V5-02-2010

INSTRUCTION MANUAL

Stationary and mobile flow and consumption measurement for compressed air and gases

VA 400 VA 400 consumption sensorconsumption sensor

Optionen: VA 400 with display* VA 400 - Max. VA 400 - HighSpeed

* Operation menu for the display with capacitive key buttons

Page 2: VA 400 consumption sensor - CS Instruments GmbH · V5-02-2010 INSTRUCTION MANUAL Stationary and mobile flow and consumption measurement for compressed air and gases VA 400 consumption

2 V5-02-2010

FUNCTIONS Page Important information before initiation 3 Safety instructions 4-5 Technical data 6 Instruments description 7 Installation description 7-8 Assembly instructions 9 Measuring ranges VA 400 - Standard 10-11 Measuring ranges VA 400 - Max. 12-13 Measuring ranges VA 400 - HighSpeed 14-15 Max. velocities of the different gases 16 Drawing/Dimensions of the instrument / Pulse output 17 Pulse output signals / pulse lengths 18 Connection of the instrument 19 Service information 20 Display functions 21-24 Calibration/adjustment 25 Warranty 25 Ordering data 25 Accessories: Measuring sections 26 EC conformity declaration 27-28 Contact 32

INTRODUCTION Dear customer, thousands of customers buy our high standard products every year. There are a few good reasons for doing so:

• The cost-performance ratio - reliable quality at a fair price. • We have the ideal solutions for your measuring tasks based on our expert

experience gained over 20 years. • Our high quality standard. • Of course, our instruments carry the CE symbol required by the EU. • We issue calibration certificates and hold seminars. • We offer a good after sales service.

Our service guarantees fast help. Measuring instrument conforms with DIN EN 61326-1 and EN 61010-1

Page 3: VA 400 consumption sensor - CS Instruments GmbH · V5-02-2010 INSTRUCTION MANUAL Stationary and mobile flow and consumption measurement for compressed air and gases VA 400 consumption

3 V5-02-2010

IMPORTANT INFORMATION BEFORE INITIATION

Please read carefully before starting the device! The consumption sensor VA 400 measures the flow velocity (calorimatric principle) in the middle of the pipe. Please observe mounting instruction and inlet section = 15 x inner diameter and outlet section = 5 x inner diameter. The final values of the measuring ranges are as follows: VA 400 standard version 92.7 m/s, please take the flow rates from the tables on pages 10-11 VA 400 max. version 185 m/s, please take the flow rates from the tables on pages 12-13 VA 400 high speed version 224 m/s, please take the flow rates from the tables on pages 14-15 1. VA 400 with display with 4… 20 mA analogue and pulse output Please enter inner diameter of the pipe! Values indicated in the display: Actual value in m³/h, m³/min etc. Counter in m³ resp. l as well as pulse output, 1 pulse per m³ resp. l are calculated according to the set diameter. Please take the analogue value for flow rate 4… 20 mA from the tables on pages 10 to 15. 4 mA always corresponds with the starting value 0 m³/h, 0 m³/min. The final value 20 mA can be taken from the tables on pages 10 to 15. Example VA 400 Standard: 1" with inner diameter 25.0 mm, 4 mA = 0 m³/h and 20 mA = 122.2 m³/h 2" with inner diameter 53.1 mm, 4 mA = 0 m³/h und 20 mA = 600.0 m³/h 2. VA 400 without Display with 4… 20 mA analogue output (without pulse output) No adjustments are necessary at the consumption sensor. The respective final values for the flow rate can be taken from the tables on the pages 10 to 15. 4 mA is always set as scaling value 0. 20 mA is the final value . Example VA 400 Standard: 1" with inner diameter 25.0 mm, 4 mA = 0 m³/h and 20 mA = 122.2 m³/h 2" with inner diameter 53.1 mm, 4 mA = 0 m³/h und 20 mA = 600.0 m³/h

!

Page 4: VA 400 consumption sensor - CS Instruments GmbH · V5-02-2010 INSTRUCTION MANUAL Stationary and mobile flow and consumption measurement for compressed air and gases VA 400 consumption

4 V5-02-2010

Please read carefully before starting the device! Warning: Do not exceed the pressure range of 50. From 10 bar we recommend to use the high-pressure protection for a safe installation and removal. Observe the measuring ranges of the sensor! Overheating destroys the sensor. Observe the admissible storage and transporation temperature as well as the permitted operating temperature (e. g. protect the instrument from direct insolation). Always observe the direction of flow when positioning the sensor! The safety ring at the sensor head must always remain undamaged and sit correctly in the destined slot. The screwed fixture must be pressure tight. The adapter sleeve must be tightened with a torque of 20 to 30 Nm. It is absolutely necessary to avoid condensation on the sensor element or water drops in the measuring air as they may cause faulty . The values of the inlet and outlet sections must not fall below the specified minimum values as this causes increased deviations in the measuring results. The manufacturer cannot be held liable for any damage which occurs as a result of non-observance or non-compliance with these instructions. Should the device be tampered with in any matter other than a procedure which is described and specified in the manu-al, the warranty is cancelled and the manufacturer is exempt from liability. The device is destined exclusively for the described application. CS Instruments GmbH offers no guarantee for the suitability for any other purpose and is not liable for errors which may have slipped into this operation manual. CS Instru-ments GmbH is also not liable for consequential damage resulting from the delivery, ca-pability or use of this device. We offer you to take back the instruments of the instruments family VA 400 which you would like to dispose of. Adjustments and calibrations should only be carried out by qualified employees from the measurement and control technology branch.

SAFETY INSTRUCTIONS

!

Page 5: VA 400 consumption sensor - CS Instruments GmbH · V5-02-2010 INSTRUCTION MANUAL Stationary and mobile flow and consumption measurement for compressed air and gases VA 400 consumption

5 V5-02-2010

SAFETY INSTRUCTIONS

Please read carefully before starting the device! The consumption sensor VA 400 works according to the calorimetric measuring principle.

