utilizing an accurate mass and retention time library to ......gradient* in ms-only mode (scan rate...

24
Utilizing an Accurate Mass and Retention Time Library to Facilitate Biomarker Discovery Nichole Reisdorph, PhD and Cole Michel, BS Mass Spectrometry Facility Skaggs School of Pharmacy and Pharmaceutical Sciences University of Colorado Anschutz Medical Campus 6/15/2017 Conflict of Interest Statement The SSPPS mass spectrometry facility is a Center of Excellence for Agilent Technologies Dr. Richard Reisdorph consults for Agilent Technologies Cole Michel Richard Reisdorph, PhD

Upload: others

Post on 20-Oct-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

  • Utilizing an Accurate Mass and Retention Time Library to Facilitate

    Biomarker Discovery

    Nichole Reisdorph, PhD and Cole Michel, BSMass Spectrometry Facility

    Skaggs School of Pharmacy and Pharmaceutical SciencesUniversity of Colorado Anschutz Medical Campus

    6/15/2017

    Conflict of Interest Statement

    • The SSPPS mass spectrometry facility is a Center of Excellence for Agilent Technologies

    • Dr. Richard Reisdorph consults for Agilent Technologies

    Cole Michel Richard Reisdorph, PhD

  • Research Program

    Core Facility

    Informatics

    Training

    Respiratory disease

    International Training Program

    Data analysis and tool optimization

    Proteomics, lipidomics, metabolomics, targeted analyses (lipid mediators, steroids, amino acids, etc)

    Overview

    • Background: Quantitative proteomics • Addressing challenges with choice of methodology

    – Nano vs Standard flow• Comparison using commercially available cell extract

    • Application to clinical research project– Analyzing Cerebrospinal Fluid (CSF) to understand Multiple

    Sclerosis (MS) phenotypes– Accurate mass retention time library vs spectral counting– (MS-only quantitation vs MS/MS quantitation)

  • Background- Quantitative proteomics

    • Objective was to develop robust and reproducible method for quantitation of complex protein samples

    • Cells > tissues > biofluids• Pros/Cons of labeling vs non-labeling methods• Pros/Cons of MS vs MS/MS

    – Accurate Mass Retention Time Library– Spectral Counting

    • Quadrupole Time of Flight

    What strategy do we choose?

    • Use of biofluids in research limits options – sample limited, challenging to multiplex

    • Compare standard flow versus nanoflow using the same sample and assess:– Reproducibility– Coverage (# of peptides and proteins identified)

    • Compare MS-based quantitation to MS/MS quantitation– Not conducted as part of standard vs nano flow experiments

  • Nanoflow LC/MS• Improved analytical sensitivity,

    especially when sample limited• Longer column re-equilibration

    times • Requires cleaner samples-

    clogging can be an issue• Column capacity may be reached

    when loading complex samples• Requires more skill to maintain

    and use• HPLC Chip vs Nanoflow columns

    Standard Flow LC/MS• Equivalent or better detection

    limit when sample is not limited• Faster analyses with quicker

    column re-equilibration• More forgiving of matrix• Higher loading and peak capacity• Easiest to use and maintain• Greater commercial column

    selection available • Best run-to-run reproducibility

    Nano vs Standard Flow: Pros and Cons

    ++

    ++

    ++ +

    +++ ++ +

    ++

    +++

    ++

    + ++

    Sample Information• Pierce Hela protein digest standard (P/N: 88328)• Tryptic digest• Re-suspended @ 1 ug/uL with 3% ACN in 0.1% formic acid• Triplicate injections in each configuration

    Column InformationNanoflow LC/MS Standard Flow AJS* LC/MS

    • Analytical Column: Thermo 75 µm x 250 mm Acclaim PepMap RSLC Column (P/N: 164941)

    • Trapping Column: Thermo 75 µm x 20 mm Acclaim PepMap 100 Trap Column (P/N: 164535)

    • Analytical Column: Agilent 2.1 x 25 cm Acclaim PepMap RSLC Column (P/N: 164941)

    • Trapping Column: None

    *Agilent Jet Stream on 6550 QTOF

    Note: AJS parameters were provided by Dr. Vadi Bhat from Agilent.Mr. Cole Michel optimized the nanoflow methods.

