webinar - real time airborne viable particle detection 2012...data to support monitoring position...
TRANSCRIPT
Real Time AirborneViable Particle Detection
TechnologyOverviewandApplicabilityToPharmaceuticalQualitySystems
© 2012, TSI Incorporated
OverviewIntroductionTechnologyOverviewLaserInducedFluorescenceCriticalOperationalParameters
ApplicationScenariosApplicationroadmapReal‐timeevaluationofaseptictechniquesRootcauseinvestigationsQualityriskmanagement
© 2012, TSI Incorporated
Introduction: The Desired State“Amaximallyefficient,agile,flexiblepharmaceuticalmanufacturingsectorthatreliablyproduceshighqualitydrugproductswithoutextensiveregulatoryoversight”
Anopportunityforalessrestrictiveregulatoryparadigmanddrivesinnovationandcontinualimprovementinmanufacturingprocesses
QbD ICH Q8(R2)
QRM ICH Q9
PQS ICH Q10
PQSICH Q10
QbDICHQ8(R2) QRM
ICH Q9
PAT
PAT
© 2012, TSI Incorporated
Pursuit of Faster Results:Traditional Microbial Monitoring
Process Environment
Review Levels
Above LimitBelow Limit
Sampling Incubation Results Reaction
2 - 14 Days
© 2012, TSI Incorporated
Pursuit of Faster Results:Lab-Based RMM
2-14 Days
Majority of RMM’s Reduce Incubation Time
8-48 hours
© 2012, TSI Incorporated
Pursuit of Faster Results:Real-Time Viable Particle Detection
+ BioTrak™Detector• NewcategoryofRMM• Opticalspectroscopy• LaserInducedFluorescence(LIF)
+ Real‐timeresults
Seconds
© 2012, TSI Incorporated
Risk Analysis
Root Cause Investigations
Segregate Product
TrainingQualification
QRM ICH Q9
CAPA
QbD ICH Q8
PAT
Environmental Monitoring in Real-Time
Real-time
viable particle
data
KnowledgeInformation
FMS
Respond in real-
time
© 2012, TSI Incorporated
ISO 21501-4 compliant particle counter 28.3 lpm
Three Stages of TSI’s BioTrak™ Real-Time Particle Counter
Particle Counter
Viability Detector
Particle Concen-
trator
Collection filter
HEPA filter
37mm Collection Filter for speciation of analyzed particles
Viability detector-Laser Induced Fluorescence
© 2012, TSI Incorporated
• Proven TSI AeroTrak 9510 Airborne Particle Counter Engine
• ISO 21501-4 Compliant Specification and Calibration
• Full Airborne Particle Counter Functionality
Airborne Particle Counter
Particle Counter
Viability Detector
Particle Concen-
trator
Collection filter
HEPA filter
© 2012, TSI Incorporated
Particle Concentrator
Clean air and small particles
Most particles of interest
The flow is slowed down for the viability detector using a concentrator
28.3 lpm flow is too fast viability detection
© 2012, TSI Incorporated
Particle Concentrator
What impact does the concentrator have on the particle transport
efficiency?
Air Samplers also have particle transport
efficiency concerns.