Burnable gases If this consumption sensor is used for measurement of burnable gases (e. g. natural gas) we explicitly point out that the sensor has no DVGW (= German Technical Association for Gas and Water) admission, however, it can be used for natural gas. A DVGW admission is not mandatory.

The consumption sensor VA 400 corresponds with the latest state of technology and can generally be used for burnable and non-burnable gases. For the use in e. g. natural gas the sensor will be calibrated in natural gas. The calibrati-on protocol (inspection certificate) is included in the scope of delivery. The area outside the pipe (environment of the sensor) is not allowed to be an explosive area.

The installation has to be done by authorized expert staff.

!

Page 6: VA 400 consumption sensor - CS Instruments GmbH · V5-02-2010 INSTRUCTION MANUAL Stationary and mobile flow and consumption measurement for compressed air and gases VA 400 consumption

6 V5-02-2010

TECHNICAL DATA

Measured variables: Volume flow (mass flow), total consumption, velocity Reference: DIN 1945, ISO 1217 at 20°C and 1000 mbar) (Standard factory setting) - other standards can be adjustec ex factory or by means of the CS Service Software Principle of measurement: calorimetric measurement Sensor: Pt45, Pt1000 Measuring medium: Air (Standard factory setting) Nitrogen, argon, CO2, oxygen, nitrous oxide, natural gas Operating temperature: -30 ... 110°C probe tube -30 ... 80 °C housing Operating pressure: up to 50 bar Analogue output: 4 ... 20 mA (max. burden < 500 Ohm) Scaling: 0 to maximum flow rate (see tables on pages 10-15) Accuracy: 0.06 mA Pulse output: see pages 18 to 19 Power supply: 12 to 30 VDC Power input: max. 80 mA at 24 VDC Accuracy: ± 3% m.v. with meas. section ± 2% m.v. (option via 5 point ISO precision calibration) These data are just valid in connection with a measuring section. Accuracy: ± 4% m.v. without meas. section ± 3% m.v. (option via 5 point ISO precision calibration) These data are just valid in case of correctly programmed inner diameter Display: 128 x 64 pixel, with backlight Measured values in maximum 6 digits, Counter max. up to 1,999,999,999, l resp. m³ drops then back to 0 Units: Standard factory settings m³/h , m³ selectable via software or pushbuttons at the display: m³/h, m³/min, l/min, l/s, kg/s, kg/min, kg/h, cfm, m/s Screw-in thread: G 1/2" Material: Probe tube and screwing: stainless steel 1.4301

Page 7: VA 400 consumption sensor - CS Instruments GmbH · V5-02-2010 INSTRUCTION MANUAL Stationary and mobile flow and consumption measurement for compressed air and gases VA 400 consumption

7 V5-02-2010

INSTRUMENTS DESCRIPTION VA 400 is a consumption sensor for consumption measurement of compressed air and gases. The version with integrated display indicates the actual consumption e. g. in m³/h and m³. Special features: - Option integrated display - units freely selectable in the display - consumption counter resettable - inner pipe diameter adjustable - Depth scale for accurate installation - Usable from tube diameter 1/2" (DN 15) - Easy installation under pressure - 4...20 mA analogue output for m³/h resp. m³/min - Pulse output for m³ Programming via Service Software SFA 300. - Analogue output 4...20 mA scalable - Switching between m³/h, m³/min, ft/min, l/min, l/s, cfm, m/s - Reading out the service data

Determining the point of installation In order to maintain the accuracy stipulated in the data sheets, the sensor must be inserted in the centre of a straight pipe section with an undisturbed flow progression. An undisturbed flow progression is achieved if the sections in front of the sensor (inlet) and behind the sensor (outlet) are sufficiently long, absolutely straight and without any obstruc-tions such as edges, seams, curves etc. Careful attention must be paid to the design of the outlet section as obstructions can cause counter-flow turbulence as well as turbulence in the direction of the flow.

L = Length of the entire measuring section L1 = Length of inlet section L2 = Length of outlet section D = Diameter of measuring section

INSTALLATION DESCRIPTON

LL 1 L 2

D

flow

D/2 D/2

D/2

D/2

Page 8: VA 400 consumption sensor - CS Instruments GmbH · V5-02-2010 INSTRUCTION MANUAL Stationary and mobile flow and consumption measurement for compressed air and gases VA 400 consumption

8 V5-02-2010

The following table shows the equalising sections necessary in relation to existing ob-structions:

INSTALLATION DESCRIPTION

Flow obstruction in front of the measuring section

Minimum length inlet (L1)

Minimum length outlet (L2)

Slight curve (bend < 90°) 12 x D 5 x D

Reduction (pipe narrows towards the meas. section) 15 x D 5 x D

Expansion (pipe expands towards the meas. section) 15 x D 5 x D

90° bend or T piece 15 x D 5 x D

2 bends á 90° on one level 20 x D 5 x D

2 bends á 90° 3-dimensional change of direction 35 x D 5 x D

Shut-off valve 45 x D 5 x D

Table of inlet and outlet sections

The respective minimum values required are indicated here. If it is not possible to observe the stipulated equalising sections, considerable deviations in measuring results must be expected. Sensor installation

Hint for the installation with ball valve: Ball valve R 1/2“, DN 15 Passage ball valve minimum Ø15 mm

Consider the flow direction which should match with the arrows at the connector head.

