INTRODUCTION
WHAT IS CALIBRATION ?
• Calibration is set of operations that establish, under specified conditions, the
relationship between values of quantities indicated by a measuring instrument or
measuring system (Inspection, Measurement and Test Equipment-IMTE) , or values
represented by a material measure or a reference material, and the corresponding
values realized by standards.
• The goal of calibration is to guarantee the accuracy of the measurements which
are done by measuring and monitoring instruments (IMTE) that affect the
comformity of measurements, in accordance with ISO/TS management standards.
Calibration
Chief Engineering
ERDEMIR CALIBRATION CENTER
• ERDEMIR Calibration Center (ECC) started its operations in mid’s 1997.
• Primarily, Calibration Center makes periodic calibrations of 6300 IMTE which are in
Erdemir’s calibration inventory,
• ECC has been accredited on November 18, 2013 related TS EN ISO / IEC 17025:
2012 standard.
• All calibrations are traceable to international standards.
• Under ECC, there are five laboratories; these are Electronic, Temperature,
Dimensional, Pressure and Gas laboratories.
• All laboratories are dust, temperature and humidity controlled.
• Air conditioning of laboratories are maintained by separately established computer-
controlled air-conditioning plants.
Calibration
Chief Engineering
MISSION OF ECC
To realize periodic traceable calibrations of IMTEs that affect quality in ERDEMIR
in accordance with the Quality Management Systems and if there is no calibration
ability in Erdemir then to have them calibrated other firms.
Calibration
Chief Engineering
ECC OPERATIONS
1. ERDEMIR internal calibration services cover ;
By using general calibration procedures and related systems /
calibration work instructions,
• Adding IMTEs into inventory,
• Removing from inventory,
• Programming of calibrations ,
• Re-call for calibration,
• Calibrations of IMTEs,
• According to outcomes of calibration, evaluation of comformity and
certification labels of IMTEs
Calibration
Chief Engineering
ECC OPERATIONS ;
ERDEMIR external calibration services cover ;
providing calibration services in charge made, as demand,
to ERDEMIR’s subsidiaries ( ISDEMIR, CELBOR, STEEL SERVICE,
ERENCO),
to ERDEMIR’s suppliers (Companies doing business with Erdemir),
and to third parties.
Organizing calibration services that are available from
outside ERDEMIR .
• 42 units ECC calibrators (for monitoring),
• 57 units IMTEs that there is no calibration ability in ECC
(withdrawal/hardness/impact test machines, radiation
measuring instruments, etc.)
Calibration
Chief Engineering
ECC OPERATIONS
In ERDEMIR,
• 139 units IMTEs that are calibrated or to have these IMTEs calibrated by
Electronic Automation Maintenance teams,
• to have 6 units IMTEs calibrated by Auxiliary Operations and
Environment Management Department
are programmed to be calibrated and tracking of reporting operations.
Calibration
Chief Engineering
GENERAL INTRODUCTION
Erdemir Calibration Laboratories consist of five separate
laboratories below,
- Dimensional,
- Electronic,
- Temperature,
- Pressure,
- Gas Detector
and a production department.
All calibration areas are environmental controlled. (temperature, humidity and
particles)
Each calibration area has its specific min-max limitations of temperature and
humidity.
Calibration is not made when these limitations are gone beyond.
Calibration
Chief Engineering
PRODUCTION CONTROL CENTER
Registration, acceptance and post-delivery operations of equipments that are brought
to our laboratories for calibration are all carried out in production control center.
Calibration
Chief Engineering
Instrument / reference material types to be calibrated;
- Micrometer, Micrometer Setting Rod, Caliper, Dial indicator, Bore Gauge.
-Height gage, Feeler gauge, Steel Rule ,
-Tape Measure, Steel gauge.
- Steel Square, proctactor, precision level.
- Balance, Mass.
- Torque wrench.
Calibration Environment working conditions ;
* Relative Humidity % 45±25 RH and Temperature 20 ± 1°C .
* Limit value of particles number larger than 5 µm in 1 feet 3 volume is 700.
* Enlightenment values between 500-1000 lux.
