version 1.2 september 2012 · qui-1025 isdp manual 13 5. isdp operation for dry purging only if the...
TRANSCRIPT
QUI-1025 ISDP Manual
ULTRA Internal Standard / Dry Purge Manual
Version 1.2
September 2012
QUI-1025 ISDP Manual
Contents
1.1.1.1. IntroductionIntroductionIntroductionIntroduction ........................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................ 1111
2.2.2.2. Gas connectionsGas connectionsGas connectionsGas connections............................................................................................................................................................................................................................................................................................................................................................................................................................................................................................ 2222
2.1. Internal standard calibration gas..........................................................................................................2
2.2. Supply of purge (carrier) gas to the ISDP..............................................................................................3
3.3.3.3. Software setupSoftware setupSoftware setupSoftware setup .................................................................................................................................................................................................................................................................................................................................................................................................................................................................................................... 4444
4.4.4.4. Operation of the ISDPOperation of the ISDPOperation of the ISDPOperation of the ISDP............................................................................................................................................................................................................................................................................................................................................................................................................................................................ 6666
5.5.5.5. ISDP operISDP operISDP operISDP operation for Dry Purging onlyation for Dry Purging onlyation for Dry Purging onlyation for Dry Purging only ............................................................................................................................................................................................................................................................................................................................................................13131313
QUI-1025 ISDP Manual 1
1. Introduction
The internal standard / dry purge (ISDP) accessory for the Markes UltrA thermal desorption autosampler is
used to load a gas phase internal standard onto sorbent tubes loaded into the autosampler flowpath. It is also
used to dry purge humid samples. Standard and/or purge gases are introduced onto the tube from the
sampling end - i.e. gas flows onto the tube in the same direction as that used to collect sample. The tube and
flow path are leak tested before purge and/or standard gases are introduced. If a leak is detected the tube is
immediately returned intact to the tube tray (without purging or standard addition).
Following internal standard introduction / dry purge, the process of 2-stage thermal desorption proceeds
normally with pre-purge, tube desorption, etc.
QUI-1025 ISDP Manual 2
2. Gas connections
2.1. Internal standard calibration gas
The ISDP accessory requires a pressurised cylinder containing an appropriate calibration gas. The cylinder
must be equipped with appropriate (inert) step-down gas pressure regulation fitted with a length of 1/-8-inch
or 1/16-inch clean, stainless steel connection tubing. (Note that the length and volume of pressure regulation
equipment and connecting tubing should be kept as small as possible.) The pressure of internal standard gas
required to be delivered to UltrA is typically in the range (10-20) psig, and should be set (just (~2 psig) below
the pressure of the purge gas - see below).
A suitable internal standard would contain one or more gas-phase organic components that behave in a
similar way to the target analytes of interest, but are not found in the sample. Deuterated toluene is commonly
selected as an internal standard for BTX analysis on single-bed sorbent tubes. A mixture of deuterated n-
butane and deuterated n-octane could be used for calibration of a tube packed with two sorbents - e.g.
Carbograph 1 and Carboxen 1000 - used to monitor volatiles from n-C3 to n-C12.
The volume of calibration gas introduced onto the tube is determined by a sampling loop on a gas sample
valve incorporated into the ISDP. The standard loop supplied with the ISDP is 1 ml volume. This means that
the concentration of internal standard gas in the pressurised cylinder should be such that a 1 ml volume of
internal standard gas introduces a similar mass of internal standard compound as the mass of target analytes
collected during a typical monitoring exercise. For example, if the ISDP is to be used for calibration of 10 L
volumes of real air containing compounds at ~1 ppb concentration, the concentration of internal standard
should be approximately 10 ppm. This is because the mass of component in 1 ml of internal standard gas at
10 ppm is approximately equal to the mass of analyte(s) that would be collected from 10 L of air at
approximately 1 ppb.
The standard gas cylinder is connected to an 1/8-inch,
stainless steel bulkhead fitting located on the rear
panel of the autosampler labelled A.
Note that if 1/16-inch narrow bore connecting tubing is
being used appropriate reducing unions will be
required.
As soon as the cylinder has been connected to the
bulkhead on ULTRA it may be opened. An on/off
solenoid valve located downstream of the sampling
loop prevents flow out of the cylinder with the ULTRA–
UNITY 2 in standby mode.
The flow of calibration gas through the loop is
controlled using a needle valve B and and measured at
vent C.
Pneumatic gas is connected via 4 mm plastic tubing to
the autosampler at 50 psi.
A
B
C
QUI-1025 ISDP Manual 3
2.2. Supply of purge (carrier) gas to the ISDP
The purge gas is used to sweep the contents of the loop onto the sorbent tube in the sampling direction and
also to purge the entire ISDP flow path, keeping it free of contamination. The pressure of the purge (carrier)
gas is regulated using conventional pressure pneumatics (and is usually set just above (~2 psig higher than)
the pressure of the calibration gas and just below (~2 psig lower than) the pressure of the carrier gas
supplying the UNITY thermal desorber and GC-column.) Purge gas pressure is typically set in the range 12-22
psig.
An independently regulated supply of pure purge gas -
usually carrier gas - is also required to be connected to
the appropriate 1/8-inch brass bulkhead on the rear
panel of the autosampler D.
The purge gas flow can be measured and adjusted
using the appropriate vent F and needle valve E shown.
D
E
F
QUI-1025 ISDP Manual 4
3. Software setup
To access the various control parameters, the ISDP accessory must be enabled in the software.
Go to View > Options
Go to the Configuration tab select and select
‘Enable’ in the Internal Standard section.
Ensure the tube ‘Tube Load Temperature’ is
low enough to prevent a loaded tube heating
up significantly during the leak test, prior to
internal standard addition or dry purging.