© 2012, TSI Incorporated
Laser Induced FluorescenceEx
cita
tion
Wav
elen
gth
Emission Wavelength
Laser Induced Fluorescence (LIF)
Viable cell metabolites tryptophan, NADH, and flavin’s (riboflavin) fluoresce when
excited by ultra-violet light
“Advanced Trigger Development” Jeyes et al, Lincoln Laboratory Journal Volume 17, Number 1, 2007
© 2012, TSI Incorporated
Laser Induced Fluorescence• Wanttodifferentiatethemicroorganismsfromthenonviableparticleswithfluorescenceproperties
• Difficulttaskevenwithfullexcitationandemissionspectruminformation
• Notpossibletoimplementfullspectroscopyinaffordableportableinstruments
• TSICollaborativedevelopmentsince1990’s
• TSIhasfundamentalsingleparticleLIFpatents*
*TSI Patents 5,701,012; 5,895,922; 6,831,279
© 2012, TSI Incorporated
Laser Induced Fluorescence
© 2012, TSI Incorporated
Viability Detector
Particle Counter
Viability Detector
Particle Concen-
trator
Collection filter
HEPA filter
• Patented* Laser Induced Fluorescence (LIF) Viability Detection
• TSI collaborative technology development over the past 20 years
*Patents 5,701,012; 5,895,922 and 6,831,279
© 2012, TSI Incorporated
Discrimination-Impact of Additional Optical Parameters
1610/18/2012© TSI Incorporated
Single Fluorescence Channel
Two Fluorescence Channels
© 2012, TSI Incorporated
Discrimination-Another way to visualize
1710/18/2012© TSI Incorporated
+ Algorithms are used to calculate a discrimination parameter+ Particles with a discrimination value higher than a threshold
are classified as viable+ Particles with optical signatures near the intersection have
potential to be misclassified (False Positive or False Negative)
A
B
AB
Discrimination Parameter
© 2012, TSI Incorporated
Discrimination
1810/18/2012© TSI IncorporatedDiscrimination Parameter
BetterDiscrimination+ Numberofanalysis
parameters+ Algorithms+ Orthogonal
measurements
© 2012, TSI Incorporated
Sensitivity
Discrimination Parameter
Discrimination Parameter
© 2012, TSI Incorporated
LIF Dual Sensitivity
+ Abilitytoadjustinstrumentresponsetoapplicationspace
“Clean” Environments• Low number of viable particles• Low number of total particles• Lower number of particles with fluorescent characteristics• Unacceptable to miss viable particles• HIGH SENSITIVITY
“Dirty” Environments• Higher number of viable particles• Higher number of total particles• Don’t want excessive false positives• Higher limit of viable particles acceptable to miss some• LOW SENSITIVITY
© 2012, TSI Incorporated
Collection Filter
37mm Collection Filter: Provides sample for speciation of optically analyzed particles
Particle Counter
Viability Detector
Particle Concen-
trator
Collection filter
HEPA filter
© 2012, TSI Incorporated
Real-Time Viable Particle Counting -Important Considerations+ Needtounderstandtheproduct’scapabilitiesandlimitations
• EffectivenessofLIFMeasurement‐ Discriminationcapability‐ FalsePositivesversusFalseNegatives
• Functionality‐ Easeofuse‐ Totalparticulatesizingcapability‐ Viableparticledetectioncapability‐ Dataavailability
• Particlecapturecapability• RegulatoryChallenges
© 2012, TSI Incorporated
Characterization of Real-Time Viable Particle Counter
+ Needstobecharacterizedwithexistingmethods
+ Comparabilitystudies:efficiencyoftheinstrumentundertestandthereferencesamplermustbeconsidered
+ Desirecorrelationtoexistingmethods
© 2012, TSI Incorporated
Regulatory Guidance-Challenges
Compendial Method
Rapid Method
ISO 14698-1 USP 1223, EP 5.1.6, PDA TR33
Real-Time Aerosol Viable Detection
Active Air Sampling Comparison: Equivalent or Better
© 2012, TSI Incorporated
Applicability+ LIFresultswillbedifferentthancurrentactivesamplerbasedresults
+Whatisthevalue?• Real–timedata• Continuousdata• Characterizationofprocesses• Trainingfeedback
+ Evolutiontoreal‐timerelease
© 2012, TSI Incorporated
Application Road Map“The Innovation Highway”
Vendor Validation
Comparability Studies
Operator Training
Real-time
release
Root cause Investigations
Monitoring Locations+ Zone Qualification
Validated in Process
Measurement
Immediate benefit Long term benefit
Confiden
ce in
the measuremen
t
© 2012, TSI Incorporated
Risk Analysis
Root Cause Investigations
Segregate Product
TrainingQualification
QRM ICH Q9
CAPA
QbD ICH Q8
PAT
Environmental Monitoring in Real-Time
Real-time
viable particle
data
KnowledgeInformation
FMS
Respond in real-
time
© 2012, TSI Incorporated
New Information: Grade C Example
Grade C Monitoring LocationIn Operation
© 2012, TSI Incorporated
0.00
500.00
1000.00
1500.00
2000.00
2500.00
3000.00
3500.00
4000.00
4500.00
5000.00
08:24 09:36 10:48 12:00 13:12
TCNT
#/M^3
11 AM
ParticleCountscomplianttoISO21501‐4
0.5 µm Counts
5 µm Counts
Root Cause Investigations
Particle Counter
Viability Detector
Particle Concen-
trator
Collection filter
HEPA filter
© 2012, TSI Incorporated
0.00
10.00
20.00
8:24 9:36 10:48 12:00 13:12
VCNT
Viable Counts
12:30PM
Particleviabilitydetector
Root Cause Investigations
Particle Counter
Viability Detector
Particle Concen-
trator
Collection filter
HEPA filter
#/M^3
© 2012, TSI Incorporated
Root Cause investigations
0.00
500.00
1000.00
1500.00
2000.00
2500.00
3000.00
3500.00
4000.00
4500.00
5000.00
08:24 09:36 10:48 12:00 13:12
TCNT
0.00
10.00
20.00
8:24 9:36 10:48 12:00 13:12
VCNT
12:30 PM11 AM
2 different conclusions…..