  • Method Information (HPLC)Nanoflow LC/MS Standard Flow AJS LC/MS

    • Gradient (~115 min)• Mobile Phase A = 0.1% Formic Acid in H2O• Mobile Phase B = 0.1% formic acid in 90% ACN

    • TCC: 65⁰C• Effective Flowrate: 0.2 mL/min• Loading Flowrate: N/A• Sample on Column: 15 µg (15µL)• Date Sample Run: January 17th 2017

    • Gradient (~112 min)• Mobile Phase A = 0.1% Formic Acid in H2O• Mobile Phase B = 0.1% formic acid in 90% ACN

    • TCC: 42⁰C• Effective Flowrate: 335 nL/min• Loading Flowrate: 3.2 µL/min• Sample on Column: 1.2 µg (1.2µL)• Date Sample Run: January 18th 2017

    Method Information (MS/MS)Gas Temp: 175⁰CDrying Gas: 11 L/minVCap: 1300 VFragmentor: 360 VOct 1 RF Vpp: 750 VMode: Auto MS/MSMS Mass Range: 290-1700 m/zMS Acquisition Rate: 10MS/MS Mass Range: 50-1700 m/zMS/MS Acquisition Rate: 3Isolation Width: narrowAdvanced Decision Engine: True (B.08 software)Collision Energy: Use formula

    Nanoflow LC/MS Standard Flow AJS LC/MS

    Charge Slope Offset

    2 3.1 1

    3 3.6 ‐4.8

    >3 3.6 ‐4.8

    Gas Temp | Sheath Gas Temp: 250⁰C | 250⁰C

    Drying Gas | Sheath Gas: 14 L/min | 11 L/min

    Vcap | Nozzle: 3500 V | 0 V

    Fragmentor: 360 V

    Oct 1 RF Vpp: 750 V

    Mode: Auto MS/MS

    MS Mass Range: 290-1700 m/z

    MS Acquisition Rate: 10

    MS/MS Mass Range: 50-1700 m/z

    MS/MS Acquisition Rate: 3

    Isolation Width: narrow

    Advanced Decision Engine: True (B.08 software)

    Collision Energy: Use formula

    Nebulizer: 35 psigCharge Slope Offset

    2 3.1 1

    3 3.6 ‐4.8

    >3 3.6 ‐4.8

  • Method Information (MS/MS)Nanoflow LC/MS Standard Flow AJS LC/MS

    Max Precursors Per Cycle: 20Abs. Threshold: 3000 countsRel. Threshold (%): 0.01%Active Exclusion: True

    Exclude after: 1 spectraReleased After: 0.4 min

    Isotope Model: PeptidesCharge State Preferences: Purity Stringency: 100%Purity Cutoff: 30%ADA Target: true, 40,000 counts/spectrumUse MS/MS accumulation time limit: trueReject Precursors that cannot reach target TIC: false

    Inactive Active1                  

    unk2                       3                     >3

    Sort by Abundance only

    Max Precursors Per Cycle: 20Abs. Threshold: 3000 countsRel. Threshold (%): 0.01%Active Exclusion: True

    Exclude after: 1 spectraReleased After: 0.15 min

    Isotope Model: PeptidesCharge State Preferences: Purity Stringency: 100%Purity Cutoff: 30%ADA Target: true, 25,000 counts/spectrumUse MS/MS accumulation time limit: trueReject Precursors that cannot reach target TIC: true

    Inactive Active1                  

    unk2                       3                     >3

    Sort by Abundance only

    Tuning: High analytical sensitivity slicer, low mass range 50-1700, extended 2Ghz922 m/z: Quad narrow isolation efficiency 65%, abundance 10.9 M, Res 22K

    Note: AJS parameters were provided by Dr. Vadi Bhat from Agilent.Mr. Cole Michel optimized the nanoflow methods.

