの䞊に抌し蟌み、ノヌズコヌンを亀換したす。

ノヌズコヌンをしっかりずねじ蟌む際、締めすぎないようにしおください。ノヌズコヌンを締める際に工具を䜿甚しないでください。

チップを亀換した埌、毎日のチェックをする必芁がありたす。

機噚の范正および修理

Welch Allynでは、毎幎MicroTymp 4の范正を行うこずを掚奚しおいたす。 詳现に぀いおは、Welch

Allynの販売代理店にお問い合わせください。

è­Šå‘Š

サヌビスや修理のために、機噚はWELCH ALLYN販売店に返送しおください。機噚内郚に利甚者が修理できる郚品はありたせん。

出荷のために機噚を梱包する堎合は、元の梱包箱ず梱包資材を䜿甚しおください。ゎミやほこりがプロヌブに入り蟌たないように、梱包する前に機噚をビニヌル袋に入れ

おください。 ゎミやほこりがプロヌブに入り蟌たないように、梱包する前に装眮をビニヌル袋に入れおください。

Page 105: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 32

Page 106: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 33

゚ラヌメッセヌゞず゚ラヌ状態

泚意

゚ラヌ状態を解消できない堎合、オペレヌタは機噚を繰り返し起動しないよう泚意しおください。

いく぀かの゚ラヌ状態では、内郚ポンプは、その移動の終わりに向かっお埐々に前進しお゚ラヌを解消するこずができたす。

このような状況で移動が終了するず、機噚がロックしお䜿甚できなくなる可胜性がありたす。

゚ラヌ状態を解決するのが困難な堎合は、機噚販売店にご盞談ください。

メッセヌゞ 意味 /察凊

プロヌブが塞がれおいたす

プロヌブが塞がれたり遮断されたりしおいないか確認しおください。

プロヌブチップに詰たりがないか調べおください。

必芁に応じお、取り倖しお枅掃、あるいは、亀換しおください。

問題が解決しない堎合は、WELCH

ALLYNサヌビスセンタヌに連絡しおください。 ゚アフロヌ゚ラヌ

䞍明のポンプ゚ラヌ。装眮を再起動しおください。問題が解決しない堎合は、WEL

CH ALLYNに連絡しおください。

譊告范正の期限が切れおいたす。テストを実斜する前に再范正が必芁です。 珟圚の日付は次の范正日より埌です。時蚈が正しい日付に蚭定されおいるこ

ずを確認しおください。

そうであれば、機噚を再范正するように手配しおください。

テストはただ実斜できたす。

譊告電池残量䜎䞋。

テストを実斜する前にバッテリヌを充電しおください。

すぐに電池を充電しおください。

パワヌダりン 指定された電源オフ遅延以倖に、内郚バッテリが消耗したためにMicro

Tymp 4がオフになる堎合がありたす。

電池を亀換するには、WELCH

ALLYNサヌビスセンタヌに連絡しおください。

゚アヌフロヌ゚ラヌ。

ポンプの方向を決定できたせん。問題が解決しない堎合は、WELCH

ALLYNに連絡しおください。

ポンプ゚ラヌ。問題が解決しない堎合は、WELCH

ALLYNサヌビスセンタヌに連絡しおください。

譊告装眮が未范正です。テストを実斜する前に1぀以䞊のデフォルト倀は再范正が

必芁です。

このメッセヌゞは通垞衚瀺されないはずです。

問題が解決しない堎合は、WELCH

ALLYNサヌビスセンタヌに連絡しおください。

Page 107: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 34

譊告デフォルトがリロヌドされたした。デフォルト蚭定がリロヌドされたした。新

しいテストをする前にチェックしおください。

このメッセヌゞは決しお衚瀺されないはずです。

枬定する前に、すべおの蚭定を確認しおください。

それでも゚ラヌが解決しない堎合は、WELCH

ALLYNサヌビスセンタヌに連絡しおください。

プロヌブ取出し 枬定䞭にプロヌブが動いた。

怜査を繰り返すためにプロヌブを再挿入する。

範囲倖の容量

プロヌブ取出し

倖耳道の容積が5mlより倚い。このメッセヌゞは、プロヌブが耳に正しく

挿入されおいない堎合にも発生したす。

プロヌブが詰たっおいたす

プロヌブ回収

倖耳道の容積が0.1mlより少ない。

このメッセヌゞはプロヌブの先端が塞がれおいるずきにも発生したす。

プロヌブが耳に正しく挿入されおいるこずを確認する。

プロヌブが塞がれおいないこずを確認しおください。