Measure from here to the tube center Read the insertion position here

The sensor head must be placed in the centre of the pipe. Therefore the probe shaft has a scale. To determine the right position measure the length from the marked position to the centre of the tube. The probe then will be inserted up to this measure and then tightened up.

mm

Page 9: VA 400 consumption sensor - CS Instruments GmbH · V5-02-2010 INSTRUCTION MANUAL Stationary and mobile flow and consumption measurement for compressed air and gases VA 400 consumption

9 V5-02-2010

ASSEMBLY INSTRUCTIONS

Assembly instructions Safety information must be observed. Assembly is carried out by inserting the connection thread (1/2“ thread, SW 27) into the connection piece. The sensor is then inserted to the required immersion depth and aligned according to the direction of air flow. A depth scale engraved on the probe tube, a flow alignment arrow and an aligning device will be of help to you. Once the sensor has been aligned, the adapter sleeve must be tightened with the stipulated torque (SW 17). Attention: Alignment of the sensor must not be modified when tightening the connection thread and adapter sleeve. In this case please check the immersion depth and alignment again and correct if necessary. The angular deviation should not be greater than ± 2° in re-lation to the ideal position as otherwise the measuring accuracy will decrease. Commissioning The valid measuring range and delivery configuration are programmed by the manufacturer on the basis of the user’s specifications. The stationary flow and air consumption measuring devices from the VA 400/ DS 300 se-ries function according to the “plug and play” principle. The device is ready for operation as soon as the power supply is connected. .

Page 10: VA 400 consumption sensor - CS Instruments GmbH · V5-02-2010 INSTRUCTION MANUAL Stationary and mobile flow and consumption measurement for compressed air and gases VA 400 consumption

10 V5-02-2010

Measuring ranges depending on the inner diameter of the pipe The sensor VA 400 - standard version with or without display, has a maximum measuring range of 92.7 m/sec. The flow rate is programmed to: Inner diameter: 53.1 mm This corresponds with an analogue output 4… 20 mA of 0…600 m³/h 0…10 m³/min 0…10000 l/min 0…166.6 l/sec 0…92.7 m/sec. In case of the version with display the inner diameter 25.00 has to be set first if the sensor is used in other pipes, e. g. 1", 25 mm. The analogue output for 1" can be taken from the table below: 4… 20 mA = 0…122.2 m³/h 0...2.04 m³/min 0…2036.6 l/min 0…33.94 l/sec 0…92.7 m/sec. In case of the version with display please adjust the respective inner diameter (see page 21).

Inner diameter of the pipe

Flow (final value of measuring range) max.

Inch mm m³/h m³/min l/min l/s m/s

1/4" 6.0 4.7 0.08 78.7 1.31 92.7 10.0 15.1 0.25 251.1 4.19 92.7 15.0 38.9 0.65 648.6 10.81 92.7

1/2" 16.1 45.6 0.76 760.8 12.68 92.7 3/4" 21.7 89.1 1.48 1484.9 24.75 92.7 1" 25.0 122.2 2.04 2036.3 33.94 92.7 26.0 132.9 2.21 2214.3 36.90 92.7 27.3 147.5 2.46 2457.5 40.96 92.7 28.5 162.0 2.70 2699.6 44.99 92.7 30.0 180.9 3.01 3014.8 50.25 92.7

1 1/4" 32.8 218.8 3.65 3646.2 60.77 92.7 36.0 266.3 4.44 4437.6 73.96 92.7 36.3 270.7 4.51 4511.9 75.20 92.7

1 1/2" 39.3 320.1 5.34 5335.7 88.93 92.7 40.0 332.5 5.54 5541.4 92.36 92.7 41.9 366.7 6.11 6111.0 101.8 92.7 43.1 389.4 6.49 6490.4 108.1 92.7

45.8 441.9 7.37 7365.7 122.7 92.7 2" 50.0 530.6 8.84 8844.1 147.4 92.7 51.2 557.1 9.29 9285.1 154.7 92.7 53.1 600.0 10.00 10000 166.6 92.7 54.5 632.8 10.55 10546 175.7 92.7 57.5 707.8 11.80 11797 196.6 92.7 60.0 773.6 12.89 12892 214.8 92.7 64.2 888.9 14.81 14814 246.9 92.7

MEASURING RANGES STANDARD

Referred to DIN 1945/ISO 1217 (20 °C, 1000 mbar) and compressed air

Page 11: VA 400 consumption sensor - CS Instruments GmbH · V5-02-2010 INSTRUCTION MANUAL Stationary and mobile flow and consumption measurement for compressed air and gases VA 400 consumption

11 V5-02-2010

Measuring ranges depending from the inner diameter of the pipe

MEASURING RANGES STANDARD

Sonde VA 400 - standard version up to 92.7 m/sec.

Inner diameter of the pipe

Flow (final value of measuring range) max.

Inch mm m³/h m³/min l/min l/s m/s

2 1/2" 65.0 913.5 15.22 15224 253.7 92.7 70.3 1071.1 17.85 17851 297.5 92.7 71.1 1095.6 18.26 18260 304.3 92.7 76.1 1258.2 20.97 20969 349.4 92.7

3" 80.0 1390.4 23.17 23173 386.2 92.7 81.0 1425.4 23.76 23756 395.9 92.7 82.5 1480.5 24.67 24674 411.2 92.7 84.9 1569.8 26.16 26162 436.0 92.7 90.0 1766.1 29.44 29435 490.6 92.7

4" 100.0 2183.1 36.38 36384 606.4 92.7 107.1 2507.1 41.78 41784 696.4 92.7 110.0 2644.7 44.08 44077 734.6 92.7

5" 125.0 3423.3 57.1 57055 950.9 92.7 133.7 3921.1 65.4 65351 1089.2 92.7

6" 150.0 4941.4 82.4 82356 1372.6 92.7 159.3 5579.8 93.0 92996 1549.9 92.7 182.5 7323.4 122.1 122055 2034.3 92.7 190.0 7947.1 132.5 132451 2207.5 92.7