    Spectrum Mill Parameters (General Workflow for Iterative Searching)

    Extraction- Default settings

    MS/MS Search #1* Auto-Validation #1 Auto-Validation #2

    MS/MS Search #2 Auto-Validation #1 Auto-Validation #2

    MS/MS Search #3 Auto-Validation #1 Auto-Validation #2

    MS/MS Search #4 Auto-Validation #1 Auto-Validation #2

    Protein Summary Results (X% Validation Yield of Filtered Spectra)Software compiles results

    *See next slide for settings

  • Spectrum Mill Parameters (MS/MS Search #1)

    Same parameters used for AJS and Nano data files

    Spectrum Mill Parameters

    Unchanged• SwissProt > Humans• Carbamidomethylation• 2 missed cleavages

    Search 1 Search 2 Search 3 Search 4

    Enzyme Trypsin Trypsin Trypsin, non-specific N term

    Trypsin, non-specific C term

    Lysine Carbamylated Acetyl K - -

    Ox Meth X X - -

    Pyroglutamicacid

    X X - -

    Deamidated X X - -

    Phos S, T, Y X X - -

    Search Mode Variable Mod Variable Mod Identity Identity

  • Spectrum Mill Parameters (Auto-Validation #1)

    Spectrum Mill Parameters (Auto-Validation #2)

    Nanoflow vs. Standard flow (AJS)Results from HeLA cell digest

  • SpectrumMill Summary

    AJS – 15 ug on column

    Nano – 1.2 ug on column

    Results could be different due to MS/MS parameters, eg multiple peptide hits.

    Peptide Comparison

    12492 12181 12573

    1725817449 17852 17341

    23808

    0

    5000

    10000

    15000

    20000

    25000

    Inj 1 Inj 2 inj 3 Total

    Dis

    tinc

    t Pep

    tide

    s

    AJS vs. Nano Source Distinct Peptides

    AJS Nano

  • Protein Comparison

    2217 2176 2207

    27993170 3217 3029

    4115

    0

    500

    1000

    1500

    2000

    2500

    3000

    3500

    4000

    4500

    Inj 1 Inj 2 inj 3 Total

    Prot

    eins

    Ide

    ntifi

    ed

    AJS vs. Nano Source - Proteins

    AJS Nano

    Total Ion Chromatograms of Triplicate Injections

    7x10

    0

    0.1

    0.2

    0.3

    0.4

    0.5

    0.6

    0.7

    0.8

    0.9

    1

    1.1

    1.2

    1.3

    1.4

    1.5

    1.6

    1.7

    1.8

    1.9

    2

    2.1

    2.2

    2.3

    2.4+ESI TIC MS(all) Hela_Jet_Stream_20ug_3.d

    1 1

    Counts vs. Acquisition Time (min)5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100 105 110 115 120 125 130 135

    7x10

    0

    0.2

    0.4

    0.6

    0.8

    1

    1.2

    1.4

    1.6

    1.8

    2

    2.2

    2.4

    2.6

    2.8

    3

    3.2

    3.4

    3.6

    3.8

    4+ESI TIC MS(all) HELA_1ug_condition_4.d

    1 1

    Counts vs. Acquisition Time (min)5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100 105 110 115 120 125 130 135

    AJS

    Nano

  • Total Ion Chromatograms Overlaid (same scale)

    7x10

    0

    0.2

    0.4

    0.6

    0.8

    1

    1.2

    1.4

    1.6

    1.8

    2

    2.2

    2.4

    2.6

    2.8

    3

    3.2

    3.4

    3.6

    3.8

    4

    4.2

    +ESI TIC MS(all) Hela_Jet_Stream_20ug_1.d

    1 1

    Counts vs. Acquisition Time (min)5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100 105 110 115 120 125 130 135

    Nano TIC | AJS TIC

    Nano

    AJS

    Base Peak Chromatograms (Scale Locked)

    6x10

    0

    0.2

    0.4

    0.6

    0.8

    1

    1.2

    1.4

    1.6

    1.8

    2

    2.2

    2.4

    2.6+ESI BPC Scan Frag=360.0V Hela_Jet_Stream_20ug_1.d

    1 1

    6x10

    0

    0.2

    0.4

    0.6

    0.8

    1

    1.2

    1.4

    1.6

    1.8

    2

    2.2

    2.4

    2.6+ESI BPC Scan Frag=360.0V HELA_1ug_condition_3.d

    1 1

    Counts vs. Acquisition Time (min)5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100 105 110 115 120 125 130 135

