version 1003 state of the art of indoor calibration of pyranometers and pyrheliometers
TRANSCRIPT
Version 1003
State of the art of indoor calibration of pyranometers and pyrheliometers
2Indoor calibration
Main points
• Most field pyranometers are calibrated indoors
• Many procedures for indoor calibration
• Not all optimally connected to ISO 98-3 GUM
• Industry requires straightforward approach
3
Industry
• Meteorology - Solar renewable energy • Site assessment• Installation performance• Professionalisation / IEC
4
Future
• A few high accuracy outdoor calibrations
• A lot of indoor facilities• Accredited labs
5
Conclusion
• Points for discussion• Normal Incidence NI calibration is
preferred (Diffuse Sphere Source DSS not)
• Uncertainty & accuracy of reference can be optimised
• Indoor calibration complies with GUM• Pyrheliometer indoor calibration must
be allowed by ISO
6
Myself
• Kees VAN DEN BOS• Director / owner Hukseflux Thermal
Sensors• Last 20 years sensor design
7
Hukseflux DR01 pyrheliometer
• Founded 1993
• Thermal sensors
• 15 employees
• 5 radiometry
8
9Hukseflux 2010
10Reolith thermal properties on moon rover
11Snow thermal conductivity
My interest
• Hukseflux company cannot work with outdoor calibration
• Our customers want a understandable accuracy statement
• Feedback• More questions than answers
12
Background
• Most pyranometers and pyrheliometers have indoor calibration
• Exception: highest accuracy (BSRN, outdoor)
• Exceptions on national level: Japan, China, … (outdoor)
13
Background
• Cost, time, weather; outdoor calibration is unacceptable to industry
•DISADVANTAGE: Indoor methods only work with reference type = field type
14
Present status (excerpt)
• Eppley, US Weather Bureau: indoor integrating diffuse source
• Kipp, Hukseflux: indoor normal incidence
• EKO: outdoor tracker with collimation tube
• KNMI: indoor (network) and outdoor (BSRN)
15
16ISO 9060
17ISO 9060
Background
• Measurement uncertianty is a function of:
• Characterisation / class • Calibration (+characterisation and
class)• Measurement & maintenance
conditions• Environmental conditions
(+characterisation and class)
18
Background
• Indoor calibration covered by ISO 9847
• Present ASME: “Indoor Transfer of Calibration from Reference to Field Pyranometers”
19
20ISO 9846
21ISO 9847 also indoor
22ISO 98-3 GUM
Hierarchy of Traceability
• A: Reference calibration (uncertainty)• B: Correction of reference to indoor
conditions (uncertainty)• C: Indoor calibration of field
instrument (uncertainty)• Indoor calibration uncertainty
estimate (A+B+C)• Field measurement uncertainty
estimate
23
Hierarchy of Traceability
• A: Reference calibration (uncertainty)• B: Correction of reference to indoor
conditions (uncertainty)• C: Indoor calibration of field
instrument (uncertainty)• Indoor calibration uncertainty
estimate (A+B+C)
24
25ISO 98-3 GUM
26Hierarchy of traceability
27
28Indoor calibration Normal Incidence NI
29ISO 98-3 GUM
30
Hierarchy of Traceability
• KNMI TR 235 "uncertainty in pyranometer and pyrheliometer measurements at KNMI in De Bilt".
31
Hierarchy of Traceability
• A: Reference calibration (uncertainty)• B: Correction of reference to indoor
conditions (uncertainty)• C: Indoor calibration of field
instrument (uncertainty)• Indoor calibration uncertainty
estimate (A+B+C)• Field measurement uncertainty
estimate
32
33
34ISO 98-3 GUM
NI Hierarchy of Traceability
• A: Reference calibration (uncertainty) (conditions and class)
• B: Correction of reference to indoor conditions (uncertainty)
• C: Indoor calibration of field instrument (uncertainty)
• Indoor calibration uncertainty estimate (A+B+C)
• Field measurement uncertainty estimate (conditions & class)
35
Strange…
• Errors in reference calibration re-appear in measurement errors
• Counted double• At least systematic errors (Zero offset
A and directional errors) can be avoided.
36
One step back
• Calibration with restricted conditions results in lower uncertainty
• See yesterday’s presentation by Ibrahim Reda
37
One step back
• Present reference works well if calibrated pyranometers are used:
• Outdoor / unventilated• At same latitude
38
39
One step back
• Present approach does NOT work well calibrated if instruments are used:
• As indoor reference• At different latitudes• Ventilated
40
Typical secondary standard calibration• Irradiance 800 W/m2
• 40 to 60 degrees angle of incidence, + / - 30 degrees azimuth• Zero offset A: -9 +/- 3 W/m2 (larger
than ISO9060)• Directional: +/- 10 W/m2 @ 1000
W/m2 , now estimated +/- 5 W/m2
41
Typical calibration
• PMOD specified uncertainty +/- 1.3%• Systematic error -1%? Type B.
42
NI reference improved
• Restricted conditions• Zero offset A: -9 +/- 3 W/m2 (larger
than ISO9060)• Directional: +/- 10 W/m2
• Solution 1: ventilation• Solution 2: single angle of incidence
43
For consideration
• Japanese collimated tube with tilt correction and ventilation
• Tilted sun-shade method
44
45
Diffuse Sphere Source DSS
• Uniformity of sphere top-edge (experimental -13%)
• Weighing for non uniform source requires weighing of reference with source
• Diffuse sphere: weighing requires weiging of field instrument with source. Complicated!
• Normal incidence: weighing of field instrument is not necessary
46
DSS Hierarchy of Traceability
• A: Reference calibration (uncertainty) (conditions and class)
• B: Correction of reference to indoor conditions (uncertainty)
• C: Indoor calibration of field instrument (uncertainty) (conditions and class)
47
DSS Hierarchy of Traceability
• Indoor calibration uncertainty estimate (A+B+C)
• Field measurement uncertainty estimate (conditions & class)
• Additional uncertainty under C compared to NI calibration
• Bottom line: DSS has less restricted conditions than NI
48
49
Conclusion
• Indoor calibration offers only acceptable solution for manufacturers and “general users” in solar industry
• Indoor calibration fits within ISO 98-3 GUM
• detailed statements about field measurement still need to be agreed upon
50
Conclusion
• Indoor calibration: Normal Incidence calibration is preferred (Diffuse Sphere Source is not)
• Accuracy and precision of reference can be optimised to serve as indoor calibration reference (restricted: single angle, ventilated)
• Pyrheliometer indoor calibration must be added /allowed by ISO
51