· web viewhowever, the method of using a titanium rod has a clear advantage over the prosthetic...
TRANSCRIPT
Implantable End Pad for Lower Limb ProsthesisMike DiCicco, Stephen Osterhoff, Trevor Taormina – Projects 1, Dr Nasir
Biomedical Engineering Program, Lawrence Technological University, MI 48075
Abstract: Roughly 185,000 lower limb amputations occur each year in the United States. Amputations are performed relatively easy and come with little complications. However, complications arise during the socket fitting process. Due to the lack of load transfer, often time there is pain and discomfort leading to a decrease in quality of life. A new product known as the implantable end pad transfers weight bearing loads back onto the skeletal structure. The research plan along with the methods is reviewed as well as anticipated challenges, future research, impact, responsibilities, and the costs associated are discussed. Over the course of this study, material characterization and mechanical testing will form a basis for the improvement of transtibial amputations
Keywords: Amputation, Implantable End Pad, Transtibial, Characterization, Mechanical Testing
In America alone, there are roughly 185,000 lower limb amputations that occur each year
according to the Amputee Coalition of America. More specifically, 131,000 amputations occur
that are below the knee joint [1]. The majority of patients that are subjected to these amputations
fall under two main categories: disease or trauma. In fact, about 70% of patients that receive an
amputation are for diseases like vascular disease or diabetes while about 22% are seen from
trauma accidents [2].
The amputation process is a relatively simple, defined, and cost efficient process.
During the procedure, the surgeon often sections the tibia near the midline of the bone; that is,
roughly 8 to 10 inches from the knee joint. This is because when a prosthetic is formed for the
patient, enough bone length must be kept to provide proper stability and momentum during the
walking gait. In addition, enough bone length must be removed so that an efficient prosthetic is
able to implemented along with the necessary hardware. Once the bone is cut, the end of the
1 | P a g e
section is smoothed out to eliminate any jagged edges that could damage the surrounding tissue.
Existing muscle tissue is wrapped over the bone tightly and is sutured closed using skin flaps.
In the current market, the most notable technique for prosthetic limbs is attaching to the
patient via a socket. The socket is customized for each patient after recovery. During the
molding process, the prosthetist “shapes” the limb accordingly for optimized fit and to keep the
tibia from swinging within the socket. However as time passes, the socket can become loose due
to the change in the limb’s geometry. The changes can arise from the type of loading that is
exerted on the socket as well as fluid retention within the body. For example, if a patient has a
high intake of sodium throughout the day, more water will be retained in the body which will
change the limb ever so slightly causing a bad fitting with the socket. Bone bridging between the
tibia and fibula is another technique which has been used to assist with load distribution.
However, this technique does not have enough research to become a standard practice for
amputations. It can be seen through literature that building a bridge between the tibia and fibula
does not appear to have better outcomes than a standard transtibial amputation surgery [3].
Another newer process that is up and coming is attaching a titanium rod directly to the tibia and
connecting the prosthetic to the rod. The limitation with this process is whenever a medical
2 | P a g e
Figure 1: (left) Prosthetic socket fitting (middle) Titanium Rod fixation (right) Tibia-Fibula bone bridging
device protrudes from inside the body through the skin barrier, it is much more likely to cause an
infection at the site. However, the method of using a titanium rod has a clear advantage over the
prosthetic socket because of the load capabilities. In the prosthetic socket, there is no load
distribution that occurs between the tissue and the skeletal structure. The entire load in this case
is distributed over the soft tissue which often leads to bone spurs, skin sores, and other
complications that will reduce the overall quality of life for an estimated 80% of patients [1].
Implantable End Pad
To solve the patient’s complications and discomfort, the proposed solution is to develop a
product that is able to dampen or transfer weight bearing loads back to the skeletal structure.
