USP DISSOLUTION STUDIES

Page 2: Usp dissolution studies

DISSOLUTION

A process in which a solid substance solubilizes in a

given solvent. i.e. mass transfer from the solid

surface to the liquid phase.

USP DISSOLUTION APPARATUS

• Apparatus 1 - Basket (37º)• Apparatus 2 - Paddle (37º)• Apparatus 3 - Reciprocating Cylinder (37º)• Apparatus 4 – Flow-Through Cell (37º)

Page 3: Usp dissolution studies

• Apparatus 5 – Paddle over Disk (32º), Transdermal Delivery

System, use paddle and vessel from Apparatus 2 with a

stainless steel disk assembly to hold the transdermal on the

bottom of vessel.

• Apparatus 6, Cylinder (32º), Transdermal Delivery System,

use Apparatus 1 except replace the basket shaft with a

stainless

steel cylinder element.

• Apparatus 7, Reciprocating Holder, for transdermal delivery

systems and also a variety of dosage forms

Page 4: Usp dissolution studies

APPARATUS 1:The assembly consists of the following :A covered vessel, Transparent material, A motor,A metallic drive shaft, A cylindrical basket,The water bath permits the holding of temp inside vessel

at 37±0.5 7C

The vessel is a cylindrical with hemispherical bottom with dimension and capacity of.

Liter Height Inside diameter

1 160-210 mm 98-106 mm

2 280 – 300 mm 98-106 mm

4 280 – 300 mm 145-155 mm

Page 5: Usp dissolution studies

The shaft is positioned so that its axis is not more than 2mm at any point from the vertical axis of the vessel and rotates smoothly and with out significant wobble.

Use 40mm mesh cloth. A basket having a gold coating 0.0001 inch thick may be used.

The dosage unit is place in a dry basket at the beginning of each test.The distance between the inside bottom of the vessel and the basket is

maintained at 25± 2mm.

APPARATUS 2• The assembly from apparatus 1. Except paddle formed from a blade. a shaft is used as the stirring element The vertical center line of the blade is flush with the bottom of the

shaft. The distance of 25±2 mm between the blade and the inside bottom of

the vessel.

Page 6: Usp dissolution studies

•The paddle, blade and shaft

may be coated with a

suitably inert coating.

•The dosage unit is allowed

to sink to the bottom of the

vessel before rotation of the

blade.

•A small, loose piece of

nonreactive material such as

wire helix may be attached

to dosage units in order to

prevent floating.

Page 7: Usp dissolution studies

DISSOLUTION MEDIUM

• If the dissolution medium is buffered solution adjust its pH within 0.05 unit of pH specified.

TIME: • If single time specification is given, the test may be

concluded in a shorter period .• If two or more times are specified, specimen are to

be withdrawn only at the stated times within a tolerance of ±2 %

PROCEDURES• Place the stated volume of the dissolution

medium in the vessel.• Equilibrate the dissolution medium to 37±o.5 7c

• Place 1 tablet or 1 capsule in the apparatus.

Page 8: Usp dissolution studies

• Operate the apparatus

• Within the time interval, withdraw a specimen from

zone midway between the surface of the dissolution

medium and the top of the rotating basket or blade ,

not less than 1cm from vessel wall.

• Replace the aliquots withdrawn for analysis with

equal volume of dissolution medium at 37 7c.

Page 9: Usp dissolution studies

INTERPRETATION:

Stage Number Tested

Acceptance Criteria

S1 6 each unit is not less than Q+5%

S2 6 average of 12 units (s₁+s₂) is

equal to or greater than Q, and

no unit is less than Q-15%.

S3 12 average of 12 units (s₁+s₂+s₃) is

equal to or greater than Q, and

no unit is more than 2 unit are

less than Q-15%, and no unit is

less than Q-25%.

Page 10: Usp dissolution studies

APPARATUS 3The assembly consists of:

a set of cylindrical,flat-bottomed glass vessels;a set of glass reciprocatingcylindrical;stainless steel fitting; screen;a motor and drive.

• The vessel is immersed in water

bath holding temp at 37± 0.5 7

Page 11: Usp dissolution studies

PROCEDURE:

• Place the stated volume of the dissolution medium in each vessel .

• Equilibrate the dissolution medium to 37 + 0.5° • Place 1 dosage-form unit in each of the six

reciprocating cylinders.• Operate the apparatus• During the upward and downward stroke, the

reciprocating cylinder moves through a total distance of 9.9 to 10.1 cm

• Within the time interval specified withdraw a portion of the solution under test from a zone midway between the surface of the dissolution medium and the bottom of each vessel.

• Perform The analysis as directed in the individual monograph.