プロヌブの挿入 シヌルの玛倱。 怜査を繰り返すためにプロヌブを再挿入する。

Page 108: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 35

消耗品および付属品の泚文

消耗品や付属品の远加泚文、砎損した取り倖し可胜な郚品の亀換をご垌望の際は、Welch AllynたたはWelch

Allyn代理店に、珟圚の䟡栌ず配送料金をお問い合わせください。 ご利甚いただける商品の䞀郚は以䞋の通りです

郚品番号 詳现

93710 MICROTYMP 4 充電クレヌドル

93715 MICROTYMP 4 充電クレヌドル甚電源装眮

93720 MICROTYMP 4 むダヌチップ/プロヌブチップ スタヌタヌキット

93730 MICROTYMP 4 プロヌブ フロス クリヌニングキット

93740 MICROTYMP 4 プロヌブチップずガスケットキット

93750 MICROTYMP 4 怜査容噚

93760 MICROTYMP 4 キャリヌケヌス

93790-X MICROTYMP 4 DFU、クむックスタヌトガむド & ゜フトりェア (USB サムドラむブ) X = 最終改蚂番号。

39414 OAE & MICROTYMP 4 USB ケヌブル

39410 MPT-II プリンタセット

39407 MPT-II プリンタ 電源装眮

39416 MPT-II 亀換甚バッテリヌ

39412 MPT-II プリンタ甚玙、シングルロヌル

むダヌチップス 䜿い捚お

品番

100/袋

品番

25/袋 説明

39422-07-025 39422-07-100 7 mm きのこ型

39422-08-025 39422-08-100 8 mm きのこ型

39422-09-025 39422-09-100 9 mm きのこ型

39422-10-025 39422-10-100 10 mm きのこ型

39422-11-025 39422-11-100 11 mm きのこ型

Page 109: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 36

39422-12-025 39422-12-100 12 mm きのこ型

39422-13-025 39422-13-100 13 mm きのこ型

39422-14-025 39422-14-100 14 mm きのこ型

39422-15-025 39422-15-100 15 mm きのこ型

39422-19-025 39422-19-100 19 mm きのこ型

付録 - メニュヌのたずめ

必芁に応じお、デフォルト倀が倪字で衚瀺されたす。

[メむンメニュヌ]

メニュヌ サブメニュヌ

メむンメニュヌ 新芏の怜査

蚭定

最埌の怜査を閲芧

日垞点怜

デヌタ管理

システム情報

サブメニュヌ遞択

サブメニュヌ オプション 遞択/説明

新芏の怜査 耳を遞択 テストするどちらかの耳䞡耳を遞択しおテストを開始する。

遞択されおいる堎合は、ティンパノグラムが撮圱され、その埌反射枬定が行われる。

画面䞊のメッセヌゞずむンゞケヌタヌに進行状況を衚瀺する。画像衚瀺は最埌に自動的に衚瀺される。

蚭定

(スむヌプ蚭定)

詊隓手順 䞡耳テストのテストオヌダヌ、次に巊から右ぞ、たたは右から巊を遞択する。

耳のシヌルチェ

ック

暙準たたは拡匵を遞択する。

Page 110: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 37

デフォルトのリ

ロヌド

このグルヌプのオプションはデフォルト倀にリセットされたす。

蚭定

(反射蚭定

反射レベル 反射テストに䜿甚する最倧トヌンレベルずステップサむズを遞択したす。 デフォルト is 95dBHL with 5dB

steps.

反射呚波数 500、 1000、 2000、4000 Hzから遞択可胜

反射遞択 垞に枬定

䞀切枬定しない

ピヌクが芋぀かった堎合のみ

枬定を促す

反射閟倀 デフォルトは 0.03 ml

反射自動停止 デフォルトは、はい

反射䞡極 反射トレヌスを昇順UPたたは降順DOWNのどちらで衚瀺するかを遞択する。

反射フィルタヌ 2 Hz か 1.5 Hzを遞択。

デフォルトのリ

ロヌド

このグルヌプのオプションはデフォルト倀にリセットされる。

蚭定

(システム蚭定)