This product is known as the implantable end pad (IEP) by Advanced Amputee Solutions, LLC
(AAS). This product will absorb the excessive forces that are
placed on the tissue and eliminate the complications that normally
come with prosthetic sockets. This is important because then the
prosthetic that fits over the limb will be able to work more
efficiently and cause less fit-up issues from limb changes.
A good deal of the benefits comes from its unique shape. The
unique tear drop shape, shown in Figure 3, is based on how
prosthetic sockets are fitted once the prosthetist shapes limb
accordingly. The half spherical bottom portion provides a comfortable fit inside the limb while
the cone shaped top provides stability. With the incorporation of an offset bone insertion site, it
not only provides stability on the top end but it also increases the amount of material on the
anterior side of the tibia. This is important because with the natural walking gait, the tibia is
subjected to higher loads on the anterior side.
3 | P a g e
Figure 3: Implantable End Pad
Research Plan
Literature Review
The first step in our research plan is to do a comprehensive literature review. In this
review, we aim to look at applications which the material has been used. Another area of
research is ISO (International Organization of Standardization) and ASTM (American Society
for Testing and Materials) standards required by the Food and Drug Administration (FDA) to
bring a product to market.
The material chosen was Bionate®, a polycarbonate-polyurethane composite
thermoplastic. From this, research and data about the polymer material was done about
biocompatibility and material characteristics. In literature provided by the polymer supplier
DSM, Bionate® has passed biocompatibility tests and maintains a FDA master file verifying
material biocompatibility with the FDA [4]. We also researched applications which the
Bionate® polymer was used. We have found that it has been used in load bearing prosthetic
products manufactured by Active Implants®. Active Implants® utilizes this material for hip and
knee prosthetics [5]. Material characterization studies were also researched and have found to
list various mechanical properties of the material [6]. From this research we determined
Bionate® to be a suitable polymer to use due to its biocompatibility and prior use in load bearing
prosthetics.
In our ISO and ASTM standards research, we have determined picked out key standards
to look at. ISO 10993 defines biocompatibility requirements needed prior to a clinical study.
ISO 14871 defines risk management assessments needed to determine the safety of the medical
device. ISO 13485 defines the quality management requirements needed for producing the
device. These ISO standards largely define requirements needed for later stages of producing the
4 | P a g e
implantable end pad and are less relevant during the design phase. One ASTM standard we have
found is ASTM D695-02. This ASTM standard defines test methods for compression properties
of rigid plastics. This standard outlines the basic procedure we will follow for our mechanical
testing.
Computational Modeling
The second step of research is to do computation modeling via finite element analysis
(FEA). To achieve this, a variety of software programs and packages will be used that are freely
available to LTU students. The first program is MIMICS®, which takes computed tomography
(CT) scans of the body and generates 3d models of the selected tissue or body part. MIMICS®
will be used to generate a 3D model of the tibia to be used to generate the bone interface portion
of the IEP. From here various computer aided design (CAD) software can be used to create and
edit the IEP model. The created IEP model can then be imported into COMSOL®, where finite
element analysis can take place. The FEA process will allow various designs of the IEP to be
tested under simulated loading cycles. This will allow for areas of stress to be visualized, from
here the model can be altered to reduce areas of high stress where failure can occur. This
process will allow us to create an optimal design without having to spend the time testing a real
world models.
Compression Testing
The final part of our research plan is to conduct mechanical testing. Using the ASTM
standard D695-02, we intend to conduct fatigue testing with material samples provided by DSM.
In addition to this, implant fixation methods are to be tested as well. We plan to test suturing and
bone cement fixation, to characterize how well each fixation method works. This is important
for the product because we believe that the next point of failure after material is at the point of
5 | P a g e
fixation. During the fixation test, cyclical stress mimicking the natural loading parameters of the
tibia will be applied to the bone and IEP device. After the compression test, analysis of the
fixation interface will be observed for degradation. In addition to this strain gages will be
utilized to determine the stress experienced by the material to help determine points of failure.