ABILITIES OF DIMENSIONAL
CALIBRATION CENTER
Calibration
Chief Engineering
FEATURES OF DIMENSIONAL
LABORATORY CALIBRATORS NAME / WORK RANGE /
BRAND / MODEL MEASURE RANGE ACCURACY
SIZE AMPLIFIER /STARRETT / 715 0-1.3 mm ±2 mm
GAUGE SET /STARRETT / RC88MAA 0.5mm-10mm Grade 00 (DIN861)
GAUGE SET /STARRETT / SC81AA 0.1001mm-1.000mm Grade 00 (DIN861)
GAUGE SET /STARRETT / RS112MA1 0.5mm-25mm Grade 0 (DIN861)
GAUGE SET /MITUTOYO / 516-107 2.5mm-25 mm Grade 1 (DIN861)
GAUGE SET /MITUTOYO / 611 Serisi 50mm-800mm Grade 1 (DIN861)
HEIGHT GAGE /STARRETT / DHG1225 25-1225 mm ±1.3 mm
HEIGHT GAGE /STARRET / 258M 27-450 mm ±2 mm
COMPARATOR CALIBRATOR /MITUTOYO / 170-102 0-25 mm ±2 mm
I-CHECKER COMPARATOR CALIBRATOR 0-100 mm
Vertical
±(0.2+L/100)μm L: mm.
Horizantal
±(0.3+L/100)μm L: mm.
RING GAUGE /MITUTOYO / 177-290 F 40 mm ±2 mm
SURFACE PLATE /STARRETT / 80610 300x450 mm Grade 00 (DIN876)
SURFACE PLATE /JFA 2400x1800 Grade 0 (DIN876)
SINE BAR /BRUNSON / 470 0-1000 sec ±0.5 sec
ANGLE GAUGE SET /STARRETT / AG6.SI 0-99 deg ±0.5 sec
SQUARENESS CYCLINDER /HELIOS / 380/0/7 90 deg ±5 mm/ 800mm
SQUARENESS GAUGE /HELIOS / 377/0/12 90 deg ±11 mm/ 350mm
SINE PLATE /MITUTOYO / 930-626 0-60 deg ±15 sec
TORQUEMETER CALIBRATOR /DATA INS. / 6500-T4 0.0015-1.11 ft.lb ±%0.2
TORQUEMETER CALIBRATOR /CDI / 2000-600-0 10-2000 ft.lb ±%0.25
STD. WEIGHT SET /ZWIEBEL / XPCOF1 1g-10 kg ±0.1mg...±50mg (F1 class)
STANDARD WEIGHT /SARTORIUS / YWC 723-00 20 kg ±100 mg (F1 class)
STANDARD WEIGHT /SARTORIUS / YWC 753-01 50 kg ±250 mg (F1 class)
SCALE /M.TOLEDO / AT200 0-205 g ±0.15 mg
SCALE /M.TOLEDO / PR5003 0-5100 g ±0.003 g...±0.01 g
SCALE /M.TOLEDO / KB 50-2 0-60 kg ± 1 g
Calibration
Chief Engineering
SOME CALIBRATORS USED IN
DIMENSIONAL LABORATORY
87 units parallel gauge
set (0 class, ISO 3650) Micrometer gauge calibration kit
Calibration
Chief Engineering
SOME CALIBRATORS USED IN
DIMENSIONAL LABORATORY
Angle Generator (Brunson 470), Electronic Height Meter (Starrett DHG1225),
Squareness Gauge (Helios)
Calibration
Chief Engineering
Mitutoyo I-Checker Dial Indicator
Calibration System
SOME CALIBRATORS USED IN
DIMENSIONAL LABORATORY
SOME CALIBRATORS USED IN
DIMENSIONAL LABORATORY
Comparators Time
Calibrator
1g.-10 kg. OIML F1 class mass set
Calibration
Chief Engineering
SOME CALIBRATORS USED IN
DIMENSIONAL LABORATORY
50 kg. OIML F1 class mass Precision Balance (Capacity 200 g.)
Calibration
Chief Engineering
ABILITIES OF ELECTRONIC
CALIBRATION LABORATORY
Instrument types to be calibrated;
- Multimeters, pens ammeter, panelmeter,
- Oscilloscope, resistance boxes, current-voltage source,
- DC power supply,
- Recorder, process calibrator.
Calibration enviromental conditions ;
* Relative Humidity, % 45±25 RH and Temperature 23 ± 5°C dir.
* Limit value of particles number larger than 5 µm in 1 feet 3 volume is 700.
* Enlightenment values between 500 and 1000 lux.