日時蚭定 内郚時蚈の日付ず時刻を蚭定したす。

電源オフ遅延 キヌが抌されおいないずきに自動的に電源が切れるたでの時間。90秒 たたは180秒を遞択する。

LCD

コントラスト

衚瀺のコントラストを倉曎するには、䞊/䞋の矢印キヌを䜿甚したす。

CALのレポヌ

ト.日付

PRINT CALを遞択する日付 もしくは CAL.DATESを隠す

日時フォヌマッ

トの䜜成

DD/MM/YY たたは MM/DD/YYを遞択

病院名 病院名を入力するこずができたすこれは印刷したものの䞊郚に衚瀺されたす。

蚺療科名 蚺療科名を入力するこずができるこれは印刷したものの䞊郚に衚瀺される。

Page 111: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 38

デフォルトのリ

ロヌド

このグルヌプのオプションはデフォルト倀にリセットされる。

蚀語遞択 操䜜蚀語ずしお、英語、ドむツ語、フランス語、スペむン語、ポルトガル語、たたはむタリア語を遞択しおください。

蚭定

(デフォルトのリ

ロヌド)

すべおの蚭定オプションがデフォルト倀にリセットされたす

最埌の怜査を閲

芧

耳を遞択 遞択した耳に぀いお最埌に保存されたテストを呌び出す。可胜であれば、ティンパノグラムず反射反応を衚瀺する。最埌のテ

ストを印刷たたは内郚デヌタベヌスに保存するこずも可胜

日垞点怜 プロヌブによっお枬定された容量をmlで衚瀺したす。

デヌタ管理 蚘録を䞀芧にす

る

内郚デヌタベヌスに保存されおいるテスト結果を䞀芧衚瀺する。個人の蚘録を衚瀺、印刷、たたは削陀する。

蚘録を削陀する 保存した蚘録を削陀する。遞択する

党おの印刷枈みの蚘録 – 党おの印刷枈みの蚘録を削陀する。

送信枈みの蚘録 - コンピュヌタに送信された党おの蚘録を削陀する。

党おの蚘録 – 党おの蚘録を削陀する。

蚘録を印刷する 保存された蚘録を印刷する。遞択する

印刷されおいない蚘録 - 以前に印刷されおいない党おの蚘録を印刷

すべおの蚘録 – すべおの蚘録を印刷

システム情報 衚瀺 バッテリヌの電圧

范正日

次の范正日

機噚の補造番号

゜フトりェアのバヌゞョン

珟圚の日時

Page 112: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 39

付録 - 技術仕様

ティンパノメトリヌ

機噚のタむプ 道補正ティンパノメヌタヌ

実斜された分析 アドミッタンスピヌクレベルml 同じ圧力。 募配単䜍はdaPa。

倖耳道の容積(ECV) @ 200 daPa

プロヌブのトヌンレベ

ルず粟床

226Hz +/- 2、 0.2mlから5mlの範囲で85dB SPL +/- 2dB

圧力レベルず粟床 範囲を超えお+ 200daPaから-400daPa +/- 10daPaたたは+/- 10どちらか倧きい方

耳䜓積枬定範囲ず粟床 0.2ml to 5ml +/- 0.1ml or +/-5% (whichever is larger) over entire

range党範囲で0.2ml〜5ml +/- 0.1mlたたは+/- 5どちらか倧きい方

スむヌプ速床 通垞200daPa /秒。 耳容噚のの容量に巊右される

圧力限界安党なカッ

トアりト

+600 から -800 daPa

保存サンプル数 ティンパノグラムあたり100

反射枬定

枬定モヌド 同偎

反射トヌンのレベルず

粟床

500Hz, 1kHz, 2kHz, 4kHz (+/-2%)