    AJS BPC | Nano BPC

    Nano

    AJS

  • Conclusions

    • Triplicate injections of HeLA digest showed good overlay of TIC for both Nano and AJS

    • Nanoflow has greater than 20x the sensitivity of AJS– AJS@15 ug: 17,258 peptides and 2799 proteins– Nano@ 1.2 ug: 23,808 peptides and 4115 proteins– 1.37 fold more peptides and 1.47 fold more proteins

    • While AJS settings could be re-optimized, Nanoflow clearly provides good proteome coverage

    • Next step is to evaluate reproducibility over time

    • Note: 25 ug is considered optimal for the AJS system; our goal was to evaluate a 10x difference in loading.

    31593388 3481

    4234

    0

    500

    1000

    1500

    2000

    2500

    3000

    3500

    4000

    4500

    Injection 1 Injection 2 Injection 3 Total

    Prot

    eins

    Ide

    ntifi

    ed

    Nano Source 1.5 ug Injections - Proteins

    Nano

    1845319157 19002

    24412

    0

    5000

    10000

    15000

    20000

    25000

    30000

    Injection 1 Injection 2 Injection 3 Total

    Dis

    tinc

    t Pep

    tide

    s

    Nano Source 1.5 ug Injections - Distinct Peptides

    Nano

    Triplicate 1.5 ug Injection Hela Standard

    Average 3342Std Dev 1665% CV

    Average 18870Std Dev 3701% CV

  • 8x10

    0

    0.1

    0.2

    0.3

    0.4

    0.5

    0.6

    0.7

    0.8

    0.9

    1

    1.1

    1.2

    1.3

    1.4

    1.5

    +ESI TIC Scan Frag=360.0V Hela_05ug_1-r019.d

    1 1

    Counts vs. Acquisition Time (min)5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100 105 110 115 120 125 130 135

    Reproducibility studies

    • Overlaid TIC of 18 replicate 500 ng Injections with a 120 min gradient* in MS-only mode (Scan rate 1.5 spec/sec)

    Pierce Hela protein digest standard (P/N: 88328) re-suspended @ 0.5ug/uL with 3%ACN in 0.1%FA Analytical Column: Thermo 75 µm x 250 mm Acclaim PepMap RSLC Column (P/N: 164941)Trapping Column: Thermo 75 µm x 20 mm Accalim PepMap 100 Trap Column (P/N: 164535)

    *68 hour or ~3 days

    Overlaid Base Peak Chromatograms7x10

    0

    0.1

    0.2

    0.3

    0.4

    0.5

    0.6

    0.7

    0.8

    0.9

    1

    1.1

    1.2

    1.3

    1.4

    1.5

    1.6

    +ESI BPC Scan Frag=360.0V Hela_05ug_1-r019.d

    Counts vs. Acquisition Time (min)20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100 105 110 115 120 125 130

    7x10

    0

    0.1

    0.2

    0.3

    0.4

    0.5

    0.6

    0.7

    0.8

    0.9

    1

    1.1

    1.2

    1.3

    1.4

    1.5

    1.6

    +ESI BPC Scan Frag=360.0V Hela_05ug_1-r019.d

    Counts vs. Acquisition Time (min)55 55.5 56 56.5 57 57.5 58 58.5 59 59.5 60 60.5 61 61.5 62 62.5 63 63.5 64 64.5 65 65.5 66 66.5 67 67.5 68 68.5 69 69.5 70 70.5 71 71.5 72 72.5 73 73.5 74 74.5 75

    RT drift was

  • CV Analysis

    CV cut‐off  Compounds Total compounds Percentage5% 4944 15325 32.3%10% 11789 15325 76.9%15% 13040 15325 85.1%20% 13590 15325 88.7%30% 14297 15325 93.3%40% 14616 15325 95.4%50% 14813 15325 96.7%