A typical value for this temperature is 50 °C.
QUI-1025 ISDP Manual 5
In Standard 2(3) stage desorption methods
several options are now available.
1. Dry Purge
A flow of carrier gas for dry purging sampled
tubes with a time in minutes
2. Inject standard
Upon selection of Inject Standard two
parameters become available; the inject
standard time and the loop fill time.
The Inject Standard time is the time, in
minutes, for which the internal standard loop is
purged with carrier gas to transfer the
standard to the sample tube.
The loop fill time is the amount of time, in
minutes, that the internal standard gas is
purged through the standard loop prior to
loading onto a sample tube. The value should
be long enough to ensure that, given the flow
of calibration gas set using the needle valve,
the regulators, connecting tubing and gas loop
are completely purged with calibration gas.
Enter the other desorption method parameters
for the method as described in the main UNITY
2 Operators manual.
QUI-1025 ISDP Manual 6
4. Operation of the ISDP
The ISDP accessory is controlled within the main
software. The process of internal standard addition
involves the stages illustrated schematically below.
QUI-1025 ISDP Manual 7
Whether the ISDP is being used for dry purge only or
internal standard addition, the tube loaded into
ULTRA is first leak tested to ensure the integrity of
the tube seals and all other parts of the flow path.
QUI-1025 ISDP Manual 8
If the tube passes the leak test and if standard
addition is required, the software then actuates a
solenoid valve to sweep calibration gas from the
cylinder and into the loop. Exhaust gases from this
part of the process are purged to vent. As described
above, It is important to allow sufficient loop fill time
to ensure that the composition of gas standard in
the loop at the end of loop fill time is representative
of that in the cylinder.
Given that the total void volume of the cylinder
pressure regulators, connection tubing to ULTRA and
the gas loop itself may be 10 ml or more, and given
also that ten times the volume of gas is needed to
completely purge one gas with another, calibration
gas flows and loop fill times are typically set at 50
ml/min and 2 minutes respectively.
NoteNoteNoteNote: After initial setup this is the stage where the
internal standard flow can be adjusted – it cannot
be adjusted while in standby.
QUI-1025 ISDP Manual 9
At the end of the loop fill time (if applicable) the gas
sampling valve moves and the supply of purge
(carrier) gas is opened to sweep (inject) the contents
of the loop onto the tube from the sampling end or
to begin the dry purge phase. The purge gas flow
used to load the standard onto the sorbent tube is
set by manually adjusting the purge flow needle
valve on the ULTRA bulkhead, see Section 2.2.
QUI-1025 ISDP Manual 10
At the end of the 'Inject Standard’ time / ‘Dry Purge’
time, the sequence continues with the next step of
conventional 2-stage thermal desorption - i.e. pre-
purge.
Note that the normal purpose of the pre-purge stage
is to purge air from the adsorbent prior to
desorption. However, given that purge (carrier) gas
has been purged through the tube during 'Std
Injection' / 'Dry Purge', the pre-purge time may be
minimised to 0.1 minutes to reduce the TD cycle
time if required.
QUI-1025 ISDP Manual 11
Following pre-purge, the sample tube is desorbed in
the normal way.
QUI-1025 ISDP Manual 12
Just prior to the end of tube desorption, the short
length of narrow-bore connecting tubing used to
deliver calibration gas during the 'Inject Standard’
phase is purged with carrier gas. This is stage is
called 'Line Purge'. It is reported in the status bar of
the operating software as 'ULTRA cleanup'. The
primary tube desorption time can be extended to
account for the line purge time if required, but this is
not usually necessary.
The final thermal desorption steps are identical to
normal ULTRA – UNITY 2 operation.
After desorption the tube is unloaded at the
temperature specified and returned to the tray as
normal - See main ULTRA Operators manual.
QUI-1025 ISDP Manual 13
5. ISDP operation for Dry Purging only
If the ISDP is only to be used for dry purging tubes, no cylinder of gas-phase internal standard is required.
However, an independently regulated supply of pure purge (carrier gas) must still be connected to the
appropriate 1/8-inch brass bulkhead on the back of UltrA.
In this mode, there is no equivalent time to the loop fill time used during internal standard addition, but pure
purge (carrier) gas does flow through the ISDP accessory and onto the sampling end of the sorbent tube
loaded into the autosampler, in a stage called 'Dry Purge' which is analogous to the 'Inject Standard’ time
described above.
As above, the flow of purge gas is measured and controlled by manually adjusting the purge flow needle valve
on the bulkhead at the back of UltrA and the 'Dry Purge’ time is set within the software.
Note that dry purging of a sample tube loaded into UltrA will not take place until that tube has successfully
passed the leak test.
After the dry purge stage has been completed, UltrA –UNITY 2 will resume the normal steps of 2-stage thermal
desorption operation. Note that if the ISDP is being used for dry purging tubes, the flowpath through the ISDP
accessory is purged briefly with purge (carrier) gas towards the end of tube desorption to ensure all parts of
the flow path are kept clean and free of contaminant ingress.
As the operation of the ISDP accessory is determined by entry of the 'Dry Purge', 'Inject Standard’ time or 'Std
Inject Inactive' parameters as described above, it is possible to generate an automatic thermal desorption
sequence containing multiple methods - some of which include standard addition, some dry purge and some
with neither.
Note that selection of a long (e.g. 10 minute) 'Inject Standard’ time effectively also dry purges tubes that are
being loaded with internal standard. The minimum volume of purge gas required to adequately transfer
calibration standard from the gas loop to the sample tube is at least 50 ml and 'Inject Standard’ and flows of
calibration gas, must be set accordingly.