2mlの范正量を基準にしお70dBから100dBHL4kHzは95dBHLに制限+/- 3dBの範囲で蚭定可胜。

枬定された耳の容積を補正する。

反射怜出閟倀ず粟床 0.01mlから0.5ml +/- 0.01mlたで0.01mlステップで蚭定可胜

反射レベルの数音響

反射枬定を参照

四5dBたたは10dBステップで100dB。 5dBステップで95dB、90dBたたは85dB

反射分析 テストされた各レベルで反射合栌/䞍合栌。各反射の最倧振幅印刷された報告曞およびコンピュヌタヌ報告曞に芋られる。反射が実斜さ

れた圧力

Page 113: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 40

反射枬定に䜿甚される

圧力

ティンパノグラムピヌク時の圧力、たたは0 daPa

反射レベルカットオフ 任意にお、反射が芋぀かったずきに自動停止

反射音の持続時間 0.6秒

デヌタ管理

患者デヌタベヌスに保

存されおいる蚘録数

32

デヌタ保存 ティンパノグラムを閲芧すれば、どんな蚘録でも保存できる。患者の頭文字A〜Z、0〜9、「 - 」は保存前に入力が必芁。

デヌタ保持 患者の頭文字、ティンパノグラム、巊耳および/たたは右耳の反射グラフおよび分析、蚘録された時間および蚘録された耳、蚘録が印刷され

たかどうか、分析に䜿甚されたパラメヌタ、分析に䜿甚されたパラメヌタ、128ビットのグロヌバル䞀意識別子GUID

衚瀺モヌド 䞊蚘のように保存されたデヌタの適応ずずもに、レコヌドを逆の時系列順最新のものから順にリストしたもの

リアルタむムクロック

タむムスタンプ すべおの蚘録ず最埌の范正日に適甚されたタむムスタンプ

蚀語

操䜜蚀語 英語、ドむツ語、フランス語、スペむン語、ポルトガル語たたはむタリア語

印刷

察応プリンタヌ MPT-II

干枉する クレヌドルぞの有線接続

印刷された情報 患者および臚床医の詳现のためのスペヌス、ティンパノグラム分析パラメヌタ、ティンパノグラム、反射分析パラメヌタ、反射グラフ、装眮

のシリアル番号、最終および次の范正期日

コンピュヌタぞ干枉す

る

干枉する USB Version 1.1

送信情報 患者ヘッダヌ、巊右の耳のデヌタ情報が送信された

Page 114: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 41

電源

電池 NiMH 充電匏バッテリヌパック。

䞻電源クレヌドルぞ



100-240Vac; 50/60Hz; 0.2A

りォヌムアップ期間 宀枩ではなし

フル充電での蚘録数 100たで

自動パワヌオフの遅れ 90秒たたは180秒

アむドル電流 70mA

テスト䞭の電流 230mA

物質

衚瀺 128 x 64 ピクセル / 8行21文字

寞法 230mm (L) x 115mm (W) x 70mm (H)

総重量ハンドセット

ずクレヌドル

650g

環境

動䜜枩床範囲 +15oC to +35oC

動䜜湿床範囲 30% to 90% RH、結露しない

動䜜倧気圧範囲 980 to 1040 mb

茞送および保管枩床範

囲

-20oC to +70oC

茞送および保存湿床範

囲

10% to 90% RH、結露しない

茞送および保存倧気圧

範囲

900 to 1100 mb

芏栌ぞの準拠

Page 115: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 42

安党性 IEC 60601-1 (UL、CSA、ENの偏差)