    Filtered on Raw Data

    CV cut‐off  Compounds Total compounds Percentage5% 14484 15325 94.5%10% 14704 15325 95.9%15% 14717 15325 96.0%20% 14718 15325 96.0%30% 14942 15325 97.5%40% 15029 15325 98.1%50% 15077 15325 98.4%

    Filtered on Normalized data 

    • Data extracted using Profinder (v 8.0)• CV analysis performed in MPP on

    features that were found in 100% of all samples (to remove artifacts due to extraction)

    • Since these are all the same sample, we expect to approach 100% of all compounds having CVs < 20%; this would indicate excellent analytical platform reproducibility

    Typical Peak Shape Characteristics for Most Peptides in a 120 min Run

    5x10

    0

    0.5

    1

    1.5

    2

    2.5

    3

    3.5

    4

    4.5

    5

    5.5

    6

    6.5

    7

    7.5

    8

    8.5

    9

    +ESI EIC(592.9749) Scan Frag=360.0V Hela_05ug_1-r019.d

    51.619

    Counts vs. Acquisition Time (min)50.6 50.8 51 51.2 51.4 51.6 51.8 52 52.2 52.4 52.6 52.8 53 53.2 53.4 53.6

    Height: ~7.7 x 105FWHM: 0.30 minPeak width at 5%: 0.66 min (51.83 – 51.17 min)Peak Mid-Point: 51.50 minTailing Factor: 1.0

    AAYLQETGKPLDETLK; 1776.938 MH+; Annexin A1

  • Summary- Nanoflow method development

    • Objective was to develop robust and reproducible method for quantitation of complex protein samples

    • Good proteome coverage compared to AJS• 18 replicate injections showed excellent reproducibility

    – Close to 90% of peptides had CVs < 20% (96% when using normalized data)

    • System is robust- can switch from AJS to nano in under an hour with no loss of signal.

    Application: CSF proteomics

    • Objective was to develop robust and reproducible method for quantitation of complex protein samples

    • Cerebrospinal fluid (CSF) and plasma are complex mixtures• High abundant proteins make up >95% of CSF proteome• Clinical research studies can result in hundreds of samples

    over time• Can be some differences in processing and storing• AMRT library approach vs MS/MS quantitation

  • CSF/Plasma

    MS/MS AMRT library

    Immunodepletion(MARS Hu 14)

    Protein Digestion

    High pH RPLC

    Low abundance proteins

    Protein Digestion

    Low abundanceproteins

    MS QuantMS/MS Quant

    MS/MS Quant

    MS only Quant

    Strategy

    Sample Preparation Workflow > AMRT library development

    Pooled CSF from control and MS patientsConcentrated using 3kDa molecular weight cut off filter > Immunodepletion using MARS Hu-14 column Tryptic digest using FASP methodHigh pH reverse phase liquid chromatography (hp-RPLC) > 45 fractions were collected Equal volumes of non-adjacent fractions were concatenated for a final total of 15 fractions.Spike in retention time standards > Analyze using MS/MS to build library

    LTQ MS + immunodepletion: 171 proteins in 2007 http://pubs.acs.org/doi/pdfplus/10.1021/pr070293w

    LTQ + immunodepletion: 2800-3500 proteins in various sampleshttps://www.ncbi.nlm.nih.gov/pubmed/24039694 (2013)

  • AMRT Library LC/MS/MS Method

    • The 15 concatenated high pH-RPLC fractions were analyzed in MS/MS mode on the Agilent 6550 QTOF with the Nano-adapter and nanosource.

    • Column = 0.075 x 250 mm Acclaim PepMap RSLC• Trap = 0.075 x 20 mm PepMap 100, nanoviper• Mobile Phase A = 0.1% Formic Acid in H2O• Mobile Phase B = 0.1% formic acid in 90% aq. ACN• Primary Flow = 0.120 mL/min; Effective Flow = 330

    nL/min; Loading Flow = 3.2 µL/min• Column Temperature = 42⁰C• Injection Volume = 2.0 uL or 480 ng on column• 1290 Infinity II LC system from Agilent: Binary Pump,

    Multisampler, MCT, Capillary Pump and Nanoadapter

    AMRT Library Database Searching and Exclusion List Method

    • Exclusion lists were utilized to achieve deeper coverage of the CSF Proteome.