8" 200.0 8816.2 146.9 146936 2448.9 92.7 206.5 9398.5 156.6 156642 2610.7 92.7

10" 250.0 13775 229.6 229587 3826.5 92.7 260.4 14945 249.1 249086 4151.4 92.7

12" 300.0 19836 330.6 330606 5510.1 92.7 309.7 21139 352.3 352331 5872.2 92.7 339.6 25418 423.6 423646 7060.8 92.7 388.8 33317 555.3 555291 9254.9 92.7 500.0 55101 918.4 918350 15305 92.7 600.0 79345 1322 1322424 22040 92.7 700.0 107998 1800 1799966 29999 92.7 800.0 141058 2351 2350976 39182 92.7 900.0 178527 2975 2975455 49590 92.7 1000.0 220404 3673 3673401 61223 92.7

Referred to DIN 1945/ISO 1217 (20 °C, 1000 mbar) and compressed air

Page 12: VA 400 consumption sensor - CS Instruments GmbH · V5-02-2010 INSTRUCTION MANUAL Stationary and mobile flow and consumption measurement for compressed air and gases VA 400 consumption

12 V5-02-2010

Measuring ranges depending from the inner diameter of the pipe

MEASURING RANGES MAX

The sensor VA 400 - max. version with or without display, has a maximum measuring range of 185.0 m/sec. The flow rate is programmed to: Inner diameter: 53.1 mm This corresponds with an analogue output 4… 20 mA of 0…1197.59 m³/h 0…19.96 m³/min 0…19959.88 l/min 0…332.66 l/sec 0…185.0 m/sec. In case of the version with display the inner diameter 25.00 has to be set first if the sensor is used in other pipes, e. g. 1", 25 mm. The analogue output for 1" can be taken from the table below: 4… 20 mA = 0…243.88 m³/h 0... 4.06 m³/min 0…4064.73 l/min 0…67.75 l/sec 0…185.0 m/sec. In case of the version with display please adjust the respective inner diameter (see page 21).

Inner diameter of the pipe

Flow (final value of measuring range) max.

Inch mm m³/h m³/min l/min l/s m/s

1/4" 6.0 9.42 0.16 156.92 2.62 185.0 10.0 30.08 0.50 501.28 8.35 185.0 15.0 77.68 1.29 1294.61 21.58 185.0

1/2" 16.1 90.98 1.52 1516.31 25.27 185.0 3/4" 21.7 177.84 2.96 2963.94 49.40 185.0 1" 25.0 243.88 4.06 4064.73 67.75 185.0 26.0 265.20 4.42 4419.99 73.67 185.0 27.3 294.72 4.91 4912.02 81.87 185.0 28.5 323.32 5.39 5388.74 89.81 185.0 30.0 361.08 6.02 6017.98 100.30 185.0

1 1/4" 32.8 436.69 7.28 7278.17 121.30 185.0 36.0 531.48 8.86 8857.96 147.63 185.0 36.3 541.06 9.02 9017.70 150.29 185.0

1 1/2" 39.3 639.84 10.66 10664.07 177.73 185.0 40.0 663.68 11.06 11061.30 184.35 185.0 41.8 728.41 12.14 12140.14 202.34 185.0 43.1 777.34 12.96 12955.60 215.93 185.0 45.8 882.17 14.70 14702.79 245.05 185.0

2" 50.0 1059.23 17.65 17653.79 294.23 185.0 51.2 1112.05 18.53 18534.19 308.90 185.0 53.1 1197.59 19.96 19959.88 332.66 185.0 54.5 1263.13 21.05 21052.15 350.87 185.0 57.5 1414.66 23.58 23577.72 392.96 185.0 60.0 1544.12 25.74 25735.30 428.92 185.0 64.2 1774.33 29.57 29572.14 492.87 185.0

Referred to DIN 1945/ISO 1217 (20 °C, 1000 mbar) and compressed air

Page 13: VA 400 consumption sensor - CS Instruments GmbH · V5-02-2010 INSTRUCTION MANUAL Stationary and mobile flow and consumption measurement for compressed air and gases VA 400 consumption

13 V5-02-2010

Measuring ranges depending from the inner diameter of the pipe

MEASURING RANGES MAX

Sensor VA 400 - max. version up to 185.0 m/sec.

Inner diameter of the pipe

Flow (final value of measuring range) max.

Inch mm m³/h m³/min l/min l/s m/s

2 1/2" 65.0 1821.03 30.35 30350.57 505.84 185.0

70.3 2137.86 35.63 35631.08 593.85 185.0

71.1 2186.80 36.45 36446.65 607.44 185.0 76.1 2511.24 41.85 41853.97 697.57 185.0

3" 80.0 2778.58 46.31 46309.59 771.83 185.0 82.5 2958.51 49.31 49308.50 821.81 185.0 84.9 3133.15 52.22 52219.09 870.32 185.0 90.0 3525.11 58.75 58751.80 979.20 185.0

4" 100.0 4357.22 72.62 72620.27 1210.34 185.0 107.1 5003.91 83.40 83398.43 1389.97 185.0 110.0 5278.56 87.98 87976.01 1466.27 185.0

5" 125.0 6824.50 113.74 113741.61 1895.69 185.0 133.7 7807.53 130.13 130125.42 2168.76 185.0

6" 150.0 9839.04 163.98 163984.07 2733.07 185.0 159.3 11096.91 184.95 184948.45 3082.47 185.0 182.5 14581.94 243.03 243032.33 4050.54 185.0 190.0 15805.08 263.42 263418.04 4390.30 185.0

8" 200.0 17533.48 292.22 292224.67 4870.41 185.0 206.5 18691.68 311.53 311527.93 5192.13 185.0

10" 250.0 27428.75 457.15 457145.91 7619.10 185.0 260.4 29793.76 496.56 496562.71 8276.05 185.0