Methods and Tasks
In this project there are two main parts that will take place. The first part of the project
will model the IEP within the COMSOL® software package and simulate loading to find areas
of high stress and strain. This will be used to quickly design an IEP device that evenly
distributes loading and minimizes stress and strain of the material. In the second part the
experiment, test samples of material will undergo fatigue compression to determine which two
durometer of material that holds up the best to the stress. In addition to this, an injection molded
IEP device will be created and then attached to a bone model with various fixation methods. The
IEP and bone would then experience cyclical compression to see how well the fixation method
withstands the loading.
Computational Modeling
In the computational model portion of the project, finite element analysis software
COMSOL® will simulate the experiment. In the beginning stages of this, simple models created
directly in COMSOL® will be utilized to quickly set up loading parameters (Figure 4).
Once material properties and loading parameters have been correctly set, more complex and
accurate models can be imported into the software to produce better results (Figure 5).
In addition to using and modifying this scanned IEP geometry, several other software
packages will be used to generate the
geometry of the bone and IEP hole
6 | P a g e
shape. MIMICS® will be used to produce 3D models of sectioned tibias and generate the bone
interface portion (hole) of the IEP. From here, various CAD software import models from
MIMICS® and combine it with the IEP model. From here, the IEP and bone models are
imported into the COMSOL® software and the simulation will be run. Based on results of the
simulation the IEP model can also be modified in the CAD software to change shape and size.
This will allow us to rapidly design and test different IEP designs to find the best one to use in
the compression testing.
Specific testing that will be conducted in the FEA software will be cyclical compression
of the bone and device. The device will be loaded to approximately 3.3 times the body weight of
a 95th percentile based male weight. The 3.3 times body weight parameter was the max load
placed on the tibia in a normal adult male as reported by one scientific study [7]. This
compression will load to the peak value and unloaded in cycles and timeframes consistent with
human gait cycles. In addition to using these loading parameters for the computational
modeling, the same loading parameters will be used for the real would cyclical compression
study as well.
Compression Testing
In the compression testing study, mechanical testing equipment will be used to test both
material samples and injection molded IEP models. For material sample testing, sample disks
(see figure 6) will be tested to natural testing parameters as established in the previous section.
The sample disks will be tested before degradation of the material will be characterized. To
characterize the material degradation, optical microscopy and SEM imaging will observe the
material for deformation a fracturing. From this
process, two appropriate durometers of the
7 | P a g e
material will be chosen from the select four. When the appropriate Bionate® durometers have
been chosen, the IEP models will be created using injection molding techniques. Protocol for
injection molding the IEP has not been established, although outside contracting and in house
fabrication processes are being explored. Once the IEP models have been created, they will be
attached to either sawbone or pig bone samples using one of three fixation techniques: no
fixation (control), PMMA bone cement and surgical suturing. Each of these attachment methods
will be tested a minimum of two times, for a total of at least six trials for each durameter – 12
devices in total. Each trial will be placed in an electromechanical compression testing machine
located in CIMR laboratory at LTU. In addition to this each fixture will have strain gages placed
in high strain areas determined by the FEA process. This will allow for us to see if FEA
modeling was accurate in describing the stresses present. The fixtures will then be tested
according the natural testing parameters as established in the previous section. After cycling the
fixtures, optical microscopy and SEM imaging will be used to characterize the damage at the
fixation site to the cement, sutures and Bionate® material.
From this data, an analysis and report will conclude our findings about the material,
design and fixation method viability. The report will detail the key results, limitations and areas
of interest for further research.
Deliverables
The deliverables of this project will be broken down into 3 sections, computational
modeling, mechanical testing, and prototype development.