    • Each of the 15 fractions was injected 3 times. – The first run was searched in SpectrumMill and summarized using

    the peptide summary with a cut-off score of 9 and SPI of 50%. – This summary was used to generate an exclusion list, this was

    imported into QTOF software and peptides on the list did not have MS/MS performed for subsequent runs

    – This was repeated prior to the 3rd injection• AMRT library built using in-house scripts (subsequently used

    MPP)

  • Sample Preparation and LCMS Workflow > Sample analysis

    Individual CSF samples from control and MS patientsConcentrated using 3kDa molecular weight cut off filter > Immunodepletion using MARS Hu-14 column Tryptic digest using FASP methodSpike in retention time standardsAnalyze using LC/MS and identify peptides/proteins using AMRT library

    MS-Only Quant Method

    Digested CSF samples were loaded onto a 2cm PepMAP 100, nanoviper trapping column and chromatographically resolved on-line using a 0.075 x 250 mm, 2.0µ Acclaim PepMap reverse phase nano column (Thermo Scientific) using a 1290 Infinity II LC system equipped with a nanoadapter(Agilent).

    • Column = 0.075 x 250 mm Acclaim PepMap RSLC • Trap = 0.075 x 20 mm PepMap 100, nanoviper• Mobile Phase A = 0.1% Formic Acid in H2O• Mobile Phase B = 0.1% formic acid in 90% aq. ACN• Primary Flow = 0.120 mL/min; Effective Flow = 330

    nL/min; Loading Flow = 3.2 µL/min• Column Temperature = 42⁰C• Injection Volume = 3.2uL or 480 ng on column• 1290 Infinity II LC system from Agilent: Binary Pump,

    Multisampler, MCT, Capillary Pump and Nanoadapter

  • Import into MPPImport into MPP

    • Mass Profiler Professional

    • Normalize

    FilterFilter

    • Present in 60-90%

    • CV RT• CV MH+• CV

    Intensity

    Compare intensityCompare intensity

    • PCA• ANOVA• Flag

    differentially abundant ions

    Export mass list Export mass list

    • Search MS/MS library to identify peptides and proteins of interest

    1) Extract peptides MH+, RT, Peak Height (and AUC), z for each molecule

    2) Data Analysis

    MS-only quant data analysis workflow

    MS Quant Data Extraction

    • Data Extracted with Profinder V.8.0– Recursive workflow (MFE MPP FBIon)– Parameters

    • MFE = Input Range 10-72.5 min; Height: > 8000 counts; + ions: H+, Na+, K+; charge state: 2-6; two or more ions required; RT window 5% + 0.8 min; mass window 20 ppm + 0.2mDa; Absolute Height >13,000; MFE score > 80

    • FBIon = + ions: H+ Na+ K+; charge state 2-6; mass and RT required for match; mass tolerance 20 ppm; RT tolerance 0.6 min; Height 10000 counts; Agile 2 Algorithm used for integration

    • Converted CEF Files area to height w/perl script

  • MS Quant Data Analysis

    • Data analyzed using Mass Profiler Professional V.14.8– Used AMRT Library and ID Browser to annotate peptides

    • ANOVA w/o multiple testing correcting (MTC) was performed on all groups to filter results to those most relevant = ANOVA list– P < 0.05 and FC > 2.0 were used as thresholds– Only proteins with > 2 peptides were included

    • Using ANOVA list, a series of moderated T-test or ANOVA were performed. – If more than 30% of peptides were significant with similar trends, a

    protein was considered a valid hit.– Fold Changes were calculated by transforming the normalized transformed

    data back to raw data and summing all peptides that were identified and linked to the protein regardless of the significance of the peptide.

    RESULTS

  • Quality Control

    • QC is used to evaluate sensitivity, variability in retention time, and percent CV (%CV)

    • Pooled QC of all CSF samples* for MS and MS/MS quant– A total of 7 QC samples were injected over 4 days of analysis.