Calibration
Chief Engineering
FEATURES OF ELECTRONIC
LABORATORY CALIBRATORS
WORK RANGE /
MEASURE RANGE ACCURACY
0-1100 V DC ±(9 ppm +500 mV) (1)
0-1100 V AC ±(90 ppm +4 mV) (2)
FLUKE / 5725 A 0-11 A DC ±(360 ppm +480 mA) (1)
0-11 A AC ±(460 ppm +170 mA) (3)
0-100 MW ±110 ppm (1)
0-30 MHz ±(%10.5 +15 mV) (1)
0-1020 V DC ±(%0.0055 +1500 mV) (1)
0-1020 V AC ±(%0.20 +500 mV) (1)
0-11 A DC ±(%0.06 +330 mA) (1)
0-11 A AC ±(%0.33 +2000 mA) (1)
0-330 MW ±(%0.5 +16.5 kW) (1)
0-2 MHz ±(25 ppm +15 mHz) (1)
0-1.1 mF ±(%1 +300 hF) (1)
T/C All T/C ranges ±0.84 ºC (1)
RTD All RTD ranges ±0.25 ºC (1)
1.8mV-105Vp-p(Square) ±(%0.25 +100 mV) (1)
10 Hz-10 kHz(Square) ±(25 ppm +15 mHz)
50kHz-250MHz(L.Sine.) ±(25 ppm +15 mHz)
5 mV-5.5 V(L.Sinewave) ±(%4 +300 mV) (1)
2ns-5s (Time Marker) ±(25 ppm +15 mHz) (1)
1.8mV-55Vp-p (Wave) ±(%3 p-p +100 mV)
0-1000 V DC ±(14 ppm+0.1ppm range)(1)
0-1000 V AC ±(%0.04+%0.002range) (4)
0-1 A DC ±(115ppm+10ppm range)(1)
0-1 A AC ±(%0.08+%0.02range) (5)
0-1 GW ±(%1+10 ppm range) (1)
1Hz-10 MHz ±%0.05 (1)
0-1000 V DC ±(%0.025+2 digit)
0-1000 V AC ±(%0.2+10 digit) (6)
0-10 A DC ±(%0.2+7 digit)
0-10 A AC ±(%0.5+10 digit) (6)
0-100 MW ±(%2+2 digit)
5Hz-1 MHz ±(%0.05+1 digit) (7)
NAME /
CALIBRATOR /
CALIBRATOR /
FLUKE / 5700 A
HP / 3458A
Osilo
sko
p
MULTIMETER /
FLUKE / 45
BRAND/MODEL
FLUKE / 5500 A
MULTIMETER /
Calibration
Chief Engineering
SOME CALIBRATORS USED IN
ELECTRONIC LABORATORY
Fluke 5700A-5725A Calibrator,
HP 3458A Multimeter
Fluke 5500A Calibrator,
HP 3458A Multimeter
Calibration
Chief Engineering
SOME CALIBRATORS USED IN
ELECTRONIC LABORATORY
Precise Resistance Box Fluke 45 Multimeter
Calibration
Chief Engineering
ABILITIES OF TEMPERATURE
CALIBRATION LABORATORY
Instrument types to be calibrated;
- Radiation thermometer (pyrometer), blackbody
furnace
- Digital or glass thermometer, thermo couple
- Resistance thermometer (pt-100), drying oven /
oven,
- Calibration bath,
- Temperature indicator with electrical signal input.
Calibration enviromental conditions;
* 23 ±5 °C temperature and % 45 ± 25 RH relative
humidity.
* Enlightenment values between 500 and 1000 lux.
Oil Bath
Calibration
Chief Engineering
FEATURES OF TEMPERATURE
LABORATORY CALIBRATORS
NAME / WORK RANGE /
BRAND / MODEL MEASURE RANGE ACCURACY
BLACKBODY OVEN /
MIKRON / M36050-1100 ºC ±(%0.2 + 1 ºC)
BLACKBODY OVEN /
IRCON / 54332350-1200 ºC ±%0.5
BLACKBODY OVEN
MIKRON / M330300-1700 ºC ±(%0.25 + 1 digit)
PYROMETER /
RAYTEK / RAYMX2D-30...+900 ºC ±%1 or ±1 ºC
PYROMETER /
MIKRON / M190-QTS200-1600 ºC ±0.8 ºC ...±2.6 ºC
CLB. BATH /
HART SCIEN./ 7025-20...+110 ºC ±0.5 ºC
CLB. BATH /
HART SCIEN./ 603535-300 ºC ±0.5 ºC
OVEN /
HERAEUS / TPK0-1200 ºC ±2.5 ºC
RTD / -200...0 ºC ±0.010 ºC
HART SCIEN./ 5614 0-500 ºC ±0.033 ºC
REF. THERMOCOUPLE / 0-600 ºC ±1.5 ºC
ERDEMİR / S Type 600-1767 ºC ±%0.25
TEMPERATURE SCANNER All T/C ranges ±0.21 ºC ... ±0.43 ºC
HART SCIEN./ 1560 0-25 W ±0.001W
25-400 W ±40 ppm
-100 ºC ...+600 ºC ±0.006 ºC ...±0.04 ºC
Calibration
Chief Engineering
SOME CALIBRATORS USED IN
TEMPERATURE LABORATORY
Miscellaneous Ovens (Black body) and
Reference Pyrometers
Temperature
Scanner
Calibration
Chief Engineering
ABILITIES OF PRESSURE
CALIBRATION LABORATORY
Device types to be calibrated;
- Pressure Transmitter, Pressure Transducer,
- Pressure meter, manometers (cistern, well type),
- Vacuum / Pressure Calibrator.