Computation Modeling
The computational modeling of this project will include the use of modeling programs
and a simulation program. The modeling programs are MIMICS®, SOLIDWORKS®, and
8 | P a g e
CATIA®. In MIMICS®, three dimensional models can be made from CT scans. This allows an
accurate representation of a human tibia for this project. Within SOLIDWORKS®/CATIA®, a
three dimensional model of the implantable end pad can be created. This model can then be
manipulated to optimize the performance of the end pad. This is one deliverable from modeling,
this will allow AAS to have a base model that they can change for future applications. Once the
model has been developed it can be imported into COMSOL®. COMSOL® is a program that is
used for finite element analysis. A test can be set, in this case a compression test, and initial
parameters of the material and test cycle can be inputted. This will then show profiles of where
stress occurs on the end pad and can be used to determine where strain gauges should be placed
to give relevant data.
Mechanical Testing
The first part of mechanical testing will consist of a material fatigue test that will help
determine what durometers of material should be used. This in combination with the
computational modeling will help us to verify that the correct material has been chosen. The
second part of mechanical testing will be to test fixation techniques between the end pad and the
bone. From this it can be determined that what is the best way to attach the end pad to the tibia
and though the use of strain gauges valuable data about the deformation of the material.
Prototype Development
Prototypes of the implantable end pad will be made from the durometers of material that
is chosen.
Anticipated Challenges
9 | P a g e
Some of the anticipated challenges we feel we might come up against along the way
while working on this projects starts with literature dealing with the FDA. The FDA has strict
regulations on medical devices before they are able to be marketed. Interpreting these
regulations in order to produce a quality product from FDA approval testing standards could
pose an issue. Another challenge deals with finite element modeling. The combination of
different programs, such as CATIA® or MIMICS®, and inputting data into COMSOL®, can
produce complications between files accurately. In addition, this modeling needs to ensure
accurate data amongst computational modeling and real mechanical testing. Mechanical testing
with the different methods of fixation – suturing and PMMA bone cement – that properly
simulates real applications could be difficult as well as understanding the machine interface of
the Instron instrument. And finally, during these mechanical testing, time constraints are bound
to occur due to the complexity and duration of fatigue and cyclical tests. For example, for a
typical 10 year study, to simulate the longevity of a medical device, it is placed under 10 million
cycles. This amount of cycles can last up to a month and half of continuous testing, leading to
clear time constraints.
Impact
The implantable end pad aims to help patients that underwent lower limb amputation,
specifically below the knee. Currently the only options a prosthetist and an amputee have is to
load all of the amputee’s weight onto the soft tissue at the site of amputation. This is a painful
experience for the amputee, because the body naturally works by loading the skeletal structure.
By implanting the end pad onto the bone, the load can then be transferred back onto the skeletal
structure. This will revolutionize the world of prosthetics. This will allow the prosthetist to
develop a better fitting socket that puts the load back where it should be. The stages that this
10 | P a g e
group is focusing on will help to develop the product for the company AAS. The data provided
will help show that this product will be a success.
Future Directions
There are many items that can be addressed for future plans. The things that can be
addressed are environmental testing, dual durometer material, customized bone interface, method
for mass production, and new socket design.
Environmental Testing
An aspect that still needs to be looked at in terms of the implantable end pad is how the
material will act when it is placed within the body. A simulation of this can be done with
environmental testing. This will have to be done in a few stages. The first is by creating an
environmental chamber that will simulate the human body. After a simple environment test,
mechanical loads can be placed on the end pad while it is in the environmental chamber.
Dual Durometer Material
One potential factor that could be looked at is creating an end pad that has a dual
durometer. This means that the end pad would be made out of a material that has two different
stiffness. This was suggested by the material supplier DSM. This could be accomplished by
molding an internal portion then adding another layer on top of that. This could possibly give a
better load for the bone.
Customized Bone Interface
Another focus that the CEO of AAS would like to see is a customized bone interface.
This means making the insertion site on the end pad fit to the particular person that it is being
used by. This can be used by some of the techniques that were used in this project. Computed
tomography scans can be taken of the bone before it is to be amputated. These scans can then be
11 | P a g e
loaded into the program MIMICS® and a 3D model of the bone can be created. This can then be
imported into CAD and an end pad model could be made to fit that bone. The end pad could
then be made through a process like 3D printing.