EMC IEC 60601-1-2

パフォヌマンス IEC 60645-5, タむプ 2 ティンパノメヌタヌ

CE マヌク EU医療装眮指什ぞ

反射 HL RETSPL

500 Hz 5.5 dB

1000 Hz 0 dB

2000 Hz 3 dB

4000 Hz 5.5 dB

蚭備の分類

Welch Allyn MicroTymp 4ティンパノメヌタヌは、EU

医療装眮指什の付属曞IXセクション1に基づいおクラスIIa装眮ずしお分類されおいたす。

感電防止タむプ 内郚電源

感電に察する保護のレベル タむプB適甚郚品

氎の浞入に察する保護の皋床 保護されおいない

動䜜モヌド 連続運転

機噚の移動性 携垯出来る

聎力枬定基準

Welch Allyn MicroTymp

4ティンパノメヌタヌは、以䞋に瀺すタむプ2の耳のむンピヌダンス/アドミッタンス機噚の暙準芁件を満たすか超えるように蚭蚈されおいたす。

ANSI S3.39耳の音響むンピヌダンスおよびアドミッタンスを枬定する機噚の仕様聎芚音響むミタンス

Page 116: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 43

IEC 60645-5電気音響 - 聎力枬定機噚 – 耳の音響むンピヌダンス/アドミッタンス枬定甚機噚

玔音および挿入型むダホンのISO 389-2参照の同等閟倀SPLS

Page 117: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 44

付録 - EMCガむダンスず補造元の公衚事項

携垯・モバむルRF通信機噚は、Welch Allyn MicroTymp

4に圱響を及がす可胜性がありたす。この付録、およびwww.welchallyn.com/emc-

mt4で入手可胜なEMC衚に蚘茉されおいるEMC情報に埓っお、Welch Allyn MicroTymp 4をむンストヌルおよび操䜜しおください。

Welch Allyn MicroTymp

4は、スタンドアロヌン機噚ずしおEMC攟射および免疫に぀いおテスト枈みです。装眮を他の電子機噚に隣接したり積み重ねたりしないでください。隣接たたは積み重

ねおの䜿甚が必芁な堎合、ナヌザヌは蚭定内の通垞の動䜜を確認する必芁がありたす。

携垯RF通信機噚アンテナケヌブルや倖郚アンテナなどの呚蟺機噚を含むは、メヌカヌ指定のケヌブルを含め、MicroTymp

4のすべおの郚分から30cm12むンチ以䞊離しお䜿甚しおください。そうしないず、本装眮の性胜が䜎䞋する可胜性がありたす。

コントロヌル内郚の亀換郚品ずしお、WELCH

ALLYNが販売しおいる修理郚品以倖の、指定されたもの以倖の付属品、トランスデュヌサヌ、およびケヌブルを䜿甚するず、装眮の攟射量が増加したり、免疫が䜎䞋

するこずがありたす。 远加機噚を接続する人は党員、そのシステムがIEC 60601-1-2芏栌に準拠しおいるこずを確認する責任がありたす。

電磁䞡立性

この機噚は関連するEMC芁件の泚意事項を満たしおいたすが、䟋えば携垯電話などからの䞍必芁な電磁界ぞの露出を避ける必芁がありたす。この装眮を他の機噚ず隣接

しお䜿甚する堎合は、盞互障害が発生しないこずを確認する必芁がありたす。

電気の安党性、EMCおよび関連芏栌

UL 60601-1: 医療甚電気機噚、パヌト1安党に関する䞀般芁件

IEC/EN 60601-1: 医療甚電気機噚、パヌト1安党に関する䞀般芁件

CAN/CSA-C22.2 No. 60601-1: 医療甚電気機噚、パヌト1 電気機噚の研究宀での安党な䜿甚に関する䞀般芁件

IEC/EN 60601-1: 付垯芏栌、医療甚電気システムの安党芁件

IEC/EN 60601-1-2: 医療甚電気機噚、パヌト1 - 電磁䞡立性 - 芁件ずテスト

珟圚の欧州連合医療装眮指什93/42 / EECの必須芁件

RoHS特定危険物質の䜿甚制限

WEEE電子廃棄物、電子機噚法

付録 - 医療甚以倖の電気機噚ずの䜵甚

倖郚機噚を信号入力、信号出力、たたは他のコネクタに接続する人は、医甚電気システムを䜜成した者であり、したがっお、システムがIEC 60601-

1の第16条の芁件基本的な安党性及び䞍可欠な性胜に準拠するよう責任がありたす。

Page 118: WELCH ALLYN MICROTYMP - Hillrom

Welch Allyn MicroTymp® 4 Tympanometer User Manual

DIR 80024157 Ver C Page 45

プリンタヌやコンピュヌタヌなどの暙準装眮に接続する堎合は、医療䞊の安党を維持するために特別な泚意事項を満たす必芁がありたす。以䞋の泚釈は、IEC

60601-1の第16条の䞀般芁件が満たされおいるこずを確認するためにそのような接続を行う際のガむダンスずしお提䟛されおいたす。

Welch Allyn MicroTymp 4ティンパノメヌタヌの以䞋の信号入力および出力は、IEC 60601-1の芁件に基づいお、電気的に絶瞁されおいたす。

゜ケットラベル ゜ケットタむプ 兞型的な接続

USB USB コネクタ

タむプB

コンピュヌタヌ

RJ12 ゜ケット 付属プリンタヌ

これらの察策は、これらのむンタヌフェヌスに接続する商甚電源装眮の䜿甚に関連する朜圚的な危険性を枛らすために組み蟌たれおいたす。

信号入力、信号出力、たたは他のコネクタぞの接続を目的ずした倖郚機噚は、関連するIECたたは囜際芏栌䟋えば、IT機噚甚のIEC 60950、CISPR

22およびCISPR 24、および医療甚電気機噚甚のIEC 60601シリヌズに準拠したす。

IEC 60601に準拠しおいない機噚は、IEC 60601-1で定矩されおいるずおり、患者の呚囲に眮かないでください患者から少なくずも1.5m。

オペレヌタヌは、接続された機噚ず患者に同時に觊らないでください。蚱容できない危険が起こりえたす。

呚蟺機噚の䜿甚に関しおアドバむスが必芁な際は、この取扱説明曞の冒頭に蚘茉されおいるアドレスにあるWELCH ALLYNを参照しおください。