Calibration enviromental conditions ;
* 23 ± 3°C temperature and % 45 ± 25 RH relative humidity.
* Limit value of particles number larger than 5 µm in 1 feet 3 volume is
700.
* Enlightenment values between 500 and 1000 lux.
Calibration
Chief Engineering
FEATURES FACTS OF PRESSURE
LABORATORY CALIBRATORS NAME / WORK RANGE /
BRAND / MODEL MEASURE RANGE ACCURACY
Hook Gage /
Dywer -550...+550 mmH2O ±0.025 mmH2O
Deadweigth Tester (Havalı) /
Budenberg / 451A-1...+1 bar ±%0.008
Deadweigth Tester (Havalı) /
Budenberg / 452A0.1-7 bar ±%0.008
Deadweigth Tester (Yağlı) /
Ametek / DM-T-1000-700 bar ±%0.025
Digital Pressure Calibrator / -1 bar...-200 mbar ±%0.025
Druck / DPI605 -200...+200 mbar ±0.05 mbar
200 mbar - 20 bar ±%0.025
0-70 mbar ±0.035 mbar
70-350 mbar ±%0.05
0-70 bar ±0.011 bar
70-350 bar ±%0.08
Digital Pressure Calibrator / -400...+400 mbar ±(%0.04+0.07 mbar)
Beamex / PC106 -0.95...+4 bar ±(%0.04+0.3 mbar)
0-20 bar ±(%0.04+0.001 bar)
0-60 bar ±(%0.04+0.005 bar)
Digital Pressure Calibrator /
D&H / DH 3650-60 bar ±(%0.04 FS+%0.02)
Digital Pressure Calibrator /
Ametek / CE PPC 35BAR0-35 bar ±%0.1 FS
Analog Test Manometer /
Budenberg / 215F0-250 bar ±%0.25 FS
Analog Test Manometer /
Budenberg / 215F0-600 bar ±%0.25 FS
Calibration
Chief Engineering
SOME CALIBRATORS USED IN
PRESSURE LABORATORY
Pressure Calibrator
(Druck) Pressure Calibrator
(Desgranges Huot)
Calibration
Chief Engineering
SOME CALIBRATORS USED IN
PRESSURE LABORATORY
Reference Analog
Manometer Low Pressure Calibrator
(Hook Gage)
Calibration
Chief Engineering
FEATURES OF GAS DETECTOR
LABORATORY CALIBRATORS
Device types to be calibrated;
-CO - O2 - CH4 Gas Detector
(OLDHAM MX2000 etc.)
- CH4 Gas Detector (OLDHAM TX11 etc.)
- O2 Gas Detector (OLDHAM OX11 etc.)
- H2 Gas Detector (OLDHAM TX11 etc.)
Calibration environmental conditions;
23 ± 5 °C temperature and % 45 ± 25 RH relative humidity.
Calibration
Chief Engineering
USED CERTIFIED CALIBRATION GASES
GAS NAME ACTUAL VALUE
COMPANY THAT
TUBE WAS
PURCHASED
TRACEBILITY
CH4 % 50 LEL ELİTE S&T UK /
NMI, NIST
CO 250 ppm,
300 ppm ELİTE
S&T UK /
NMI, NIST
H2 % 50 LEL ELİTE S&T UK /
NMI, NIST
O2
% 18,0
hacimsel ELİTE
S&T UK /
NMI, NIST
C3H8 % 50 LEL DRAGER S&T UK /
NMI, NIST
REFERENCE GASES USED IN GAS
DETECTOR CALIBRATION
Example of Reference Gas Tubes Information Table
Calibration
Chief Engineering
Certified
CO - O2 - H2 - CH4
Gas tubes Portable Gas Detectors in
Different Brands
EXAMPLES OF REFERENCE GAS
TUBES AND GAS DETECTORS
Calibration
Chief Engineering
Calibration
Chief Engineering
Electronic Automation Department
Calibration Center Chief Engineering
THANK YOU