OR
#/M^3#/M^3
© 2012, TSI Incorporated
0.00
500.00
1000.00
1500.00
2000.00
2500.00
3000.00
3500.00
4000.00
4500.00
5000.00
08:24 09:36 10:48 12:00 13:12
TCNT
RootcauseInvestigationsRoot Cause Investigations
#/M^3
© 2012, TSI Incorporated
Root Cause Investigations Collection Filter
Particle Counter
Viability Detector
Particle Concen-
trator
Collection filter
HEPA filter
© 2012, TSI Incorporated
RootCauseInvestigations
ICH Q8(R2) – Improved Process Knowledge
0.00
10.00
20.00
8:24 10:48 13:12
The BioTrak detector high temporal resolution means you know when a microbial
excursion occurred
Leads toImproved root cause
investigations
© 2012, TSI Incorporated
IdentifyingMicrobiologicalContaminationSources
ICH Q8(R2) – Improved Process Knowledge
InvestigationToolQuickly determinethesourceofairbornemicrobiologicalcontamination“GeigerCounter“mode– UserconfigurableBeepforViablecounts
© 2012, TSI Incorporated
RootCauseInvestigations
Leads to
ICH Q8(R2) – Improved Process Knowledge
Understanding the source of airborne microbiological contamination will help
determine the cause
Improved root cause
investigations
© 2012, TSI Incorporated
RootCauseInvestigations
Leads to
ICH Q8(R2) – Improved Process Knowledge
Improved root cause
investigations
More effective CAPA
© 2012, TSI Incorporated
QualificationTool:Earlyriskbasedreleaseof(manufacturing)areas‐Supportedbyexistingmethods:resultsdayslater
ICH Q9 – Quality Risk Management
Loc 2Loc 1
Loc 3 Loc 4
Loc 5
6m
Grade B
Loc 6
6mBioTrak
Real-time OPC and
Airborne Viables
OPC Monitoring
location
Environmental Monitoring
location
© 2012, TSI Incorporated
RiskAnalysis:DatatosupportmonitoringpositionrationaleforSettlePlates,ActiveAirSamplingandParticleMonitoring:Monitoringlocationfinetuning
ICH Q9 – Quality Risk Management
Loc 1Loc 1
Grade BBioTrak
Real-time OPC andAirborne Viables
?OPC Monitoring
location
Environmental Monitoring
location
© 2012, TSI Incorporated
Training‐ ImmediatenotificationofpoorpracticeGowning
ICH Q9: Quality Risk Management
© 2012, TSI Incorporated
Training‐ ImmediatenotificationofpoorpracticeAseptictechnique
ICH Q9: Quality Risk Management
© 2012, TSI Incorporated
Training‐ ImmediatenotificationofpoorpracticePoorcleanroompractice
ICH Q9: Quality Risk Management
© 2012, TSI Incorporated
Innovation HighwayReal-Time Release
Vendor Validation
Comparability Studies
Operator Training
Goal
Root cause Investigations
Monitoring Locations+ Zone Qualification
Validated in Process
Measurement
Short to medium term benefit Long term benefit
Confiden
ce in
the measuremen
t