12" 300.0 39544.48 659.07 659074.72 10984.58 185.0 309.7 42143.03 702.38 702383.91 11706.40 185.0 339.6 50673.25 844.55 844554.17 14075.90 185.0 400.0 70301.30 1171.69 1171688.40 19528.14 185.0 500.0 109845.79 1830.76 1830763.12 30512.72 185.0 600.0 158177.93 2636.30 2636298.89 43938.31 185.0 700.0 215297.74 3588.30 3588295.71 59804.93 185.0 800.0 281205.22 4686.75 4686753.58 78112.56 185.0 900.0 355900.35 5931.67 5931672.51 98861.21 185.0 1000.0 439383.15 7323.05 7323052.48 122050.87 185.0

Referred to DIN 1945/ISO 1217 (20 °C, 1000 mbar) and compressed air

Page 14: VA 400 consumption sensor - CS Instruments GmbH · V5-02-2010 INSTRUCTION MANUAL Stationary and mobile flow and consumption measurement for compressed air and gases VA 400 consumption

14 V5-02-2010

Measuring ranges depending from the inner diameter of the pipe

MEASURING RANGES HIGH SPEED

The sensor VA 400 - high speed version with or without display, has a maximum measuring range of 224.0 m/sec. The flow rate is programmed to: Inner diameter: 53.1 mm This corresponds with an analogue output 4… 20 mA of 0… 1450.06 m³/h 0… 24.17 m³/min 0… 24167.64 l/min 0…402.79 l/sec 0… 224.0 m/sec. In case of the version with display the inner diameter 25.00 has to be set first if the sensor is used in other pipes, e. g. 1", 25 mm. The analogue output for 1" can be taken from the table below: 4… 20 mA = 0… 295.30 m³/h 0... 4.92 m³/min 0… 4921.62 l/min 0… 82.03 l/sec 0… 224.0 m/sec. In case of the version with display please adjust the respective inner diameter (see page 21).

Inner diameter of the pipe

Flow (final value of measuring range) max.

Inch mm m³/h m³/min l/min l/s m/s

1/4" 6.0 11.40 0.19 190.00 3.17 224.0 10.0 36.42 0.61 606.96 10.12 224.0 15.0 94.05 1.57 1567.53 26.13 224.0

1/2" 16.1 110.16 1.84 1835.96 30.60 224.0 3/4" 21.7 215.33 3.59 3588.77 59.81 224.0 1" 25.0 295.30 4.92 4921.62 82.03 224.0 26.0 321.11 5.35 5351.77 89.20 224.0 27.3 356.85 5.95 5947.52 99.13 224.0 28.5 391.48 6.52 6524.74 108.75 224.0 30.0 437.20 7.29 7286.64 121.44 224.0

1 1/4" 32.8 528.75 8.81 8812.49 146.87 224.0 36.0 643.52 10.73 10725.32 178.76 224.0 36.3 655.12 10.92 10918.73 181.98 224.0

1 1/2" 39.3 774.73 12.91 12912.18 215.20 224.0 40.0 803.59 13.39 13393.14 223.22 224.0 41.8 881.96 14.70 14699.41 244.99 224.0 43.1 941.21 15.69 15686.78 261.45 224.0 45.8 1068.14 17.80 17802.30 296.71 224.0

2" 50.0 1282.52 21.38 21375.40 356.26 224.0 51.2 1346.48 22.44 22441.40 374.02 224.0 53.1 1450.06 24.17 24167.64 402.79 224.0 54.5 1529.41 25.49 25490.17 424.84 224.0 57.5 1712.89 28.55 28548.16 475.80 224.0 60.0 1869.63 31.16 31160.58 519.34 224.0 64.2 2148.38 35.81 35806.27 596.77 224.0

Referred to DIN 1945/ISO 1217 (20 °C, 1000 mbar) and compressed air

Page 15: VA 400 consumption sensor - CS Instruments GmbH · V5-02-2010 INSTRUCTION MANUAL Stationary and mobile flow and consumption measurement for compressed air and gases VA 400 consumption

15 V5-02-2010

Measuring ranges depending from the inner diameter of the pipe

MEASURING RANGES HIGH SPEED

Sensor VA 400 - high speed version up to 224.0 m/sec.

Inner diameter of the pipe

Flow (final value of measuring range) max.

Inch mm m³/h m³/min l/min l/s m/s

2 1/2" 65.0 2204.93 36.75 36748.79 612.48 224.0 70.3 2588.55 43.14 43142.50 719.04 224.0 71.1 2647.80 44.13 44129.99 735.50 224.0 76.1 3040.63 50.68 50677.24 844.62 224.0

3" 80.0 3364.33 56.07 56072.15 934.54 224.0 82.5 3582.20 59.70 59703.26 995.05 224.0 84.9 3793.65 63.23 63227.43 1053.79 224.0 90.0 4268.24 71.14 71137.32 1185.62 224.0

4" 100.0 5275.76 87.93 87929.41 1465.49 224.0 107.1 6058.78 100.98 100979.72 1683.00 224.0 110.0 6391.34 106.52 106522.31 1775.37 224.0

5" 125.0 8263.17 137.72 137719.57 2295.33 224.0 133.7 9453.44 157.56 157557.27 2625.95 224.0

6" 150.0 11913.22 198.55 198553.68 3309.23 224.0 159.3 13436.25 223.94 223937.58 3732.29 224.0 182.5 17655.97 294.27 294266.18 4904.44 224.0 190.0 19136.96 318.95 318949.42 5315.82 224.0

8" 200.0 21229.73 353.83 353828.78 5897.15 224.0 206.5 22632.08 377.20 377201.39 6286.69 224.0

10" 250.0 33211.03 553.52 553517.21 9225.29 224.0 260.4 36074.61 601.24 601243.50 10020.73 224.0

