version 1003 state of the art of indoor calibration of pyranometers and pyrheliometers

51
Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

Upload: wilfrid-stevens

Post on 18-Dec-2015

214 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

Version 1003

State of the art of indoor calibration of pyranometers and pyrheliometers

Page 2: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

2Indoor calibration

Page 3: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

Main points

• Most field pyranometers are calibrated indoors

• Many procedures for indoor calibration

• Not all optimally connected to ISO 98-3 GUM

• Industry requires straightforward approach

3

Page 4: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

Industry

• Meteorology - Solar renewable energy • Site assessment• Installation performance• Professionalisation / IEC

4

Page 5: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

Future

• A few high accuracy outdoor calibrations

• A lot of indoor facilities• Accredited labs

5

Page 6: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

Conclusion

• Points for discussion• Normal Incidence NI calibration is

preferred (Diffuse Sphere Source DSS not)

• Uncertainty & accuracy of reference can be optimised

• Indoor calibration complies with GUM• Pyrheliometer indoor calibration must

be allowed by ISO

6

Page 7: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

Myself

• Kees VAN DEN BOS• Director / owner Hukseflux Thermal

Sensors• Last 20 years sensor design

7

Page 8: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

Hukseflux DR01 pyrheliometer

• Founded 1993

• Thermal sensors

• 15 employees

• 5 radiometry

8

Page 9: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

9Hukseflux 2010

Page 10: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

10Reolith thermal properties on moon rover

Page 11: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

11Snow thermal conductivity

Page 12: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

My interest

• Hukseflux company cannot work with outdoor calibration

• Our customers want a understandable accuracy statement

• Feedback• More questions than answers

12

Page 13: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

Background

• Most pyranometers and pyrheliometers have indoor calibration

• Exception: highest accuracy (BSRN, outdoor)

• Exceptions on national level: Japan, China, … (outdoor)

13

Page 14: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

Background

• Cost, time, weather; outdoor calibration is unacceptable to industry

•DISADVANTAGE: Indoor methods only work with reference type = field type

14

Page 15: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

Present status (excerpt)

• Eppley, US Weather Bureau: indoor integrating diffuse source

• Kipp, Hukseflux: indoor normal incidence

• EKO: outdoor tracker with collimation tube

• KNMI: indoor (network) and outdoor (BSRN)

15

Page 16: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

16ISO 9060

Page 17: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

17ISO 9060

Page 18: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

Background

• Measurement uncertianty is a function of:

• Characterisation / class • Calibration (+characterisation and

class)• Measurement & maintenance

conditions• Environmental conditions

(+characterisation and class)

18

Page 19: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

Background

• Indoor calibration covered by ISO 9847

• Present ASME: “Indoor Transfer of Calibration from Reference to Field Pyranometers”

19

Page 20: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

20ISO 9846

Page 21: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

21ISO 9847 also indoor

Page 22: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

22ISO 98-3 GUM

Page 23: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

Hierarchy of Traceability

• A: Reference calibration (uncertainty)• B: Correction of reference to indoor

conditions (uncertainty)• C: Indoor calibration of field

instrument (uncertainty)• Indoor calibration uncertainty

estimate (A+B+C)• Field measurement uncertainty

estimate

23

Page 24: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

Hierarchy of Traceability

• A: Reference calibration (uncertainty)• B: Correction of reference to indoor

conditions (uncertainty)• C: Indoor calibration of field

instrument (uncertainty)• Indoor calibration uncertainty

estimate (A+B+C)

24

Page 25: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

25ISO 98-3 GUM

Page 26: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

26Hierarchy of traceability

Page 27: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

27

Page 28: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

28Indoor calibration Normal Incidence NI

Page 29: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

29ISO 98-3 GUM

Page 30: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

30

Page 31: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

Hierarchy of Traceability

• KNMI TR 235 "uncertainty in pyranometer and pyrheliometer measurements at KNMI in De Bilt".

31

Page 32: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

Hierarchy of Traceability

• A: Reference calibration (uncertainty)• B: Correction of reference to indoor

conditions (uncertainty)• C: Indoor calibration of field

instrument (uncertainty)• Indoor calibration uncertainty

estimate (A+B+C)• Field measurement uncertainty

estimate

32

Page 33: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

33

Page 34: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

34ISO 98-3 GUM

Page 35: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

NI Hierarchy of Traceability

• A: Reference calibration (uncertainty) (conditions and class)

• B: Correction of reference to indoor conditions (uncertainty)

• C: Indoor calibration of field instrument (uncertainty)

• Indoor calibration uncertainty estimate (A+B+C)

• Field measurement uncertainty estimate (conditions & class)

35

Page 36: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

Strange…

• Errors in reference calibration re-appear in measurement errors

• Counted double• At least systematic errors (Zero offset

A and directional errors) can be avoided.

36

Page 37: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

One step back

• Calibration with restricted conditions results in lower uncertainty

• See yesterday’s presentation by Ibrahim Reda

37

Page 38: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

One step back

• Present reference works well if calibrated pyranometers are used:

• Outdoor / unventilated• At same latitude

38

Page 39: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

39

Page 40: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

One step back

• Present approach does NOT work well calibrated if instruments are used:

• As indoor reference• At different latitudes• Ventilated

40

Page 41: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

Typical secondary standard calibration• Irradiance 800 W/m2

• 40 to 60 degrees angle of incidence, + / - 30 degrees azimuth• Zero offset A: -9 +/- 3 W/m2 (larger

than ISO9060)• Directional: +/- 10 W/m2 @ 1000

W/m2 , now estimated +/- 5 W/m2

41

Page 42: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

Typical calibration

• PMOD specified uncertainty +/- 1.3%• Systematic error -1%? Type B.

42

Page 43: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

NI reference improved

• Restricted conditions• Zero offset A: -9 +/- 3 W/m2 (larger

than ISO9060)• Directional: +/- 10 W/m2

• Solution 1: ventilation• Solution 2: single angle of incidence

43

Page 44: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

For consideration

• Japanese collimated tube with tilt correction and ventilation

• Tilted sun-shade method

44

Page 45: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

45

Page 46: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

Diffuse Sphere Source DSS

• Uniformity of sphere top-edge (experimental -13%)

• Weighing for non uniform source requires weighing of reference with source

• Diffuse sphere: weighing requires weiging of field instrument with source. Complicated!

• Normal incidence: weighing of field instrument is not necessary

46

Page 47: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

DSS Hierarchy of Traceability

• A: Reference calibration (uncertainty) (conditions and class)

• B: Correction of reference to indoor conditions (uncertainty)

• C: Indoor calibration of field instrument (uncertainty) (conditions and class)

47

Page 48: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

DSS Hierarchy of Traceability

• Indoor calibration uncertainty estimate (A+B+C)

• Field measurement uncertainty estimate (conditions & class)

• Additional uncertainty under C compared to NI calibration

• Bottom line: DSS has less restricted conditions than NI

48

Page 49: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

49

Page 50: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

Conclusion

• Indoor calibration offers only acceptable solution for manufacturers and “general users” in solar industry

• Indoor calibration fits within ISO 98-3 GUM

• detailed statements about field measurement still need to be agreed upon

50

Page 51: Version 1003 State of the art of indoor calibration of pyranometers and pyrheliometers

Conclusion

• Indoor calibration: Normal Incidence calibration is preferred (Diffuse Sphere Source is not)

• Accuracy and precision of reference can be optimised to serve as indoor calibration reference (restricted: single angle, ventilated)

• Pyrheliometer indoor calibration must be added /allowed by ISO

51