    • Waters MassPrep Peptide Retention Time Mix (RT Mix) was spiked in to control for retention time shifts. – For library building, MS quant, and MS/MS quant– Maximum shift in retention time at any given point is 0.4 – 0.5

    minutes.

    * Experimental samples, not library

    QC of RT Mix: Library Building

    Angiotensin Frag. 1-7 [300.4960 m/z

    (Z=3)]

    Bradykinin [354.1944 m/z

    (Z=3)]

    Angiotensin II [349.5188 m/z

    (Z=3)]

    Angiotensin I [432.9022 m/z

    (Z=3)]

    Renin Substrate [440.4886 m/z

    (Z=4)]

    Enolase T35 [624.9956 m/z

    (Z=3)]

    • 42 injections of 15 concatenated fractions

    • 6 MassPrep Peptides were identified in each fraction

    • Some variability due to different amounts being spiked (to assess fold-change differences) and possible degradation of peptides

  • QC: Retention time reproducibility over 2 weeks

    Black = MS-only Quant: 21 Sample injections [scan rate 1.5]run 02/09/17 Red = Library Building: 42 injections of concatenated fractions [scan rate 10] run 01/25/2017

    Angiotensin I [432.9022 m/z

    (Z=3)]

    Bradykinin [354.1944 m/z

    (Z=3)]

    Angiotensin Frag. 1-7 [300.4960 m/z

    (Z=3)]

    Renin Substrate [440.4886 m/z

    (Z=4)]

    Enolase T35 [624.9956 m/z

    (Z=3)]

    Angiotensin II [349.5188 m/z

    (Z=3)]

    Results: Spectrum Mill Summary of CSF MS/MS Library

    Total Distinct Peptides

    Total protein groups with > 2

    Distinct Peps

    Total possible proteins with > 2 Distinct Peps

    Total protein groups

    Total possible proteins

    Total Proteins Spectrum Mill Cut-offs

    50044 4651 5005 7614 7968 9600 fdr < 1%, Peptide score > 0 & SPI > 0%, Protein Score > 0

    39240 2964 3199 3941 4176 5239 fdr < 1%, Peptide score > 8 & SPI > 50%, Protein Score > 10

    36531 2506 2715 3632 3841 4931 fdr < 1%, Peptide score > 9 & SPI > 50%, Protein Score > 10

    32875 2006 2189 3425 3609 4703 fdr < 1%, Peptide score > 10 & SPI > 50%, Protein Score > 10

    38405 2964 3199 3106 3341 3903 fdr < 1%, Peptide score > 8 & SPI > 50%, Protein Score > 13

    35572 2506 2715 2673 2882 3406 fdr < 1%, Peptide score > 9 & SPI > 50%, Protein Score > 13

    31680 2006 2189 2231 2415 2832 fdr < 1%, Peptide score > 10 & SPI > 50%, Protein Score > 13

    28794 1833 1995 2086 2248 2638 fdr < 1%, Peptide score > 10 & SPI > 70%, Protein Score > 13

  • Summary

    • Overall strategy proved effective for quantitating proteins from complex biofluid (CSF)

    • Immunodepletion > high pH RPLC > MS/MS on Nano 6550 system resulted in almost 40,000 peptides and over 5,000 proteins in our AMRT library

    • Nano system is robust and reproducible• Initial pilot study showed good coverage compared to

    previous studies (~3500 vs 5000 proteins)• Next step is to increase sample sizes in all categories

    Analytical Position

    Binary Pump

    NanoSplitter (1:400)

    To wasteTrap Column

    Analytical Column

    Auto sampler

    Cap Pump

    Detector

    Loading Position

    Binary Pump

    NanoSplitter (1:400)

    To wasteTrap Column

    Analytical Column

    Auto sampler

    Cap Pump

    Detector Switching Valve in TCC at 42oC

    Switching Valve in TCC at

    42oC

    Nano Configuration with Binary and Cap Pumps

  • http://www.ucdenver.edu/academics/colleges/pharmacy/Research/CoreFacilities/Spectrometry/Pages/default.aspx