12" 300.0 47880.89 798.01 798014.80 13300.25 224.0 309.7 51027.24 850.45 850454.04 14174.23 224.0 339.6 61355.72 1022.60 1022595.32 17043.26 224.0 400.0 85121.58 1418.69 1418692.98 23644.88 224.0 500.0 133002.47 2216.71 2216707.78 36945.13 224.0 600.0 191523.55 3192.06 3192059.20 53200.99 224.0 700.0 260684.83 4344.75 4344747.24 72412.45 224.0 800.0 340486.31 5674.77 5674771.91 94579.53 224.0 900.0 430927.99 7182.13 7182133.20 119702.22 224.0 1000.0 532009.87 8866.83 8866831.11 147780.52 224.0

Referred to DIN 1945/ISO 1217 (20 °C, 1000 mbar) and compressed air

Page 16: VA 400 consumption sensor - CS Instruments GmbH · V5-02-2010 INSTRUCTION MANUAL Stationary and mobile flow and consumption measurement for compressed air and gases VA 400 consumption

16 V5-02-2010

MAXIMUM VELOCITY OF THE DIFFERENT GASES

[J/kg*K] Luft/Gas

VA 400 Standard VA 400 Max. VA 400 HighSpeed

Max. velocity [m/s] Max. velocity

[m/s] Max. velocity [m/s]

Reference DIN1945/ ISO 1217: 20°C, 1000 mbar (reference at calibration of the sensors)

Air ## ## 92.7 185.0 224.0

Adjustment to DIN 1343: 0°C, 1013.25 mbar

Air ## ## 85.2 170.1 206.0

Argon Ar ## ## 144.9 289.2 350.2

Carbon dioxyde CO2 ## ## 91.7 183.1 221.6

Nitrogen N2 ## ## 82.4 164.5 199.2

Oxygen O2 ## ## 88.4 176.4 213.6

Nitrous oxide N2O ## ## 90.9 181.4 219.7

Natural gas NG ## ## 54.8 109.4 132.5

Maximum velocities of VA 400 for different gases and reference standards

Please note: The consumption sensor VA 400 corresponds with the latest state of technology and can generally be used for burnable and non-burnable gases. If this consumption sensor is used for measurement of burnable gases (e. g. natural gas) we explicitly point out that the sensor has no DVGW admission, however, it can be used for burnable gases. A DVGW admission is not mandatory.

For the use in e. g. natural gas the sensor will be calibrated in natural gas. The calibration protocol (inspection certificate) is included in the scope of delivery.

The area outside the pipe (environment of the sensor) is not allowed to be an explosive area.

Page 17: VA 400 consumption sensor - CS Instruments GmbH · V5-02-2010 INSTRUCTION MANUAL Stationary and mobile flow and consumption measurement for compressed air and gases VA 400 consumption

17 V5-02-2010

Connector plug A Connector plug B

DRAWING OF THE INSTRUMENT/INSTRUMENTS DIMENSIONS

mm

Pulse output: 1 pulse per m³ (factory setting) With the option display VA 400 or in case of connection to a multifunction measuring instrument DS 300 the pulse output to litre, kg, cf is set via the setting of the consumption unit litre/min, kg/h, cfm. Active pulse to connector plug A (PIN 5)

In case of the version VA 400 without display the pulse is an active signal (not galvanically isolated) Pulse +P = +VB (12...30 VDC), max. current I = 10 mA, to connector plug A. If VA 400 is connected to a multifunction measuring instrument DS 300 DS 300 will convert the active pulse signal into a galvanically isolated pulse. Pulse diagram please see page 18. Galvanically isolated pulse to plug B (PIN 4 and PIN 5)

In case of the version VA 400 with display there is a galvanically isolated pulse available on connector plug B (PIN 4 and 5) in addition to the connector plug A (PIN 5). An isolated contact is clo-sed for the duration of the pulse (semi-conductor relay galvanically isolated by optocoupler). Maximum switching capacity: 30 VDC, 20 mA Pulse diagrams please see page 18.

Page 18: VA 400 consumption sensor - CS Instruments GmbH · V5-02-2010 INSTRUCTION MANUAL Stationary and mobile flow and consumption measurement for compressed air and gases VA 400 consumption

18 V5-02-2010

Pulse output signal indication Pulse lengths depending on consumption

Internal pulse receiver: The numbers of m³ per second are summed up and indicated after one second. Pulse lengths depending on consumption

Pulse output: max. voltage pulse +P = +VB (12 .. 30 VDC), active signal max. current I = 10 mA

PULSE OUTPUT / PULSE LENGTHS

Table of pulse lengths and maximum flow quantities

Pulse length [ms]

[m³/h]

[m³/min]

[l/min]

120 10,800 180 180 60 21,600 360 360 30 39,600 660 660

[cfm]

180 360 660

[kg/h]

10.800 21.600 39.600

[kg/min]

180 360 660

[kg/s]

3 6

11 10 129,600 2,160 2,160 2,160 129,600 2,160 36

Max. flow 129,600 2,160 2,160 2,160 129,600 2,160 36

Maximum 36 pulses can be given per second. Please observe: If the maximum flow is exceeded there will be no more signals! In this case please set the units e. g. from l/min to m³/h. !

Page 19: VA 400 consumption sensor - CS Instruments GmbH · V5-02-2010 INSTRUCTION MANUAL Stationary and mobile flow and consumption measurement for compressed air and gases VA 400 consumption

19 V5-02-2010

CONNECTION OF THE INSTRUMENT

If no connection cable ( 0553 0104, 0553 0105 ) is ordered the sensor will be sup-plied with a M12 connector plug. the user can connect the supply and signal cables as indicated in the connection diagram.