Page 12: Usp dissolution studies

APPARATUS 4

• The assembly consists of a reservior and a pump for the dissolution medium; a flow-through cell; a water bath.

• The pump forces the dissolution medium upwards through the flow-through cell.

• The pump has a delivery range between 240 and 960 mL per hour, with standard flow rates of 4, 8, and 16 mL per minutes.

• The cell is immersed in a water bath and the temperature is maintained at 37 + 0.5°

• The apparatus uses a clamp mechanism and two O-rings for the fixation of the cell assembly.

Page 13: Usp dissolution studies

PROCEDURE

• Place the glass beads into the cell specified in the monograph.

• Place 1 dosage form unit on top of the beads.

• Assemble the filter head and fix the parts together by means of a suitable clamping device.

• Introduce by the pump the dissolution medium warmed to 37 + 0.5° through the bottom of the cell.

• Collect the eluate by fractions at each of the times stated.

• Perform the analysis as directed in the individual monograph.

Page 14: Usp dissolution studies

APPARATUS 5

• The paddle and vessel assembly from apparatus 2 with the addition of stainless steel disk assembly.

• The temperature is maintained at 32 + 0.5°C

• A distance of 25 + 2 mm between the paddle blade and the surface the disk assembly is maintained during the test.

• The disk assembly is designed to minimize any dead volume between the disk assembly and the bottom of the vessel.

Page 15: Usp dissolution studies

PROCEDURE

• Place the stated volume of dissolution medium in the vessel assemble the apparatus without the disk assembly and equilibrate the medium to 32 + 0.5°C

• Apply the transdermal system disk assembly.• The system may be attached to the disk by a

suitable adhesive.• Place the disk assembly flat at the bottom of the

vessel with the release surface facing up and parallel to of the paddle blade and surface of the dissolution medium.

• Operate the apparatus • At each sampling time interval, withdraw a

sample from a zone midway between the surface of the dissolution medium and the top of the blade, not less than 1 cm from the vessel wall.

• Perform the analysis on each sampled aliquot as directed in the monograph.

Page 16: Usp dissolution studies

APPARATUS 6• The vessel assembly from apparatus 1 except to

replace the basket and shaft with a stainless steel cylinder stirring element and to maintain the temperature at 32 + 0.05°C

• The dosage unit is placed on the cylinder at the beginning of each test.

• The distance between the inside bottom of the vessel and the cylinder is maintained at 25 + 2 mm during the test.

PROCEDURES• Place the stated volume of the dissolution medium in

the vessel.• Equilibrate the dissolution medium to 32 + 0.5°C• Prepare the test system prior to test as follows.• Remove the protective liner from the system and place

the adhesive side on a piece of cuprophan.• Cuprophan covered side down, on a clean surface, and

apply a suitable adhesive to the exposed cuprophan borders.

Page 17: Usp dissolution studies

• Dry for 1 minutes.

• Press the cuprophan covering to remove trapped air

bubbles.

• Place the cylinder in the apparatus and immediately

rotate at the rate specified

• Within the time interval specified, withdraw a quantity

of dissolution medium for analysis from a zone midway

between the surface of the dissolution medium and

the top of the rotating cylinder, not less than 1 cm

from the vessel wall.

• Perform the analysis as directed in the individual

monograph.

Page 18: Usp dissolution studies

APPARATUS 7

• The assembly consists of a set of volumetrically calibrated or tared solution containers.

• A motor and drive assembly to reciprocate the system vertically

• A set of suitable sample holders.• The solution containers are partially immersed in a

suitable water bath, inside the temperature of the containers at 32 + 0.5°C

SAMPLE PREPARATION – A (Coated tablet drug delivery system)

• Attach each system to be tested to a suitable sample holder, (e.g gluing system edge with 2-cyano acrylate glue onto the end of plastic rod or by placing the system into a small nylon net bag at end of a plastic rod).

Page 19: Usp dissolution studies

SAMPLE PREPARATION-B (Transdermal drug delivery system)

• Press the system into a dry, unused piece of cuprophan.• Taking care to eliminate air bubbles between the

substrate release surface • Attach the system to a suitable sized sample holder with

a suitable O-ring.• Trim the excess substrate with a sharp blade.

SAMPLE PREPARATION-C (Other drug delivery system)

PROCEDURE• Suspend each sample holder from a vertical

reciprocating shaker such that each system is continuously immersed in measured volume of dissolution medium.

• Reciprocating a frequency of about 30 cycles per minutes with an amplitude of about 2 cm.

• Remove solution containers from the bath, cool to room temperature.

• Perform the analysis as directed in the monograph.