Method for Mass Production
Mass production is an area that needs to be investigated. Once the prototype models have
been made a decision needs to be made of what will be the most efficient way to produce the end
pad.
New Socket Design
A potential focus of this project it designing a new type of socket for the amputee once
the end pad has been implanted. Since the load transfer will be placed back onto the bone, a new
socket can be designed and patented.
Roles and Responsibilities
This team consists of Michael DiCicco, Trevor Taormina, and Stephen Osterhoff. For
our technical expertise, we have a well-rounded group consisting of our advisors Gordon
Maniere CEO of AAS, Dr. Nasir, and Dr. Meyer. Along with our vast contributors of Tech
Hwy, WEC Group, and a surgeon consultant Dr. Jon Ilhas of DMC and St. Mary’s Mercy
Hospital, we have a very dedicated team. The roles each of the engineering team members
overlap so we are all involved in the process of completing our goals. However, each of us are
leads in areas that allow us to accomplish these goals. Michael is most responsible for collecting
the data from the various tests, designing and refining the implantable end pad design, and
developing the protocol for our tests. Trevor is responsible for documenting our progress, the
lead on the mechanical testing, and the interpretation of the FDA documentation. Stephen is
12 | P a g e
responsible for our finite element analysis, fabricating the prototype device, and also FDA
interpretation as well. Each of these tasks highlights sections of our individual skillsets so we
are able to work independently, but also lead in group work.
Cost Analysis
These costs are based on what is currently available on the market and are subject to
change. The test equipment that is planning on being used is an Instron. Currently within
Lawrence Technological University there are two of these machines that are available for use.
The material that is being used for the IEP has been provided by the company DSM and they are
continuing to work in conjunction with this project to make sure all material needs are met. Test
fixtures will be made out of metal and any other parts that are needed (i.e. screws, nuts, bolts).
For fabrication of the test fixture Lawrence Tech has a fabrication lab that is available for use.
Strain gages will be bought prewired to save time and ensure quality. The electronic setup will
consist of a strain gauge monitoring system and any extra wires that may be needed. Proposals
will be made to LESA and NCIIA for the funding of this project.
Table 1: Projected cost analysis for project objectives
13 | P a g e
Materials Costs
Testing Equipment Provided by Lawrence Tech
IEP Material Provided by DSM
Test Fixtures $300-$500
Strain Gages $30-40 ea.
Electronic Setup (Strain Gages)
$100
Total $430-$640
**Planning to apply for funding
LESA, NCIIA Program
References
[1].."Understanding Limb Loss in the United States." Understanding Limb Loss in the United
States. Amputee Coalition, 27 Nov. 2012. Web. 23 Oct. 2013.
[2].. "Causes of Amputation." MossRehab.com. Web. 14 Dec. 2013.
[3].. Pinzur, M. S., J. Beck, R. Himes, and J. Callaci. "Distal Tibiofibular Bone-Bridging in
Transtibial Amputation." The Journal of Bone and Joint Surgery 90.12 (2008): 2682-687.
Print.
[4].."DSM in Medical." DSM. N.p., n.d. Web. 14 Dec. 2013.
[5].."Implant Materials." Implant Materials. N.p., n.d. Web. 14 Dec. 2013.
[6] Geary, C., C. Birkinshaw, and E. Jones. "Characterisation of Bionate polycarbonate
polyurethanes for orthopaedic applications." Journal of Materials Science: Materials in
14 | P a g e
Medicine 19.11 (2008): 3355-3363.
[7]..Wehner, Tim, Lutz Claes, and Ulrich Simon. "Internal Loads in the Human Tibia during
Gait."Clinical Biomechanics 24.3 (2009): 299-302
15 | P a g e