Pin 1 Pin 2 Pin 3 Pin 4 Pin 5

Connector plug A NC (SDI)

-VB +VB I+ 4..20 mA

NC (pulse)

active not isolated

Connector plug B only assigned with display NC NC NC

pulse isolated

pulse isolated

colours connection cables 0553.0104 (5 m) 0553.0105 (10 m)

brown white blue black grey

colours pulse cables 0553.0106 (5 m) 0553.0107 (10 m)

brown white blue black grey

-VB Negative supply voltage 0 V

+VB Positive supply voltage 12...30 VDC - smoothed

I + Current signal 4..20 mA - actual flow

Plug A PIN 1 (SDI)

is only required for connection to DS 300

Pulse Pulse for consumption

NC (SDI) Is not allowed to be put to potential and/or earth. Please cut off cables and isolate them

Plug A PIN 5

only for connection to DS 300

!

Legend:

Connector plug A View reverse side (clamp side)

Connector plug B PIN 4 and 5 (only assigned if VA 400 has a display) View reverse side (clamp side)

max. 30 VDC, 20 mA

black

blue

white black

grey

Page 20: VA 400 consumption sensor - CS Instruments GmbH · V5-02-2010 INSTRUCTION MANUAL Stationary and mobile flow and consumption measurement for compressed air and gases VA 400 consumption

20 V5-02-2010

SERVICE INFORMATION

Maintenance The sensor head should be checked regularly for dirt and cleaned if necessary. Should dirt, dust or oil accumulate on the sensor element, a deviation will occur in the measuring value. An annual check is recommended. Should the compressed air be heavily soiled this interval must be shortened. Cleaning of the sensor head The sensor head can be cleaned by carefully moving it to and fro in warm water with a small amount of washing-up liquid. Avoid physical intervention on the sensor (e. g. using a sponge or brush). If soiling cannot be removed, service and maintenance must be carried out by the manufacturer. Re-calibration If no customer specifications are given then we recommend to carry out calibration every 12 months. For this purpose the sensor must be sent to the manufacturer. Spare parts and repair For reasons of measuring accuracy spare parts are not available. If parts are faulty they must be sent to the supplier for repair. If the measuring device is used in important company installations we recommend to keep a spare measuring system ready. Calibration certificates Calibration certificates are issued by the manufacturer on request. This is a fee-paying service. Precision is tested with DKD (= German calibration authority) certificated volume flow meters.

Page 21: VA 400 consumption sensor - CS Instruments GmbH · V5-02-2010 INSTRUCTION MANUAL Stationary and mobile flow and consumption measurement for compressed air and gases VA 400 consumption

21 V5-02-2010

DISPLAY FUNCTION

Display VA 400 can display up to 3 measured values. This are volumetric flow/mass flow, velocity and total consumption. For better reading only one measured value is displayed. The measured val-ues are toggled during normal operation mode every 3 seconds. The following changes can be made : - Diameter - Unit volume flow (m³/min, m³/h, l/min, l/s, kg/s, kg/min, kg/h, cfm) - Display of measured value (volume flow, velocity, total consumption) - Zero consumption - Display contrast - Display turnable by 180 ° Settings ex works: - Volume flow in m³/h ( If the sensor is calibrated for reference DIN 1343, the unit is Nm³/h. This unit is only adjustable in the factory or with the CS Service Software.) - Total consumption in m³ - Diameter If not ordered by customer, the diameter will be preset to 53.1 mm.

On the top of the VA 400 are the capacitive key buttons to operate the display menu.

“Up“ ( ) “Enter“ ( ↵ )

Page 22: VA 400 consumption sensor - CS Instruments GmbH · V5-02-2010 INSTRUCTION MANUAL Stationary and mobile flow and consumption measurement for compressed air and gases VA 400 consumption

22 V5-02-2010

Connection VA 400

After power on, the display will go through an initialisation procedure and will finally show the ac-tual online values.

Slave mode Changing the settings only with DS 300

Master mode Changing the settings with push buttons VA 400.

DISPLAY FUNCTION

Page 23: VA 400 consumption sensor - CS Instruments GmbH · V5-02-2010 INSTRUCTION MANUAL Stationary and mobile flow and consumption measurement for compressed air and gases VA 400 consumption

23 V5-02-2010

Configurations-Settings

Display currently set diameter

A short click on the “Enter”-button shows the currently set diameter. Input diameter In order to change the diameter value, keep the "Enter“ button pressed for 3 seconds. After input of the unlock code the menu begins with diameter setting. The first digit will start blinking and can be changed with the "Up" key. Every digit has to be confirmed with the „Enter“ button.

DISPLAY FUNCTION

Page 24: VA 400 consumption sensor - CS Instruments GmbH · V5-02-2010 INSTRUCTION MANUAL Stationary and mobile flow and consumption measurement for compressed air and gases VA 400 consumption

24 V5-02-2010

DISPLAY FUNCTION

Menu You can exit the configuration process by pressing and keeping the “Enter” key for 3 sec or no key event for 20 sec. The configuration before this picture will be save and effected late.

Page 25: VA 400 consumption sensor - CS Instruments GmbH · V5-02-2010 INSTRUCTION MANUAL Stationary and mobile flow and consumption measurement for compressed air and gases VA 400 consumption

25 V5-02-2010

ORDERING DATA

Order no. Description

0695.4001 VA 400 consumption sensor without display including plug

Z695.4000 Display

Z695.4003 Max. version 185 m/s

Z695.4002 High speed version 224 m/s

0553.0104 Connection cable for VA/FA Series 400, 5 m with M12 plug

0553.0105 Connection cable for VA/FA Series 400, 10 m with M12 plug

ZSL.0120 Length 120 mm

ZSL.0160 Length 160 mm

ZSL.0220 Length 220 mm

ZSL.0300 Length 300 mm

ZSL.0400 Length 400 mm

0500.3000 DS 300 display for wall mounting

Options:

WARRANTY

If you have reason for complaint we will of course repair any faults free of charge if it can be proven that they are manufacturing faults. The fault should be reported immediately after it has been found and within the warranty time guaranteed by us. Excluded from this warranty is damage caused by improper use and non adherence to the instruction manual. The warranty is also cancelled once the instrument has been opened - as far as this has not been mentioned in the instruction manual for maintenance purposes - or if the serial number in the instrument has been changed, damaged or removed. The warranty time for the VA 400 is 12 months. If no other definitions are given the accessory parts have a warranty time of 6 months. Warranty services do not extend the warranty time. If in addition to the warranty service necessary repairs, adjustments or similar are carried out the warranty services are free of charge but there is a charge for other services such as transport and packaging costs. Other claims, especially those for damage occurring outside the instrument, are not included unless responsibility is legally binding. After sales service after the warranty time has elapsed We are of course there for you even after the warranty time has elapsed. In case of mal-functions please send us the instrument with a short-form description of the fault. Please do not forget to indicate your telephone number so that we can call you in case of any questions.

CALIBRATION

According to DIN ISO certification of the measuring instruments we recommend to calibrate and if applicable to adjust the instruments regularly from the manufacturer. The calibration intervals should comply with your internal specificaton. According to DIN ISO we recom-mend a calibration interval of one year for the instrument VA 400.

Page 26: VA 400 consumption sensor - CS Instruments GmbH · V5-02-2010 INSTRUCTION MANUAL Stationary and mobile flow and consumption measurement for compressed air and gases VA 400 consumption

26 V5-02-2010

MEASURING SECTIONS

G 1/2"-outer screw neck

30 mm 30 mm

Total length

Outer thread: G x Outer thread: G x

LA

DA

Measurement section for VA 300 probes

Outer dia.DA Length LA Pipe

21.3 * 2.6 mm, Stahl 1.4301

26.9 * 2.6 mm, Stahl 1.4301

33.7 * 3.2 mm, Stahl 1.4301

42.4 * 3.2 mm, Stahl 1.4301

48.3 * 3.2 mm, Stahl 1.4301

60.3 * 3.6 mm, Stahl 1.4301

76.1 * 3.6 mm, Stahl 1.4301

ThreadG x

21.3 mm

26.9 mm

33.7 mm

42.4 mm

48.3 mm

60.3 mm

76.1 mm

350 mm

430 mm

530 mm

660 mm

750 mm

930 mm

1170 mm G 2 1/2 "

G 2"

G 1 1/2"

G 1 1/4"

G 1"

G 3/4"

G 1/2"

Total length

500 mm

600 mm

750 mm

900 mm

1000 mm

1250 mm

1500 mm

Strömungsrichtung / Direction of air flow

On request CS Instrument will supply the following measuring sections for VA 400 consumption sensors:

G 1/2"-outer screw neck

Length: L1

LADA

Measurement section for VA 300 probes with flange connection

Outer dia.DA Length L1 Pipe

88.9 * 2.0 mm, Stahl 1.4301

114.3 * 2.0 mm, Stahl 1.4301

139.7 * 3.0 mm, Stahl 1.4301

168.3 * 3.0 mm, Stahl 1.4301

Total length= L1 + 2 * DIN flange

Weld neck flange acc. to DIN 2633 Weld neck flange acc. to DIN 2633

DIN - flange

= DA

88.9 mm

114.3 mm

139.7 mm

168.3 mm

1750 mm

2000 mm

2750 mm

3000 mm

DN 80 / 88.9

DN 100 / 114.3

DN 125 / 139.7

DN 150 / 168.3

Total length = L1 + 2 * DIN-flange

1750 + (2*50) = 1850 mm

2000 + (2*52) = 2104 mm

2750 + (2*55) = 2860 mm

3000 + ( 2*55) = 3110 mm

LA

1330 mm

1700 mm

2050 mm

2450 mm

Strömungsrichtung / Direction of air flow

Measuring section for VA 400 flow sensors with flange connection:

Measuring section for VA 400 flow sensors:

Page 27: VA 400 consumption sensor - CS Instruments GmbH · V5-02-2010 INSTRUCTION MANUAL Stationary and mobile flow and consumption measurement for compressed air and gases VA 400 consumption

27 V5-02-2010

DECLARATION OF CONFORMITY

Page 28: VA 400 consumption sensor - CS Instruments GmbH · V5-02-2010 INSTRUCTION MANUAL Stationary and mobile flow and consumption measurement for compressed air and gases VA 400 consumption

28 V5-02-2010

DECLARATION OF CONFORMITY

Page 29: VA 400 consumption sensor - CS Instruments GmbH · V5-02-2010 INSTRUCTION MANUAL Stationary and mobile flow and consumption measurement for compressed air and gases VA 400 consumption

29 V5-02-2010

NOTES

Page 30: VA 400 consumption sensor - CS Instruments GmbH · V5-02-2010 INSTRUCTION MANUAL Stationary and mobile flow and consumption measurement for compressed air and gases VA 400 consumption

30 V5-02-2010

NOTES

Page 31: VA 400 consumption sensor - CS Instruments GmbH · V5-02-2010 INSTRUCTION MANUAL Stationary and mobile flow and consumption measurement for compressed air and gases VA 400 consumption

31 V5-02-2010

NOTES

Page 32: VA 400 consumption sensor - CS Instruments GmbH · V5-02-2010 INSTRUCTION MANUAL Stationary and mobile flow and consumption measurement for compressed air and gases VA 400 consumption

32 V5-02-2010

CONTACT

Sales office SOUTH

Zindelsteiner Str. 15

78052 Villingen-Schwenningen

Phone +49 (0) 7705 97 89 9-0 Fax +49 (0) 7705 97 89 9-20

[email protected] www.cs-instruments.com

Sales office NORTH

Am Oxer 28c 24955 Harrislee

Phone +49 (0) 461 700 20 25

Fax +49 (0) 461 700 20 26

[email protected] www.cs-instruments.com